Protec FI-75 Operation Manual
Text of the manual
PROTEC GmbH & Co. KG
In den Dorfwiesen 14, D-71720 Oberstenfeld, Germany
Telephone: +49-7062-9255-0 e-mail: [email protected]
Machine No.:
Type:
Installationsdatum:
Stand:
03-2017/1.0
Subject to change
English
Film Processor
Operation Manual
ID: 5870-0-0022
CP - OPTI 2
Table of Contents
English
Introduction .......................................................................................................... 2
Intended Use......................................................................................................... 2
Technical Specifications ..................................................................................... 3
Safety Instructions............................................................................................... 4
Installation .......................................................................................................... 5
Initial Operation.................................................................................................. 7
Operation
Short Overview and Control Panel .............................................................. 9
Switching on the machine............................................................................ 11
Stand-by mode ............................................................................................ 11
Bath temperature......................................................................................... 11
Display “Infeeding film”................................................................................ 11
Manual pumping .......................................................................................... 11
Anti-crystallisation function / Time replenishment ....................................... 11/12
Automatic replenishment............................................................................. 12
Rollfilms and Paper Films............................................................................ 12
Care
Daily Care.................................................................................................... 13
Weekly Care ................................................................................................ 13
Thorough Cleaning...................................................................................... 14
Maintenance / Disposal ..................................................................................... 15
Problems and Solutions
Advice on Film Defects................................................................................ 18
Advice on Machine Errors ........................................................................... 19
Service Manual as appendix, see page 20
Additional informations for FI-75 mammo see page 3; 23; 33
Copyright
© 2017 by PROTEC®. All rights reserved. Any reproduction that violates the limita-
tions set forth by copyright law, needs to be authorised in writing by PROTEC®.
Information on Liability
This manual has been checked for correctness. The instructions and specifications
were correct at the time it was published. Future models may have modifications
without prior notice.
PROTEC® does not take responsibility for damage caused directly or indirectly by
error, omission or non-conformity of the manual.
Introduction
The FI-75 processor is a compact, automatic table-top processor. Due to the
precision roller transport system, both sheet and roller films can be processed. The
film materials are developed, fixed, rinsed and dried. The FI-75 has been fit-
ted with an automatic film-registration and a Stand-by mode. The developing solu-
tions are temperature-regulated, circulated and automatically replenished.
This Operation Manual contains the most important instructions for installation, op-
eration and servicing of this machine. Please read the provided information careful-
ly to ensure reliable and satisfactory operation of your FI-75.
Intended Use
The X-ray film processor FI-75 is intended exclusively for the purpose de-
scribe in the introduction above.
CP - OPTI
Technical Specifications
X-ray film processorss (MDD class I) are employed in “medical” applications (med-
ical products directive) and “non-medical” applications (electrical appliances and
EMC directives).
Intended use includes observing the operating instructions, the installation instruc-
tions and adherence to the safety notes.
Any application differing from intended use voids the guarantee by PROTEC®.
The owner of the machine will be liable for damages resulting from unintended use
or faulty application.
Intended use includes adherence to all statutory regulations concerning occupa-
tional safety and radiation protection applicable at the operating site.
Technical Specifications
Film transport:
Continuous roller transport system
Film formats:
In general: Sheet and roll films up to 35.8 cm (14.1 ’’)
width;
Roll films with leader from 70 mm (2.8 ’’) width;
Smallest film format 10x10 cm (4x4 ’’).
Mammography type 1171: For processing mammog-
raphy films.
-
-
Processing capactiy:
111 films 24x30 cm (10x12 “) per hours (standard
model, film fed in crosswise)
Process time:
Standard 105 s
Mammography 135 s (*Option 90 s /167 s)
Linear speed:
Standard 48 cm/min. (19 in/min)
Mammography 37 cm/min (14.5 in/min)
*Option 56 cm/min (22 in/min) / 30 cm/min (11.8 in/min)
Developer time:
Standard 29 s Mammography 37 s *Special 25 s /46 s
Tank capacities:
Developer, fixer and washing 5 litres each (1.3 gal)
Circulation system:
Developer and fixer are continiuously circulated by a circulation pump
Replenishment:
Automatic replenishment by film detection, in relation
to film length
Developer temperature:
Adjustable 28 - 37 °C (82.4 - 98.6 °F)
Fixer temperature:
Adjusted
to
developer
temperature
by
heat
exchanger.
Water connection:
Permissible water pressure 2 - 10 bar (29 - 145 psi),
permissible water temperature 5 - 30 °C (41 - 86 °F).
Water consumption:
1.9 litres per minute (0.5 gal/min) when processing.
Drain capacity:
7 litres per minute (1.85 gal/min).
Noise level:
Less than 58 dB(A).
Heat emission:
Stand-by: 0.1 kJ/s Processing: 1.4 kJ/s
Environmental condi-
tions:
1) Temperature 18 - 40 °C (51.6 - 104 °F), venti-
lated room, room temperature should be lower
than set bath temperature.
2) Relative humidity lower than 80% up to 31 °C
(80 °F), linear decreasing to 50% at 40°C (104°F)
3) Height above sea level less than 2000 m (6666 ft.)
4) Indoor use
CP - OPTI
4
Technical Specifications
English
* Depending on machine type and used gears processors have different speeds.
** Height incl. optional working table resp. base cabinet.
Safety Instructions
Pollution degree:
2
System protection:
IP 20
Electrical connections:
Electrical specification are indicated on model nameplate
Type 107x-1-0000: 230 V~ ±10 %, 8.8 A, 50 Hz.
Type 107x-2-0000: 230 V~ ±10 %, 8.8 A, 60 Hz
Type 107x-3-0000: 110 / 120 V~ ±10 %, 12 A, 60Hz
Machine tested for overvoltage category II
according to IEC 1010 (EN 61010, VDE 0411)
Power consumption:
Stand-by: 0.12 kWh;
Processing: 1.4 kWh
Weight (processor):
Empty 35 kg (77 lbs)
Filled 50 kg (110 lbs)
Dimensions (LxWxH):
77x59x42 (** 112) cm 30.3x23.2x16.5 (** 44.1)’’
Floor space required:
0.45 m2 (4.8 sqft)
To ensure the safe operation of this processor, installation and use should always
conform to the instructions contained in this manual.
The developer and fixer chemicals used in the processor should be handled
according to the manufacturers instructions. In general: Non-diluted chemicals
are caustic. For this reason, chemicals should be handled very carefully. Avoid
contact with skin, always wear protective clothing, gloves and glasses when han-
dling the chemicals - for example, when mixing and refilling. Also when taking the
racks out for cleaning or servicing. In case of chemicals getting into the eyes,
rinse eyes immediately with cold, running water for approximately 15 minutes,
and contact a doctor afterwards. Inhalation of chemicals can be dangerous to
your health and should be avoided. For this reason, always ensure that the room
in which the processor is installed is adequately ventilated.
Environmental regulations regarding the storage and disposal of waste chemi-
cals should be obtained from the local water authorities and complied with.
Before opening the processor switch off the unit and unplug it from the electrical
socket. Service and repairs must be performed by trained service technicians
only. Use only manufacturers replacement parts.
CP - OPTI
Installation
Installation
1. Requirements for installation
a.
Fresh water connection: Stop cock with 3/4" outer-thread diameter (washing
machine connection), Water pressure 2 - 10 bar (29 - 145 psi).
b.
Drainage connection: Plastic tube - inner diameter 50 mm (2”) or larger. A
ventilated syphon which serves as odour preventor should be included in the
planning. The drainage tubes should be installed with a fall of minimum 5 %.
Local Water Authorities regulations should be complied with.
c.
Electrical connection: Fused wall socket with earth connection according to
electrical data (see technical specifications, page 9). It is also required to
install an earth-leakage switch (with 25 A / 30 mA nominal error current).
2. Transport
Due to the weight and dimensions of the film processor FI-75 should always
be carried by two persons. To do so, hold the machine at the sides on the bottom
(see figure). While putting the machine down, watch the position of the levelling feet
to prevent damaging these.
3. Installation of processor
Unpack the Processor. Remove cover and transport securing brackets on the sides
of the roller racks. Remove roller racks - start with the dryer rack.
In 220-240 V version the FI-75 is delivered as a tabletop processor with a
two part floorplate. If the machine is upgraded with the optional stand or cabinet,
the narrow plastic part has to be removed.
In 110-120 V version the FI-75 is delivered as a tabletop processor with a
three part floorplate. If the machine is upgraded with the optional stand or cabinet,
both small PVC-parts have to be removed and main plate must be reconnected.
Table-top installation
In the event that the processor is to be installed on a work top or table, the four ad-
justable feet should be screwed on.
Installing on processor stand or base cabinet
In the event that the processor is to be installed on the stand or cabinet (optional
accessory), the processor will be mounted directly to it. Mount processor according
to manual included with stand or cabinet (the adjustable feet inside the accessory
bag are not required).
Finally the processor needs to be levelled:
Place spirit level across the sidewalls of processor and adjust the levelling feet ac-
cordingly. Replace the racks into the processor and close the latches.
Electrical connections should be carried out according to reg-
ulations by an electrician.
Attention!
Machine should not be installed on table-top without adjusta-
ble feet, as this would block the ventilation openings under the
machine and cause overheating.
CP - OPTI
6
Installation
English
4. Connecting the processor
Water connection: Connect the water hose which comes out from the processor’s
back to the fresh water supply.
All other hoses (see diagram page 7): Connect the enclosed hoses according to
colour system onto the front of the machine. Put hose clip (enclosed in accessory
bag) over hose end, before attaching to connection. Warm up hose end (with hot
water or hot-air fan) and push it onto the respective connection. Finally push clip
over hose and connection.
Cut hoses to required length. Then integrate the stop cocks into the three drainage
hoses in such a position, that they are easy to reach.
Connect the suction pipes to the hose ends for the replenishment tanks using hose
clips. Put suction pipes through cover opening into respective replenishment tanks
and snap them in.
The overflow and drainage hoses from the developer and fixer should be guided
into their respective collecting containers.
The overflow and the drainage of the water can either be guided into the drainage
syphon or into respective collection containers.
min. 70 cm
Drainage and
collecting con-
tainers
CP - OPTI
Installation
Hose Connections
Very important:
The hose piping should be straight (without the hoses going
up and down) with a constant fall. The hoses should be as
short as possible and without bends and kinks. This is very
important for the water overflow hose. Bad piping work will
cause the machine to overflow!
Inform yourself of the local water board regulations regarding
drainage. These regulations may differ from information in
this manual, but they should be complied with.
If the machine is installed table-top, ensure that the table is
stable enough and does not wobble.
Danger of Overflow!
Use the included cable binders (accessory bag) to secure the
hoses. Fix all hose ends which guide into syphon or collecting
container, so that they do not drop into the liquid.
DEV
FIX
H O
2
Pay attention to the correct colour connections:
Developer:
red;
Fixer:
blue;
Water:
clear
Replenishment
Overflow
Outlet
Replenishment tanks
Replenishment tanks
Overflow
Overflow
Overflow
Outlet
Outlet
Outlet
Developer (red)
Water (clear)
Fixer (blue)
CP - OPTI
8
Initial Operation
English
Initial Operation
1. Test run
a.
Close the three drainage stop cocks and fill the tanks and replenishment con-
tainers with water. Open water inflow tap. Connect electrical socket and
switch the machine on. Water now flows into water tank. The circulation pump
activates, however the hosing of the machine must be ventilated.
b.
Ventilation of the replenishment pump:
Set temperature dial to position „Manual pumping“. Keep dial to this position
until no more air bubbles rise in the tanks.
c.
Ventilation of the circulation pump:
If air is in the circulation pump, a loud running noise can be heard. Switch the
machine off again. Open the stop cocks of the developer and fixer for five sec-
onds and then close them again. Then switch the machine on again. Repeat
this procedure until no more air bubbles are visible in the developer and fixer
baths and until the circulation pump runs quietly.
d.
Check all hose connections for leakage. Switch machine off and drain water
out.
2. Fill processor with chemicals
Prepare chemicals inside the replenishment containers according to manufacturers
instructions.
Fill up processor manually
By using a suitable container, pour chemicals into the respective tanks. First the fix-
er and then the developer. Caution: when filling, be sure that chemicals do no
splash from one bath into another. When fixer solution is mixed with developer so-
lution, the developer chemical is destroyed.
Snap each suction pipe into the respective cover of it’s replenishment container and
close it carefully. Place containers under processor.
Using replenishment pump
Filling of processor can also be done by use of the replenishment pump (this takes
much more time). The chemicals containers need to be filled manually with at least
1.0 litres of fresh chemicals, to ensure the pumps will not run dry. Snap each suction
pipe into the respective cover of it’s replenishment container and close it carefully.
Place containers under processor. Now set the temperature adjusting knob to po-
sition “Manual pumping”. Leave the knob in this position until the tanks have been
filled. After 20 minutes this function stops automatically - to restart a cycle, turn dial
to another position and return to „Manual pumping“. ›‹Limitation: Limitation: The
function may fill up tanks of developer and fixer to different levels. This may be due
to different causes. If this happens, then use a suitable container to fill up the tanks
completely.
Important!
Processor should not be run dry!
Upon commissioning and every refilling the pumps must be
vented.
Warning, hot surface!
CP - OPTI
Operation
Operation
Short Overview and Control Panel
*Only machines with cassette box:
When processing roll films in cassettes, pull approx. 10 cm of film out of the cas-
sette and fold the corners. Place cassette into cassette box and feed film into the
infeed. Notice that on processors with cassette box, the film switch is located in the
middle of the infeed.
LED displays
•
Infeeding film
Wait with the next film until light goes off.
•
Power
When power is on the LED lights up.
•
Bath temperature
Flashes when temperature is not reached.
Temperature dial
Manual pumping
Important!
Safety function stops film transportation when cover is re-
moved. Therefore keep cover placed on the machine when
processing films.
28
32
33
34
35
36
37
29
30
31
®
Film input
Light protection cover
Power switch
Control panel
Roll film cassette
Film
switch
Developing
Replenishment Developer
CP - OPTI
10
Operation
English
Before use...
1.
Close water-drainage stop cock.
2.
Open water tap.
3.
Switch processor on.
4.
Check liquid level in replenishment and drainage collecting containers.
5.
Wait until the developer temperature is reached. If the temperature has not
been reached, the bath temperature light is flashing.
6.
Run cleaning films through processor.
Working procedure
7.
Processing films:
Open light protection cover. Important: Put film first on left side of infeed tray
and then feed in. During processing films please watch the display „Infeeding
film“. If this is lit, wait until it goes off again and an acoustic signal can be
heard, before inserting the next film.
After work...
8.
Switch processor off.
9.
Close water tap.
10. Open water-drainage stop cock and drain water out of the machine.
Attention:
Upon first operation and each refilling of a devleoper check
the function of the circulation pump and vent the pump if
required (see page 8)
Do not place any
object on the proc-
essor.
Film exit
Fixing
Washing
Drying
Fresh water inflow
Replenishment fixer
CP - OPTI
Operation
Stand-by mode
When no film is being processed, the machine switches to Stand-by. The chemicals
remain at a constant temperature. The film transport and water inflow activate at in-
tervals to avoid crystalisation of the chemicals on the transport rollers. Entry of the
next film is possible at any time.
Switching on the machine
The processor is switched on at the main switch on the front side. After switching
on, the water tank will be filled up and a replenishment cycle will be carried out. The
developer bath will be heated up. If the temperature has not been reached, the bath
temperature light is flashing. When the machine is ready and the developer temper-
ature has been reached, a long signal sound is heard.
Bath temperature
The processor heats up the chemicals automatically to the dial-set temperature.
When this temperature is reached for the first time after switching on the machine,
a long signal tone can be heard. This is also the case when the temperature is
changed by the dial switch.
If the difference between actual and dial-set temperature is more than 1 °C, the
bath temperature LED will flash. If a film has been fed in and the temperature is not
reached, a warning sound will be heard.
Display “Infeeding film”
If films are fed into the processor without clearance one after an other, then this may
cause a film jam. The light „Infeeding film“ is on during infeeding a film. After the film
has been fed in completely the light goes off. Additionally an acoustic signal indi-
cates that the next film can be fed in.
Manual pumping
The function „Manual pumping“ activates the replenishment pump. and it pumps
chemicals into the tanks in addition to automatic replenishment. To do so, turn the
temperature dial to the position “Manual pumping”. This function is available only in
standby mode (when there is no film in the unit). If the dial is left in this position, the
pumping function is stopped automatically after 20 minutes. To restart a cycle turn
dial to a temperature and return to “Manual pumping”.
Please note: Films cannot be processed when dial is set to “Manual pumping”. Also
bath temperature is set automatically to 28 °C during “Manual pumping”. After end-
ing the manual pumping the temperature must be set again by turning the dial. It
takes some time until the temperature is reached (look at the bath temperature dis-
play).
Anti-crystalisation function
During the stand-by mode, the film transport, the dryer ventilation, the dryer heating
and the water inflow are activated every 20 minutes for a period of 20 seconds. This
prevents the build-up of crystals on the rollers.
Time replenishment (Anti-oxidation function)
Also during the stand-by, the developer chemicals are subject to change which
causes their deterioration. By means of the time replenishment, a replenishment cy-
cle is activated after 60 minutes without replenishment. The pump runs for
15 seconds. With this function, the quality of the developer chemicals are main-
tained even when standing idle for long periods.
Both chemical pumps - fixer and developer - are driven parallel
by only one motor and therefore they run always together.
CP - OPTI 12
Operation
English
Automatic replenishment
Depending on the processed amount of films the chemicals are replenished auto-
matically. This is done by pumping chemicals from the replenishment containers.
By means of the film detection switch at the film-infeed, the surface of the proc-
essed films is calculated and after approx. 0.25 m2 a replenishment cycle is auto-
matically activated for 40 seconds. The replenishment volume per cycle (with pump
set at 100 %) is 150 ml. The table below shows the replenishment volume in
ml per m2 film surface, dependent on the film width and the pump setting.
*Setting at 50 Hz current resp. settings at 60 Hz current are in brackets
**Standard setting
Rollfilms and Paper Films
Rollfilms can be transported into the machine when a leader film (10x10 cm) is at-
tached, that has been prepared with a chemical resistant adhesive tape.
Replenishment Rate
Film width
Setting of replenishment pump*
100 % (85 %)**
75 % (62 %)
35 cm
600 ml/m2
450 ml/m2
24 cm
870 ml/m2
650 ml/m2
18 cm
1150 ml/m2
875 ml/m2
Rollfilms without leader films and paper should be folded on the corners as per
diagram displayed on the right.
Please note:
High or low room temperature may influence the function of
the film processor machine.
CP - OPTI
Care
Care
Daily Care
Before use...
•
Remove dirt and dust from film-infeed with soft cloth.
•
Run 2 - 3 cleaner films through processor to remove all accumulated dirt and
dust from the rollers.
•
Check the liquid level in the replenishment containers and if necessary refill.
After use...
•
When work has been completed at the end of the day, drain water out of the
machine. This avoids the growth of algae in the water.
Weekly Care
The developer chemicals cause residue buildup in the machine. This residue has a
negative effect on the developing process of the film material. For this reason the
processor should be regularly cleaned. Proceed as follows:
1.
Switch machine off and remove cover.
2.
Loosen the securing latches (red, blue and beige) of the drive shafts of each
roller rack at the right side.
3.
Remove the roller-racks. First of all remove the large dryer-rack (beige). The
racks are easier to remove and insert when they are slightly tilted. Then
remove the fixer (blue) and developer (red) racks in sequence.
4.
Rinse all racks thoroughly under warm running water and then leave to drain
off. Use a soft sponge (do not use scouring-pad, as this would scratch the roll-
ers!) and remove the pollution from the rollers.
5.
Replace the racks: Red = Developer, Blue = Fixer. Beige = Washing/Drying.
Ensure that the racks are firmly installed and do not forget to close the secur-
ing latches on the drive shafts.
6.
Replace machine cover and ensure it is securely closed.
7.
Clean processor outer shell with damp cloth. Do not use aggressive cleaners
or solvents.
Attention: Do not let any liquid drop inside the processor or
run over the control. Liquids may cause damage to the proc-
essor.
Please note:
When removing the Rinsing / Drying roller-rack, ensure that no
water gets into the film dryer air channel.
CP - OPTI
14
Care
English
Thorough Cleaning
Every three months (maybe earlier) a thorough cleaning is necessary, depending
on the quantity of films processed. Tank cleaners are available for developer and
water baths. The fixer bath is cleaned with water. When preparing chemical tank
cleaners, follow manufacturers instructions explicitly.
How to proceed:
1.
Switch the machine off and empty all tanks by opening the stop cocks.
2.
Remove machine cover. When all tanks are emptied, close stop cocks again.
Now fill the fixer-tank with water. Prepare cleaner solutions for developer and
water baths and fill into respective tanks.
3.
Remove suction pipes from the replenishment containers and place them in a
water filled bucket. Attention: Do not add chemical cleaners here!
4.
Close machine cover and switch machine on.
5.
Start film transportation and keep running for 10 to 20 minutes. Place a film in
the infeed so that it activates the film switch but will not be pulled into the
processor. During the operation with water, the installed roller racks will be
cleaned.
6.
Important: After completion of tank cleaning, the tank should be rinsed thor-
oughly with clean water. To do this, fill the machine with fresh water twice and
each time, let the machine run for a 10 minute period. Empty the tanks and re-
close the stop cocks.
7.
Take out the roller-racks and rinse them thoroughly with running water.
Remove remaining dirt from the rollers by using a sponge and clean thor-
oughly. Doing this, the rollers can be turned by turning the drive shaft.
Remove the sheet metal covers from the dryer rack and clean the rack with
soap (dishwashing liquid). Reinstall the roller-racks in the machine.
8.
Refill the tanks with respective chemicals. Replace the suction pipes into the
replenishment containers. In certain circumstances the circulation system
must be ventilated: see page 8, item 1c.
9.
For quality check, process test films.
Before you go on holiday...
or in the event that your processor will not be in use for longer than two weeks, all
the chemicals have to be emptied out of the tanks. In case you don't want to do a
complete tank cleaning at once, then fill the tanks after emptying, with water.
Attention: Do not use alcohol containing solvents to clean the
machine!
The colour changes in the baths is normal; it is caused by the
properties of the chemicals!
CP - OPTI
Maintenance / Disposal
Maintenance / Disposal
Maintenance Protocol
Installation
Parameters Set
Maintenance work performed (see page 16)
Name:
Machine type:
Serial number:
Technician:
Training:
by:
Telephone:
Date:
Guarantee until:
Developer temp.:
Dryer temp.:
Cycle time:
Dev. reg. volume:
Dev. reg. volume:
Anti-oxidation:
Developer:
Fixer:
Film type:
Changed by:
Date:
Developer temp.:
Dryer temp.:
Cycle time:
Dev. reg. volume:
Dev. reg. volume:
Anti-oxidation:
Developer:
Fixer:
Film type:
Changed by:
Date:
Developer temp.:
Dryer temp.:
Cycle time:
Dev. reg. volume:
Dev. reg. volume:
Anti-oxidation:
Developer:
Fixer:
Film type:
Changed by:
Date:
Maintenance
work
performed
Maintenance
work
performed
Maintenance
work
performed
Maintenance
work
performed
Maintenance
work
performed
Date:
Date:
Date:
Date:
Date:
Name:
Name:
Name:
Name:
Name:
next maintenance:
next maintenance:
next maintenance:
next maintenance:
next maintenance:
Maintenance
work
performed
Maintenance
work
performed
Maintenance
work
performed
Maintenance
work
performed
Maintenance
work
performed
Date:
Date:
Date:
Date:
Date:
Name:
Name:
Name:
Name:
Name:
next maintenance:
next maintenance:
next maintenance:
next maintenance:
next maintenance:
Maintenance
work
performed
Maintenance
work
performed
Maintenance
work
performed
Maintenance
work
performed
Maintenance
work
performed
Date:
Date:
Date:
Date:
Date:
Name:
Name:
Name:
Name:
Name:
next maintenance:
next maintenance:
next maintenance:
next maintenance:
next maintenance:
CP - OPTI
16
Maintenance / Disposal
English
Recommended Maintenance Work:
1.
Functional check
film intake / film transport / replenishment / bath heating / dryer heating / water supply
2.
Cleaning
2.1.
Switch off machine, remove cover
2.2.
Empty all three tanks
2.3.
Close drain cocks and fill tanks with water
2.4.
Install cover, switch machine on
2.5.
Fill two additional vessels with water, put suction pipes into these vessels and activate replenishment for at
least two minutes (to remove residues of chemicals from replenishing hoses)
2.6.
Switch machine on for a few minutes
2.7.
Switch machine off
2.8.
Empty all tanks
2.9.
Prepare tank cleaning agent for developer and water tank according to manufacturer´s instructions
2.10. Fill developer and water tank with tank cleaning agent (do not use the replenishment pumps to do so)
2.11. Fill fixer tank with water
2.12. Place suction pipes into empty tanks
2.13. Install cover, switch machine on
2.14. Wait until the operating temperature is reached, approx. 30 °C (observe information concerning temperature,
time, cleaning procedure contained in the datasheet of the tank cleaning agent)
2.15. Activate the transport (activate film intake switch)
2.16. After approx. 15 minutes (observe information concerning temperature, time, cleaning procedure contained in
the datasheet of the tank cleaning agent) switch film transport off
2.17. Remove cover, neutralise developer tank (observe information concerning temperature, time, cleaning proce-
dure contained in the datasheet of the tank cleaning agent)
2.18. Switch machine off
2.19. Empty all three tanks
2.20. Fill machine with water and switch it on
2.21. Put suction pipes into vessels with water
2.22. Activate regeneration pumps for at least five minutes
2.23. Check all pumps for tightness
2.24. Switch machine off
2.25. Drain tanks
2.26. Fill tanks 3/4 with water
2.27. Switch machine on
2.28. Activate replenishment pumps manually until tanks overflow
2.29. Activate film transport for a few minutes
2.30. Switch machine off and drain all three tanks
Attention:
Never start the machine up unless it is filled with liquid!
Attention:
Do not use chlorine containing cleaning agents!
CP - OPTI
Maintenance / Disposal
2.31. Repeat item 2.20 to 2.30 if required (observe information e. g. concerning temperature, time and cleaning pro-
cedure outlined in tank cleaner datasheet)
2.32. Remove roller racks from the machine and remove dirt under flowing water using a soft rag or sponge
2.33. Remove risidual dirt particles from the tanks
2.34. Clean all toothed gear wheels, axles, bearings and rollers, check them for damage (replace if required)
2.35. Remove light protection flap and wipe its underside using a soft rag
2.36. Clean inlet plate using a soft rag
2.37. Reinstall light protection flap
2.38. Align roller racks and re-insert them in machine
2.39. Fill machine with chemicals
2.40. Switch machine on
2.41. Adjust bath temperature to previously adjusted value
2.42. Feed cleaning film (approx. 4 pieces)
2.43. Check function as described under item 1.
2.44. Approx. 15 minutes after reaching of the bath temperature, check it by measuring and recalibrate if required
(see operating instructions page 25)
2.45. Perform sensitometric test
3.
Perform constancy tests based on the applicable national standards (e.g. IEC
61223-2-1 and DIN 6868-2).
Be sure to dipose properly of used machines.
Used machine contain valueable materials that should be recycled
and turned over for proper treatment. Please be sure to turn used
machines over to approved recycling centers.
CP - OPTI
18
Problems and Solutions
English
Problems and Solutions
Advice on Film Defects
Your processor has been constructed for long term use. If however irregularities
might occur, you will find help to locate the problem below. Please check the listed
points, before calling your service-technician.
Films do not have enough density
•
Bath temperature is too low.
•
Developing time too short.
•
Exposure time is too short.
•
Replenishment rate of developer too low.
•
Developer chemicals are exhausted or too strong diluted: Renew.
•
Fixer solution has been mixed into developer: Renew. Clean and rinse bath
well before refilling.
•
Circulation is broken down.
Too high a density
•
Bath temperature too high.
•
Developing time too long.
•
Exposure time is too long.
•
Replenishment rate of developer too high.
•
Developer chemicals are too high diluted: Renew.
•
After renewing chemicals: Starter is missing.
•
Circulation is broken down.
Films will not dry
•
If warm air comes out of the air channel in the dryer, chemicals and film type
should be checked.
•
Fixer solution is exhausted or diluted.
Film has a yellow-green surface
•
Not fixed correctly. Check the film type and fixer chemistry.
•
Fixer solution is exhausted or diluted. Replenishment rate of fixer is too low.
Scratches, pressure marks, dirt on film
•
Prior to processing films, run cleaner films through the processor.
•
Pressure marks caused by careless handling, finger nails etc.
•
Rollers are polluted. Clean tanks and roller racks.
Cloudy film
•
Level in developer is too low.
•
First guide bar of fixer rack is dirty (condensate or crystals). Clean roller-
racks.
•
Developer is old or circulation not working.
•
Try processing films by infeeding them with emulsion side up.
CP - OPTI
Problems and Solutions
Advice on Machine Errors
Machine does not switch on
•
Ensure that electrical plug is firmly inserted into socket.
•
Ensure that electrical socket has power supply by testing with an appliance
(e.g. tablelight).
Film will not feed in but the dryer fan works
•
Place cover firmly on machine, ensuring that the cover switch on the right
front side has been activated.
Machine does not start automatically
•
Film switch left side at the infeed has not been pressed down. Feed in the film
to activate the switch.
Developer temperature too low
•
The temperature dial is set between two positions.
Replenishment pump does not pump
•
Check whether the replenishment containers are full and that the end of the
suction pipe is positioned under the liquid level.
•
Check whether there is air in the replenishment pipes. If this is the case, then
check the pipe connections.
Rinsing water does not flow
•
Open water inflow tap.
•
Water pressure in the water system is too low: Minimum pressure 2 bar
(29 psi).
Water tank overflows
•
Water drainage hose (overflow) is bent. The hose end should be positioned
above the drainage level in the syphon (see diagram on page 7).
•
Check water drainage in the tank and hose for blockage and pollution. The
drainage hoses should have a constant fall.
The film does not transport correctly
•
Film is fed in and gets caught in the machine: Check the positioning of the
racks in the machine and make sure that the latches are closed.
Important notes:
Ensure correct seating of roller racks; keep the lock closed at
all times.
Don’t operate processor with empty replenishment tanks.
After a long machine shut down check bath level and refill if
necessary.
CP - OPTI
20
English
Service Manual
Table of Contents
Installation Data ................................................................................................. 21
Trouble Shooting ............................................................................................... 23
Spare Parts List.................................................................................................. 29
Electric Diagrams............................................................................................... 35
CP - OPTI
Installation Data
Installation Data
75cm (29.5’‘)
70cm (27.6’‘)
42cm (16.5’‘)
77cm (30.3’‘)
59cm (23.2’‘)
59cm (23.2’‘)
Optional worktable or
closed cabinet
(accessories)
1
2
4
5
7
3
30cm
(12’‘)
130cm (51’‘)
85cm (34’‘)
20cm (8’‘)
50cm (20’‘)
60x40cm (24x16’‘)
6
8
1.
Wall socket
220-240 V, 16 A resp. 110 V,
15 A (depending on machine
model). Power lead should be
equipped with Earth-Leakage
Switch, 25 A / 30 mA nominal
error-current. In addition, a
power control switch can be
installed.
2.
Fresh water connection 3/4”
with stop cock, permissible
pressure 2-10 bar, water tem-
perature 5-25 °C.
3.
Wasserabfluß PVC-Rohr ∅ 50
mm (2“) mit Siphon.
4.
Drainage resp. collecting con-
tainers for used developer.
5.
Drainage resp. collecting con-
tainers for used fixer.
6.
Storing space for replenish-
ment tanks: Below machine or
externally.
7.
Ventilation of darkroom is nec-
essary.
8.
Sink with freshwater and flexi-
ble hose. Inner dimensions
minimum (LxWxH)
60x40x30 cm (24x16x12“).
Sink
Darkroom
Measures and positions are recommendations
CP - OPTI
Trouble Shooting
Trouble Shooting
Summary
1
Algae ............................................................................................................30
1.1 Excessive algae growth in water tank....................................................30
2
General.........................................................................................................31
2.1 Unit does not switch on..........................................................................31
3
Drive .............................................................................................................31
3.1 Film transport does not run, fan runs.....................................................31
3.2 Machine does not start automatically ....................................................31
3.3 Machine doesn't stop automatically.......................................................31
3.4 Drive motor does not run .......................................................................31
3.5 Machine stops before film comes out ....................................................31
3.6 Relation between processing time and developer temperature.............32
3.7 Changing the transport speed ...............................................................32
4
Baths ............................................................................................................33
4.1 No circulation in bath .............................................................................33
4.2 Developer temperature too high ............................................................33
4.3 Developer temperature too low..............................................................33
4.4 Developer temperature too low, fixer temperature too high...................33
4.5 Removing the turning knob....................................................................33
4.6 Calibration of the developer temperature ..............................................33
5
Film defects .................................................................................................34
5.1 Films will not dry ....................................................................................34
5.2 The film does not transport correctly .....................................................34
5.3 Scratches, pressure marks, dirt on film .................................................34
6
Replenishment ............................................................................................35
6.1 Replenishment pump does not work or works insufficiently ..................35
6.2 Regeneriermenge zu hoch oder zu niedrig............................................35
6.3 Adjust replenishment pump ...................................................................35
7
Dryer.............................................................................................................35
7.1 Dryer fan does not run or runs with too low speed ................................35
8
Water ............................................................................................................35
8.1 Rinsing water does not flow...................................................................35
8.2 Water tank overflows .............................................................................35
1
Algae
1.1 Excessive algae growth in water tank
•
Algae growth inside the water tank is not only anoying, it causes increased
cleaning work and leaves residue on the films. When algae growth increases,
countermeasures are in demand:
•
When work has been completed at the end of the day, drain water out of the
machine.
•
Clean dryer-water rack regularly. Use soft sponge and soap to remove resi-
due from the rollers.
•
Install a particle filter system in the fresh water supply for the processor.
•
If water tank overflows due to algae growth blocking the overflow hose, then
the overflow hose can directly be connected to the connection at the water
tank inside the machine.
•
If no other solutions can be found, then usage of Anti-Algae-Agents can be a
great improvement (automatic dispensers work the best). However, it is
known that cleaning agents containing chlorine may corrode rubber rollers
and high-grade steel in the tank area (check before use).
CP - OPTI 23
Trouble Shooting
English
2
General
2.1 Unit does not switch on
•
Ensure that electrical socket has power supply.
•
Check machine fuse in main switch.
•
While power switch is on, check the following components: Voltage on contact
of main switch - if no voltage: - change main switch. Check input voltage at
electronics. If the voltage is normal, exchange the electronics. if no voltage:
check the cable wiring harness.
3
Drive
3.1 Film transport does not run, fan runs
•
When placing processor cover on, the cover switch should be activated, re-
adjust if necessary.
•
Cover switch has no current passage when activated: Replace.
•
Check screwing of chain wheel on motor- and driveshaft.
3.2 Machine does not start automatically
•
Film switch is not correctly positioned or operator wire is bent. Re-adjust film
switch and operator wire.
•
Check following parts: Film switch, wiring of film switch and electronic.
•
Check wiring from electronic to the connections of components (motor, fan,
dryer heating, solenoid valve). If the connections have no fault then electronic
is possibly defective.
3.3 Machine doesn't stop automatically
•
Display „Infeeding film“ is permanently illuminated: Wire band of film switch is
jammed. Readjust wire.
•
Check following parts: Film switch, wiring of film switch and electronic.
3.4 Drive motor does not run
•
Check cover switch.
•
Check drive motor: If voltage can be detected on motor, then motor is defec-
tive.
•
Dryer fan runs but no voltage on motor: Interruption in the wiring.
3.5 Machine stops before film comes out
The cycle time is the processing time which runs off after a film has passed the film
switch. Activate the switch in the infeed tray with a film and remove the film. Meas-
ure the time until the processor stops automatically.
Following times can be measured (+/- 5%):
Change cycle time if necessary. This can be accomplished by changing the position
of the jumper on the upper side of the electronics.
Pleasse use as replacement fuses only the PROTEC®s. These
fuses are optimized for use under existing conditions.
Standard machine:
125 s
Jumper 2-3
Mammography:
155 s
Jumper 1-2
Optional type:
190 s
Jumper removed
CP - OPTI
Trouble Shooting
3.6 Relation between processing time and developer temperature
The following chart demonstrates guide value relations between developer temper-
atures and processing times. Variations are possible depending on the various films
and chemicals. Changing the transport speed see 3.7.
3.7 Changing the transport speed
The processing speed can be changed by changing the gear wheels. To do this the
tanks need to be emptied and the machine be turned over. After removing the drive
motor the chain gears can be changed. Please note that the jumper on the PCB
needs to be placed to the indicated position.
Following gear combinations are available:
t = Processing time in seconds
Gm = Chain gear on motor shaft
Gs = Chain gear on drive shaft
Processing time “Dry to Dry”
Developer temperature
105 s
32 °C - 34 °C
118 s
32 °C - 34 °C
135 s
31 °C - 33 °C
143 s
30 °C - 32 °C
167 s
30 °C - 32 °C
220-240 V, 50 Hz versions
110-120 V, 50 Hz versions
220-240 V, 60 Hz versions
t
Gm
Gs
Jumper
position
Devel-
oper time
(s)
Infeed
speed
(cm/min)
t
Gm
Gs
Jumper
position
Devel-
oper time
(s)
Infeed
speed
(in/min)
105
t=17
t=16
2-3
29
48
105
t=16
t=18
2-3
29
48
118
t=16
t=17
1-2
32
43
118
t=14
t=17
1-2
32
43
135
t=14
t=17
1-2
37
37
135
t=12
t=17
1-2
37
37
143
t=14
t=18
removed
39
35
143
t=12
t=18
removed
39
35
167
t=12
t=18
removed
46
30
167
t=10
t=18
removed
46
30
110-120 V, 60 Hz versions
t
Gm
Gs
Jumper
position
Devel-
oper time
(s)
Infeed
speed
(in/min)
105
(90)
t=14
t=16
2-3
29
18,9
113
t=14
t=17
2-3
31
17,6
124
t=12
t=16
1-2
34
16,0
132
t=12
t=17
1-2
36
15,0
139
t=12
t=18
1-2
38
14,3
CP - OPTI
25
Trouble Shooting
English
4
Baths
4.1 No circulation in bath
•
Circulation pump works but no circulation in bath: Air lock in heating and cir-
culation system. Ventilate pump (see page 8).
•
Particles in the pump chamber. The pump chamber can be easily opened by
removing the four clips. When closing again ensure that the rubber seal is
positioned correctly and not damaged.
•
Check connection of pump, circulation pump possibly defective.
4.2 Developer temperature too high
•
Check attachment of temperature sensor. This should be firmly positioned on
tube and completely covered with foam rubber.
•
Check sensor: Check voltage between pin 3 (brown) and pin 2 (white).
between 0.1 and 0.5 V.
•
If the sensor has no fault then electronic is defective.
4.3 Developer temperature too low
•
Check circulation pump. Air lock in the circulation pump: Ventilate pump (see
page 13). If no circulationcan be detected: Check wiring of circulation pump,
pump possibly defective.
•
Bath is not heated: Check temperature safety switch on heat-exchanger.
Check heating element: impedance across the element should reach
approx. 66 W.
•
Check temperature sensor (see 4.2).
•
If no error can be found then electronic is possibly defective.
4.4 Developer temperature too low, fixer temperature too high
•
Air lock in the circulation pump: Ventilate circulation (see page 8).
4.5 Removing the turning knob
•
Using flat pliers pull the lever from the knob, while holding the bottom part
(knob) down with the other hand.
•
Open the screw of the collet (Attention: don’t open completely) and pull the
knob out.
•
When reinstalling the knob turn axis on PCB to end position counter clock-
wise. Fix the knob in that the pointer is at position of „Manual pumping“.
4.6 Calibration of the developer temperature
•
temperature deviations of +/-1.5 °C can be compensated for by adjusting the
potentiometer on the control board. It can be reached after removing the turn-
ing knob (see page on right hand side) from above. Turning clockwise
decreases the temperature.
Only remove the electronics when the turning knob has been
removed!
26 CP - OPTI
Trouble Shooting
5
Film defects
5.1 Films will not dry
•
No air comes out of air channel: Check wiring of dryer fan, fan is possibly
defective.
•
Cold air comes out of air channel: Check wiring of heating element in the air
channel, heating element possibly defective.
•
Hot air comes out of air channel, but the film is still not dried to satisfaction.
Check chemicals and film type. If this leads to no solution then the transport
speed of the machine can be reduced (see Changing the transport speed,
page 24).
5.2 The film does not transport correctly
•
Check the positioning of the racks in the machine and make sure that the
latches are closed.
•
Check the roller racks: Position of the guide elements, rollers are in correct
position and are not loose, flat springs are not bent, all gears are in place.
•
Motor runs: The worm gear of the drive shaft should be secured with a splint
to avoid twisting. Check the screws and positioning of the chain and chain
wheel.
5.3 Scratches, pressure marks, dirt on film
•
Straight scratches in the infeed direction indicate faulty guide elements.
Check each rack and straighten up the guide elements. If mechanically dam-
aged, replace the guide elements.
•
Pressure marks caused due to polluted or damaged rollers. Check rollers for
visible damage. Rubber rollers sometimes swell up. Exchange defective roll-
ers.
-
+
28
°C
32
33
34
35
36
37
29
30
31
CP - OPTI
27
Trouble Shooting
English
6
Replenishment
6.1 Replenishment pump does not work or works insufficiently
6.2 Regeneriermenge zu hoch oder zu niedrig
•
The replenishment rate can be changed by adjusting the stroke of the pump.
To do this, the eccentric on the replenishment pump must be adjusted. Maxi-
mum pump capacity is 240 ml/min (100 %).
6.3 Adjust replenishment pump
•
For the adjustment of the eccentric first open the allan screw on the big
eccentric with the red line. If screw is not reachable, then start the „Manual
pumping“ (dial switch) for a short time. If the screw is reachable fastly turn
back the dial switch on a temperature position.
•
Turn the eccenter so that the red line will be at the desired position and fasten
again the allan screw.
7
Dryer
7.1 Dryer fan does not run or runs with too low speed
•
Check the correct connecting of the fan cables: bl = blue; bk = black; br =
brown.
•
If the fan is connected improperly, then the fan runs only half power.
8
Water
8.1 Rinsing water does not flow
•
Water pressure in the water system is too low: Minimum pressure 2 bar
(29 psi).
•
Valve activates, no flow passage - filter at inflow is blocked.
•
Check green water inlet hose inside the machine.
8.2 Water tank overflows
•
Water drainage hose (overflow) should have a constant fall. The hose end
should be positioned above the drainage level in the syphon.
•
Check water drainage in the tank and hose for blockage and pollution.
•
When extreme algae growth is registered, the overflow can be connected
directly to the fitting of the water tank.
•
Check filter in the suction pipe (repl. container) and clean it if necessary.
•
Replenishment pump sucks air in. Check hoses and connections.
•
Check eccentric position. Flow rate reads 240 ml/min at maximum eccentric
position 100%.
•
(60 Hz: 240 ml/min at 85 %)
•
Activate the „Manual pumping“ and while on, check the voltage of connection
X2 on the power PCB. If no voltage can be registered - exchange power PCB.
Minimum setting must not be below 75 %.
•
Clean valves inside connection tube of pump.
Install valve-insert correctly: Pay attention to flow-through di-
rection!
CP - OPTI
Spare Parts
Pos.
Order No.
Description:
1
1170-0-0200
Cover complete
2
2006-0-0005
Drain stop cock 10 mm
3
1170-0-2000
Replenishment tank 12 l dev.
1170-0-2100
Replenishment tanke 12 l fix.
4
1170-0-1760
Saugrohr m. Filter f. 12l Behälter rund
6
2004-0-0003
Electrical power lead 220-240 V
2004-0-0021
Electrical power lead 220-240 V
7
2018-0-0001
Water inlet tube
8
1267-0-0000
Processor stand
-
1267-0-0010
Closed base cabinet
9
1101-0-2000
Replenishment tank 25 l dev.
1101-0-2100
Replenishment tank 25 l fix.
10
1101-0-1700
Saugrohr m. Filter f. 25l Behälter
11
1101-0-4100
Floating cover developer
-
2018-0-0012
Hose 10 x 2 mm, celar, reinforced
-
2018-0-0009
Hose 10 x 2 mm, blue, reinforced
-
2018-0-0008
Hose 10 x 2 mm, red, reinforced
-
2018-0-0005
Hose 4 x 1 mm, green
-
2018-0-0021
Hose 9 x 2 mm, red, clear
-
2018-0-2022
Hose 9 x 2 mm, blue, clear
-
2022-0-0014
Tube clamp Snap
-
2022-0-0019
Wire tube clamp
-
2022-0-0026
Wire tube clamp
-
2022-0-0028
Wire tube clamp
-
1101-0-4600
Floating balls, 300 pcs.
-
1101-0-4800
Floating balls, 200 pcs.
1
2
3
4
6
7
8
9
10
11
CP - OPTI
29
Spare Parts
English
Pos.
Order No.
Description:
20
0202-1-0008
Replenishment pump 2KBA 220-240 V, 50/60 Hz
0202-6-0008
Replenishment pump 2KBA 115 V, 50/60 Hz
20a
0002-1-0008
Kegelventil f. Pos 20
21
2002-1-0013
Circulation pump 220-240 V, 50/60 Hz
2002-6-0013
Circulation pump 110 V, 50/60 Hz
22
1170-0-1400
Mains plug combi 220-240 V UL
23
2010-0-0010
Fusible, gold, T 10 A / 250 V UL
22+23
2028-0-0036
Geräteschalter therm. Electronic 110-120 V
24
1170-0-0702
Angle connection (grey)
25
1170-4-1300
Control unit 110 V °F
1170-5-1300
Control unit 220-240 V
0170-6-1300
Control unit 110-120 V
26
0170-0-2400
Micro-switch with operator (cover)
0170-4-2400
Micro-switch with operator (cover) UL
27
2007-0-0010
Operator for micro-switch
28
1101-0-0704
Angle fitting
-
0016-0-0002
PU - glue kit
29
2021-0-0001
Screw-in connector
-
1170-0-1250
Wiring harness V2 230 V
-
1170-0-1251
Wiring harness V2 115 V
20a
20
29
21
(53)
(30)
(61)
28
(58)
24
22-23
25
20
26
27
CP - OPTI
Spare Parts
Pos.
Order No.
Description:
30
2001-0-0003
Main drive motor 220-240 V, 50 Hz
2001-6-0003
Main drive motor 120 V, 50/60 Hz
31
1170-0-1101
Motor bracket
32
1170-0-1501
Drive shaft worm-gear
33
1170-0-1503
Worm-gear
34
1170-0-1502
Supporting bock
35
1170-0-1506
Chain wheel t=12
1170-0-1504
Chain wheel t=14
1170-0-1505
Chain wheel t=16
1170-0-1102
Chain wheel t=17
1170-0-1507
Chain wheel t=18
37
2037-0-0002
Chain 6 mm with coupler link
38
3000-9-4013
Splint pin 2.0x20 mm, inox
39
0170-0-3100
Lichtschutzdeckel kpl.
40
1170-0-3104
Sealing wedge
35
37
35
34
38
33
32
30
31
40
39
CP - OPTI
31
Spare Parts
English
Pos.
Order No.
Description:
50
1170-0-1301
Air channel
51
0170-0-1300
Heating elment 230 V, 1100 W
0170-6-1310
Heating elment 110 V, 900 W
standard model
0170-5-1300
Heating element 230 V, 1000 W
53
2008-5-0007
Dryer fan 220-240 V, 50/60 Hz
2008-6-0007
Dryer fan 115 V, 50/60 Hz
54
0170-5-1900
Solenoid valve 220-240 V, 50/60 Hz
0170-6-1900
Solenoid valve 115 V, 50/60 Hz
55
1101-0-0121
Securing bracket
56
0170-0-0800
Micro-rotary-switch for film-detection with operator
58
1170-0-0804
Bracket for micro-rotary-switch
59
1170-0-0105
Film feed tray
1172-0-0105
Film feed tray graphic arts
60
2003-5-0002
Heating element 230 V, 800 W
2003-6-0002
Heating element 120 V, 400 W
61
1130-0-2101
Heat exchanger
62
2005-0-0005
Temperature safety switch mounted on heat exchanger
63
0190-0-2200
Temperature sensor
65
1170-0-1302
Channel dryer heating
55
54
56
58
60
61
62
63
59
53
65
50
51
CP - OPTI
Spare Parts
Standard roller racks
104
123
106
107
103
105
109
122
120
119
104
117
121
100
102
112
111
113
110
114
115
116
104
103
108
CP - OPTI
33
Spare Parts
English
Pos.
Order No.
Description:
Mammography Processor
-
1171-0-0600
Roller rack, dryer
-
-
-
Standard Processor
-
1170-0-0300
Roller rack, developer
-
1170-0-0400
Roller rack, fixer
-
1170-0-0600
Roller rack, dryer
100
0170-0-0301
Dryer side plate left w. shafts
1170-0-0301
Side plate, dev. (right)
0170-0-0401
Dryer side plate left w. shafts (left)
1170-0-0401
Seitenteil F. (rechts)
102
1140-0-3800
Guide bar straight, short
103
1140-0-4500
Guide bar with nose
104
1140-0-3700
Guide bar, curved
105
1170-0-0304
Flat spring 55
106
3079-8-5013
Screw M4x10, A4
107
3009-3-4023
Hexagon nut M4, A4
108
1140-0-0301
PU-roller 35 ground
109
1170-0-0310
Drive shaft rack
110
1101-0-0302
Gear t = 16, D-hole
111
1101-0-0304
Gear t = 32, round hole
112
1101-0-0303
Gear t = 32, D-hole
113
1170-0-0302
Worm wheel
114
1101-0-0305
Bushing without clearance
115
1101-0-0317
Bushing big, black
116
2014-0-0001
Circlip
117
0170-0-0601
Dryer side plate left w. shafts
1170-0-0602
Dryer side plate right
119
1140-0-0302
Rubber roller 35
120
1140-0-0605
Air jet (35)
121
1170-0-0604
Dryer plate, large
122
1170-0-0603
Dryer plate, small
123
1170-0-0303
Flat spring 35