ABB 4623 Models 4623 and 4628
Conductivity Transmitters
To Meet United States
Pharmacopœia (USP 23)
Regulations
4600
µS/cm 25°C
0 .50
A1
A2
4600
0 . 50
A1
A2
µS/cm 25°C
ABB Automation is an established world force in the design and manufacture of
instrumentation for industrial process control, flow measurement, gas and liquid analysis and
environmental applications.
As a part of ABB, a world leader in process automation technology, we offer customers
application expertise, service and support worldwide.
We are committed to teamwork, high quality manufacturing, advanced technology and
unriva
INTRODUCTION .......................................................... 2
2
PREPARATION ............................................................ 2
2.1
Checking the Code Number .............................. 2
2.1.1
Wall-/Pipe-mounted Instruments ......... 2
2.1.2
Panel-mounted Instruments ................ 2
2.1.3
Conductivity Cells ................................ 3
3
MECHANICAL INSTALLATION ..................................4
The 4623 and 4628 conductivity transmitters and associated
measuring cells have been designed to meet United States
Pharmacopœia (USP 23) requirements for continuous
monitoring and control of the conductivity of demineralised
water, and de-ionised water.
The 4623 model is a wall-mounted instrument and the 4628
model a panel-mounted, DIN sized instrument. Both
instruments have a single programmable conductivity input
channel and a single temperature inpu
ABB 4623 Conductivity Cells – Fig. 2.3
Model 2278/305
Basic Type No.
Version
Constant (K)
Process
Connection
Type
Temperature
Compensation
Code Characters
1, 2
3, 4, 5
6
7
8
22
Electrolytic
conductivity
measuring
cells
0
2in. Tri-clover
Hygienic fitting
0
None
5
Pt100
resistance
thermometer
78/ Stainless steel
3
0.05
2
PREPARATION
Fig. 2.3 Checking the Conductivity Cell Code Number
Instruments – Fig. 3.1
Caution.
•
Mount in a location free from excessive vibration.
•
Mount away from harmful vapours and/or dripping fluids.
Information. It
is
preferable
to
mount
the
instrument at eye level, allowing an unrestricted view of
the front panel displays and controls
3.1.2
Conductivity Cells – Fig. 3.2
Caution. Ensure that the integral cable (where
applicable) does not hang against hot or abrasive objec
ABB 4623 Wall-/Pipe-mounted Instruments – Figs. 3.3 and 3.4
Mark fixing centres
(see Fig. 3.3)
Drill suitable
holes
Fix instrument to
wall using
suitable fixing
1
2
3
Position ‘U’ bolts on pipe
Position plates over ‘U’ bolts
Secure transmitter to mounting plate
Secure plates
1
2
3
4
160 (6.3)
250
(9.84)
69 (2.72)
Fixing Centres
Allowance for
Cable Bends
200 (7.9)
68 (2.68)
214
(8.43)
232
(9.13)
Fixing Centres
42
(1.65)
61 (23/8) O.D.