Freestyle Precision Neo Operation & User’s Manual
Read Online
Read the Freestyle Precision Neo Operation & User’s Manual online using the built-in PDF viewer.
Text of the manual
Blood Glucose and Ketone Monitoring System
FreeStyle Precision Neo Pharmacy Guide
Time Setup
Press to turn on meter.
Press and hold the time on the
screen until the screen changes.
Set Time Format (12 hr or 24 hr clock)
Press or to change.
Press to continue.
Note: System time and date are
preset. Users or HCPs may need to
update settings to correct the hour.
Note: There is no “AM”
in the 12 hour format.
Set Minutes
The minutes blink.
Press or to set the minutes.
Press to continue.
Set Hour
The hour blinks.
Press or to set the hour.
Press to continue.
Meter date and time set up is quick and easy.
IMPORTANT: Only health care professionals can set up and change the insulin calculator settings.
Refer to package insert/label accompanied with your product for detailed instructions and indications for use.
FreeStyle and related brand marks are trademarks of Abbott Diabetes Care Inc. in various jurisdictions.
The month blinks.
Press or to set the month.
Press to continue.
Set Year
Press or to set the year.
Press to save.
Setup is complete.
Set Day
The day blinks.
Press or to set the day.
Press to continue.
Enabling Insulin Logging
How to Log Insulin Doses
While on the Home screen, press and
hold until appears.
Insulin dose logging is now enabled.
Press to turn on meter.
The patient can enable insulin
logging in one simple step.
IMPORTANT: Only health care professionals can set up and change the insulin calculator settings.
Refer to package insert/label accompanied with your product for detailed instructions and indications for use.
FreeStyle and related brand marks are trademarks of Abbott Diabetes Care Inc. in various jurisdictions.
Logging an Insulin Dose
To log a dose, press
Press the button for the
dose of insulin to be logged.
Long-acting
insulin
Meal
insulin
Press or to enter
actual dose amount.
Press to log the dose.
means the dose was taken.
Pressing again will un-log the dose.
How the patient logs their insulin doses.
Blood Glucose Trend Indicators
If a blood glucose result is lower than
3.9 mmol/L, a appears
Blood Glucose and Ketone Monitoring System
You will see a blinking after 2 low
readings in the same 3-hour period
over 5 days
If a blood glucose result is higher
than 13.3 mmol/L, a appears
You will see a blinking after 3 high
readings in the same 3-hour period
over 5 days.
Blood glucose trend indicators
are already switched on. There
is no need to enable.
IMPORTANT: Only health care professionals can set up and change the insulin calculator settings.
Refer to package insert/label accompanied with your product for detailed instructions and indications for use.
FreeStyle and related brand marks are trademarks of Abbott Diabetes Care Inc. in various jurisdictions. ME604