ECD HCA1 Instruction Manual
Text of ECD HCA1 Instruction Manual
HCA1 Chlorine Plus Photometer System
Instruction Manual
Ideal for the measurement of Free and Total Chlorine in Drinking Water,
Waste Water, Environmental or other clear water samples
Parts Order/Reorder Information
Part#
Description
1000040-1
Meter, case and 2 bottles test strips, Free chlorine and total
Chlorine
1000040-2
Meter, case and 2 bottles of test strips, Free chlorine
1000040-3
Meter, case and 2 bottles of test strips, Total chlorine
1000040-4
Meter, case and 2 bottles of test strips, Chlorine Dioxide
1000040-5
Meter, case and 2 bottles of test strips, High Range chlorine
1000040-6
Meter, case and 2 bottles of test strips,
HCA1 Chlorine Plus Specifications ........................................................ 4
HCA1 Range and Accuracy .................................................................... 4
HCA1 General Information .................................................................... 5
Accuracy and Calibration ...................................................................... 5
Compliance Verification for Free and Total Chlorine Testing .......
HCA1 Chlorine Plus Specifications
Measurement method:
Photometric
Light source:
Light Emitting Diode (LED)
Wavelength:
525 nm
Transmission Range:
0.0 – 100%
Photometric precision:
± 0.1/0.01 %T
Display:
3 digit LCD with annunciators
Cell path length:
20 mm
Cell Chamber:
Molded PET plastic, non-removable
Sample requirement:
4 ml
Temperature range:
0 - 50°C
Power Supply:
(4) AAA alkaline batteries, not included
Battery Life:
>2000 tests with alkaline b
HCA1 General Information
In order to save power, the meter is designed to turn off after 3 minutes (timed from
the last button pressed).
Should the meter turn off in the middle of a test, the last stored zero in the meter will
remain valid when the meter is turned on again. Also, the test result is stored in
memory for easy retrieval.
The HCA1 Chlorine Plus meter is controlled by three buttons:
1. ZERO/ON: When first pressed, this button turns the meter
calibration, the Ready Snap® 3 Calibration Verification Standard part #490903) is
available for purchase.
Compliance Verification for Free and Total Chlorine Testing
This DPD test system is accepted by most health departments because this test is USEPA
(DIN Standard 38 408 G4, ISO 7393/2) accepted for testing requirements for Free and
Total Chlorine. The Chlorine Plus meter uses a wavelength of 525nm; and the
compliance requirement is that the colorimet