LAMBDA DG-4 & 5 PLUS/USB OPERATION MANUAL – REV. 1.01B3 (20190607)

72 Table 5-2. Cable specifications. Type Max. Length Ferrite at Controller End Cabling Parallel DB-25 6 ft (1.829 m) Fair-Rite #0443164-251 (See Note 1) Dielectric separation of circuits. Foil shielding Serial (RS-232) DB-9 10 ft (3.048 m) Fair-Rite #0443167-251 (See Note 1) Connected to metal faceplates of connectors on both ends. (See Note 2.) Note 1: Fair-Rite Products Corp., P.O. Box J, One Commercial Row, Wallkill, NY, 12589 USA Note 2: The supplied 9-pin serial RS-232 cable is a “straight-through” cable (Pin 1 of on ...

LAMBDA DG-4 & 5 PLUS/USB OPERATION MANUAL – REV. 1.01B3 (20190607)

33 Press 1 to activate the option and return to the MAIN MENU or 0 to abort the routine and return to the MAIN MENU. The status of the TURBO-BLANKING mode will be indicated on the first line of the MAIN MENU. 3.7.10 Display the First Four Filter Values Mode [MODE/3] To display the current Input and Output Galvanometer Values for the first four FILTER NUMBERS press the MODE key. The following menu will be displayed: To bring up the values display, press 3. Press ANY key to return to the MAIN MENU. TURBO-BLANKING IS TURNED OFF, ...

LAMBDA DG-4 & 5 PLUS/USB OPERATION MANUAL – REV. 1.01B3 (20190607)

14 on and the lamp is ignited. If you are only configuring the Lambda DG-4 controller or a computer interface, you do not need to switch on the LAMP before turning on the MAIN switch. The following message will be briefly displayed on the front panel when the MAIN switch is turned on: After a few seconds the MAIN MENU will be displayed: At this point, the Controller subsystem is ready for operation. When the unit is first turned on, the FILTER NUMBER display will read “OFF”. This indicates that the filter selecting mirror ...

LAMBDA DG-4 & 5 PLUS/USB OPERATION MANUAL – REV. 1.01B3 (20190607)

ix 5.1.4 Reinstalling the Lamp/Housing Assembly to the System ................................................... 63 APPENDIX A. LIMITED WARRANTY ............................................................................................ 65 APPENDIX B. ACCESSORIES ......................................................................................................... 67 B.1. Liquid Light Guides and Other Accessories ............................................................................... 67 B.2. Microscope Mounting A ...

LAMBDA DG-4 & 5 PLUS/USB OPERATION MANUAL – REV. 1.01B3 (20190607)

55 routine that could be used to monitor the status of the DG-4 (the example assumes that #1 has already been opened as the serial port to which the DG-4 is connected. Listing 4-4. Reading the Lambda DG-4's status. SENDITS: GOSUB SERCLEAR ‘clears receive buffer IF DATANUM < 255 THEN PRINT #1, CHR$(DATANUM); ‘sends ‘command GOSUB SERIN DTIME = 10 RETURN SERIN: D = 0 ‘Check the SERIAL PORT If characters are waiting SERIN1: SerialInput$ = “” ‘blanks the input variable IF LOC(1) > 0 THEN SerialInput$ = INPUT$(1, #1) ...

LAMBDA DG-4 & 5 PLUS/USB OPERATION MANUAL – REV. 1.01B3 (20190607)

47 4.1.3 Command Codes All commands to the Lambda DG-4 correspond to an 8-bit binary value (not all of the possible 256 values are utilized). The binary values that are used can be classified in the following way: Table 4-8. Command code types. Value range (Decimal, hexadecimal, binary) Command Type 0 – 15 00 - 0F 00000000 - 00001111 MOVE ON COMMAND to the corresponding FILTER NUMBER (i.e., immediately after receiving the command). 16 - 31 10 - 1F 00010000 - 00011111 MOVE ON TRIGGER to the corresponding FILTER NUMBER (move ...

LAMBDA DG-4 & 5 PLUS/USB OPERATION MANUAL – REV. 1.01B3 (20190607)

20 When the ENTER key is then pressed, FILTER NUMBER 8 is then displayed correctly. Method 2: The main menu will look like this if the initial FILTER NUMBER = 10 The numeric key 0 is pressed first. The numeric key 8 is pressed next with no further action required. FILTER NUMBER 8 is then displayed correctly. 3.7.4 Move on Sync Pulse Mode [MODE/4] If the Move on Sync Pulse Mode is active, the controller will advance to the next filter in a previously programmed FILTER NUMBER sequence (Ring Buffer) when an internal video synch ...

LAMBDA DG-4 & 5 PLUS/USB OPERATION MANUAL – REV. 1.01B3 (20190607)

24 Figure 3-4. Optical pathway based on filter position. The position of the Output Galvanometer must be coordinated with that of the Input Galvanometer to complete the optical pathway to the Light Guide Port (shown above). The Output Galvanometer angle is also expressed in an arbitrary unit whose value ranges from 0 to 4095. The discrete Output Galvanometer Value that defines the optimal light path for each of the FILTER POSITIONS has been determined at the factory, and can be found on the Optical Alignment Data Sheet. The f ...

Sutter Instrument Lambda DG-4Plus/USB user guide recommended for: DEADML805, MB572, ELED STRIP RGBW, SLR-1750, KWL 224-03.

⚠️ Download Safety

The Work Light Operation manual (98 pages) is completely safe to download (last scan date: 30/05/2026).

Related PDF Instructions for Sutter Instrument Lambda DG-4Plus/USB:

Work Light Equipment: Frequently Requested

Have some Questions about Sutter Instrument Lambda DG-4Plus/USB Device?

Get valuable and actionable advices from other Sutter Instrument Lambda DG-4Plus/USB device users. Describe your problem or question, regarding the using of Lambda DG-4Plus/USB, in as much detail as possible, and other users will be able to answer you.