NUCLEONIX SYSTEMS PVT. LTD

I N S T R U C T I O N M A N U A L PRE - AMPLIFIER TYPE : PA320 NUCLEONIX SYSTEMS PRIVATE LIMITED Plot No : 162 A & B, PHASE II, I.D.A.Cherlapally, Hyderabad - 500 051 Ph: 91-040-27263701/30918055, FAX : 27262146, e-mail : [email protected], [email protected] File Name : PA320_MAN Date : 20-08-2013 ...

NUCLEONIX SYSTEMS PVT. LTD

CONTENTS UNPACKING CHAPTER I INTRODUCTION CHAPTER II SPECIFICATIONS CHAPTER III INSTALLATION CHAPTER IV OPERATION CHAPTER V BLOCK DIAGRAM & ITS DESCRIPTION CHAPTER VI CIRCUIT DIAGRAM ...

NUCLEONIX SYSTEMS PVT. LTD

UNPACKING Charge sensitive Pre-amplifier Type : PA320 has been thoroughly tested and is despatched in ready to use condition. However, on unpacking and prior to operation, it is advisable to check visually and make sure that there is no visible damage caused in transit. If any damage to the instrument is observed, do not switch O ...

NUCLEONIX SYSTEMS PVT. LTD

NUCLEONIX Charge sensitive Pre -amplifier PA320 is designed for use with gas filled proportional counters. Fully solid-state design and has a choice of two gain factors 1 or 3. Its mechanical construction enables the detector to be directly connected where as the low output impedance can drive fairly long connecting cable upto t ...

NUCLEONIX SYSTEMS PVT. LTD

Detector Load Resistor : 10 M Ohms Output impedance : 100 Ohms Charge Sensitivity : (0.22x10-12volts/ coulomb. Gain can be set additionally by a toggle switch to 1 or 3 position. Rise Time : Better than 100 nano seconds for both polarities of the signal. Decay Time Constant : 50 micro sec. (approx). Charge Sensitive Device : N-Chan ...

NUCLEONIX SYSTEMS PVT. LTD

CHAPTER - III INSTALLATION A. CONTROLS / CONNECTORS : EHT (MHV socket) : High Voltage connector, at which the EHT supply is provided to connect to the X-ray proportional counter / gas proportional / He / BF3 / Neutron detector for detector biasing. Output (BNC) : BNC Co-axial connector at which output from the Pre - Amplifieris ge ...

NUCLEONIX SYSTEMS PVT. LTD

CHAPTER - IV OPERATION A. OPERATING INSTRUCTIONS: Charge sensitive preamplifier PA320 is designed for connecting directly to any x-ray proportional counter/ gas proportional detectors, Krypton filled mossbauer detectors, He, BF3 Neutron counters etc. It is essential to have short cable between detector & pre-amplfier PA320. Ofcours ...