Quick Installation Guide
00825-0100-4662, Rev CC
December 2012
Rosemount 8732
Start
Step 1: Pre-Installation
Step 2: Handling
Step 3: Mounting
Step 4: Installation
(Flanged Sensors)
(Wafer Sensors)
(Sanitary Sensors)
Step 5: Grounding
Step 6: Wiring
Step 7: Basic Configuration
Product Certifications
End
Rosemount 8732E
Quick Installation Guide
00825-0100-4662, Rev CC
December 2012
Rosemount 8732
2
© 2013 Rosemount Inc. All rights reserved. All marks property of owner.
IMPORTANT NOTICE
This document provides basic installation guidelines for the Rosemount® 8732. It does
not provide instructions for detailed configuration, diagnostics, maintenance, service,
troubleshootin
Quick Installation Guide
00825-0100-4662, Rev CC
December 2012
Rosemount 8732
3
STEP 1: PRE-INSTALLATION
Before installing the Rosemount 8732 Magnetic Flowmeter Transmitter, there are several
pre-installation steps that should be completed to make the installation process easier:
• Identify the options and configurations that apply to your application
•
Quick Installation Guide
00825-0100-4662, Rev CC
December 2012
Rosemount 8732
4
Environmental Considerations
To ensure maximum transmitter life, avoid excessive heat and vibration. Typical problem
areas:
• high-vibration lines with integrally mounted transmitters
• warm-climate installations in direct sunlight
• outdoor installations in cold climates.
Quick Installation Guide
00825-0100-4662, Rev CC
December 2012
Rosemount 8732
5
Changing Hardware Switch Settings
In most cases, it is not necessary to change the setting of the hardware switches. If you
need to change the switch settings, complete the steps outlined in the manual.
Electrical Considerations
Before making any electrical connections to the Ros
Quick Installation Guide
00825-0100-4662, Rev CC
December 2012
Rosemount 8732
6
STEP 2: HANDLING
Handle all parts carefully to prevent damage. Whenever possible, transport the system to
the installation site in the original shipping containers. PTFE-lined sensors are shipped with
end covers that protect it from both mechanical damage and normal unrestrained
Quick Installation Guide
00825-0100-4662, Rev CC
December 2012
Rosemount 8732
7
STEP 3: MOUNTING
Upstream/Downstream Piping
To ensure specification accuracy over widely varying process conditions, install the sensor a
minimum of five straight pipe diameters upstream and two pipe diameters downstream from
the electrode plane (see Figure 3).
Installations with
Quick Installation Guide
00825-0100-4662, Rev CC
December 2012
Rosemount 8732
8
Sensor Location
The sensor should be installed in a location that ensures the sensor remains full during
operation. Vertical installation allows upward process fluid flow and keeps the
cross-sectional area full, regardless of flow rate. Horizontal installation should be restricte