PIE PIE 830PM Manual
Read Online
Read the PIE PIE 830PM Manual online using the built-in PDF viewer.
Text of the manual
Multifunction Process Calibrator
mA • V • pH • TC • Ω • RTD • Freq • Pressure
Operating Instructions
PIE 830PM
Panel Mounted
Multifunction Process Calibrator
mA • V • pH • TC • Ω • RTD • Freq • Pressure
Operating Instructions
Copyright © 2018 All rights reserved • 830PM-9002 - Rev A 19 June 2018
Switches & Knobs........................................................................... 2
Connections.................................................................................... 2
MAIN Menus - Functions, Units & Ranges...................................... 3
FEATURE Menu - Stepping & Ramping / Auto Off / Backlight........ 4
Storing EZ-CHECK Outputs & Automatic Stepping........................ 5
Functions and Hookup Diagrams
2-Wire Transmitter Calibration
Simulate sensor while powering 2-Wire Transmitters.................. 6, 7
Universal Isolated Transmitter
Use the 830 to replace a 2-Wire Transmitter...............................8-11
Milliamp
Simulate 2 Wire Transmitters................................................... 12, 13
Source mA, Read mA............................................................. 14, 15
Power/Measure Transmitters......................................................... 16
Using Ground Leak Detection.................................................. 17, 18
Voltage & Millivolt
Source V & mV; Read V & mV................................................ 19, 20
pH
Source pH..................................................................................... 21
Thermocouple
Source T/C & Read T/C Sensors............................................ 22, 23
Resistance
Source Resistance, Read Resistance & Continuity................ 24, 25
RTD
Source RTD & Read RTD Sensors......................................... 26, 27
Frequency
Source KHz, Hz & CPM; Read KHz, Hz & CPM..................... 28, 29
Pressure
Read Pressure............................................................................... 30
Pressure Module Ranges & Specifications................................... 31
Pressure Module........................................................................... 32
Warranty
Accessories, Warranty & Additional Information........................... 33
Specifications
General .................................................................................... 34-36
Thermocouple Ranges & Accuracies.......................................37, 38
RTD Ranges & Accuracies............................................................ 39
Installation
Panel Cutout Dimensions.............................................................. 40
Mounted in Panel.................................................................... 41, 42
Practical Instrument Electronics
Tel: 585.872.9350 • Fax: 585.872.2638 • [email protected] • www.piecal.com
The PIE Model 830PM is more than a multifunction calibrator. It is also a loop
detective that is able to diagnose common problems that other test equipment
just can’t find. Have a flooded junction box or unknown ground faults? Our Loop
Diagnostic technology will detect it. Or use the LoopScope to see at a glance all
the parameters - milliamps, voltage and resistance - in the loop.
Stop throwing away perfectly good transmitters only to find the problem is
somewhere else in the loop. Setup the PIE Model 830PM as an isolated universal
transmitter and turn on the Loop Diagnostics. The display will tell you EXACTLY
the sensor input, the current output and if there is any uncontrolled current in the
loop due to a ground fault, corrosion bridge or moisture. If the control system sees
no problem when the 830PM is acting as a transmitter THEN you can replace the
faulty transmitter. With RTDs the 830PM automatically detects which of the 2, 3 or
4 wires are connected and unbroken quickly alerting you to a sensor problem.
Diagnostic features are covered by US Patent #7,248,058.
Become a troubleshooting technician with Patented Diagnostic Technology -
Available only with PIE Calibrators!
Get more tools in a smaller calibrator
Combines eight single function calibrators
plus a milliamp calibrator with loop supply
plus a loop troubleshooter into a compact panel mounted enclosure!
Milliamp • Voltage • Frequency • pH • Ohms • Thermocouples • RTDs • Pressure
Loop Diagnostics • Transmitter Supply
1
Detect ‘hidden’ loop problems Quickly diagnose troublesome ground faults
& current leakage with patented Loop Diagnostic technology. These problems are
undetectable with other instruments!
2
Check all loop parameters at once with LoopScope™ Simultaneously
displays current, voltage and resistance to let you know the condition of a
live loop. Finds problems with power supplies & loops with too many
loads. Patented by PIE!
3
Automatically detect 2, 3, or 4 wire RTDs Trouble shoot sensor connections
and find broken wires with patented technology. LCD indicates which of the four
wires are connected to an RTD sensor.
4
Swap out faulty transmitters to diagnose control issues Easily setup the 830
as a universal isolated transmitter for Thermocouple, RTD, frequency, millivolt,
resistance or pressure. If the loop is back under control you know the
transmitter was faulty. Turn on the LoopScope to simultaneously see the loop
current, loop voltage and loop loads for troubleshooting.
5
Troubleshoot wiring problems without a multimeter Built in continuity check-
er with ‘beeper’ quickly finds broken wires or shorts in instrumentation wiring.
Also handy for checking operation of relays and controller outputs.
Practical Instrument Electronics
Tel: 585.872.9350 • Fax: 585.872.2638 • [email protected] • www.piecal.com
q EZ-CHECK™ SWITCH
SOURCE: Instantly output two preset settings by
moving the EZ-CHECK™ switch to the “LO” position
or “HI” position. For fast three point checks select
the “SET” position. The 830PM will remember
the last “SET” value, even with the power off.
These values can easily be changed to suit the
calibration requirements. The values stored in the HI
and LO positions are also used for Auto Stepping.
READ: Slide the switch to the SET position. The
830PM will display the current input value. Slide the
switch to HI and the highest value measured since
turn-on or reset will be displayed; slide the switch to
LO and the lowest value measured since turn-on or
reset will be displayed.
w SOURCE/OFF/READ Switch
Select “SOURCE” to output mA, V, pH, T/C, Ω,
RTD or Hz.
Select “READ” to read mA, V, T/C, Ω, RTD, Hz or
pressure.
e EZ-DIAL™ KNOB
SOURCE: Turn the knob to adjust the output level.
Turn clockwise to increase the output, counter
clockwise to decrease the output in one least
significant digit step at a time. Push down and turn
the EZ-DIAL knob for faster dialing.
Press and hold the knob for two seconds to store
desired EZ-Check™ HI/LO points in SOURCE
mode. Continue to press and hold the knob for two
more seconds to start the automatic ramping.
READ: Press and hold to transfer the current
temperature into the EZ-Check™ MIN/MAX
points. This clears the MIN/MAX readings
which will update as the input value changes.
Double Click the EZ-DIAL knob to change the
function of the calibrator and to select ranges,
units and other user settings.
Basic Operation
5
MAX
READ
MIN
READ
6
-
+
-
3
2
-2
OFF
HI
SET
LO
SOURCE
Hz SYNC
EZ-DIAL
Double Click
Menu
Push & Turn for
Fast Dialing
Push & Hold to
Store/Step
piecal.com
EZ-CHECK™
PRACTICAL INSTRUMENT ELECTRONICS
MODEL 830PM
Process Calibrator
Loop Diagnostics · Transmitter Supply
mA · V ∙ pH · TC · Ω · RTD · Hz · Pressure
Pressure
Module
All mA∙ Read V
TC
Source & Read mV
Hz
Source V
pH
RTD • Ohms
OVERLOAD
4
1
+
+1
5 6
+ -
HI
mA
20.000
SOURCE
Test leads with shielded or non shielded banana
plugs may be plugged into the protected banana
jacks labeled 1 through 6.
PIE external pressure modules may be plugged
into the jack marked Pressure Module
Connections
Simulating or reading thermocouples requires
the use of thermocouple or extension grade
thermocouple wire terminated with a miniature
thermocouple plug.
Practical Instrument Electronics
Tel: 585.872.9350 • Fax: 585.872.2638 • [email protected] • www.piecal.com
Configuration
Double Click Menus - Main Page
Move w POWER SWITCH to “SOURCE” or
“READ”.
Turn the e DIAL KNOB to move through the two
pages of menus. Shown are the MAIN menus for
each function. Turn the e knob to scroll thru the
menus and press the e knob to select.
Available choices are shown in grey.
Source mA & Simulate
2 Wire Transmitters
Read mA, Power/Measure Transmitters
& Leak Detect
Source V
Read V & mV
Source pH
Source & Read Frequency
DOUBLE CLICK
EZ-DIAL KNOB
FOR CONFIGURATION
V#.##
Double Click Menus - Main Page
Source & Read Thermocouples
Source RTD
Read RTD
Source Ohms
Read Ohms
Read Pressure
MAIN
> EXIT
FUNCTION
mA
MODE
SOURCE 2W SIM
UNITS
mA %
HART 250Ω
ON OFF
> EXIT
FUNCTION
mA
MODE
READ PWR MEAS
UNITS
mA %
HART 250Ω
ON OFF
> EXIT
FUNCTION V
RANGE 10V 100mV 1V
> EXIT
FUNCTION V
RANGE 10V 1V 60V 100mV
> EXIT
FUNCTION pH
> EXIT
FUNCTION FREQ
RANGE 20KHZ 10000HZ 1000HZ 2000CPM
MAIN
> EXIT
FUNCTION T/C
UNITS
°C °F
T/C TYPE
J K E T R S B N L U G C D P
COLD JUNC ON OFF
>EXIT
FUNCTION RTD
UNITS °C °F
RTD Pt 100 a=3850 [*RTD Types - See Read RTD]
>EXIT
FUNCTION RTD
UNITS
°C °F
RTD Pt 100 α=3850, α=3902, α=3916, α=3926
Pt 1000 α=3850;
Cu 10 α=4274, Cu 50 α=4280
Ni 120 α=6720, Ni 110 α=5801
> EXIT
FUNCTION OHMS
RANGE 400Ω 4000Ω
> EXIT
FUNCTION OHMS
RANGE 400Ω 4000Ω CONT
> EXIT
FUNCTION PRESSURE
UNITS psi inH2O* ftH2O* mmH2O*
cmH2O* mH2O* inHg mHg cmHg mmHg
torr kg/cm2 kg/m2 hPa kPa MPa bar mbar
atm oz/in2 lb/ft2
*Engineering unit available at 4°C, 20°C and
60°F.
Practical Instrument Electronics
Tel: 585.872.9350 • Fax: 585.872.2638 • [email protected] • www.piecal.com
Operating Instructions
Double Click Menu - STEPPING, AUTO OFF & BACKLIGHT
To change the Automatic Stepping settings
Double click the e DIAL KNOB at any time the
unit is on and the following typical display (will
be different for each FUNCTION) will appear for
15 seconds:
Turn the e DIAL KNOB to move to the second
menu (FEATURES) page.
Turn the e DIAL KNOB to move through the
menu. Press the e DIAL KNOB to toggle
between OFF and ON or to change the STEPS/
RAMP and the STEP/RAMP TIME settings.
These settings are remembered even with the
power off.
EXIT MENU - exits this menu immediately and
saves any changes. Menu will automatically exit
after 15 seconds of inactivity.
AUTO OFF - If AUTO OFF is ON, the unit will
turn off after 30 minutes of inactivity to save
battery life. If AUTO OFF is OFF the unit will stay
on until the POWER SWITCH is moved to the off
position.
BACKLIGHT - If BACKLIGHT is ON the
backlight will light all the time the unit is powered
up. For maximum battery life turn the backlight
off when using the calibrator in areas with
enough ambient light to read the display.
FEATURES
> EXIT
(2/2)
AUTO OFF
ON
BACKLIGHT
ON
STEPS/RAMP 3
STEP/RAMP TIME 5
MAIN
> EXIT
(1/2)
FUNCTION
mA
MODE
SOURCE
UNITS
mA
HART 250Ω
ON
STEPS/RAMP - pressing the knob will cycle
through 2, 3, 5, 11 and RAMP. The endpoints of
the steps or ramp are based on the values
stored in the HI and LO EZ-CHECK outputs.
2 steps will automatically switch between the
values stored in the HI & LO EZ-CHECK (0 &
100%).
3 steps between the HI, Midpoint and LO
EZ-CHECK (0, 50 & 100%).
5 steps between the HI and LO EZ-CHECK
in 25% increments (0, 25, 50, 75 & 100%).
11 steps between the HI and LO EZ-CHECK
in 10% increments (0, 10, 20...80, 90 &100%).
RAMP continuously ramps up and down
between the HI and LO EZ-CHECK outputs.
STEP/RAMP TIME - pressing the knob will cycle
through 5, 6, 7, 8, 9, 10, 15, 20, 25, 30 and 60
seconds.
To start the Automatic Stepping
Start automatic stepping or ramping by placing
the EZ-CHECK Switch into the HI or LO position
then press and hold the e DIAL KNOB for 6
seconds (the word STORE will appear on the
display after 3 seconds and continue to press
the EZ-DIAL KNOB) until the word STEP
appears on the display. The word STEP will
appear on the display anytime the selected
automatic function is running. Stop the stepping
or ramping by again pressing and holding the e
DIAL KNOB for 3 seconds.
Practical Instrument Electronics
Tel: 585.872.9350 • Fax: 585.872.2638 • [email protected] • www.piecal.com
To change the Automatic Stepping settings
Double click the e DIAL KNOB at any time the unit is on and the menu will appear for 15 seconds.
Turn the e DIAL KNOB to move through down
to the third (FEATURES) menu. Press the e
DIAL KNOB to toggle between OFF and ON or
to change the STEPS and the STEP TIME
settings. These settings are remembered even
with the power off.
EXIT MENU - exits this menu immediately and saves any changes. Menu will automatically exit after
15 seconds of inactivity.
STEPS - pressing the knob will cycle through 2, 3, 5 and 11 then reverse direction. The endpoints of
the steps are based on the values stored in the HI and LO EZ-CHECK outputs.
2 steps will automatically switch between the values stored in the HI & LO EZ-CHECK (0 &
100%).
3 steps between the HI, Midpoint and LO EZ-CHECK (0, 50 & 100%).
5 steps between the HI and LO EZ-CHECK in 25% increments (0, 25, 50, 75 & 100%).
11 steps between the HI and LO EZ-CHECK in 10% increments (0, 10, 20...80, 90 &100%).
RAMP continuously between the HI and LO EZ-CHECK.
STEP TIME - pressing the knob will cycle through 5, 6, 7, 8, 9, 10, 15, 20, 25, 30 and 60 seconds.
To start the Automatic Stepping
Start automatic stepping or ramping by placing the EZ-CHECK Switch into the HI or LO position
then press and hold the e DIAL KNOB for 6 seconds (the word STORE will appear on the display
after 3 seconds and continue to press the DIAL KNOB) until the word STEPPING appears on the
display. The word STEPPING will appear on the display anytime the selected automatic function is
running. Stop the stepping by again pressing and holding the e DIAL KNOB for 3 seconds.
Automatic Stepping
Storing EZ-CHECK Outputs
STORING HI and LO EZ-CHECK Outputs
Choose this function to provide a simulated thermocouple signal into controllers, temperature
transmitters, indicators or any other input device that measure thermocouple sensors.
1) Store your high (SPAN) output temperature by moving the EZ-CHECK switch to the HI position
and turn the e EZ-Dial knob until the desired temperature is on the display. Press and hold the
EZ-Dial knob until STORED appears to store the value. Release the EZ-Dial knob.
2) Store your low (ZERO) output temperature by moving the EZ-CHECK switch to the LO position
and turn the e EZ-Dial knob until the desired temperature is on the display. Press and hold the
EZ-Dial knob until STORED appears to store the value. Release the EZ-Dial knob.
3) Instantly output your SPAN and ZERO temperature outputs by moving the EZ-CHECK switch
between HI and LO. You may also select any third temperature output (such as mid-range) using
the SET position on the EZ-CHECK switch.
FEATURES
> EXIT (3/3)
AUTO OFF
ON OFF
BACKLIGHT
ON OFF
STEPS/RAMP 2 3 5 11 RAMP
STEP/RAMP TIME 5 6 7 8 9 10 15 20 25 30 60
Practical Instrument Electronics
Tel: 585.872.9350 • Fax: 585.872.2638 • [email protected] • www.piecal.com
Operating Instructions
Calibrate a 2-Wire Transmitter by sourcing the input while monitoring the output.
Works with SOURCE pH, T/C, DC V, OHMS, RTD, FREQ and READ PRESSURE.
Move the power switch w to READ and Double click the e DIAL KNOB and the MAIN menu for the
function in use will appear for 15 seconds:
Turn the e DIAL KNOB to move to the second (or third) menu page so the word mA DISPLAY
appears at the top of the menu.
Turn the e DIAL KNOB to move through the menu. Press the e DIAL KNOB to toggle between OFF
and ON or to change the MODE setting.
HART 250Ω - turn on the 250Ω resistor if you are powering up a HART transmitter. This provides
the loop load required for HART communication.
EXIT MENU - exits this menu immediately and saves any changes. Menu will automatically exit after
15 seconds of inactivity.
MAIN
> EXIT
FUNCTION
T/C
UNITS
°C
T/C TYPE
K
COLD JUNC
ON
mA DISPLAY
> EXIT
MODE
OFF READ PWRM READ% PWRM%
HART 250Ω
OFF ON
LEAK DETECT OFF
MODE - pressing the knob will cycle through READ, PWRM, READ% ,PWRM% and OFF.
READ turns on the mA display and indicates current passing through the loop proportional to the
input of the transmitter which is controlled by the output of the 830PM. Choose READ% if your
would like the mA display in percent of 4-20 milliamps.
PWRM is POWER MEASURE which uses the internal loop supply of the 830PM to power up the
transmitter while indicating the current passing through the loop proportional to the input of
transmitter which is controlled by the output of the 830PM. Choose PWRM% if your would like the
mA display in percent of 4-20 milliamps.
If PWRM or PWRM% is selected an additional menu selection of LEAK DETECT will appear.
When LEAK DETECT is turned on the 830PM will display LEAK on the display if there is more
than 0.015 mA of uncontrolled current in the loop. This may be due to a faulty transmitter, corrosion
causing a bridge to ground or moisture present at some connection point. When LEAK DETECT
tests a loop with leakage the loop mA signal will be affected.
Practical Instrument Electronics
Tel: 585.872.9350 • Fax: 585.872.2638 • [email protected] • www.piecal.com
Operating Instructions
PWRM LEAK
18.300 mA
830.0 °c
TYPE K
(Enlarged Display of the 830PM
detecting a current leakage in the loop)
Calibrate a 2-Wire Transmitter by sourcing the input while monitoring the output.
5
MAX
READ
MIN
READ
6
-
+
-
3
2
-2
OFF
HI
SET
LO
SOURCE
Hz SYNC
EZ-DIAL
Double Click
Menu
Push & Turn for
Fast Dialing
Push & Hold to
Store/Step
piecal.com
EZ-CHECK™
PRACTICAL INSTRUMENT ELECTRONICS
MODEL 830PM
Process Calibrator
Loop Diagnostics · Transmitter Supply
mA · V ∙ pH · TC · Ω · RTD · Hz · Pressure
Pressure
Module
All mA∙ Read V
TC
Source & Read mV
Hz
Source V
pH
RTD • Ohms
OVERLOAD
4
1
+
+1
Typical
2-Wire
Transmitter
-OUT+
REF
+ IN -
5 6
+ -
°C
Type K
830.0
PWRM
18.300 mA
Thermocouple
Wire
Practical Instrument Electronics
Tel: 585.872.9350 • Fax: 585.872.2638 • [email protected] • www.piecal.com
Operating Instructions
Universal Isolated Transmitter
Swap out a transmitter to diagnose control issues.
The 830PM acts as an isolated universal T/C, mV, OHMS, RTD, FREQ and
PRESSURE transmitter. Choose this function to temporarily replace a transmitter
when you suspect the transmitter is faulty or to diagnose the parameters of the
loop.
Move the power switch w to READ and Double click the e DIAL KNOB and the
MAIN menu for the function in use will appear. Turn the e DIAL KNOB until the
pointer is at FUNCTION and press the e KNOB until the desired FUNCTION
appears.
Turn the e DIAL KNOB to move through the menu. Press the e DIAL KNOB to
toggle the units, range, T/C Type or RTD curve.
Turn the e DIAL KNOB to move to the second (or third) menu page so the word
XXX XMTR appears at the top of the menu.
Turn the e DIAL KNOB to move through the menu. Press the e DIAL KNOB to
toggle between DOWN and UP, OFF and ON, or to change the MODE setting.
MAIN
> EXIT
FUNCTION
T/C
UNITS
°C
T/C TYPE
K
COLD JUNC
ON
MAIN
> EXIT
FUNCTION
RTD
UNITS
°C
RTD Pt 100 a=3850
MAIN
>EXIT
FUNCTION
V
RANGE
100mV
MAIN
> EXIT
FUNCTION
FREQ
RANGE
20KHZ
MAIN
>EXIT
FUNCTION OHMS
RANGE 400Ω
MAIN
>EXIT
FUNCTION PRESSURE
UNITS
psi
XXX XMTR
> EXIT
MODE OFF LINEAR NONLINEAR SQ ROOT
BURNOUT DOWN UP
LOOPSCOPE OFF ON
Practical Instrument Electronics
Tel: 585.872.9350 • Fax: 585.872.2638 • [email protected] • www.piecal.com
Operating Instructions
MODE - pressing the knob will cycle through OFF, LINEAR, NONLINEAR. When
setup to read pressure SQ ROOT replaces NONLINEAR.
LINEAR turns on the mA display and regulates the loop current linear with the
input signal. For thermocouples & RTDs this is linear relative to the temperature
between zero and span of the sensor input. This mimics the operation of a
digital temperature transmitter or other linearizing transmitters.
NONLINEAR (applies only to thermocouples and RTDs) turns on the mA
display and regulates the loop current linear with base units (millivolts or
ohms) of zero and span of the sensor input. This mimics the operation of an
analog temperature transmitter without linearizing.
SQ ROOT turns on the mA display and regulates the loop current with square
root extraction of zero and span of the pressure module input. This mimics the
operation of a differential pressure transmitter setup for square root extraction.
BURNOUT selects the failure mode of the calibrator when the input signal is lost or
beyond the input scale. Down sets the output to 1.0 mA and UP to 21.5 mA when
sensor failure is detected.
LOOPSCOPE turns on the Loop Diagnostic display which simultaneously
indicates the current, voltage and resistance of the loop. When enabled the
LOOPSCOPE also indicates errors by flashing the LED and writing diagnostic
messages on the display. Messages include LOW LOOP VOLT, HIGH LOOP
VOLT and HIGH LOOP IMPED. With LOOPSCOPE on the 830PM occasionally
performs a test on the loop. During this short test the loop mA signal will be
affected.
Practical Instrument Electronics
Tel: 585.872.9350 • Fax: 585.872.2638 • [email protected] • www.piecal.com
Operating Instructions
Setting up the 830PM as a transmitter
Configure your 830PM to the same Span (URV) and Zero (LRV) by storing Span
& Zero setting with the q EZ-CHECK switch.
1) Store your SPAN input by moving the q EZ-CHECK switch to the MAX posi-
tion and turning the e EZ-Dial knob until the desired output value is on the
display. Press and hold the EZ-Dial knob until STORED appears to store the
value. Release the EZ-Dial knob.
2) Store your ZERO input by moving the EZ-CHECK switch to the MIN position
and turning the EZ-Dial knob until the desired output value is on the display.
Press and hold the EZ-Dial knob until STORED appears to store the value.
Release the EZ-Dial knob.
3) Move the EZ-CHECK switch to the READ position. The 830PM constantly
measures the input from the process sensor and regulates the current in the
4-20 mA loop.
Once the 830PM is setup and connected to the process sensor and the 4-20 mA
loop it acts just like a calibrated isolated transmitter and will display the input signal
from the sensor and the regulated 4-20 mA output signal. XMTR is shown on the
display along with L, N or S for Linear, Nonlinear or Square Root respectively.
12.000 mA
SUPPLY 23.4 V
LOAD 987 Ω
500.0 °c
TYPE K
XMTRL
12.000 mA
500.0 °c
TYPE K
830PM as a Transmitter
830PM with LOOPSCOPE
For full loop diagnostics turn on LOOPSCOPE which indicates the current,
SUPPLY (voltage) and LOAD (resistance) of the loop. It also indicates errors by
flashing the LED and writing diagnostic messages including LOW LOOP VOLT,
HIGH LOOP VOLT and HIGH LOOP IMPED.
Practical Instrument Electronics
Tel: 585.872.9350 • Fax: 585.872.2638 • [email protected] • www.piecal.com
Operating Instructions
Connecting the 830PM in place of a transmitter
Connect the 830PM in place of the transmitter. The 4-20 mA loop connects to
the mA jacks of the 830PM and the sensor (or pressure module) connects to the
other jacks of the 830PM.
5
MAX
READ
MIN
READ
6
-
+
-
3
2
-2
OFF
HI
SET
LO
SOURCE
Hz SYNC
EZ-DIAL
Double Click
Menu
Push & Turn for
Fast Dialing
Push & Hold to
Store/Step
piecal.com
EZ-CHECK™
PRACTICAL INSTRUMENT ELECTRONICS
MODEL 830PM
Process Calibrator
Loop Diagnostics · Transmitter Supply
mA · V ∙ pH · TC · Ω · RTD · Hz · Pressure
Pressure
Module
All mA∙ Read V
TC
Source & Read mV
Hz
Source V
pH
RTD • Ohms
OVERLOAD
4
1
+
+1
Receiver
Powers External
2-Wire Transmitter
Power Supply
Typical 12 to 24 V DC
(2 to 100 V for 820PM)
Typical
2-Wire
Transmitter
- IN +
REF
+OUT-
500.0°C
12.000mA
27.4 V
830 Ω
5 6
+ -
Note: With LOOPSCOPE on the 830PM occasionally performs a test on the
loop. During this short test the loop mA signal will be affected.
Practical Instrument Electronics
Tel: 585.872.9350 • Fax: 585.872.2638 • [email protected] • www.piecal.com
Operating Instructions
2 Wire SIM mA, 2 Wire SIM % (Percent of 4 to 20 mA)
Choose this function to simulate a 2 Wire Transmitter output from 0.000 to
24.000 milliamps. Operates in loops with power supply voltages from 2 to 60
VDC.
Move the power switch w to SOURCE then Double Click the EZ-DIAL knob to
get into the Menu. Turn the knob to scroll through the settings and press the knob
to make your selection. Select mA for the FUNCTION and 2W SIM for the
MODE. Choose either mA or % and whether you need the 250Ω HART resistor
active in the loop.
If you would also like to see the condition of the loop turn the e DIAL KNOB to
move to the second (or third) menu page so the word mA DIAGNOSTIC
appears at the top of the menu.
LOOPSCOPE turns on the Loop Diagnostic display which simultaneously
indicates the current, voltage and resistance of the loop. When enabled the
LOOPSCOPE also indicates errors by flashing the LED and writing diagnostic
messages on the display. Messages include LOW LOOP VOLT, HIGH LOOP
VOLT and HIGH LOOP IMPED. With LOOPSCOPE on the 830PM occasionally
performs a test on the loop. During this short test the loop mA signal will be
affected.
mA DIAGNOSTIC
> EXIT
LOOPSCOPE OFF ON
2-WIRE
SUPPLY 25.4 V
LOAD 485 Ω
12.000 mA
2-WIRE
12.000 mA
830PM with LOOPSCOPE
830PM Simulating a Transmitter
Practical Instrument Electronics
Tel: 585.872.9350 • Fax: 585.872.2638 • [email protected] • www.piecal.com
Operating Instructions
2 Wire SIM mA, 2 Wire SIM % (Percent of 4 to 20 mA)
Connect the output leads of the 830PM to the inputs of the device being
calibrated, making sure to check polarity. Red lead from jack (5) to the plus (+)
input and black lead from jack (6) to the minus (-) input. Open loops and signals
above the maximum scale are limited by protection circuitry with “ERROR” or
“OVER RANGE” flashed on the display and the red OVERLOAD LED lit.
Instantly output your SPAN and ZERO output settings by moving the EZ-CHECK
switch between HI and LO (defaults to 20 & 4 mA). You may also select any third
output setting (such as mid-range) using the SET position on the EZ-CHECK
switch. The output is adjusted in 0.001 mA (0.01%) increments by turning the
knob e. Press and turn the knob for faster dialing with 0.100 mA (1.00%)
increments.
Note: With LOOPSCOPE on the 830PM occasionally performs a test on the
loop. During this short test the loop mA signal will be affected.
5
MAX
READ
MIN
READ
6
-
+
-
3
2
-2
OFF
HI
SET
LO
SOURCE
Hz SYNC
EZ-DIAL
Double Click
Menu
Push & Turn for
Fast Dialing
Push & Hold to
Store/Step
piecal.com
EZ-CHECK™
PRACTICAL INSTRUMENT ELECTRONICS
MODEL 830PM
Process Calibrator
Loop Diagnostics · Transmitter Supply
mA · V ∙ pH · TC · Ω · RTD · Hz · Pressure
Pressure
Module
All mA∙ Read V
TC
Source & Read mV
Hz
Source V
pH
RTD • Ohms
OVERLOAD
4
1
+
+1
Receiver
Powers External
2-Wire Transmitter
Power Supply
Typical 12 to 24 V DC
(2 to 100 V for 820PM)
Typical
2-Wire
Transmitter
- IN +
REF
+OUT-
5 6
+ -
%
mA
28.20
PWRM
Practical Instrument Electronics
Tel: 585.872.9350 • Fax: 585.872.2638 • [email protected] • www.piecal.com
Operating Instructions
SOURCE mA / SOURCE % (Percent of 4 to 20 mA)
Choose this function to provide an output from 0.000 to 24.000 milliamps or
-25.00 to 125.00%. The compliance voltage is a nominal 24 VDC to provide the
driving power to your milliamp receivers.
Move the power switch w to SOURCE then Double Click the EZ-DIAL knob to
get into the Double Click Menu. Turn the knob to scroll through the settings and
press the knob to make your selection. Select mA for the FUNCTION and
SOURCE for the MODE. Choose either mA or % and whether you need the
250Ω HART resistor active in the loop.
Connect the output leads of the 830PM to the inputs of the device being
calibrated, making sure to check polarity. Red lead from jack (5) to the plus (+)
input and black lead from jack (6) to the minus (-) input. Open loops and signals
above the maximum scale are limited by protection circuitry with “ERROR” or
“OVER RANGE” flashed on the display and the red OVERLOAD LED lit.
Instantly output your SPAN and ZERO output settings by moving the EZ-CHECK
switch between HI and LO (defaults to 20 & 4 mA). You may also select any third
output setting (such as mid-range) using the SET position on the EZ-CHECK
switch. The output is adjusted in 0.001 mA (0.01%) increments by turning the
knob e. Press and turn the knob for faster dialing with 0.100 mA (1.00%)
increments.
5
MAX
READ
MIN
READ
6
-
+
-
3
2
-2
OFF
HI
SET
LO
SOURCE
Hz SYNC
EZ-DIAL
Double Click
Menu
Push & Turn for
Fast Dialing
Push & Hold to
Store/Step
piecal.com
EZ-CHECK™
PRACTICAL INSTRUMENT ELECTRONICS
MODEL 830PM
Process Calibrator
Loop Diagnostics · Transmitter Supply
mA · V ∙ pH · TC · Ω · RTD · Hz · Pressure
Pressure
Module
All mA∙ Read V
TC
Source & Read mV
Hz
Source V
pH
RTD • Ohms
OVERLOAD
4
1
+
+1
Milliamp Receiver Input
Controller
Transmitter
Computer
Logger, I/P, DCS
5 6
+ -
HI
mA
20.000
SOURCE
Practical Instrument Electronics
Tel: 585.872.9350 • Fax: 585.872.2638 • [email protected] • www.piecal.com
Operating Instructions
READ mA, READ % (Percent of 4 to 20 mA)
Choose this function to measure from from 0.000 to 24.000 milliamps or -25.00
to 125.00%.
Move the power switch w to READ then Double Click the EZ-DIAL knob to get
into the Double Click Menu. Turn the knob e to scroll through the settings and
press the knob to make your selection. Select mA for the FUNCTION and READ
for the MODE. Choose either mA or % and whether you need the 250Ω HART
resistor active in the loop.
Connect the red input lead from jack (5) of the 830PM to the more positive point
of the break and the black input lead from jack (6) to the more negative point.
Signals below 0 mA or open circuits are indicated by 0.000 mA (-25.00%) on the
display. Signals above 24 mA are current limited by protection circuitry with
“OVER RANGE” on the display and the red OVERLOAD LED lit.
The 830PM measures the input signal and constantly updates the display with
the current reading. Move the EZ-CHECK switch q to MAX to see the highest
reading and to MIN to see the lowest reading. Press and hold the knob e to clear
the MAX and MIN readings.
5
MAX
READ
MIN
READ
6
-
+
-
3
2
-2
OFF
HI
SET
LO
SOURCE
Hz SYNC
EZ-DIAL
Double Click
Menu
Push & Turn for
Fast Dialing
Push & Hold to
Store/Step
piecal.com
EZ-CHECK™
PRACTICAL INSTRUMENT ELECTRONICS
MODEL 830PM
Process Calibrator
Loop Diagnostics · Transmitter Supply
mA · V ∙ pH · TC · Ω · RTD · Hz · Pressure
Pressure
Module
All mA∙ Read V
TC
Source & Read mV
Hz
Source V
pH
RTD • Ohms
OVERLOAD
4
1
+
+1
Milliamp Output Signal
Controller
Transmitter
P/I
DCS
5 6
+ -
mA
8.301
READ
Practical Instrument Electronics
Tel: 585.872.9350 • Fax: 585.872.2638 • [email protected] • www.piecal.com
Operating Instructions
Power/Measure mA, Power/Measure % (Percent of 4 to 20 mA)
Choose this function to simultaneously supply power to a 2 Wire Transmitter
while displaying the 4 to 20 mA output of the transmitter.
Move the power switch w to READ then Double Click the EZ-DIAL knob to get
into the Double Click Menu. Turn the knob e to scroll through the settings and
press the knob to make your selection. Select mA for the FUNCTION and PWR
MEAS for the MODE. Choose either mA or % and whether you need the 250Ω
HART resistor active in the loop.
Disconnect one or both input wires from the device to be calibrated. Connect the
red source lead of the 830PM from jack (5) to the plus (+) input of the device and
the black source lead from jack (6) to the minus (-).
The 830PM supplies a nominal 24 volts DC at 24 mA to the 2 Wire Transmitter.
The current passed by the transmitter will be accurately displayed by the 830PM.
Calibrate the transmitter in the usual manner and disconnect the 830PM. Signals
above 24 mA are current limited by protection circuitry with “OVER RANGE” on
the display and the red OVERLOAD LED lit.
5
MAX
READ
MIN
READ
6
-
+
-
3
2
-2
OFF
HI
SET
LO
SOURCE
Hz SYNC
EZ-DIAL
Double Click
Menu
Push & Turn for
Fast Dialing
Push & Hold to
Store/Step
piecal.com
EZ-CHECK™
PRACTICAL INSTRUMENT ELECTRONICS
MODEL 830PM
Process Calibrator
Loop Diagnostics · Transmitter Supply
mA · V ∙ pH · TC · Ω · RTD · Hz · Pressure
Pressure
Module
All mA∙ Read V
TC
Source & Read mV
Hz
Source V
pH
RTD • Ohms
OVERLOAD
4
1
+
+1
Transmitter Input
Sensor
Process Signal
Simulated Input
Typical
2-Wire
Transmitter
- IN +
REF
+OUT-
5 6
+ -
%
mA
28.30
PWRM
Practical Instrument Electronics
Tel: 585.872.9350 • Fax: 585.872.2638 • [email protected] • www.piecal.com
Operating Instructions
Using Ground Leak Detection
mA OUT, % OUT (Percent of 4 to 20 mA)
Find current leaks in loops caused by ground faults, moisture or corrosion. The
830PM simultaneously supplies power to a 2 Wire Transmitter (or loop with a
transmitter) while displaying the 4 to 20 mA output and the amount of current
leaking in the loop.
1) Move the power switch w to READ then Double Click the EZ-DIAL knob to
get into the Menu. Turn the knob e to scroll through the settings and press the
knob to make your selection. Select mA for the FUNCTION and PWR MEAS
for the MODE. Choose either mA or %.
2) Turn the knob e until the following menu appears.
3) Turn the knob e to scroll through the settings and press the knob to make
your selection. Turn on the LEAK DETECT.
4) Connect the red source lead from the mA (+) jack of the 830PM to the plus
(+) input of the device and the black source lead from the mA (-) to the minus
(-).
The 830PM supplies a nominal 24 volts DC at 24 mA to the 2 Wire Transmitter
or loop. The current passed by the transmitter will be accurately displayed by the
830PM along with an indication of leakage current at the top of the display. If
there is an uncontrolled loop, a transmitter with upscale burnout and bad or
missing sensor or a short the display shows "OVER RANGE”
mA DIAGNOSTIC
> EXIT
LEAK DETECT OFF ON
Practical Instrument Electronics
Tel: 585.872.9350 • Fax: 585.872.2638 • [email protected] • www.piecal.com
Operating Instructions
Using Ground Leak Detection
Typical
2-Wire
Transmitter
+ IN -
REF
+OUT-
Transmitter Input
Sensor
Process Signal
Simulated Input
12.280
PROCESS
INDICATOR
12.280mA
PWRM LEAK
5 6
+ -
00.28 mA
Note: Many installed transmitters will normally
indicate 0.01 to 0.02 mA leakage without significant
control problem. Unstable readings may indicate
loose connections or the presence of moisture.
Typical Error Conditions
The 830PM is supplying the
loop voltage. There is a control
loop error. This may be a
transmitter (set for upscale
burnout) with a bad or missing
sensor, or a short in the loop.
The 830PM is supplying the loop
voltage. A calibrated transmitter is
limiting the loop current to 12.00
mA. An additional 0.28 mA is not
controlled by the transmitter and
is leaking somewhere in the loop.
PWRM LEAK
00.28 mA
12.280 mA
PWRM LEAK
mA
OVER RANGE
mA
Practical Instrument Electronics
Tel: 585.872.9350 • Fax: 585.872.2638 • [email protected] • www.piecal.com
Operating Instructions
SOURCE mV / V
Choose this function to provide an output from -20.000 to 99.999 mV, -500.00 to
999.99 mV or from 0.000 to 10.250 V. The source current is a nominal 20 mA to
provide the driving power to your voltage receivers.
Move the power switch w to SOURCE then Double Click the EZ-DIAL knob to
get into the Double Click Menu. Turn the knob to scroll through the settings and
press the knob to make your selection. Select V for the FUNCTION and 1V, 10V
or 100 mV for the RANGE.
Connect the output leads of the 830PM to the inputs of the device being
calibrated, making sure to check polarity. Connect the Red lead from jack (1) to
the plus (+) input and black lead from jack (2) to the minus (-) input.
Instantly output your SPAN and ZERO output settings by moving the EZ-CHECK
switch between HI and LO. You may also select any third output setting (such as
mid-range) using the SET position on the EZ-CHECK switch. The output is
adjusted in 0.001 mV, 0.01 mV or 0.001 V increments by turning the knob e.
Press and turn the knob for faster dialing with 0.100 mV, 1.00 mV or 0.100 V
increments.
5
MAX
READ
MIN
READ
6
-
+
-
3
2
-2
OFF
HI
SET
LO
SOURCE
Hz SYNC
EZ-DIAL
Double Click
Menu
Push & Turn for
Fast Dialing
Push & Hold to
Store/Step
piecal.com
EZ-CHECK™
PRACTICAL INSTRUMENT ELECTRONICS
MODEL 830PM
Process Calibrator
Loop Diagnostics · Transmitter Supply
mA · V ∙ pH · TC · Ω · RTD · Hz · Pressure
Pressure
Module
All mA∙ Read V
TC
Source & Read mV
Hz
Source V
pH
RTD • Ohms
OVERLOAD
4
1
+
+1
Voltage Receiver Input
Controller
Transmitter
Computer
Logger
DCS
V
5.000
HI
2 +
- 1
Practical Instrument Electronics
Tel: 585.872.9350 • Fax: 585.872.2638 • [email protected] • www.piecal.com
Operating Instructions
Read mV / V
Choose this function to measure from -99.999 to 99.999 millivolts, -999.99 to
+999.99 mV, 0.000 to 10.250 V dc or 0.00 to 60.00 V dc.
Move the power switch to READ then Double Click the EZ-DIAL knob to get into
the Menu. Turn the knob to scroll through the settings and press the knob to
make your selection. Select V for the FUNCTION and 1V, 10V, 60V or 100 mV
for the RANGE.
Connect the red input lead (+) of the 830PM to the more positive point of the
break and the black input to the more negative point. When measuring up to 100
mV connect the red lead to jack (1) and the black lead to jack (2). When
measuring up to 1V, 10 V or 60 V connect the red lead to jack (5) and the black
lead to jack (6).
Signals above the maximum scale are limited by protection circuitry with “OVER
RANGE” on the display and the red OVERLOAD LED lit.
The 830PM measures the input signal and constantly updates the display with
the current reading. Move the EZ-CHECK switch q to MAX to see the highest
reading and to MIN to see the lowest reading. Press and hold the knob e to clear
the MAX and MIN readings.
MAX
READ
MIN
READ
5
6
-
+
-
3
2
-2
Voltage Output Signal
Controller
Transmitter
Power Supply
OFF
HI
SET
LO
SOURCE
mV
8.20
2 +
- 1
Hz SYNC
PRACTICAL INSTRUMENT ELECTRONICS
MODEL 820PM
Process Calibrator
mA · V ∙ TC · Ω · RTD · Hz
Double Click
Menu
Push & Turn for
Fast Dialing
Push & Hold to
Store/Step
piecal.com
EZ-CHECK™
All mA∙ Read V
TC
Source & Read mV
Hz
Source V
RTD • Ohms
OVERLOAD
4
1
+
+1
Read 1V, 10 V
& 60 V
Practical Instrument Electronics
Tel: 585.872.9350 • Fax: 585.872.2638 • [email protected] • www.piecal.com
Choose this function to provide an output from 0.000 to 14.000 pH @ 25°C
(77°F) which corresponds to 414.12 to -414.12 mV. The source current is a
nominal 20 mA to provide the driving power to your pH receivers.
Move the power switch w to SOURCE then Double Click the EZ-DIAL knob to
get into the Menu. Turn the knob to scroll through the settings and press the knob
to make your selection. Select pH for the FUNCTION.
Connect the output leads of the 830PM to the inputs of the device being
calibrated, making sure to check polarity. Red lead from the mV (+) jack of the
830PM to the plus (+) input and black lead from the mV (-) jack to the minus (-)
input.
Instantly output your SPAN and ZERO output settings by moving the EZ-CHECK
switch between HI and LO. You may also select any third output setting (such as
mid-range) using the DIAL position on the EZ-CHECK switch. The output is
adjusted in 0.001 pH increments by turning the knob e. Press and turn the knob
for faster dialing with 0.100 pH increments.
5
MAX
READ
MIN
READ
6
-
+
-
3
2
-2
OFF
HI
SET
LO
SOURCE
Hz SYNC
EZ-DIAL
Double Click
Menu
Push & Turn for
Fast Dialing
Push & Hold to
Store/Step
piecal.com
EZ-CHECK™
PRACTICAL INSTRUMENT ELECTRONICS
MODEL 830PM
Process Calibrator
Loop Diagnostics · Transmitter Supply
mA · V ∙ pH · TC · Ω · RTD · Hz · Pressure
Pressure
Module
All mA∙ Read V
TC
Source & Read mV
Hz
Source V
pH
RTD • Ohms
OVERLOAD
4
1
+
+1
pH Receiver Input
Controller
Transmitter
Analyzer
Operating Instructions
pH SOURCE
Simulate pH probes into transmitters & analyzers
Use the pH simulator to verify proper operation of pH devices before you place
a probe into a calibrated buffer. Adjusting the pH transmitter or analyzer without
a probe allows you to make sure the device is calibrated and operating correctly.
The 830PM simulates 0.000 to 14.000 pH @ 25°C corresponding to 414.12 to
-414.12 mV.
Once the pH instrument has been adjusted against the 830PM reconnect the pH
probe and check it against the proper buffer (typically 7 pH). If the instrument
zero point requires more than the manufacturer’s recommendations (typically
within 0.5 pH) it is time to clean or replace the probe.
Practical Instrument Electronics
Tel: 585.872.9350 • Fax: 585.872.2638 • [email protected] • www.piecal.com
Operating Instructions
Source Thermocouple
Choose this function to provide a simulated thermocouple signal into controllers,
temperature transmitters, indicators or any input devices that measure
thermocouple sensors.
Move the power switch w to SOURCE then Double Click the EZ-DIAL knob to
get into the Double Click Menu. Turn the knob to scroll through the settings and
press the knob to make your selection. Select T/C for the FUNCTION, °F or °C
for the UNITS, T/C Type (J, K, T, E, R, S, B, N, G, C, D, L (J-DIN), U (T-DIN) or
P (Platinel II) and internal COLD JUNC ON or OFF (ON is the default).
Connect the 830PM to the inputs of the device being calibrated using the proper
type of thermocouple wire via the miniature thermocouple socket.
Instantly output your SPAN and ZERO output settings by moving the EZ-CHECK
switch between HI and LO. You may also select any third output setting (such as
mid-range) using the SET position on the EZ-CHECK switch. The output is
adjusted in 0.1° increments by turning the knob e. Press and turn the knob for
faster dialing with 10.0° increments.
5
MAX
READ
MIN
READ
6
-
+
-
3
2
-2
OFF
HI
SET
LO
SOURCE
Hz SYNC
EZ-DIAL
Double Click
Menu
Push & Turn for
Fast Dialing
Push & Hold to
Store/Step
piecal.com
EZ-CHECK™
PRACTICAL INSTRUMENT ELECTRONICS
MODEL 830PM
Process Calibrator
Loop Diagnostics · Transmitter Supply
mA · V ∙ pH · TC · Ω · RTD · Hz · Pressure
Pressure
Module
All mA∙ Read V
TC
Source & Read mV
Hz
Source V
pH
RTD • Ohms
OVERLOAD
4
1
+
+1
Instrument with T/C Input
Controller
Temperature Transmitter
Temperature Indicator
Temperature Trip or Alarm
Thermocouple Wire
T/C
+
-
TYPE K °C
830.0
LO
Practical Instrument Electronics
Tel: 585.872.9350 • Fax: 585.872.2638 • [email protected] • www.piecal.com
Operating Instructions
Read Thermocouple Sensors
Choose this function to measure temperatures with a thermocouple probe or
sensor.
Move the power switch w to READ then Double Click the EZ-DIAL knob to get
into the Double Click Menu. Turn the knob to scroll through the settings and
press the knob to make your selection. Select T/C for the FUNCTION, °F or °C
for the UNITS, T/C Type (J, K, T, E, R, S, B, N, G, C, D, L (J-DIN), U (T-DIN) or
P (Platinel II) and COLD JUNC ON or OFF (ON is the default).
Connect the 830PM to the inputs of the device being calibrated using the proper
type of thermocouple wire via the miniature thermocouple socket. If no sensor is
connected, a wire is broken or the sensor is burned out, OPEN TC will appear
on the display. Signals above the maximum scale are limited by protection
circuitry with “OVER RANGE” on the display and the red OVERLOAD LED lit.
The 830PM measures the input signal and constantly updates the display with
the current reading. Move the EZ-CHECK switch q to MAX to see the highest
reading and to MIN to see the lowest reading. Press and hold the knob e to clear
the MAX and MIN readings.
5
MAX
READ
MIN
READ
6
-
+
-
3
2
-2
OFF
HI
SET
LO
SOURCE
Hz SYNC
EZ-DIAL
Double Click
Menu
Push & Turn for
Fast Dialing
Push & Hold to
Store/Step
piecal.com
EZ-CHECK™
PRACTICAL INSTRUMENT ELECTRONICS
MODEL 830PM
Process Calibrator
Loop Diagnostics · Transmitter Supply
mA · V ∙ pH · TC · Ω · RTD · Hz · Pressure
Pressure
Module
All mA∙ Read V
TC
Source & Read mV
Hz
Source V
pH
RTD • Ohms
OVERLOAD
4
1
+
+1
Thermocouple Wire
T/C
+
-
TYPE K °F
1830.1
Practical Instrument Electronics
Tel: 585.872.9350 • Fax: 585.872.2638 • [email protected] • www.piecal.com
Operating Instructions
Source Resistance
Choose this function to provide a simulated resistance into any device that
measures resistance.
Move the power switch w to SOURCE then Double Click the EZ-DIAL knob to
get into the Double Click Menu. Turn the knob to scroll through the settings and
press the knob to make your selection. Select OHMS for the FUNCTION, 400Ω
or 4000Ω for the RANGE.
Disconnect all sensor wires from the devices to be calibrated and connect the
830PM to the inputs of the device using 2, 3 or 4 wires. For 2 Wire resistacne
connect the read lead into jack (1) and the black lead into jack (2). For 3 wire
resistance add a black lead plugged into jack (3) and for 4 wire resistance add
a red lead plugged into jack (4).
Instantly output your SPAN and ZERO output settings by moving the EZ-CHECK
switch between HI and LO. You may also select any third output setting (such as
mid-range) using the SET position on the EZ-CHECK switch. The output is
adjusted in 0.01Ω/0.1Ω increments by turning the knob e. Press and turn the
knob for faster dialing with 1.00Ω/10.0Ω increments.
5
MAX
READ
MIN
READ
6
-
+
-
3
2
-2
OFF
HI
SET
LO
SOURCE
Hz SYNC
EZ-DIAL
Double Click
Menu
Push & Turn for
Fast Dialing
Push & Hold to
Store/Step
piecal.com
EZ-CHECK™
PRACTICAL INSTRUMENT ELECTRONICS
MODEL 830PM
Process Calibrator
Loop Diagnostics · Transmitter Supply
mA · V ∙ pH · TC · Ω · RTD · Hz · Pressure
Pressure
Module
All mA∙ Read V
TC
Source & Read mV
Hz
Source V
pH
RTD • Ohms
OVERLOAD
4
1
+
+1
Instrument with
Resistance Input
Controller
Transmitter
PLC
Ω
250.00
1
+
-
2
4
+
-
3
Practical Instrument Electronics
Tel: 585.872.9350 • Fax: 585.872.2638 • [email protected] • www.piecal.com
Operating Instructions
Read Resistance & Check Continuity
Choose this function to measure resistance or check continuity.
Move the power switch w to READ then Double Click the EZ-DIAL knob to get
into the Double Click Menu. Turn the knob to scroll through the settings and
press the knob to make your selection. Select OHMS for the FUNCTION, 400Ω,
4000Ω or Continuity for the RANGE.
Disconnect all wires from the sensor and connect the 830PM to the resistor or
sensor using 2, 3 or 4 wires. The 830PM automatically detects how many wires
are connected using a patented circuit and indicates each wire that is connected.
Any wires that are not connected or broken are indicated by the 830PM. This is
useful for troubleshooting the sensor.
If continuity is selected, resistance is measured up to 400.00Ω. The beeper will
sound and )))) appears on the display when resistances < 100.00Ω are measured.
Signals above the maximum scale are limited by protection circuitry with “OVER
RANGE” on the display and the red OVERLOAD LED lit.
The 830PM measures the input signal and constantly updates the display with
the current reading. Move the EZ-CHECK switch q to MAX to see the highest
reading and to MIN to see the lowest reading. Press and hold the knob e to clear
the MAX and MIN readings.
5
MAX
READ
MIN
READ
6
-
+
-
3
2
-2
OFF
HI
SET
LO
SOURCE
Hz SYNC
EZ-DIAL
Double Click
Menu
Push & Turn for
Fast Dialing
Push & Hold to
Store/Step
piecal.com
EZ-CHECK™
PRACTICAL INSTRUMENT ELECTRONICS
MODEL 830PM
Process Calibrator
Loop Diagnostics · Transmitter Supply
mA · V ∙ pH · TC · Ω · RTD · Hz · Pressure
Pressure
Module
All mA∙ Read V
TC
Source & Read mV
Hz
Source V
pH
RTD • Ohms
OVERLOAD
4
1
+
+1
Ω
830.90
1
+
-
2
Practical Instrument Electronics
Tel: 585.872.9350 • Fax: 585.872.2638 • [email protected] • www.piecal.com
Operating Instructions
Source RTD
Choose this function to provide a simulated RTD signal into controllers,
temperature transmitters, indicators or any input devices that measure RTD
sensors.
Move the power switch w to SOURCE then Double Click the EZ-DIAL knob to
get into the Double Click Menu. Turn the knob to scroll through the settings and
press the knob to make your selection. Select RTD for the FUNCTION, °F or °C
for the UNITS and RTD (Choose from one of Platinum 100Ω, or 1000Ω, Copper
10Ω or 50Ω, or Nickel 120Ω curves). Note: Pt 100Ω 3850 is the most common
RTD type.
Disconnect all sensor wires from the devices to be calibrated and connect the
830PM to the inputs of the device using 2, 3 or 4 wires. For 2 Wire RTDs connect
the red lead into jack (1) and the black lead into jack (2). For 3 wire RTDs add a
black lead plugged into jack (3) and for 4 wire RTDs add a red lead plugged into
jack (4).
Instantly output your SPAN and ZERO output settings by moving the EZ-CHECK
switch between HI and LO. You may also select any third output setting (such as
mid-range) using the SET position on the EZ-CHECK switch. The output is
adjusted in 0.1° increments by turning the knob e. Press and turn the knob for
faster dialing with 10.0° increments.
5
MAX
READ
MIN
READ
6
-
+
-
3
2
-2
OFF
HI
SET
LO
SOURCE
Hz SYNC
EZ-DIAL
Double Click
Menu
Push & Turn for
Fast Dialing
Push & Hold to
Store/Step
piecal.com
EZ-CHECK™
PRACTICAL INSTRUMENT ELECTRONICS
MODEL 830PM
Process Calibrator
Loop Diagnostics · Transmitter Supply
mA · V ∙ pH · TC · Ω · RTD · Hz · Pressure
Pressure
Module
All mA∙ Read V
TC
Source & Read mV
Hz
Source V
pH
RTD • Ohms
OVERLOAD
4
1
+
+1
Instrument with RTD Input
Controller
Temperature Transmitter
Temperature Indicator
Temperature Trip or Alarm
Pt 100 α=3850 °C
LO 830.0
1
+
-
2
-
3
Practical Instrument Electronics
Tel: 585.872.9350 • Fax: 585.872.2638 • [email protected] • www.piecal.com
Operating Instructions
Read RTD Sensors
Choose this function to measure temperatures with an RTD probe or sensor.
Move the power switch w to READ then Double Click the EZ-DIAL knob to get
into the Double Click Menu. Turn the knob to scroll through the settings and press
the knob to make your selection. Select RTD for the FUNCTION, °F or °C for the
UNITS and RTD (Choose from one of Platinum 100Ω, or 1000Ω, Copper 10Ω or
50Ω, or Nickel 120Ω curves). Note: Pt 100Ω 3850 is the most common RTD type.
Disconnect all wires from the sensor and connect the PIE 830PM to the inputs of
the device using 2, 3 or 4 wires. For 2 Wire RTDs connect the read lead into jack
(1) and the black lead into jack (2). For 3 wire RTDs add a black lead plugged into
jack (3) and for 4 wire RTDs add a red lead plugged into jack (4). The PIE 830PM
automatically detects how many wires are connected using a patented circuit and
indicates each wire that is connected. Any wires that are not connected or broken
are indicated by the 830PM. This is useful for troubleshooting the sensor.
Signals above the maximum scale are limited by protection circuitry with “OVER
RANGE” on the display and the red OVERLOAD LED lit.
The 830PM measures the input signal and constantly updates the display with
the current reading. Move the EZ-CHECK switch q to MAX to see the highest
reading and to MIN to see the lowest reading. Press and hold the knob e to clear
the MAX and MIN readings.
MAX
READ
MIN
READ
5
6
-
+
-
3
2
-2
OFF
HI
SET
LO
SOURCE
Pt 100 α=3850 °F
LO 820.1
1
+
-
2
-
3
Hz SYNC
PRACTICAL INSTRUMENT ELECTRONICS
MODEL 820PM
Process Calibrator
mA · V ∙ TC · Ω · RTD · Hz
Double Click
Menu
Push & Turn for
Fast Dialing
Push & Hold to
Store/Step
piecal.com
EZ-CHECK™
All mA∙ Read V
TC
Source & Read mV
Hz
Source V
RTD • Ohms
OVERLOAD
4
1
+
+1
Practical Instrument Electronics
Tel: 585.872.9350 • Fax: 585.872.2638 • [email protected] • www.piecal.com
Operating Instructions
Source Frequency
Choose this function to provide a frequency signal into any input devices that
measure frequency.
Move the power switch w to SOURCE then Double Click the EZ-DIAL knob to
get into the Double Click Menu. Turn the knob to scroll through the settings and
press the knob to make your selection. Select FREQ for the FUNCTION and
20KHZ, 10000HZ, 1000HZ or 2000CPM for the RANGE.
Disconnect all input wires from the devices to be calibrated and connect the
830PM to the input of the device matching polarity. Connect the Red lead from
jack (1) to the plus (+) input and black lead from jack (2) to the minus (-) input.
The green HZ SYNC LED pulses in synch with the output pulses and may be
used to calibrate optical pickups. The output signal is a zero crossing square
wave with a fixed amplitude of 6 V peak-to-peak from -1 and + 5 V.
Instantly output your SPAN and ZERO output settings by moving the EZ-CHECK
switch between HI and LO. You may also select any third output setting (such as
mid-range) using the SET position on the EZ-CHECK switch. The output is
adjusted in 1 count increments by turning the knob e. Press and turn the knob
for faster dialing with 100 count increments.
5
MAX
READ
MIN
READ
6
-
+
-
3
2
-2
OFF
HI
SET
LO
SOURCE
Hz SYNC
EZ-DIAL
Double Click
Menu
Push & Turn for
Fast Dialing
Push & Hold to
Store/Step
piecal.com
EZ-CHECK™
PRACTICAL INSTRUMENT ELECTRONICS
MODEL 830PM
Process Calibrator
Loop Diagnostics · Transmitter Supply
mA · V ∙ pH · TC · Ω · RTD · Hz · Pressure
Pressure
Module
All mA∙ Read V
TC
Source & Read mV
Hz
Source V
pH
RTD • Ohms
OVERLOAD
4
1
+
+1
Frequency Receiver Input
Flowmeter
Transmitter
Computer
Logger
Controller
DCS
KHz
20.000
HI
2 +
- 1
Practical Instrument Electronics
Tel: 585.872.9350 • Fax: 585.872.2638 • [email protected] • www.piecal.com
Operating Instructions
Read Frequency
Choose this function to count frequency.
Move the power switch w to READ then Double Click the EZ-DIAL knob to get
into the Double Click Menu. Turn the knob to scroll through the settings and
press the knob to make your selection. Select FREQ for the FUNCTION and
20KHZ, 10000HZ, 1000HZ or 2000CPM for the RANGE.
Disconnect all input wires from the devices to be calibrated and connect the
830PM to the output of the device matching polarity. 830PM to the input of the
device matching polarity. Connect the Red lead from jack (1) to the plus (+) input
and black lead from jack (2) to the minus (-) input.
The green HZ SYNC LED pulses in synch with the input frequency.
Signals above the maximum scale are limited by protection circuitry with “OVER
RANGE” on the display and the red OVERLOAD LED lit.
The 830PM measures the input signal and constantly updates the display with
the current reading. Move the EZ-CHECK switch q to MAX to see the highest
reading and to MIN to see the lowest reading. Press and hold the knob e to clear
the MAX and MIN readings.
5
MAX
READ
MIN
READ
6
-
+
-
3
2
-2
OFF
HI
SET
LO
SOURCE
Hz SYNC
EZ-DIAL
Double Click
Menu
Push & Turn for
Fast Dialing
Push & Hold to
Store/Step
piecal.com
EZ-CHECK™
PRACTICAL INSTRUMENT ELECTRONICS
MODEL 830PM
Process Calibrator
Loop Diagnostics · Transmitter Supply
mA · V ∙ pH · TC · Ω · RTD · Hz · Pressure
Pressure
Module
All mA∙ Read V
TC
Source & Read mV
Hz
Source V
pH
RTD • Ohms
OVERLOAD
4
1
+
+1
Frequency Output Signal
Flowmeter
Flow Sensor
Variable Speed Drive
Controller
Transmitter
Hz
820.00
2 +
- 1
Practical Instrument Electronics
Tel: 585.872.9350 • Fax: 585.872.2638 • [email protected] • www.piecal.com
Operating Instructions
Read Pressure
Choose this function to measure pressure in one of 32 different engineering
units using a PIE Pressure Module.
1) Move the power switch w to READ then Double Click the EZ-DIAL knob to get
into the Menu. Turn the knob to scroll through the settings and press the knob to
make your selection. Select PRESSURE for the FUNCTION and make your
choice of UNITS to match the pressure instrument to be checked.
2) Remove the covers from the ends of the connector on the pressure module
cable and the pressure connector of the 830PM. Align the white arrows and plug
the cable into the 830PM.
3) Connect pressure hoses, fittings & pumps (if required) to the pressure
instrument to be checked.
4) Press and hold the e E-Z DIAL KNOB for 2 seconds (after MAX/MIN RESET
appears) to 'Zero' or 'Tare' the pressure. The display will briefly display 'TARE
ON' then a '0' appears on the display indicating that all measurements are
relative to the pressure measured when the calibrator was zeroed. Press and
hold the e E-Z DIAL KNOB for 2 seconds (after MAX/MIN RESET appears)
when you want to turn the 'Tare' off.
The 830PM measures the pressure
and constantly updates the display with
the
current
reading.
Move
the
EZ-CHECK switch q to MAX to see the
highest reading and to MIN to see the
lowest reading. Press and hold the
knob e for 1 second to clear the MAX
and MIN readings.
To disconnect the pressure module
gently pull up on the groved area just
above the white arrow.
5
MAX
READ
MIN
READ
6
-
+
-
3
2
-2
OFF
HI
SET
LO
SOURCE
Hz SYNC
EZ-DIAL
Double Click
Menu
Push & Turn for
Fast Dialing
Push & Hold to
Store/Step
piecal.com
EZ-CHECK™
PRACTICAL INSTRUMENT ELECTRONICS
MODEL 830PM
Process Calibrator
Loop Diagnostics · Transmitter Supply
mA · V ∙ pH · TC · Ω · RTD · Hz · Pressure
Pressure
Module
All mA∙ Read V
TC
Source & Read mV
Hz
Source V
pH
RTD • Ohms
OVERLOAD
4
1
+
+1
psi
015.03
Port for Pressure Module
Practical Instrument Electronics
Tel: 585.872.9350 • Fax: 585.872.2638 • [email protected] • www.piecal.com
Operating Instructions
Read Pressure
Media Compatibility
Non-isolated DN sensors: clean, dry, non-corrosive, non-condensing gases only
Isolated DI sensors: any media compatible with 316L SS & Viton®
Isolated GI, CI & AI sensors: any media compatible with 316L SS
Pressure Module Accuracy
±0.025% of full scale including all effects of linearity, repeatability and hysteresis
from -20° to +50°C (-4° to +122°F)
* The DN0010 sensor accuracy is ±0.050% of full scale
32 Engineering Units:
PSI • inches, feet, mm, cm and meter of H2O @ 4°C, 20°C & 60°F inches,
meter, cm and mm of Hg @ 0°C
torr • kg/cm2 • kg/m2 • Pa • hPa • kPa • MPa • Bar • mBar • ATM • oz/in2 • lb/ft2
Sensor Code
Application
DNxxxx
Differential, Non-isolated
0 to 0010*, 0028, 0200, 0415, 2000” H2O
DIxxxx
Differential, Isolated
0 to 0001, 0005, 0015, 0030, 0100, 0300, 0500 PSID
GIxxxx
Gauge, Isolated
0 to 0015, 0030, 0050, 0100, 0300, 0500, 1000, 3000 PSIG
CIxxxx
Compound, Isolated
-14.7 to +0015, 0030, 0050, 0100, 0300, 0500, 1000, 3000 PSIG
AIxxxx
Absolute, Isolated
0 to 0017, 0038, 0100, 1000 PSIA
Optional Pressure Modules
Practical Instrument Electronics
Tel: 585.872.9350 • Fax: 585.872.2638 • [email protected] • www.piecal.com
Pressure Module
Practical Instrument Electronics
Tel: 585.872.9350 • Fax: 585.872.2638 • [email protected] • www.piecal.com
Warranty
Our equipment is warranted against defective material and workmanship
(excluding batteries) for a period of three years from the date of shipment.
Claims under warranty can be made by returning the equipment prepaid to our
factory. The equipment will be repaired, replaced or adjusted at our option. The
liability of Practical Instrument Electronics (PIE) is restricted to that given under
our warranty. No responsibility is accepted for damage, loss or other expense
incurred through sale or use of our equipment. Under no condition shall Practical
Instrument Electronics, Inc. be liable for any special, incidental or consequential
damage.
Accessories
INCLUDED:
AC Adaptor (Part No. 020-0101 for 100 to 120V, 020-0100 for 220/230V ac), Certificate of
Calibration, Six (6) Mounting Screws
PIE Multifunction Wire Kit
Part No. 020-0820
1 Red & 1 Black Lead with Banana Plug & Alligator Clips
2 Red & 2 Black Leads with Banana Plugs & Spade Lugs
OPTIONAL:
T/C Wire Kit 1* for Types J, K, T & E
Part No. 020-0202
T/C Wire Kit 2* for Types B, R/S & N
Part No. 020-0203
* Thermocouple extension wire, stripped on one end with a corresponding miniature
thermocouple male connector on the other end.
Additional Information
PIE Calibrators are designed, assembled and calibrated in the USA. This
product is calibrated on equipment traceable to NIST and includes a Certificate
of Calibration. Test Data is available for an additional charge.
Practical Instrument Electronics recommends a calibration interval of one
year. Contact your local representative for recalibration and repair services.
Practical Instrument Electronics
Tel: 585.872.9350 • Fax: 585.872.2638 • [email protected] • www.piecal.com
Specifications
General
Operating Temp Range
-20 to 60 °C (-5 to 140 °F)
Storage Temp Range
-30 to 60 °C (-22 to 140 °F)
Temperature effect
≤ ± 0.01 %/°C of Full Scale; Cold Junction Sensor ≤ ±25 ppm/°C
Relative Humidity Range
10 % ≤RH ≤90 % (0 to 35 °C), Non-condensing
10 % ≤RH≤ 70 % (35 to 60 °C), Non-condensing
Isolation: Voltage
Common Mode
60V rms between all mA functions/Read V DC and Source V DC/
thermocouple/RTD/Ohms//Frequency/Pressure
50/60 Hz, 100 dB
Normal Mode Rejection
50/60 Hz, 50 dB
Noise
≤ ± ½ Least Significant Digit from 0.1 to 10 Hz
Size
5.63x3.00x1.60” 143x76x41mm (LxWxH)
Weight
12.1 ounces, 0.34 kg with boot & batteries
Batteries
Four “AA” Alkaline 1.5V (LR6)
Battery life
Read Functions: ≥ 20 hrs; Pressure ≥ 7 hrs
Source mA: ≥ 14 hours @ 12 mA into 250Ω
Pwr/Meas mA: ≥ 12 hours at 20 mA
Source V, Ω, T/C, RTD & Hz: ≥ 20 hours
Low Battery
Low battery indication with nominal 1 hour of operation left
Protection against
misconnection
Over-voltage protection to 60 vrms (rated for 30 seconds)
Red LED indicates OVERLOAD or out of range conditions
Display
High contrast graphic liquid crystal display, 0.315” (8.0 mm) high digits,
backlighting
Read mA
Ranges and Resolution
0.000 to 24.000 mA or -25.00 to 125.00% of 4-20 mA
Accuracy
≤ ± (0.02 % of Reading + 0.003 mA)
Voltage burden
≤ 2V at 24 mA
Overload/Current limit
protection
25 mA nominal
Source mA / Power & Measure Two Wire Transmitters
Ranges and Resolution
Same as Read mA
Accuracy
≤ ± (0.02 % of Reading + 0.003 mA)
Loop compliance voltage
≥ 24 DCV @ 20.00mA
Loop drive capability
1200 Ω at 20 mA for 15 hours nominal;
950 Ω with Hart Resistor enabled
mA 2-Wire Transmitter Simulation
Accuracy
Same as Source/Power & Measure
Voltage burden
≤ 2V at 20 mA
Overload/Current limit
24 mA nominal
Loop voltage limits
2 to 60 VDC (fuse-less protected)
Practical Instrument Electronics
Tel: 585.872.9350 • Fax: 585.872.2638 • [email protected] • www.piecal.com
Specifications
DC Voltage Read
Ranges and Resolution
±99.999 mV, ±999.99mV, 0 to 10.250 V, 0.00 to 60.00 V DC
Accuracy
≤ ± (0.02 % of Reading + 0.01% Full Scale)
Input resistance
≥ 1 MΩ
Source V dc
Ranges & Resolution
-20.000 to 99.999 mV, -500.00 to 999.99 mV, 0.000 to 10.250V
Accuracy
≤ ± (0.02 % of Reading + 0.01% Full Scale)
Source Current
≥ 24 mA
Sink Current
> 16 mA
Output Impedance
< 1 Ohm
Short Circuit Duration
Infinite
Thermocouple Source
Accuracy
≤ ± (0.02 % of Reading + 0.01 mV)
Cold Junction Compensation
± 0.05°C; Thermistor traceable to NIST for 11 years
Output Impedance
< 1 Ohm
Source Current
> 20 mA (drives 80 mV into 10 Ohms)
Thermocouple Read
Accuracy & Cold Junction
Compensation
Same as Thermocouple Source
Input Impedance
> 1 Megohms
Open TC Threshold; Pulse
10K Ohms; <5 µamp pulse for 300 milliseconds (nominal)
RTD, OHMS and Continuity Read
Resistance Ranges
0.00 to 401.00, 0.0 to 4010.0 Ohms
Accuracy
±(0.025% of Reading + 0.075 Ohms)
Excitation Current
1.0 mA to 401.0 Ohms, 0.5 mA to 4010 Ohms (nominal)
Continuity
0.0 to 401.0 Ohms; Beeps from 0.0 to 100.0 Ohms
pH Source
Range and Resolution
-414.00 to +414.00 pH
Accuracy in mV
≤ ± (0.02 % of Reading in mV + 0.1 mV)
Accuracy in pH
≤ ± 0.003 pH @ 25°C
Practical Instrument Electronics
Tel: 585.872.9350 • Fax: 585.872.2638 • [email protected] • www.piecal.com
Specifications
RTD and OHMS Source
Accuracy
From 1 to 10.2 mA
External Excitation Current
Below 1 mA of External
Excitation Current
±(0.025% of Full Scale + 0.075 Ohms)
±(0.025% of Full Scale+0.075 Ohms +
Resistance Ranges
0.00 to 401.00, 0.0 to 4010.0 Ohms
Allowable Excitation Current Range
<401 Ohm:10.2 mA max; steady or pulsed/intermittent
401 to 4000 Ohms: 1 mA max; steady or pulsed/intermittent
Pulsed Excitation Current Compatibility
DC to 0.01 second pulse width
0.025 mV
mA Excitation Current )
Frequency Source
Ranges
1 to 2000 CPM, 0.01 to 999.99 Hz, 0.1 to 9999.9 Hz, 0.001
to 20.000 kHz
Accuracy
± (0.02 % of Reading + 0.01% Full Scale)
Output Waveform
Square Wave, Zero Crossing -1.0 to +5 V peak-to-peak ±10%
Risetime (10 to 90% of amplitude)
< 10 microseconds
Output Impedance
< 1 Ohm
Source Current
> 1 mA rms at 20 kHz
Short Circuit Duration
Infinite
Optical Coupling
Green LED (HZ SYNC) flashes at output frequency
Frequency Read
Ranges & Accuracy
Same as Frequency Source
Accuracy
± (0.02 % of Reading + 0.01% Full Scale)
Trigger Level
1 V rms, dc coupled
Input Impedance
> 1 Meg Ohm + 60 pF
Practical Instrument Electronics
Tel: 585.872.9350 • Fax: 585.872.2638 • [email protected] • www.piecal.com
T/C
Type
Degrees C
Range
Accuracy
°C
Degrees F
Range
Accuracy
°F
T/C
Material
ISA/ANSI
Color
J
-200.0 to -50.0
±0.5°
-328.0 to -58.0
±1.0°
+Iron
-Connstantan
Jacket
White
Red
Black
-50.0 to 300.0
±0.2°
-58.0 to 572.0
±0.4°
300.0 to 900.0
±0.3°
572.0 to 1652.0
±0.6°
900.0 to 1200.0
±0.4°
1652.0 to 2192.0
±0.8°
K
-230.0 to -50.0
±1.2°
-382.0 to -58.0
±2.2°
+Chromel®
-Alumel®
Jacket
Yellow
Red
Yellow
-50.0 to 550.0
±0.3°
-58.0 to 1022.0
±0.6°
550.0 to 1000.0
±0.5°
1022.0 to 1832.0
±0.8°
1000.0 to 1371.1
±0.6°
1832.0 to 2500.0
±1.1°
T
-260.0 to -230.0
±2.9°
-436.0 to -382.0
±5.2°
+Copper
-Constantan
Jacket
Blue
Red
Blue
-230.0 to -210.0
±1.0°
-382.0 to -346.0
±1.9°
-210.0 to -50.0
±0.8°
-346.0 to -58.0
±1.4°
-58.0 to 50.0
±0.3°
-58.0 to 122.0
±0.6°
50.0 to 400.0
±0.2°
122.0 to 752.0
±0.4°
E
-240.0 to -200.0
±0.9°
-400.0 to -328.0
±1.7°
+Chromel
-Constantan
Jacket
Purple
Red
Purple
-200.0 to 0.0
±0.5°
-328.0 to 32.0
±0.8°
0.0 to 350.0
±0.2°
32.0 to 662.0
±0.3°
350.0 to 1000.0
±0.3°
662.0 to 1832.0
±0.6°
R
-18.3 to 100.0
±2.1°
-1.0 to 212.0
±3.8°
+Pt/13Rh
-Platinum
Jacket
Black
Red
Green
100.0 to 500.0
±1.3°
212.0 to 932.0
±2.4°
500.0 to 1400.0
±1.0°
932.0 to 2552.0
±1.8°
1400.0 to 1767.8
±1.2°
2552.0 to 3214.0
±2.0°
S
-18.3 to 100.0
±2.0°
-1.0 to 212.0
±3.7°
+Pt/10Rh
-Platinum
Jackrt
Black
Red
Green
100.0 to 350.0
±1.4°
212.0 to 662.0
±2.5°
350.0 to 1600.0
±1.1°
662.0 to 2912.0
±2.0°
1600.0 to 1767.8
±1.3°
2912.0 to 3214.0
±2.4°
B
315.6 to 600.0
±3.2°
600.0 to 1122.0
±5.7°
+Pt/30Rh
-Pt/6Rh
Jacket
Grey
Red
Grey
600.0 to 850.0
±1.7°
1122.0 to 1562.0
±3.1°
850.0 to 1100.0
±1.3°
1562.0 to 2012.0
±2.4°
1100.0 to 1820.0
±1.1°
2012.0 to 3308.0
±2.0°
Thermocouple Ranges & Accuracies
Table based on ≤ ± (0.02 % of Reading + 0.01 mV)
Note: Doesn’t include cold junction error of ±0.05°C
Practical Instrument Electronics
Tel: 585.872.9350 • Fax: 585.872.2638 • [email protected] • www.piecal.com
T/C
Type
Degrees C
Range
Accuracy
°C
Degrees F
Range
Accuracy
°F
T/C
Material
ISA/ANSI
Color
N
-230.0 to -150.0
±1.9°
-382.0 to -238.0
±3.4°
+Nicrosil
-Nisil
Jacket
Orange
Red
Orange
-150.0 to -50.0
±0.7°
-238.0 to -58.0
±1.2°
-50.0 to 950.0
±0.4°
-58.0 to 1742.0
±0.8°
950.0 to 1300.0
±0.5°
1742.0 to 2372.0
±1.0°
G
(W)
100.0 to 350.0
±1.7°
212.0 to 662.0
±3.0°
+Tungsten
-W26/Re
Jacket
White
Red
White/Blue
350.0 to 1700.0
±0.8°
662.0 to 3092.0
±1.5°
1700.0 to 2000.0
±1.0°
3092.0 to 3632.0
±1.8°
2000.0 to 2320.0
±1.1°
3632.0 to 4208.0
±2.1°
C
(W5)
-1.1 to 100.0
±0.8°
30.1 to 212.0
±1.4°
+W5/Re
-W26/Re
Jacket
White
Red
White/Red
100.0 to 1000.0
±0.7°
212.0 to 1832.0
±1.3°
1000.0 to 1750.0
±1.2°
1832.0 to 3182.0
±2.1°
1750.0 to 2320.0
±2.0°
3182.0 to 4208.0
±3.5°
D
(W3)
-1.1 to 150.0
±1.0°
30.1 to 302.0
±1.8°
+W3/Re
-W25/Re
Jacket
White
Red
White/Yellow
150.0 to 1100.0
±0.7°
302.0 to 2012.0
±1.3°
1100.0 to 1750.0
±1.0°
2012.0 to 3182.0
±1.8°
1750.0 to 2320.0
±2.0°
3182.0 to 4208.0
±3.6°
P
Platinel®
0.0 to 600.0
±0.3°
32.0 to 1112.0
±0.6°
+Pd55/Pt31/Au14
-Au65/Pd35
Jacket
Yellow
Red
Black
600.0 to 900.0
±0.4°
1112.0 to 1652.0
±0.8°
900.0 to 1200.0
±0.6°
1652.0 to 2192.0
±1.1°
1200.0 to 1395.0
±0.7°
2192.0 to 2543.0
±1.2°
DIN Colors
L
J-DIN
-200.0 to -50.0
±0.4°
-328.0 to -58.0
±0.7°
+Iron
-Connstantan
Jacket
Red
Blue
Blue
-50.0 to 300.0
±0.2°
-58.0 to 572.0
±0.4°
300.0 to 900.0
±0.3°
572.0 to 1652.0
±0.5°
U
T-DIN
-200.0 to -50.0
±0.6°
-328.0 to -58.0
±1.1°
+Copper
-Constantan
Jacket
Red
Brown
Brown
-50.0 to 50.0
±0.3°
-58.0 to 122.0
±0.5°
50.0 to 550.0
±0.2°
122.0 to 1022.0
±0.4°
550.0 to 600.0
±0.3°
1022.0 to 1112.0
±0.5°
Thermocouple Ranges & Accuracies
Table based on ≤ ± (0.02 % of Reading + 0.01 mV)
Note: Doesn’t include cold junction error of ±0.05°C
Practical Instrument Electronics
Tel: 585.872.9350 • Fax: 585.872.2638 • [email protected] • www.piecal.com
RTD Ranges & Accuracies
RTD Accuracy Based on ±(0.025% of Reading in Ohms + 0.075 Ohms)
RTD
Type
Degrees C
Range
°C
Degrees F
Range
°F
Pt 100 Ohm
DIN/IEC/JIS 1989
1.3850 (ITS-90)
-200.0 to 0.0
0.0 to 340.0
340.0 to 640.0
640.0 to 850.0
±0.2°
±0.3°
±0.4°
±0.5°
-328.0 to 32.0
248.0 to 644.0
644.0 to 1184.0
1184.0 to 1562.0
±0.4°
±0.6°
±0.8°
±1.0°
Pt 100 Ohm (Burns)
1.3902
-200.0 to 10.0
10.0 to 350.0
350.0 to 650.0
650.0 to 850.0
±0.2°
±0.3°
±0.4°
±0.5°
328.0 to 50.0
50.0 to 662.0
662.0 to 1202.0
1202.0 to 1562.0
±0.4°
±0.6°
±0.8°
±0.9°
Pt 100 Ohm
(Old JIS 1981)
1.3916
-200.0 to 20.0
20.0 to 360.0
360.0 to 650.0
650.0 to 850.0
±0.2°
±0.3°
±0.4°
±0.5°
-328.0 to 68.0
68.0 to 680.0
680.0 to 1202.0
1202.0 to 1562.0
±0.4°
±0.6°
±0.8°
±0.9°
Pt 100 Ohm
(US Lab)
1.3926
-200.0 to 20.0
20.0 to 360.0
360.0 to 660.0
660.0 to 850.0
±0.2°
±0.3°
±0.4°
±0.5°
-328.0 to 68.0
68.0 to 680.0
680.0 to 1220.0
1220.0 to 1562.0
±0.4°
±0.6°
±0.8°
±0.9°
Pt 1000 Ohm
DIN/IEC/JIS 1989
1.3850
-200.0 to 0.0
0.0 to 340.0
340.0 to 640.0
640.0 to 850.0
±0.2°
±0.3°
±0.4°
±0.5°
-328.0 to 32.0
248.0 to 644.0
644.0 to 1184.0
1184.0 to 1562.0
±0.4°
±0.6°
±0.8°
±1.0°
Copper 10 Ohm
1.4274 (Minco)
-200.0 to 260.0
±2.0°
-328.0 to 500.0
±3.6°
Copper 50 Ohm
1.4280
-50.0 to 150.0
±0.4°
-58.0 to 302.0
±0.8°
Ni 120 Ohm 1.6720
(Pure)
-80.0 to 260.0
±0.1°
-112.0 to 500.0
±0.3°
Ni 110 Ohm
(Bristol 7 NA)
1.5801
-100.0 to 260.0
±0.2°
-148.0 to 500.0
±0.4°
Practical Instrument Electronics
Tel: 585.872.9350 • Fax: 585.872.2638 • [email protected] • www.piecal.com
0.565”
0.995”
4.375”
7.755”
8.185”
8.85”
0.196”
0.35”
4.40”
4.554”
4.75”
0.125” Dia
(x6)
Bezel
Cutout
Panel Cutout Dimensions
Practical Instrument Electronics
Tel: 585.872.9350 • Fax: 585.872.2638 • [email protected] • www.piecal.com
Mounted in Panel
Practical Instrument Electronics
Tel: 585.872.9350 • Fax: 585.872.2638 • [email protected] • www.piecal.com
Rear view of 830PM mounted in a panel with AC Adapter
Mounted in Panel