Text of Agilent Technologies 300A Operating Manual

Title & Document Type: 300A Harmonic Wave Analyzer

Manual Part Number: N/A Revision Date: 11/30/1955 HP References in this Manual This manual may contain references to HP or Hewlett-Packard. Please note that Hewlett-Packard's former test and measurement, semiconductor products and chemical analysis businesses are now part of Agilent Technologies. We have made no changes to this manual copy. The HP XXXX referred to in this document is now the Ag

This is a combined Operation, Instruction,

and Service Manual for all ~ Mode1300A Harmonic Wave Analyzers. This manual contains complete operation and servicing instructions for the 300A and may be used in place of thl~ Instruction and Operating Manual originally supplied with each ins trument. Copyright 1956 by Hewlett-Packard Company The information contained in this booklet is inte

Agilent Technologies 300A MODEL 300A HARMONIC WAVE ANALYZER

INSTRUCTION & OPERATING MANUAL TABLE OF CONTENTS SECTION I GENERAL Page 1-1 1-2 GENERAL DESCRIPTION SPECIFICATIONS • • • · . . . · . · .· . III III SECTION II OPERATING INSTRUCTIONS 2 -1 23D VOLT OPERATION. · · · · · · · · · · · • III 2 -2 CONTROLS & TERMINALS · · · · · · · · · · • IV 2 -3 OPERATION. . . . . · · · · · · · · · · • • · · · · · IV 2-4 CALIBRATION ADJUST

INSTRUCTION & OPERATING MANUAL

SECTION I GENERAL 1-1 GENERAL DESCRIPTION The Model 300A Harmonic Wave Analyzer is a selective voltmeter designed to measure the individual components of complex waves. The selectivity can be varied by means of selective amplifiers to measure either closely or widely spaced harmonics. The instrument cover s the audio spectrum from 30 to 16,000 cps. It has a wide voltage range so that full scale readings may be ob- tained

CONTROLS AND TERMINALS

All controls and terminals are fully des cribed on pages 4 and 5 of the r5j:; No. 300A-2 Service ManuaL CAUTION THIS INSTRUMENT IS ACCURATE AT AMBIENT TEMPERATURES OF APPROXIMATELY 55 TO 95 DEGREES FAHRENHEIT. OTHER AMBI- ENT TEMPERATURES MAY NECESSITATE REALIGNMENT OF THE SELECTIVE AMPLIFIER SYSTEM. 2-3 OPERATION The operation of the Model 300A is divided into two parts. the calibration adjustment procedure and the measuremen

METER SE. SITIVITY at 500 (5. a on meter scale) full scale millivolts.

Meter pointer at 2. a therefor e meter actually indicates 200 millivolts. 200 millivolts x 100 (meter multiplier factor) is equal to 20, 000 millivolts or 20 volts. 2-5 MEASUREMENT PROCEDURE (Contld.) v .. £. Turn the frequency dial to 160 cps (second harmonic of 80 cps) and set the METER SEl SITIVITY and METER MULTIPLIER controls to obtain a read- able meter indication. Use t

Agilent Technologies 300A selectivity or at "30" for maximum selectivity.

Us e mlmmum selectivity for voltages with a fundamental frequency higher than 100 to 300 cps. Use maximum selectivity for voltages with a fundamental frequency lower than 100 to 300 cps. This system eliminates the necessity of determining whether an intermediate degree of selectivity offers sufficient attenuation for the case at hand. Occasionally, however, it is desirable to use

Agilent Technologies 300A To remove the instrument from the cabinet it is necessary to unscrew the

eight large Phillips head screws on ~he control panel and slide the instrument forward out of the cabinet. In some older instruments, it may be necessary to remove the wire connect- ing the bottom chassis to the metal plate in the bottom of the cabinet. This wire will be found on the rear of the instrument. 4-2 TUBE REPLACEMENT Refer to p