Endress+Hauser prosonic flow 92 Technical Information

Text of Endress+Hauser prosonic flow 92 Technical Information

Portable Ultrasonic Flow Measuring

System prosonic flow 92 Temporary volume flow measurement of liquids with “Clamp On” sensors Features and benefits • Portable transmitter, battery operated, for temporary measurements • Built-in data logger with a capacity of 40000 measuring values • Site Setup operating menus for straightforward commissioning of up to 20 measurement sites • “Clamp On” sensors,

Prosonic Flow 92

2 Endress+Hauser Function and system design Measuring principle Prosonic Flow operates on the principle of transit time difference. An acoustic (ultrasonic) signal is transmitted in both directions from one measuring sensor to the other. As the signal propagation velocity of the waves is less when the waves travel against the direction of flow than along the direction of flow, a transit time difference occurs. This difference is directly p
Prosonic Flow 92 Endress+Hauser 3 Sensor arrangement Prosonic Flow W/U The Prosonic Flow W sensors offer a choice between 1, 2 and 4 traverses and the Prosonic Flow U sensors 2 traverses. Please note that the more reflection points there are in the pipe, the more the signal strength decreases. (Example: 2 traverses = 1 reflection point, etc.) Prosonic Flow W: 1 = 1 traverse, 2 = 2 traverses, 4 = 4 traverses Prosonic Flow U: 2 = 2 traverses Therefore, to ma
Prosonic Flow 92 4 Endress+Hauser Measuring system The measuring system consists of the following transmitter and flow measuring sensors. Overview of components: 1 = Prosonic Flow 92 transmitter; 2 = Prosonic Flow W sensors, sensor holders and spacing ruler; 3 = Prosonic Flow U sensor assembly; 4 = Sensor connecting cables; 5 = Power adapter; 6 = Tensioning bands; 7 = Coupling medium; 8 = Carrying case Measuring transmitter Prosonic Flow 92 • For operati
Prosonic Flow 92 Endress+Hauser 5 Input Measured variable Flow velocity (differential delay proportional to flow velocity) Measuring range Prosonic Flow W/U sensors typically v = 0...7 m/s at the specified measuring accuracy Operable flow range over 70 : 1 Input signal Current input: 4...20 mA, galvanically not isolated, for input of externally measured variable (Display and data logging) Output Output signal Current output: active 4...20 mA, RL = 0...1 k�
Prosonic Flow 92 6 Endress+Hauser Power supply by power adapter A = View A – 1 = 17.5 V DC connector – 2 = AC power adapter for power supply and recharging the battery: 100...240 V AC, 47...63 Hz, Power consumption ≤12 W Don’t use other power adapters, or it may result in an accident or in damage of the device. The following mains connector adapters are delivered with the power supply: 1 = European standard 2 = GB standard 3 = USA/Japan standard 4
Prosonic Flow 92 Endress+Hauser 7 Electrical connection: sensor cable connection A = View A – 1 = Upstream sensor cable – 2 = Downstream sensor cable Electrical connection: analog input/output connection A = View A – 1 = Analog input/output connector – 2 = Analog output wires; 2.1 = red (+); 2.2 = black (-) – 3 = Analog input wires; 3.1 = red (+); 3.2 = black (-) Potential equalisation Special measures for potential equalisation are not necessary

Endress+Hauser Measuring Instruments Owners Instructions: