Extech Instruments MN25 Operation & User’s Manual
| Brand: | Extech Instruments |
| Category: | Multimeter |
| Model: | MN25 (Operation & User’s Manual) |
| File Type: | |
| File Size: | — |
| Pages: | 15 |
Text of Extech Instruments MN25 Operation & User’s Manual
Extech Instruments MN25 MN25 Version 2.2 Aug 2003
2
Warranty
EXTECH INSTRUMENTS CORPORATION warrants this instrument to be
free of defects in parts and workmanship for one year from date of shipment
(a six month limited warranty applies on sensors and cables). If it should
become necessary to return the instrument for service during or beyond the
warranty period, contact the Customer Service Department at (781) 890-
7440 ext. 210 for authorization or visit our website at www.exte
MN25 Version 2.2 Aug 2003 3
Safety
This meter has been designed for safe use, but must be
operated with caution. The rules listed below must be carefully
followed for safe operation.
1.
NEVER apply voltage or current to the meter that exceeds
the specified maximum:
Input Limits
Function
Maximum Input
V DC, V AC
600V DC/AC, 200Vrms on 200mV range
mA DC
200mA 250V fast acting fuse
A DC, A AC
10A 250V fast acting fuse(30 seconds
max every 15 minutes)
Resistan
MN25 Version 2.2 Aug 2003 4
Safety Symbols
NEVER operate the meter unless the back
cover and the battery and fuse doors are in
place and fastened securely. This symbol
adjacent to another symbol, terminal or
operating device indicates that the operator
must refer to an explanation in the Operating
Instructions to avoid personal injury or
damage to the meter.
This WARNING symbol indicates a potentially
hazardous situation, which if not avoided,
could result
MN25 Version 2.2 Aug 2003 5
Specifications
Function
Range
Resolution Accuracy
200mV
0.1mV
2000mV
1mV
20V
0.01V
200V
0.1V
DC Voltage
(V DC)
600V
1V
± (0.5% reading + 2
digits)
200V
0.1V
AC Voltage
(V AC)
600V
1V
± (1.5% reading + 5 digits
(Frequency 50/60Hz)
200mA
100µA
± (1.2% reading + 2
digits)
DC Current
(A DC)
10A
10mA
± (2.0% reading + 2
digits)
200mA
100µA
± (1.8% reading + 5
digits)
AC Current
(A AC)
10A
10mA
± (3.0% reading + 5
digits)
2
MN25 Version 2.2 Aug 2003 6
Display
3 ½ digit, 2000 count LCD, 0.5” digits
Overrange indication
“1” is displayed
Diode Test
Test current of 1mA maximum, open
circuit voltage 2.8V DC typical
Continuity Check
Audible signal will sound if the
resistance is less than approximately
30Ω
Battery Test current
9V (6mA); 1.5V (100mA)
Input Impedance
>1MΩ
ACV Bandwidth
45Hz to 450Hz
DCA voltage drop
200mV
Polarity
Automatic (no indication for positive
pola
MN25 Version 2.2 Aug 2003 7
Meter Description
1.
LCD Display
2.
Function switch
3.
10A jack
4.
COM jack
5.
Positive jack
Note: Tilt stand, fuse and battery
compartment are on rear of unit.
LCD Display Indicators
•)))
Continuity
Diode test
m
milli ( volts, amps)
k
kilo (ohms)
Ω
ohms
V DC
volts direct current
V AC
volts alternating current
A DC
amps direct current
A AC
amps alternating current
BAT
Low battery and battery test
1
2
5
3 4 5
MN25 Version 2.2 Aug 2003 8
Operation
WARNING: Risk of electrocution. High-voltage circuits, both AC
and DC, are very dangerous and should be measured with great
care.
1. ALWAYS turn the function switch to the OFF position when
the meter is not in use.
2. If “OL” appears in the display during a measurement, the
value exceeds the range you have selected. Change to a
higher range.
NOTE: On some low AC and DC voltage ranges, with the test
leads not connec
MN25 Version 2.2 Aug 2003 9
AC Voltage Measurements
WARNING: Risk of Electrocution. The probe tips may not be
long enough to contact the live parts inside some 240V outlets
for appliances because the contacts are recessed deep in the
outlets. As a result, the reading may show 0 volts when the
outlet actually has voltage on it. Make sure the probe tips are
touching the metal contacts inside the outlet before assuming
that no voltage is present.
CAUTION: D