Key Pages of Document

Table of Contents

English Manual Updates 1 Deutsch Handbuchaktualisierung 5 Español Actualización del Manual 9 Français Mises à jour des guides 13 Italiano Aggiornamenti del manuale 17 日日日 マママママの更更 21 한국어 설명서 업데이트 25 简体中文 手册更新 29 繁體中文 手冊更新 33 Русский Изменения в р ...

These manual updates apply to the 34401A User’s Guides listed below. Updates are also provided for Edition

5 of the 34401A Service Guide (p/n 34401-90013) starting on page two. Page numbers correspond to the English version of the manual. • 34401-90413 (User’s Guide – Japanese) • 34401-90416 (User’s Guide – Korean) • 34401-90417 (User ...

Chapter 8: Page 216 - DC Characteristics

Change: Continuity 24-Hour 90 Day 1 Year 0.002 + 0.010 0.008 + 0.020 0.010 + 0.020 To: Continuity 24-Hour 90 Day 1 Year 0.002 + 0.030 0.008 + 0.030 0.010 + 0.030 Chapter 8: Pages 216-217 - DC Characteristics Add footnote [12] under “Function” paramete ...

Chapter 1: Pages 12-13 - DC Characteristics and Measuring/Operating Characteristics

Add footnote [12] under the ‘Function’ parameter “Diode Test” (page 12) and the following text to page 13: “[12] Accuracy specifications are for the voltage measured at the input terminals only. 1mA test current is typical. Variation in the current source will cre ...

Chapter 7: Page 132 - 34401-66501 Main PC Assembly (A1) Replaceable Parts

Change: U502 34401-88861 Prog EPROM (1818-5187) 28480 34401-88861 (See above) 1818-5187 IC 2M OTP ROM 150-ns CMOS 34335 AM27C020-155JC To: U502 34401-88806 Prog EPROM (1818-5680) 28480 34401-88806 (See above) 1818-5680 IC 2M-BIT OTP 120-ns CMOS 04078 M27C2001-12 ...

Diese Handbuchaktualisierungen gelten für die im Folgenden aufgeführten Benutzerhandbücher 34401A. Ab

Seite 2 finden Sie Aktualisierungen für die Ausgabe 5 des 34401A Service Guide (P/N 34401-90013). Die Seitenzahlen entsprechen der englischen Handbuchversion. • 34401-90413 (Benutzerhandbuch – Japanisch) • 34401-90416 (Benutzerhandbuch – Kore ...

Kapitel 8: Seite 210 – DC-Kenndaten

Ändern von: Durchgang 24-Std. 90 Tage 1 Jahr 0.002 + 0.010 0.008 + 0.020 0.010 + 0.020 in: Durchgang 24-Std. 90 Tage 1 Jahr 0.002 + 0.0 ...