ABB 4623 Models 4623 and 4628

Conductivity Transmitters To Meet United States Pharmacopœia (USP 23) Regulations 4600 µS/cm 25°C 0 .50 A1 A2 4600 0 . 50 A1 A2 µS/cm 25°C

ABB Automation is an established world force in the design and manufacture of

instrumentation for industrial process control, flow measurement, gas and liquid analysis and environmental applications. As a part of ABB, a world leader in process automation technology, we offer customers application expertise, service and support worldwide. We are committed to teamwork, high quality manufacturing, advanced technology and unriva

INTRODUCTION .......................................................... 2

2 PREPARATION ............................................................ 2 2.1 Checking the Code Number .............................. 2 2.1.1 Wall-/Pipe-mounted Instruments ......... 2 2.1.2 Panel-mounted Instruments ................ 2 2.1.3 Conductivity Cells ................................ 3 3 MECHANICAL INSTALLATION ..................................4

The 4623 and 4628 conductivity transmitters and associated

measuring cells have been designed to meet United States Pharmacopœia (USP 23) requirements for continuous monitoring and control of the conductivity of demineralised water, and de-ionised water. The 4623 model is a wall-mounted instrument and the 4628 model a panel-mounted, DIN sized instrument. Both instruments have a single programmable conductivity input channel and a single temperature inpu

ABB 4623 Conductivity Cells – Fig. 2.3

Model 2278/305 Basic Type No. Version Constant (K) Process Connection Type Temperature Compensation Code Characters 1, 2 3, 4, 5 6 7 8 22 Electrolytic conductivity measuring cells 0 2in. Tri-clover Hygienic fitting 0 None 5 Pt100 resistance thermometer 78/ Stainless steel 3 0.05 2 PREPARATION Fig. 2.3 Checking the Conductivity Cell Code Number

Instruments – Fig. 3.1

Caution. • Mount in a location free from excessive vibration. • Mount away from harmful vapours and/or dripping fluids. Information. It is preferable to mount the instrument at eye level, allowing an unrestricted view of the front panel displays and controls 3.1.2 Conductivity Cells – Fig. 3.2 Caution. Ensure that the integral cable (where applicable) does not hang against hot or abrasive objec

ABB 4623 Wall-/Pipe-mounted Instruments – Figs. 3.3 and 3.4

Mark fixing centres (see Fig. 3.3) Drill suitable holes Fix instrument to wall using suitable fixing 1 2 3 Position ‘U’ bolts on pipe Position plates over ‘U’ bolts Secure transmitter to mounting plate Secure plates 1 2 3 4 160 (6.3) 250 (9.84) 69 (2.72) Fixing Centres Allowance for Cable Bends 200 (7.9) 68 (2.68) 214 (8.43) 232 (9.13) Fixing Centres 42 (1.65) 61 (23/8) O.D.