Fluke 5500A Getting Started

Text of Fluke 5500A Getting Started

© 2001 Fluke Corporation. All rights reserved. Printed in U.S.A.

All product names are trademarks of their respective companies.

LIMITED WARRANTY & LIMITATION OF LIABILITY

Each Fluke product is warranted to be free from defects in material and workmanship under normal use and service. The warranty period is one year and begins on the date of shipment. Parts, product repairs and services are warranted for 90 days. This warranty extends only to the original buyer or end-user customer of a Fluke authorized reseller, and does not apply to fuses, disposable batteries or to any product w

LIMITE DE GARANTIE ET LIMITE DE RESPONSABILITE

La société Fluke garantit l'absence de vices des matériaux et à la fabrication de ce produit dans des conditions normales d'utilisation et d'entretien. La période de garantie est de un an et prend effet à la date d'expédition. Les pièces, les réparations de produit et les services sont garantis pour un période de 90 jours. Cette garantie ne s'applique qu'à l'acheteur d'origine ou à l'utilisateur fi

Fluke 5500A BEFRISTETE GARANTIEBESTIMMUNGEN & HAFTUNGSBESCHRÄNKUNG

Für jedes Produkt, das Fluke herstellt, leistet Fluke eine Garantie für einwandfreie Materialqualitßt und fehlerfreie Ausführung unter normalen Betriebs- und Wartungsbedingungen. Der Garantiezeitraum gilt für ein Jahr und beginnt mit dem Lieferdatum. Die Garantiebestimmungen für Ersatzteile, Instandsetzungs- und Wartungsarbeiten gelten für einen Zeitraum von 90 Tagen. Diese Garantie wird aussch

GARANTÍA LIMITADA Y LIMITACIÓN DE RESPONSABILIDAD

Se garantiza que cada uno de los productos de Fluke no tiene defectos de material y mano de obra si es objeto de una utilización y un mantenimiento normales. El período de garantía es de un año y comienza a partir de la fecha de envío. Las piezas, reparaciones y mantenimiento del producto están garantizados durante 90 días. Esta garantía se concede exclusivamente al comprador original o al cliente

This is an IEC safety Class 1 product. Before using, the ground wire in the

line cord or rear panel binding post must be connected to an earth ground for safety. Interference Information This equipment generates and uses radio frequency energy and if not installed and used in strict accordance with the manufacturer’s instructions, may cause interference to radio and television reception. It has been type tested and found to comply with the limits

SAFETY TERMS IN THIS MANUAL

This instrument has been designed and tested in accordance with IEC publication 1010-1 (1992-1), Safety Requirements for Electrical Measuring, Control and Laboratory Equipment, and ANSI/ISA-582.01-1994, and CAN/CSA-C22.2 No. 1010.1-92. This User Manual contains information, warning, and cautions that must be followed to ensure safe operation and to maintain the instrument in a safe condition. Use of this equipment in a manner no

USE THE PROPER POWER CORD

Use only the power cord and connector appropriate for the voltage and plug configuration in your country. Use only a power cord that is in good condition. Refer power cord and connector changes to qualified service personnel. DO NOT OPERATE IN EXPLOSIVE ATMOSPHERES To avoid explosion, do not operate the instrument in an atmosphere of explosive gas. DO NOT REMOVE COVER DURING OPERATION To avoid personal injury or death, do not remo

Fluke 5500A Table of Contents

Title Page Introduction ....................................................................................................... 1 Instruction Manuals........................................................................................... 2 5500A Getting Started Manual ..................................................................... 3 5500A Operator Manual ..........................................................................

5500A Phase Specifications.......................................................................... 27

Calculating Power Uncertainty ..................................................................... 28 Additional Specifications .................................................................................. 29 Frequency Specifications .............................................................................. 29 Harmonics

The Fluke Model 5500A Multi-Product Calibrator (Figure 1) is a precise instrument that

calibrates a wide variety of electrical measuring instruments. With the 5500A Calibrator, you can calibrate precision multimeters that measure ac or dc voltage, ac or dc current, ac or dc power, resistance, capacitance, and temperature. With the Oscilloscope Calibration option, you can use the 5500A Calibrator to calibrate analog and digita

Fluke 5500A The 5500A Calibrator is a fully programmable precision source of the following:

• DC voltage from 0 V to +1020 V. • AC voltage from 1 mV to 1020 V, with output from 10 Hz to 500 kHz. • AC current from 0.01 µA to 11.0 A, with output from 10 Hz to 10 kHz. • DC current from 0 to +11.0 A. • Resistance values from a short circuit to 330 MΩ. • Capacitance values from 330 pF to 1100 µF. • Simulated output for three types

Where to Go from Here

3 5500A Getting Started Manual This 5500A Getting Started Manual contains a brief introduction to the 5500A Manual Set, instructions on how to get your calibrator prepared for operation and a complete set of specifications. 5500A Operator Manual The 5500A Operator Manual provides complete information for installing the 5500A Calibrator and operating it from the front panel keys and in remote configurations. Th

The 5500A Calibrator may be operated at the front panel in the local mode, or remotely

using RS-232 or IEEE-488 ports. For remote operations, several software options are available to integrate 5500A operation into a wide variety of calibration requirements. Local Operation Typical local operations include front panel connections to the Unit Under Test (UUT), and then manual keystroke entries at the front panel to

Preparing for Operation

5 Unit Under Test PC or Terminal 5500A 5500A SERIAL 1 FROM HOST port COM port RS-232 Remote Operation using the SERIAL 1 FROM HOST port PC or Terminal SERIAL 1 FROM HOST port COM port SERIAL 2 TO UUT port RS-232 Remote Operation using the SERIAL 1 FROM HOST and SERIAL 2 TO UUT ports F1-02.eps Figure 2. RS-232 Remote Connections Preparing for Operation The following paragraphs provide instructions for unpacki

Fluke 5500A Table 1. Standard Equipment

Item Model or Part Number Calibrator Line Power Cord 5500A CD-ROM (contains Operator and Service Manual) 5500A Getting Started Manual 5500A Operator Reference Guide 5500A Remote Programming Reference Guide Certificate of Calibration 5500A See Table 2 1627768 945159 945097 105783 Form G749 Replacing the Fuse CAUTION To prevent possible damage to the instrument, verify the correct fuse is installed for the

Connecting to Line Power

7 Connecting to Line Power Warning To avoid shock hazard, connect the factory supplied three- conductor line power cord to a properly grounded power outlet. Do not use a two-conductor adapter or extension cord; this will break the protective ground connection. Use the rear-panel ground terminal for a protective grounding wire if there is any question as to instrument earth grounding. The calibrator is ship

CAUTION FOR FIRE PROTECTION

REPLACE ONLY WITH A 250V FUSE OF INDICATED RATING MAINS SUPPLY / 120V 100V / 240V 220V T2.5A 250V (SB) T1.25A 250V (SB) 0V (SB) 120 120 CHANGING LINE VOLTAGE CHANGING LINE FUSE 240 LINE VOLTAGE INDICATOR F2-01.eps Figure 4. Accessing the Fuse and Selecting Line Voltage

Each Model 5500A Calibrator is warranted to the original purchaser for a period of 1

year beginning on the date received. The warranty is located at the front of this manual. To locate an authorized service center, call Fluke using any of the phone numbers listed below, or visit us on the World Wide Web: www.fluke.com USA: 1-888-99-FLUKE (1-888-993-5853) Canada: 1-800-36-FLUKE (1-800-363-

Fluke Test Equipment Owners Instructions: