Extech Instruments 401025 Operation & User’s Manual
Read Online
Read the Extech Instruments 401025 Operation & User’s Manual online using the built-in PDF viewer.
Text of the manual
Congratulations on your purchase of Extech’s Digital Light Meter. This professional meter, with proper
care, will provide years of safe reliable service.
Meter Description
1.
LCD Display
2.
Data Hold Switch
3.
Power Off/Range Switch
4.
Analog Output Terminal
5.
Battery Compartment (rear)
6.
LUX/Fc Switch & Response Switch
7.
Light Sensor
8.
Zero adjust
Operation
1.
Select Units (Lux or Ft-candle) and Response Time (Fast or Slow) on the slide switch.
Typical selection is Slow and Fc using the gray lettering.
2.
Select the maximum range on the "Range Switch"
3.
Hold the "Light Sensor" so that the sensor faces the light source to be measured.
4.
The Display will indicate measured values. Use a range display multiplier if on the Lux
20,000 and 50,000 ranges or on the Fc 5000 range.
5.
To "hold" a measurement, slide the "Data Hold Switch" to the "hold" position. The reading
will "freeze" in the display until the "Data Hold Switch" is released.
Note 1: An Over Range indication is a display of "I ". If this occurs, switch to a higher range.
Note 2: For measurements made on the Fc 5000 range, the displayed reading must be multiplied
by 10.
Note 3: For measurements made on the Lux 20000 or 50000 range, the displayed reading must
be multiplied by 10 and 100 respectively.
Note 4: The meter will indicate values above the maximum ranges. The accuracy of these
measurements is unknown.
Range Display Multipliers
Range
Units
Multiplier
200
Fc
Direct reading
2000
Fc & Lux
Direct reading
5000
Fc
Reading x10
20,000
Lux
Reading x10
50,000
Lux
Reading x100
Example: If a measurement on the 5000 Fc range displays 350, then the actual measured value
is; 350 X 10 = 3500 Fc.
Selecting a Measurement Range
The meter has three measurement ranges (0-200, 0-2000, and 0-5000 Fc) and (0-2000, 0-20000, and
0-50000) Lux). The proper range selection will produce the most accurate reading. Always select the
range that produces the maximum number of digits without exceeding the maximum count for that
particular range. For example, a reading of 1456 Fc should be read on the 0 - 2000 range, not the 0-
5000 range.
Zero procedure
The meter zero (display with no light input) may change with time. Occasional checking and
adjustment may be required.
1.
Completely cover the sensor to block out any light.
2.
Set the range switch to the lowest Lux or Fc range
3.
Using a small screwdriver, adjust the “Zero” control for a zero display. The last digit may
change slightly. This is normal and does not affect the accuracy of the meter.
Analog Output
The analog output jacks on the front panel produce a 0.1mV DC per digit signal that can be used for
recording or datalogging purposes.
Lighting Type Correction Factors
The 401025 light meter is calibrated under a precise "Standard tungsten light source of 2856
OK". If the
meter is to be used under a different type of light the correction factor of from the table below should
be applied to the readings obtained.
Mercury Lamp
x1.14
Fluorescent Lamp
X0.92 to 1.12
Daylight
x1.00
Sodium
x1.22
Metal Halide
x1.00
Replacing the Battery
When the left corner of the LCD display shows "LO BAT", it indicates the battery output is below the
design limit and the battery needs to be replaced. However, reliable measurement can still be taken for
another few hours before the tester becomes inaccurate.
1. Open the Battery Cover at the back of tester and remove the battery.
2. Replace with a 9V battery and install the cover.
You, as the end user, are legally bound (EU Battery ordinance) to return all used
batteries, disposal in the household garbage is prohibited! You can hand over your
used batteries / accumulators at collection points in your community or wherever batteries
/ accumulators are sold!
Disposal: Follow the valid legal stipulations in respect of the disposal of the device at the
end of its lifecycle
13 mm (0.5") LCD (Liquid Crystal Display).
Measurement
Lux, Ft-candle (Fc).
Ranges
Lux: 0 to 50,000 Lux, 3 ranges. Foot-candle: 0-5,000 Fc, 3
ranges.
Sensor
Exclusive photo diode & color correction filter, spectrum
designed to meet C. I. E.
Zero Adj.
Manual adjustment.
Sampling Time
Approx. 0.4 sec.
Response Time
Fast: 0.25s; Slow: 1s
Over input indication
Indication of "1 "
Analog Output
0.1 mV/1 digit, max. output :200mV.
Operating Temperature
0
OC to 50
OC (32
OF to 122
OF)
Operating Humidity
Less than 80% RH.
Power Supply
006P DC 9V battery
Power Current
Approx. 2 mA DC
Weight
220 g / 0.52 LB
Dimension
Main instrument: 163x70x30mm (6.4x2.8x1.2 inch).
Sensor Probe: 85x55x12 mm (3.2x2.2x0.5 inch).
Optional Accessories
Vinyl pouch carrying case, 409996
Range Specifications
Lux
Range
In-range Display
Resolution
Accuracy
2,000 Lux
0-1,999 Lux
1 Lux
± (5% + 2digits)
20,000 Lux
2,000-19,990 Lux
10 Lux
± (5 % + 2digits)
50,000 Lux
20,000-50,000 Lux
100 Lux
±- (5 % + 2digits)
Foot-candle (Fc)
Range
In-range Display
Resolution
Accuracy
200 Fc
0-199.9 Fc
0.1 Fc
± (5% + 2digits)
2,000 Fc
200-1,999 Fc
1 Fc
± (5% + 2digits)
5,000 Fc
2,000-5,000 Fc
10 Fc
± (5% + 2digits)
Frequency Spectrum
300 400 500 600 700 800
1
0.8
0.6
0.4
0.2
0
Wavelength (nm)
Appendix A: Typical Light Levels
Lux
Foot
Candles
Lux
Foot
Candles
Factories
Home
20-75
2-7
Emergency Stairs,
Warehouse
100-150
10-15
Washing
75-150
7-15
Exit/Entrance Passages
150-200
15-20
Recreational
Activities
150-300
15-30
Packing Work
200-300
20-30
Drawing Room,
Table
300-750
30-75
Visual Work: Production
Line
300-500
30-50
Makeup
750-1,500 75-150
Typesetting: Inspection
Work
500-1,500
50-150
Reading, Study
1,500-
3,000
150-300
Electronic Assembly,
Drafting
1,000-
2,000
100-200
Sewing
Office
Restaurant
75-100
7-10
Indoor Emergency Stairs
75-150
7-15
Corridor Stairs
100-200
10-20
Corridor Stairs
150-300
15-30
Entrance, Wash
Room
200-750
20-75
Conference, Reception
Room
300-750
30-75
Cooking/Dining
Room
750-1,500 75-150
Clerical Work
750-1,500
75-150
Show Window
1,500-
2,000
150-2000
Typing, Drafting
Store
Hospital
75-150
7-15
Indoors
30-75
3-7
Emergency Stairs
150-200
15-20
Corridor/Stairs
75-100
7-10
Stairs
200-300
20-30
Reception
100-150
10-15
Sick Room,
Warehouse
300-500
30-50
Display Stand
150-200
15-20
Waiting Room
500-750
50-75
Elevator
200-750
20-75
Medical Exam Room
750-1,500 75-150
Show Window, Packing
Table
750-1,500
75-150
Operating Room
1,500-
3,000
150-300
Storefront, Show Window
5,000-
10,000
500-1000
Eye Inspection
Appendix B - Common Terms and Conversion Factors
1 lux (lx)
10-4 lm/cm2
10-4 phot (ph)
9.290 x 10-2 lm/ft2
Illuminance (Visible Flux Density)
1 lm/m2 =
9.290 x 10-2 foot-candles
Luminance (Visible Flux Density per Solid Angle)
1 lm/m2/sr =
1 candela/m2
Luminous Intensity (Visible Flux per Solid Angle)
1 lm/sr =
1 candella
Luminous Flux ( Visible Flux)
1 lumen (lm) =
1.464 x 10-3 watts @ 555 nm
Copyright © 2012 Extech Instruments Corporation (a FLIR company)
All rights reserved including the right of reproduction in whole or in part in any form.
www.extech.com