Tenma 72-7765 Operating Manual
Text of the manual
500mA/250V fast type
fuse φ 5x20mm
0.1µA
1µA
10µA
100µA
1mA
10mA
(2%+5)
(2.5%+5)
D. AC Current
Un-Fused
Remark:
l At 10A Range: For continuous measurement 10 seconds and interval not less
than 15 minutes.
l Input impedance: approx. 10MΩ.
l Measurement voltage drop: Full range at 400mV.
Model 72-7765: OPERATING MANUAL
250V DC or AC
Model 72-7765: OPERATING MANUAL
G. Transistor hFE Test
Ibo 10µΑ
Vce 1.5V
hFE
250V DC or AC
F. Diodes and Continuity Measurement
Displays approximate forward
voltage drop
Buzzer beeps at <100Ω
1mV
0.1Ω
Test NPN and PNP type transistors.
Display range: 0-1000β
Model 72-7765: OPERATING MANUAL
This section provides basic maintenance information including battery and fuse
replacement instructions.
Warning
Do not attempt to repair or service your Meter unless you are qualified to do
so and have the relevant calibration, performance test, and service information.
To avoid electrical shock or damage to the Meter, do not get water inside the
case.
A. General Service
l Periodically wipe the case with a damp cloth and mild detergent. Do not use
abrasives or solvents.
l To clean the terminals with cotton bar with detergent, as dirt or moisture in the
terminals can affect readings.
l Turn the Meter to OFF position when it is not in use and take out the battery
when not using for a long time.
l Do not store the Meter in a place of humidity, high temperature, explosive,
inflammable and strong magnetic field.
Model 72-7765: OPERATING MANUAL
B. Replacing the Battery (see figure 8)
Warning
To avoid false readings, which could lead to
possible electric shock or personal injury,
replace the battery as soon as the battery
indicator “ ”appears.
To replace the battery:
1. Disconnect the connection between the test
leads and the circuit under test, and remove
the test leads from the input terminals of the
Meter.
2. Turn the Meter to OFF position.
3. Remove the screw from case bottom, and
separate the case bottom from the case top.
(figure 8)
SCREW
Model 72-7765: OPERATING MANUAL
C. Replacing the Fuses (see figure 8)
Warning
To avoid electrical shock,arc blast, personal injury or damage to the Meter,
use specified fuses ONLY in accordance with the following procedure.
To replace the Meter’s fuse:
1. Disconnect the connection between the test leads and the circuit under test,
and remove the test leads from the input terminals of the Meter.
2. Turn the Meter to OFF position.
3. Remove the screws from case bottom, and separate the case bottom from the
case top.
Maintenance(4)
4. Remove the battery from the battery compartment.
5. Replace the battery with a new 1.5V AAA battery.
6. Rejoin the case bottom and case top, and reinstall
the screw.
Model 72-7765: OPERATING MANUAL
** END **
39
4. Remove the fuse by gently prying one end loose, and then take out the fuse
from its bracket.
6. Rejoin the case bottom and case top, and reinstall the screw.
Replacement of the fuses is seldom required. Burning of a fuse always results from
improper operation.
5. Install ONLY replacement fuses with the identical type and specification as
follows and make sure the fuse is fixed firmly in the bracket.
Fuse 1:500mA 250 V, fast type, 5x20mm.
Maintenance(5)
Model 72-7765: OPERATING MANUAL
Model 72-7765: OPERATING MANUAL
Tenma Test Equipment
405 S. Pioneer Blvd.
Springboro,Ohio 45066
www.tenma.com
C. DC Current Measurement (see figure 4)
Warning
Never attempt an in-circuit current measurement
where the voltage between terminals and ground
is greater than 60V . If the fuse burns out during
measurement, the Meter may be damaged or the
operator may be injured. Use proper terminals,
function, and range for the measurement. When
the test leads are connected to the current
terminals, do not parallel them across any circuit.
Measurement Operation (4)
(figure 4)
To measure current, do the following:
1. Turn off power to the circuit. Discharge all high-
voltage capacitors.
black red
17
Model 72-7765: OPERATING MANUAL
Measurement Operation (5)
2. Insert the red test lead into the VΩmA or 10A terminal and the black test lead
into the COM terminal.
3. Set the rotary switch to an appropriate measurement position in A range. DC
measurement is default or press SELECT button to select DC measurement mode.
4. Break the current path to be tested. Connect the red test lead to the more positive
side of the break and the black test lead to the more negative side of the break.
5. Turn on power to the circuit. The measured value shows on the display.
Note
l If the value of current to be measured is unknown, use the maximum measurement
position (10A) and reduce the range step by step until a satisfactory reading is
obtained.
l When current measurement has been completed, disconnect the connection
between the testing leads and the circuit under test.
18
Model 72-7765: OPERATING MANUAL
Measurement Operation (6)
19
D. AC Current Measurement (see figure 4)
Warning
Never attempt an in-circuit current measurement where the open-circuit voltage
between the circuit and ground is greater than 500V.
If the fuse burns out during measurement, the Meter may be damaged or the
operator may be injured. Use proper terminals function, and range for the
measurement. When the test leads are connected to the current terminals,
do not parallel them across any circuit.
1. Turn off power to the circuit. Discharge all high-voltage capacitors.
2. Turn the rotary switch to suitable range, press SELECT to change from DC to AC
measurement.
3. Insert the red test lead into the VΩmA or 10A terminal and the black test lead
into the COM terminal.
4. Set the rotary switch to an appropriate measurement position in A range.
Model 72-7765: OPERATING MANUAL
Measurement Operation (7)
20
5. Break the current path to be tested. Connect the red test lead to the more positive
side of the break and the black test lead to the more negative side of the break.
6. Turn on power to the circuit.
The measured value shows on the display.
Model 72-7765: OPERATING MANUAL
Measurement Operation (8)
(figure 5)
black red
21
E. Resistance Measurement (see figure 5)
Warning
To avoid damage to the Meter or to the devices
under test, disconnect circuit power and discharge
all the high-voltage capacitors before measuring
resistance.
The Ω ranges are : 400.0Ω, 4.000kΩ, 40.00kΩ,
400.0kΩ, 4.000MΩ, and 40.00MΩ,
1. Insert the red test lead into the VΩmA terminal and
the black test lead into the COM terminal.
2. Set the rotary switch to the Ω range.
3. Connect the test leads across with the object being
measured. The measured value shows on the display.
Model 72-7765: OPERATING MANUAL
Measurement Operation (9)
22
Note
l The test leads can add 0.1Ω to 0.3Ω of error to resistance measurement. To
obtain precision readings in low-resistance measurement, that is the range of
200Ω, short-circuit the input terminals beforehand and record the reading
obtained (called this reading as X). (X) Is the additional resistance from the test
lead. Then use the equation:
measured resistance value (Y) - (X) = precision
readings of resistance.
l For high-resistance measurement (>1MΩ), it normally requires several seconds
to obtain a stable reading.
l When resistance measurement has been completed, disconnect the connection
between the testing leads and the circuit under test.
Model 72-7765: OPERATING MANUAL
Measurement Operation (10)
(figure 6)
23
F. Diode and Continuity Test (see figure 6)
Testing Diodes
Warning
To avoid damage to the Meter or to the
devices under test, disconnect circuit power
and discharge all the high-voltage capacitors
before diodes.
Use the diode test to check diodes, transistors,
and other semiconductor devices. The diode test
sends a current through the semiconductor junction,
and then measures the voltage drop across the junction.
A good silicon junction drops between 0.5V and 0.8V.
black red
Model 72-7765: OPERATING MANUAL
Measurement Operation (11)
24
To test a diode out of a circuit, connect the Meter as follows:
1. Insert the red test lead into the VΩmA terminal and the black test lead into the
COM terminal.
2. Set the rotary switch to Diode measurement is default or press SELECT
button to select diode measurement mode.
3. For forward voltage drop readings on any semiconductor component, place the
red test lead on the component’s anode and place the black test lead on the
component’s cathode.
The measured value shows on the display.
Note
l In a circuit, a good diode should still produce a forward voltage drop reading of
0.5V to 0.8V; however, the reverse voltage drop reading can vary depending on
the resistance of other pathways between the probe tips.
l Connect the test leads to the proper terminals as said above to avoid error display.
The LCD will display “OL” indicating open-circuit for wrong connection. The unit
of diode is Volt (V), displaying the positive-connection voltage-drop value.
Model 72-7765: OPERATING MANUAL
Measurement Operation (12)
25
l When diode testing has been completed, disconnect the connection between the
testing leads and the circuit under test.
Testing for Continuity
To test for continuity, connect the Meter as below:
1. Insert the red test lead into the VΩ mA terminal and the black test lead into the
COM terminal.
2. Set the rotary switch to .
3. Press SELECT button to select continuity measurement mode.
4. Connect the test leads across with the object being measured.
The buzzer sounds if the resistance of a circuit under test is less than 100Ω.
Model 72-7765: OPERATING MANUAL
Measurement Operation (13)
(figure 7)
26
G. Transistor hFE Measurement (see figure 7)
Warning
To avoid damages to the Meter or to the devices
under test, do not input any current over 60V DC
or 30V AC.
1. Check that the transistor is PNP or NPN type.
2. Connect the transistor to be measured to the
corresponding jacks.
3. LCD display hFE reference value.
Model 72-7765: OPERATING MANUAL
To preserve battery life, the Meter automatically turns off after 30minutes of inacjivity.
The Meter can be activated by pressing the SELECT button or turning the rotary
switch.
Sleep Mode
Model 72-7765: OPERATING MANUAL
l Maximum Voltage between any Terminals and Ground: 500V rms.
l
Fused Protection for VΩmA Input Terminal: 500mA, 250V fast type, φ5x20 mm
l
10A Terminal: Un-fused.
l Range: Auto ranging
l Maximum Display: Display: 3999.
l Measurement Speed: Updates 3 times / second.
l Temperature:
Operating: 0
oC~40
oC (32
oF~104
oF).
Storage: -10
oC~50
oC (14
oF~122
oF).
l Relative Humidity: 75% @ 0
oC - 30
oC; 50% @ 31 - 40
oC.
l Altitude: Operating: 2000 m.
Storage: 10000 m.
General Specifications(1)
Model 72-7765: OPERATING MANUAL
l Battery Type: Two piece of 1.5V AAA Battery.
l Low Battery : Display: .
l Negative reading: Display: “-“.
l Overloading: Display: “0L”.
l Dimensions (HxWxL): 130 x 73.5 x 35mm.
l Weight: Approx. 156g (battery included).
l Safety/Compliances: IEC61010 CAT I 600V, CAT II 300V overvoltage and double
insulation standard.
l Certification:
Model 72-7765: OPERATING MANUAL
A. DC Voltage
Remark: Input impedance: 10MΩ.
100µV
1mV
10mV
100mV
1V
Accuracy: (a% reading + b digits), guarantee for 1 year.
Operating temperature: 23
oC -5
oC.
Relative humidity: <75%.
Range Resolution
400mV
4V
40V
400V
500V
Accuracy
Overload Protection
(0.8%+3)
(0.8%+1)
500V DC or AC
(1%+3)
Model 72-7765: OPERATING MANUAL
500V DC or AC
1mV
10mV
100mV
1V
(1.5%+5)
B. AC Voltage
Remarks:
l Input impedance: approx. 10MΩ.
l Displays effective value of sine wave (mean value response).
l Frequency response 40Hz ~ 400Hz.
Model 72-7765: OPERATING MANUAL
C. DC Current
Remark:
l At 10A Range: For continuous measurement 10 seconds and interval not less
than 15 minutes.
l Input impedance: approx. 10MΩ.
l Measurement voltage drop: Full range at 400mV.
Range Resolution
400µA
4000µA
40mA
400mA
4A
10A
Accuracy
Overload Protection
(1%+2)
500mA/250V fast type
fuse φ 5x20mm
0.1µA
1µA
10µA
100µA
1mA
10mA
(1.2%+2)
(1.5%+5)
Un-Fused
Model 72-7765: OPERATING MANUAL
Rules For Safe Operation
International Electrical Symbols
The Meter Structure
Functional Buttons
LCD Display
Measurement Operation
A.
DC Voltage Measurement
B.
AC Voltage Measurement
C. DC Current Measurement
D.
AC Current Measurement
E.
Resistance Measurement
F.
Diode and Continuity Test
Table of Contents (1)
G. Transistor hFE Measurement
Model 72-7765: OPERATING MANUAL
Table of Contents (2)
Title
Sleep Mode
General Specifications
Accuracy Specifications
A.
DC Voltage
B.
AC Voltage
C.
DC Current
D.
AC Current
E.
Resistance
F.
Diodes and Continuity
Measurement
G.
Transistor hFE Test
Maintenance
A. General Service
B. Replacing the Battery
C. Replacing the Fuses
Page
27
28
30
30
31
32
33
34
35
35
36
36
37
38
Model 72-7765: OPERATING MANUAL
This Operating Manual covers information on safety and cautions. Please read
the relevant information carefully and observe all the Warnings and Notes strictly.
Warning
To avoid electric shock or personal injury, read the “ Safety Information” and
“ Rules for Safety Operation” carefully before using the Meter.
The Model 72-7765 Multimeter (hereafter referred as “the Meter”) is a 3 3/4 digits
with steady operations, fashionable design and highly reliable hand-held measuring
instrument. The Meter can measure AC/DC voltage, AC/DC Current, Resistance,
Transistor hFE, Diode and Continuity. It is an ideal tool for maintenance.
Model 72-7765: OPERATING MANUAL
Open the package case and take out the Meter. Check the following items carefully
for missing or damaged parts:
In the event you find items missing or damaged, please contact your dealer
immediately.
Unpacking Inspection
Item
1
2
3
Description
Qty
1 piece
1 pair
1 piece
English Operating Manual
Test Lead
Holster
Model 72-7765: OPERATING MANUAL
This Meter complies with the standards IEC61010: in pollution degree 2, overvoltage
category (CAT I 600V, CAT II 300V) and double insulation.
CAT. I: Signal level, special equipment or parts of equipment, telecommunication,
electronic, etc., with smaller transient overvoltages than overvoltages CAT. II.
CAT. II: Local level, appliance, PORTABLE EQUIPMENT etc., with smaller transient
overvoltages than CAT. III
Use the Meter only as specified in this operating manual, otherwise the protection
provided by the Meter may be impaired.
Model 72-7765: OPERATING MANUAL
In this manual, a Warning identifies conditions and actions that pose hazards to the
user, or may damage the Meter or the equipment under test. A Note identifies the
information that user should pay attention to.
International electrical symbols used on the Meter and in this Operating Manual are
explained on page 10.
Model 72-7765: OPERATING MANUAL
To avoid possible electric shock or personal injury, and to avoid possible
damage to the Meter or to the equipment under test, adhere to the following
rules:
l Before using the Meter inspect the case. Do not use the Meter if it is damaged
or the case (or part of the case) is removed. Look for cracks or missing plastic.
Pay attention to the insulation around the connectors.
l Inspect the test leads for damaged insulation or exposed metal. Check the
test leads for continuity. Replace damaged test leads with identical model
number or electrical specifications before using the Meter.
l Do not apply more than the rated voltage, as marked on the Meter, between
the terminals or between any terminal and grounding.
l The rotary switch should be placed in the proper position, and no
change of range shall be made during measurement to prevent damage
of the Meter.
Rules For Safe Operation (1)
Model 72-7765: OPERATING MANUAL
l When the Meter is working at an effective voltage over 60V DC or 42V
rms AC, special care should be taken for there is danger of electric shock.
l Use the proper terminals, function, and range for your measurements.
l Do not use or store the Meter in an environment of high temperature,
humidity, explosive, flammable and strong magnetic field. The performance
of the Meter may deteriorate after dampened.
l When using the test leads, keep your fingers behind the finger guards.
l Disconnect circuit power and discharge all high-voltage capacitors before
testing resistance, continuity, diodes and current.
l Before measuring current, check the Meter’s fuses and turn off power to
the circuit before connecting the Meter to the circuit.
l Replace the battery as soon as the battery indicator “ ” appears. With a low
battery, the Meter might produce false readings that can lead to electric
shock and personal injury.
Rules For Safe Operation (2)
Model 72-7765: OPERATING MANUAL
l Remove test leads from the Meter and turn the Meter power off before
opening the case.
l When servicing the Meter, use only the same model number or identical
electrical specifications replacement parts.
l The internal circuit of the Meter shall not be altered at will to avoid damage
of the Meter and any accident.
l Soft cloth and mild detergent should be used to clean the surface of the
Meter when servicing. No abrasive and solvent should be used to prevent
the surface of the Meter from corrosion, damage and accident.
l The Meter is suitable only for indoor use.
l Turn the Meter off when it is not in use and take out the battery when not
using for a long time.
l Periodically check the battery as it may leak after some time.If leakage is
apparent,the battery should be immediately replaced to prevent damage.
Rules For Safe Operation (3)
Model 72-7765: OPERATING MANUAL
International Electrical Symbols
Continuity Test
AC Current
DC Current
Double Insulated.
Low Battery
Diode.
Fuse.
Warning. Refer to the Operating Manual
Conforms to Standards of European Union.
Ground
AC or DC
Model 72-7765: OPERATING MANUAL
The Meter Structure (figure 1)
1) LCD Display
2) SELECT Button
3) Rotary Switch
4) COM Input Terminal
5) 10A Input Terminal
6) Other Input Terminal
7) Transistor Input Terminal
The Meter Structure
(figure 1)
Model 72-7765: OPERATING MANUAL
Selecting for DC and AC current measurement, Continuity Test and Diode Test. The
buzzer sounds when switching from one function to the other.Press this button to
activate the Meter when it is under Sleep Mode.
Model 72-7765: OPERATING MANUAL
Measurement Operation (1)
Warning
To avoid harm to you or damage to the Meter
from electric shock, please do not attempt to
measure voltages higher than 500V although
readings may be obtained.
The DC Voltage ranges are : 400.0mV, 4.000V,
40.00V, 400.0V and 500V
1. Insert the red test lead into the VΩmA terminal
and the black test lead into the COM terminal.
2. Set the rotary switch to V range.
3. Connect the test leads across with the object being
measured.
The measured value shows on the display.
(figure 3)
A. DC Voltage Measurement (see figure 3)
black red
Model 72-7765: OPERATING MANUAL
Measurement Operation (2)
l DCV Measurement is autoranging, the Meter has an input impedance of approx.
10MΩ. This loading effect can cause measurement errors in high impedance
circuits. If the circuit impedance is less than or equal to 10kΩ, the error is
negligible (0.1% or less).
l When DC voltage measurement has been completed, disconnect the connection
between the testing leads and the circuit under test.
B. AC Voltage Measurement (see figure 3)
Warning
To avoid harm to you or damage to the Meter from electric shock, please
do not attempt to measure voltages higher than 500V although readings may
be obtained.
The AC voltage ranges are : 4.000V, 40.00V, 400.0V and 500V
1. Insert the red test lead into the VΩmA terminal and the black test lead into the
COM terminal.
Note:
Model 72-7765: OPERATING MANUAL
Measurement Operation (3)
2. Set the rotary switch to V range.
3. Connect the test leads across with the object being measured.
The measured value shows on the display.
l DCV Measurement is autoranging, the Meter has an input impedance of approx.
10MΩ. This loading effect can cause measurement errors in high impedance
circuits. If the circuit impedance is less than or equal to 10kΩ, the error is negligible
(0.1% or less).
l When DC voltage measurement has been completed, disconnect the connection
between the testing leads and the circuit under test.
Note:
Model 72-7765: OPERATING MANUAL