Title & Document Type: 300A Harmonic Wave Analyzer
Manual Part Number: N/A
Revision Date: 11/30/1955
HP References in this Manual
This manual may contain references to HP or Hewlett-Packard. Please note that
Hewlett-Packard's former test and measurement, semiconductor products and chemical
analysis businesses are now part of Agilent Technologies. We have made no changes to
this manual copy. The HP XXXX referred to in this document is now the Ag
This is a combined Operation, Instruction,
and Service
Manual for all ~ Mode1300A Harmonic Wave Analyzers.
This manual contains complete operation and servicing
instructions for the 300A and may be used in place of
thl~ Instruction and Operating Manual originally supplied
with each ins trument.
Copyright 1956 by Hewlett-Packard Company
The information contained in this booklet
is inte
Agilent Technologies 300A MODEL 300A HARMONIC WAVE ANALYZER
INSTRUCTION & OPERATING MANUAL
TABLE OF CONTENTS
SECTION I
GENERAL
Page
1-1
1-2
GENERAL DESCRIPTION
SPECIFICATIONS
•
•
•
· . . . · . · .· .
III
III
SECTION II
OPERATING INSTRUCTIONS
2 -1
23D VOLT OPERATION. · · · · · · · · · ·
· •
III
2 -2
CONTROLS & TERMINALS · · · · · · · · ·
· •
IV
2 -3
OPERATION. . . . . · · · · · · · · · · •
• · · · · ·
IV
2-4
CALIBRATION ADJUST
INSTRUCTION & OPERATING MANUAL
SECTION I
GENERAL
1-1
GENERAL DESCRIPTION
The Model 300A Harmonic Wave Analyzer is a selective voltmeter designed to
measure the individual components of complex waves.
The selectivity can be
varied by means of selective amplifiers to measure either closely or widely
spaced harmonics.
The instrument cover s the audio spectrum from 30 to
16,000 cps.
It has a wide voltage range so that full scale readings may be ob-
tained
CONTROLS AND TERMINALS
All controls and terminals are fully des cribed on pages 4 and 5 of the r5j:;
No. 300A-2 Service ManuaL
CAUTION
THIS INSTRUMENT IS ACCURATE AT AMBIENT TEMPERATURES OF
APPROXIMATELY 55 TO 95 DEGREES FAHRENHEIT.
OTHER AMBI-
ENT TEMPERATURES MAY NECESSITATE REALIGNMENT OF THE
SELECTIVE AMPLIFIER SYSTEM.
2-3
OPERATION
The operation of the Model 300A is divided into two parts. the calibration
adjustment procedure and the measuremen
METER SE. SITIVITY at 500 (5. a on meter scale) full scale millivolts.
Meter pointer at 2. a therefor e meter actually indicates 200 millivolts.
200 millivolts x 100 (meter multiplier factor) is equal to 20, 000 millivolts
or 20 volts.
2-5
MEASUREMENT PROCEDURE (Contld.)
v
..
£.
Turn the frequency dial to 160 cps (second harmonic of 80 cps) and set the
METER SEl SITIVITY and METER MULTIPLIER controls to obtain a read-
able meter indication.
Use t
Agilent Technologies 300A selectivity or at "30" for maximum selectivity.
Us e mlmmum selectivity
for voltages with a fundamental frequency higher than 100 to 300 cps.
Use
maximum selectivity for voltages with a fundamental frequency lower than
100 to 300 cps.
This system eliminates the necessity of determining whether
an intermediate degree of selectivity offers sufficient attenuation for the case
at hand.
Occasionally, however,
it is desirable to use
Agilent Technologies 300A To remove the instrument from the cabinet it is necessary to unscrew the
eight large Phillips head screws on ~he control panel and slide the instrument
forward out of the cabinet.
In some older instruments, it may be necessary to remove the wire connect-
ing the bottom chassis to the metal plate in the bottom of the cabinet.
This
wire will be found on the rear of the instrument.
4-2
TUBE REPLACEMENT
Refer to p