1 Safety Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 1.1 Designated Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 1.2 Installation, Commissioning, and Operation . . . . . . . . . . . 4 1.3 Product Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 1.4 Operational Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 1.5 Electrostatic Charge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 1.6 Notes on Safety Conventions and Symbols . . . . . . . . . . . . 7 2 Identification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 2.1 Device Designation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 2.2 Ordering Structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 2.3 Scope of Delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16 2.4 Certificates and Approvals . . . . . . . . . . . . . . . . . . . . . . . 16 2.5 Registered Trademarks . . . . . . . . . . . . . . . . . . . . . . . . . . 16 3 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17 3.1 Incoming Acceptance, Transport, and Storage . . . . . . . . . 17 3.2 Terms Related to Tank Measurements . . . . . . . . . . . . . . 18 3.3 Design and Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . 19 3.4 Necessary Tools for Installation . . . . . . . . . . . . . . . . . . . . 20 4 Mounting of NMS5 . . . . . . . . . . . . . . . . . . . . . . . . . . 21 4.1 Application Drawing for Tank . . . . . . . . . . . . . . . . . . . . . 21 4.2 Mounting without Guide System . . . . . . . . . . . . . . . . . . 22 4.3 Mounting with Stilling Well . . . . . . . . . . . . . . . . . . . . . . 22 4.4 Mounting NMS5 with Guide Wire . . . . . . . . . . . . . . . . . 24 4.5 Preparation for Mounting of NMS5 . . . . . . . . . . . . . . . . . 28 4.6 Electrostatic Charge . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29 4.7 Installation for Wire Drum and Displacer . . . . . . . . . . . . 29 5 Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34 5.1 Wiring Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34 5.2 Terminal Assignment . . . . . . . . . . . . . . . . . . . . . . . . . . . 35 5.3 Temperature Input System . . . . . . . . . . . . . . . . . . . . . . . 42 6 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44 6.1 Touch Control and Programming Matrix . . . . . . . . . . . . 44 6.2 HOME Position . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47 6.3 Access Code . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49 6.4 Operation Command and New Operation Status . . . . . . . 50 6.5 Operation of Displacer . . . . . . . . . . . . . . . . . . . . . . . . . . 52 6.6 Calculation of Level and Densities . . . . . . . . . . . . . . . . . 53 7 Commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55 7.1 Initial Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55 7.2 Settings for NMT 53x Connections . . . . . . . . . . . . . . . . . 58 7.3 Settings for Connection to NRF560 . . . . . . . . . . . . . . . . . 59 7.4 Liquid Level Calibration . . . . . . . . . . . . . . . . . . . . . . . . . 60 7.5 Remote Communications . . . . . . . . . . . . . . . . . . . . . . . . 63 7.6 Density and Density Profile Measurements . . . . . . . . . . . 70 7.7 Interface Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . 77 7.8 Sealing of NMS5 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78 8 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79 9 Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80 9.1 Calibration Chamber . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80 9.2 Power and Control Switch . . . . . . . . . . . . . . . . . . . . . . . 80 9.3 Ball Valve NHV4A . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81 9.4 Reducing Flange NHF4 . . . . . . . . . . . . . . . . . . . . . . . . . . 81 10 Diagnosis and Troubleshooting . . . . . . . . . . . . . . . . 82 10.1 Selection of Diagnostic Code and History . . . . . . . . . . . . . 82 10.2 Error and Status Messages . . . . . . . . . . . . . . . . . . . . . . . . 83 10.3 Flowcharts for Diagnostics and Troubleshooting . . . . . . . 85 10.4 Setting after Parts Replacement . . . . . . . . . . . . . . . . . . . . 91 10.5 Intelligent Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93 10.6 Spare Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94 10.7 Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97 10.8 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97 10.9 Gauge Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98 10.10 Contact Addresses of Endress+Hauser . . . . . . . . . . . . . . . 98 11 Displacer and Measuring Wire. . . . . . . . . . . . . . . . . 99 11.1 Shape, Diameter, and Material . . . . . . . . . . . . . . . . . . . . 99 11.2 Alarm History Display . . . . . . . . . . . . . . . . . . . . . . . . . . 100 11.3 List of Alarm Messages . . . . . . . . . . . . . . . . . . . . . . . . . 101 11.4 Hall Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101 11.5 Initial Weight Calibration . . . . . . . . . . . . . . . . . . . . . . . 103 12 Technical Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107 13 Matrix . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 108 13.1 Programming Matrix . . . . . . . . . . . . . . . . . . . . . . . . . . . 108 13.2 Description of Programming Matrix . . . . . . . . . . . . . . . . 118 14 Appendix. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 154 14.1 RS 485 MODBUS (COM- 5) Terminator . . . . . . . . . . . . 154 14.2 Measurement Wire Replacement . . . . . . . . . . . . . . . . . . 156 14.3 Displacer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 158 Table of Contents

Proservo NMS5 of intelligent tank gauge is designed for high-accuracy liquid level measurement in storage and

process applications. NMS5 performs the tank inventory management, loss control, total cost saving, and safe operation. NMS5 is designed for the purpose of single or multi-task installations, converting a wide range of measurement functions. 1.2 Installation, Commissioning, and Operation • Mounting, electrical installation, start-up, and maintenance of the instrument may only be performed by trained personnel authorized by the operator of the facility. • Personnel must read and understand these installation instructions before performing the procedures. • The instrument may only be operated by personnel who are authorized and trained by the operator of the facility. All instructions in this manual must be observed. • The installer must make sure that the measuring system is correctly wired according to the wiring diagrams. The measuring system must be grounded. • Observe all law and regulations applicable and valid for your country and pertaining to the opening and repairing of electrical devices. 1.3 Product Requirements Power Source Check the voltage of the power supply before connecting it to the product. It should be the exact voltage required for proper operation of the product. Use in Hazardous Areas When using the product in the first or second-class hazard location (Zone 1 or Zone 2) be sure to use an intrin- sically safe or pressure and explosion-proof apparatus. Take the utmost care during the installation, wiring, and piping of such apparatus to ensure the safety of the system. For safety reasons, maintenance or repairs on the product while it is being used with such apparatus should only be performed by qualified personnel. External Connection When an external connection is required, the product should be protectively grounded before it is connected to a measurement object or an external control circuit. Caution! To avoid injury or damage to devices: Release pressure in the device before removing plugs, drum housing cover, or calibration windows. 1 Safety Instructions

Measuring systems for use in hazardous locations are accompanied by separate "Ex documentation", which

is an integral part of this Operating Manual. Strict compliance with installation instructions and - ratings, as directed in this supplementary documentation, is mandatory. • Use the explosion proof type for measurement in areas where explosion hazards are present. • Devices employed in areas having explosion hazards should be installed and wired in keeping with explo- sion proof regulations. • Devices installed in areas having explosion hazards must not be opened when the power is on. Tighten the cable gland firmly. • Device maintenance and repair is restricted to meet explosion proof regulations. • Ensure that all personnel are properly qualified. • Observe the certification requirements as well as national and local regulations. Power Supply Confirm that local power supply voltage and frequency meet device specifications before attempting to operate. Refer to "12.1 Technical Specifications". Power cable • Use the power supply cable supplied with the device by the manufacturer, or other appropriate cable as specified in the instruction manual. • The power source should have a ground terminal, and the power supply cable should have a ground line. Refer to "5 Wiring". Grounding Do not unground device while the power supply is turned on. Doing so may cause a short circuit, resulting in damage to devices, as well as personal injury from electric shock. Caution! Be certain to ground NMS5 before connecting input and output to other systems. Designated NMS5 Use NMS5 series is designed to measure liquid level in storage tanks or other containment facilities. Auxiliary devices may be connected, as specified in this manual. Performance of such aftermarket devices, however, is not guaranteed. Refer to manufacture supplied instructions when connecting aftermarket devices. If device is used for any purpose other than intended, device damage, as well as personal injury or death may result. Devices are supplied with IEC class 1 (ground terminal). Caution! Changes or modifications other than those expressly approved by Endress+Hauser are strictly prohibited. Unauthorized modifications can cause malfunction or damage, resulting in serious injury or death. 1 Safety Instructions

Mounting with a stilling well is recommended for use in a tank which contains flammable liquid with conduc-

tivity at 10-8 S/m or less. In case of installation without a stilling well, allow enough stilling time according to the table below before lowering the level gauge down to the liquid surface. When a stilling well is used, the stilling time is shown as in column “<10” in the table below. conductivity (S/cm) Reference: Stilling Time (min.) Liquid Volume in Tank (m3) ≤10 10~50 50~100 ≥5000 ≥10-8 ≥1 ≥1 ≥1 ≥2 10-12 ~10-8 ≥2 ≥3 ≥10 ≥30 10-14~10-12 ≥4 ≥5 ≥60 ≥120 ≤10-14 ≥10 ≥10 ≥120 ≥240 (Japan National Institute of Occupational Safety and Health) 1 Safety Instructions

Notes on Safety Conventions and Symbols

In order to highlight safety-relevant or alternative operating procedures in the manual, the following conventions have been used, each indicated by a corresponding symbol in the margin. Safety Conventions Warning! A warning highlights actions or procedures which, if not performed correctly, will lead to personal injury, a safety hazard or destruction of the instrument Caution! Caution highlights actions or procedures which, if not performed correctly, may lead to personal injury or incorrect functioning of the instrument Note! A note highlights actions or procedures which, if not performed correctly, may indirectly affect operation or may lead to an instrument response which is not planned Explosion protection Device certified for Use in Explosion Hazardous Area If the device has this symbol embossed on its name plate it can be installed in an explosion hazardous area Explosion Hazardous Areas Symbol used in drawings to indicate explosion hazardous areas. Devices located in and wiring entering areas with the designation “explosion hazardous areas” must conform with the stated type of protection. Safe Area (non-explosion hazardous area) Symbol used in drawings to indicate, if necessary, non-explosion hazardous areas. Devices located in safe areas still require a certificate if their outputs run into explosion hazardous areas Explosion protection Direct Voltage A terminal to which or from which a direct current or voltage may be applied or supplied Alternating Voltage A terminal to which or from which an alternating (sine-wave) current or voltage may be applied or supplied Grounded Terminal A grounded terminal, which as far as the operator is concerned, is already grounded by means of an earth grounding system Protective Grounding (earth) Terminal A terminal which must be connected to earth ground prior to making any other connection to the equipment Equipotential Connection (earth bonding) A connection made to the plant grounding system which may be of type e.g. neutral star or equipotential line according to national or company practice EX EX 1 Safety Instructions

The following technical data are given on the instrument nameplate:

Figure. 1: ATEX Approval Type Ex d/Ex d [ia] Figure. 2: ATEX Approval Type Ex d � � � � � � � � � � � � � � � � � � �� XA00578G-A/08 Supply Measuring range m Displacer weight g Displacer dia. mm� Measure wire dia. mm� Density range ~ Manufacturing date Test date Tester Zero point of liquid level gauge is mm under the reference point. Only read level when indication "BAL" is present. Certification no. Tank ID Ambient temperature: ~ +60°C Process temperature: -200 ~ +200°C KEMA 05 ATEX 2071 Ex d IIB T6...T3 ATEX II 1/2 G Order code Ser. no. 0044 PROSERVO NMS53 Made in Japan Yamanashi 406-0846 Endress+Hauser Yamanashi Co.,Ltd. IP67, NEMA 4X Warning: Do not open the cover when energized. � � � � � � � � � � � � � � � � � � Type code Order code Serial number ������������������������� Measuring range ���������������� ������������������ ����������������������� ������������������������������� ������������������������������� ������������������� ������������������������ �������������� ��������������������� �������������������������������� ������������������������� ������������������������� ������������������������� NP- ### - # (*2) ��������������������������������������������������������������������� ����������������������� ���������������������� � � � � � � � � � � � � � � � � � KEMA 05 ATEX 2071 Ex d IIC T6...T3 ATEX II 1/2 G NP-2551-4 XA00578G-A/08 Certification no. Tank ID Ambient temperature: -20 ~ +60°C Process temperature: -200 ~ +200°C Order code Ser. no. 0044 PROSERVO NMS53 Made in Japan Yamanashi 406-0846 Endress+Hauser Yamanashi Co.,Ltd. IP67, NEMA 4X Warning: Do not open the cover when energized. Supply Measuring range m Displacer weight g Displacer dia. mm� Measure wire dia. mm� Density range ~ Manufacturing date Test date Tester Zero point of liquid level gauge is mm under the reference point. Only read level when indication "BAL" is present. � � � � � � � � � � � � � � � � � Type code Order code Serial number ������������������������� Measuring range ���������������� ������������������ ����������������������� ������������������������������� ������������������������������� ������������������� ������������������������ �������������� ��������������������� ������������������������� ������������������������� ������������������������� 2 Identification

FM Approval Type XP

Figure. 4: FM Approval Type XP-AIS � � � NP-2346-4 Temperarure Class T4 APPROVED F M XP Class I, Div. 1, Gp. C, D XP Class I, Zone 1, Gp. IIB DIP Class II, III, Div. 1, Gp. E, F, G Ambient temperature: -20 � +60� NEMA 4X, IP67 Order code Ser. no. Input rated Ambient temperature -20 ~ +60 Input Output Relay contact rated 30 V AC, 2A; 42 V DC, 2A, 60 W. Measuring range m Manufacturing date Warning: Keep cover tight while circuits are alive. Caution: Use supply wires suitable for 70� minimum. Caution: A seal shall be installed within 50 mm of the enclosure. PROSERVO NMS53 Made in Japan Yamanashi 406-0846 Endress+Hauser Yamanashi Co.,Ltd. � � � � � � � � � Type code Order code Serial number ������������������������� ������������������������� �������������������������� � � � Install: per drawing Ex421-439- NP-2421-5 APPROVED F M Temperarure class T4 Ambient temperature: -20 � +60� XP-AIS Class I, Div. 1, Gp. C, D XP-AIS Class I, Zone 1, Gp. IIB DIP Class II, III, Div. 1, Gp. E, F, G XP-ANI Class I, Div. 2, Gp. C, D Order code Ser. no. Input rated Ambient temperature -20 ~ +60 Input Output Relay contact rated 30 V AC, 2A; 42 V DC, 2A, 60 W. Measuring range m Manufacturing date Warning: Keep cover tight while circuits are alive. Caution: Use supply wires suitable for 70� minimum. Caution: A seal shall be installed within 50 mm of the enclosure. PROSERVO NMS53 Made in Japan Yamanashi 406-0846 Endress+Hauser Yamanashi Co.,Ltd. � � � ���Type code ���Order code ���Serial number ���������������������������� ���������������������������� ����������������������������� ���XA version number � 2 Identification

CSA Approval Type Ex d[ia]

Figure. 6: CSA Approval Type Ex d ���Type code ���Order code ���Serial number ���Input rating ���������������������������� ����������������������������� ���Measuring ragne ���XA version number NP-2526-2 Ex d[ia] IIB T4; Class I, Zone 1; Class I, Div. 1, Gp. C, D ; [Ex ia] Clsss II, Div. 1, Gp. E, F, G ; Class III Ambient temperature: -20 � +60 � MWP psi R 216422 Type 4X, IP67 Intrinsically Safe Outputs when installed per drawing Ex540-742 Order code Ser. no. Input rated Input Output Relay contact rated 30 V AC, 2A; 42 V DC, 2A, 60 W. Measuring range m Manufacturing date Warning: Keep cover tight while circuits are alive. Caution: Use supply wires suitable for 70� minimum. Caution: A seal shall be installed within 50 mm of the enclosure. PROSERVO NMS53 Made in Japan Yamanashi 406-0846 Endress+Hauser Yamanashi Co.,Ltd. 5 2 3 1 4 6 7 ⑧ � � � NP-2342-3 Order code Ser. no. Input rated Ambient temperature: -20 � +60� MWP psi Exd IIB T4; Class I, Divsion 1, Groups C, D ; Clsss II, Divsion 1, Groups E, F, G ; Class III R 216422 Type 4X, IP67 Input Output Relay contact rated 30 V AC, 2A; 42 V DC, 2A, 60 W. Measuring range m Manufacturing date Warning: Keep cover tight while circuits are alive. Caution: Use supply wires suitable for 70� minimum. Caution: A seal shall be installed within 50 mm of the enclosure. PROSERVO NMS53 Made in Japan Yamanashi 406-0846 Endress+Hauser Yamanashi Co.,Ltd. � � � � Type code Order code Serial number Input rating ������������������������� �������������������������� Measuring range � � � � � � � 2 Identification

TIIS Approval Type Ex d

Figure. 8: IEC Approval Type Ex d -20°C � � � � � � � � � � � � � � � � � � �������Ex proof model : �������Protection class : �� / Rating ���� � Power supply: ���������/ Data communication: DC 48 V 300 mA������� ��������/ Contact input :� DC 30 V 0.6 W ������� ������� / Contact output:� AC 250 V 1.5 A �����������������������������������DC 30 V 9 W �����/Ambient temperature: -20 � +60 � � � � �������������������������������� �����������������������������HIV�����0.05M��Km���� ������������������� Note: Be sure to cut off the power and cool down this instrument before opening the cover. In case of conduit wiring method, shall be use HIV wire (insulation resistance over 0.05M-ohm.Km) Zero point of liquid level gauge is mm under the reference point. Only read level when indication "BAL" is present. NP-2645-1 IP67 NEMA 4X Order code Ser. no. PROSERVO NMS53 ��������������� Yamanashi 406-0846 Made in Japan Endress+Hauser Yamanashi Co.,Ltd. BA 00401G/08 Measuring range m Displacer weight g Displacer dia. mm� Measure wire dia. mm� Density range ~ Manufacturing date Test date Tester Certification no. Tank ID � � � � � � � � � � � � � � � � � � Type code Order code Serial number Ex-proof model Protection class ������������������������� Measuring range Displacer weight Displacer diameter Measuring wire diameter Density measurement lower limit Density measurement upper limit Date of manufacture Date of manufacture test Name of tester ������������������������� ������������������������� ������������������������� � � � � � � � � � � � � � � � � � � Ex d IIB T6...T3 Ga/Gb IECEx KEM 09.0062 NP-2619-2 XA00582G-A/08 Ambient temperature: ~ +60°C Process temperature: -200 ~ +200°C Order code Ser. no. 0044 PROSERVO NMS53 Made in Japan Yamanashi 406-0846 Endress+Hauser Yamanashi Co.,Ltd. IP67, NEMA 4X Warning: Do not open the cover when energized. Supply Measuring range m Displacer weight g Displacer dia. mm� Measure wire dia. mm� Density range ~ Manufacturing date Test date Tester Zero point of liquid level gauge is mm under the reference point. Only read level when indication "BAL" is present. Certification no. Tank ID � � � � � � � � � � � � � � � � � � Type code Order code Serial number ������������������������� Measuring range ���������������� ������������������ ����������������������� ������������������������������� ������������������������������� ������������������� ������������������������ �������������� ��������������������� ���������������� ������������������������� ������������������������� ������������������������� -20°C 2 Identification

IEC Approval Type Ex d -40°C

Figure. 10: IEC Approval Type Ex d IIC � � � � � � � � � � � � � � � � � � Ex d IIB T6...T3 Ga/Gb IECEx KEM 09.0062 NP-2620-2 XA00582G-A/08 Ambient temperature: ~ +60°C Process temperature: -200 ~ +200°C Order code Ser. no. 0044 PROSERVO NMS53 Made in Japan Yamanashi 406-0846 Endress+Hauser Yamanashi Co.,Ltd. IP67, NEMA 4X Warning: Do not open the cover when energized. Supply Measuring range m Displacer weight g Displacer dia. mm� Measure wire dia. mm� Density range ~ Manufacturing date Test date Tester Zero point of liquid level gauge is mm under the reference point. Only read level when indication "BAL" is present. Certification no. Tank ID � � � � � � � � � � � � � � � � � � Type code Order code Serial number ������������������������� Measuring range ���������������� ������������������ ����������������������� ������������������������������� ������������������������������� ������������������� ������������������������ �������������� ��������������������� ���������������� ������������������������� ������������������������� ������������������������� -40°C � � � � � � � � � � � � � � � � � Ex d IIC T6...T3 Ga/Gb IECEx KEM 09.0062 NP-2621-2 XA00582G-A/08 Ambient temperature: -20 ~ +60°C Process temperature: -200 ~ +200°C Order code Ser. no. 0044 PROSERVO NMS53 Made in Japan Yamanashi 406-0846 Endress+Hauser Yamanashi Co.,Ltd. IP67, NEMA 4X Warning: Do not open the cover when energized. Supply Measuring range m Displacer weight g Displacer dia. mm� Measure wire dia. mm� Density range ~ Manufacturing date Test date Tester Zero point of liquid level gauge is mm under the reference point. Only read level when indication "BAL" is present. Certification no. Tank ID � � � � � � � � � � � � � � � � � Type code Order code Serial number ������������������������� Measuring range ���������������� ������������������ ����������������������� ������������������������������� ������������������������������� ������������������� ������������������������ �������������� ��������������������� ������������������������� ������������������������� ������������������������� 2 Identification

Drum Housing Pressure Rating; Material:

1 0… 200mbar/20kPa/2.9psi; Alu 2 0… 200mbar/20kPa/2.9psi; stainless ste> 4 0… 5.88bar/588kPa/85.28psi; Alu 5 0… 5.88bar/588kPa/85.28psi; stainless steel 6 0… 24.5bar/2.45MPa/355.34psi; stainless steel 9 Special version, TSP-no. to be spec. 020 Approval: 0 Weather proof, IP 67 NEMA 4X 1 TIIS Ex d IIB T4 5 FM XP Cl. I Div. 1 Gr. C-D N FM XP-AIS Cl. I Div.1 Gr. C-D 6 CSA Cl. I Div. 1 Gr. C-D O CSA Ex d[ia] Cl. I Div. 1 Gr. C-D G ATEX II 1/2G Ex d IIB T6...T3 Q ATEX II 1/2G Ex d IIC T6...T3 S ATEX II 1/2G Ex d IIB T6...T3, -40°C J ATEX II 1/2G Ex d (ia) IIB T6...T3 U ATEX II 1/2G Ex d (ia) IIB T6...T3, -40°C A IEC Ex d ia IIB T6...T3 Ga/Gb B IEC Ex d IIB T6...T3 Ga/Gb C IEC Ex d IIC T6...T3 Ga/Gb D IEC Ex d ia IIB T6...T3 Ga/Gb, -40°C E IEC Ex d IIB T6...T3 Ga/Gb, -40°C 9 Special version, TSP-no. to be spec. 030 Application: A Liquid Level B PTB (<1mm) type approval, liquid level C NMi (<1mm) type approval, liquid level D Multi measurement, liquid level, I/F level, bottom, density E PTB (<1mm) type approval, liquid level, I/F level, bottom, density F NMi (<1mm) type approval, liquid level, I/F level, bottom, density G Density profile multi measurement, liquid level, I/F level, bottom, density H PTB (<1mm)type approval, density profile, liquid level, I/F level, bottom, density J NMi (<1mm) type approval, density profile, liquid level, I/F level, bottom, density Y Special version, TSP-no. to be spec. 040 Output 1: F Not selected A 2-way 2-wire (V1 protocol) J 2-way 2-wire (MDP protocol) B 2-way 2-wire (BBB protocol) C 2-way 2-wire (MIC, RS232C protocol) D 2-way 2-wire (MIC protocol) G HART active H HART passive L Whessmatic 550, overvoltage protection M Mark Space N Enraf BPM P Modbus RS 485 Y Special version, TSP-no. to be spec. 050 Output 2: 0 Not selected 1 4 x relay SPST 2 2 x 4-20mA 3 4 x relay SPST, 2 x 4 - 20mA 4 2 x relay SPST, Overspill prevention TÜV 5 4 x relay SPST, 1 x 4 - 20mA 9 Special version, TSP-no. to be spec. NMS5- Product designation (part 1) 2 Identification

HART (NMT5xx, NRF560, pressure transmitter)

1 1 x spot temperature Pt100, HART (NRF560, pressure transmitter) 2 2 x operation contact, HART (NMT5xx, NRF560, pressure transmitter) 3 1 x spot temperature Pt100, 2 x operation contact, HART (NMT5xx, NRF560, pressure transmitter) 4 1 x status, HART (NMT5xx, NRF560, pressure transmitter) 5 1 x spot temperature Pt100, 1 x status, HART (NMT5xx, NRF560, pressure transmitter) 6 1 x spot temperature Pt100, 1 x status, 2 x operation contact, HART (NMT5xx, NRF560, pressure transmitter) 9 Special version, TSP-no. to be spec. 070 Measuring Range; wire: C 0 - 28m; SUS316L, 0.15mm L 0 - 36m; SUS316L, 0.15mm N 0 - 47m; SUS316L, 0.15mm H 0 - 16m; PFA>SUS316, 0.4mm K 0 - 16m; Alloy C, 0.2mm M 0 - 22m; Alloy C, 0.2mm Y Special version, TSP-no. to be spec. 080 Cable entry: E 4 x thread G1/2 F 4 x thread G3/4 G 4 x thread NPT1/2 H 4 x thread NPT3/4 L 4 x thread M20 M 4 x thread M25 Y Special version, TSP-no. to be spec. 090 Process connection: A 10K 80A RF, flange JIS B2220 C 10K 80A FF, flange JIS B2220 U 10K 150A RF, flange JIS B2220 E 20K 80A RF, flange JIS B2220 G 3" 150Ibs RF, flange JIS B2220 J 3" 300lbs RF,flange ANSI B16.5 W 4" 300lbs RF, flange ANSI B16.5 T 6" 150lbs RF, flange ANSI B16.5 L DN80 PN10 B1, flange EN1092-1 (DIN2527 B) N DN80 PN25 B1, flange EN1092-1 (DIN2527 B) Q 80A 150lbs RF, flange JPI 7S-15 S 80A 300lbs RF (apply for NMS5-6), flange JPI 7S-15 Y Special version, TSP-no. to be spec. 100 Power supply: 3 85 - 264 VAC, 50/60 Hz 4 20 - 62 VDC, 20 - 55 VAC, 50/60 Hz 9 Special version, TSP-no. to be spec. 110 Displacer: N Cylindrical 30 mm, SUS316 K Cylindrical 40 mm, SUS316 D Cylindrical 50 mm, SUS316 W Cylindrical 30 mm, PTFE V Cylindrical 40 mm, PTFE U Cylindrical 50 mm, PTFE T Cylindrical 50 mm, AIloy C B Conical 50 mm, PTFE R Conical 70 mm, SUS316 S Conical 110 mm, SUS316 Y Special version, TSP-no. to be spec. NMS5- Product designation (part 2) 2 Identification

O-ring; chamber finishing:

0 NBR; not selected 1 Silicon rubber; not selected 5 Silicone rubber; FEP coated 3 PTFE (wire drum FKM); not selected 4 PTFE (Wire drum FKM); FEP coated 6 CR; not selected 2 FKM; not selected 7 FKM; FEP coated 8 FFKM; not selected A FFKM; FEP coated 9 Special version, TSP-no. to be spec. 130 Options: A not selected C Rc3/8 cleaning nozzle D Rc3/8 gas purging nozzle E Guide wire installation G Relief valve H Relief valve, pressure gauge J Sunshade L Rc3/8 gas purging nozzle, sunshade M Rc3/8 cleaning nozzle, sunshade N Rc3/8 gas purging nozzle, guide wire installation P Rc3/8 cleaning nozzle, guide wire installation Q Guide wire installation, sunshade R Relief valve, guide wire installation S Relief valve, pressure gauge, guide wire installation T Relief valve, sunshade U Relief valve, pressure gauge, sunshade Y Special version, TSP-no. to be spec. NMS5- Complete product designation 2 Identification

Scope of Delivery

Caution! It is extremely important to follow the instructions concerning the unpacking, transportation and storage of measuring instruments provided in the chapter “Incoming Acceptance, Transportation, Storage”. The scope of delivery consists of: • Assembled Instrument Accompanying documentation: • Installation Instructions (this manual) • Safety Instructions (XA) • Functional Safety Manual (SD), Supplied with 4-20mA and Overspill Prevention Relay Output 2.4 Certificates and Approvals CE Mark, Declaration of Conformity The device is designed to meet state-of-the-art safety requirements, has been tested and left the factory in a condition in which it is safe to operate. The device complies with the applicable standards and regulations as listed in the EC declaration of conformity and thus complies with the statutory requirements of the EC directives. Endress+Hauser confirms the successful testing of the device by affixing to it the CE mark. 2.5 Registered Trademarks HART® Registered trademark of HART Communication Foundation, Austin, USA ToF® Registered trademark of the company Endress+Hauser GmbH+Co. KG, Maulburg, Germany FieldCare® Registered trademark of the company Endress+Hauser Flowtec AG, Rheinach, CH 2 Identification

Incoming Acceptance, Transport, and Storage

3.1.1 Incoming Acceptance Check the packing and contents for any signs of damage. Check the shipment, and make sure that nothing is missing and that the items match your order. 3.1.2 Transport Caution! • Follow the safety instructions and conditions of transportation for instruments in excess of 18kg (40 lbs.). • Do not lift the measuring instrument by its housing during transportation. 3.1.3 Storage Pack the measuring instrument so that it is protected against impacts during storage and transportation. The original packing material provides the optimum protection for this. The allowed storage temperature is -40°C to +85°C (-40°F to +185°F) 3 Installation

Terms Related to Tank Measurements

Figure 11: Terms Related to Tank Measurements Lower Layer Middle Layer Upper Layer Upper Interface Lower Interface Liquid Level Nozzle NMS5 Reference Line Dip-point Offset Upper Limit Stop Level Upper Limit Alarm Ullage Level Gauge Reference Length Tank Height Upper Interface Level Lower Interface Level Lower Limit Alarm Lower Limit Stop Level Dipping Reference Point (Dipping Data Place) Liquid Depth (Innage Level) Measurement Hatch (Dipping Reference) Bottom of Tank 3 Installation

Design and Dimensions

Figure 12: NMS5 Dimension NMS5 - 1/ NMS5 - 4 � � � � � � � � � � � Electrical compartment Terminal box Display (LCD) Cable entery (A, B, C, D) Wire drum housing Flange Measuring wire Displacer Calibration window Window cover Touch control Max. 39.7mm Travel Distance of Displacer 34.6mm for 28m Wire (Φ0.15mm) 39.7mm for 36m Wire (Φ0.15mm) ��������������������������������������������������� Displacer position at bottom of tank NMS5 - 2/ NMS5 - 5 / NMS5 - 6 Max. 41.6mm Travel Distance of Displacer 34.6mm for 28m Wire (Φ0.15mm) 39.7mm for 36m Wire (Φ0.15mm) 41.6mm for 47m Wire (Φ0.15mm) ��������������������������������������������������� Displacer position at bottom of tank 3 Installation

Necessary Tools for Installation

The following tools are required when installing NMS5. Figure. 13: Tools to Be Prepared 24,26,30,32 350mm 3mm and 5mm B mm 3 1.25 , 2.0 sq sq Box End Wrench Crescent Wrench Allen Wrench (Hex Key) Screw Driver  ■ Flat Head  ■ Phillips Wire Terminal Water Pump Pliers Wire Cutters or Terminal Pliers Density Calibration Test Weight 3 Installation

Mounting of NMS5

The following installation procedures are available for NMS5. • Mounting without guide system. • Mounting with stilling well • Mounting with guide wire 4.1 Application Drawing for Tank Mounting with stilling well or guide wire is required for the following applications: • Floating Roof Tank • Covered Floating Roof Tank • Tank with Strong Agitator or Heavy Turbulence Mounting without any guide system covers all cases that are not listed above. Refer to Figure 14 of applications (with and without mounting stilling well, with mounting guide wire) Figure. 14: Application for Each Tank Floating Roof Tank Fixed Roof Tank Without Stilling Well Fixed Roof Tank With Stilling Well Covered Floating Roof Tank Spherical or Bullet Tank With Guide Wire Fixed Roof Tank Spherical or Bullet Tank Floating Roof Tank ����������������������������������������������� Covered Floating Roof Tank 4 Mounting of NMS5

Mounting without Guide System

NMS5 is mounted on a nozzle of the tank roof without any guide system (Refer to Figure. 14: Application for Each Tank). The mounting preparations require the observance of some recommendations for setting the noz- zle and the minimum measuring level. Figure. 15: Mounting without Guide System 4.3 Mounting with Stilling Well Stilling Well Diameter The stiling well diameter that is required to protect the measuring wire without disturbing its operation varies depending on the tank height. The stiling well could either be constant diameter, or thinner at its upper part and thicker at its lower part. The figure 16 shows two examples of the latter case, namely an asymmetric pipe and a concentric pipe. Figure. 16: Mounting with Stilling Well, Asymmetric Pipe, and Connection Pipe Displacer NMS5 d D1 Valve Asymmetric Pipe Concentric Pipe Displacer Maintenace Chamber NMS5 Asymmetric Pipe (Side View) Concentric Pipe (Front View) NMS5 4 Mounting of NMS5

The valve is required when mounting NMS5 on pressurized liquid tanks.

NMS5 must be mounted on the asymmetric pipe in the direction shown on Figure 16. To calculate the required pipe diameters, the formula below should be used. The variables and constants have the following meanings. D1 Inner diameter of the upper part of the pipe D2 Inner diameter of the lower part of the pipe L Length of the pipe (from the flange of NMS5 to bottom of the stilling well) meters v Deviation of the pipe from the vertical per length (mm/m) d Diameter of the displacer e Lateral shift of the displacer per length due to the groove of the wire drum • Upper Diameter of Stilling Well D1 > d + 10 mm D1 should be 3" or more. • Lower Diameter of Stilling Well - Asymmetric Pipe D2 > d+ eL + 2vL+ 10mm - Concentric Pipe D2 > d + 2eL + 2vL + 10mm Recommendations for Mounting Stilling Well Note! Follow the recommendations for mounting NMS5 with stilling well: • Keep the pipe connection welds smooth. • While drilling holes into the pipe, keep the interior surface of the holes clear of metal chips and burrs. • Coat or paint the interior surface of the pipe to avoid rust. • Keep the pipe as perfectly vertical as possible. Check this by a plumb. • Install the asymmetric pipe under the valve and fit the centers of the NMS5 and the valve. • Set the center of the lower part of the asymmetric pipe to the direction of the displacer motion. • Observe recommendations as per API MPMS chapter 3.1B. • Confirm grounding between NMS NMS5 and tank nozzle. Feature: 070 Measuring Range/Material/Diameter of Wire e = Lateral Shift (mm) C 0-28m: SUS316L, 0.15mm 0.37 H 0-16m: PFA>SUS316, 0.4mm 0.40 K 0-16m: Alloy C, 0.2mm 0.47 L 0-36m: SUS316L, 0.15mm 0.33 M 0-22m: Alloy C, 0.2mm 0.47 N 0-47m: SUS316L, 0.15mm 0.30 4 Mounting of NMS5

Mounting NMS5 with Guide Wire

It is also possible to guide the displacer by a guide wire to prevent lateral motion. Guide Wire Installation Procedure 1. Secure the guide wire [4] to the wire hook. 2. Secure the springs [3] to the reducer plate [2] (see Figure 17). 3. Fix NMS5 [1] temporarily so that springs are located at the outside the reducer plate. 4. Perform weight calibration. 5. Fix NMS5 to the reducer plate. 6. Put displacer ring [7] through guide wire and set the displacer [5] so that marking ring [6] is located on the reference position. This completes the guide wire installation procedure. Figure. 17: Mounting with Guide Wire NMS5 [1] Chamber for Maintenance Displacer (Ø50 mm, Ø70mm, and Ø110mm) [5] Bottom of Tank Welding Anchor Hook Plate Caliblation Chanber NHC4HP (for High Pressure) NHC4LP (for Low Pressure) Ball Valve NHV4A (with ANSI Flange) NHV4J (with JIS Flange) 172 300 Ø150 CALIBRATION CHAMBER (for Low Pressure) Flange Flange Measuring Wire Tank Guide Wire [4] Sleeve Spring [3] Reducer Plate [2] Spring [3] Sleeve Guide Wire Sleeve Crimp Tool Flange 6” 100mm Reference Position Marking Ring [6] *2 *1 *1 *2 Accessory can be ordered separately. *1 Reducer Plate : 6mm for Low Pressur Reducer Flange : for High Pressure, (depending on standard) Ring [7] 4 Mounting of NMS5

Weight Calibration Method for Guide Wire Installation

Figure. 18: Weight Calibration for NMS with Guide Wire Weight Calibration Procedure 1. Fix NMS5 temporarily so that NMS5 is located at the outside the reducer plate to perform weight calibration. 2. Replace the measuring wire to an appropriate length. • During standard weight calibration, the displacer moves up until it touches the mechanical stop in the housing. • After weight calibration is finished, the displacer stops 70mm (the value set in the matrix (configurable, GVH378)) below the mechanical stop. This is the Reference Position. 3. During this special weight calibration, the guide rings of the displacer touches the bottom face of the flange. 4. The displacer goes down 70mm and stops. NMS5 remembers the distance of 70mm as Reference Position, and weight table calibration resumes. 5. When weight calibration is complete, return NMS5 back to its original mounting position on the tank nozzle. This completes the weight calibration procedure. 70mm Reference Position 4 Mounting of NMS5

Type of Tanks

Fixed Roof Tank with Stilling Well Figure. 19: Stilling Well in Fixed Roof Tank Pressurized Tank with Stilling Well and Ball Valve Figure. 20: Stilling Well in Pressurized Tank Measuring Wire Displacer Stilling Well E - + NMT539 NMS5 E-+ Ball Valve Reducer Meas. Wire Displacer Stilling Well Proservo NMS5 Prothermo NMT539 4 Mounting of NMS5

Floating Roof Tank and /or Covered Floating Roof Tank

Figure. 21: Stilling Well in Floating Roof Tank/Covered Floating Roof Tank Note! When NMS5 is installed in a floating roof tank, ensure to use a stilling well. Figure. 22: Stilling Well in Floating Roof Tank E - + Measuring Wire Displacer Stilling Well NMS5 NMT539 E - + Measuring Wire Displacer NMS5 4 Mounting of NMS5

Preparation for Mounting of NMS5

4.5.1 Flange A nozzle and a flange should be prepared prior to mounting NMS5 on the tank. The flange size and the rating of NMS5 vary depending on the customer’s specifications. Note! • Check the flange size of NMS5. • Mount the flange on the top of the tank. The deviation of the flange from the horizontal plane should not exceed +/- 1 degree. • When mounting NMS5 on a long nozzle, ensure that the displacer does not touch the inner side of the noz- zle when the nozzle is in the vertical position. Figure. 23: Allowable Inclination of Mounting Flange Note! When NMS5 is installed without a guide system, follow the recommendations below: • Set the mounting nozzle in the sector between 45 and 90 deg (or -45 and -90 deg.) apart from the inlet pipe of the tank. This prevents heavy swing of the displacer caused by wave or turbulence of the inlet liquid. • Set the mounting nozzle at 500mm or more away from the tank wall. Ensure that the measurement is not influenced by changes of the ambient temperature. • Set the minimum measuring level at 500mm or more above the top of the inlet pipe. This protects the displacer from direct flow of the inlet liquid. • If the stilling well can not be mounted in the tank due to the shape or condition of a tank, attaching a guide system is recommended. Consult E+H services for further information. Figure. 24: Recommended Setting of Mounting Nozzle and Minimum Measuring Level Caution! • Before pouring liquid into the tank, confirm that liquid flowing through the inlet of the pipe does not touch the displacer directly. • When discharging liquid of the tank, ensure that the displacer does not get caught in the liquid current and sucked into the outlet pipe. Mounting Nozzle max. +1° max. -1° -90° 90° -45° 45° 0° Inlet Pipe Mounting Nozzle Mounting Nozzle Minimum Measuring Level �������� Inlet Pipe �������� 4 Mounting of NMS5

When liquid measured by NMS5 has a conductivity at 10-8 s/cm or less, it is quasi-nonconductive. In this case,

using a stilling well or guide wire is recommended. This releases the electrostatic charge on liquid surface. (refer to "1.5 Electrostatic Charge" details on a stiling time for a volume of a charged object). 4.7 Installation for Wire Drum and Displacer When NMS5 is delivered, there are 2 ways of packing procedures depending on mounting methods of a dis- placer. 1. For All-in-one method, the displacer is mounted on the measuring wire of NMS5. Follow the instructions to remove the packing attached to NMS5. 2. Displacer shipped separately. It is necessary to install the displacer in the measuring wire inside NMS5. Follow the instructions below. 4.7.1 All-in-one As NMS5 is protected by some packaging materials, remove them as follows prior to mounting NMS5 (refer to Figure 25: Removing Procedure for All-in-one). Packaging Material Removing Procedure Step Procedure Notes 1 1. Hold the gauge so that it stays horizontal against the flange. 2. Cut the fixing bands. 3. Remove the displacer holder. • Do not tilt the gauge after removing the displacer holder. 2 4. Mount NMS5 on the nozzle . • Be sure that the measuring wire hangs vertically. • Confirm that there are no kinks or other malfunctions in the calibration window. 3 5. Remove the drum housing cover from NMS5 unit. 6. Loosen two screws on the wire drum holder and remove the holder and the bracket. • Be sure not to lose O-ring and fixed bolts for the cover for the drum housing. 4 7. Remove the tape from the wire drum care- fully. • Remove the tape by hands to avoid damage to wire drum. • Make sure that the measuring wire is wrapped so that it fits correctly in grooves. • If not, rewrap the wire so it fits correctly in grooves. 5 8. Attach the drum housing cover. • Confirm that O-ring is in groove. 6 9. Turn on the main power and confirm that NMS5 shows "STOP" and that level shows "16050 ~16060 mm". 10. Set "LEVEL" command, then set "STOP" where the displacer stops 16000 mm or less near the 15950 mm level. 11. Set "UP" command and the displacer will automatically stop at 16000 mm. 12. Set "STOP" command. 4 Mounting of NMS5

Removing Packing Materials for All-in-one

Displacer Holder Fixing Band Tape Wire Drum Holder Set Screw Remove a wire drum holder using a straight slot screwdriver. Drum Housing Cover 4 Mounting of NMS5

Displacer Shipped Separately

It is necessary to install the displacer on the measuring wire inside NMS5. Procedure A : Install Displacer prior to Installing NMS5 on Tank Figures Procedure Notes Figure. 26: Displacer Installation - 1A 1. Put NMS5 on the blocks or a pedestal. 2. Confirm that there is enough space under NMS5. • Be careful not to drop NMS. Figure. 27: Displacer Installation - 2A 3. Remove M6 bolts. 4. Remove the wire drum cover and bracket. 5. Remove the cover, the wire drum bracket, and the wire drum. 6. Remove the tape. 7. Mount the wire drum, the wire drum bracket, and the cover again. • Handle the measuring wire with care. Figure. 28: Displacer Installation - 3A 8. Rotate the wire drum clockwise until the measuring wire ring appears out side the flange. 9. Hook the displacer on the ring of the measuring wire. 10. Secure the displacer to the measuring wire by using the securing wire. 11. Install the ground wire of the displacer (refer to Appendix "14.3.1 Displacer Installation"). 12. Rotate the wire drum counterclockwise until measuring wire ring appears in the calibration window. • Wire drum rotates every one-fifth revolution. • Displacer moves 300mm each one revolution. Figure. 29: Displacer Installation - 4A 13. Mount NMS5 on the nozzle of the tank top. 14. Confirm that the displacer does not touch inside the nozzle inside. 15. Replace the drum housing cover. • Confirm that the measuring wire is wrapped appropriately in the grooves. • If not, rewind the wire correctly in the grooves. Wire Drum Wire Drum Bracket M6 Bolts Tape Cover Measuring Wire Ring Securing Wire Displacer Ground Wire Nozzle NMS5 4 Mounting of NMS5

Displacer Shipped Separately (Diameter 50mm Displacer)

In the case of the diameter 50mm displacer, the displacer can be installed from the calibration window. Note! It is not possible to install the displacer from the calibration window in the following cases. • Except Ø50mm, SUS and alloy C, cylindrical displacer • Except 0.15mm, SUS measuring wire Procedure B : Install Displacer after Installing NMS5 on Tank Figures Procedures Notes Figure. 30: Displacer Installation - 1B 1. Remove the calibration window cover. • Ensure not to kink and damage the measuring wire. Figure. 31: Displacer Installation - 2B 2. Remove M6 bolts. 3. Remove the cover and bracket. 4. Remove the wire drum from the drum housing. 5. Remove the tape that is securing the wire. • Handle the measuring wire with care. Figure. 32: Displacer Installation - 3B 6. Hold the wire drum by using single hand and drop the measuring wire down at approximately 50cm. 7. Secure the wire temporarily with the tape. 8. Insert the wire ring into the drum housing. 9. Pull the wire ring so that the ring comes out from the calibration window. • Handle the measuring wire with care. Figure. 33: Displacer Installation - 4B 10. Insert the wire drum temporarily into the drum housing. 11. Hook the displacer on the wire ring. 12. Secure the displacer to the measuring wire by using the securing wire. 13. Install the ground wire for the displacer (refer to "14.3.1 Displacer Installation" for details). • Handle the measuring wire with care. It may cause damage by kink. Calibration Window NMS5 Wire Drum Wire Drum Bracket M6 Bolts Tape Cover Tape Wire Drum Wire Ring Calibration Window Wire Drum Securing Wire Ground Wire Displacer 4 Mounting of NMS5

Displacer Installation - 5B

14. Hold the wire drum and the displacer by hands. 15. Remove the wire drum from the drum housing and drop the measuring wire down approximately 50 cm. 16. Hold the wire drum up and place the displacer into the calibration window. 17. Hold the displacer at center of calibration window. 18. Hold other hand (wire drum) up and add tension to measuring wire not to drop displacer down rapidly. • Wire drum rotates every one-fifth revolution. • Displacer move over 300mm each one revolution. Figure. 35: Displacer Installation - 6B 19. Release the hand from the displacer. 20. Remove the tape from the wire drum. 21. Insert the wire drum into the drum housing. 22. Set the wire drum bracket. 23. Rotate the wire drum with the hand a couple of times and check that the displacer does not touch inside the nozzle. 24. Pull the ring of the measuring wire up so that it can be seen from the calibration window. 25. Replace the covers of the wire drum and the calibration window. • Confirm that the measuring wire is wrapped correctly in grooves before turning on the power. • If not, rewind the wire correctly in grooves. Figures Procedures Notes Wire Drum Displacer Displacer Window Cover Cover Wire Drum Bracket Wire Drum 4 Mounting of NMS5

The electrical connections of NMS5 are shown in Figures 36-39.

Note! The power supply cable should have the following specifications: • PVC, PE, or equivalently isolated • 600 V insulation voltage or equivalent. • Temperature rating 80 or high The size of the core is defined by the core resistance, voltage drop, and required power consumption. The max- imum power consumption of NMS5 is 50 VA. Caution! • Connect the ground terminal to the ground line inside or outside the terminal box. • Use a wire 4mm2 or more for ground. • Use cables and wires of sufficient and appropriate size and length to make a solid connection at each terminal required. Warnung! Do not stretch cable and wires. This may lead to failure, loss of function, and/or damage to the device and facility. Warnung! Cut the trim cables and wires for appropriate length. Do not leave the extra cables and wires in the electrical compartment. This may lead to failure, loss of function, and /or damage to the device and facility. 5 Wiring

Ex d, Output 1 : V1, Modbus RS485, HART, Enraf BPM

Figure. 36: Electrical Connection of NMS5 NMS5 - X X X X X X X X X X X 1 5 6 G Q S B C E N P F A J B C D G H 14 AL4 COM 18 CTR2 20 OT1+ 22 OT2+ AL4 15 CTR1 17 N.C. 19 OT1- 21 OT2- 23 4 8 AL1 10 AL2 12 AL3 AL3 13 AL2 11 AL1 9 RC B/- 7 5 6 RC A/+ 24 NRF+ A 25 NRF- B 26 b 1 2 3 ARS L ARS G ARS POWL POWN POWG ARSL ARSG ARSN NRF+ NRF- RCA/+ RCB/- ALM1/OSP ALM1/OSP ALM2/OSP ALM2/OSP ALM3 1 2 3 L N G AC Power Supply 24 25 26 + - + - + - ALM3 14 15 16 17 18 19 20 21 22 + - + - 23 4 5 6 7 8 9 10 11 12 13 ALM4 ALM4 COM CTR1 CTR2 OUT1+ OUT1- OUT2+ OUT2- TMPA/NMT+ TMPB/NMT- TMPb Fuse F1 L N G Lightning Arrester NRF or Non IS HART Devices V1, Modbus RS485, HART, Enraf BPM Alarm Contact (Option) External contact Input (Option) Alarm Contact (Option) DC4...20mA Output (Option) DC4...20mA Output (Option) NMT or HART Device (NonIS) Ground or Pt100 ohm Input N 16 L N GND Spot Hoist Common NMT+NMT- Alarm Contact (Option) Alarm Contact (Option) Order Code Example 5 Wiring

Ex d [ia], Output 1 : V1, Modbus RS485, HART, Enraf BPM

Figure. 37: Electrical Connection of NMS5 (with IS HART Connection) 14 AL4 COM 18 CTR2 20 B+ 22 MK AL4 15 CTR1 17 N.C. 19 B- 21 SP 23 4 P_B+ 8 AL1 10 AL2 12 AL3 AL3 13 AL2 11 AL1 9 7 P_B- 5 6 24 P_A+ A 25 P_A- B 26 b 1 2 3 ARS L ARS G ARS N POWL POWN POWG ARSL ARSG ARSN P_B+ P_B- ALM1 / OSP ALM1 / OSP ALM2 / OSP ALM2 / OSP ALM3 1 2 3 L N G AC Power Supply 24 25 26 + - + - + - ALM3 14 15 16 17 18 19 20 21 22 + - 23 4 5 6 7 8 9 10 11 12 13 ALM4 ALM4 COM CTR1 CTR2 B+ B- MK SP TMP A /P_A+ TMP B/P_A- TMP b Fuse F1 L N G Lightning Arrester NRF or Non-IS HART Devices Alarm Contact (Option) Alarm Contact (Option) Alarm Contact (Option) External Contact Input (Option) Alarm Contact (Option) Common Hoist Stop Ground NMT or HART Device (IS) or Pt100 ohm (IS) Input N L GND 16 V1, Modbus, RS485, HART, Enraf BPM DC4...20mA Output (Option) DC4...20mA Output (Option) Order Code Example NMS5 - X X X X X X X X X X X J N O U A D N P F A J B C D G H 5 Wiring

Exd, Output 1: Whessoe Matic 550 (WM550) Protocol

Figure. 38: Electrical Connection of NMS5 (with WM550 Protocol/without IS HART Connection) NMS5 - X X L X X X X X X X X X 1 5 6 G Q S B C E 14 AL4 COM 18 CTR2 20 Ch1+ 22 Oh2+ AL4 15 CTR1 17 N.C. 19 Ch1- 21 Oh2- 23 4 8 AL1 10 AL2 12 AL3 AL3 13 AL2 11 AL1 9 7 5 6 24 A 25 B 26 b 1 2 3 ARS L ARS G ARS N POWL POWN POWG ARSL ARSG ARSN ALM1/OSP ALM1/OSP ALM2/OSP ALM2/OSP ALM3 1 2 3 L N G AC power Supply 24 25 26 + - + - + - ALM3 14 15 16 17 18 19 20 21 22 + - + - 23 4 5 6 7 8 9 10 11 12 13 ALM4 ALM4 COM CTR1 CTR2 Ch1+ Ch1- Ch2+ Ch2- TMPA /NMT+ TMPB/NMT- TMP b Fuse F1 L N G Lightning Arrester NRF or Non IS HART Devices Alarm Contact (Option) Alarm Contact (Option) Alarm Contact (Option) External Contact Input (Option) Alarm Contact (Option) Whessoe Matic 550 Channel 1 Whessoe Matic 550 Channel 2 Common Ground or Pt100 ohm Input NMT or HART Device (Non IS) L N GND NMT+ NMT- Stop Hoist 16 P_B+ P_B- NRF+ NRF- Order Code Example 5 Wiring

Exd [ia], Output 1: Whessoe Matic 550 (WM550) Protocol

Figure. 39: Electrical Connection of NMS5 (with WM550 Protocol/ IS HART Connection) NMS5 - X X L X X X X X X X X X J N O U A D Order Code Example 14 AL4 COM 18 CTR2 20 Ch1+ 22 Oh2+ AL4 15 CTR1 17 N.C. 19 Ch1- 21 Oh2- 23 4 P_B+ 8 AL1 10 AL2 12 AL3 AL3 13 AL2 11 AL1 9 7 P_B- 5 6 24 P_A+ A 25 P_A- B 26 b 1 2 3 ARS L ARS G ARS N POWL POWN POWG ARSL ARSG ARSN P_B+ P_B- ALM1/OSP ALM1/OSP ALM2/OSP ALM2/OSP ALM3 1 2 3 L N G AC power Supply 24 25 26 + - + - + - ALM3 14 15 16 17 18 19 20 21 22 + - + - 23 4 5 6 7 8 9 10 11 12 13 ALM4 ALM4 COM CTR1 CTR2 Ch1+ Ch1- Ch2+ Ch2- TMPA/P_A+ TMPB/P_A- TMP b Fuse F1 L N G Lightning Arrester Alarm Contact (Option) Alarm Contact (Option) Alarm Contact (Option) External Contact Input (Option) Alarm Contact (Option) Whessoe Matic 550 Channel 1 Whessoe Matic 550 Channel 2 Common Ground NMT or HART Device (IS) NRF or Non-IS HARTDevices Pt100 ohm (IS) Input L GND N 16 or Spot Hoist 5 Wiring

Exd, Output 1: Mark/Space (M/S) Protocol

Figure. 40: Electrical Connection of NMS5 (with M/S Protocol without IS HART Connection) NMS5 - X X M X X X X X X X X X 1 5 6 G Q S B C E 14 AL4 COM 18 CTR2 20 B+ 22 MK AL4 15 CTR1 17 N.C. 19 B- 21 SP 23 4 8 AL1 10 AL2 12 AL3 AL3 13 AL2 11 AL1 9 7 5 6 24 A 25 B 26 b 1 2 3 ARS L ARS G ARS N POWL POWN POWG ARSL ARSG ARSN ALM1 / OSP ALM1 / OSP ALM2 / OSP ALM2 / OSP ALM3 1 2 3 L N G AC power Supply 24 25 26 + - + - + - ALM3 14 15 16 17 18 19 20 21 22 + - 23 4 5 6 7 8 9 10 11 12 13 ALM4 ALM4 COM CTR1 CTR2 B+ B- MK SP TMP A /NMT+ TMP B/NMT- TMP b Fuse F1 L N G Lightning Arrester NRF or Non IS HART Devices Alarm Contact (Option) Alarm Contact (Option) Alarm Contact (Option) External contact Input (Option) Alarm Contact (Option) Common Hoist Stop Ground NMT or HART Device (Non IS) Mark / Space or Pt100 ohm Input NMT+ NMT- L GND N 16 P_B+ P_B- NRF+ NRF- Order Code Example 5 Wiring

Exd [ia], Output 1: Mark/Space (M/S) Protocol

Figure. 41: Electrical Connection of NMS5 (with M/S Protocol with IS HART Connection) NMS5 - X X M X X X X X X X X X J N O U A D 14 AL4 COM 18 CTR2 20 B+ 22 MK AL4 15 CTR1 17 N.C. 19 B- 21 SP 23 4 P_B+ 8 AL1 10 AL2 12 AL3 AL3 13 AL2 11 AL1 9 7 P_B- 5 6 24 P_A+ A 25 P_A- B 26 b 1 2 3 ARS L ARS G ARS N POWL POWN POWG ARSL ARSG ARSN P_B+ P_B- ALM1 / OSP ALM1 / OSP ALM2 / OSP ALM2 / OSP ALM3 1 2 3 L N G AC power Supply 24 25 26 + - + - + - ALM3 14 15 16 17 18 19 20 21 22 + - 23 4 5 6 7 8 9 10 11 12 13 ALM4 ALM4 COM CTR1 CTR2 B+ B- MK SP TMP A /P_A+ TMP B/P_A- TMP b Fuse F1 L N G Lightning Arrester NRF or Non-IS HART Devices Alarm Contact (Option) Alarm Contact (Option) Alarm Contact (Option) External Contact Input (Option) Alarm Contact (Option) Common Hoist Stop Ground NMT or HART Device (IS) Mark / Space or Pt100 ohm (IS) Input N L GND 16 Order Code Example 5 Wiring

Using the external operation switch, ensure that the settings of the system are the same as shown in the fol-

lowing logic setting table. This will allow user to input operation signals. (Hoist-CTR1)(Stop-CTR2) 5.2.7 Input and Output Input – Contact Switch – HART – Pt 100 Output – V1 – WM550 – M/S – ENRAF BPM – MODBUS – HART – ANALOG 4-20mA – Overspill Prevention (OSP) Warnung! • The cable used for input and/or output must be 24 AWG or more and must be screened or steel armored. A twisted pair is required for HART and/or RS 485 signal. • Two or three cores for mains, two cores for digital output and two cores for HART input are normally used for the cabling of NMS5. The instrument has max. four cable entries. • Check the cable size and the number of cables prior to ordering NMS5. 5.2.8 Cable Gland When there are extra cable entries not be used, mount threaded plugs for waterproof. Refer to "Safety Instruction XA00578G-A". Caution! When ordering TIIS Ex d specification, cable glands are attached with NMS5. Ensure to use the cable glands. CTR 1 CTR 2 OPERATION OFF OFF LEVEL ON OFF HOIST OFF ON STOP ON ON INTERFACE 5 Wiring

Temperature Input System

There are two types of temperature input method and two types of Ex protection system as Ex d or Ex [ia] in NMS5. 5.3.1 Ex d NMT Input + HART Input Ex d HART sensor and NMT53x Ex d are connected in this system. Figure. 42: Ex d NMT Input and HART Input 5.3.2 Ex d Pt100 Spot Temperature Input +HART Input Ex d HART sensor and Ex d Pt100 sensor are connected in this system. Figure. 43: Ex d Pt100 Spot Temperature Input and HART Input 26 4 24 25 J3 1A 1 3 5 J1 1B Power Module POW Ex d NRF 560 NMT 53x Ex d Ex d HART TCB CPU Module Terminal module HART Sensor Ex d HART Sensor Ex d 5 26 4 5 24 25 J3 1A 1 3 5 J1 1B Spot Temp. Module A B b Ex d HART HART+ HART- TCB CPU Module NRF 560 Ex d Pt 100 Power Module POW Ex d HART Sensor Ex d HART Sensor Ex d 5 Wiring

Ex ia NMT Input + HART Input

Ex d HART sensor and NMT53x Ex ia are connected in this system. Figure. 44: Ex ia NMT Input and HART Input 5.3.4 Ex ia Pt100 Spot Temperature Input + HART Input Ex d HART sensor and Ex ia Pt100 sensor are connected in this system. Figure. 45: Ex ia Pt100 Spot Temperature Input and HART Input 26 4 5 24 25 J3 1A 1 3 5 J1 1B Power Module POW Ex ia NRF 560 NMT 53x Ex ia Ex d HART TCB CPU Module Terminal Module Separation Wall Ex ia HART COM-4 Module Ex ia Ex ia Power Feed Ex d Power feed Ex d HART Sensor Ex d 26 4 5 24 25 J3 1A 1 3 5 J1 1B Spot Temp. Module A B b HART+ HART- TCB-4 CPU Module NRF 560 Ex ia Pt 100 Power Module POW Ex ia Separation Wall Ex ia Ex ia Ex ia Power Feed Ex d Power Feed Ex d HART Sensor Ex d HART Sensor Ex d Ex d HART Plastic Case Ex ia HART 5 Wiring

Touch Control and Programming Matrix

6.1.1 Display and Operating Elements Display During normal operation, NMS5 has an illuminated LCD that shows the level, the temperature, and the status of the device at "HOME" position. For the display of the other data and the programming of the parameters for operation, NMS5 uses a convenient programming matrix. Operating Elements NMS5 is operated by three visual operating elements, namely the keys "E", "+", and "-" . They are actuated when the appropriate field on the protective glass of the front is touched with the finger ("touch control"). The corresponding transmitting and receiving diodes are not affected by external influences, e.g. direct sunlight. The software and hardware installed in NMS5 rule out any malfunction that may be caused in this way. Even in explosive hazardous areas, the explosion-proof housing of the touch control ensures a safe access to the data. Figure. 46: Display 6.1.2 Functions of Operating Elements The programming matrix consists of matrix groups, namely one "static" matrix and additional "dynamic" matri- ces. They are described in detail in Section "13 Matrix". The individual matrix groups, function groups, and functions within the programming matrix can be selected by alternately touching the operating elements. Figure. 47: Displays Illuminated LCD 3 Optical Operating Elements "Touch Control" Infrared Receiving Diaode Infrared Transmitting Diaode LEVEL U-D 0.0°C LEVEL U-D 0.0°C LEVEL Data Temp. Data Displacer Status Symbol Gauge Status / Error MORE FUNCTION GVH:Position Group Message Key Guide G:0 VH:3* > GROUP SELECT < E Enter Matrix Go Back HOME Display Function Group Display Input Number Key Guide G:0 VH:39 ACCESS CODE 3 E - + Measurement Data G:0 VH:20 OPERATION 15955 LEVEL Matrix Display Matrix Display 6 Operation

Figure. 48: Matrix Construction

0 1 2 3 4 5 6 7 8 9 0 1 2 3 4 5 6 7 8 9 Measured Value 1 0 Measured Value 1 0 Measured Value 2 1 Measured Value 2 1 Operation 2 Operation 2 More Function 3 Calibration More Function 3 Device Data Level Data 4 Contact Output 4 Calibration 5 Analog Out 5 Adjustment 6 parts Data 6 Auto Wire Calib. 7 Input Signal 7 Auto Calib. Displ. 8 Communication 8 Display 9 Status 9 0 1 2 3 4 5 6 7 8 9 0 1 2 3 4 5 6 7 8 9 Measured Value 1 0 Measured Value 1 0 Measured Value 2 1 Measured Value 2 1 Operation 2 Operation 2 More Function 3 Service More Function 3 Temperature 4 Temperature Data 4 Gauge data 5 Element Temp. 5 System Data 6 Element Position 6 Service 7 NMT Adjustment 7 Sensor Value 8 Set Data NMT 8 Sensor Data 9 Device Data 9 0 1 2 3 4 5 6 7 8 9 0 1 2 3 4 5 6 7 8 9 Measured Value 1 0 Measured Value 1 0 Measured Value 2 1 Measured Value 2 1 Operation 2 Operation 2 More Function 3 HART Dev (1) More Function 3 HART Dev (2) Measured Value 4 Measured Value 4 P.V. Setting 5 P.V. Setting 5 Sensor Speci�c 6 Sensor Speci�c 6 Alarm 7 Alarm 7 Self Diagnostic 8 Self Diagnostic 8 Device Data 9 Device Data 9 0 1 2 3 4 5 6 7 8 9 0 1 2 3 4 5 6 7 8 9 Measured Value 1 0 Measured Value 1 0 Measured Value 2 1 Measured Value 2 1 Operation 2 Operation 2 More Function 3 Adjust Sensor More Function 3 Tank Pro�le Adjust Sensor 4 Pro�le Operation 4 HART Error Rate 5 Status/Data 5 Unit 6 Density 1-10 6 HART Line 7 Density 11-16 7 Interface Adjust 8 Position 1-10 8 9 Position 11-16 9 0 1 2 3 4 5 6 7 8 9 Measured Value 1 0 Static Matrix Dynamic Matrix Measured Value 2 1 Operation 2 More Function 3 Interface Pro�le 4 Status/Data 5 Density 1-10 6 Density 11-16 7 Position 1-10 8 Position 11-16 9 Meas. Wire & Drum 6 Operation

LCD returns to HOME position when no key is touched for 10 minutes or more. Digits are incremented or

decremented using by + or - key. When touching + or - key continuously, then the minimum digit changes first. After one cycle of the minimum, the second minimum changes. After one cycle of the second follows the third minimum, and so on. When releasing a finger from the touch control, then the procedure starts again from the minimum digit (Analogy of mechanical counter). Figure. 49: Selection Matrix Groups, Function Groups, and Functions with Programming Matrix Key Functions • Access to the programming matrix (touch the key for 3 seconds or more). • Return to HOME position (touch the key for 3 seconds or more). • Move the status horizontally in the function group to select functions. • Save parameters or access codes. • Move the status vertically to select function groups. • Selec or set parameters. • Set access codes. + - 4 5 6 7 8 9 4 5 6 7 8 9 H V LCD ( HOME Position) When touching on the matrix for 3 seconds or longer, the screen returns to HOME position. 1 CALIBRATION 3 SERVICE 4 TEMPERATURE 5 HART DEVICE (1) 6 HART DEVICE (2) 7 ADJ. SENSOR 8 TANK PROFILE 9 INTERFACE PROFILE 0 1 2 3 4 5 6 7 8 9 0 1 2 3 4 5 6 7 8 9 4 4 4 4 7 ADJ. SENSER 3 SERVICE 4 TEMPERATURE 5 HART DEVICE(1) 6 HART DEVICE(2) > 3 s 2 DEVICE DATA MATRIX OF G0 V3H0 4 2 DEVICE DATA 4 1 CALIBRATION 5 6 7 8 9 5 5 5 5 5 6 6 6 6 6 7 7 7 7 7 8 8 8 8 8 9 9 9 9 9 When touching "E" on the matrix for 3 seconds or longer, the screen moves to V0 position. 8 TANK PROFILE 9 INTERFACE PROFILE 6 Operation

After turning on the power, LCD shows the current data on HOME position.

Figure. 50: HOME Position The letters A, B, C and D stand for the areas where information on measured values and status of the device are displayed: The meanings of message appearing on LCD (display) is explained in the following table. Area Information A Current Level B Current Temperature C Gauge Status D Displacer Status Gauge Status Description G - RE Reference position for measuring UP The measuring wire is hoisted. STOP The measuring wire stops. LIQU The liquid is being measured. U - IF The upper interface level is being measured. LIF The lower interface level is being measured. BOTM The tank bottom level is being measured. U - DE The upper liquid density is being measured. M - DE The middle liquid density is being measured. B - DE The bottom liquid density is being measured. CAN Resetting RELE.OVER TENS (over tension error). TEAC Calibration is being measured. blank Measuring is not possible. 6 Operation

If no LCD operation, NMS5 will turn off the backlight of LCD 12 hour later. Touching LCD again after this

time will turn on the backlight. Displacer Status Symbol Meaning BAL Balance The displacer is resting on the liquid surface or interface and in bal- anced status. T - B Temporary Balance Automatic weight calibration is being carried out. U - U Unbalance Up The displacer is being hoisted and in unbalanced status. U - D Unbalance Down The displacer is being lowered and in unbalanced status. R - U Balance Up The displacer is being hoisted and in correction of balance. R - D Balance down The displacer is being lowered and in correction of balance. LOW The displacer is resting at the lower stop. 6 Operation

The access code is to ensure the confidentiality of the setup data. Three security levels are available along with

access codes. The higher levels include the lower ones. e.g. If access code 50 is specified for a function, then code 51 also enables editing. A function that requires access code 51, on the contrary, cannot be edited by code 50. However, 777 is only used to change I.S. terminal configuration. Figure. 51: Order code which require Access Code 777 Setting Access Code Security Level Access Code 0 None 1 For Operator 50 2 For Engineer 51/530/777 Item Procedure Remarks Static Matrix Figure. 52: Access Code 1. At the static matrix "MORE FUNCTION", select GVH=039 "ACCESS CODE" 2. The default value is "0". Touch the "+" key. 3. The first digit increases to 9, then the second digit increases. Stop touching "+" once you reach "50". 4. "50" is blinking. Gently touch "+" key again to change the first digit from 0 to 1. Now you have "51". 5. Here touch "E"; "EDITING ENABLE" will be displayed. • When touching "E" while displaying an access code except 0, 50, or 51 "EDITING LOCKED" will appear. • If an access code has not been selected before performing any settings, the screen will automatically change to show "ACCESS CODE" • Operation Commands can be sent, and displayed data read, by remote systems, depending on your NMS5 specification. NMS5- X X X X X X X X X X X J N N P F A J B C D E G MORE FUNCTION ACCESS CODE 039 6 Operation

Operation Command and New Operation Status

6.4.1 Operation Commands Operation Commands can be sent the NMS5 from a host system. The following table explain the command codes. 6.4.2 New Operation Status The following table shows the new operation status, which is available when "NEW NMS STATUS", matrix position GVH=272, is selected to "ENABLED". Code Command Remarks 0 LEVEL 1 UP 2 STOP After weight calibration, STOP is set as a default operation command. 3 BOTTOM LEVEL 4 UPPER INTERF. LEVEL 5 MIDD.INTERF.LEVEL 6 UPPER DENSITY 7 MIDDLE DENSITY 8 DENSITY BOTTOM 9 REPEATABILITY 10 WATER DIP Code Meaning NMS Display 0 No definition - 1 Displacer at reference position REFERENCE 2 Displacer hoisting up UP 3 Displacer going down DOWN 4 Displacer stop STOP 5 Level measurement, balanced LEVEL 6 Upper I/F level, balanced UPPER INTERF. LEV. 7 Middle I/F level, balanced MIDD. INTERF. LEV. 8 Bottom meas. Balanced BOTTOM LEVEL 9 Upper density finished UPPER DENSITY 10 Middle density finished MIDDLE DENSITY 11 Bottom density finished DENSITY BOTTOM 12 Release over tension RELE. OVER TENS. 13 Calibration activated CAL. ACTIVE 14 Seeking level LEVEL SEEKING 15 Following level LEVEL FOLLOWING 16 Seeking upper density UPP.DEN.SEEKING 17 Seeking middle density MID.DEN.SEEKING 18 Seeking bottom density BOT. DEN. SEEKING 19 Seeking upper I/F level UPP. INT. SEEKING 20 Following upper I/F level UPP. INT. FOLLOWING 6 Operation

Seeking middle I/F level

MID.INT.SEEKING 22 Following middle I/F level MID.INT.FOLLOWING 23 Seeking bottom level BOTTOM SEEKING 24 Not initialized NO INITIALIZE 25 Stopped at upper pos. UPPER STOP 26 Stopped at lower pos. LOWER STOP 27 Repeatability testing REPEATABILITY 28 Seeking water level WATER SEEKING 29 Water level, balanced WATER LEVEL 30 Following water level WATER FOLLOWING 31 Over-/under tension, Z-phase, ADC error EMERGENCY ERROR 32 GVH157 SERVICE MODE = ON MAINTENACE Code Meaning NMS Display 6 Operation

Operation of Displacer

The operation of the displacer for level, bottom level, interface level, and density measurement is possible by touch control. Item Procedure Remarks Figure. 53: Static Matrix Figure. 54: Touch Control 1. Select function groupGVH=020 OPERATION 2. Select item OPERATION. The LCD shows the command given to NMS5and the displacer position. • Set access code to 50. • If editing has previously been enabled by a valid access code, then the request for the code will not appear. Figure. 55: Operation of Displacer 3. The following commands are available at this position: LEVEL UP STOP BOTTOM LEVEL UPPER INTERF. LEV* MIDD.INTERF.LEV* UPPER DENSITY* MIDDLE DENSITY* DENSITY BOTTOM* WATER DIP* REPEATABILITY TEST • Interface and density measurements are not available unless specifically requested. • Density profile measurement is explained in section "7 Commissioning". • The optional operation commands are available when such options are set. OPERATION OPERATION Level Bottom Level Upper Interface Level Middle Interface Level Density Bottom Water Dip Middle Density Upper Density 6 Operation

Calculation of Level and Densities

The section specifies the formula used by NMS5 to calculate levels and densities. Figure. 56: Calculation of Levels and Densities Surface and Interface Levels While the measured surface or interface level is constant, the displacer rests at the balancing position. The ten- sion on the measuring wire is proportional to the displacer weight diminished by the buoyancy forces in both layers: A rise or fall of the level will increase or decrease the submerged volume. If this change exceeds the volume tolerance set at matrix position GVH=345 "VOLUME TOLERANCE", then the corresponding change will actuate the motor of the NMS5 till the balancing condition is fulfilled again. Tank Bottom Level For bottom level measurement, the balancing condition is defined as Level Formula Liquid Level T = W - Vвρu Upper Surface T = W - Vвρm - (V - Vв) ρu Middle Surface T = W - Vвρb - (V - Vв) ρm Measuring Wire Tension T Gas Density Negligible Measured Displacer Weight W Upper Liquid Density �� Measured Displacer Weight Wu Middle Liquid Density �� Measured Displacer Weight Wm Bottom Liquid Density �� Measured Displacer Weight Wb Balance Volume VB Displacer Volume V 6 Operation

The upper, middle, and bottom densities are calculated by the following formulas.

• Upper density (Pu) and Density profile • Middle density (Pm) • Bottom density (Pb) Draft The draft varies depending on the shape of the displacer. For cylindrical shape, the draft is where the variables and constants have the following meanings. Figure. 57: Variables and Constants for Displacer The surface or interface level should be within the cylindrical part of the displacer and approximately in the middle of its total height. ρu = V W - Wu ρm = V Wu - Wm ρu + ρb = V Wm - Wb ρm + D = (V2 - V1) / A x 10 + h = (V1 - V2 - V3) / A x 10 + h + h1 V1 V2 h A Displacer Balance Volume (69cm3) Displacer Lower Hemisphere Volume Height of Lower Cone Cross-section of Displacer V1 V2 h A Balance Volume (69 cm3) Volume of Lower Cone (24.46 cm3) Displacer Bottom Stub Height (20 mm) Cross-section of Displacer (20.26 cm2) For example: standard displacer Ø50.8 mmCylindrical shape D = (69 - 24.46 - 0.12) / 20.26 x 10 + 20 + 3 = 44.92 V3 Displacer Bottom Stub Volume V3 Volume of End of Cone (0.12 cm3) h Height of End of Cone 1 1 d h V2 D h1 V3 Liquid Level or Interface 6 Operation

All of the initial settings are required for NMS5 in most case. However, depending on NMS5 specification,

some settings may not be required. Set Access code 51 at GVH039 before applying the initial settings. 7.1.1 System Calendar/Clock (GVH193 to 197) Calendar/Clock values are set at factory (Japan Standard Time) prior to delivery. Change the data to reflect local time. 7.1.2 Density Values (GVH005 to 007) Caution! Always set GVH=005 Upper Density for actual density data, in LPG and any application where the actual den- sity is 0.7000 g/ml or less. failure to set this data may result in level gauge malfunction. Density values for 3 liquid phases are set to 1.000 g/ml prior to delivery. Change the data to reflect actual den- sity values. For tanks with only one liquid phase, set Upper Density. For tanks with 2 or 3 clear phases, set Middle and Bottom densities too. Note: minimum difference between phase settings should be at least 0.100 g/ml. *GVH005£GVH006£GVH007 Example: • GVH005 Upper Density: 0.758 • GVH006 Middle Density: 0.880 • GVH007 Density Bottom: 1.000 7.1.3 Tank Height (GVH140) Tank height value is set to default value at factory prior to delivery. Tank height is the reference height, usually a gauging hatch on the tank that is used during manual level measurements. Change GVH140 to equal the reference height. Note! GVH141 "Dip Point Offset" will automatically change to reflect the difference in height between the reference height and the NMS5 reference position. Figure. 58: Tank Height Tank Height Dip Point Offset (Difference between the dipping hole height and the reference line of the calibration window) 7 Commissioning

Upper/Lower Stop (GVH161/162)

Upper and Lower Stop determine the highest and lowest point of displacer movement. These data are set to 16000mm and 0mm respectively at factory prior to delivery. Change these data to the desired actual Upper and Lower limit values. 7.1.5 Communications Address (GVH285) Several types of digital output communication protocols are available when ordered from Endress + Hauser. NMS5 is delivered with communications hardware and/or software installed. Only minimal matrix parameter setting, Access code 51, is required. Communications protocol is displayed and selected at G2V8H6 Protocol. NMS5 has been preset at factory, it is not necessary to change the setting. At GVH285 select the desired device address for NMS5. Address range: 0 to 9, 00 to FF, or 1 to 247 for Mod- bus. Note! FF is fixed for MIC protocol. WM550 and M/S address setting is done via dip switches on the communications module inside NMS5. For Rackbus, termination resistors should be set on NMS5 at end of loop. Enraf BPM address range is 00 to 99. Refer to Sect. 7.5 for Address Setting details. 7.1.6 Proactive Safety Function (GVH157/158/159) Safety Level Default by Output Protocol Matrix Setting G1V5H7 Service Mode: default = OFF. Select OFF if GVH158 Prosafety = ON. Select ON only when performing maintenance on NMS5. Caution! Selecting ON disables the Proactive Safety function! Access code 530 G1V5H8 Prosafety: default = ON. Select OFF only if the Proactive Safety function will not be used. Caution! Failure to confirm ON disables the Proactive Safety function! Access code 530 G1V5H9 Safety Level: default = 65000.0mm (WM550.), or = 99999.0 mm (other protocols) Possible data values depend on receiver specification (see table below) Access code 530 Protocol Data Range Data Format Modbus 0.0 to 99999.0mm Float V1 0 to 99999.9mm ASCII / 6 digit MDP 0 to 99999mm BCD / 5digit WM550 0 to 65000mm 16bit Mark/Space 19.999m/32.699m 20bit BCD / 5 digit Enraf BPM 0 to 99999.9mm 6 or 7 byte Rackbus 0 to 99999.9mm Float HART 0 to 99999.9mm Float 7 Commissioning

Analogue Output (GVH250 to 256)

When specified and ordered from Endress+Hauser, NMS5 is equipped with analogue output hardware installed. Function settings may be changed as follows: 7.1.8 Contact Relay Alarm Output (GVH240 to 247) When specified and ordered from Endress+Hauser, NMS5 is equipped with contact relay alarm output hard- ware installed. Function settings may be changed as follows: Function Setting Assign Output 1 Assign level or temperature to channel 1 output. Adjust 4mA Set desired value at which level or temperature outputs 4 mA. Adjust 20mA Set desired value at which level or temperature outputs 20 mA Assign Output 2 Assign level or temperature to channel 2 output. Adjust 4mA Set desired value at which level or temperature outputs 4 mA Adjust 20mA Set desired value at which level or temperature outputs 20 mA. Device at Alarm Select type of output for alarm Function Setting Select Relay Select relay number from 1, 2, 3 or 4. Assign Relay Select output definition from range of choices: None, Level, Liquid Temperature, Caution, Warning, Emergency Error, Balance Signal. Relay Function Select High or Low. Switching Point Set value at which relay is activated. Hysteresis Set hysteresis value for selected relay. Relay on Alarm Select relay of Normal Open or Normal Closed. On Delay Time Set time delay value for alarm output start. Off Delay Time Set delay value for alarm output stop. 7 Commissioning

Settings for NMT 53x Connections

The following settings are required to display NMT 539 data on NMS5 screens. Caution! For NMT535/539/532 connections, refer to the appropriate operating manual for the device. Item Procedure Remarks Figure. 59: Matrix Group: Service Static Matrix Figure. 60: Static Matrix 1. In Static Matrix "MORE FUNCTION", invoke GVH=030 "MATRIX OF" and select "SERVICE". 2. Invoke the Dynamic Matrix GVH=362 "CONNECTION NMT" screen. 3. Use the "+" and "-" keys to display "AVERAGE" and press "E" to set data. 4. Press "E" to return to "SYSTEM DATA" and press the "-" key to return to "MORE FUNCTION". 5. Invoke Static Matrix "GVH=030 MATRIX OF". NMS5 matrix is divided into matrix groups. Select "TEMPERATURE" from these groups. 6. "EDITING ENABLED" is displayed on the LCD. 7. The average liquid temperature is displayed on Dynamic Matrix screen GVH=440. 8. The temperature of each contact is displayed on Dynamic Matrix screens GVH=450 to GVH=459. • Set access code 51. • NMS HART connection (terminal 24 & 25) must have corresponding IS input in order to configure NMT IS version. • The Static Matrix GVH=010 "LIQUID TEMP." screen also displays the average temperature. MORE FUNCTION SYSTEM DATA CONNECTION NMT 3 6 2 MATRIX OF 0 3 0 MORE FUNCTION MATRIX OF 0 3 0 7 Commissioning

Settings for Connection to NRF560

To connect NRF560, NMS5 requires the following settings. Caution! Turn on the power to NMS5 first. Item Procedure Remarks Figure. 61: Matrix Group: Service 1. In Static Matrix "MORE FUNCTION", invoke GVH=030 "MATRIX OF" and select "SERVICE". 2. Invoke the Dynamic Matrix GVH=361 "CONNECTION NRF" screen. 3. Use the "+" and "-" keys to select either "CONTACT 1" or "CONTACT 2". • Set access code 51. • CONTACT 1... NRF 560 software version 1.6x and earlier (those NRF 560 that indicate no software version correspond to connection type 1). • CONTACT 2... NRF 560 software version 1.8x and later. MORE FUNCTION SYSTEM DATA CONNECTION NRF 3 6 1 MATRIX OF 0 3 0 7 Commissioning

Liquid Level Calibration

7.4.1 Opened Tank Preparation Select LEVEL at GVH020 Operation. The displacer will descend to the liquid surface and balance. When "BAL" is displayed, the displacer stopped moving. When measuring the liquid level manually, use an approved method. Set Level (GVH150) for Opening Tank Input the value at GVH150 after measuring the dipping level. Figure. 62: Calibration Level (Manual Dipping Level) When there is not liquid in the tank; Figure. 63: Calibration Level (Empty Tank) Note! When there is no liquid in the tank, level calibration is not 0mm on the measurement principle. When 0mm level calibration is needed, refer to GVH004(BOTTOM LEVEL) or GVH142(DISPLAC. DRAFT). Manual Dipping Level Draft Calibration Level B A B=45mm Note: Ø50mm (Dtandard Displacer) 140ml (Displacer Volume) 70ml (Balance Volume) A= Depend on Tank Design 7 Commissioning

Level Calibration Procedure for Closed Tank

Closed tanks (e.g. LPG) cannot be hand-dipped. Follow the procedure shown below. Figure. 64: Level Calibration 1 1. Check the NMS parameter 2. Set the Level gauge at GVH020(BOTTOM) Figure. 65: Draft Level of Displacer 3. Check that the displacer is balance and LCD (display) shows Balance. 4. Calculate the displacer balance position. • In case of Ø50 standard displacer, when balance volume(GVH343) is reference value (a half of displace volume), the balance point is 45mm from the bottom of the tank. Note! Refer to Appendix "14.3 Displacer". GVH ITEM 005 UPPER DENSITY 340 WIRE DRUM CIRC. 342 DISPLACER WEIGHT 343 DISPLACER VOLUME 344 BALANCE VOLUME BOTTOM plate (datum plate) BALANCE position for displacer 7 Commissioning

Level Calibration 2

5. Set Level Calibration to 45mm at GVH150 (SET LEVEL) in G1 MATRIX (Calibration).The bottom is ZERO. • If the bottom plate is not Zero (ex; X mm), adjust the level value by adding to the value. LEVEL = X + 45(mm) This completes the level calibration procedure. 0mm 45mm 45mm 7 Commissioning

The desired loop address for most digital protocols can be set at G2V8H5 (Address). The allowable range for

setting varies depending on the protocol installed in NMS5. The range has 0-9, 00-FF (FF is fixed for MIC pro- tocol), or 1-247 for Modbus. • WM550 and Mark Space protocols addresses must be set by switches on the communications module inside NMS5. • MODBUS: Termination resistors should be set on BODBUS modules (refer to Appendix "13.1 RS-485 MODBUS (COMM-5) Terminator"). • Enraf BPM: Address range is 00-99. A-F is not available. 7.5.2 Whessoe Matic 550 (WM550) Communication Board Setting Jumper Setting Jumper settings for WM550 communication board To loaded software into processor [IC1] J3 has to be open. If software testing is required; J4 has to be open. If software reset is required; J6 has to be short. Polling Address Setting Caution! Polling addresses are set by mechanically at SW1 on WM550 communication board (not by acccessing NMS programming matrix). Check all polling addresses before setting. The following table shows about address setting. Note! For current loop setting, refer a operating manual for Whessoe 1098 or RTU 8130 operating manual. Jumper Function Default Condition J3 (Mode) Use EPROM [IC4]->short Short J4 (Test) Software testing Short J6 (Reset) Reset Open J7 (Watch dog) Watch dog setting Short Switch Position Value 1 1 2 2 3 4 4 8 5 16 6 COMPATIBILITY MODE ON 1 2 3 4 5 6 Polling address = 1 ON 1 2 3 4 5 6 Polling address = 5 ON 1 2 3 4 5 6 Polling address = 9 ON 1 2 3 4 5 6 Polling address = 3 ON 1 2 3 4 5 6 Polling address = 7 Setting Example 7 Commissioning

Mark/Space(M/S) Communication Board Setting

Jumper Setting Jumper settings for M/S communication board Polling Address Setting Caution! Polling addresses are set by mechanically at SW1 (1-8) and SW2(1-2) on Mark Space communication board (not by accessing NMS programming matrix). Check all polling addresses before setting. The following table shows address setting. Function Setting Each function is set at SW2-* shown as follows. Jumper Function Default condition J3 (Mode) Use EPROM [IC4]->short Short J3 (Reset) Reset Open J3 (WD) Watch dog setting Short Switch Position Value SW1-1 1 SW1-2 2 SW1-3 4 SW1-4 8 SW1-5 16 SW1-6 32 SW1-7 64 SW1-8 128 SW2-1 256 SW2-2 512 Switch Function Default Condition SW2-3 ON: Data transmission under unbalanced condition ON SW2-4 ON: low baud rate setting OFF SW2-5 ON: Feet data OR 0-20m, OFF: 0-30m OFF SW2-6 ON: Measured data converted to feet OFF SW2-7 ON: Temperature data (57 bits) ON SW2-8 ON: Deg. F measured temperature unit, OFF: Deg. C OFF ON 1 2 3 4 5 6 7 8 ON 1 2 3 4 5 6 7 8 SW - 1 SW - 2 Address Setting Function Setting Setting Example (example above:267) 7 Commissioning

ENRAF Bi Phase Communication Board (COM-3) Setting

Jumper Setting COM-3 Communication Board Jumper Setting Communication Setting Select "V1/ENRAF BPM" at the matrix position GVH=286. Communication Setting Polling address is set at GVH=285 "ADDRESS" of NMS5 matrix. The allowable range is from 00 to 99. Caution! A to F for polling address can not use. Communication Setting • Switch S1: Reset switch • Connector J3: Communication port for debugging Jumper Function Default Condition JP1 (Mode) ROM type setting 1-2 short: 27C4096 2-3 short: 27C1024 2-3 short JP2 (Reset) CPU mode setting (Fixed) 1-2 open 3-4 short 5-6 short J3 (WD) Baud rate setting 1-2 short: 1200 bps 1-2 open: 2400 bps 1-2 open 7 Commissioning

Setting of analog output 4-20mA is available when ordered from Endress+Hauser. NMS5 is delivered with ana-

log output hardware installed. For the following matrix parameter settings, Access code 51, is required. Note! If "LEVEL", confirm GVH158 (Prosafety) is "NO". Matrix Item Setting G2V5H0 Assign Output Assign analog output (Level or Temperature) for channel 1. G2V5H1 Adjust 4mA Set level or temperature value for 4mA output on channel 1. Available only when G2V5H0= "Level" or "Liquid Temp" Level: 0 to 99999mm Temperature: -999 to 999°C Default: 0mm / 0°C G2V5H2 Adjust 20mA Set level or temperature value for 20mA output on channel 1. Avail- able only when G2V5H0 = "Level" or "Liquid Temp." Level: 0 to 99999mm Temperature: -999 to 999°C Default: 0mm / 0°C G2V5H3 Assign Output 2 Assign analog output (Level or Temperature) for channel 2. G2V5H4 Adjust 4mA Set level or temperature value for 4mA output on channel 2. Available only when G2V5H3 = "Level" or "Liquid Temp." Level: 0 to 99999mm Temperature: -999 to 999°C Default: 0mm / 0°C G2V5H5 Adjust 20mA Set level or temperature value for 20mA output on channel 2. Available only when G2V5H3 = "Level" or "Liquid Temp." Level: 0 to 99999mm Temperature: -999 to 999°C Default: 0mm / 0°C G2V5H6 Device at Alarm Select type of output for alarm. Select from OFF, HOLD current output, Maximum value or Minimum value. Default : OFF 7 Commissioning

Contact Relay Alarm Output

The specification of alarm for output contact relay is delivered as a part of NMS5. This specification allows a maximum of 4 output contact relay settings. Set access code 51 in case of the following matrix parameters. Description of Error Message Matrix Item Setting G2V4H0 Select Relay Use +/- and E keys to select to activate Contact Output Relays 1 to 4. Default value = 1. G2V4H1 Assign Relay Select output definition from range of choices: None, Level, Liquid Temp, Caution, Warning, Emergency Error, and Balance Signal. Default value = NONE. G2V4H2 Relay Function Select High or Low function, available only when G2V4H1 = "Level" or "Liquid Temp." Default value = HIGH. G2V4H3 Switching Point Set level (0-99999 mm) at which relay is activated, available only when G2V4H1 = "Level" or "Liquid Temp." Default value = 0 mm. G2V4H4 Hysterisis Set hysterisis value (0-99999 mm) for selected relay, available only when G2V4H1 = "Level" or "Liquid Temp." Default value = 0 mm. G2V4H5 Relay on Alarm Select from Normal Open or Normal Closed, available only when G2V4H1 = "Level" or "Liquid Temp." Default value = NORMAL OPENED. G2V4H6 On Delay time Set delay time (0-999 seconds) for alarm output start, available only when G2V4H1 = "Level" or "Liquid Temp." Default value = 0 seconds. G2V4H7 Off Delay Time Set delay time (0-999 seconds) for alarm stop, available only when G2V4H1 = "Level" or "Liquid Temp." Default value = 0 seconds. Error Type Descriptions LEVEL Liquid level exceeded the designated upper or lower limit. LIQUID TEMPERATURE Liquid temperature exceeded upper or lower limit. CAUTION Auto wire calibration error; Auto displacer calibration error WARNING Weight value exceeded the designated alarm sending limit (upper limit: GVH=162 or lower limit: GVH=163). (e.g. local communication error, LCD error, communication error, EEROM data error) EMERGENCY ERROR Z-phase no input error, ADC sensor error, communication IC error, A-phase no input error, Driver error, power failure 7 Commissioning

I/O - 3

Alarm Output, Standard (Order Code Position 050= 1,2,3 or 5) 4 relays with potential-free change-over contacts, freely assignable to measured value Hysteresis, Alarm Output Switch points and switching hysteresis freely adjustable, residual current fail-safe mode: minimum or maximum, selectable Operation Output Logic Switching Capacity, Alarm Output • Max. 250VAC, 2A / 62.5VA • Max. 220VDC, 2A / 60 W • FM/CSA: 30VAC, 2A/42VDC, 2A, 60W Alarm Output, Overspill Prevention (Order Code Position 050=4) 2 relays with potential- free changeover contacts, assignable to Level Hysteresis, Alarm Output Switch points and switching hysteresis freely adjustable, residual current fail-safe mode: minimum or maximum, selectable Operation Output Logic Switching Capacity • U max.: 200VDC/200VppAC • I max.: 0.5A, DC or peak AC • P max.: 15W Relay Condition Initial Configuration (normal status) Open Closed On alarm Closed Open On device Error Closed Open On abnormal power supply (5% lower than, or 5% higher than specified power supply) Custody Transfer Closed Custody Transfer Open Power outage Hold last condition Relay Condition Initial Configuration (normal status) Closed On alarm Open On device Error Open On abnormal power supply (5% lower than, or 5% higher than specified power supply) Open Power outage Open 7 Commissioning

2 photo couplers, for external input from controller (tumbler switch,

DCS, etc.) Operation output logic Input voltage 15VDC, active circuit (supplied by NMS5) Input current Approximately 5mA Gauge Status CTR1 CTR2 LEVEL UP STOP INTERFACE LEVEL 0 (OFF) 1 (ON) 0 (OFF) 1 (ON) 0 (OFF) 0 (OFF) 1 (ON) 1 (ON) 7 Commissioning

Density and Density Profile Measurements

Two different types of density measurement are available when ordered from Endress+Hauser. • Spot density measurement for up to 3 liquid phases • Density profile measurement 1. Tank profile for up to 16 points for total liquid height. 2. Tank profile for upper phase liquid only. 3. Tank profile based on a manually entered profile parameter. 7.6.1 Spot Density Measurement Spot measurement for up to 3 liquid phases is an option that is available when ordered from Endress+Hauser. The following preconditions are required in order to assure safe and accurate operation. • NMS5 must be configured for Spot density measurement as ordered from Endress+Hauser. For information regarding upgrading your NMS5, contact Endress+Hauser. • Density values: Figure. 67: Density Measurement – Density values can be viewed and changed at the following matrix positions: G0V0H5 Upper Density G0V0H6 Middle Density G0V0H7 Density Bottom • The displacer is balanced at liquid level and the liquid surface is not moving. For spot density measurement, the following parameters can be set or selected. However the factory default settings also can be used. Operation Select the measurement of the density desired from UPPER DENSITY, MIDDLE DENSITY, or DENSITY BOT- TOM at G0V2H0 "OPERATION". NMS5 raises the displacer up out of the liquid, measures its weight, and sends the displacer down to measure the density. After density is measured, the displacer remains at that posi- tion until UP or LEVEL command are selected at G0V2H0. Matrix Item Setting G1V4H3 Displacer Raise Density Set distance (0-300mm) for displacer to rise the level position above during measuring density. Default setting = 150 mm. Observe the rule: G1V4H3 + G1V4H4 = 300 n, where n = integer. G1V4H4 Displacer Submerg Density Set distance (0-300mm) for displacer to sink the level position above during measuring density. Default setting = 150 mm. Observe the rule: G1V4H3 + G1V4H4 = 300 n, where n = integer. - Upper Density < Middle Density < Bottom Density Middle Density - Upper Density 0.100 g/cm3 Bottom Density - Middle Density 0.100 g/cm3 e.g. 0.760 g/cm3 e.g. 0.880 g/cm3 e.g. 0.1000 g/cm3 (Example) 7 Commissioning

Density Profile Measurement

7.6.2 Density Profile Measurement Tank Profile Density Measurement In addition to the preconditions, the following additional preconditions are required in order to assure safe and accurate operation for the density measurement of the tank profile. Access Code 51 is required to edit parameters. • NMS5 must be configured for Spot & Profile density measurement as ordered from Endress+Hauser. Contact Endress+Hauser for the information regarding upgrading NMS5. • The standard tank bottom level (GVH004) is zero. If not, the formula used for calculating the interval between profiles is GVH000 / GVH841. The displacer touches the bottom of the tank before measuring a lower level (e.g. 884mm). • If Bottom Level is not 0 mm, select Manual I/F Profile at GVH840. Set GVH843 I/F Manual Level = actual bottom level e.g. 884mm. In this case, the intervals will be calculated based on GVH000 - GVH843/ GVH841. • The density profile can be operated when the liquid is still in the tank. However, depending on the height of liquid level, it may take 1 hour or more to operate the density profile. Note! Additional matrix parameter settings are required for remote communications via WM550 communications protocol. Refer to Section "13. Matrix, (Dynamic Matrix, Device Data : G2), WM550 DENS. SEL.". Matrix Item Setting G1V5H4 Safe Density Select the desired resultant condition when density profile measurement fails due to displacer reaching the low-limit for density profile operation (set in G1V5H5). Selecting "ON" will result in STOP operation. Select- ing "IGNOR" will result in "LEVEL" operation, displacer will return to liq- uid level. Selecting "OFF" will leave the displacer at the position where density profile measurement failed, and in "DENSITY" operation. G1V5H5 Den Ope Level Set the lower limit for displacer movement during density profile operation. Default setting = 300 mm. Upper Layer (Gas or Liquid) Lower Layer (Liquid) A B 7 Commissioning

A. Tank profile for up to 16 points for total liquid height

G8V4H0 Ope Select: Select 1: Tank Profile. • B. Tank profile for upper phase liquid only G8V4H0 Ope Select: Select 2: I/F Profile. • C. Tank profile for based on a manually entered profile parameter. G8V4H0 Ope Select: Select 3: MANU.I/F Profile. When choosing A, B, and C above, the following parameters can be set or select. The factory default settings also can be used. Initial Setting of Tank Profile (Density Profile) Measurement All of necessary set up can be done at "G8 Tank Profile" matrix. There are six factors to configured prior to the actual operation. G8V4H0 OPE. SELECT Matrix Item Setting G8V4H1 Ope Point Set the number of measurement points (2-16) for Tank Profile mea- surement. Default setting = 2. G8V4H3 I/F Manu. Level Manually set the water interface level (0-99999.9mm) to be referenced during the tank profile measurement. Default setting = 0 mm. Only used with C, tank profile for based on a manually entered profile parameter. G8V4H4 Bal. Level Set the allowance for level movement (1.0 -99.9 mm) prior to Tank Profile operation. If the level movement exceeds this value, profile operation is canceled. Setting 99.9mm allows Profile measurement to proceed, regardless of level movement. Default setting = 2.0 mm. G8V4H5 Up Wait Time Set waiting time (1-31 minutes) for displacer weight to be measured in air, when making the weight table at start of profile operation. Default setting = 1 minute. G8V4H6 Liq Wait Time Set the time (1-31 minutes) for displacer to stop at each measurement position during profile operation. Default setting = 1 minute. G8V4H7 Ope Wait Time Set maximum waiting time (1-31 minutes) for conditions in G8V4H4 (Bal Level) to be satisfied before profile operation begins. E.g. if the level movement exceeds setting at G8V4H4, NMS5 is in "standby" for the time set in this matrix. After the maximum time elapses, profile operation is canceled. Default setting = 1 minute. Item Procedure Remarks Figure. 69: G8V4H0 1. Select G0V3H0 "MATRIX OF" in the row "MORE FUNCTION" on Static Matrix, then select "Tank profile". • Set access code to 50. 2. Select function group "PROFILE OPE" and G8V4H0 "OPE SELECT". • The row of PROFILE OPE is the only function group that needs configuration for the density profile operation. 3. Select 1: "TANK PROFILE" by pressing + or - key. 4. Press E to configure. • Selection of "0: SPOT" enables Upper Density measurement instead of Density Profile measurment. Static Matrix MORE FUNCTION Matrix of TANK PROFILE PROFILE OPE. OPE. SELECT STATUS DATA DENSITY 1-10 DENSITY 11-16 Position 1-10 Position 10-16 Dynamic Matrix Function Group GVH=840 GVH=030 Matrix Group : TANK PROFILE 7 Commissioning

G8V4H1 OPE. POINT

G8V4H4 BAL. LEVEL G8V4H5 UP WAIT TIME Item Procedure Remarks Figure. 70: G8V4H1 1. Select G8V4H1 "OPE POINT " to determine number of density measuring points. • Set access code to 51. Default value is 2 points and selectable up to 16 points. 2. Enter required number by pressing + or - key, and select E to enter. • Whatever the number of measuring point is selected, the measuring interval & point between each point are automati- cally calculated within the liquid level. Item Procedure Remarks Figure. 71: G8V4H4 1. Select G8V4H4 "BAL LEVEL" to determine the liquid stability range in order to start the operation. • Set access code to 51. NMS software has a func- tion to scan & record the liquid level every 5 min- utes. 2. Enter desired value by pressing + or - key and select E to enter. Input value of 99.9mm represents that the measuring procedure starts regardless of level condition. • NMS5 only starts the actual measurement when the liquid level is within the entered value compared to the recorded level data 5 minutes ago. This function ensures the safe operating condition to avoid filling or draining liquid from the tank. Item Procedure Remarks Figure. 72: G8V4H5 1. Select G8V4H4 "UP WAIT TIME" to determine the displacer weight calibration interval in air. • Set access code to 51. The displacer raises approxi- mately 500mm above liq- uid surface after confirm- ing the stability of liquid level within "BAL LEVEL" function. 2. Enter desired waiting time by pressing + or - key, and select E to enter. • The displacer weight is calibrated at 8 different positions in air and the interval between those positions are adjustable within this function to allow excess liquid to drip off the displacer to achieve higher accuracy. PROFILE OPE. OPE. POINT Function Group Matrix Group: TANK PROFILE TANK PROFILE Function Group BAL. LEVEL GVH=844 Matrix Group : TANK PROFILE TANK PROFILE Function Group UP.WAIT TIME GVH=845 Matrix Group : TANK PROFILE 7 Commissioning

G8V4H6 LIQ. WAIT TIME

G8V4H7 OPE. WAIT TIME Item Procedure Remarks Figure. 73: G8V4H6 1. Select G8V4H6 "LIQ. WAIT TIME" to determine the displacer weight calibration interval between each measuring point in the liquid. • Set access code to 51. A small turbulence made by the movement of displacer itself may cause the liquid surrounding displacer to be unstable for the accurate measurement. 2. Enter desired waiting time by pressing + or - key, and select E to enter. • This function allows displacer to hold in each calculated position before actual measurement is proceeded based on the entered value. Item Procedure Remarks Figure. 74: G8V4H7 1. Select G8V4H7 "OPE. WAIT TIME" to determine the operation holding time limit to accept "BAL. LEVEL" before the actual operation starts. • Set access code to 51. Entire operation begins only when the condition described in "BAL. LEVEL" is achieved. 2. Enter desired waiting time by pressing + or - key, and select E to enter. • This function determines how long for the NMS to wait if "BAL. LEVEL" can not be established before quitting entire operation. TANK PROFILE Function Group LIQ. WAIT TIME GVH=846 Matrix Group : TANK PROFILE TANK PROFILE Function Group OPE. WAIT TIME GVH=847 Matrix Group : TANK PROFILE 7 Commissioning

Operation of Tank Profile (Density Profile) Measurement

After all the initial settings are completed, the actual operation can be proceeded by one simple command. The operation status regarding this function is displayed at G8V5H0 and the liquid level condition is also displayed at G8V5H1. Terms of status display and descriptions are listed below. Operations Status (G8V5H0, OPE. STATUS) Level Condition Status (G8V5H1, LEVEL CONDITION) Status Contents 0: Accepting Accepting density profile command 1: Standby Ready to execute density profile command 2: In operation Executing density profile command 3: OPR. END Density profile operation completed successfully 4: UN_BALANCE ERR Fail: Required conditions for density profile operation not satisfied 5: IPR. ERR. STOP Fail: Abnormal conditions occurred during density profile operation Status Contents 0: Off Level Meas. Active operation is not LEVEL (Profile cannot start). 1: Stable Liquid level/surface is stable (Profile can start). 2: Unstable Liquid level/surface is unstable (Profile cannot start). 3: Ignore condition Ignores liquid level/surface condition (profile cannot start). Item Procedure Remarks Figure. 75: Static Matrix 1. Select G0V2H0 "OPERATION" in the Static Matrix and select "UPPER DENSITY". • Set access code to 50. • The operation command "UPPER DENSITY" is shared between spot and Tank Profile density measurement depends on G8V4H0 configuration. 2. Press E to start the operation. • The operation must start when the NMS is in Level measuring mode. 3. Unless there is any incompatible circumstances during the displacer automatically returns to the level measurement after the last measuring point. • To minimize the risk of safe operation, the displacer immediately returns to the Level measurement regardless of completing various data process. Static Matrix GVH=020 OPERATION OPERATION LEVEL UP STOP BOTTOM LEVEL MIDDLE INTER LEVEL UPPER DENSITY MIDDLE DENSITY DENSITY BOTTOM REPEATABLITY WATER DIP 7 Commissioning

Density Profile Operation Diagram

Figure. 76: Density Profile Operation Procedure 1. Set NMS5 so that it can measures the liquid level. 2. Raises the displacer and measure each weight of 8 measuring points in the air. 3. Measure the density at the setting position in the liquid level. 4. Return to level measurement. 5. Save the data in G8 matrix accordingly. This completes the operation procedure. Caution! The average density and temperature data regarding density profile function (G8 Matrix) are completely differ- ent from the data on G0 static matrix which displays the conventional upper density (spot) and temperature (continuously scanned) data. 8 points of air weight measurement (within 300mm circumference of wire drum) Nozzle Proservo NMS5 �������� First Displacer Weight Calculation Line (Approximately 500mm from Liquid Level) No. 1 Measuring Point No. 2 Measuring Point No. 3 Measuring Point No. 4 Measuring Point No. 5 Measuring Point No. 6 Measuring Point No. 7 Measuring Point If required density measuring points are 7, the method to calculate each point from liquid level in NMS with ver 4.25 softwear is: (7 points +1) Liquid level x measuring point number Ex. if the liquid level is 4,000 mm and needs to determine the position of No. 5 measuring point from the liquid surface; (7 +1) 4,000 x 5 = 2,500 mm ������������������������������ Max.16 point ����������������������������������������� Max.16 point Liquid Level 7 Commissioning

NMS5 can measure interface levels via the following operations.

• Upper Interface Level: returns the level data for the interface between top 2 liquid phases. Displacer remains at interface level in UPPER INTERFACE operation. • Middle Interface Level: returns the level data for the interface between bottom 2 liquid phases. Displacer remains at interface level in MIDDLE INTERFACE operation. • Water Dip: returns the level data for the interface between top 2 liquid phases. Displacer returns to liquid level in LEVEL operation. The following preconditions are required in order to assure safe and accurate operation. • Density values: Figure. 77: Interface Measurement Density values can be viewed and changed at the following matrix positions: G0V0H5 Upper Density G0V0H6 Middle Density G0V0H7 Density Bottom Note! Additional matrix parameter settings are required for remote communications via WM550 communications protocol. Refer to Section "13. Matrix, (Dynamic Matrix, Device Data : G2), WM550 ALM. SELECT.". Operation: After confirming the above pre-conditions, use the "+" or "-" keys to select the desired operation at matrix posi- tion G0V2H0 OPERATION. When the desired operation is displayed, press the "E" key. e.g. 0.760 g/cm3 e.g. 0.880 g/cm3 e.g. 0.1000 g/cm3 (Example) - Upper Density < Middle Density < Bottom Density Middle Density - Upper Density 0.100 g/cm3 Bottom Density - Middle Density 0.100 g/cm3 7 Commissioning

Sealing of NMS5

NMS5 can be sealed, and data can be made unchangeable for custody transfer purposes as follows. Sealing of NMS5 Procedure 1. Set OPE. DENSITY at GVH=278 as the liquid density (if the density is expected to change during operation, set the average value.). 2. Turn off the power supply, open NMS5. 3. Turn the micro switch to the "on position" on the printed circuit board TCB (see figure below). NMS5 will then enter "write-protect" mode and not accept any further level, weight, or density data changes. This completes the sealing of NMS5 procedure. Figure. 78: The printed circuit board TCB 2 20 9 1 6 1 13 1 1 CNB CN5 CN7 CN6 CN14 SW1 U1 4 1 5 4 1 5 8 5 8 1 4 5 8 1 4 U20 U19 U15 U13 U10 14 1 15 28 1 16 8 9 1 7 8 U14 U9 U21 U7 U4 U11 U12 1 14 7 8 1 16 U24 U6 9 8 C16 C8 C7 U2 DC-DC Converter C35 U3 C38 ON OFF CN9 CN2 CN3 TCB CD_L423 Protect Switch 7 Commissioning

NMS5 measuring instrument requires no special maintenance.

Exterior Cleaning When cleaning the exterior of measuring devices, always use cleaning agents which will not cause corrosion or other damage to the housing surface seals. Seals Replacement NMS5 O-rings may need to be replaced periodically, and more frequently in extreme climates. The period between changes will depend on climate severity, ambient temperatures and process. Repairs The Endress+Hauser repair policy is based on the fact that the measuring devices have a modular design and that customers are able to undertake repairs themselves. Spare parts are contained in corresponding kits along with their related replacement instructions. Endress+Hauser provides spare parts for repairs of NMS5, which are located with their order numbers on later pages (refer to "10.6 Spare Parts"). Contact Endress+Hauser ser- vice representatives for further assistance regarding service and spare parts. Repairs to Ex-approved devices When performing repairs on Ex-approved devices, note the following: • Repairs of Ex-approved devices may only be performed by trained personnel or by Endress+Hauser Service. • Comply with the prevailing standards, national Ex-area regulations, safety instructions (XA) and other relevant rules. • Only use original spare parts provided by Endress+Hauser. • When ordering spare parts, note the device information on the nameplate. Replace parts only with parts that have the same device information. • Perform repairs according to the instructions. When completing repairs, perform the specified routine test on the device. • Only Endress+Hauser service representatives may convert a certified device into a different certified variant. • Document all repair work and conversions. Replacement After replacing NMS5 or electronic module, the adjustment of a new sensor is required. 8 Maintenance

Refer to the separate "Technical Information for NMS5".

NHC4HP: high pressure version NHC4LP: low pressure version Figure. 79: Calibration Chamber 9.2 Power and Control Switch Figure. 80: Power and Control Switch NHS8 172 300 Ø150 CALIBRATION CHAMBER Flange Flange 8 hole Power + Control Swith SPECIFICATIONS RATED VOLTAGE AC110 / 220V RATED CURRENT 6A / 5A ALUMINUM ALLY CASTING HOUSING TERMINAL BOX IRON CASTING STRUCTURE EXPLOSION PROOF d3aG5 NO.OF CERTIFICATE T3875 CONNECTION DIAGRAM Term.No Power Supply NMS Term.No NHS8 1 2 3 4 5 6 7 1 2 16 17 18 Handle position Wiring LIFT INST STOP OFF 7 6 4 3 5 2 1 OFF STOP INST LIFT Cable Entry PF (G)1/2 PF (G) 3/4 PF (G)1 h (mm) B 17 20 24 140 140 145 Cable entry 193 113 65 Grounding LIFT STOP OFF 2 - Ø7 120 B INST h Power L Power N COM CRT1 CRT2 9 Accessories

Ball Valve NHV4A

NHV4A: Standard ball valve with ANSI flanges NHV4J: Standard ball valve with JIS flanges Figure. 81: Ball Valve 9.4 Reducing Flange NHF4 Figure. 82: Reducing Flange Ball Body cap Body 190 153 400 Low pressure (JPI example) High pressure (ANSI example) Stud Bolt Flange Stud Bolt Flange 9 Accessories

Diagnosis and Troubleshooting

NMS5 has a superb self-diagnosis function which monitors its operation. If an error is occurred, the corre- sponding message is displayed on the LCD. The selected matrix position and the error message are displayed every few seconds alternatively. Data can be accessed while the selected matrix position is displayed. The error messages are stored in the memory of NMS5. Matrix position GVH=037 provides the diagnostic history. 10.1 Selection of Diagnostic Code and History Item Procedure Remarks Static Matrix : G0 Figure. 83: Static Matrix Touch Control Figure. 84: Touch Control 1. On "MORE FUNCTION" , select item GVH=036 "DIAGNOSTIC CO." 2. Previous static matrix records are sequentially displayed in Static Matrix screen GVH=037, start- ing with the latest record. Up to 100 alarm records can be saved. If the num- ber of records exceeds 100, it is sequentially overwritten starting with the oldest record. For example, 973192238 2402 means that an error occurred at 22:38 on March 19, 1997, when the instrument temperature was 24 oC and this is the second error since NMS5 was installed. Display includes the year, month, day, hour, minute, instrument temperature, and an error sequential number, in this order. • Item GVH=037 only shows error message, calendar, and pointer, but no label on the LCD. MORE FUNCTION DIAGNOSTIC CO (Erroneous Mess 0 3 6 0 3 7 10 Diagnosis and Troubleshooting

Error and Status Messages

Message Cause Remedy Error Code -49.5 ºC Indication of shorted element in NMT temperature sensor Check GVH=450 thru GVH=459 to see element. 358.0 ºC Element is not enabled, or not installed in NMT temperature sensor . Check GVH=482 Total number of elements 359.0 ºC Indication of open element in NMT temperature sensor Check GVH=450 thru GVH=459 to see element. ADC/SENSOR ERROR The signal from the A/D converter is out of range. Consult E+H Service. 107 ADJ. XXXCOUNTER (X=A, I, Z or combina- tion) Error of level data check by A, I, and/ or Z phase encoder. If the message appears frequently, then consult E+H Ser- vice. A PHASE NO INPUT The input of the A phase signal from the encoder is not available. Replace the detector unit. Consult E+H Service. 122 BELOWBOT. POINT Indicates NMS displacer position is below lowest of NMT temperature sensor. DEVICE ERROR: DEV 1 (or 2) The HART device 1 (or 2) gives an error signal. Check the HART device 1 (or 2) . 130, 131 DEVICE ERROR: NMT NMT 53x gives an error signal. Check the connection of NMT 53x. Check the register of the temeperature sensor at position GVH=362 "CONNEC- TION NMT." 250 DEVICE ERROR: NMT NMT 53x temperature sensor is not available. Check NMT 53x. Error messages are available in NMT 53x manual. 233 DEVICE ERROR: NRF NRF 560 gives an error signal. Check NRF 560. Error messages are available in NRF 560 manual. 240 DISPL CALIB ERROR The deviation of automatic weight calibration exceeds its set limit. Check build-up or deposit on the displacer. 120 ELEM 0 RANGE OVER Indicates the reference resistance ele- ment (ºC) on NMT electronics is out- of- tolerance. IMPOSSIBILITY Operation is impossible because there is no weight table. LOCAL ERROR: DEV1 (or 2) NMS5 cannot access the local HART device 1 (or 2). Check the connection of the HART device to NMS5. Check the registration of the device(s) in matrix group G5/6. 232 LOCAL ERROR: NMT The signal from NMT 53x tempera- ture sensor is not available. Check the connection of NMT 53x. Check the register of the temperature sensor at position GVH=362 "CONNEC- TION NMT." 111 LOCAL ERROR: NRF NMS5 cannot access NRF 560. Check the connection of NRF 560 and register of the tank side monitor at position GVH=361 "CONNECTION NRF." 113 MINTENANCE GVH157 Service Mode = ON Set GVH157 = OFF 199 MEM. ERROR Defect in the memory that is specially used for custody transfer sealing. Replace the CPU board. Consult E+H Service. 201 MPU RESET Power to the CPU module fell below minimum value. MPU: START ACT* Power to NMS5 was turned OFF/ ON. If the message occurs frequently, then consult E+H Service. 10 Diagnosis and Troubleshooting

These error histories are available (Erroneous Message) at position GVH=037.

OVERTENSION The tension on the measuring wire exceeds the upper limit set at position GVH=162 "OVER TENS. SET." Check if the displacer motion is blocked by clogging or sticking. To release overtensioning, access position GVH=371 "RELE. OVER TENS." 101 POWER FAILURE The supply voltage falls below the allowed value. Check the power source. 124 RAM FAILURE CPU RAM failure ROM ERROR EEPROM Consult E+H Service. 132 SIFA ERROR Error between communication board and CPU board for digital output. Replace the CPU board. Consult E+H service. 114 TEMP BELOW RANGE Indicates the measured temperature data is below the fixed range of the NMT temperature sensor. TEMP COM OPEN Indicates an open common line in the NMT temperature sensor TEMP COM SHORT Indicates a shorted common line in the NMT temperature sensor TEMP OVER RANGE Indicates the measured temperature data is highter than the fixed range of the NMT temperature sensor. UNDERTENSION The tension on the measuring wire falls below the lower limit set at position GVH=163 "UNDER TENS. SET." Check if the measuring wire is cut or the displacer is lost. In this case, check the installation of the NMS5. 102 WIRE CALIB ERROR The deviation of automatic wire calibration exceeds its set limit. Check wire and wire drum. 115 Z PHASE NO INPUT The input of the Z phase signal from the encoder is not available. Replace the detector unit. Consult E+H Service. 106, 112 MPU:XXXX* (XXXX=text) CPU error The error might happen occasionally and might be regis- tered in GVH=037 (Erroneous Message). However, it is normally negligible. If it occurs frequently, then consult E+H Service. OPE.CODE ERROR An illegal operation command is accessed. If the message occurs frequently, then consult E+H Ser- vice. LCD CHECK Error between LCD (touch control) unit and CPU boaed for digital output. Replace the touch control. 121 GAUGE TEMP. The temperature inside the gauge exceeds the limit. Check if the ambient temperature stays within the limit. If the application is a high temperature tank, then take meadures to avoid heat tranfer from the tank to NMS5. DRIVER ERROR A motor driver error has occurred Consult E+H Service Message Cause Remedy Error Code 10 Diagnosis and Troubleshooting

Flowcharts for Diagnostics and Troubleshooting

For up to 10 elements, check G4V5H0 thru H9 to �nd shorted element. For more than 10 elements, select element at G4V7H0, view temperature at G4V7H3. (Level display) -49.5 C Wire ring at normal position ? May require complete NMT exchange. Send all Temperature matrix (G4) data to Endress+Hauser 359.0 C TEMP. NO. n ADJ(A, I, Z) COUNTER YYMMDDhhmm TT ## Raise displacer to red line in calibration window, or into maintenance chamber. G3V5H0 = G3V5H1 = 0? 3V5H0 = G3V5H1 = 0? Displacer stopped at Upper Stop, G1V6H0? Meaning: Shorted element in NMT temperature Meaning: Element not enabled, or not installed in NMT temperature sensor This is an indicator, not an error code. Verify total number of elements at G4V8H2. Diagnostic : -49.5 C Diagnostic : 358.0 C (at G4V5_or G4V7H3) Diagnostic : 359.5 C (Level display) 359.5C Meaning: Open element in NMT temperature sensor For or up to 10 elements, check G4V5H0 thru H9 to �nd open element. For more than 10 elements, select element at G4V7H0, view temperature at G4V7H3. May require complete NMT exchange. Send all Temperature matrix (G4) data to Endress+Hauser Diagnostic : ADC SENSOR ERROR ADC SENSOR ERROR YYMMDDhhmm TT ## Meaning: The A/D signal output from the detector unit, or input to CPU, is abnormal. Exchange detector unit, adjust Hall sensors, recalibrate. Make connections �rm. TCB CN5, CN7 loose? Diagnostic : ADJ A, I or Z COUNTER Automatic software adjustment working normally. Sometimes seen in turbulent applications. Noise interface to internal software counters. Eliminate noise, adjust Hall sensors, recalibrate. Raise displacer manually (G3V7H6) until red line in window. Diagnostic : A PHASE NO INPUT A PHASE NO INPUT YYMMDDhhmm TT ## Meaning: The A-phase signal output from the encoder, or input to CPU. is abnormal. 20 outputs = 360 degrees. Exchange detector unit, adjust Hall sensors, recalibrate. Make connections �rm. TCB CN5, CN7 loose? OK? Meaning: Automatic software adjustment working normally. Sometimes seen in turbulent applications. Yes No Yes No Yes No Yes No No Yes No Yes 10 Diagnosis and Troubleshooting

Adjust G4V8H6 from

0 -500 mm. This will create an offset for temperature elements. (FLASHING ON ANY DISPLAY) Meaning: Shorted element in NMT temperature Meaning: An error has occurred at the HART device. Refer to HART device Operating Manual Diagnostic : BELOW BOT. POINT Diagnostic : DEVICE ERROR: DEV1 (2) Diagnostic : DEVICE ERROR : NRF Meaning: An error has occurred at the NRF tank side indicator. Refer to NRF Operating Manual. Diagnostic : DISPL CALIB ERROR Increase G1V8H5, or adjust Hall sensors and recalibrate. Diagnostic : ELEM 0 RANGE OVER Meaning: The A-phase signal output from the encoder, or input to CPU. is abnormal. 20 outputs = 360 degrees. Record Temperature matrix data, exchange temperature electronics. Loose connector J3 behind NMS terminal block. Check connector. G4V4H7 = 0.0 C ? Meaning: Measured weight of the displacer exceeds allowable limit: G1V8H4 >=G1V8H5 SW v 4.20 or Later ? Set G4V9H8 = ON Note!: If displacer position gose lower than (Elem 0 position + G4V8H6), BELOW BOT. POINT will display. (FLASHING ON ANY DISPLAY) Meaning: An error has occurred at the NMT temperature sensor. Check G4V8H0 for NMT Diagnostics codes, refer to NMT Operating Manual. Diagnostic : DEVICE ERROR: NMT (FLASHING ON ANY DISPLAY) (FLASHING ON ANY DISPLAY) (FLASHING ON ANY DISPLAY) Buildup on displacer ? Clean displacer, wire, wire drum, adjust Hall sensors, recalibrate (FLASHING ON ANY DISPLAY) G4V4H7 <+/-9.9 C ? At G4V7H1 select element 0, then at G4V4H7. Repeat for all elements. Diagnostic : IMPOSSIBILITY Meaning: Operation impossible because a valid weight table does not exist. Adjust Hall sensors and make weight table. IMPOSSIBILITY OPERATION Yes No Yes No Yes Yes No No 10 Diagnosis and Troubleshooting

Check HART bus wiring:

DEV to NMS cut or short ? DEV (+)(-) to NMS (+)(-) ? Voltage = DC16.5 - 25.5V ? (FLASHING ON ANY DISPLAY) Meaning: An error has occurred at the HART device. Diagnostic : LOCAL ERROR DEV1(2) Diagnostic : LOCAL ERROR NRF Check HART bus wiring: NRF to NMS cut or short ? NMS (+)(-) to NRF (+)(-) ? NMT to NMS cut or short ? Voltage = DC16.5 - 25.5V ? Meaning: An error has occurred at the NRF tank side indicator. G5V4H9 = ON ? Set G5(6)V4H9 = ON Check HART device address with hand-held. HART DEV1 = 4 (�xed) HART DEV2 = 5 (�xed) Meaning: An error has occurred at the NMT temperature sensor. Check G4V8H0 for NMT Diagnostics codes, refer to NMT Operating Manual. Diagnostic : LOCAL Error NMT (FLASHING ON ANY DISPLAY) (FLASHING ON ANY DISPLAY) G3V6H1 correct ? Make correct setting at G3V6H1. Diagnostic : MEM ERROR Exchange TCB module, adjust Hall sensors, recalibrate. MEM. ERROR YYMMDDhhmm TT ## Yes No Yes No G5V4H9 = ON ? Refer to HART device operating manual. No Check G3V6H1: NRF >=v 1.82, CONTACT2 NRF < v 1.82, CONTACT1 OK ? OK ? Exchange FCB module in NRF. Exchange POW module in NRF No No OK ? Exchange electronics module in NMT. Check HART bus wiring: NRF to NMS cut or short ? NMS (+)(-) to NMT (+)(-) ? NMT to NMS cut or short ? Voltage = DC16.5 - 25.5V ? No Meaning: Power to the CPU module fell below minimum value. Diagnostic : MPU : RESET OK ? Noise, Isolate and eliminate source of noise. No MEM. ERROR YYMMDDhhmm TT ## Frequent ? Exchange TCB module, adjust Hall sensors, recalibrate. No 10 Diagnosis and Troubleshooting

Check local power supply

if frequent. MPU: START ACT YYMMDDhhmm TT ## Meaning: Power supply has been turned (off and) on. Indicates normal operation. Diagnostic : MPU: START ACT Diagnostic : ROM ERROR Meaning: CPU ROM failure. Meaning: Tension on measuring wire is more than value in G1V6H3. Set Operation = STOP Set G3V7H1 = ON Diagnostic : OVERTENSION (FLASHING ON ANY DISPLAY) Diagnostic : TEMP. BELOW RAMGE Check NMT speci�cation. Check G4V5H0 thru H9. Check temperature in tank. Exchange TCB module, adjust Hall sensor, recalibrate Check measuring wire, displacer, stilling pipe for cause of overtension. No Meaning: Open element in NMT temperature sensor. Diagnostic : TEMP. COM. OPEN OK ? Enter correct number of temperature elements at G4V8H2. G4V8H2 data correct ? Record Temperature matrix data, exchange temperature electronics. No ROM ERROR YYMMDDhhmm TT ## Diagnostic : SIFA ERROR Meaning: CPU SIFA failure. Exchange TCB module, adjust Hall sensor, recalibrate SIFA ERROR YYMMDDhhmm TT ## Measuring wire cut ? Find and �x cause of cut wire. Replace displacer, adjust Hall sensor, recalibrate. Yes Diagnostic : POWER FAILURE POWER FAILURE YYMMDDhhmm TT ## Check local power supply. Meaning: Supply voltage has fallen below about AC95V. Diagnostic : RAM FAILURE Exchange TCB module, adjust Hall sensors, recalibrate. RAM FAILURE YYMMDDhhmm TT ## Meaning: CPU RAM failure. Data below range ? Record Temperature matrix data, exchange temperature electronics. Normal operation. Error will clear when actual temperature is within speci�cation. Meaning: Measured temperature data for one or more elements is below �xed range of NMT speci�cation. (FLASHING ON ANY DISPLAY) (FLASHING ON ANY DISPLAY) May require complete NMT exchange. Send all Temperature matrix (G4) data to Endress+Hauser. Yes No Yes No 10 Diagnosis and Troubleshooting

Enter correct number of

temperature elements at G4V8H2. Meaning: Shorted element in NMT temperature sensor. Diagnostic : TEMP. COM. SHORT Meaning: Measured temperature data for or more elements is over �xed range of NMT speci�cation. Check NMT speci�cation. Check G4V5H0 thru H9. Check temperature in tank. Diagnostic : TEMP. OVER RANGE (FLASHING ON ANY DISPLAY) Diagnostic : UNDERTENTION Set Operation = STOP Set G3V7H1 = ON Record Temperature matrix data, exchange temperature electronics. No Meaning: Measured length of the measuring wire exceeds allowable limit: G1V7H4 >= G1V7H5 Diagnostic : WIRE CALIB ERROR Increase G1V8H5, or adjust Hall sensors and recalibrate. Buildup on measuring wire ? Clean displacer, wire, wire drum, adjust Hall sensors, recalibrate. No Data over range ? Normal operation. Error will clear when actual temperature is within speci�cation. Yes Measuring wire cut ? Check measuring wire, displacer, stilling pipe for cause of undertension. Find and �x cause of cut wire. Replace displacer, adjust Hall sensors, recalibrate. Meaning: Tension on measuring wire is more less value in G1V6H3. (FLASHING ON ANY DISPLAY) (FLASHING ON ANY DISPLAY) Yes No Yes (FLASHING ON ANY DISPLAY) G4V8H2 data correct ? Record Temperature matrix data, exchange temperature electronics. OK ? May require complete NMT exchange. Send all Temperature matrix (G4) data to endress+Hauser. No No Yes Meaning: The Z-phase signal output from the encoder, or input to CPU, is abnormal. 1 output = 360 degrees. Diagnostic : Z PHASE NO INPUT Exchange detector unit, adjust Hall sensors, recalibrate. TCB CN5, CN7 loose ? Make connections �rm. No Z PHASE NO INPUT YYMMDDhhmm TT ## Yes OK ? No 10 Diagnosis and Troubleshooting

After calibration G3V7H0 Measured Weight = 0.0 minus value.

Diagnostic : Calibration unsuccessful Diagnostic : No response to OPERATION 0.0g MEASURED WEIGHT G3V4H0, H1, H2 correct ? Input correct values for circumference, wire and displacer weight, sdjust Hall sensors, recalibrate. Adjust Hall sensors, recalibrate. If G7V4H0 thru H4(any) = 0, and does not respond to adjustment, exchange detector unit. No No Yes Hall sensor adjustment OK ? G3V9H7 a = b = 65000 ? Calibration OK ? Exchange TCB module, adjust Hall sensors and recalibrate. Forgot to pick up displacer at step 3 of Weight Table Recalibration. Do again. Exchange detector unit, adjust Hall sensors and recalibrate. Yes Yes No No LEVEL OPERATION G0V2H4 = LEVEL ? At NRF tank side indicator set OPERATION = LEVEL No No Yes G0V2H5 = LEVEL ? EPROM OK in TCB module Operation OK ? Exchange TCB module, adjust Hall sensors and recalibrate. Inspect EPROM U2 in TCB module. Straighten any bent legs. EPROM should be inserted �rm and level in socket. Exchange detector unit, adjust Hall sensors and recalibrate. Yes No G0V2H0 = LEVEL, G0V2H1 0 = (not LEVEL) (not LEVEL) OPERATION STATUS G3V7H0 G3V4H2 (+/- 2g)? Adjust Hall sensors and recalibrate. At host CPU and /or interface set OPERATION = LEVEL No No Yes G0V2H0 G0V2H0 10 Diagnosis and Troubleshooting

Setting after Parts Replacement

After any part of NMS5 is replaced, and before starting NMS5 calibration, specify the data on the following equipment. • Circumferential length of wire drum (indicated on face of wire drum) • Weight of displacer (indicated on the body of the displacer) • Volume of displacer (indicated on the body of the displacer) • Balance volume (indicated on the body of the displacer or 50% of volume) • Density of measured liquid (up to three phases) • Height of tank with NMS5 mounted on it (refer to Section "7 Commissioning"). Item Procedure Remarks Matrix Group: SERVICE 1. In Static Matrix "MORE FUNCTION" bring up GVH=030 "MATRIX OF" and select "SER- VICE. 2. Select Dynamic Matrix GVH=340 "WIRE DRUM CIRC." Check whether the displayed value is equal to the value marked on the wire drum. If not, adjust the displayed value. • Set the access code at 51. Figure. 85: Dynamic Matrix Note! The weight and volume of the displacer are marked on its bottom. Figure. 86: Displacer • The balance volume is the volume of that part of the displacer that is immersed in the liquid when the displacer is balanced in the liquid. 1. Select the Dynamic Matrix screen GVH=341. 2. Set Dynamic Matrix GVH=342 "DIS- PLACER WEIGHT" at the value marked on the displacer. 3. Set Dynamic Matrix GVH=343 "DIS- PLACER VOLUME" to the value marked on the displacer. 4. Set Dynamic Matrix GVH=344 "BALANCE VOLUME" at half the value set in "DIS- PLACER VALUME." "This setting is provided to the approximate posi- tion of the displacer when it becomes stationary in the liquid. • For calculation methods of the draft position. MORE FUNCTION DISPLACER WEIGHT 3 4 2 MATRIX OF 0 3 0 DISPLACER VOLUME BALANCE VOLUME 3 4 3 3 4 4 MEAS. WIRE & DRUM 10 Diagnosis and Troubleshooting

Maintenance Prediction Function

The history of maintenance is displayed at matrix screen GVH=265 "(Parts Overused Date)." LCD screen dis- plays the following contents. • Total operation time for parts managed according to this value • Total number of rotations of the wire drum for parts managed according to this value Reference Part Management Value Display Part Reference (conducting time/ number of rotations) 1) POWER UNIT 2) DISPLAY UNIT 3) MOTOR UNIT 4) WIRE UNIT 5) BEARINGS UNIT 6) SHAFT UNIT Power supply unit LCD indicator Motor/driver unit Measuring wire Drum bearing metal Drum shaft 43,800 hours (about 5 years) 61,300 hours (about 7 years) 43,800 hours (about 5 years) 240,000 rotations 145,000 rotations 240,000 rotations 10 Diagnosis and Troubleshooting

Spare parts are contained in kits. Spare parts for NMS5 which can be ordered from Endress+Hauser are shown

with their order numbers in the diagram below. Contact Endress+Hauser service representatives for further assistance. 10.6.1 NMS5 Figure. 87: Spare parts T-AMP Zone 0 NMS5-2/5-5/5-6 NMS5-2/5-5/5-6 NMS5-1/5-4 Accessories 11 �� 11 �� �� 11 21 21 �� 21 21 �� �� 31 32 33 34 35 36 37 37 37 �� 42 43 44 51 52 52 52 52 52 53 53 53 �� cable between terminal and module �� NMS5 ����� ���Wire Hook Welding Part Bottom of Tank 10 Diagnosis and Troubleshooting

Order Code List

Aluminum Drum Housing Drum Housing Calibration Window Order Code Parts Name Order Code Parts Name 11 Oring 70106039 Upgrade kit HART active output 017803-0034 O-ring, cover drum housing Alu, NBR 56004534 Communication module WM550, arrestor 70106019 O-ring, cover drum housing Alu, CR 52013662 Upgrade kit WM550 70106020 O-ring, cover drum housing Alu, FKM 56004409 Communication module Mark Space 70106021 O-ring, cover drum housing Alu, Silicon 52013658 Upgrade kit Mark Space 56004345 O-ring, cover drum housing Alu, PTFE 56004355 Communication module ENRAF BPM, COM-3 017803-0066 O-ring, cover drum housing, SS, FKM 52013649 Upgrade kit Enraf BPM 017803-0035 O-ring, cover drum housing, SS, silicone 34 Electronic Module, Contact Output 70109105 O-ring, cover drum housing, SS, NBR 017800-0007 Alarm output, 2x contact, I/O 3 70109106 O-ring, cover drum housing, D158-d3.5 CR 70109108 Alarm output, 2x contact, I/O 3, TUV OSP 70109107 O-ring, cover drum housing, D158-d3.5 PFA 52013654 Upgrade kit I/O-3 017803-0033 O-ring, calibration window, NBR 35 Electronic Module, A/O 70106018 O-ring, calibration window, CR 017800-0017 Output 4-20mA, I/O 5, TCB4 017803-0041 O-ring, calibration window, silicone 52013656 Upgrade kit I/O-5 017803-0036 O-ring, calibration window, FKM 36 Electronic Module, Ex [ia] HART Communication 56004518 O-ring, calibration window, PTFE 56004356 Communication module EXI HART, COM-4 017803-0030 O-ring, display cover, NBR 37 Electronic Module, I/O Spot Temperature Input 017803-0040 O-ring, terminal cover, NBR 56004442 Oper. I./C.O., spot temperature 71070894 PTFE packin+ring, calibr. window, Zone0 70106035 Upgrade kit oper. I./C.O., Ex d temp. 21 Cover 70106032 Upgrade kit oper. input/contact out 017800-0111 Cover, display module, Aluminium 56004441 Operation input/contact out, carrier 017800-0112 Cover, terminal box, Aluminum, NMS 56004491 Spot temp. with Chassis 017800-0113 Cover, drum housing, Alu, NMS531/NMS534 56004490 Spot temp. input T-AMP UP grade KIT 56004307 Cover, drum housing, SS, NMS535/536 70106027 Spot temp. Ex ia, carrier 56004308 Cover, drum housing, SS, NMS532 40 Fuse 56004295 Calibration window, Alu, NMS531/NMS534 70106027 Fuse 250VAC T2A50, NMS, 10 pcs. 56004296 Calibration window, SS, NMS532/NMS535 70106499 Fuse 20-62VDC T3A15, NMS, 10 pcs. 56004297 Calibration window, SS, NMS536 70106500 Fuse cover 10set, NMS 71070909 Calibration window, SS, Zone 0, NMS532/535/536 41 Connector, Cable 31 Electronic Module, Power Module 52013660 Wiring harness,Ter.-COM-1/RS485/BPM mod. 70103940 POW-6, HV, non Ex i 52013650 Wiring harness, Term.-HART passive mod. 70103942 POW-6, HV, Ex i 70106037 Wiring harness, Term.-HART active mod. 70103941 POW-6, LV, non Ex i 52013657 Wiring harness, Terminal-Mark Space mod. 70103943 POW-6, LV, Ex i 52013648 Wiring harness, Terminal-ENRAF BPM mod. 32 Electronic Module, Main CPU 52013653 Wiring harness, Terminal-I/O-3 module 70103937 TCB-6, CPU module, upgrade of TCB-4 52013652 Wiring harness I/O-3-TCB 33 Electronic Module, Remote Communication 52013655 Wiring harness, Terminal-I/O-5 module. 70103937 COM-1, Communication module, V1 70106036 Wiring harness, Terminal-COM-4 module 52013661 Upgrade kit V1 70106033 CPU, T-AMP Ex d module cable 71086763 Modbus module RS485, COM-5 70106034 Wiring harness,Terminal-T-AMP Ex d mod. 70106270 Upgrade kit Modbus RS485 70105996 Wiring harness,Terminal-T-AMP Ex ia mod. 56004484 COM-6 HART, passive remote commu. module 70105995 CPU, T-AMP Ex ia module cable 52013651 42 Terminal 56004485 56004508 �������������������������� 56004509 ������������������������������� 10 Diagnosis and Troubleshooting

PCB CPU Carrier set complete, spacer

56004257 5x measuring wire, 28m, d=0.15mm, 316L 017800-0021 Display module 2line, 3key optical 5x measuring wire, 16m, d=0.2mm, Alloy C 70103938 Display module 4line, 3key optical 56004914 Measuring wire, 0.2mm, AlloyC, 16m 56004256 5x measuring wire, 16m, d=0.4mm, PTFE 017800-0043 Detector unit NMS, TCC-3 56004912 Measuring wire, 0.4mm, PTFE coated, 16m 70103944 Detector unit NMS, TCC-3, V1 017800-0241 Wire ring, 316 017800-0242 Wire ring, PTFE coated 56004523 Wire drum + wire Mes., 28m 017800-0243 017800-0252Wire ring, AlloyC 56004524 ������������������������������� 017800-0252 Displacer, 50mm, cylindrical, SUS316 56004520 Wire drum, wire Alloy, 16m 56004339 Displacer, 50mm, cylindrical, guide ring 71114404 Wire drum, 28m wire, CR 017800-0253 Displacer, 50mm, conical, PTFE solid 71114405 Wire drum, 28m wire, silicone 70105994 Displacer, 50mm, cylindrical, PTFE solid 71114406 ������������������������������ 017800-0271 Displacer, 50mm, cylindrical, AlloyC 71114407 ������������������������������������ Displacer, 70mm, conical, SUS316, NMi 71114408 Wire drum, 16m AlloyC wire, CR 017800-0267 Displacer, 110mm, conical, SUS316, low 71114409 Wire drum, 16m AlloyC wire, silicone 70104684 Displacer, 110mm, conical, 316, high 71133077 Wire drum + 36m wire, SUS316L 71083820 Displacer, 50mm, cylindrical, 316, Zone0 71133079 Wire drum + 36m wire SUS316L, CR 71134069 Displacer, 30mm, cylindrical, 316, Zone0 71133078 Wire drum + 36m wire SUS316L, Silicone 71134070 Displacer, 40mm, cylindrical, 316, Zone0 71134099 Wire drum + 47m wire. 71134071 Displacer, ring, 50mm, cylind.,316, Zone0 71134100 Wire drum + 47m wire, CR. 71134067 Displacer, ring, 70mm, cylind.,316, Zone0 71134101 Wire drum + 47m wire, Silicone Displacer, ring, 110mm, cylind.,316, Zone0 56004525 Wire drum bracket, Alu Displacer, 50mm, cylind., AlloyC, Zone0 56004527 Wire drum bracket, SS 56004526 ���������������������������������� 71067825 Calibration weight 50gr, NMS5 017800-0221 Bearing, wire drum bracket, PTFE 71134095 Wire hook, SUS, L=100 Wire hook, SS400, L=100 71114403 Measuring wire 0.15mm 28m SUS316L, ring SUS316 44 Display Module 51 Detector, Servo Unit 52 Wire Drum 53 Measuring Wire, Displacer 56004255 017800-0263 71134068 71134072 75 Accessory 71134096 10 Diagnosis and Troubleshooting

The following procedure must be performed before returning NMS5 transmitter to Endress+Hauser e.g. for

repair or calibration. • Remove all residue. Pay special attention to the gasket grooves and crevices where fluid may be present. This is especially important if the fluid is corrosive, poisonous, carcinogenic, radioactive, or otherwise hazardous. • Always enclose a duly completed "Declaration of Hazardous Material and De-contamination" form (a copy of the “Declaration of Hazardous Material and De-contamination” is included at the end of this operating manual). Only then can Endress+Hauser transport, examine, and repair a returned device. • Enclose special handling instructions if necessary, for example a safety data sheet as per EN 91/155/EEC. Additionally specify: • An exact description of the application • The chemical and physical characteristics of the instrument • A short description of the error that occurred (specify the error code where possible) • Operating time of the device Caution! • Hazardous materials may be attached to damaged parts of NMS5 or its plastic material. Unless hazardous materials are completely removed from NMS5, no repair request is accepted. • Incomplete cleaning of the instrument may result in waste disposal or cause harm to personnel (burns, etc.). Any costs arising from this will be charged to the operator of the instrument. 10.8 Disposal In case of disposal, separate the various components according to their materials. 10 Diagnosis and Troubleshooting

10.9 Gauge Software

10.10 Contact Addresses of Endress+Hauser The addresses of Endress+Hauser are given on the back cover of this operating manual. If you have any questions, do not hesitate to contact E+H representatives. Software version / Date Software changes Documentation changes V2.13, from 9.96 Original Release BA001N/08/en/09.98 V2.20, from 10.97 T2:TCB-2, Modified V1: R&S command BA001N/08/en/11.97 V4.06, from 04.98 HART master, T4:TCB-4 V4.06, from 04.98 HART master T&W, T4:TCB-4 V4.08X, from 08.98 upper stop speed reduction 3-step V4.20, from 08.98 WM550, Commuwin II display, elem. Error fix BA001N/08/en/12.99 V4.20, from 09.98 T&W BA001N/08/en/12.99 V4.22, from 05.00 HART line selection add to static matrix: gas temp. minus data remote communication level 6 digit data pro- cessing memory clear volume calculation matrix deleted V4.23B, from 09.00 standard release, alarm output modify, NMT level selection V4.24, from 04.01 standard release BA001N/08/en/03.01 V4.24 OSP, from 06.01 Over Spill protection V4.24T&W, from 07.01 T&W spec. V4.24 T2, from 02.02 TCB-2 CPU, NMT, V1 level measurement V4.25, from 09.02 Density profile function BA001N/08/en/02.02 V4.27, from 09.04 Level hold matrix, Error display on Home screen only BA001N/08/en/11.04 V4.27 T6/T&O, from11.05 W&M and Overspill prevention BA001N/08/en/03.06 V4.27 A, from 05.07 standard release V4.27 B, from 09.08 standard release BA1001N/08/en/12.08 V4.27 C, from 04.09 Proactive Safety function BA1001N/08/en/04.09 V4.27 E, from 03.10 DTM support BA00401G/08/en/01.11 V4.27 F, from 05.11 SIL BA00401G/08/en/02.11 V4.27 G, from 05.12 4-20mA long range support BA00401G/08/en/03.13 10 Diagnosis and Troubleshooting

Displacer and Measuring Wire

11.1 Shape, Diameter, and Material 11.1.1 Displacer There are several types of displacer available for NMS5. • The standard type has cylindrical shape and a diameter of 50 mm. Diameters from 30 to 50 mm are optional. • Cylindrical shape is used for viscous liquids. It is also effective if the stilling well is not smooth on its interior surface. Displacer weight and volume depend on the application. Thin displacers are suited for level measurement, thicker ones for bottom level, interface level, and density measurement. A counterweight is optional for heavy turbulence (inquire). Displacers of three different materials are provided. • The standard material is stainless steel SUS316. • Alloy C and PFA are optional for corrosive liquids. • Solid PFA, however, is not applied for flammable liquids. The following size of displacer will be supplied, when you order custody transfer approval. Note! NMi (Netherlands).................70mm PTB (Germany).......................110mm 11.1.2 Measuring Wire • The standard material of the measuring wire is stainless steel SUS316L. • Alloy C and PFA coated stainless steel SUS316 are for corrosive liquids. The following specification of measuring will be supplied when ordering the custody transfer approval. Note! • Select SUS316L (0.15mm) for NMi and PTB specifications. • Refer to the separate "Techenical Information TI00452G" for displacer selection information and Appendix "14.3 Displacer" in this chapter. 11 Displacer and Measuring Wire

Alarm History Display

Item Procedure Remarks Matrix Group: DEVICE DATA Figure. 88: Device Data Touch Control Figure. 89: LCD (Display) 1. Bring up Static Matrix GVH=034 *ALARM CONTACT." Static Matrix Figure. 90: Static Matrix 2. Previous static matrix records are sequentially displayed in Static Matrix screen GVH=035, starting with the latest record. Up to 100 alarm records can be saved. If the number of records exceeds 100, it is sequentially overwritten starting with the oldest record. The display includes the year, month, day, hour, min- ute, instrument temperature, and an error sequential number, in this order. For example, 97 3192238 2402 means that an error occurred at 22:38 on March 19, 1997, when the instrument temperature was 24 oC and this is the second error since NMS5 was installed. MORE FUNCTION ALARM CONTACT 0 3 4 MORE FUNCTION (Alarm Message) 0 3 5 11 Displacer and Measuring Wire

List of Alarm Messages

11.4 Hall Sensors After exchanging CPU module, detector unit, wire drum, measuring wire (all or part), it is necessary to adjust the hall sensors and make new weight table calibration. 1. Move the displacer to the side of calibration window or the maintenance chamber window (top of the tank) or put the displacer on a work bench so that the displacer can be touched. Figure. 91: Hall Sensor Adjustment 1 2. Confirm that NMS5 is level and no vibration or other disturbance occurred. • Access Code 530 • Set Operation = STOP • Set GVH030 = Adj. Sensor 3. Set A Span and Record Sensor Count (Wa2) at GVH74. Figure. 92: Hall Sensor 2 MESSAGE Cause of Alarm UPPER LIMIT LEVEL The level has risen above the set alarm operation value. LOWER LIMIT LEVEL The level has fallen below the set alarm operation value. UPPER LIMIT TEMP. The temperature has risen above the set alarm operation value. LOWER LIMIT TEMP. The temperature has fallen below the set alarm operation value. 530 ACCESS CODE STOP OPERATION ADJ. SENSOR MATRIX OF 29XX 25000 ADJ. A SPAN index count sensor count (Ex. Wa2= 25000) 29XX 15000 ADJ. A SPAN (Ex. Wa1= 15000) 11 Displacer and Measuring Wire

Do not let wire fall off the wire drum.

4. Lift the displacer until Sensor Count is stable and record the value of Sensor Count (Wa1). • |Wa2 - Wa1 | = 10000 (+/-100)? 5. If Yes, go to step 7 "GVH740 A Zero". 6. If No, return the displacer to the original position at GVH741 • If |Wa2 - Wa1 |<10,000, increase Index Count. • If |Wa2 - Wa1 |>10,000, decrease Index Count. • Repeat until |Wa2 - Wa1 | = 10000 (+/-100). 7. Adjust A Zero Sensor Count = 21000 (+/-100) at GVH740. Figure. 93: Hall Sensor Adjustment 3 8. Record the value of Sensor Counter (Wa2) 9. Lift the displacer until Sensor Count is stable and record the value of Sensor Count (Wa2). • |Wa1 - Wa2 | = 10000 (+/-100)? 10. If yes, go to step 12. "GVH743 B Span". 11. If no, repeat Steps 3and 7. 12. Set B Span at GVH743 and repeat steps above. Figure. 94: Hall Sensor Adjustment 4 13. Adjust B Zero Sensor Count = 11000 (+/- 100) at GVH742. Figure. 95: Hall Sensor Adjustment 5 This completes the hall sensor adjustment. 29XX 21000 ADJ. A ZERO (Ex Wa2=2100) (Ex Wa1=1100) 29XX 11000 ADJ. A ZERO 31XX 11000 ADJ. B SPAN 31XX 21000 ADJ. B SPAN (Ex Wa1=1100) 31XX 21000 ADJ. B SPAN 11 Displacer and Measuring Wire