Perel EVM1400 Operation & User’s Manual
Read Online
Read the Perel EVM1400 Operation & User’s Manual online using the built-in PDF viewer.
Key Pages of Document
MULTIMÈTRE 5 IN 1 5-IN-1 MULTIMETER MULTÍMETRO 5 EN 1 5-IN-1 MULTIMETER MULTÍMETRO 5 EM 1 USER MANUAL NOTICE D’EMPLOI GEBRUIKERSHANDLEIDING MANUAL DEL USUARIO BEDIENUNGSANLEITUNG MANUAL DO UTILIZADOR ...
1. Introduction To all residents of the European Union Important environmental information about this product This symbol on the device or the package indicates that disposal of the device after its lifecycle could harm the environment. Do not dispose of the unit (or batteries) as unsorted municipal waste; it should be taken to a specialized company for recycling. This device should be returned to your distr ...
The data hold function makes the meter stop updating the display. Press HOLD to hold the reading. appears on the display. Press again to switch off the function. • Function Selection Press FUNC. when measuring voltage or current to switch between AC and DC; press the button to select between resistance, capacitance, diode and continuity measuring. • Manual Ranging and Autoranging Modes The meter defau ...
1. Connect the black test lead to the COM jack and the red test lead to the V jack. 2. Set the rotary selector at the desired V range. 3. Press FUNC. to select the AC range. 4. Connect the test leads with the source or the load under measurement. 5. Read the displayed value. 6. The meter indicates an overrange when OL is displayed. Select a higher range. Select the highest range when the value to be measured is unknown. � ...
1. Connect the black test lead to the COM jack and the red test lead to the jack. 2. Set the rotary selector at the desired range. 3. Press FUNC. to select the range. 4. Connect the test leads to the capacitor being measured and read the displayed value. Note: • For high capacity values, the meter may take a few seconds before stabilizing the reading. • For low capacity values, press REL to make the LCD ...
1. Set the rotary selector at the desired °C range. 2. The LCD will display the current ambient temperature. 3. Connect the black test lead of the “K” type thermocouple to the COM jack and the red test lead to the °C jack. 4. Touch the object to be measured with the thermocouple. • Humidity Measurement 1. Set the rotary selector at the desired %RH range. 2. The LCD will display the current ambient humi ...
• Set the rotary switch to the OFF position. • Disconnect all test leads and/or any connectors from the terminals. • Open the battery compartment using an appropriate screwdriver. • Remove the fuse by sliding it out of its bracket. • Install a replacement fuse (F 500mA/250V, Ø 5 x 20mm). • Close the battery compartment. c. Battery Replacement • Set the rotary switch to the OFF position. • Disconnect a ...
4V 1mV 40V 10mV 400V 100mV 600V 1V ± (1.0% of rdg + 2 digits) Overload protection: 250VDC or 250VAC RMS for 400mV range; 600VDC or 600VAC RMS for 4V~600V ranges Input impedance: 10MΩ Max. input voltage: 600VDC Note that the meter may display an unsteady reading when the test leads do not touch the circuit due to the high sensitivity. AC Voltage Range Resoluti ...
4kΩ 1Ω 40kΩ 10Ω 400kΩ 100Ω 4MΩ 1kΩ 40MΩ 10kΩ ± (2.0% of rdg + 5 digits) Open circuit voltage: 0.25V Overload protection: 250VDC or 250VAC RMS Capacitance Range Precision Accuracy 4nF 1pF ± (3.0% of rdg + 3 digits) 40nF 10pF 400nF 0.1nF 4µF 1nF 40µF 10nF 200µF 100nF ± (8.0% of rdg + 10 digits) Overload protection: 250VDC or 250VAC RMS ...
0.1% ± 3.0% Hz range Frequency response: 0~200kHz Input voltage range: 0.5V – 10V AC RMS Overload protection: 250VDC or 250VAC RMS V range Frequency response: 0~40kHz Input voltage range: 0.5VAC – 600VAC RMS Input impedance: 10MΩ Max. input voltage: 600VDC or 600VAC RMS mA range Frequency response: 0~40kHz Input current range: 100mA – 400mA AC RMS Max. input current: 400mA DC or ...
voltage of diode Forward DC Current: ± 1mA Reverse DC Voltage: ± 1.5V Overload Protection: 250VDC or 250VAC RMS Continuity Range Function built-in buzzer will sound if resistance < 40Ω Open circuit voltage: ± 0.5V Overload Protection: 250VDC or 250VAC RMS Use this device with original accessories only. Velleman nv cannot be held responsible in the event of damage ...
• Si la gamme de la valeur à mesurer est inconnue, instaurer le multimètre sur la gamme la plus élevée. • Déconnecter les câbles de mesure du circuit avant de tourner le sélecteur de gammes. • Les impulsions de tension de forte amplitude peuvent endommager le mètre lors de mesurages sur des postes de télévision ou des circuit ...
Le multimètre affiche la valeur relative de toutes les fonctions, excepté le mesurage de fréquence. 1. Sélectionner une fonction et raccorder les cordons de mesure au circuit. 2. Enfoncer REL pour mémoriser la valeur et pour activer le mode de mesurage relatif. La différence entre la valeur de référence et la valeur mesurée s’affiche. 3. Maintenir REL enfoncé pendant 2 secondes pour revenir au mode normal ...
1. Raccorder le câble de mesure noir à la connexion COM et le câble de mesure rouge à la connexion A (max. 400mA) ou 10A (max. 10A). 2. Placer le sélecteur rotatif sur la gamme A désirée. 3. Sélectionner la gamme CA avec la touche FUNC.. 4. Raccorder les câbles de mesure au circuit. 5. Lire la valeur affichée. 6. Le mètre indique une gamme trop petite en affichant OL. Sélectionner une gamme plus larg ...
1. Raccorder le câble de mesure noir à la connexion COM et le câble de mesure rouge à la connexion Ω. 2. Placer le sélecteur rotatif sur la gamme désirée. 3. Sélectionner la gamme avec la touche FUNC. 4. Raccorder les cordons de mesure au circuit. Le ronfleur sonne lorsqu’il y a continuité, c.à.d. la résistance < 40Ω. Remarque : • Le mètre indique OL lorsque la résistance mesurée > 400Ω. ...
1. Placer le sélecteur rotatif sur la gamme dB désirée. 2. Le LCD affiche le niveau sonore actuel. • Mesure de l’éclairement lumineux 1. Placer le sélecteur rotatif sur la gamme Lux désirée. 2. Placer le capteur du mètre en face de la source lumineuse à mesurer. 3. Le LCD affiche le niveau de l’éclairement lumineux actuel. 5. Nettoyage et entretien Ne pas réparer ou entretenir le multimètre ...
• Insérer une nouvelle pile (6F22). • Refermer le compartiment des piles. 6. Spécifications techniques Précision spécifiée un an après calibrage à une température de service entre 18°C ~ 28°C avec un taux d’humidité relative de 0% ~ 75%. Précision notée comme ± (% de l’affichage + nombre de digits secondaires). Conditions ambiantes 1000V CAT. II et 600V CAT. III Degré de pollution 2 Altitude < ...
40V 10mV 400V 100mV 600V 1V ± (1.5% + 5 digits) Protection surcharge : 600VCC ou 600VAC RMS pour gammes 4V~600V Impédance d’entrée : 10MΩ Tension d’entrée max. : 600VCA Plage de fréquence : 40Hz~500Hz Réponse : Moyenne, calibrée en RMS de l’onde sinusoïde Le multimètre affiche une valeur instable tant que les cordons de mesure ne sont par raccordés ...
99.99Hz 0.01Hz ± (1.5% + 5 digits) 999.9Hz 0.1Hz 9.999kHz 1Hz 99.99kHz 10Hz ± (2.0% + 5 digits) 199.9kHz 100Hz > 200kHz référence Gamme Hz Plage de mesure : 0~200kHz Plage de tension d’entrée : 0.5VCA – 10VCA RMS Protection surcharge : 250VCC ou 250VCA RMS Gamme V Plage de mesure : 0~40kHz Plage de tension d’entrée : 0.5VCA – 600VCA RMS Impédance ...