Simpson 150-2 About this Manual
To the best of our knowledge and at the time written, the information con-
tained in this document is technically correct and the procedures accurate
and adequate to operate this instrument in compliance with its original ad-
vertised specifications.
Notes and Safety Information
This Operator’s Manual contains warning symbols which alert the user to
check for hazardous conditions. These appear throughout this manual where
applicable,
Simpson 150-2 SHOCK HAZARD: As defined in American National Standard, C39.5, Safety
Requirements for Electrical & Electronic Measuring & Controlling Instrumen-
tation, a shock hazard shall be considered to exist at any part involving a
potential in excess of 30 volts RMS (sine wave) or 42.4 volts DC or peak and
where a leakage current from that part to ground exceeds 0.5 milliampere,
when measured with an appropriate measuring instrument defined in Sec-
tion 11.6.1 o
Simpson 150-2 Items And Accessories .......................................................................7
1.2
Safety Considerations ........................................................................ 7
1.3
Technical Data ....................................................................................7
2.
INSTALLATION ..............................
The Simpson 150-2 Amp-Clamp (hereafter referred to as the 150-2 or the
Amp-Clamp) allows current measurements without breaking the current line
and interrupting the working appliance under test. In effect, the 150-2 works
as a transformer, containing a split core for accommodation of the current
conductor under test. The transformer split core is made of two sections, hinged
at one end, and pivoted so that it can be attached around the
These amp clamps are calibrated for opera-
tion on 60 Hz circuits, however, they can be
calibrated for operation on 50 Hz circuits or
other frequencies on special order.
EFFECT OF CONDUCT OR
POSITION:
The effect of moving current conductor from
center to any position within core jaws:
�3%.
TEST CONDUCTOR SIZE
CHECK:
Round wires and buss bars, 1-1/8 inch
maximum outside diameter. Rectangular
and square-shaped conductors to 1 inch
s
The following contains instructions for installation and shipping of the 150-2.
Included are unpacking and inspection procedures, warranty, shipping and
care.
2.1
Unpacking And Inspection
Examine the shipping carton for signs of damage before unpacking. If un-
damaged unpack the instrument and inspect it for possible damage. Check
that all furnished items and accessories are included (Table 1-2). Check the
electrical performance as soon
CONTROLS AND CONNECTORS
All operating and adjustment controls and connectors are described in this
section, along with a list (Table 3-1) describing their functions. Become famil-
iar with each item before operating the Amp-Clamp.
Table 3-1. Controls and Connectors
RANGE SELECTOR
SWITCH:
The range selector switch has six positions. Each
position is marked for two AC current ranges. The
two current ranges selected are indicated in a re-
cessed sec
Simpson 150-2 This section contains information required to operate the Instrument in a safe
and proper manner.
Amp-Clamp core and circuit housings are made of high quality insulating
material and designed to hold the stresses of daily service involvements.
BE EXTREMELY CAREFUL WHEN MEASURING HIGH MAGNITUDE CUR-
RENTS OR APPROACHING WIRES OR APPLIANCES CARRYING HIGH
POWER LINE VOLTAGES.
4.1
AC Current Measurements :
1.
Before starting measurements, conne
Simpson Multimeter Owners Instructions:
| Device Model | Type of Document | Pages | Size | Views | Downloads | Updated |
|---|---|---|---|---|---|---|
8455A |
User Guide: Simpson 8455A (3X3E78, Upd.06-06-2026) |
15 | 998 | 190 | 06-06-2026 | |
260 Series 5P |
Simpson 260 Series 5P User Guide (Operator's manual), @RI93BD |
21 | 52 | 13 | 24-08-2026 | |
260-AFP-1 |
Simpson 260-AFP-1 User Guide (Handbook), @414981 |
19 | 1344 | 283 | 22-06-2026 | |
150-2 |
Simpson Multimeter Operator's manual (File: PDF, 15-04-2026) |
16 | 88 | 17 | 15-04-2026 | |
270 series 5 |
Multimeter #GLN87Q |
58 | 554 | 100 | 01-09-2026 | |
260 series 5 |
Multimeter #P7N4PP |
18 | 1322 | 252 | 06-08-2026 |