Extech Instruments UV510 Operation & User’s Manual

Brand: Extech Instruments
Category: Measuring Instruments
Model: UV510 (Operation & User’s Manual)
File Type: PDF
File Size: —
Pages: 6
Text of Extech Instruments UV510 Operation & User’s Manual

Extech Instruments UV510 UV Light Meter

Model UV510 Additional User Manual Translations available at www.extech.com

Thank you for selecting the Extech Model UV510 UV Light Meter. This device measures UVA

ultraviolet light between 320-390nm. This device is shipped fully tested and calibrated and, with proper use, will provide years of reliable service. Please visit our website (www.extech.com) to check for the latest version and translations of this User Manual, Product Updates, Product Registration, and Customer Support. Features �

LCD Backlight button and ▲ button

5. Range button 6. Zero button 7. SET and ▼ button 8. Record button (max/min) 9. Hold button 10. Light sensor cover Note: Battery compartment and tripod mount on rear of meter LCD Description 1. Primary Display 2. MAX/MIN Record Icon 3. MAX Icon 4. MIN Icon 5. Display Hold Icon 6. Battery Indicator 7. Auto Power OFF 8. Unit o

Taking a Measurement

1. Short press the power button. If the LCD does not switch on, check the batteries located in the rear battery compartment. 2. With the sensor cover attached to the meter, short press the zero button, to zero the display, before taking any measurements. 3. Remove the sensor cover. 4. Aim the light sensor at the UV light source. 5. View the measurement on the meter’s display. 6. Use the Range R bu

Press the R (record) button again to display the minimum value recorded. The

icon will appear on the display. 4. Press the H (hold) button to clear the recorded data. The previous data will be erased and the meter will start recording new data. 5. Long press the R (record) button to exit the record mode. Auto Power OFF (APO) In order to conserve battery life, the meter will automatically shut off after approximately 10 minutes

Extech Instruments UV510 Backlit LCD 35 x 30mm (1.4 x 1.2“)

Light sensor Photo sensor with cosine correction filter Measurements Ultraviolet Light (UVA) UV sensor spectrum 320 to 390 nm Input Overload Display “----“ Tripod mount On rear panel Power 3 x 1.5V ‘AAA’ batteries Power Consumption Approx. 5mA DC Automatic power off After approx. 10 minutes of inactivity Operating temperature 0 to 50oC (32 to 122oF) Operating humi

Extech Instruments Measuring Instruments Owners Instructions:

Device Model Type of Document Pages Size Views Downloads Updated
ExStik II EC400

User Manual: Extech Instruments ExStik II EC400 (MX14H6, Upd.16-06-2026)

11 1152 208 16-06-2026
DC400

#53R931: DC400 Measuring Instruments Operation & user’s manual

4 1063 213 07-10-2026
MA440

Extech Instruments MA440 Guide (Operation & user’s manual), @6W7429

21 629 120 10-10-2026
UV505

Extech Instruments Measuring Instruments Operation & user’s manual (File: PDF, 06-09-2026)

5 172 38 06-09-2026
ExStik FL700

#7951DK: ExStik FL700 Measuring Instruments Operation & user’s manual

11 396 96 25-07-2026
EX613

User Guide: Extech Instruments EX613 (3Q7OBG, Upd.15-09-2026)

14 99 15 15-09-2026
407910

User Guide: Extech Instruments 407910 (IS35OY, Upd.18-05-2026)

7 1322 318 18-05-2026
380801

PDF User Guide (@6913G7), Extech Instruments 380801 Measuring Instruments (05-06-2026)

16 23 6 05-06-2026
461893

Measuring Instruments #U4IWQ6

4 291 65 20-04-2026

Measuring Instruments: Similar Manuals