Endress+Hauser prosonic flow 92 Technical Information
Text of Endress+Hauser prosonic flow 92 Technical Information
Portable Ultrasonic Flow Measuring
System
prosonic flow 92
Temporary volume flow measurement of liquids
with “Clamp On” sensors
Features and benefits
• Portable transmitter, battery operated,
for temporary measurements
• Built-in data logger with a capacity of
40000 measuring values
• Site Setup operating menus for
straightforward commissioning of up to
20 measurement sites
• “Clamp On” sensors,
Prosonic Flow 92
2
Endress+Hauser
Function and system design
Measuring principle
Prosonic Flow operates on the principle of transit time difference.
An acoustic (ultrasonic) signal is transmitted in both directions from one measuring sensor to the
other. As the signal propagation velocity of the waves is less when the waves travel against the
direction of flow than along the direction of flow, a transit time difference occurs. This difference
is directly p
Prosonic Flow 92 Endress+Hauser
3
Sensor arrangement Prosonic Flow W/U
The Prosonic Flow W sensors offer a choice between 1, 2 and 4 traverses and the Prosonic Flow
U sensors 2 traverses. Please note that the more reflection points there are in the pipe, the more
the signal strength decreases. (Example: 2 traverses = 1 reflection point, etc.)
Prosonic Flow W: 1 = 1 traverse, 2 = 2 traverses, 4 = 4 traverses
Prosonic Flow U: 2 = 2 traverses
Therefore, to ma
Prosonic Flow 92 4
Endress+Hauser
Measuring system
The measuring system consists of the following transmitter and flow measuring sensors.
Overview of components:
1 = Prosonic Flow 92 transmitter; 2 = Prosonic Flow W sensors, sensor holders and spacing ruler;
3 = Prosonic Flow U sensor assembly; 4 = Sensor connecting cables; 5 = Power adapter;
6 = Tensioning bands; 7 = Coupling medium; 8 = Carrying case
Measuring transmitter
Prosonic Flow 92
• For operati
Prosonic Flow 92 Endress+Hauser
5
Input
Measured variable
Flow velocity (differential delay proportional to flow velocity)
Measuring range
Prosonic Flow W/U sensors typically v = 0...7 m/s at the specified measuring accuracy
Operable flow range
over 70 : 1
Input signal
Current input: 4...20 mA, galvanically not isolated, for input of externally measured variable
(Display and data logging)
Output
Output signal
Current output: active 4...20 mA, RL = 0...1 k�
Prosonic Flow 92 6
Endress+Hauser
Power supply by power adapter
A = View A
– 1 = 17.5 V DC connector
– 2 = AC power adapter for power supply and recharging the battery:
100...240 V AC, 47...63 Hz, Power consumption ≤12 W
Don’t use other power adapters, or it may result in an accident or in damage of the device.
The following mains connector adapters are delivered with the power supply:
1 = European standard
2 = GB standard
3 = USA/Japan standard
4
Prosonic Flow 92 Endress+Hauser
7
Electrical connection:
sensor cable connection
A = View A
– 1 = Upstream sensor cable
– 2 = Downstream sensor cable
Electrical connection:
analog input/output
connection
A = View A
– 1 = Analog input/output connector
– 2 = Analog output wires; 2.1 = red (+); 2.2 = black (-)
– 3 = Analog input wires; 3.1 = red (+); 3.2 = black (-)
Potential equalisation
Special measures for potential equalisation are not necessary