Tenma 72-7740 Operating Manual
Read Online
Read the Tenma 72-7740 Operating Manual online using the built-in PDF viewer.
Text of the manual
Rules For Safe Operation
Title
Page
Table of Contents
1
D.
Continuity Test
International Electrical Symbols
AC Voltage Measurement
Capacitance Measurement
The Use of Relative Value Mode
The Meter Structure
Turning on the Display Backlight
General Specifications
Resistance
Overview
Safety Information
Rotary Switch
Functional Buttons
Display Symbols
Measurement Ranges
A.
B.
Measurement Operation
A.
B.
C.
D.
Testing for Continuity
E.
F.
G.
H.
I.
Temperature Measurement
J.
Operation of Hold Mode
The POWER Button
The BLUE Button
Sleep Mode (Model 72-7740 only)
Accuracy Specifications
A.
DC Voltage
B.
AC Voltage
C.
Selecting a Measurement Range
Manual Ranging and Autoranging
DC Voltage Measurement
Measuring Resistance
Frequency Measurement
DC or AC Current Measurement
Testing Diodes
Measuring Duty Cycle
Model 72-7740/72-7745: OPERATING MANUAL
Frequency & Duty Cycle
DC Current
28
28
28
29
29
30
31
31
32
33
35
35
35
Page
2
J.
AC Current
Maintenance
A.
General Service
B.
Testing the Fuses
C.
Replacing the Battery
D.
Replacing the Fuses
RS232C Serial Port (Model 72-7745)
A.
RS232C Port Cable
B.
Setting of RS232C Serial Ports
C.
System Requirements for
Installing the serial Interface Program
Diode Test
I.
E.
F.
G.
H.
Temperature
Model 72-7740/72-7745: OPERATING MANUAL
This Operating Manual covers information on safety and
cautions. Please read the relevant information carefully
and observe all the Warnings and Notes strictly.
Warning
To avoid electric shock or personal injury, read the
“Safety Information” and “Rules for Safe Operation”
carefully before using the Meter.
Digital Multimeter Model 72-7740 (hereafter referred to
as “the Meter”) have autorange and manual range options
with maximum reading 3999. The enclosure structure
design adopted advanced “co-injection” technique in
order to provide sufficient insulation.
Model 72-7740/72-7745: OPERATING MANUAL
CAT IV: Primary supply level, overhead lines, cable
systems etc.
4
Unpacking Inspection
Open the package case and take out the Meter. Check the
following items carefully to see any missing or damaged part:
In the event you find any items missing or damaged,
please contact your dealer immediately.
Safety Information
This Meter complies with the standards IEC61010: in
pollution degree 2, overvoltage category (CAT. III 1000V,
CAT. IV 600V;)and double insulation.
CAT. III: Distribution level, fixed installation, with smaller
transient overvoltages than CAT. IV
Use the Meter only as specified in this operating manual,
otherwise the protection provided by the Meter may be
impaired.
In this manual, a Warning identifies conditions and
actions that pose hazards to the user, or may damage
the Meter or the equipment under test.
A Note identifies the information that user should pay
attention to.
International electrical symbols used on the Meter and
in this Operating Manual are explained on page 6.
Point Contact Temperature Probe
9V Battery (NEDA1604, 6F22 or 006P)
(installed)
Model 72-7740/72-7745: OPERATING MANUAL
Rules For Safe Operation
To avoid possible electric shock or personal injury, and to
avoid possible damage to the Meter or to the equipment
under test, adhere to the following rules:
l Before using the Meter inspect the case. Do not use the
Meter if it is damaged or the case (or part of the case) is
removed. Look for cracks or missing plastic. Pay
attention to the insulation around the connectors.
l Inspect the test leads for damaged insulation or exposed
metal. Check the test leads for continuity. Replace
damaged test leads with identical model number or
electrical specifications before using the Meter.
l Do not apply more than the rated voltage, as marked on
the Meter, between the terminals or between any
terminal and ground.
l The rotary switch should be placed in the correct
position and no change of range made during
measurement , to prevent damage of the Meter.
l When working at an effective voltage over 60V
DC or 30V rms AC, special care should be taken
there is danger of electric shock.
l Use the proper terminals, function, and range for your
measurements.
l Do not use or store the Meter in an environment of high
temperature, humidity, explosive, flammable and
strong magnetic field. The performance of the Meter
may deteriorate after dampened.
l When using the test leads, keep your fingers behind the
finger guards.
l Disconnect circuit power and discharge all high-voltage
capacitors before testing resistance, continuity, diodes,
current, or capacitance.
l Before measuring current, check the Meter’s fuses and
turn off power to the circuit before connecting the Meter
to the circuit.
l Replace the battery as soon as the battery indicator " "
appears. With a low battery, the Meter might produce
false readings that can lead to electric shock and
personal injury.
l Remove test lead, temperature probe, RS232C interface
Warning
5
Model 72-7740/72-7745: OPERATING MANUAL
cable and test clip from the Meter and turn the Meter
power off before opening the case.
l When servicing the Meter, use only the same model
number or identical electrical specifications replacement
parts.
l The internal circuit of the Meter shall not be altered at
will to avoid damage of the Meter and any accident.
l Soft cloth and mild detergent should be used to clean the
surface of the Meter when servicing. No abrasive and
solvent should be used to prevent the surface of the Meter
from corrosion, damage and accident.
l
l
l
l
The Meter is suitable for indoor use.
Under the environment with high (+/-4kV) electrostatic
discharge, the Meter may not be operated as normal
condition. The user may require resetting the Meter.
Remove the battery when not used for a prolonged
period to avoid damage to the Meter.
Please constantly check the battery as it may leak
when it has been using for some time, replace the
battery as soon as leaking appears. A leaking battery
will damage the Meter.
6
International Electrical Symbols
AC (Alternating
Current)
Deficiency of Built-In
Battery
DC (Direct Current)
Continuity Test
AC or DC
Diode
Grounding
Capacitance Test
Double Insulated
Fuse
Warning. Refer to the
Operating Manual
Conforms to Standards
of European Union
Model 72-7740/72-7745: OPERATING MANUAL
l Periodically check the battery as it may leak after
some time. If leakage is apparent,the battery should
be immediately replaced to prevent damage.
The Meter Structure (see figure 1)
(figure 1)
1 LCD Display
2 Functional Buttons
3 Rotary Switch
4 HzV
Input Terminal:
Input for voltage, frequency/duty cycle, resistance, diode,
continuity and capacitance measurements.
5 COM Input Terminal:
Return terminal for all measurements.
Input Terminal:
Input for 0.1 A to 400.0mA current measurements and
temperature testing.
7 10A Input Terminal:
Input for 0.01A to 10.00A current measurements.
7
Model 72-7740/72-7745: OPERATING MANUAL
The table below provides information about rotary switch
positions.
Rotary
Switch
Position
Function
V
DC voltage measurement range from 400.0mV to
1000V or
AC voltage measurement range from 4.000V to
750.0V.
Continuity test.
Diode test.
Resistance measurement range from 400.0
Temperature in celsius
Hz Frequency measurement range from 10.00Hz to
10.00MHz.
AC or DC current measurement range from
A.
mA
AC or DC current measurement range from
40.00mA to 400.0mA.
A
AC or DC current measurement range from
4.000A to 10.00A.
to 40.00M
.
Model 72-7740/72-7745: OPERATING MANUAL
Functional Buttons (1)
9
The table below provides information about functional
button operation.
Button
Measuring
Function
Operation Performed
POWER
Any rotary
switch
position
Turn the power on and off.
V
Switches between AC and DC
voltage; the Meter beeps. DC is
default.
Switches between continuity and
diode and resistance and
capacitance measurements; the
Meter beeps.
Resistance is default.
Switches between AC and DC
current range from 400.0 A to
4000 A; the Meter beeps. DC is
default.
mA
Switches between AC and DC
current range from 40.00mA to
400.0mA; the Meter beeps. DC is
default.
A
Switches between AC and DC
current range from 4.000A to
10.00A; the Meter beeps. DC is
default.
Disable Sleep Mode feature.
(BLUE)
Model 72-7740/72-7745: OPERATING MANUAL
l Press RANGE to enter the
manual ranging mode; the Meter
beeps.
Manually selecting a range causes
the Meter to exit the Hold and REL
modes.
l Press RANGE to step through
the ranges available for the
selected function; the Meter beeps.
l Press and hold RANGE for 2
seconds to return to autoranging;
the Meter beeps.
Hz
1. Press to start the frequency
counter; the Meter beeps.
2. Press again to enter duty cycle
mode; the Meter beeps.
3. Press again to return to the
frequency counter mode; the Meter
beeps.
Hz %
V
,
,
mA
or
A
1. Press to start the frequency
counter; the Meter beeps.
2. Press again to enter duty cycle
mode; the Meter beeps.
3. Press again to return to the
previous measurement mode; the
Meter beeps.
REL
Any rotary
switch
position
except Hz
Press
to enter
and exit the REL mode in any
measuring mode except in
frequency/duty cycle; the Meter
beeps.
Any rotary
switch
position
Press HOLD H to enter and exit
the Hold mode in any mode, the
Meter beeps.
REL
Functional Buttons (2)
HOLD H
10
Model 72-7740/72-7745: OPERATING MANUAL
Display Symbols (1)
(see figure 2)
Number Symbol Meaning
1AC
Indicator for AC voltage or current.
Model 72-7740: The displayed value is the
mean value.
Model 72-7745:The displayed value is
the true rms value.
2 TRMS Model 72-7745: Indicator for true rms
value.
3
The Meter is in the auto range mode in
which the Meter automatically selects the
range with the best resolution.
4 RS232C Data output. It is always on the LCD, but
data output is only in progress when the
Meter is connected to the computer via the
included RS232C Interface Cable.
5
% Percent: Used for duty cycle measurements.
6
Data hold is active.
7
The REL is on to display the present value
minus the stored value.
8
The battery is low.
Warning To avoid false readings,
which could lead to possible electric shock
or personal injury, replace the battery as
soon as the battery indicator appears.
9
Centigrade. The unit of temperature.
10
Diode test.
11
The continuity buzzer is on.
18
1
2
3
4
5
6
7
8
9
10
11
12 ~ 16
17
(figure 2)
11
Model 72-7740/72-7745: OPERATING MANUAL
, k ,
M
:
k
:
M
:
Ohm. The unit of resistance.
kilohm. 1 x 10 or 1000 ohms.
Megaohm. 1 x 10 or 1,000,000 ohms.
nF
F:
nF:
Farad. The unit of capacitance.
Microfarad. 1 x 10 or 0.000001
farads.
Nanofarad. 1 x 10 or 0.000000001
farads.
Hz,
kHz,
MHz
Hz:
kHz:
MHz:
Hertz. The unit of frequency in
cycles/second.
Kilohertz. 1 x 10 or 1,000 hertz.
Megahertz. 1 x 10 or1,000,000
hertz.
V, mV V:
mV
Volts. The unit of voltage.
Millivolt. 1 x 10 or 0.001 volts.
12~16
A, mA, A:
mA:
Amperes (amps). The unit of current.
Milliamp. 1 x 10 or 0.001 amperes.
Microamp. 1x 10 or 0.000001
amperes.
17
Indicates negative reading.
18
The input value is too large for the selected
range.
Measurement Ranges (1)
The Meter has both manual range and autorange options:
12
Display Symbols (1) (see figure 2)
A measurement range determines the highest value the Meter
can measure. Most Meter functions have more than one range.
See “Accuracy Specifications.”
A.Selecting a Measurement Range
Choosing the correct measurement range is important:
l If the range is too low for the input, the Meter displays to
indicate an overload.
l If the range is too high, the Meter will not display the most
accurate measurement.
B. Manual Ranging and Autoranging
Model 72-7740/72-7745: OPERATING MANUAL
l In the autorange mode, the Meter selects the best range for
input signals.
This allows you to switch test points without having to reset
the range.
l In the manual range mode, you may select the range.
This allows you to override autorange and lock the Meter in a
specific range.
The Meter defaults to the autorange mode in measurement
functions that have more than one range. When the Meter is in
the autorange mode,
To enter and exit the manual range mode:
1. Press RANGE.
The Meter enters the manual range mode and
turns off.
Pressing RANGE increments the range. When the highest
range is reached, the Meter wraps to the lowest range.
Measurement Ranges (2)
Note
l If you manually change the measurement range after entering
the REL or Hold recording modes, the Meter exits these
modes.
l Under frequency/duty cycle and capacitance measurement,
only autorange mode is available.
2. To exit the manual range mode, press and hold RANGE for
two seconds.
The Meter returns to the autorange mode and
is displayed.
is displayed.
13
Model 72-7740/72-7745: OPERATING MANUAL
Measurement Operation (1)
A.DC Voltage Measurement
(see figure 3)
(figure 3)
To avoid harm to you or damage to the Meter from
electric shock, please do not attempt to measure voltages
higher than 1000VDC / 750VAC rms although readings
may be obtained.
The DC Voltage ranges are: 400.0mV, 4.000V, 40.00V, 400.0V
and 1000V. To measure DC voltage, connect the Meter as
follows:
1. Insert the red test lead into the
terminal and the black
test lead into the COM terminal.
2. Set the rotary switch to
; DC measurement is default or
press BLUE button to select DC measurement mode.
3. Connect the test leads across with the object being measured.
The measured value shows on the display.
Note
l In each range, the Meter has an input impedance of 10M
.
This loading effect can cause measurement errors in high
impedance circuits. If the circuit impedance is less than or
equal to 10k
, the error is negligible (0.1% or less).
l When DC voltage measurement has been completed,
disconnect the connection between the test leads and the
circuit under test, and move the test leads away from the
input terminals of the Meter.
Warning
HzV
V
14
Model 72-7740/72-7745: OPERATING MANUAL
Measurement Operation (2)
15
B. AC Voltage Measurement
(see figure 4)
(figure 4)
the true root mean square value converter needs a longer
time to stabilize.
To avoid harm to you or damage to the Meter from
electric shock, please do not attempt to measure voltages
higher than 1000VDC / 750VAC rms although readings
may be obtained.
Warning
The AC voltage ranges are: 4.000V, 40.00V, 400.0V and
750.0V. To measure AC Voltage, connect the Meter as follows:
1. Insert the red test lead into the
terminal and the black
test lead into the COM terminal.
2. Set the rotary switch to
and press BLUE button to select
AC measurement.
3. Connect the test leads across with the object being measured.
The measured value shows on the display.
Note
l In each range, the Meter has an input impedance of 10M
.
This loading effect can cause measurement errors in high
impedance circuits. If the circuit impedance is less than or
equal to 10k
, the error is negligible (0.1% or less).
l When AC voltage measurement has been completed,
disconnect the connection between the testing leads and the
circuit under test, and remove the testing leads away from the
input terminals of the Meter.
l True root mean square (TRMS)value stability period(Model
72-7745): When the reading obtained is less than 100 digits,
When there is no input voltage, the maximum reading
displayed is 10 digits.
HzV
V
Model 72-7740/72-7745: OPERATING MANUAL
To avoid damage to the Meter or to the devices under test,
disconnect circuit power and discharge all the high-voltage
Warning
C.Measuring Resistance
(see figure 5)
Measurement Operation (3)
(figure 5)
from the reading.
The resistance ranges are: 400.0 , 4.000k
, 40.00k
,
400.0k
, 4.000M and 40.00M
. To measure resistance,
connect the Meter as follows:
1. Insert the red test lead into theHzV terminal and the black
test lead into theCOM terminal.
2. Set the rotary switch to
; resistance measurement
( ) is default or pressBLUE button to select
measurement
mode.
3. Connect the test leads across with the object being measured.
The measured value shows on the display.
16
capacitors before measuring resistance.
Note
l The test leads can add 0.1
to 0.2
of error to resistance
measurement. To obtain precision readings in low-resistance
measurement, that is the range of 400.0
, short-circuit the
input terminals beforehand, using the relative measurement
function buttonREL
to
measured when the testing leads are short-circuited
automatically subtract the value
Model 72-7740/72-7745: OPERATING MANUAL
reading with shorted test leads is not
0.5
, check for
loose test leads, incorrect function selection, or enabled Data
Hold function.
l For high-resistance measurement (
), it is normal to
take several seconds to obtain a stable reading.
l The LCD displays indicating open-circuit for the tested
resistor or the resistor value is higher than the maximum
range of the Meter.
l When resistance measurement has been completed,
disconnect the connection between the test leads and the
circuit under test, and remove the test leadsfrom the
input terminals of the Meter.
D.Testing for Continuity
(see figure 6)
Measurement Operation (4)
(figure 6)
To avoid damage to the Meter or to the device under test,
disconnect circuit power and discharge all the high-voltage
capacitors before testing for continuity.
To test for continuity, connect the Meter as below:
1. Insert the red test lead into theHzV
terminal and the black
test lead into the COM terminal.
2. Set the rotary switch to
measurement mode.
3. The buzzer sounds if the resistance of a circuit under test is
less than around 70
Warning
and press BLUE button
to select
Note
l The LCD displays
indicating the circuit being tested is
open.
17
Model 72-7740/72-7745: OPERATING MANUAL
E. Testing Diodes
(see figure 7)
To avoid possible damage to the Meter and to the device
under test, disconnect circuit power and discharge all high-
voltage capacitors before testing diodes.
Use the diode test to check diodes, transistors, and other
semiconductor devices. The diode test sends a current through
the semiconductor junction, and then measures the voltage drop
across the junction. A good silicon junction drops between
0.5V and 0.8V.
Warning
l When continuity testing has been completed, disconnect the
connection between the test leads and the circuit under
test, and remove the test leads from the input terminals
of the Meter.
3. For forward voltage drop readings on any semiconductor
component, place the red test lead on the component’s anode
and place the black test lead on the component’s cathode.
The measured value shows on the display.
Note
l In a circuit, a good diode should still produce a forward
To test a diode out of a circuit, connect the Meter as follows:
Insert the red test lead into theHzV
terminal and the black
testlead into the COM terminal.
2. Set the rotary switch to
measurement mode.
and press BLUE button
to select
Measurement Operation (5)
(figure 7)
1.
18
Model 72-7740/72-7745: OPERATING MANUAL
Measurement Operation (6)
F.Capacitance Measurement
(see figure 8)
voltage drop reading of 0.5V to 0.8V; however, the reverse
voltage drop reading can vary depending on the resistance of
other pathways between the probe tips.
l Connect the test leads to the proper terminals as said above to
avoid error display. The LCD will display
indicating
diode being tested is open or polarity is reversed.The unit of
diode is Volt (V), displaying the forward voltage drop
readings.
l When diode testing has been completed, disconnect the
connection between the testing leads and the circuit under
test, and remove the testing leads away from the input
terminals of the Meter.
(figure 8)
Warning
Insert the red test lead into theHzV
terminal and the
black test lead into the COM terminal.
2.Set the rotary switch to
measurement mode.
and press BLUE
button to select nF
3. Connect the test leads across with the object being
measured. The measured value shows on the display.
Note
l For testing polarited capacitors, connect the red clip to
To avoid damage to the Meter or to the equipment under
test, disconnect circuit power and discharge all high-voltage
capacitors before measuring capacitance. Use the DC
Voltage function to confirm that the capacitor is discharged.
1.
19
Model 72-7740/72-7745: OPERATING MANUAL
input terminals of the Meter.
input terminals of the Meter.
Measurement Operation (7)
20
anode & black clip to cathode instead of using test leads as
mentioned above.
l To minimize the effect of capacitance stored in the test leads,
the test lead should be as short as possible. To measure a
small value of capacitance, use REL mode to remove the
leads capacitance. Remaining voltage, insulated impedance,
& dielectric absorption from the capacitor may cause slight
measurement error.
l Higher value capacitors will require additional time for
measurement.The 100 F range requires up to 15 seconds.
l The LCD displays
indicating the tested capacitor is
shorted or it exceeds the maximum range.
l When capacitance measurement has been completed,
disconnect the connection between the test leads and the
circuit under test and remove the test leads from the
The measured value shows on the display.
The measurement range is from 10Hz to 10MHz. To measure
frequency, connect the Meter as follows:
1. Insert the red test lead into theHzV
terminal and the black
test lead into the COM terminal.
2. Set the rotary switch to Hz; frequency measurement (Hz) is
default or press Hz % to select Hz measurement mode.
3. Connect the test leads across with the object being measured.
G. Frequency Measurement
(see figure 9)
Note
l When frequency measurement has been completed,
disconnect the connection between the test leads and the
circuit under test, and remove the test leads from the
(figure 9)
Model 72-7740/72-7745: OPERATING MANUAL
Range Signal Requirement Frequency Range
4V
1.0V 5Hz~10kHz
5.0 5Hz~20kHz
45V 45Hz~4kHz
420 45Hz~1.6kHz
45mA 5Hz~5kHz
Measurement Operation (8)
A
l To obtain a stable reading when measuring input scope
30V rms frequency signal:
Set the rotary switch to V .
Then press Hz% to select Hz measurement mode to
obtain frequency value.
When input scope 30V rms, please follow the above
step 2. carrying out the measurement.
l When making frequency measurement at voltage or current
range, please mind the following signal requirement table:
4A 45Hz~1kHz
l To obtain a stable reading when measuring input scope
30V rms frequency signal:
Set the rotary switch to V .
Then press Hz% to select % measurement mode to
obtain
duty cycle value.
When input scope 30V rms, please follow the above
step 2. carrying out the measurement.
H. Measuring Duty Cycle (See figure 9)
The duty cycle measurement range is: 0.1%~99.9%. To
measure duty cycle, do the following:
1. Set up the Meter to measure frequency.
2. To select duty cycle, press Hz % until the % symbol is shown
on the display.
3. Connect the test leads across with the object being measured.
The measured value shows on the display.
Note
l The LCD displays 000.0% indicating the input signal is high
or low level.
l When duty cycle measurement has been completed,
disconnect the connection between the test leads and the
circuit under test, and remove the test leads from the
input terminals of the Meter.
21
Model 72-7740/72-7745: OPERATING MANUAL
Measurement Operation (9)
I. Temperature Measurement(see figure 10)
The temperature measurement range is
measure temperature, connect the Meter as follows:
1. Insert the red temperature probe into the
and the black temperature probe into theCOM terminal.
2. Set the rotary switch to
3. Place the temperature probe to the object being measured.
The measured value shows on the display.
Never attempt an in-circuit current measurement where the
open-circuit voltage between the circuit and ground is
greater than 250V.
If the fuse burns out during measurement, the Meter may be
damaged or the operator may be injured. Use proper
terminals, function, and range for the measurement. When
the test leads are connected to the current terminals, do
not parallel them across any circuit.
Warning
J. DC or AC Current Measurement
(see figure 11)
Note
l The Meter automatically displays the temperature value
inside the Meter when there is no temperature probe
connection.
l The included point contact temperature probe can only be used
up to 250 . For any measurement higher than that, rod type
temperature probe must be used.
l When temperature measurement has been completed,
disconnect the connection between the test leads and the
circuit under test, and remove the test leads from the
input terminals of the Meter.
(figure 11)
To
22
Model 72-7740/72-7745: OPERATING MANUAL
(figure 10)
Measurement Operation (10)
auto ranging.
The current measurement has 3 measurement positions on the
rotary switch: A
, mA
and A
.
The A
has a 400.0 A and 4000 A range, with auto ranging;
the mA
has a 40.00mA and 400.0mA range, with auto
ranging; A
position has a 4.000A and 10.00A range, with
To measure current, do the following:
1. Turn off power to the circuit. Discharge all high-voltage
capacitors.
2. Insert the red test lead into the
or 10A terminal
and the black test lead into the COM terminal.
Use the10A terminal and A
range if the current value to
be tested is an unknown.
3. Set the rotary switch to µA
, mA
, or A
.
4. The Meter defaults to DC current measurement mode. To
toggle between DC and AC current measurement function,
press BLUE button.
Model 72-7740:
AC current is displayed as effective
value of sine wave (mean value response).
Model 72-7745 : AC current is displayed as true rms value.
5. Break the current path to be tested. Connect the red test lead
to the more positive side of the break and the black test lead
to the more negative side of the break.
6. Turn on power to the circuit.
The measured value shows on the display.
l For safety sake, each measurement time for high current
should be less than 10 seconds and the interval time between
2 measurements should be greater than 15 minutes.
l When current measurement has been completed, disconnect
the connection between the test leads and the circuit under
test, and remove the test leads from the input terminals
of the Meter.
Note
l (TRMS) stability period: (Model 72-7745)
When the reading obtained is less than 100 digits, the true
root mean square value converter needs a longer time to
stabilize. When there is no input voltage, the maximum reading
displayed is 10 digits.
23
Model 72-7740/72-7745: OPERATING MANUAL
oC
Operation of Hold Mode
To avoid possibility of electric shock, do not use Hold mode
to determine if circuits are without power. The Hold mode
will not capture unstable or noisy readings.
The Hold mode is applicable to all measurement functions.
l PressHOLD H to enter Hold mode; the Meter beeps.
l Press HOLD H again orRANGE or Hz % or turn the rotary
switch to exit Hold mode; the Meter beeps.
l In Hold mode, isdisplayed.
Warning
PressingHOLD H in REL mode makes the Meter stop
updating. PressingHOLDH again to resume updating.
To enter or exit REL mode:
l Use rotary switch to select the measurement function before
selectingREL . If measurement function changes manually
after REL
is selected, the Meter exits the REL mode.
l PressREL
to enter REL mode, auto ranging turns off except
under capacitance testing mode, and the present measurement
range is locked and display “0” as the stored value.
l PressREL
again or turn the rotary switch to reset the
stored value and exit REL mode.
The Use of Relative Value Mode
The REL mode applies to all measurement functions except
frequency/duty cycle measurement. It subtracts a stored value
from the present measurement value and displays the result.
For instance, if the stored value is 20.0V and the present
measurement value is 22.0V, the reading would be 2.0V. If a
new measurement value is equal to the stored value then display
0.0V.
24
Model 72-7740/72-7745: OPERATING MANUAL
The POWER button
This is a self-lock switch used to turn on or off the power of the
Meter.
The BLUE button
It is used for selecting the required measurement function when
more than one function exists at one position of the rotary
switch.
Turning on the Display Backlight
In order to avoid the hazard arising from mistaken readings
in insufficient light or poor vision, please use Display
Backlight function.
l Press and hold HOLD H
for over 2 seconds to turn the
Display Backlight on.
l Pressand hold HOLD H again for over 2 seconds to turn the
Display Backlight off, otherwise it will stay on continuously.
Sleep Mode (Model 72-7740)
To preserve battery life, the Meter automatically turns off after
30 minutes of inactivity.
To disable the Sleep Mode function, press BLUE button while
turning on the Meter.
Warning
25
Model 72-7740/72-7745: OPERATING MANUAL
Temperature: Operating : 0
oC~40
oC (32
oF ~104
oF).
Storage : -10
oC~50
oC (14
oF~122
oF).
: 75% @ 0
oC - 30
oC; 50% @ 31
oC - 40
oC;
UL & CUL
General Specifications
l Maximum Voltage between
any Terminals and Grounding:
1000V.
l
Fused Protection for
Input Terminal:
l
Fused Protection for 10AInput Terminal:
Glass fuse, 10A, 250V, fast type, 5x20mm.
Glass fuse, 0.5A, 250V, fast type, 5x20mm.
l Maximum Display: Digital: 3999
l Measurement Speed: Updates 3 times/second.
l
l Relative Humidity:
l Altitude: Operating: 2000 m.
Storage : 10000 m.
l Battery Type: One piece of 9V (NEDA1604 or 6F22 or
006P).
l Low Battery : Display
l Dimensions (HxWxL): 177 x 85 x 40 mm.
l Weight: Approximate 300g (battery included).
l Safety/Compliances: IEC61010 CAT.III 1000V,
CAT.IV 600V overvoltage and double insulation standard.
l Certifications: ,
26
Model 72-7740/72-7745: OPERATING MANUAL
Accuracy: (a% reading + b digits), guarantee for 1 year.
Operating temperature: 23
oC 5
oC.
Relative humidity: 75%.
Temperature coefficient: 0.1 x (specified accuracy)/1
oC
A. DC Voltage
Range
Resolution
Accuracy
Overload
Protection
400mV
0.1mV
(0.8%+3)
4V
40V
400V
(0.8%+1)
1000V
1V
(1%+3)
1000V DC
750V AC rms
continuous.
1mV
10mV
100mV
l Frequency response 40Hz~400Hz.
400
0.1
1
4k
40k
10
400k
100
4M
1k
Accuracy Specifications (1)
Remarks: Input impedance
C. Resistance
Range
Resolution
Accuracy
Overload
Protection
(1%+2)
(1.2%+2)
(1.5%+2)
1000Vp
Remarks: Open circuit voltage approximate 0.45V.
40M
10k
Measure at
REL mode
(1.2%+2)
B. AC Voltage
Range Resolution Accuracy
Overload
Protection
4V
40V
400V
(1%+5)
750V
1V
(1.2%+5)
1000V DC
750V AC rms
continuous.
Remarks:
l Input impedance
10M
.
l Model 72-7740: displays effective value of sine wave
(mean value response).
Model 72-7745: displays true rms value.
100mV
10mV
1mV
27
Model 72-7740/72-7745: OPERATING MANUAL
D Continuity Test
Range
Resolution
Accuracy
Overload
Protection
400.0
0.1
Approximate 100
1000Vp
Remarks:
l Buzzer beeps continuously.
l Open circuit voltage approximate 0.45V.
E. Diode Test
Range Resolution
Overload Protection
Diode 1mV 1000Vp
Remarks:
l Open circuit voltage approximate 1.48V.
l Displays approximate forward voltage drop reading
0.5V~0.8V.
F. Capacitance
Resolution
Accuracy
Overload
Protection
Measure at REL mode
(3%+10)
(3%+5)
(4%+5)
1000Vp
G. Frequency & Duty Cycle
Range
Resolution Accuracy
Overload
Protection
10Hz~10MHz
0.1%~99.9% 0.01%
1000Vp
Remarks:
l 10Hz~10MHz Range:
l 0.1%~99.9%:
Reading is only for reference purpose.
Accuracy Specifications (2)
300mV rms input sensitivity 30V rms;
600mV rms input sensitivity 30V rms.
1MHz:
1MHz:
Range
(0.1%+3)
28
Model 72-7740/72-7745: OPERATING MANUAL
For continuous measurement 10 seconds and interval not
less than 15 minutes.
I. DC Current
Glass fuse 0.5A, 250V,
fast type Glass fuse, 5x20 mm.
Glass fuse 10A, 250V,
fast type Glass fuse, 5x20 mm.
Overload Protection:
Glass fuse 0.5A, 250V, fast type, 5x20mm.
Remarks:
Accuracy Specifications (3)
400 A
0.1 A
l 4A & 10A Range:
29
Model 72-7740/72-7745: OPERATING MANUAL
fast type Glass fuse, 5x20 mm.
10A, 250V,
fast type Glass fuse, 5x20 mm.
Accuracy Specifications (4)
J. AC Current
400 A
0.1 A
30
and interval not less than 15 minutes.
Remarks:
Model 72-7740 : displays effective value of sine wave
(mean value response).
Model 72-7745 : displays true rms value.
l 4A & 10A Range:
For continuous measurement 10 seconds
Model 72-7740/72-7745: OPERATING MANUAL
to the10A terminal.
MAINTENANCE
This section provides basic maintenance information including
battery and fuse replacement instruction.
performance test, and service information.
water inside the case.
field.
To test the fuse:
Do not attempt to repair or service your Meter unless you
are qualified to do so and have the relevant calibration,
Warning
To avoid electrical shock or damage to the Meter, do not get
A. General Service
l
l Periodically wipe the case with a damp cloth and mild
detergent. Do not use abrasives or solvents.
l To clean the terminals with cotton bar with detergent, as dirt
or moisture in the terminals can affect readings.
l Turn off the power of the Meter when it is not in use and take
out the battery when not using for a long time.
Take out the battery when it is using for a long time.
l Do not use or store the Meter in a place of humidity, high
temperature, explosive, inflammable and strong magnetic
B. Testing the Fuses
fuses with identical amperage, voltage, and speed ratings.
To avoid electrical shock or personal injury, remove the test
leads and any input signals before replacing the battery or
fuse.
To prevent damage or injury, install ONLY replacement
Warning
1. Set the rotary switch to
2. Plug a test lead into the terminalHzV
and touch the probe
and pressBLUE
button to select
Model 72-7740/72-7745: OPERATING MANUAL
If the Meter does not work while the fuse is all right, send it to
your dealer for repair.
C. Replacing the Battery (see figure 12)
To avoid false readings, which could lead to electric
shock or personal injury, replace the battery as soon as the
battery indicator “
”
appears.
Make sure the test leads are disconnected from the circuit
being tested before opening the case bottom.
To replace the battery:
1.
2.
3.
4.
5.
Press the POWER to turn the Meter off and remove all
connections from the terminals.
Remove the screw from the battery compartment, and
separate the battery compartment from the case bottom.
Remove the battery from the battery compartment.
Replace the battery with a new 9V battery (NEDA1604,
6F22 or 006P)
Rejoin the case bottom and battery compartment, and
reinstall the screw.
Warning
l If the Meter beeps, the fuse is good.
l If the display shows
, replace the fuse and test again.
l If the display shows any other value, have the Meter serviced
and contact your dealer immediately.
(figure 12)
possible
32
Model 72-7740/72-7745: OPERATING MANUAL
Fuse 1: Glass fuse 0.5A, 250V, fast type, 5x20mm.
Fuse 2: Glass fuse 10A, 250V, fast type, 5x20mm.
To avoid electrical shock, arc blast, personal injury or
damage to the Meter, use specified fuses ONLY in
accordance with the following procedure.
To replace the Meter’s fuse:
Press the POWER to turn the Meter off and remove all
connections from the terminals.
Remove the screw from the battery compartment, and
separate the battery compartment from the case bottom.
Remove the 2 rubber feet and 2 screws from the case bottom,
and separate the case top from the case bottom.
Remove the fuse by gently prying one end loose, then take
out the fuse from its bracket.
Install ONLY replacement fuses with the identical type and
specification as follows and make sure the fuse is fixed
firmly in the bracket.
Warning
(figure 13)
1.
2.
3.
4.
5.
D. Replacing the Fuses (see figure 13)
33
Model 72-7740/72-7745: OPERATING MANUAL
always results from improper operation.
Replacement of the fuses is seldom required. Burning of a fuse
6.
7. Rejoin the case bottom and case top, and reinstall the 2
screws and 2 rubber feet.
Rejoin the battery compartment and the case top, and
reinstall the screw.
34
Model 72-7740/72-7745: OPERATING MANUAL
Tenma Test Equipment
405 S. Pioneer Blvd.
Springboro,Ohio 45066
www.tenma.com
Model 72-7740/72-7745: OPERATING MANUAL