PIE 830PM
Panel Mounted
Multifunction Process Calibrator
mA • V • pH • TC • Ω • RTD • Freq • Pressure
Operating Instructions
PIE 830PM
Panel Mounted
Multifunction Process Calibrator
mA • V • pH • TC • Ω • RTD • Freq • Pressure
Operating Instructions
Copyright © 2018 All rights reserved • 830PM-9002 - Rev A 19 June 2018
Contents
Basic Operation
Switches & Knobs........................................................................... 2
Connections.................................................................................... 2
MAIN Menus - Functions, Units & Ranges...................................... 3
FEATURE Menu - Stepping & Ramping / Auto Off / Backlight........ 4
Storing EZ-CHECK Outputs & Automatic Stepping........................ 5
Functions and Hookup Diagrams
2-Wire Transmitter Calibration
Simulate sensor while powering 2-Wire Transmitters.................. 6, 7
Universal Isolated Transmitter
Use the 830 to replace a 2-Wire Transmitter...............................8-11
Milliamp
Simulate 2 Wire Transmitters................................................... 12, 13
Source mA, Read mA............................................................. 14, 15
Power/Measure Transmitters......................................................... 16
Using Ground Leak Detection.................................................. 17, 18
Voltage & Millivolt
Source V & mV; Read V & mV................................................ 19, 20
pH
Source pH..................................................................................... 21
Thermocouple
Source T/C & Read T/C Sensors............................................ 22, 23
Resistance
Source Resistance, Read Resistance & Continuity................ 24, 25
RTD
Source RTD & Read RTD Sensors......................................... 26, 27
Frequency
Source KHz, Hz & CPM; Read KHz, Hz & CPM..................... 28, 29
Pressure
Read Pressure............................................................................... 30
Pressure Module Ranges & Specifications................................... 31
Pressure Module........................................................................... 32
Warranty
Accessories, Warranty & Additional Information........................... 33
Specifications
General .................................................................................... 34-36
Thermocouple Ranges & Accuracies.......................................37, 38
RTD Ranges & Accuracies............................................................ 39
Installation
Panel Cutout Dimensions.............................................................. 40
Mounted in Panel.................................................................... 41, 42
Page 1
Practical Instrument Electronics
Tel: 585.872.9350 • Fax: 585.872.2638 •
[email protected] • www.piecal.com
The PIE Model 830PM is more than a multifunction calibrator. It is also a loop
detective that is able to diagnose common problems that other test equipment
just can’t find. Have a flooded junction box or unknown ground faults? Our Loop
Diagnostic technology will detect it. Or use the LoopScope to see at a glance all
the parameters - milliamps, voltage and resistance - in the loop.
Stop throwing away perfectly good transmitters only to find the problem is
somewhere else in the loop. Setup the PIE Model 830PM as an isolated universal
transmitter and turn on the Loop Diagnostics. The display will tell you EXACTLY
the sensor input, the current output and if there is any uncontrolled current in the
loop due to a ground fault, corrosion bridge or moisture. If the control system sees
no problem when the 830PM is acting as a transmitter THEN you can replace the
faulty transmitter. With RTDs the 830PM automatically detects which of the 2, 3 or
4 wires are connected and unbroken quickly alerting you to a sensor problem.
Diagnostic features are covered by US Patent #7,248,058.
Become a troubleshooting technician with Patented Diagnostic Technology -
Available only with PIE Calibrators!
Get more tools in a smaller calibrator
Combines eight single function calibrators
plus a milliamp calibrator with loop supply
plus a loop troubleshooter into a compact panel mounted enclosure!
Milliamp • Voltage • Frequency • pH • Ohms • Thermocouples • RTDs • Pressure
Loop Diagnostics • Transmitter Supply
1
Detect ‘hidden’ loop problems Quickly diagnose troublesome ground faults
& current leakage with patented Loop Diagnostic technology. These problems are
undetectable with other instruments!
2
Check all loop parameters at once with LoopScope™ Simultaneously
displays current, voltage and resistance to let you know the condition of a
live loop. Finds problems with power supplies & loops with too many
loads. Patented by PIE!
3
Automatically detect 2, 3, or 4 wire RTDs Trouble shoot sensor connections
and find broken wires with patented technology. LCD indicates which of the four
wires are connected to an RTD sensor.
4
Swap out faulty transmitters to diagnose control issues Easily setup the 830
as a universal isolated transmitter for Thermocouple, RTD, frequency, millivolt,
resistance or pressure. If the loop is back under control you know the
transmitter was faulty. Turn on the LoopScope to simultaneously see the loop
current, loop voltage and loop loads for troubleshooting.
5
Troubleshoot wiring problems without a multimeter Built in continuity check-
er with ‘beeper’ quickly finds broken wires or shorts in instrumentation wiring.
Also handy for checking operation of relays and controller outputs.