Endress+Hauser Proservo NMS5 Operating Instructions Manual
Text of the manual
Table of Contents
1
Safety Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.1
Designated Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.2
Installation, Commissioning, and Operation . . . . . . . . . . . 4
1.3
Product Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.4
Operational Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.5
Electrostatic Charge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
1.6
Notes on Safety Conventions and Symbols . . . . . . . . . . . . 7
2
Identification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.1
Device Designation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.2
Ordering Structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
2.3
Scope of Delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
2.4
Certificates and Approvals . . . . . . . . . . . . . . . . . . . . . . . 16
2.5
Registered Trademarks . . . . . . . . . . . . . . . . . . . . . . . . . . 16
3
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
3.1
Incoming Acceptance, Transport, and Storage . . . . . . . . . 17
3.2
Terms Related to Tank Measurements . . . . . . . . . . . . . . 18
3.3
Design and Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . 19
3.4
Necessary Tools for Installation . . . . . . . . . . . . . . . . . . . . 20
4
Mounting of NMS5 . . . . . . . . . . . . . . . . . . . . . . . . . . 21
4.1
Application Drawing for Tank . . . . . . . . . . . . . . . . . . . . . 21
4.2
Mounting without Guide System . . . . . . . . . . . . . . . . . . 22
4.3
Mounting with Stilling Well . . . . . . . . . . . . . . . . . . . . . . 22
4.4
Mounting NMS5 with Guide Wire . . . . . . . . . . . . . . . . . 24
4.5
Preparation for Mounting of NMS5 . . . . . . . . . . . . . . . . . 28
4.6
Electrostatic Charge . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
4.7
Installation for Wire Drum and Displacer . . . . . . . . . . . . 29
5
Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
5.1
Wiring Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
5.2
Terminal Assignment . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
5.3
Temperature Input System . . . . . . . . . . . . . . . . . . . . . . . 42
6
Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
6.1
Touch Control and Programming Matrix . . . . . . . . . . . . 44
6.2
HOME Position . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
6.3
Access Code . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
6.4
Operation Command and New Operation Status . . . . . . . 50
6.5
Operation of Displacer . . . . . . . . . . . . . . . . . . . . . . . . . . 52
6.6
Calculation of Level and Densities . . . . . . . . . . . . . . . . . 53
7
Commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
7.1
Initial Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
7.2
Settings for NMT 53x Connections . . . . . . . . . . . . . . . . . 58
7.3
Settings for Connection to NRF560 . . . . . . . . . . . . . . . . . 59
7.4
Liquid Level Calibration . . . . . . . . . . . . . . . . . . . . . . . . . 60
7.5
Remote Communications . . . . . . . . . . . . . . . . . . . . . . . . 63
7.6
Density and Density Profile Measurements . . . . . . . . . . . 70
7.7
Interface Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . 77
7.8
Sealing of NMS5 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
8
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
9
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
9.1
Calibration Chamber . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
9.2
Power and Control Switch . . . . . . . . . . . . . . . . . . . . . . . 80
9.3
Ball Valve NHV4A . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
9.4
Reducing Flange NHF4 . . . . . . . . . . . . . . . . . . . . . . . . . . 81
10
Diagnosis and Troubleshooting . . . . . . . . . . . . . . . . 82
10.1
Selection of Diagnostic Code and History . . . . . . . . . . . . . 82
10.2
Error and Status Messages . . . . . . . . . . . . . . . . . . . . . . . . 83
10.3
Flowcharts for Diagnostics and Troubleshooting . . . . . . . 85
10.4
Setting after Parts Replacement . . . . . . . . . . . . . . . . . . . . 91
10.5
Intelligent Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93
10.6
Spare Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94
10.7
Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
10.8
Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
10.9
Gauge Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98
10.10
Contact Addresses of Endress+Hauser . . . . . . . . . . . . . . . 98
11
Displacer and Measuring Wire. . . . . . . . . . . . . . . . . 99
11.1
Shape, Diameter, and Material . . . . . . . . . . . . . . . . . . . . 99
11.2
Alarm History Display . . . . . . . . . . . . . . . . . . . . . . . . . . 100
11.3
List of Alarm Messages . . . . . . . . . . . . . . . . . . . . . . . . . 101
11.4
Hall Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101
11.5
Initial Weight Calibration . . . . . . . . . . . . . . . . . . . . . . . 103
12
Technical Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107
13
Matrix . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 108
13.1
Programming Matrix . . . . . . . . . . . . . . . . . . . . . . . . . . . 108
13.2
Description of Programming Matrix . . . . . . . . . . . . . . . . 118
14
Appendix. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 154
14.1
RS 485 MODBUS (COM- 5) Terminator . . . . . . . . . . . . 154
14.2
Measurement Wire Replacement . . . . . . . . . . . . . . . . . . 156
14.3
Displacer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 158
Table of Contents
Proservo NMS5 of intelligent tank gauge is designed for high-accuracy liquid level measurement in storage and
process applications.
NMS5 performs the tank inventory management, loss control, total cost saving, and safe operation.
NMS5 is designed for the purpose of single or multi-task installations, converting a wide range of measurement
functions.
1.2
Installation, Commissioning, and Operation
•
Mounting, electrical installation, start-up, and maintenance of the instrument may only be performed by
trained personnel authorized by the operator of the facility.
•
Personnel must read and understand these installation instructions before performing the procedures.
•
The instrument may only be operated by personnel who are authorized and trained by the operator of the
facility. All instructions in this manual must be observed.
•
The installer must make sure that the measuring system is correctly wired according to the wiring diagrams.
The measuring system must be grounded.
•
Observe all law and regulations applicable and valid for your country and pertaining to the opening and
repairing of electrical devices.
1.3
Product Requirements
Power Source
Check the voltage of the power supply before connecting it to the product. It should be the exact voltage
required for proper operation of the product.
Use in Hazardous Areas
When using the product in the first or second-class hazard location (Zone 1 or Zone 2) be sure to use an intrin-
sically safe or pressure and explosion-proof apparatus. Take the utmost care during the installation, wiring, and
piping of such apparatus to ensure the safety of the system. For safety reasons, maintenance or repairs on the
product while it is being used with such apparatus should only be performed by qualified personnel.
External Connection
When an external connection is required, the product should be protectively grounded before it is connected
to a measurement object or an external control circuit.
Caution!
To avoid injury or damage to devices:
Release pressure in the device before removing plugs, drum housing cover, or calibration windows.
1 Safety Instructions
Measuring systems for use in hazardous locations are accompanied by separate "Ex documentation", which
is an integral part of this Operating Manual. Strict compliance with installation instructions and - ratings, as
directed in this supplementary documentation, is mandatory.
•
Use the explosion proof type for measurement in areas where explosion hazards are present.
•
Devices employed in areas having explosion hazards should be installed and wired in keeping with explo-
sion proof regulations.
•
Devices installed in areas having explosion hazards must not be opened when the power is on. Tighten the
cable gland firmly.
•
Device maintenance and repair is restricted to meet explosion proof regulations.
•
Ensure that all personnel are properly qualified.
•
Observe the certification requirements as well as national and local regulations.
Power Supply
Confirm that local power supply voltage and frequency meet device specifications before attempting to operate.
Refer to "12.1 Technical Specifications".
Power cable
•
Use the power supply cable supplied with the device by the manufacturer, or other appropriate cable as
specified in the instruction manual.
•
The power source should have a ground terminal, and the power supply cable should have a ground line.
Refer to "5 Wiring".
Grounding
Do not unground device while the power supply is turned on. Doing so may cause a short circuit, resulting in
damage to devices, as well as personal injury from electric shock.
Caution!
Be certain to ground NMS5 before connecting input and output to other systems.
Designated NMS5 Use
NMS5 series is designed to measure liquid level in storage tanks or other containment facilities.
Auxiliary devices may be connected, as specified in this manual. Performance of such aftermarket devices,
however, is not guaranteed. Refer to manufacture supplied instructions when connecting aftermarket devices.
If device is used for any purpose other than intended, device damage, as well as personal injury or death may
result. Devices are supplied with IEC class 1 (ground terminal).
Caution!
Changes or modifications other than those expressly approved by Endress+Hauser are strictly prohibited.
Unauthorized modifications can cause malfunction or damage, resulting in serious injury or death.
1 Safety Instructions
Mounting with a stilling well is recommended for use in a tank which contains flammable liquid with conduc-
tivity at 10-8 S/m or less.
In case of installation without a stilling well, allow enough stilling time according to the table below before
lowering the level gauge down to the liquid surface. When a stilling well is used, the stilling time is shown as
in column “<10” in the table below.
conductivity
(S/cm)
Reference: Stilling Time (min.)
Liquid Volume in Tank (m3)
≤10
10~50
50~100
≥5000
≥10-8
≥1
≥1
≥1
≥2
10-12 ~10-8
≥2
≥3
≥10
≥30
10-14~10-12
≥4
≥5
≥60
≥120
≤10-14
≥10
≥10
≥120
≥240
(Japan National Institute of Occupational Safety and Health)
1 Safety Instructions
Notes on Safety Conventions and Symbols
In order to highlight safety-relevant or alternative operating procedures in the manual, the following
conventions have been used, each indicated by a corresponding symbol in the margin.
Safety Conventions
Warning!
A warning highlights actions or procedures which, if not performed correctly,
will lead to personal injury, a safety hazard or destruction of the instrument
Caution!
Caution highlights actions or procedures which, if not performed correctly, may
lead to personal injury or incorrect functioning of the instrument
Note!
A note highlights actions or procedures which, if not performed correctly, may
indirectly affect operation or may lead to an instrument response which is not
planned
Explosion protection
Device certified for Use in Explosion Hazardous Area
If the device has this symbol embossed on its name plate it can be installed in an
explosion hazardous area
Explosion Hazardous Areas
Symbol used in drawings to indicate explosion hazardous areas. Devices located
in and wiring entering areas with the designation “explosion hazardous areas”
must conform with the stated type of protection.
Safe Area (non-explosion hazardous area)
Symbol used in drawings to indicate, if necessary, non-explosion hazardous
areas. Devices located in safe areas still require a certificate if their outputs run
into explosion hazardous areas
Explosion protection
Direct Voltage
A terminal to which or from which a direct current or voltage may be applied or
supplied
Alternating Voltage
A terminal to which or from which an alternating (sine-wave) current or voltage
may be applied or supplied
Grounded Terminal
A grounded terminal, which as far as the operator is concerned, is already
grounded by means of an earth grounding system
Protective Grounding (earth) Terminal
A terminal which must be connected to earth ground prior to making any other
connection to the equipment
Equipotential Connection (earth bonding)
A connection made to the plant grounding system which may be of type e.g.
neutral star or equipotential line according to national or company practice
EX
EX
1 Safety Instructions
The following technical data are given on the instrument nameplate:
Figure. 1:
ATEX Approval Type Ex d/Ex d [ia]
Figure. 2:
ATEX Approval Type Ex d
�
�
�
�
�
�
�
�
�
�
�
�
�
�
�
�
�
�
��
XA00578G-A/08
Supply
Measuring range m
Displacer weight g
Displacer dia. mm�
Measure wire dia. mm�
Density range ~
Manufacturing date
Test date
Tester
Zero point of liquid level gauge is mm
under the reference point.
Only read level when indication "BAL" is present.
Certification no.
Tank ID
Ambient temperature: ~ +60°C
Process temperature: -200 ~ +200°C
KEMA 05 ATEX 2071
Ex d IIB T6...T3
ATEX II 1/2 G
Order code
Ser. no.
0044
PROSERVO NMS53
Made in Japan
Yamanashi 406-0846
Endress+Hauser Yamanashi Co.,Ltd.
IP67, NEMA 4X
Warning: Do not open the cover when energized.
�
�
�
�
�
�
�
�
�
�
�
�
�
�
�
�
�
�
Type code
Order code
Serial number
�������������������������
Measuring range
����������������
������������������
�����������������������
�������������������������������
�������������������������������
�������������������
������������������������
��������������
���������������������
��������������������������������
�������������������������
�������������������������
�������������������������
NP- ### - # (*2)
���������������������������������������������������������������������
�����������������������
����������������������
�
�
�
�
�
�
�
�
�
�
�
�
�
�
�
�
�
KEMA 05 ATEX 2071
Ex d IIC T6...T3
ATEX II 1/2 G
NP-2551-4
XA00578G-A/08
Certification no.
Tank ID
Ambient temperature: -20 ~ +60°C
Process temperature: -200 ~ +200°C
Order code
Ser. no.
0044
PROSERVO NMS53
Made in Japan
Yamanashi 406-0846
Endress+Hauser Yamanashi Co.,Ltd.
IP67, NEMA 4X
Warning: Do not open the cover when energized.
Supply
Measuring range m
Displacer weight g
Displacer dia. mm�
Measure wire dia. mm�
Density range ~
Manufacturing date
Test date
Tester
Zero point of liquid level gauge is mm
under the reference point.
Only read level when indication "BAL" is present.
�
�
�
�
�
�
�
�
�
�
�
�
�
�
�
�
�
Type code
Order code
Serial number
�������������������������
Measuring range
����������������
������������������
�����������������������
�������������������������������
�������������������������������
�������������������
������������������������
��������������
���������������������
�������������������������
�������������������������
�������������������������
2 Identification
FM Approval Type XP
Figure. 4:
FM Approval Type XP-AIS
�
�
�
NP-2346-4
Temperarure Class T4
APPROVED
F M
XP Class I, Div. 1, Gp. C, D
XP Class I, Zone 1, Gp. IIB
DIP Class II, III, Div. 1, Gp. E, F, G
Ambient temperature: -20 � +60�
NEMA 4X, IP67
Order code
Ser. no.
Input rated
Ambient temperature -20 ~
+60
Input
Output
Relay contact rated
30 V AC, 2A; 42 V DC, 2A, 60 W.
Measuring range m
Manufacturing date
Warning: Keep cover tight while circuits are alive.
Caution: Use supply wires suitable for 70� minimum.
Caution: A seal shall be installed within 50 mm of the enclosure.
PROSERVO NMS53
Made in Japan
Yamanashi 406-0846
Endress+Hauser Yamanashi Co.,Ltd.
�
�
�
�
�
�
�
�
�
Type code
Order code
Serial number
�������������������������
�������������������������
��������������������������
�
�
�
Install: per drawing Ex421-439-
NP-2421-5
APPROVED
F M
Temperarure class T4
Ambient temperature: -20 � +60�
XP-AIS Class I, Div. 1, Gp. C, D
XP-AIS Class I, Zone 1, Gp. IIB
DIP Class II, III, Div. 1, Gp. E, F, G
XP-ANI Class I, Div. 2, Gp. C, D
Order code
Ser. no.
Input rated
Ambient temperature -20 ~
+60
Input
Output
Relay contact rated
30 V AC, 2A; 42 V DC, 2A, 60 W.
Measuring range m
Manufacturing date
Warning: Keep cover tight while circuits are alive.
Caution: Use supply wires suitable for 70� minimum.
Caution: A seal shall be installed within 50 mm of the enclosure.
PROSERVO NMS53
Made in Japan
Yamanashi 406-0846
Endress+Hauser Yamanashi Co.,Ltd.
�
�
�
���Type code
���Order code
���Serial number
����������������������������
����������������������������
�����������������������������
���XA version number
�
2 Identification
CSA Approval Type Ex d[ia]
Figure. 6:
CSA Approval Type Ex d
���Type code
���Order code
���Serial number
���Input rating
����������������������������
�����������������������������
���Measuring ragne
���XA version number
NP-2526-2
Ex d[ia] IIB T4; Class I, Zone 1;
Class I, Div. 1, Gp. C, D ; [Ex ia]
Clsss II, Div. 1, Gp. E, F, G ;
Class III
Ambient temperature: -20 � +60 �
MWP psi
R
216422
Type 4X, IP67
Intrinsically Safe Outputs when
installed per drawing Ex540-742
Order code
Ser. no.
Input rated
Input
Output
Relay contact rated
30 V AC, 2A; 42 V DC, 2A, 60 W.
Measuring range m
Manufacturing date
Warning: Keep cover tight while circuits are alive.
Caution: Use supply wires suitable for 70� minimum.
Caution: A seal shall be installed within 50 mm of the enclosure.
PROSERVO NMS53
Made in Japan
Yamanashi 406-0846
Endress+Hauser Yamanashi Co.,Ltd.
5
2
3
1
4
6
7
⑧
�
�
�
NP-2342-3
Order code
Ser. no.
Input rated
Ambient temperature: -20 � +60�
MWP psi
Exd IIB T4;
Class I, Divsion 1, Groups C, D ;
Clsss II, Divsion 1, Groups E, F, G ;
Class III
R
216422
Type 4X, IP67
Input
Output
Relay contact rated
30 V AC, 2A; 42 V DC, 2A, 60 W.
Measuring range m
Manufacturing date
Warning: Keep cover tight while circuits are alive.
Caution: Use supply wires suitable for 70� minimum.
Caution: A seal shall be installed within 50 mm of the enclosure.
PROSERVO NMS53
Made in Japan
Yamanashi 406-0846
Endress+Hauser Yamanashi Co.,Ltd.
�
�
�
�
Type code
Order code
Serial number
Input rating
�������������������������
��������������������������
Measuring range
�
�
�
�
�
�
�
2 Identification
TIIS Approval Type Ex d
Figure. 8:
IEC Approval Type Ex d -20°C
�
�
�
�
�
�
�
�
�
�
�
�
�
�
�
�
�
�
�������Ex proof model :
�������Protection class :
�� / Rating
����
� Power supply:
���������/ Data communication: DC 48 V 300 mA�������
��������/ Contact input :� DC 30 V 0.6 W �������
������� / Contact output:� AC 250 V 1.5 A
�����������������������������������DC 30 V 9 W
�����/Ambient temperature: -20 � +60 �
� � � ��������������������������������
�����������������������������HIV�����0.05M��Km����
�������������������
Note: Be sure to cut off the power and cool down this instrument before opening
the cover.
In case of conduit wiring method, shall be use HIV wire (insulation resistance
over 0.05M-ohm.Km)
Zero point of liquid level gauge is mm
under the reference point.
Only read level when indication "BAL" is present.
NP-2645-1
IP67
NEMA 4X
Order code
Ser. no.
PROSERVO NMS53
���������������
Yamanashi 406-0846
Made in Japan
Endress+Hauser Yamanashi Co.,Ltd.
BA 00401G/08
Measuring range m
Displacer weight g
Displacer dia. mm�
Measure wire dia. mm�
Density range ~
Manufacturing date
Test date
Tester
Certification no.
Tank ID
�
�
�
�
�
�
�
�
�
�
�
�
�
�
�
�
�
�
Type code
Order code
Serial number
Ex-proof model
Protection class
�������������������������
Measuring range
Displacer weight
Displacer diameter
Measuring wire diameter
Density measurement lower limit
Density measurement upper limit
Date of manufacture
Date of manufacture test
Name of tester
�������������������������
�������������������������
�������������������������
�
�
�
�
�
�
�
�
�
�
�
�
�
�
�
�
�
�
Ex d IIB T6...T3 Ga/Gb
IECEx KEM 09.0062
NP-2619-2
XA00582G-A/08
Ambient temperature: ~ +60°C
Process temperature: -200 ~ +200°C
Order code
Ser. no.
0044
PROSERVO NMS53
Made in Japan
Yamanashi 406-0846
Endress+Hauser Yamanashi Co.,Ltd.
IP67, NEMA 4X
Warning: Do not open the cover when energized.
Supply
Measuring range m
Displacer weight g
Displacer dia. mm�
Measure wire dia. mm�
Density range ~
Manufacturing date
Test date
Tester
Zero point of liquid level gauge is mm
under the reference point.
Only read level when indication "BAL" is present.
Certification no.
Tank ID
�
�
�
�
�
�
�
�
�
�
�
�
�
�
�
�
�
�
Type code
Order code
Serial number
�������������������������
Measuring range
����������������
������������������
�����������������������
�������������������������������
�������������������������������
�������������������
������������������������
��������������
���������������������
����������������
�������������������������
�������������������������
�������������������������
-20°C
2 Identification
IEC Approval Type Ex d -40°C
Figure. 10:
IEC Approval Type Ex d IIC
�
�
�
�
�
�
�
�
�
�
�
�
�
�
�
�
�
�
Ex d IIB T6...T3 Ga/Gb
IECEx KEM 09.0062
NP-2620-2
XA00582G-A/08
Ambient temperature: ~ +60°C
Process temperature: -200 ~ +200°C
Order code
Ser. no.
0044
PROSERVO NMS53
Made in Japan
Yamanashi 406-0846
Endress+Hauser Yamanashi Co.,Ltd.
IP67, NEMA 4X
Warning: Do not open the cover when energized.
Supply
Measuring range m
Displacer weight g
Displacer dia. mm�
Measure wire dia. mm�
Density range ~
Manufacturing date
Test date
Tester
Zero point of liquid level gauge is mm
under the reference point.
Only read level when indication "BAL" is present.
Certification no.
Tank ID
�
�
�
�
�
�
�
�
�
�
�
�
�
�
�
�
�
�
Type code
Order code
Serial number
�������������������������
Measuring range
����������������
������������������
�����������������������
�������������������������������
�������������������������������
�������������������
������������������������
��������������
���������������������
����������������
�������������������������
�������������������������
�������������������������
-40°C
�
�
�
�
�
�
�
�
�
�
�
�
�
�
�
�
�
Ex d IIC T6...T3 Ga/Gb
IECEx KEM 09.0062
NP-2621-2
XA00582G-A/08
Ambient temperature: -20 ~ +60°C
Process temperature: -200 ~ +200°C
Order code
Ser. no.
0044
PROSERVO NMS53
Made in Japan
Yamanashi 406-0846
Endress+Hauser Yamanashi Co.,Ltd.
IP67, NEMA 4X
Warning: Do not open the cover when energized.
Supply
Measuring range m
Displacer weight g
Displacer dia. mm�
Measure wire dia. mm�
Density range ~
Manufacturing date
Test date
Tester
Zero point of liquid level gauge is mm
under the reference point.
Only read level when indication "BAL" is present.
Certification no.
Tank ID
�
�
�
�
�
�
�
�
�
�
�
�
�
�
�
�
�
Type code
Order code
Serial number
�������������������������
Measuring range
����������������
������������������
�����������������������
�������������������������������
�������������������������������
�������������������
������������������������
��������������
���������������������
�������������������������
�������������������������
�������������������������
2 Identification
Drum Housing Pressure Rating; Material:
1
0… 200mbar/20kPa/2.9psi; Alu
2
0… 200mbar/20kPa/2.9psi; stainless ste>
4
0… 5.88bar/588kPa/85.28psi; Alu
5
0… 5.88bar/588kPa/85.28psi; stainless steel
6
0… 24.5bar/2.45MPa/355.34psi; stainless steel
9
Special version, TSP-no. to be spec.
020
Approval:
0
Weather proof, IP 67 NEMA 4X
1
TIIS Ex d IIB T4
5
FM XP Cl. I Div. 1 Gr. C-D
N
FM XP-AIS Cl. I Div.1 Gr. C-D
6
CSA Cl. I Div. 1 Gr. C-D
O
CSA Ex d[ia] Cl. I Div. 1 Gr. C-D
G
ATEX II 1/2G Ex d IIB T6...T3
Q
ATEX II 1/2G Ex d IIC T6...T3
S
ATEX II 1/2G Ex d IIB T6...T3, -40°C
J
ATEX II 1/2G Ex d (ia) IIB T6...T3
U
ATEX II 1/2G Ex d (ia) IIB T6...T3, -40°C
A
IEC Ex d ia IIB T6...T3 Ga/Gb
B
IEC Ex d IIB T6...T3 Ga/Gb
C
IEC Ex d IIC T6...T3 Ga/Gb
D
IEC Ex d ia IIB T6...T3 Ga/Gb, -40°C
E
IEC Ex d IIB T6...T3 Ga/Gb, -40°C
9
Special version, TSP-no. to be spec.
030
Application:
A
Liquid Level
B
PTB (<1mm) type approval, liquid level
C
NMi (<1mm) type approval, liquid level
D Multi measurement, liquid level, I/F level, bottom, density
E
PTB (<1mm) type approval, liquid level, I/F level, bottom, density
F
NMi (<1mm) type approval, liquid level, I/F level, bottom, density
G Density profile multi measurement, liquid level, I/F level, bottom, density
H PTB (<1mm)type approval, density profile, liquid level, I/F level, bottom, density
J
NMi (<1mm) type approval, density profile, liquid level, I/F level, bottom, density
Y
Special version, TSP-no. to be spec.
040
Output 1:
F
Not selected
A
2-way 2-wire (V1 protocol)
J
2-way 2-wire (MDP protocol)
B
2-way 2-wire (BBB protocol)
C
2-way 2-wire (MIC, RS232C protocol)
D
2-way 2-wire (MIC protocol)
G
HART active
H
HART passive
L
Whessmatic 550, overvoltage protection
M Mark Space
N
Enraf BPM
P
Modbus RS 485
Y
Special version, TSP-no. to be spec.
050
Output 2:
0
Not selected
1
4 x relay SPST
2
2 x 4-20mA
3
4 x relay SPST, 2 x 4 - 20mA
4
2 x relay SPST, Overspill prevention TÜV
5
4 x relay SPST, 1 x 4 - 20mA
9
Special version, TSP-no. to be spec.
NMS5-
Product designation (part 1)
2 Identification
HART (NMT5xx, NRF560, pressure transmitter)
1
1 x spot temperature Pt100, HART (NRF560, pressure transmitter)
2
2 x operation contact, HART (NMT5xx, NRF560, pressure transmitter)
3
1 x spot temperature Pt100, 2 x operation contact, HART
(NMT5xx, NRF560, pressure transmitter)
4
1 x status, HART (NMT5xx, NRF560, pressure transmitter)
5
1 x spot temperature Pt100, 1 x status, HART (NMT5xx, NRF560, pressure transmitter)
6
1 x spot temperature Pt100, 1 x status, 2 x operation contact, HART
(NMT5xx, NRF560, pressure transmitter)
9
Special version, TSP-no. to be spec.
070
Measuring Range; wire:
C
0 - 28m; SUS316L, 0.15mm
L
0 - 36m; SUS316L, 0.15mm
N
0 - 47m; SUS316L, 0.15mm
H
0 - 16m; PFA>SUS316, 0.4mm
K
0 - 16m; Alloy C, 0.2mm
M
0 - 22m; Alloy C, 0.2mm
Y
Special version, TSP-no. to be spec.
080
Cable entry:
E
4 x thread G1/2
F
4 x thread G3/4
G 4 x thread NPT1/2
H 4 x thread NPT3/4
L
4 x thread M20
M 4 x thread M25
Y
Special version, TSP-no. to be spec.
090
Process connection:
A 10K 80A RF, flange JIS B2220
C 10K 80A FF, flange JIS B2220
U 10K 150A RF, flange JIS B2220
E
20K 80A RF, flange JIS B2220
G 3" 150Ibs RF, flange JIS B2220
J
3" 300lbs RF,flange ANSI B16.5
W 4" 300lbs RF, flange ANSI B16.5
T
6" 150lbs RF, flange ANSI B16.5
L
DN80 PN10 B1, flange EN1092-1 (DIN2527 B)
N DN80 PN25 B1, flange EN1092-1 (DIN2527 B)
Q 80A 150lbs RF, flange JPI 7S-15
S
80A 300lbs RF (apply for NMS5-6), flange JPI 7S-15
Y
Special version, TSP-no. to be spec.
100
Power supply:
3
85 - 264 VAC, 50/60 Hz
4
20 - 62 VDC, 20 - 55 VAC, 50/60 Hz
9
Special version, TSP-no. to be spec.
110
Displacer:
N
Cylindrical 30 mm, SUS316
K
Cylindrical 40 mm, SUS316
D
Cylindrical 50 mm, SUS316
W Cylindrical 30 mm, PTFE
V
Cylindrical 40 mm, PTFE
U
Cylindrical 50 mm, PTFE
T
Cylindrical 50 mm, AIloy C
B
Conical 50 mm, PTFE
R
Conical 70 mm, SUS316
S
Conical 110 mm, SUS316
Y
Special version, TSP-no. to be spec.
NMS5-
Product designation (part 2)
2 Identification
O-ring; chamber finishing:
0
NBR; not selected
1
Silicon rubber; not selected
5
Silicone rubber; FEP coated
3
PTFE (wire drum FKM); not selected
4
PTFE (Wire drum FKM); FEP coated
6
CR; not selected
2
FKM; not selected
7
FKM; FEP coated
8
FFKM; not selected
A FFKM; FEP coated
9
Special version, TSP-no. to be spec.
130
Options:
A
not selected
C Rc3/8 cleaning nozzle
D Rc3/8 gas purging nozzle
E
Guide wire installation
G Relief valve
H Relief valve, pressure gauge
J
Sunshade
L
Rc3/8 gas purging nozzle, sunshade
M Rc3/8 cleaning nozzle, sunshade
N Rc3/8 gas purging nozzle, guide wire installation
P
Rc3/8 cleaning nozzle, guide wire installation
Q Guide wire installation, sunshade
R
Relief valve, guide wire installation
S
Relief valve, pressure gauge, guide wire installation
T
Relief valve, sunshade
U Relief valve, pressure gauge, sunshade
Y
Special version, TSP-no. to be spec.
NMS5-
Complete product designation
2 Identification
Scope of Delivery
Caution!
It is extremely important to follow the instructions concerning the unpacking, transportation and storage of
measuring instruments provided in the chapter “Incoming Acceptance, Transportation, Storage”.
The scope of delivery consists of:
•
Assembled Instrument
Accompanying documentation:
•
Installation Instructions (this manual)
•
Safety Instructions (XA)
•
Functional Safety Manual (SD), Supplied with 4-20mA and Overspill Prevention Relay Output
2.4
Certificates and Approvals
CE Mark, Declaration of Conformity
The device is designed to meet state-of-the-art safety requirements, has been tested and left the factory in a
condition in which it is safe to operate. The device complies with the applicable standards and regulations as
listed in the EC declaration of conformity and thus complies with the statutory requirements of the EC
directives. Endress+Hauser confirms the successful testing of the device by affixing to it the CE mark.
2.5
Registered Trademarks
HART®
Registered trademark of HART Communication Foundation, Austin, USA
ToF®
Registered trademark of the company Endress+Hauser GmbH+Co. KG, Maulburg, Germany
FieldCare®
Registered trademark of the company Endress+Hauser Flowtec AG, Rheinach, CH
2 Identification
Incoming Acceptance, Transport, and Storage
3.1.1
Incoming Acceptance
Check the packing and contents for any signs of damage.
Check the shipment, and make sure that nothing is missing and that the items match your order.
3.1.2
Transport
Caution!
•
Follow the safety instructions and conditions of transportation for instruments in excess of 18kg (40 lbs.).
•
Do not lift the measuring instrument by its housing during transportation.
3.1.3
Storage
Pack the measuring instrument so that it is protected against impacts during storage and transportation.
The original packing material provides the optimum protection for this.
The allowed storage temperature is -40°C to +85°C (-40°F to +185°F)
3 Installation
Terms Related to Tank Measurements
Figure 11:
Terms Related to Tank Measurements
Lower Layer
Middle Layer
Upper Layer
Upper Interface
Lower Interface
Liquid Level
Nozzle
NMS5
Reference Line
Dip-point Offset
Upper Limit Stop Level
Upper Limit Alarm
Ullage Level
Gauge Reference Length
Tank Height
Upper Interface Level
Lower Interface Level
Lower Limit Alarm
Lower Limit Stop Level
Dipping Reference Point
(Dipping Data Place)
Liquid Depth
(Innage Level)
Measurement Hatch
(Dipping Reference)
Bottom of Tank
3 Installation
Design and Dimensions
Figure 12:
NMS5 Dimension
NMS5 - 1/ NMS5 - 4
�
�
�
�
�
�
�
�
�
�
�
Electrical compartment
Terminal box
Display (LCD)
Cable entery (A, B, C, D)
Wire drum housing
Flange
Measuring wire
Displacer
Calibration window
Window cover
Touch control
Max. 39.7mm
Travel Distance of Displacer
34.6mm for 28m Wire (Φ0.15mm)
39.7mm for 36m Wire (Φ0.15mm)
���������������������������������������������������
Displacer position at bottom of tank
NMS5 - 2/ NMS5 - 5 / NMS5 - 6
Max. 41.6mm
Travel Distance of Displacer
34.6mm for 28m Wire (Φ0.15mm)
39.7mm for 36m Wire (Φ0.15mm)
41.6mm for 47m Wire (Φ0.15mm)
���������������������������������������������������
Displacer position at bottom of tank
3 Installation
Necessary Tools for Installation
The following tools are required when installing NMS5.
Figure. 13:
Tools to Be Prepared
24,26,30,32
350mm
3mm and 5mm
B
mm
3
1.25 , 2.0
sq
sq
Box End Wrench
Crescent Wrench
Allen Wrench (Hex Key)
Screw Driver
■ Flat Head
■ Phillips
Wire Terminal
Water Pump Pliers
Wire Cutters or
Terminal Pliers
Density Calibration
Test Weight
3 Installation
Mounting of NMS5
The following installation procedures are available for NMS5.
•
Mounting without guide system.
•
Mounting with stilling well
•
Mounting with guide wire
4.1
Application Drawing for Tank
Mounting with stilling well or guide wire is required for the following applications:
•
Floating Roof Tank
•
Covered Floating Roof Tank
•
Tank with Strong Agitator or Heavy Turbulence
Mounting without any guide system covers all cases that are not listed above.
Refer to Figure 14 of applications (with and without mounting stilling well, with mounting guide wire)
Figure. 14:
Application for Each Tank
Floating Roof Tank
Fixed Roof Tank
Without Stilling Well
Fixed Roof Tank
With Stilling Well
Covered Floating
Roof Tank
Spherical or Bullet Tank
With Guide Wire
Fixed Roof Tank
Spherical or Bullet Tank
Floating Roof Tank
�����������������������������������������������
Covered Floating Roof Tank
4 Mounting of NMS5
Mounting without Guide System
NMS5 is mounted on a nozzle of the tank roof without any guide system (Refer to Figure. 14: Application for
Each Tank). The mounting preparations require the observance of some recommendations for setting the noz-
zle and the minimum measuring level.
Figure. 15:
Mounting without Guide System
4.3
Mounting with Stilling Well
Stilling Well Diameter
The stiling well diameter that is required to protect the measuring wire without disturbing its operation varies
depending on the tank height. The stiling well could either be constant diameter, or thinner at its upper part
and thicker at its lower part. The figure 16 shows two examples of the latter case, namely an asymmetric pipe
and a concentric pipe.
Figure. 16:
Mounting with Stilling Well, Asymmetric Pipe, and Connection Pipe
Displacer
NMS5
d
D1
Valve
Asymmetric Pipe
Concentric
Pipe
Displacer
Maintenace Chamber
NMS5
Asymmetric Pipe (Side View)
Concentric Pipe (Front View)
NMS5
4 Mounting of NMS5
The valve is required when mounting NMS5 on pressurized liquid tanks.
NMS5 must be mounted on the asymmetric pipe in the direction shown on Figure 16.
To calculate the required pipe diameters, the formula below should be used.
The variables and constants have the following meanings.
D1
Inner diameter of the upper part of the pipe
D2
Inner diameter of the lower part of the pipe
L
Length of the pipe (from the flange of NMS5 to bottom of the stilling well) meters
v
Deviation of the pipe from the vertical per length (mm/m)
d
Diameter of the displacer
e
Lateral shift of the displacer per length due to the groove of the wire drum
•
Upper Diameter of Stilling Well
D1 > d + 10 mm
D1 should be 3" or more.
•
Lower Diameter of Stilling Well
- Asymmetric Pipe
D2 > d+ eL + 2vL+ 10mm
- Concentric Pipe
D2 > d + 2eL + 2vL + 10mm
Recommendations for Mounting Stilling Well
Note!
Follow the recommendations for mounting NMS5 with stilling well:
•
Keep the pipe connection welds smooth.
•
While drilling holes into the pipe, keep the interior surface of the holes clear of metal chips and burrs.
•
Coat or paint the interior surface of the pipe to avoid rust.
•
Keep the pipe as perfectly vertical as possible. Check this by a plumb.
•
Install the asymmetric pipe under the valve and fit the centers of the NMS5 and the valve.
•
Set the center of the lower part of the asymmetric pipe to the direction of the displacer motion.
•
Observe recommendations as per API MPMS chapter 3.1B.
•
Confirm grounding between NMS NMS5 and tank nozzle.
Feature: 070
Measuring Range/Material/Diameter of Wire
e = Lateral Shift (mm)
C
0-28m: SUS316L, 0.15mm
0.37
H
0-16m: PFA>SUS316, 0.4mm
0.40
K
0-16m: Alloy C, 0.2mm
0.47
L
0-36m: SUS316L, 0.15mm
0.33
M
0-22m: Alloy C, 0.2mm
0.47
N
0-47m: SUS316L, 0.15mm
0.30
4 Mounting of NMS5
Mounting NMS5 with Guide Wire
It is also possible to guide the displacer by a guide wire to prevent lateral motion.
Guide Wire Installation Procedure
1.
Secure the guide wire [4] to the wire hook.
2.
Secure the springs [3] to the reducer plate [2] (see Figure 17).
3.
Fix NMS5 [1] temporarily so that springs are located at the outside the reducer plate.
4.
Perform weight calibration.
5.
Fix NMS5 to the reducer plate.
6.
Put displacer ring [7] through guide wire and set the displacer [5] so that marking ring [6] is located on
the reference position.
This completes the guide wire installation procedure.
Figure. 17: Mounting with Guide
Wire
NMS5 [1]
Chamber for Maintenance
Displacer (Ø50 mm, Ø70mm, and Ø110mm) [5]
Bottom of Tank
Welding
Anchor Hook Plate
Caliblation Chanber
NHC4HP (for High Pressure)
NHC4LP (for Low Pressure)
Ball Valve
NHV4A (with ANSI Flange)
NHV4J (with JIS Flange)
172
300
Ø150
CALIBRATION CHAMBER (for Low Pressure)
Flange
Flange
Measuring Wire
Tank
Guide Wire [4]
Sleeve
Spring [3]
Reducer Plate [2]
Spring [3] Sleeve
Guide Wire
Sleeve
Crimp Tool
Flange 6”
100mm
Reference
Position
Marking Ring [6]
*2
*1
*1
*2
Accessory can be ordered separately.
*1
Reducer Plate : 6mm for Low Pressur
Reducer Flange : for High Pressure, (depending on standard)
Ring [7]
4 Mounting of NMS5
Weight Calibration Method for Guide Wire Installation
Figure. 18:
Weight Calibration for NMS with Guide Wire
Weight Calibration Procedure
1.
Fix NMS5 temporarily so that NMS5 is located at the outside the reducer plate to perform weight
calibration.
2.
Replace the measuring wire to an appropriate length.
• During standard weight calibration, the displacer moves up until it touches the mechanical stop in the
housing.
• After weight calibration is finished, the displacer stops 70mm (the value set in the matrix (configurable,
GVH378)) below the mechanical stop. This is the Reference Position.
3.
During this special weight calibration, the guide rings of the displacer touches the bottom face of the
flange.
4.
The displacer goes down 70mm and stops. NMS5 remembers the distance of 70mm as Reference
Position, and weight table calibration resumes.
5.
When weight calibration is complete, return NMS5 back to its original mounting position on the tank
nozzle.
This completes the weight calibration procedure.
70mm
Reference
Position
4 Mounting of NMS5
Type of Tanks
Fixed Roof Tank with Stilling Well
Figure. 19:
Stilling Well in Fixed Roof Tank
Pressurized Tank with Stilling Well and Ball Valve
Figure. 20:
Stilling Well in Pressurized Tank
Measuring Wire
Displacer
Stilling Well
E
-
+
NMT539
NMS5
E-+
Ball Valve
Reducer
Meas. Wire
Displacer
Stilling Well
Proservo NMS5
Prothermo NMT539
4 Mounting of NMS5
Floating Roof Tank and /or Covered Floating Roof Tank
Figure. 21:
Stilling Well in Floating Roof Tank/Covered Floating Roof Tank
Note!
When NMS5 is installed in a floating roof tank, ensure to use a stilling well.
Figure. 22:
Stilling Well in Floating Roof Tank
E
-
+
Measuring Wire
Displacer
Stilling Well
NMS5
NMT539
E
-
+
Measuring Wire
Displacer
NMS5
4 Mounting of NMS5
Preparation for Mounting of NMS5
4.5.1
Flange
A nozzle and a flange should be prepared prior to mounting NMS5 on the tank. The flange size and the rating
of NMS5 vary depending on the customer’s specifications.
Note!
•
Check the flange size of NMS5.
•
Mount the flange on the top of the tank. The deviation of the flange from the horizontal plane should not
exceed +/- 1 degree.
•
When mounting NMS5 on a long nozzle, ensure that the displacer does not touch the inner side of the noz-
zle when the nozzle is in the vertical position.
Figure. 23:
Allowable Inclination of Mounting Flange
Note!
When NMS5 is installed without a guide system, follow the recommendations below:
•
Set the mounting nozzle in the sector between 45 and 90 deg (or -45 and -90 deg.) apart from the inlet
pipe of the tank. This prevents heavy swing of the displacer caused by wave or turbulence of the inlet liquid.
•
Set the mounting nozzle at 500mm or more away from the tank wall. Ensure that the measurement is not
influenced by changes of the ambient temperature.
•
Set the minimum measuring level at 500mm or more above the top of the inlet pipe. This protects the
displacer from direct flow of the inlet liquid.
•
If the stilling well can not be mounted in the tank due to the shape or condition of a tank, attaching a guide
system is recommended. Consult E+H services for further information.
Figure. 24:
Recommended Setting of Mounting Nozzle and Minimum Measuring Level
Caution!
•
Before pouring liquid into the tank, confirm that liquid flowing through the inlet of the pipe does not touch
the displacer directly.
•
When discharging liquid of the tank, ensure that the displacer does not get caught in the liquid current and
sucked into the outlet pipe.
Mounting Nozzle
max. +1°
max. -1°
-90°
90°
-45°
45°
0°
Inlet
Pipe
Mounting
Nozzle
Mounting
Nozzle
Minimum Measuring Level
��������
Inlet
Pipe
��������
4 Mounting of NMS5
When liquid measured by NMS5 has a conductivity at 10-8 s/cm or less, it is quasi-nonconductive. In this case,
using a stilling well or guide wire is recommended. This releases the electrostatic charge on liquid surface. (refer
to "1.5 Electrostatic Charge" details on a stiling time for a volume of a charged object).
4.7
Installation for Wire Drum and Displacer
When NMS5 is delivered, there are 2 ways of packing procedures depending on mounting methods of a dis-
placer.
1.
For All-in-one method, the displacer is mounted on the measuring wire of NMS5. Follow the instructions
to remove the packing attached to NMS5.
2.
Displacer shipped separately. It is necessary to install the displacer in the measuring wire inside NMS5.
Follow the instructions below.
4.7.1
All-in-one
As NMS5 is protected by some packaging materials, remove them as follows prior to mounting NMS5 (refer to
Figure 25: Removing Procedure for All-in-one).
Packaging Material Removing Procedure
Step
Procedure
Notes
1
1.
Hold the gauge so that it stays horizontal
against the flange.
2.
Cut the fixing bands.
3.
Remove the displacer holder.
• Do not tilt the gauge after removing the
displacer holder.
2
4.
Mount NMS5 on the nozzle .
•
Be sure that the measuring wire hangs
vertically.
•
Confirm that there are no kinks or other
malfunctions in the calibration window.
3
5.
Remove the drum housing cover from NMS5
unit.
6.
Loosen two screws on the wire drum holder
and remove the holder and the bracket.
•
Be sure not to lose O-ring and fixed bolts for
the cover for the drum housing.
4
7.
Remove the tape from the wire drum care-
fully.
•
Remove the tape by hands to avoid damage
to wire drum.
•
Make sure that the measuring wire is
wrapped so that it fits correctly in grooves.
•
If not, rewrap the wire so it fits correctly in
grooves.
5
8.
Attach the drum housing cover.
•
Confirm that O-ring is in groove.
6
9.
Turn on the main power and confirm that
NMS5 shows "STOP" and that level shows
"16050 ~16060 mm".
10.
Set "LEVEL" command, then set "STOP"
where the displacer stops 16000 mm or less
near the 15950 mm level.
11.
Set "UP" command and the displacer will
automatically stop at 16000 mm.
12.
Set "STOP" command.
4 Mounting of NMS5
Removing Packing Materials for All-in-one
Displacer Holder
Fixing Band
Tape
Wire Drum Holder
Set Screw
Remove a wire drum holder
using a straight slot screwdriver.
Drum Housing Cover
4 Mounting of NMS5
Displacer Shipped Separately
It is necessary to install the displacer on the measuring wire inside NMS5.
Procedure A : Install Displacer prior to Installing NMS5 on Tank
Figures
Procedure
Notes
Figure. 26:
Displacer Installation - 1A
1.
Put NMS5 on the blocks or
a pedestal.
2.
Confirm that there is
enough space under
NMS5.
• Be careful not to drop NMS.
Figure. 27:
Displacer Installation - 2A
3.
Remove M6 bolts.
4.
Remove the wire drum
cover and bracket.
5.
Remove the cover, the
wire drum bracket, and
the wire drum.
6.
Remove the tape.
7.
Mount the wire drum, the
wire drum bracket, and
the cover again.
•
Handle the measuring wire
with care.
Figure. 28:
Displacer Installation - 3A
8.
Rotate the wire drum
clockwise until the
measuring wire ring
appears out side the flange.
9.
Hook the displacer on the
ring of the measuring wire.
10.
Secure the displacer to the
measuring wire by using
the securing wire.
11.
Install the ground wire of
the displacer (refer to
Appendix "14.3.1
Displacer Installation").
12.
Rotate the wire drum
counterclockwise until
measuring wire ring
appears in the calibration
window.
•
Wire drum rotates every
one-fifth revolution.
•
Displacer moves 300mm
each one revolution.
Figure. 29:
Displacer Installation - 4A
13.
Mount NMS5 on the
nozzle of the tank top.
14.
Confirm that the displacer
does not touch inside the
nozzle inside.
15.
Replace the drum housing
cover.
•
Confirm that the measuring
wire is wrapped
appropriately in the
grooves.
•
If not, rewind the wire
correctly in the grooves.
Wire Drum
Wire Drum Bracket
M6 Bolts
Tape
Cover
Measuring Wire Ring
Securing Wire
Displacer
Ground Wire
Nozzle
NMS5
4 Mounting of NMS5
Displacer Shipped Separately (Diameter 50mm Displacer)
In the case of the diameter 50mm displacer, the displacer can be installed from the calibration window.
Note!
It is not possible to install the displacer from the calibration window in the following cases.
•
Except Ø50mm, SUS and alloy C, cylindrical displacer
•
Except 0.15mm, SUS measuring wire
Procedure B : Install Displacer after Installing NMS5 on Tank
Figures
Procedures
Notes
Figure. 30:
Displacer Installation - 1B
1.
Remove the calibration
window cover.
•
Ensure not to kink and
damage the measuring
wire.
Figure. 31:
Displacer Installation - 2B
2.
Remove M6 bolts.
3.
Remove the cover and
bracket.
4.
Remove the wire drum
from the drum housing.
5.
Remove the tape that is
securing the wire.
•
Handle the measuring wire
with care.
Figure. 32:
Displacer Installation - 3B
6.
Hold the wire drum by
using single hand and drop
the measuring wire down
at approximately 50cm.
7.
Secure the wire
temporarily with the tape.
8.
Insert the wire ring into
the drum housing.
9.
Pull the wire ring so that
the ring comes out from
the calibration window.
•
Handle the measuring wire
with care.
Figure. 33:
Displacer Installation - 4B
10.
Insert the wire drum
temporarily into the drum
housing.
11.
Hook the displacer on the
wire ring.
12.
Secure the displacer to the
measuring wire by using
the securing wire.
13.
Install the ground wire for
the displacer (refer to
"14.3.1 Displacer
Installation" for details).
•
Handle the measuring wire
with care. It may cause
damage by kink.
Calibration Window
NMS5
Wire Drum
Wire Drum Bracket
M6 Bolts
Tape
Cover
Tape
Wire Drum
Wire Ring
Calibration Window
Wire Drum
Securing Wire
Ground Wire
Displacer
4 Mounting of NMS5
Displacer Installation - 5B
14.
Hold the wire drum and
the displacer by hands.
15.
Remove the wire drum
from the drum housing
and drop the measuring
wire down approximately
50 cm.
16.
Hold the wire drum up
and place the displacer
into the calibration
window.
17.
Hold the displacer at
center of calibration
window.
18.
Hold other hand (wire
drum) up and add tension
to measuring wire not to
drop displacer down
rapidly.
•
Wire drum rotates every
one-fifth revolution.
•
Displacer move over
300mm each one
revolution.
Figure. 35:
Displacer Installation - 6B
19.
Release the hand from the
displacer.
20.
Remove the tape from the
wire drum.
21.
Insert the wire drum into
the drum housing.
22.
Set the wire drum bracket.
23.
Rotate the wire drum with
the hand a couple of times
and check that the
displacer does not touch
inside the nozzle.
24.
Pull the ring of the
measuring wire up so that
it can be seen from the
calibration window.
25.
Replace the covers of the
wire drum and the
calibration window.
•
Confirm that the measuring
wire is wrapped correctly in
grooves before turning on
the power.
•
If not, rewind the wire
correctly in grooves.
Figures
Procedures
Notes
Wire Drum
Displacer
Displacer
Window Cover
Cover
Wire Drum Bracket
Wire Drum
4 Mounting of NMS5
The electrical connections of NMS5 are shown in Figures 36-39.
Note!
The power supply cable should have the following specifications:
•
PVC, PE, or equivalently isolated
•
600 V insulation voltage or equivalent.
•
Temperature rating 80 or high
The size of the core is defined by the core resistance, voltage drop, and required power consumption. The max-
imum power consumption of NMS5 is 50 VA.
Caution!
•
Connect the ground terminal to the ground line inside or outside the terminal box.
•
Use a wire 4mm2 or more for ground.
•
Use cables and wires of sufficient and appropriate size and length to make a solid connection at each
terminal required.
Warnung!
Do not stretch cable and wires. This may lead to failure, loss of function, and/or damage to the device and
facility.
Warnung!
Cut the trim cables and wires for appropriate length. Do not leave the extra cables and wires in the electrical
compartment. This may lead to failure, loss of function, and /or damage to the device and facility.
5 Wiring
Ex d, Output 1 : V1, Modbus RS485, HART, Enraf BPM
Figure. 36:
Electrical Connection of NMS5
NMS5 - X X X X X X X X X X X
1
5
6
G
Q
S
B
C
E
N
P
F
A
J
B
C
D
G
H
14
AL4 COM
18
CTR2
20
OT1+
22
OT2+
AL4
15
CTR1
17
N.C.
19
OT1-
21
OT2-
23
4
8
AL1
10
AL2
12
AL3
AL3
13
AL2
11
AL1
9
RC
B/-
7
5
6
RC
A/+
24
NRF+
A
25
NRF-
B
26
b
1
2
3
ARS
L
ARS
G
ARS
POWL
POWN
POWG
ARSL
ARSG
ARSN
NRF+
NRF-
RCA/+
RCB/-
ALM1/OSP
ALM1/OSP
ALM2/OSP
ALM2/OSP
ALM3
1
2
3
L
N
G
AC Power Supply
24
25
26
+
-
+
-
+
-
ALM3
14
15
16
17
18
19
20
21
22
+
-
+
-
23
4
5
6
7
8
9
10
11
12
13
ALM4
ALM4
COM
CTR1
CTR2
OUT1+
OUT1-
OUT2+
OUT2-
TMPA/NMT+
TMPB/NMT-
TMPb
Fuse
F1
L
N
G
Lightning Arrester
NRF or Non IS HART Devices
V1, Modbus RS485, HART, Enraf BPM
Alarm Contact (Option)
External contact Input
(Option)
Alarm Contact (Option)
DC4...20mA Output (Option)
DC4...20mA Output (Option)
NMT or HART
Device (NonIS)
Ground
or Pt100 ohm Input
N
16
L
N
GND
Spot
Hoist
Common
NMT+NMT-
Alarm Contact (Option)
Alarm Contact (Option)
Order Code Example
5 Wiring
Ex d [ia], Output 1 : V1, Modbus RS485, HART, Enraf BPM
Figure. 37:
Electrical Connection of NMS5 (with IS HART Connection)
14
AL4
COM
18
CTR2
20
B+
22
MK
AL4
15
CTR1
17
N.C.
19
B-
21
SP
23
4
P_B+
8
AL1
10
AL2
12
AL3
AL3
13
AL2
11
AL1
9
7
P_B-
5
6
24
P_A+
A
25
P_A-
B
26
b
1
2
3
ARS
L
ARS
G
ARS
N
POWL
POWN
POWG
ARSL
ARSG
ARSN
P_B+
P_B-
ALM1 / OSP
ALM1 / OSP
ALM2 / OSP
ALM2 / OSP
ALM3
1
2
3
L
N
G
AC Power Supply
24
25
26
+
-
+
-
+
-
ALM3
14
15
16
17
18
19
20
21
22
+
-
23
4
5
6
7
8
9
10
11
12
13
ALM4
ALM4
COM
CTR1
CTR2
B+
B-
MK
SP
TMP A /P_A+
TMP B/P_A-
TMP b
Fuse
F1
L
N
G
Lightning Arrester
NRF or Non-IS HART Devices
Alarm Contact (Option)
Alarm Contact (Option)
Alarm Contact (Option)
External Contact Input
(Option)
Alarm Contact (Option)
Common
Hoist
Stop
Ground
NMT or HART
Device (IS)
or Pt100 ohm (IS) Input
N
L
GND
16
V1, Modbus, RS485, HART, Enraf BPM
DC4...20mA Output (Option)
DC4...20mA Output (Option)
Order Code Example
NMS5 - X X X X X X X X X X X
J
N
O
U
A
D
N
P
F
A
J
B
C
D
G
H
5 Wiring
Exd, Output 1: Whessoe Matic 550 (WM550) Protocol
Figure. 38:
Electrical Connection of NMS5 (with WM550 Protocol/without IS HART Connection)
NMS5 - X X L X X X X X X X X X
1
5
6
G
Q
S
B
C
E
14
AL4
COM
18
CTR2
20
Ch1+
22
Oh2+
AL4
15
CTR1
17
N.C.
19
Ch1-
21
Oh2-
23
4
8
AL1
10
AL2
12
AL3
AL3
13
AL2
11
AL1
9
7
5
6
24
A
25
B
26
b
1
2
3
ARS
L
ARS
G
ARS
N
POWL
POWN
POWG
ARSL
ARSG
ARSN
ALM1/OSP
ALM1/OSP
ALM2/OSP
ALM2/OSP
ALM3
1
2
3
L
N
G
AC power Supply
24
25
26
+
-
+
-
+
-
ALM3
14
15
16
17
18
19
20
21
22
+
-
+
-
23
4
5
6
7
8
9
10
11
12
13
ALM4
ALM4
COM
CTR1
CTR2
Ch1+
Ch1-
Ch2+
Ch2-
TMPA /NMT+
TMPB/NMT-
TMP b
Fuse
F1
L
N
G
Lightning Arrester
NRF or Non IS HART Devices
Alarm Contact (Option)
Alarm Contact (Option)
Alarm Contact (Option)
External Contact Input
(Option)
Alarm Contact (Option)
Whessoe Matic 550
Channel 1
Whessoe Matic 550
Channel 2
Common
Ground
or Pt100 ohm Input
NMT or HART
Device (Non IS)
L
N
GND
NMT+ NMT-
Stop
Hoist
16
P_B+
P_B-
NRF+
NRF-
Order Code Example
5 Wiring
Exd [ia], Output 1: Whessoe Matic 550 (WM550) Protocol
Figure. 39:
Electrical Connection of NMS5 (with WM550 Protocol/ IS HART Connection)
NMS5 - X X L X X X X X X X X X
J
N
O
U
A
D
Order Code Example
14
AL4
COM
18
CTR2
20
Ch1+
22
Oh2+
AL4
15
CTR1
17
N.C.
19
Ch1-
21
Oh2-
23
4
P_B+
8
AL1
10
AL2
12
AL3
AL3
13
AL2
11
AL1
9
7
P_B-
5
6
24
P_A+
A
25
P_A-
B
26
b
1
2
3
ARS
L
ARS
G
ARS
N
POWL
POWN
POWG
ARSL
ARSG
ARSN
P_B+
P_B-
ALM1/OSP
ALM1/OSP
ALM2/OSP
ALM2/OSP
ALM3
1
2
3
L
N
G
AC power Supply
24
25
26
+
-
+
-
+
-
ALM3
14
15
16
17
18
19
20
21
22
+
-
+
-
23
4
5
6
7
8
9
10
11
12
13
ALM4
ALM4
COM
CTR1
CTR2
Ch1+
Ch1-
Ch2+
Ch2-
TMPA/P_A+
TMPB/P_A-
TMP b
Fuse
F1
L
N
G
Lightning Arrester
Alarm Contact (Option)
Alarm Contact (Option)
Alarm Contact (Option)
External Contact Input
(Option)
Alarm Contact (Option)
Whessoe Matic 550
Channel 1
Whessoe Matic 550
Channel 2
Common
Ground
NMT or HART
Device (IS)
NRF or Non-IS HARTDevices
Pt100 ohm (IS) Input
L
GND
N
16
or
Spot
Hoist
5 Wiring
Exd, Output 1: Mark/Space (M/S) Protocol
Figure. 40:
Electrical Connection of NMS5 (with M/S Protocol without IS HART Connection)
NMS5 - X X M X X X X X X X X X
1
5
6
G
Q
S
B
C
E
14
AL4
COM
18
CTR2
20
B+
22
MK
AL4
15
CTR1
17
N.C.
19
B-
21
SP
23
4
8
AL1
10
AL2
12
AL3
AL3
13
AL2
11
AL1
9
7
5
6
24
A
25
B
26
b
1
2
3
ARS
L
ARS
G
ARS
N
POWL
POWN
POWG
ARSL
ARSG
ARSN
ALM1 / OSP
ALM1 / OSP
ALM2 / OSP
ALM2 / OSP
ALM3
1
2
3
L
N
G
AC power Supply
24
25
26
+
-
+
-
+
-
ALM3
14
15
16
17
18
19
20
21
22
+
-
23
4
5
6
7
8
9
10
11
12
13
ALM4
ALM4
COM
CTR1
CTR2
B+
B-
MK
SP
TMP A /NMT+
TMP B/NMT-
TMP b
Fuse
F1
L
N
G
Lightning Arrester
NRF or Non IS HART Devices
Alarm Contact (Option)
Alarm Contact (Option)
Alarm Contact (Option)
External contact Input
(Option)
Alarm Contact (Option)
Common
Hoist
Stop
Ground
NMT or HART
Device (Non IS)
Mark / Space
or Pt100 ohm Input
NMT+ NMT-
L
GND
N
16
P_B+
P_B-
NRF+
NRF-
Order Code Example
5 Wiring
Exd [ia], Output 1: Mark/Space (M/S) Protocol
Figure. 41:
Electrical Connection of NMS5 (with M/S Protocol with IS HART Connection)
NMS5 - X X M X X X X X X X X X
J
N
O
U
A
D
14
AL4
COM
18
CTR2
20
B+
22
MK
AL4
15
CTR1
17
N.C.
19
B-
21
SP
23
4
P_B+
8
AL1
10
AL2
12
AL3
AL3
13
AL2
11
AL1
9
7
P_B-
5
6
24
P_A+
A
25
P_A-
B
26
b
1
2
3
ARS
L
ARS
G
ARS
N
POWL
POWN
POWG
ARSL
ARSG
ARSN
P_B+
P_B-
ALM1 / OSP
ALM1 / OSP
ALM2 / OSP
ALM2 / OSP
ALM3
1
2
3
L
N
G
AC power Supply
24
25
26
+
-
+
-
+
-
ALM3
14
15
16
17
18
19
20
21
22
+
-
23
4
5
6
7
8
9
10
11
12
13
ALM4
ALM4
COM
CTR1
CTR2
B+
B-
MK
SP
TMP A /P_A+
TMP B/P_A-
TMP b
Fuse
F1
L
N
G
Lightning Arrester
NRF or Non-IS HART Devices
Alarm Contact (Option)
Alarm Contact (Option)
Alarm Contact (Option)
External Contact Input
(Option)
Alarm Contact (Option)
Common
Hoist
Stop
Ground
NMT or HART
Device (IS)
Mark / Space
or Pt100 ohm (IS) Input
N
L
GND
16
Order Code Example
5 Wiring
Using the external operation switch, ensure that the settings of the system are the same as shown in the fol-
lowing logic setting table. This will allow user to input operation signals.
(Hoist-CTR1)(Stop-CTR2)
5.2.7
Input and Output
Input
– Contact Switch
– HART
– Pt 100
Output
– V1
– WM550
– M/S
– ENRAF BPM
– MODBUS
– HART
– ANALOG 4-20mA
– Overspill Prevention (OSP)
Warnung!
•
The cable used for input and/or output must be 24 AWG or more and must be screened or steel armored.
A twisted pair is required for HART and/or RS 485 signal.
•
Two or three cores for mains, two cores for digital output and two cores for HART input are normally used
for the cabling of NMS5. The instrument has max. four cable entries.
•
Check the cable size and the number of cables prior to ordering NMS5.
5.2.8
Cable Gland
When there are extra cable entries not be used, mount threaded plugs for waterproof.
Refer to "Safety Instruction XA00578G-A".
Caution!
When ordering TIIS Ex d specification, cable glands are attached with NMS5. Ensure to use the cable glands.
CTR 1
CTR 2
OPERATION
OFF
OFF
LEVEL
ON
OFF
HOIST
OFF
ON
STOP
ON
ON
INTERFACE
5 Wiring
Temperature Input System
There are two types of temperature input method and two types of Ex protection system as Ex d or Ex [ia] in
NMS5.
5.3.1
Ex d NMT Input + HART Input
Ex d HART sensor and NMT53x Ex d are connected in this system.
Figure. 42:
Ex d NMT Input and HART Input
5.3.2
Ex d Pt100 Spot Temperature Input +HART Input
Ex d HART sensor and Ex d Pt100 sensor are connected in this system.
Figure. 43:
Ex d Pt100 Spot Temperature Input and HART Input
26
4
24
25
J3
1A
1
3
5
J1
1B
Power
Module
POW
Ex d
NRF
560
NMT
53x
Ex d
Ex d
HART
TCB CPU Module
Terminal module
HART
Sensor
Ex d
HART
Sensor
Ex d
5
26
4
5
24
25
J3
1A
1
3
5
J1
1B
Spot
Temp.
Module
A
B
b
Ex d
HART
HART+
HART-
TCB CPU Module
NRF
560
Ex d
Pt 100
Power
Module
POW
Ex d
HART
Sensor
Ex d
HART
Sensor
Ex d
5 Wiring
Ex ia NMT Input + HART Input
Ex d HART sensor and NMT53x Ex ia are connected in this system.
Figure. 44:
Ex ia NMT Input and HART Input
5.3.4
Ex ia Pt100 Spot Temperature Input + HART Input
Ex d HART sensor and Ex ia Pt100 sensor are connected in this system.
Figure. 45:
Ex ia Pt100 Spot Temperature Input and HART Input
26
4
5
24
25
J3
1A
1
3
5
J1
1B
Power
Module
POW
Ex ia
NRF
560
NMT
53x
Ex ia
Ex d
HART
TCB CPU Module
Terminal Module
Separation Wall
Ex ia
HART
COM-4 Module
Ex ia
Ex ia Power Feed
Ex d Power feed
Ex d
HART
Sensor
Ex d
26
4
5
24
25
J3
1A
1
3
5
J1
1B
Spot
Temp.
Module
A
B
b
HART+
HART-
TCB-4 CPU Module
NRF
560
Ex ia
Pt 100
Power
Module
POW
Ex ia
Separation Wall
Ex ia
Ex ia
Ex ia Power Feed
Ex d Power Feed
Ex d
HART
Sensor
Ex d
HART
Sensor
Ex d
Ex d
HART
Plastic Case
Ex ia
HART
5 Wiring
Touch Control and Programming Matrix
6.1.1
Display and Operating Elements
Display
During normal operation, NMS5 has an illuminated LCD that shows the level, the temperature, and the status
of the device at "HOME" position.
For the display of the other data and the programming of the parameters for operation, NMS5 uses a convenient
programming matrix.
Operating Elements
NMS5 is operated by three visual operating elements, namely the keys "E", "+", and "-" . They are actuated
when the appropriate field on the protective glass of the front is touched with the finger ("touch control"). The
corresponding transmitting and receiving diodes are not affected by external influences, e.g. direct sunlight.
The software and hardware installed in NMS5 rule out any malfunction that may be caused in this way. Even
in explosive hazardous areas, the explosion-proof housing of the touch control ensures a safe access to the data.
Figure. 46:
Display
6.1.2
Functions of Operating Elements
The programming matrix consists of matrix groups, namely one "static" matrix and additional "dynamic" matri-
ces. They are described in detail in Section "13 Matrix". The individual matrix groups, function groups, and
functions within the programming matrix can be selected by alternately touching the operating elements.
Figure. 47:
Displays
Illuminated LCD
3 Optical Operating Elements "Touch Control"
Infrared Receiving
Diaode
Infrared Transmitting
Diaode
LEVEL
U-D
0.0°C
LEVEL
U-D
0.0°C
LEVEL Data
Temp. Data
Displacer Status Symbol
Gauge Status / Error
MORE FUNCTION
GVH:Position
Group Message
Key Guide
G:0
VH:3*
> GROUP SELECT < E
Enter Matrix
Go
Back
HOME Display
Function Group Display
Input Number
Key Guide
G:0
VH:39
ACCESS CODE
3
E
-
+
Measurement Data
G:0
VH:20
OPERATION
15955
LEVEL
Matrix Display
Matrix Display
6 Operation
Figure. 48: Matrix Construction
0
1
2
3
4
5
6
7
8
9
0
1
2
3
4
5
6
7
8
9
Measured Value 1 0
Measured Value 1 0
Measured Value 2 1
Measured Value 2 1
Operation
2
Operation
2
More Function
3 Calibration
More Function
3 Device Data
Level Data
4
Contact Output
4
Calibration
5
Analog Out
5
Adjustment
6
parts Data
6
Auto Wire Calib.
7
Input Signal
7
Auto Calib. Displ.
8
Communication
8
Display
9
Status
9
0
1
2
3
4
5
6
7
8
9
0
1
2
3
4
5
6
7
8
9
Measured Value 1 0
Measured Value 1 0
Measured Value 2 1
Measured Value 2 1
Operation
2
Operation
2
More Function
3 Service
More Function
3 Temperature
4
Temperature
Data
4
Gauge data
5
Element Temp.
5
System Data
6
Element Position
6
Service
7
NMT Adjustment
7
Sensor Value
8
Set Data NMT
8
Sensor Data
9
Device Data
9
0
1
2
3
4
5
6
7
8
9
0
1
2
3
4
5
6
7
8
9
Measured Value 1 0
Measured Value 1 0
Measured Value 2 1
Measured Value 2 1
Operation
2
Operation
2
More Function
3 HART Dev (1)
More Function
3 HART Dev (2)
Measured Value
4
Measured Value
4
P.V. Setting
5
P.V. Setting
5
Sensor Speci�c
6
Sensor Speci�c
6
Alarm
7
Alarm
7
Self Diagnostic
8
Self Diagnostic
8
Device Data
9
Device Data
9
0
1
2
3
4
5
6
7
8
9
0
1
2
3
4
5
6
7
8
9
Measured Value 1 0
Measured Value 1 0
Measured Value 2 1
Measured Value 2 1
Operation
2
Operation
2
More Function
3 Adjust Sensor
More Function
3 Tank Pro�le
Adjust Sensor
4
Pro�le Operation
4
HART Error Rate
5
Status/Data
5
Unit
6
Density 1-10
6
HART Line
7
Density 11-16
7
Interface Adjust
8
Position 1-10
8
9
Position 11-16
9
0
1
2
3
4
5
6
7
8
9
Measured Value 1 0
Static Matrix
Dynamic Matrix
Measured Value 2 1
Operation
2
More Function
3 Interface Pro�le
4
Status/Data
5
Density 1-10
6
Density 11-16
7
Position 1-10
8
Position 11-16
9
Meas. Wire &
Drum
6 Operation
LCD returns to HOME position when no key is touched for 10 minutes or more. Digits are incremented or
decremented using by + or - key. When touching + or - key continuously, then the minimum digit changes
first. After one cycle of the minimum, the second minimum changes. After one cycle of the second follows the
third minimum, and so on. When releasing a finger from the touch control, then the procedure starts again
from the minimum digit (Analogy of mechanical counter).
Figure. 49:
Selection Matrix Groups, Function Groups, and Functions with Programming Matrix
Key
Functions
• Access to the programming matrix (touch the key for 3 seconds or more).
• Return to HOME position (touch the key for 3 seconds or more).
• Move the status horizontally in the function group to select functions.
• Save parameters or access codes.
• Move the status vertically to select function groups.
• Selec or set parameters.
• Set access codes.
+ -
4
5
6
7
8
9
4
5
6
7
8
9
H
V
LCD ( HOME Position)
When touching on the matrix for 3 seconds or longer,
the screen returns to HOME position.
1 CALIBRATION
3 SERVICE
4 TEMPERATURE
5 HART DEVICE (1)
6 HART DEVICE (2)
7 ADJ. SENSOR
8 TANK PROFILE
9 INTERFACE PROFILE
0
1
2
3
4
5
6
7
8
9
0
1
2
3
4
5
6
7
8
9
4
4
4
4
7 ADJ. SENSER
3 SERVICE
4 TEMPERATURE
5 HART DEVICE(1)
6 HART DEVICE(2)
> 3 s
2 DEVICE DATA
MATRIX OF G0 V3H0
4
2 DEVICE DATA
4
1 CALIBRATION
5
6
7
8
9
5
5
5
5
5
6
6
6
6
6
7
7
7
7
7
8
8
8
8
8
9
9
9
9
9
When touching
"E" on the matrix
for 3 seconds or
longer, the screen
moves to V0
position.
8 TANK PROFILE
9 INTERFACE PROFILE
6 Operation
After turning on the power, LCD shows the current data on HOME position.
Figure. 50:
HOME Position
The letters A, B, C and D stand for the areas where information on measured values and status of the device
are displayed:
The meanings of message appearing on LCD (display) is explained in the following table.
Area
Information
A
Current Level
B
Current Temperature
C
Gauge Status
D
Displacer Status
Gauge Status
Description
G - RE
Reference position for measuring
UP
The measuring wire is hoisted.
STOP
The measuring wire stops.
LIQU
The liquid is being measured.
U - IF
The upper interface level is being measured.
LIF
The lower interface level is being measured.
BOTM
The tank bottom level is being measured.
U - DE
The upper liquid density is being measured.
M - DE
The middle liquid density is being measured.
B - DE
The bottom liquid density is being measured.
CAN
Resetting RELE.OVER TENS (over tension error).
TEAC
Calibration is being measured.
blank
Measuring is not possible.
6 Operation
If no LCD operation, NMS5 will turn off the backlight of LCD 12 hour later. Touching LCD again after this
time will turn on the backlight.
Displacer
Status
Symbol
Meaning
BAL
Balance
The displacer is resting on the liquid surface or interface and in bal-
anced status.
T - B
Temporary Balance
Automatic weight calibration is being carried out.
U - U
Unbalance Up
The displacer is being hoisted and in unbalanced status.
U - D
Unbalance Down
The displacer is being lowered and in unbalanced status.
R - U
Balance Up
The displacer is being hoisted and in correction of balance.
R - D
Balance down
The displacer is being lowered and in correction of balance.
LOW
The displacer is resting at the lower stop.
6 Operation
The access code is to ensure the confidentiality of the setup data. Three security levels are available along with
access codes.
The higher levels include the lower ones. e.g. If access code 50 is specified for a function, then code 51 also
enables editing. A function that requires access code 51, on the contrary, cannot be edited by code 50.
However, 777 is only used to change I.S. terminal configuration.
Figure. 51:
Order code which require Access Code 777
Setting Access Code
Security Level
Access Code
0
None
1
For Operator
50
2
For Engineer
51/530/777
Item
Procedure
Remarks
Static Matrix
Figure. 52:
Access Code
1. At the static matrix "MORE
FUNCTION", select
GVH=039 "ACCESS
CODE"
2. The default value is "0".
Touch the "+" key.
3. The first digit increases to 9,
then the second digit
increases. Stop touching "+"
once you reach "50".
4. "50" is blinking. Gently
touch "+" key again to
change the first digit from 0
to 1. Now you have "51".
5. Here touch "E"; "EDITING
ENABLE" will be displayed.
• When touching "E" while
displaying an access code
except 0, 50, or 51
"EDITING LOCKED" will
appear.
• If an access code has not
been selected before
performing any settings, the
screen will automatically
change to show
"ACCESS CODE"
• Operation Commands can
be sent, and displayed data
read, by remote systems,
depending on your
NMS5 specification.
NMS5-
X
X
X
X
X
X
X
X
X
X
X
J
N
N
P
F
A
J
B
C
D
E
G
MORE
FUNCTION
ACCESS CODE
039
6 Operation
Operation Command and New Operation Status
6.4.1
Operation Commands
Operation Commands can be sent the NMS5 from a host system. The following table explain the command
codes.
6.4.2
New Operation Status
The following table shows the new operation status, which is available when "NEW NMS STATUS", matrix
position GVH=272, is selected to "ENABLED".
Code
Command
Remarks
0
LEVEL
1
UP
2
STOP
After weight calibration, STOP is set as a default operation command.
3
BOTTOM LEVEL
4
UPPER INTERF. LEVEL
5
MIDD.INTERF.LEVEL
6
UPPER DENSITY
7
MIDDLE DENSITY
8
DENSITY BOTTOM
9
REPEATABILITY
10
WATER DIP
Code
Meaning
NMS Display
0
No definition
-
1
Displacer at reference position
REFERENCE
2
Displacer hoisting up
UP
3
Displacer going down
DOWN
4
Displacer stop
STOP
5
Level measurement, balanced
LEVEL
6
Upper I/F level, balanced
UPPER INTERF. LEV.
7
Middle I/F level, balanced
MIDD. INTERF. LEV.
8
Bottom meas. Balanced
BOTTOM LEVEL
9
Upper density finished
UPPER DENSITY
10
Middle density finished
MIDDLE DENSITY
11
Bottom density finished
DENSITY BOTTOM
12
Release over tension
RELE. OVER TENS.
13
Calibration activated
CAL. ACTIVE
14
Seeking level
LEVEL SEEKING
15
Following level
LEVEL FOLLOWING
16
Seeking upper density
UPP.DEN.SEEKING
17
Seeking middle density
MID.DEN.SEEKING
18
Seeking bottom density
BOT. DEN. SEEKING
19
Seeking upper I/F level
UPP. INT. SEEKING
20
Following upper I/F level
UPP. INT. FOLLOWING
6 Operation
Seeking middle I/F level
MID.INT.SEEKING
22
Following middle I/F level
MID.INT.FOLLOWING
23
Seeking bottom level
BOTTOM SEEKING
24
Not initialized
NO INITIALIZE
25
Stopped at upper pos.
UPPER STOP
26
Stopped at lower pos.
LOWER STOP
27
Repeatability testing
REPEATABILITY
28
Seeking water level
WATER SEEKING
29
Water level, balanced
WATER LEVEL
30
Following water level
WATER FOLLOWING
31
Over-/under tension, Z-phase, ADC error
EMERGENCY ERROR
32
GVH157 SERVICE MODE = ON
MAINTENACE
Code
Meaning
NMS Display
6 Operation
Operation of Displacer
The operation of the displacer for level, bottom level, interface level, and density measurement is possible by
touch control.
Item
Procedure
Remarks
Figure. 53:
Static Matrix
Figure. 54:
Touch Control
1.
Select function
groupGVH=020
OPERATION
2.
Select item
OPERATION. The
LCD shows the
command given to
NMS5and the
displacer position.
•
Set access code to 50.
•
If editing has previously
been enabled by a valid
access code, then the
request for the code will
not appear.
Figure. 55:
Operation of Displacer
3.
The following
commands are
available at this
position:
LEVEL
UP
STOP
BOTTOM LEVEL
UPPER INTERF. LEV*
MIDD.INTERF.LEV*
UPPER DENSITY*
MIDDLE DENSITY*
DENSITY BOTTOM*
WATER DIP*
REPEATABILITY TEST
• Interface and density
measurements are not
available unless
specifically requested.
• Density profile
measurement is
explained in section "7
Commissioning".
•
The optional operation
commands are available
when such options are
set.
OPERATION
OPERATION
Level
Bottom
Level
Upper Interface
Level
Middle Interface
Level
Density Bottom
Water
Dip
Middle Density
Upper Density
6 Operation
Calculation of Level and Densities
The section specifies the formula used by NMS5 to calculate levels and densities.
Figure. 56:
Calculation of Levels and Densities
Surface and Interface Levels
While the measured surface or interface level is constant, the displacer rests at the balancing position. The ten-
sion on the measuring wire is proportional to the displacer weight diminished by the buoyancy forces in both
layers:
A rise or fall of the level will increase or decrease the submerged volume. If this change exceeds the volume
tolerance set at matrix position GVH=345 "VOLUME TOLERANCE", then the
corresponding change will actuate the motor of the NMS5 till the balancing condition is fulfilled again.
Tank Bottom Level
For bottom level measurement, the balancing condition is defined as
Level
Formula
Liquid Level
T = W - Vвρu
Upper Surface
T = W - Vвρm - (V - Vв) ρu
Middle Surface
T = W - Vвρb - (V - Vв) ρm
Measuring Wire
Tension T
Gas
Density Negligible
Measured Displacer Weight W
Upper Liquid
Density ��
Measured Displacer Weight Wu
Middle Liquid
Density ��
Measured Displacer Weight Wm
Bottom Liquid
Density ��
Measured Displacer Weight Wb
Balance
Volume VB
Displacer
Volume V
6 Operation
The upper, middle, and bottom densities are calculated by the following formulas.
• Upper density (Pu) and Density profile
• Middle density (Pm)
• Bottom density (Pb)
Draft
The draft varies depending on the shape of the displacer. For cylindrical shape, the draft is
where the variables and constants have the following meanings.
Figure. 57:
Variables and Constants for Displacer
The surface or interface level should be within the cylindrical part of the displacer and approximately in the
middle of its total height.
ρu =
V
W - Wu
ρm =
V
Wu - Wm
ρu
+
ρb =
V
Wm - Wb
ρm
+
D = (V2 - V1) / A x 10 + h
= (V1 - V2 - V3) / A x 10 + h + h1
V1
V2
h
A
Displacer Balance Volume (69cm3)
Displacer Lower Hemisphere Volume
Height of Lower Cone
Cross-section of Displacer
V1
V2
h
A
Balance Volume (69 cm3)
Volume of Lower Cone (24.46 cm3)
Displacer Bottom Stub Height (20 mm)
Cross-section of Displacer (20.26 cm2)
For example: standard displacer
Ø50.8 mmCylindrical shape
D = (69 - 24.46 - 0.12) / 20.26 x 10 + 20 + 3
= 44.92
V3
Displacer Bottom Stub Volume
V3
Volume of End of Cone (0.12 cm3)
h
Height of End of Cone
1
1
d
h
V2
D
h1
V3
Liquid Level or Interface
6 Operation
All of the initial settings are required for NMS5 in most case. However, depending on NMS5 specification,
some settings may not be required. Set Access code 51 at GVH039 before applying the initial settings.
7.1.1
System Calendar/Clock (GVH193 to 197)
Calendar/Clock values are set at factory (Japan Standard Time) prior to delivery. Change the data to reflect
local time.
7.1.2
Density Values (GVH005 to 007)
Caution!
Always set GVH=005 Upper Density for actual density data, in LPG and any application where the actual den-
sity is 0.7000 g/ml or less. failure to set this data may result in level gauge malfunction.
Density values for 3 liquid phases are set to 1.000 g/ml prior to delivery. Change the data to reflect actual den-
sity values. For tanks with only one liquid phase, set Upper Density. For tanks with 2 or 3 clear phases, set
Middle and Bottom densities too. Note: minimum difference between phase
settings should be at least 0.100 g/ml. *GVH005£GVH006£GVH007
Example:
•
GVH005 Upper Density: 0.758
•
GVH006 Middle Density: 0.880
•
GVH007 Density Bottom: 1.000
7.1.3
Tank Height (GVH140)
Tank height value is set to default value at factory prior to delivery. Tank height is the reference height, usually
a gauging hatch on the tank that is used during manual level measurements. Change GVH140 to equal the
reference height.
Note!
GVH141 "Dip Point Offset" will automatically change to reflect the difference in height between the reference
height and the NMS5 reference position.
Figure. 58:
Tank Height
Tank Height
Dip Point Offset
(Difference between the
dipping hole height and
the reference line of
the calibration window)
7 Commissioning
Upper/Lower Stop (GVH161/162)
Upper and Lower Stop determine the highest and lowest point of displacer movement. These data are set to
16000mm and 0mm respectively at factory prior to delivery. Change these data to the desired actual Upper
and Lower limit values.
7.1.5
Communications Address (GVH285)
Several types of digital output communication protocols are available when ordered from Endress + Hauser.
NMS5 is delivered with communications hardware and/or software installed. Only minimal matrix parameter
setting, Access code 51, is required.
Communications protocol is displayed and selected at G2V8H6 Protocol. NMS5 has been preset at factory, it
is not necessary to change the setting.
At GVH285 select the desired device address for NMS5. Address range: 0 to 9, 00 to FF, or 1 to 247 for Mod-
bus.
Note!
FF is fixed for MIC protocol. WM550 and M/S address setting is done via dip switches on the communications
module inside NMS5. For Rackbus, termination resistors should be set on NMS5 at end of loop. Enraf BPM
address range is 00 to 99.
Refer to Sect. 7.5 for Address Setting details.
7.1.6
Proactive Safety Function (GVH157/158/159)
Safety Level Default by Output Protocol
Matrix
Setting
G1V5H7
Service Mode: default = OFF.
Select OFF if GVH158 Prosafety = ON.
Select ON only when performing maintenance on NMS5.
Caution!
Selecting ON disables the Proactive Safety function!
Access code 530
G1V5H8
Prosafety: default = ON.
Select OFF only if the Proactive Safety function will not be used.
Caution!
Failure to confirm ON disables the Proactive Safety function!
Access code 530
G1V5H9
Safety Level: default = 65000.0mm (WM550.), or = 99999.0 mm (other
protocols)
Possible data values depend on receiver specification (see table below)
Access code 530
Protocol
Data Range
Data Format
Modbus
0.0 to 99999.0mm
Float
V1
0 to 99999.9mm
ASCII / 6 digit
MDP
0 to 99999mm
BCD / 5digit
WM550
0 to 65000mm
16bit
Mark/Space
19.999m/32.699m
20bit BCD / 5 digit
Enraf BPM
0 to 99999.9mm
6 or 7 byte
Rackbus
0 to 99999.9mm
Float
HART
0 to 99999.9mm
Float
7 Commissioning
Analogue Output (GVH250 to 256)
When specified and ordered from Endress+Hauser, NMS5 is equipped with analogue output hardware
installed. Function settings may be changed as follows:
7.1.8
Contact Relay Alarm Output (GVH240 to 247)
When specified and ordered from Endress+Hauser, NMS5 is equipped with contact relay alarm output hard-
ware installed. Function settings may be changed as follows:
Function
Setting
Assign Output 1
Assign level or temperature to channel 1 output.
Adjust 4mA
Set desired value at which level or temperature outputs 4 mA.
Adjust 20mA
Set desired value at which level or temperature outputs 20 mA
Assign Output 2
Assign level or temperature to channel 2 output.
Adjust 4mA
Set desired value at which level or temperature outputs 4 mA
Adjust 20mA
Set desired value at which level or temperature outputs 20 mA.
Device at Alarm
Select type of output for alarm
Function
Setting
Select Relay
Select relay number from 1, 2, 3 or 4.
Assign Relay
Select output definition from range of choices: None, Level, Liquid
Temperature, Caution, Warning, Emergency Error, Balance Signal.
Relay Function
Select High or Low.
Switching Point
Set value at which relay is activated.
Hysteresis
Set hysteresis value for selected relay.
Relay on Alarm
Select relay of Normal Open or Normal Closed.
On Delay Time
Set time delay value for alarm output start.
Off Delay Time
Set delay value for alarm output stop.
7 Commissioning
Settings for NMT 53x Connections
The following settings are required to display NMT 539 data on NMS5 screens.
Caution!
For NMT535/539/532 connections, refer to the appropriate operating manual for the device.
Item
Procedure
Remarks
Figure. 59:
Matrix Group: Service
Static Matrix
Figure. 60:
Static Matrix
1.
In Static Matrix
"MORE
FUNCTION", invoke
GVH=030 "MATRIX
OF" and select
"SERVICE".
2.
Invoke the Dynamic
Matrix GVH=362
"CONNECTION
NMT" screen.
3.
Use the "+" and "-"
keys to display
"AVERAGE" and
press "E" to set data.
4.
Press "E" to return to
"SYSTEM DATA" and
press the "-" key to
return to "MORE
FUNCTION".
5.
Invoke Static Matrix
"GVH=030 MATRIX
OF". NMS5 matrix is
divided into matrix
groups.
Select
"TEMPERATURE"
from these groups.
6.
"EDITING
ENABLED" is
displayed on the
LCD.
7.
The average liquid
temperature is
displayed on
Dynamic Matrix
screen GVH=440.
8.
The temperature of
each contact is
displayed on
Dynamic Matrix
screens GVH=450 to
GVH=459.
•
Set access code 51.
•
NMS HART connection
(terminal 24 & 25) must
have corresponding IS
input in order to
configure NMT IS
version.
•
The Static Matrix
GVH=010 "LIQUID
TEMP." screen also
displays the average
temperature.
MORE
FUNCTION
SYSTEM
DATA
CONNECTION NMT
3 6 2
MATRIX OF
0 3 0
MORE
FUNCTION
MATRIX OF
0 3 0
7 Commissioning
Settings for Connection to NRF560
To connect NRF560, NMS5 requires the following settings.
Caution!
Turn on the power to NMS5 first.
Item
Procedure
Remarks
Figure. 61:
Matrix Group: Service
1.
In Static Matrix
"MORE
FUNCTION", invoke
GVH=030 "MATRIX
OF" and select
"SERVICE".
2.
Invoke the Dynamic
Matrix GVH=361
"CONNECTION
NRF" screen.
3.
Use the "+" and "-"
keys to select either
"CONTACT 1" or
"CONTACT 2".
•
Set access code 51.
•
CONTACT 1...
NRF 560
software version 1.6x
and earlier (those NRF
560 that indicate no
software version
correspond to
connection type 1).
•
CONTACT 2...
NRF 560
software version 1.8x
and later.
MORE
FUNCTION
SYSTEM
DATA
CONNECTION NRF
3 6 1
MATRIX OF
0 3 0
7 Commissioning
Liquid Level Calibration
7.4.1
Opened Tank
Preparation
Select LEVEL at GVH020 Operation. The displacer will descend to the liquid surface and balance.
When "BAL" is displayed, the displacer stopped moving.
When measuring the liquid level manually, use an approved method.
Set Level (GVH150) for Opening Tank
Input the value at GVH150 after measuring the dipping level.
Figure. 62:
Calibration Level (Manual Dipping Level)
When there is not liquid in the tank;
Figure. 63:
Calibration Level (Empty Tank)
Note!
When there is no liquid in the tank, level calibration is not 0mm on the measurement principle.
When 0mm level calibration is needed, refer to GVH004(BOTTOM LEVEL) or GVH142(DISPLAC. DRAFT).
Manual
Dipping
Level
Draft
Calibration Level
B
A
B=45mm
Note: Ø50mm (Dtandard Displacer)
140ml (Displacer Volume)
70ml (Balance Volume)
A= Depend on Tank Design
7 Commissioning
Level Calibration Procedure for Closed Tank
Closed tanks (e.g. LPG) cannot be hand-dipped. Follow the procedure shown below.
Figure. 64:
Level Calibration 1
1.
Check the NMS parameter
2.
Set the Level gauge at GVH020(BOTTOM)
Figure. 65:
Draft Level of Displacer
3.
Check that the displacer is balance and LCD (display) shows Balance.
4.
Calculate the displacer balance position.
• In case of Ø50 standard displacer, when balance volume(GVH343) is reference value (a half of displace
volume), the balance point is 45mm from the bottom of the tank.
Note!
Refer to Appendix "14.3 Displacer".
GVH
ITEM
005
UPPER DENSITY
340
WIRE DRUM CIRC.
342
DISPLACER WEIGHT
343
DISPLACER VOLUME
344
BALANCE VOLUME
BOTTOM plate (datum plate)
BALANCE position for displacer
7 Commissioning
Level Calibration 2
5.
Set Level Calibration to 45mm at GVH150 (SET LEVEL) in G1 MATRIX (Calibration).The bottom is
ZERO.
• If the bottom plate is not Zero (ex; X mm), adjust the level value by adding to the value.
LEVEL = X + 45(mm)
This completes the level calibration procedure.
0mm
45mm
45mm
7 Commissioning
The desired loop address for most digital protocols can be set at G2V8H5 (Address). The allowable range for
setting varies depending on the protocol installed in NMS5. The range has 0-9, 00-FF (FF is fixed for MIC pro-
tocol), or 1-247 for Modbus.
•
WM550 and Mark Space protocols addresses must be set by switches on the communications module inside
NMS5.
•
MODBUS: Termination resistors should be set on BODBUS modules (refer to Appendix "13.1 RS-485
MODBUS (COMM-5) Terminator").
•
Enraf BPM: Address range is 00-99. A-F is not available.
7.5.2
Whessoe Matic 550 (WM550) Communication Board Setting
Jumper Setting
Jumper settings for WM550 communication board
To loaded software into processor [IC1] J3 has to be open.
If software testing is required; J4 has to be open.
If software reset is required; J6 has to be short.
Polling Address Setting
Caution!
Polling addresses are set by mechanically at SW1 on WM550 communication board (not by
acccessing NMS programming matrix). Check all polling addresses before setting.
The following table shows about address setting.
Note!
For current loop setting, refer a operating manual for Whessoe 1098 or RTU 8130 operating manual.
Jumper
Function
Default Condition
J3 (Mode)
Use EPROM [IC4]->short
Short
J4 (Test)
Software testing
Short
J6 (Reset)
Reset
Open
J7 (Watch dog)
Watch dog setting
Short
Switch Position
Value
1
1
2
2
3
4
4
8
5
16
6
COMPATIBILITY
MODE
ON
1
2
3
4
5
6
Polling address = 1
ON
1
2
3
4
5
6
Polling address = 5
ON
1
2
3
4
5
6
Polling address = 9
ON
1
2
3
4
5
6
Polling address = 3
ON
1
2
3
4
5
6
Polling address = 7
Setting Example
7 Commissioning
Mark/Space(M/S) Communication Board Setting
Jumper Setting
Jumper settings for M/S communication board
Polling Address Setting
Caution!
Polling addresses are set by mechanically at SW1 (1-8) and SW2(1-2) on Mark Space communication board
(not by accessing NMS programming matrix). Check all polling addresses before setting. The following table
shows address setting.
Function Setting
Each function is set at SW2-* shown as follows.
Jumper
Function
Default condition
J3 (Mode)
Use EPROM [IC4]->short
Short
J3 (Reset)
Reset
Open
J3 (WD)
Watch dog setting
Short
Switch Position
Value
SW1-1
1
SW1-2
2
SW1-3
4
SW1-4
8
SW1-5
16
SW1-6
32
SW1-7
64
SW1-8
128
SW2-1
256
SW2-2
512
Switch
Function
Default Condition
SW2-3
ON: Data transmission under unbalanced condition
ON
SW2-4
ON: low baud rate setting
OFF
SW2-5
ON: Feet data OR 0-20m, OFF: 0-30m
OFF
SW2-6
ON: Measured data converted to feet
OFF
SW2-7
ON: Temperature data (57 bits)
ON
SW2-8
ON: Deg. F measured temperature unit, OFF: Deg. C
OFF
ON
1
2
3
4
5
6
7
8
ON
1
2
3
4
5
6
7
8
SW - 1
SW - 2
Address Setting
Function Setting
Setting Example (example above:267)
7 Commissioning
ENRAF Bi Phase Communication Board (COM-3) Setting
Jumper Setting
COM-3 Communication Board Jumper Setting
Communication Setting
Select "V1/ENRAF BPM" at the matrix position GVH=286.
Communication Setting
Polling address is set at GVH=285 "ADDRESS" of NMS5 matrix. The allowable range is from 00 to 99.
Caution!
A to F for polling address can not use.
Communication Setting
•
Switch S1: Reset switch
•
Connector J3: Communication port for debugging
Jumper
Function
Default Condition
JP1 (Mode)
ROM type setting
1-2 short: 27C4096
2-3 short: 27C1024
2-3 short
JP2 (Reset)
CPU mode setting (Fixed)
1-2 open
3-4 short
5-6 short
J3 (WD)
Baud rate setting
1-2 short: 1200 bps
1-2 open: 2400 bps
1-2 open
7 Commissioning
Setting of analog output 4-20mA is available when ordered from Endress+Hauser. NMS5 is delivered with ana-
log output hardware installed. For the following matrix parameter settings, Access code 51, is required.
Note!
If "LEVEL", confirm GVH158 (Prosafety) is "NO".
Matrix
Item
Setting
G2V5H0
Assign Output
Assign analog output (Level or Temperature) for channel 1.
G2V5H1
Adjust 4mA
Set level or temperature value for 4mA output on channel 1.
Available only when G2V5H0= "Level" or "Liquid Temp"
Level: 0 to 99999mm
Temperature: -999 to 999°C
Default: 0mm / 0°C
G2V5H2
Adjust 20mA
Set level or temperature value for 20mA output on channel 1. Avail-
able only when G2V5H0 = "Level" or "Liquid Temp."
Level: 0 to 99999mm
Temperature: -999 to 999°C
Default: 0mm / 0°C
G2V5H3
Assign Output 2
Assign analog output (Level or Temperature) for channel 2.
G2V5H4
Adjust 4mA
Set level or temperature value for 4mA output on channel 2.
Available only when G2V5H3 = "Level" or "Liquid Temp."
Level: 0 to 99999mm
Temperature: -999 to 999°C
Default: 0mm / 0°C
G2V5H5
Adjust 20mA
Set level or temperature value for 20mA output on channel 2.
Available only when G2V5H3 = "Level" or "Liquid Temp."
Level: 0 to 99999mm
Temperature: -999 to 999°C
Default: 0mm / 0°C
G2V5H6
Device at Alarm
Select type of output for alarm. Select from OFF, HOLD current
output, Maximum value or Minimum value.
Default : OFF
7 Commissioning
Contact Relay Alarm Output
The specification of alarm for output contact relay is delivered as a part of NMS5. This specification allows a
maximum of 4 output contact relay settings. Set access code 51 in case of the following matrix parameters.
Description of Error Message
Matrix
Item
Setting
G2V4H0
Select Relay
Use +/- and E keys to select to activate Contact Output Relays 1 to 4.
Default value = 1.
G2V4H1
Assign Relay
Select output definition from range of choices: None, Level, Liquid
Temp, Caution, Warning, Emergency Error, and Balance Signal.
Default value = NONE.
G2V4H2
Relay Function
Select High or Low function, available only when G2V4H1 = "Level"
or "Liquid Temp." Default value = HIGH.
G2V4H3
Switching Point
Set level (0-99999 mm) at which relay is activated, available only
when G2V4H1 = "Level" or "Liquid Temp." Default value = 0 mm.
G2V4H4
Hysterisis
Set hysterisis value (0-99999 mm) for selected relay, available only
when G2V4H1 = "Level" or "Liquid Temp." Default value = 0 mm.
G2V4H5
Relay on Alarm
Select from Normal Open or Normal Closed, available only when
G2V4H1 = "Level" or "Liquid Temp." Default value = NORMAL
OPENED.
G2V4H6
On Delay time
Set delay time (0-999 seconds) for alarm output start, available only
when G2V4H1 = "Level" or "Liquid Temp." Default value = 0
seconds.
G2V4H7
Off Delay Time
Set delay time (0-999 seconds) for alarm stop, available only when
G2V4H1 = "Level" or "Liquid Temp." Default value = 0 seconds.
Error Type
Descriptions
LEVEL
Liquid level exceeded the designated upper or lower limit.
LIQUID TEMPERATURE
Liquid temperature exceeded upper or lower limit.
CAUTION
Auto wire calibration error; Auto displacer calibration error
WARNING
Weight value exceeded the designated alarm sending limit (upper limit:
GVH=162 or lower limit: GVH=163). (e.g. local communication error, LCD
error, communication error, EEROM data error)
EMERGENCY ERROR
Z-phase no input error, ADC sensor error, communication IC error, A-phase
no input error, Driver error, power failure
7 Commissioning
I/O - 3
Alarm Output, Standard
(Order Code Position
050= 1,2,3 or 5)
4 relays with potential-free change-over contacts, freely assignable to
measured value
Hysteresis, Alarm Output
Switch points and switching hysteresis freely adjustable, residual
current fail-safe mode: minimum or maximum, selectable
Operation Output Logic
Switching Capacity, Alarm Output
•
Max. 250VAC, 2A / 62.5VA
•
Max. 220VDC, 2A / 60 W
•
FM/CSA: 30VAC, 2A/42VDC, 2A, 60W
Alarm Output,
Overspill Prevention
(Order Code Position 050=4)
2 relays with potential- free changeover contacts, assignable to Level
Hysteresis, Alarm Output
Switch points and switching hysteresis freely adjustable, residual
current fail-safe mode: minimum or maximum, selectable
Operation Output Logic
Switching Capacity
•
U max.: 200VDC/200VppAC
•
I max.: 0.5A, DC or peak AC
•
P max.: 15W
Relay Condition
Initial Configuration
(normal status)
Open
Closed
On alarm
Closed
Open
On device Error
Closed
Open
On abnormal power
supply (5% lower than,
or 5% higher than
specified power supply)
Custody
Transfer
Closed
Custody
Transfer
Open
Power outage
Hold last condition
Relay Condition
Initial Configuration
(normal status)
Closed
On alarm
Open
On device Error
Open
On abnormal power
supply (5% lower than,
or 5% higher than
specified power supply)
Open
Power outage
Open
7 Commissioning
2 photo couplers, for external input from controller (tumbler switch,
DCS, etc.)
Operation output logic
Input voltage
15VDC, active circuit (supplied by NMS5)
Input current
Approximately 5mA
Gauge Status
CTR1
CTR2
LEVEL
UP
STOP
INTERFACE
LEVEL
0 (OFF)
1 (ON)
0 (OFF)
1 (ON)
0 (OFF)
0 (OFF)
1 (ON)
1 (ON)
7 Commissioning
Density and Density Profile Measurements
Two different types of density measurement are available when ordered from Endress+Hauser.
•
Spot density measurement for up to 3 liquid phases
•
Density profile measurement
1.
Tank profile for up to 16 points for total liquid height.
2.
Tank profile for upper phase liquid only.
3.
Tank profile based on a manually entered profile parameter.
7.6.1
Spot Density Measurement
Spot measurement for up to 3 liquid phases is an option that is available when ordered from Endress+Hauser.
The following preconditions are required in order to assure safe and accurate operation.
•
NMS5 must be configured for Spot density measurement as ordered from Endress+Hauser.
For information regarding upgrading your NMS5, contact Endress+Hauser.
• Density values:
Figure. 67:
Density Measurement
– Density values can be viewed and changed at the following matrix positions:
G0V0H5 Upper Density
G0V0H6 Middle Density
G0V0H7 Density Bottom
•
The displacer is balanced at liquid level and the liquid surface is not moving.
For spot density measurement, the following parameters can be set or selected. However the factory default
settings also can be used.
Operation
Select the measurement of the density desired from UPPER DENSITY, MIDDLE DENSITY, or DENSITY BOT-
TOM at G0V2H0 "OPERATION". NMS5 raises the displacer up out of the liquid, measures its weight, and
sends the displacer down to measure the density. After density is measured, the displacer remains at that posi-
tion until UP or LEVEL command are selected at G0V2H0.
Matrix
Item
Setting
G1V4H3
Displacer Raise
Density
Set distance (0-300mm) for displacer to rise the level position above
during measuring density. Default setting = 150 mm. Observe the rule:
G1V4H3 + G1V4H4 = 300 n, where n = integer.
G1V4H4
Displacer
Submerg
Density
Set distance (0-300mm) for displacer to sink the level position above
during measuring density. Default setting = 150 mm. Observe the rule:
G1V4H3 + G1V4H4 = 300 n, where n = integer.
- Upper Density < Middle Density < Bottom Density
Middle Density - Upper Density 0.100 g/cm3
Bottom Density - Middle Density 0.100 g/cm3
e.g. 0.760 g/cm3
e.g. 0.880 g/cm3
e.g. 0.1000 g/cm3
(Example)
7 Commissioning
Density Profile Measurement
7.6.2
Density Profile Measurement
Tank Profile Density Measurement
In addition to the preconditions, the following additional preconditions are required in order to assure safe and
accurate operation for the density measurement of the tank profile.
Access Code 51 is required to edit parameters.
•
NMS5 must be configured for Spot & Profile density measurement as ordered from Endress+Hauser.
Contact Endress+Hauser for the information regarding upgrading NMS5.
•
The standard tank bottom level (GVH004) is zero. If not, the formula used for calculating the interval
between profiles is GVH000 / GVH841. The displacer touches the bottom of the tank before measuring a
lower level (e.g. 884mm).
•
If Bottom Level is not 0 mm, select Manual I/F Profile at GVH840. Set GVH843 I/F Manual Level = actual
bottom level e.g. 884mm. In this case, the intervals will be calculated based on GVH000 - GVH843/
GVH841.
•
The density profile can be operated when the liquid is still in the tank. However, depending on the height
of liquid level, it may take 1 hour or more to operate the density profile.
Note!
Additional matrix parameter settings are required for remote communications via WM550 communications
protocol.
Refer to Section "13. Matrix, (Dynamic Matrix, Device Data : G2), WM550 DENS. SEL.".
Matrix
Item
Setting
G1V5H4
Safe Density
Select the desired resultant condition when density profile measurement
fails due to displacer reaching the low-limit for density profile operation
(set in G1V5H5). Selecting "ON" will result in STOP operation. Select-
ing "IGNOR" will result in "LEVEL" operation, displacer will return to liq-
uid level. Selecting "OFF" will leave the displacer at the position where
density profile measurement failed, and in "DENSITY" operation.
G1V5H5
Den Ope Level
Set the lower limit for displacer movement during density profile
operation. Default setting = 300 mm.
Upper Layer (Gas or Liquid)
Lower Layer
(Liquid)
A
B
7 Commissioning
A. Tank profile for up to 16 points for total liquid height
G8V4H0 Ope Select: Select 1: Tank Profile.
•
B. Tank profile for upper phase liquid only
G8V4H0 Ope Select: Select 2: I/F Profile.
•
C. Tank profile for based on a manually entered profile parameter.
G8V4H0 Ope Select: Select 3: MANU.I/F Profile.
When choosing A, B, and C above, the following parameters can be set or select. The factory default settings
also can be used.
Initial Setting of Tank Profile (Density Profile) Measurement
All of necessary set up can be done at "G8 Tank Profile" matrix. There are six factors to configured prior to the
actual operation.
G8V4H0 OPE. SELECT
Matrix
Item
Setting
G8V4H1
Ope Point
Set the number of measurement points (2-16) for Tank Profile mea-
surement. Default setting = 2.
G8V4H3
I/F Manu.
Level
Manually set the water interface level (0-99999.9mm) to be
referenced during the tank profile measurement. Default setting = 0
mm. Only used with C, tank profile for based on a manually entered
profile parameter.
G8V4H4
Bal. Level
Set the allowance for level movement (1.0 -99.9 mm) prior to Tank
Profile operation. If the level movement exceeds this value, profile
operation is canceled. Setting 99.9mm allows Profile measurement to
proceed, regardless of level movement. Default setting = 2.0 mm.
G8V4H5
Up Wait Time
Set waiting time (1-31 minutes) for displacer weight to be measured in
air, when making the weight table at start of profile operation. Default
setting = 1 minute.
G8V4H6
Liq Wait Time
Set the time (1-31 minutes) for displacer to stop at each measurement
position during profile operation. Default setting = 1 minute.
G8V4H7
Ope Wait Time
Set maximum waiting time (1-31 minutes) for conditions in G8V4H4
(Bal Level) to be satisfied before profile operation begins. E.g. if the
level movement exceeds setting at G8V4H4, NMS5 is in "standby" for
the time set in this matrix. After the maximum time elapses, profile
operation is canceled. Default setting = 1 minute.
Item
Procedure
Remarks
Figure. 69:
G8V4H0
1.
Select G0V3H0
"MATRIX OF" in the
row "MORE
FUNCTION" on Static
Matrix, then select
"Tank profile".
•
Set access code to 50.
2.
Select function group
"PROFILE OPE" and
G8V4H0 "OPE
SELECT".
•
The row of PROFILE OPE
is the only function group
that needs configuration
for the density profile
operation.
3.
Select 1: "TANK
PROFILE" by pressing +
or - key.
4.
Press E to configure.
•
Selection of "0: SPOT"
enables Upper Density
measurement instead of
Density Profile
measurment.
Static
Matrix
MORE
FUNCTION Matrix of
TANK
PROFILE
PROFILE
OPE.
OPE.
SELECT
STATUS
DATA
DENSITY
1-10
DENSITY
11-16
Position
1-10
Position
10-16
Dynamic
Matrix
Function
Group
GVH=840
GVH=030
Matrix Group : TANK PROFILE
7 Commissioning
G8V4H1 OPE. POINT
G8V4H4 BAL. LEVEL
G8V4H5 UP WAIT TIME
Item
Procedure
Remarks
Figure. 70:
G8V4H1
1.
Select G8V4H1 "OPE
POINT " to determine
number of density
measuring points.
•
Set access code to 51.
Default value is 2 points
and selectable up to 16
points.
2.
Enter required number
by pressing + or - key,
and select E to enter.
•
Whatever the number of
measuring point is
selected, the measuring
interval & point between
each point are automati-
cally calculated within the
liquid level.
Item
Procedure
Remarks
Figure. 71:
G8V4H4
1.
Select G8V4H4 "BAL
LEVEL" to determine
the liquid stability range
in order to start the
operation.
•
Set access code to 51.
NMS software has a func-
tion to scan & record the
liquid level every 5 min-
utes.
2.
Enter desired value by
pressing + or - key and
select E to enter. Input
value of 99.9mm
represents that the
measuring procedure
starts regardless of level
condition.
•
NMS5 only starts the
actual measurement when
the liquid level is within
the entered value
compared to the recorded
level data 5 minutes ago.
This function ensures the
safe operating condition to
avoid filling or draining
liquid from the tank.
Item
Procedure
Remarks
Figure. 72:
G8V4H5
1.
Select G8V4H4 "UP
WAIT TIME" to
determine the displacer
weight calibration
interval in air.
•
Set access code to 51. The
displacer raises approxi-
mately 500mm above liq-
uid surface after confirm-
ing the stability of liquid
level within "BAL LEVEL"
function.
2.
Enter desired waiting
time by pressing + or -
key, and select E to
enter.
•
The displacer weight is
calibrated at 8 different
positions in air and the
interval between those
positions are adjustable
within this function to
allow excess liquid to drip
off the displacer to achieve
higher accuracy.
PROFILE
OPE.
OPE.
POINT
Function
Group
Matrix Group: TANK PROFILE
TANK
PROFILE
Function
Group
BAL.
LEVEL
GVH=844
Matrix Group : TANK PROFILE
TANK
PROFILE
Function
Group
UP.WAIT
TIME
GVH=845
Matrix Group : TANK PROFILE
7 Commissioning
G8V4H6 LIQ. WAIT TIME
G8V4H7 OPE. WAIT TIME
Item
Procedure
Remarks
Figure. 73:
G8V4H6
1.
Select G8V4H6 "LIQ.
WAIT TIME" to
determine the displacer
weight calibration
interval between each
measuring point in the
liquid.
•
Set access code to 51. A
small turbulence made by
the movement of displacer
itself may cause the liquid
surrounding displacer to be
unstable for the accurate
measurement.
2.
Enter desired waiting
time by pressing + or -
key, and select E to
enter.
•
This function allows
displacer to hold in each
calculated position before
actual measurement is
proceeded based on the
entered value.
Item
Procedure
Remarks
Figure. 74:
G8V4H7
1.
Select G8V4H7 "OPE.
WAIT TIME" to
determine the operation
holding time limit to
accept "BAL. LEVEL"
before the actual
operation starts.
•
Set access code to 51.
Entire operation begins
only when the condition
described in "BAL. LEVEL"
is achieved.
2.
Enter desired waiting
time by pressing + or -
key, and select E to
enter.
•
This function determines
how long for the NMS to
wait if "BAL. LEVEL" can
not be established before
quitting entire operation.
TANK
PROFILE
Function
Group
LIQ. WAIT
TIME
GVH=846
Matrix Group : TANK PROFILE
TANK
PROFILE
Function
Group
OPE. WAIT
TIME
GVH=847
Matrix Group : TANK PROFILE
7 Commissioning
Operation of Tank Profile (Density Profile) Measurement
After all the initial settings are completed, the actual operation can be proceeded by one simple command. The
operation status regarding this function is displayed at G8V5H0 and the liquid level condition is also displayed
at G8V5H1. Terms of status display and descriptions are listed below.
Operations Status (G8V5H0, OPE. STATUS)
Level Condition Status (G8V5H1, LEVEL CONDITION)
Status
Contents
0: Accepting
Accepting density profile command
1: Standby
Ready to execute density profile command
2: In operation
Executing density profile command
3: OPR. END
Density profile operation completed successfully
4: UN_BALANCE ERR
Fail: Required conditions for density profile operation not satisfied
5: IPR. ERR. STOP
Fail: Abnormal conditions occurred during density profile operation
Status
Contents
0: Off Level Meas.
Active operation is not LEVEL (Profile cannot start).
1: Stable
Liquid level/surface is stable (Profile can start).
2: Unstable
Liquid level/surface is unstable (Profile cannot start).
3: Ignore condition
Ignores liquid level/surface condition (profile cannot start).
Item
Procedure
Remarks
Figure. 75:
Static Matrix
1.
Select G0V2H0
"OPERATION" in the
Static Matrix and select
"UPPER DENSITY".
•
Set access code to 50.
•
The operation command
"UPPER DENSITY" is
shared between spot and
Tank Profile density
measurement depends on
G8V4H0 configuration.
2.
Press E to start the
operation.
•
The operation must start
when the NMS is in Level
measuring mode.
3.
Unless there is any
incompatible
circumstances during
the displacer
automatically returns to
the level measurement
after the last measuring
point.
•
To minimize the risk of safe
operation, the displacer
immediately returns to the
Level measurement
regardless of completing
various data process.
Static Matrix
GVH=020
OPERATION OPERATION
LEVEL
UP
STOP
BOTTOM LEVEL
MIDDLE INTER LEVEL
UPPER DENSITY
MIDDLE DENSITY
DENSITY BOTTOM
REPEATABLITY
WATER DIP
7 Commissioning
Density Profile Operation Diagram
Figure. 76:
Density Profile
Operation Procedure
1.
Set NMS5 so that it can measures the liquid level.
2.
Raises the displacer and measure each weight of 8 measuring points in the air.
3.
Measure the density at the setting position in the liquid level.
4.
Return to level measurement.
5.
Save the data in G8 matrix accordingly.
This completes the operation procedure.
Caution!
The average density and temperature data regarding density profile function (G8 Matrix) are completely differ-
ent from the data on G0 static matrix which displays the conventional upper density (spot) and temperature
(continuously scanned) data.
8 points of air weight measurement
(within 300mm circumference
of wire drum)
Nozzle
Proservo NMS5
��������
First Displacer Weight Calculation Line
(Approximately 500mm from Liquid Level)
No. 1 Measuring Point
No. 2 Measuring Point
No. 3 Measuring Point
No. 4 Measuring Point
No. 5 Measuring Point
No. 6 Measuring Point
No. 7 Measuring Point
If required density measuring points are 7, the method to calculate
each point from liquid level in NMS with ver 4.25 softwear is:
(7 points +1)
Liquid level
x measuring point number
Ex. if the liquid level is 4,000 mm
and needs to determine the position of No. 5 measuring point from the liquid surface;
(7 +1)
4,000
x 5 = 2,500 mm
������������������������������
Max.16 point
�����������������������������������������
Max.16 point
Liquid Level
7 Commissioning
NMS5 can measure interface levels via the following operations.
•
Upper Interface Level: returns the level data for the interface between top 2 liquid phases. Displacer remains
at interface level in UPPER INTERFACE operation.
•
Middle Interface Level: returns the level data for the interface between bottom 2 liquid phases. Displacer
remains at interface level in MIDDLE INTERFACE operation.
•
Water Dip: returns the level data for the interface between top 2 liquid phases. Displacer returns to liquid
level in LEVEL operation.
The following preconditions are required in order to assure safe and accurate operation.
•
Density values:
Figure. 77:
Interface Measurement
Density values can be viewed and changed at the following matrix positions:
G0V0H5 Upper Density
G0V0H6 Middle Density
G0V0H7 Density Bottom
Note!
Additional matrix parameter settings are required for remote communications via WM550 communications
protocol.
Refer to Section "13. Matrix, (Dynamic Matrix, Device Data : G2), WM550 ALM. SELECT.".
Operation:
After confirming the above pre-conditions, use the "+" or "-" keys to select the desired operation at matrix posi-
tion G0V2H0 OPERATION. When the desired operation is displayed, press the "E" key.
e.g. 0.760 g/cm3
e.g. 0.880 g/cm3
e.g. 0.1000 g/cm3
(Example)
- Upper Density < Middle Density < Bottom Density
Middle Density - Upper Density 0.100 g/cm3
Bottom Density - Middle Density 0.100 g/cm3
7 Commissioning
Sealing of NMS5
NMS5 can be sealed, and data can be made unchangeable for custody transfer purposes as follows.
Sealing of NMS5 Procedure
1.
Set OPE. DENSITY at GVH=278 as the liquid density (if the density is expected to change during
operation, set the average value.).
2.
Turn off the power supply, open NMS5.
3.
Turn the micro switch to the "on position" on the printed circuit board TCB (see figure below). NMS5
will then enter "write-protect" mode and not accept any further level, weight, or density data changes.
This completes the sealing of NMS5 procedure.
Figure. 78:
The printed circuit board TCB
2
20
9
1
6
1
13
1
1
CNB
CN5
CN7
CN6
CN14
SW1
U1
4
1
5
4
1
5
8
5
8
1
4
5
8
1
4
U20
U19
U15
U13
U10
14
1
15
28
1
16
8
9
1
7
8
U14
U9
U21
U7
U4
U11
U12
1
14
7
8
1
16
U24
U6
9
8
C16
C8
C7
U2
DC-DC Converter
C35
U3
C38
ON
OFF
CN9
CN2
CN3
TCB
CD_L423
Protect Switch
7 Commissioning
NMS5 measuring instrument requires no special maintenance.
Exterior Cleaning
When cleaning the exterior of measuring devices, always use cleaning agents which will not cause corrosion
or other damage to the housing surface seals.
Seals Replacement
NMS5 O-rings may need to be replaced periodically, and more frequently in extreme climates. The period
between changes will depend on climate severity, ambient temperatures and process.
Repairs
The Endress+Hauser repair policy is based on the fact that the measuring devices have a modular design and
that customers are able to undertake repairs themselves. Spare parts are contained in corresponding kits along
with their related replacement instructions. Endress+Hauser provides spare parts for repairs of NMS5, which
are located with their order numbers on later pages (refer to "10.6 Spare Parts"). Contact Endress+Hauser ser-
vice representatives for further assistance regarding service and spare parts.
Repairs to Ex-approved devices
When performing repairs on Ex-approved devices, note the following:
•
Repairs of Ex-approved devices may only be performed by trained personnel or by Endress+Hauser Service.
•
Comply with the prevailing standards, national Ex-area regulations, safety instructions (XA) and other
relevant rules.
•
Only use original spare parts provided by Endress+Hauser.
•
When ordering spare parts, note the device information on the nameplate. Replace parts only with parts that
have the same device information.
•
Perform repairs according to the instructions. When completing repairs, perform the specified routine test
on the device.
•
Only Endress+Hauser service representatives may convert a certified device into a different certified variant.
•
Document all repair work and conversions.
Replacement
After replacing NMS5 or electronic module, the adjustment of a new sensor is required.
8 Maintenance
Refer to the separate "Technical Information for NMS5".
NHC4HP: high pressure version
NHC4LP: low pressure version
Figure. 79:
Calibration Chamber
9.2
Power and Control Switch
Figure. 80:
Power and Control Switch NHS8
172
300
Ø150
CALIBRATION CHAMBER
Flange
Flange
8 hole
Power + Control Swith
SPECIFICATIONS
RATED VOLTAGE
AC110 / 220V
RATED CURRENT
6A / 5A
ALUMINUM ALLY CASTING
HOUSING
TERMINAL BOX
IRON CASTING
STRUCTURE
EXPLOSION PROOF
d3aG5
NO.OF CERTIFICATE T3875
CONNECTION DIAGRAM
Term.No
Power Supply
NMS
Term.No
NHS8
1
2
3
4
5
6
7
1
2
16
17
18
Handle
position
Wiring
LIFT
INST
STOP
OFF
7
6
4
3
5
2
1
OFF
STOP
INST
LIFT
Cable Entry
PF (G)1/2
PF (G) 3/4
PF (G)1
h (mm)
B
17
20
24
140
140
145
Cable entry
193
113
65
Grounding
LIFT
STOP
OFF
2 - Ø7
120
B
INST
h
Power L
Power N
COM
CRT1
CRT2
9 Accessories
Ball Valve NHV4A
NHV4A: Standard ball valve with ANSI flanges
NHV4J: Standard ball valve with JIS flanges
Figure. 81:
Ball Valve
9.4
Reducing Flange NHF4
Figure. 82:
Reducing Flange
Ball
Body cap
Body
190
153
400
Low pressure (JPI example)
High pressure (ANSI example)
Stud Bolt
Flange
Stud Bolt
Flange
9 Accessories
Diagnosis and Troubleshooting
NMS5 has a superb self-diagnosis function which monitors its operation. If an error is occurred, the corre-
sponding message is displayed on the LCD. The selected matrix position and the error message are displayed
every few seconds alternatively. Data can be accessed while the selected matrix position is displayed. The error
messages are stored in the memory of NMS5. Matrix position GVH=037 provides the diagnostic history.
10.1
Selection of Diagnostic Code and History
Item
Procedure
Remarks
Static Matrix : G0
Figure. 83:
Static Matrix
Touch Control
Figure. 84:
Touch Control
1. On "MORE FUNCTION"
, select item GVH=036
"DIAGNOSTIC CO."
2. Previous static matrix
records are sequentially
displayed in Static Matrix
screen GVH=037, start-
ing with the latest record.
Up to 100 alarm records
can be saved. If the num-
ber of records exceeds
100, it is sequentially
overwritten
starting with the oldest
record. For example,
973192238 2402 means
that an error occurred at
22:38 on March 19,
1997, when the
instrument
temperature was 24 oC
and this is the second
error since NMS5 was
installed. Display
includes the year, month,
day, hour, minute,
instrument
temperature, and an
error sequential
number, in this order.
•
Item GVH=037 only
shows error message,
calendar, and pointer,
but no label on the
LCD.
MORE
FUNCTION
DIAGNOSTIC CO
(Erroneous Mess
0 3 6
0 3 7
10 Diagnosis and Troubleshooting
Error and Status Messages
Message
Cause
Remedy
Error Code
-49.5 ºC
Indication of shorted element in NMT
temperature sensor
Check GVH=450 thru GVH=459 to see element.
358.0 ºC
Element is not enabled, or not
installed in NMT temperature sensor .
Check GVH=482 Total number of elements
359.0 ºC
Indication of open element in NMT
temperature sensor
Check GVH=450 thru GVH=459 to see element.
ADC/SENSOR ERROR
The signal from the A/D converter is
out of range.
Consult E+H Service.
107
ADJ. XXXCOUNTER
(X=A, I, Z or combina-
tion)
Error of level data check by A, I, and/
or Z phase encoder.
If the message appears frequently, then consult E+H Ser-
vice.
A PHASE NO INPUT
The input of the A phase signal from
the encoder is not available.
Replace the detector unit. Consult E+H Service.
122
BELOWBOT. POINT
Indicates NMS displacer position is
below lowest of NMT temperature
sensor.
DEVICE ERROR: DEV
1 (or 2)
The HART device 1 (or 2) gives an
error signal.
Check the HART device 1 (or 2) .
130, 131
DEVICE ERROR: NMT
NMT 53x gives an error signal.
Check the connection of NMT 53x. Check the register of
the temeperature sensor at position GVH=362 "CONNEC-
TION NMT."
250
DEVICE ERROR: NMT
NMT 53x temperature sensor is not
available.
Check NMT 53x. Error messages are available in NMT
53x manual.
233
DEVICE ERROR: NRF
NRF 560 gives an error signal.
Check NRF 560. Error messages are available in NRF 560
manual.
240
DISPL CALIB ERROR
The deviation of automatic weight
calibration exceeds its set limit.
Check build-up or deposit on the displacer.
120
ELEM 0 RANGE OVER
Indicates the reference resistance ele-
ment (ºC) on NMT electronics is out-
of- tolerance.
IMPOSSIBILITY
Operation is impossible because there
is no weight table.
LOCAL ERROR: DEV1
(or 2)
NMS5 cannot access the local HART
device 1 (or 2).
Check the connection of the HART device to NMS5.
Check the registration of the device(s) in matrix group
G5/6.
232
LOCAL ERROR: NMT
The signal from NMT 53x tempera-
ture sensor is not available.
Check the connection of NMT 53x. Check the register of
the temperature sensor at position GVH=362 "CONNEC-
TION NMT."
111
LOCAL ERROR: NRF
NMS5 cannot access NRF 560.
Check the connection of NRF 560 and register of the tank
side monitor at position GVH=361 "CONNECTION NRF."
113
MINTENANCE
GVH157 Service Mode = ON
Set GVH157 = OFF
199
MEM. ERROR
Defect in the memory that is specially
used for custody transfer sealing.
Replace the CPU board. Consult E+H Service.
201
MPU RESET
Power to the CPU module fell below
minimum value.
MPU: START ACT*
Power to NMS5 was turned OFF/
ON.
If the message occurs frequently, then consult E+H
Service.
10 Diagnosis and Troubleshooting
These error histories are available (Erroneous Message) at position GVH=037.
OVERTENSION
The tension on the measuring wire
exceeds the upper limit set at
position GVH=162 "OVER TENS.
SET."
Check if the displacer motion is blocked by clogging or
sticking.
To release overtensioning, access position GVH=371
"RELE. OVER TENS."
101
POWER FAILURE
The supply voltage falls below the
allowed value.
Check the power source.
124
RAM FAILURE
CPU RAM failure
ROM ERROR
EEPROM
Consult E+H Service.
132
SIFA ERROR
Error between communication board
and CPU board for digital output.
Replace the CPU board. Consult E+H service.
114
TEMP BELOW
RANGE
Indicates the measured temperature
data is below the fixed range of the
NMT temperature sensor.
TEMP COM OPEN
Indicates an open common line in the
NMT temperature sensor
TEMP COM SHORT
Indicates a shorted common line in
the NMT temperature sensor
TEMP OVER RANGE
Indicates the measured temperature
data is highter than the fixed range of
the NMT temperature sensor.
UNDERTENSION
The tension on the measuring wire
falls below the lower limit set at
position GVH=163 "UNDER TENS.
SET."
Check if the measuring wire is cut or the displacer is lost.
In this case, check the installation of the NMS5.
102
WIRE CALIB ERROR
The deviation of automatic wire
calibration exceeds its set limit.
Check wire and wire drum.
115
Z PHASE NO INPUT
The input of the Z phase signal from
the encoder is not available.
Replace the detector unit. Consult E+H
Service.
106, 112
MPU:XXXX*
(XXXX=text)
CPU error
The error might happen occasionally and might be regis-
tered in GVH=037 (Erroneous Message).
However, it is normally negligible. If it occurs frequently,
then consult E+H Service.
OPE.CODE ERROR
An illegal operation command is
accessed.
If the message occurs frequently, then consult E+H Ser-
vice.
LCD CHECK
Error between LCD (touch control)
unit and CPU boaed for digital
output.
Replace the touch control.
121
GAUGE TEMP.
The temperature inside the gauge
exceeds the limit.
Check if the ambient temperature stays within the limit.
If the application is a high temperature tank, then take
meadures to avoid heat tranfer from the tank to NMS5.
DRIVER ERROR
A motor driver error has occurred
Consult E+H Service
Message
Cause
Remedy
Error Code
10 Diagnosis and Troubleshooting
Flowcharts for Diagnostics and Troubleshooting
For up to 10 elements,
check G4V5H0 thru H9 to
�nd shorted element.
For more than 10
elements, select element at
G4V7H0, view
temperature at G4V7H3.
(Level display)
-49.5 C
Wire ring at normal
position ?
May require complete NMT
exchange. Send all
Temperature matrix (G4)
data to Endress+Hauser
359.0 C
TEMP. NO. n
ADJ(A, I, Z) COUNTER
YYMMDDhhmm TT ##
Raise displacer to red line
in calibration window, or
into maintenance
chamber.
G3V5H0 = G3V5H1 = 0?
3V5H0 = G3V5H1 = 0?
Displacer stopped at
Upper Stop, G1V6H0?
Meaning:
Shorted element in NMT
temperature
Meaning:
Element not enabled, or not
installed in NMT
temperature sensor
This is an indicator, not an
error code. Verify total
number of elements at
G4V8H2.
Diagnostic : -49.5 C
Diagnostic : 358.0 C (at G4V5_or G4V7H3)
Diagnostic : 359.5 C
(Level display)
359.5C
Meaning:
Open element in NMT
temperature sensor
For or up to 10 elements,
check G4V5H0 thru H9 to
�nd open element.
For more than 10 elements,
select element at G4V7H0,
view temperature at
G4V7H3.
May require complete
NMT exchange. Send all
Temperature matrix (G4)
data to Endress+Hauser
Diagnostic : ADC SENSOR ERROR
ADC SENSOR ERROR
YYMMDDhhmm TT ##
Meaning:
The A/D signal output from
the detector unit, or input to
CPU, is abnormal.
Exchange detector unit,
adjust Hall sensors,
recalibrate.
Make connections �rm.
TCB CN5, CN7 loose?
Diagnostic : ADJ A, I or Z COUNTER
Automatic software
adjustment working
normally. Sometimes seen
in turbulent applications.
Noise interface to internal
software counters.
Eliminate noise, adjust Hall
sensors, recalibrate.
Raise displacer manually
(G3V7H6) until red line in
window.
Diagnostic : A PHASE NO INPUT
A PHASE NO INPUT
YYMMDDhhmm TT ##
Meaning:
The A-phase signal output
from the encoder, or input
to CPU. is abnormal.
20 outputs = 360 degrees.
Exchange detector unit,
adjust Hall sensors,
recalibrate.
Make connections �rm.
TCB CN5, CN7 loose?
OK?
Meaning:
Automatic software
adjustment working
normally. Sometimes seen in
turbulent applications.
Yes
No
Yes
No
Yes
No
Yes
No
No
Yes
No
Yes
10 Diagnosis and Troubleshooting
Adjust G4V8H6 from
0 -500 mm.
This will create an offset for
temperature elements.
(FLASHING ON ANY
DISPLAY)
Meaning:
Shorted element in NMT
temperature
Meaning:
An error has occurred at the
HART device.
Refer to HART device
Operating Manual
Diagnostic : BELOW BOT. POINT
Diagnostic : DEVICE ERROR: DEV1 (2)
Diagnostic : DEVICE ERROR : NRF
Meaning:
An error has occurred at the
NRF tank side indicator.
Refer to NRF Operating
Manual.
Diagnostic : DISPL CALIB ERROR
Increase G1V8H5, or
adjust Hall sensors and
recalibrate.
Diagnostic : ELEM 0 RANGE OVER
Meaning:
The A-phase signal output
from the encoder, or input
to CPU. is abnormal.
20 outputs = 360 degrees.
Record Temperature matrix
data, exchange temperature
electronics.
Loose connector J3 behind
NMS terminal block. Check
connector.
G4V4H7 = 0.0 C ?
Meaning:
Measured weight of the
displacer exceeds
allowable limit:
G1V8H4 >=G1V8H5
SW v 4.20 or Later ?
Set G4V9H8 = ON
Note!:
If displacer position gose
lower than (Elem 0 position
+ G4V8H6), BELOW BOT.
POINT will display.
(FLASHING ON ANY
DISPLAY)
Meaning:
An error has occurred at the
NMT temperature sensor.
Check G4V8H0 for NMT
Diagnostics codes, refer to
NMT Operating Manual.
Diagnostic : DEVICE ERROR: NMT
(FLASHING ON ANY
DISPLAY)
(FLASHING ON ANY
DISPLAY)
(FLASHING ON ANY
DISPLAY)
Buildup on displacer ?
Clean displacer, wire, wire
drum, adjust Hall sensors,
recalibrate
(FLASHING ON ANY
DISPLAY)
G4V4H7 <+/-9.9 C ?
At G4V7H1 select element
0, then at G4V4H7.
Repeat for all elements.
Diagnostic : IMPOSSIBILITY
Meaning:
Operation impossible
because a valid weight table
does not exist.
Adjust Hall sensors and
make weight table.
IMPOSSIBILITY
OPERATION
Yes
No
Yes
No
Yes
Yes
No
No
10 Diagnosis and Troubleshooting
Check HART bus wiring:
DEV to NMS cut or short ?
DEV (+)(-) to NMS (+)(-) ?
Voltage = DC16.5 - 25.5V ?
(FLASHING ON ANY
DISPLAY)
Meaning:
An error has occurred at the
HART device.
Diagnostic : LOCAL ERROR DEV1(2)
Diagnostic : LOCAL ERROR NRF
Check HART bus wiring:
NRF to NMS cut or short ?
NMS (+)(-) to NRF (+)(-) ?
NMT to NMS cut or short ?
Voltage = DC16.5 - 25.5V ?
Meaning:
An error has occurred at the
NRF tank side indicator.
G5V4H9 = ON ?
Set G5(6)V4H9 = ON
Check HART device
address with hand-held.
HART DEV1 = 4 (�xed)
HART DEV2 = 5 (�xed)
Meaning:
An error has occurred at the
NMT temperature sensor.
Check G4V8H0 for NMT
Diagnostics codes, refer to
NMT Operating Manual.
Diagnostic : LOCAL Error NMT
(FLASHING ON ANY
DISPLAY)
(FLASHING ON ANY
DISPLAY)
G3V6H1 correct ?
Make correct setting at
G3V6H1.
Diagnostic : MEM ERROR
Exchange TCB module,
adjust Hall sensors,
recalibrate.
MEM. ERROR
YYMMDDhhmm TT ##
Yes
No
Yes
No
G5V4H9 = ON ?
Refer to HART device
operating manual.
No
Check G3V6H1:
NRF >=v 1.82, CONTACT2
NRF < v 1.82, CONTACT1
OK ?
OK ?
Exchange FCB module in
NRF.
Exchange POW module in
NRF
No
No
OK ?
Exchange electronics
module in NMT.
Check HART bus wiring:
NRF to NMS cut or short ?
NMS (+)(-) to NMT (+)(-) ?
NMT to NMS cut or short ?
Voltage = DC16.5 - 25.5V ?
No
Meaning:
Power to the CPU module
fell below minimum value.
Diagnostic : MPU : RESET
OK ?
Noise, Isolate and eliminate
source of noise.
No
MEM. ERROR
YYMMDDhhmm TT ##
Frequent ?
Exchange TCB module,
adjust Hall sensors,
recalibrate.
No
10 Diagnosis and Troubleshooting
Check local power supply
if frequent.
MPU: START ACT
YYMMDDhhmm TT ##
Meaning:
Power supply has been
turned (off and) on.
Indicates normal operation.
Diagnostic : MPU: START ACT
Diagnostic : ROM ERROR
Meaning:
CPU ROM failure.
Meaning:
Tension on measuring wire
is more than value in
G1V6H3.
Set Operation = STOP
Set G3V7H1 = ON
Diagnostic : OVERTENSION
(FLASHING ON ANY
DISPLAY)
Diagnostic : TEMP. BELOW RAMGE
Check NMT speci�cation.
Check G4V5H0 thru H9.
Check temperature in tank.
Exchange TCB module,
adjust Hall sensor,
recalibrate
Check measuring wire,
displacer, stilling pipe for
cause of overtension.
No
Meaning:
Open element in NMT
temperature sensor.
Diagnostic : TEMP. COM. OPEN
OK ?
Enter correct number of
temperature elements at
G4V8H2.
G4V8H2 data correct ?
Record Temperature matrix
data, exchange temperature
electronics.
No
ROM ERROR
YYMMDDhhmm TT ##
Diagnostic : SIFA ERROR
Meaning:
CPU SIFA failure.
Exchange TCB module,
adjust Hall sensor,
recalibrate
SIFA ERROR
YYMMDDhhmm TT ##
Measuring wire cut ?
Find and �x cause of cut
wire. Replace displacer,
adjust Hall sensor,
recalibrate.
Yes
Diagnostic : POWER FAILURE
POWER FAILURE
YYMMDDhhmm TT ##
Check local power supply.
Meaning:
Supply voltage has fallen
below about AC95V.
Diagnostic : RAM FAILURE
Exchange TCB module,
adjust Hall sensors,
recalibrate.
RAM FAILURE
YYMMDDhhmm TT ##
Meaning:
CPU RAM failure.
Data below range ?
Record Temperature matrix
data, exchange temperature
electronics.
Normal operation. Error
will clear when actual
temperature is within
speci�cation.
Meaning:
Measured temperature data
for one or more elements is
below �xed range of NMT
speci�cation.
(FLASHING ON ANY
DISPLAY)
(FLASHING ON ANY
DISPLAY)
May require complete NMT
exchange. Send all
Temperature matrix (G4)
data to Endress+Hauser.
Yes
No
Yes
No
10 Diagnosis and Troubleshooting
Enter correct number of
temperature elements at
G4V8H2.
Meaning:
Shorted element in NMT
temperature sensor.
Diagnostic : TEMP. COM. SHORT
Meaning:
Measured temperature data
for or more elements is over
�xed range of NMT
speci�cation.
Check NMT speci�cation.
Check G4V5H0 thru H9.
Check temperature in
tank.
Diagnostic : TEMP. OVER RANGE
(FLASHING ON ANY
DISPLAY)
Diagnostic : UNDERTENTION
Set Operation = STOP
Set G3V7H1 = ON
Record Temperature matrix
data, exchange temperature
electronics.
No
Meaning:
Measured length of the
measuring wire exceeds
allowable limit:
G1V7H4 >= G1V7H5
Diagnostic : WIRE CALIB ERROR
Increase G1V8H5, or adjust
Hall sensors and recalibrate.
Buildup on
measuring wire ?
Clean displacer, wire, wire
drum, adjust Hall sensors,
recalibrate.
No
Data over range ?
Normal operation. Error
will clear when actual
temperature is within
speci�cation.
Yes
Measuring wire cut ?
Check measuring wire,
displacer, stilling pipe for
cause of undertension.
Find and �x cause of cut
wire. Replace displacer,
adjust Hall sensors,
recalibrate.
Meaning:
Tension on measuring wire
is more less value in
G1V6H3.
(FLASHING ON ANY
DISPLAY)
(FLASHING ON ANY
DISPLAY)
Yes
No
Yes
(FLASHING ON ANY
DISPLAY)
G4V8H2 data correct ?
Record Temperature
matrix data, exchange
temperature electronics.
OK ?
May require complete
NMT exchange. Send all
Temperature matrix (G4)
data to endress+Hauser.
No
No
Yes
Meaning:
The Z-phase signal output
from the encoder, or input
to CPU, is abnormal.
1 output = 360 degrees.
Diagnostic : Z PHASE NO INPUT
Exchange detector unit,
adjust Hall sensors,
recalibrate.
TCB CN5, CN7 loose ?
Make connections �rm.
No
Z PHASE NO INPUT
YYMMDDhhmm TT ##
Yes
OK ?
No
10 Diagnosis and Troubleshooting
After calibration G3V7H0 Measured Weight = 0.0 minus value.
Diagnostic : Calibration unsuccessful
Diagnostic : No response to OPERATION
0.0g
MEASURED WEIGHT
G3V4H0, H1, H2
correct ?
Input correct values for
circumference, wire and
displacer weight, sdjust
Hall sensors, recalibrate.
Adjust Hall sensors,
recalibrate. If G7V4H0
thru H4(any) = 0, and does
not respond to adjustment,
exchange detector unit.
No
No
Yes
Hall sensor
adjustment OK ?
G3V9H7
a = b = 65000 ?
Calibration OK ?
Exchange TCB module,
adjust Hall sensors and
recalibrate.
Forgot to pick up displacer
at step 3 of Weight Table
Recalibration. Do again.
Exchange detector unit,
adjust Hall sensors and
recalibrate.
Yes
Yes
No
No
LEVEL
OPERATION
G0V2H4 = LEVEL ?
At NRF tank side indicator
set OPERATION = LEVEL
No
No
Yes
G0V2H5 = LEVEL ?
EPROM OK in
TCB module
Operation OK ?
Exchange TCB module,
adjust Hall sensors and
recalibrate.
Inspect EPROM U2 in TCB
module. Straighten any
bent legs. EPROM should
be inserted �rm and level
in socket.
Exchange detector unit,
adjust Hall sensors and
recalibrate.
Yes
No
G0V2H0 = LEVEL, G0V2H1 0 = (not LEVEL)
(not LEVEL)
OPERATION STATUS
G3V7H0
G3V4H2 (+/- 2g)?
Adjust Hall sensors and
recalibrate.
At host CPU and /or
interface set OPERATION
= LEVEL
No
No
Yes
G0V2H0
G0V2H0
10 Diagnosis and Troubleshooting
Setting after Parts Replacement
After any part of NMS5 is replaced, and before starting NMS5 calibration, specify the data on the following
equipment.
•
Circumferential length of wire drum (indicated on face of wire drum)
•
Weight of displacer (indicated on the body of the displacer)
•
Volume of displacer (indicated on the body of the displacer)
•
Balance volume (indicated on the body of the displacer or 50% of volume)
•
Density of measured liquid (up to three phases)
•
Height of tank with NMS5 mounted on it (refer to Section "7 Commissioning").
Item
Procedure
Remarks
Matrix Group: SERVICE
1. In Static Matrix "MORE
FUNCTION" bring up
GVH=030 "MATRIX
OF" and select "SER-
VICE.
2. Select Dynamic Matrix
GVH=340 "WIRE
DRUM CIRC." Check
whether the displayed
value is equal to the
value marked on the
wire drum. If not, adjust
the displayed value.
•
Set the access code at
51.
Figure. 85:
Dynamic Matrix
Note!
The weight and volume of the displacer are
marked on its bottom.
Figure. 86:
Displacer
•
The balance volume is the volume of that part
of the displacer that is immersed in the liquid
when the displacer is balanced in the liquid.
1. Select the Dynamic
Matrix screen
GVH=341.
2. Set Dynamic Matrix
GVH=342 "DIS-
PLACER WEIGHT" at
the value marked on the
displacer.
3. Set Dynamic Matrix
GVH=343 "DIS-
PLACER VOLUME" to
the value marked on the
displacer.
4. Set Dynamic Matrix
GVH=344 "BALANCE
VOLUME" at half the
value set in "DIS-
PLACER VALUME."
"This setting is provided
to the approximate posi-
tion of the displacer
when it becomes
stationary in the liquid.
•
For calculation methods
of the draft position.
MORE
FUNCTION
DISPLACER WEIGHT 3 4 2
MATRIX OF
0 3 0
DISPLACER VOLUME
BALANCE VOLUME
3 4 3
3 4 4
MEAS. WIRE
& DRUM
10 Diagnosis and Troubleshooting
Maintenance Prediction Function
The history of maintenance is displayed at matrix screen GVH=265 "(Parts Overused Date)." LCD screen dis-
plays the following contents.
• Total operation time for parts managed according to this value
• Total number of rotations of the wire drum for parts managed according to this value
Reference Part
Management Value
Display
Part
Reference
(conducting time/
number of rotations)
1) POWER UNIT
2) DISPLAY UNIT
3) MOTOR UNIT
4) WIRE UNIT
5) BEARINGS UNIT
6) SHAFT UNIT
Power supply unit
LCD indicator
Motor/driver unit
Measuring wire
Drum bearing metal
Drum shaft
43,800 hours (about 5 years)
61,300 hours (about 7 years)
43,800 hours (about 5 years)
240,000 rotations
145,000 rotations
240,000 rotations
10 Diagnosis and Troubleshooting
Spare parts are contained in kits. Spare parts for NMS5 which can be ordered from Endress+Hauser are shown
with their order numbers in the diagram below. Contact Endress+Hauser service representatives for further
assistance.
10.6.1
NMS5
Figure. 87: Spare parts
T-AMP
Zone 0 NMS5-2/5-5/5-6
NMS5-2/5-5/5-6
NMS5-1/5-4
Accessories
11
��
11
��
��
11
21
21
��
21
21
��
��
31
32
33
34
35
36
37
37
37
��
42
43
44
51
52
52
52
52
52
53
53
53
��
cable between terminal and module
��
NMS5
�����
���Wire Hook
Welding Part
Bottom of Tank
10 Diagnosis and Troubleshooting
Order Code List
Aluminum Drum Housing
Drum Housing
Calibration Window
Order Code
Parts Name
Order Code
Parts Name
11 Oring
70106039
Upgrade kit HART active output
017803-0034
O-ring, cover drum housing Alu, NBR
56004534
Communication module WM550, arrestor
70106019
O-ring, cover drum housing Alu, CR
52013662
Upgrade kit WM550
70106020
O-ring, cover drum housing Alu, FKM
56004409
Communication module Mark Space
70106021
O-ring, cover drum housing Alu, Silicon
52013658
Upgrade kit Mark Space
56004345
O-ring, cover drum housing Alu, PTFE
56004355
Communication module ENRAF BPM, COM-3
017803-0066
O-ring, cover drum housing, SS, FKM
52013649
Upgrade kit Enraf BPM
017803-0035
O-ring, cover drum housing, SS, silicone
34 Electronic Module, Contact Output
70109105
O-ring, cover drum housing, SS, NBR
017800-0007
Alarm output, 2x contact, I/O 3
70109106
O-ring, cover drum housing, D158-d3.5 CR
70109108
Alarm output, 2x contact, I/O 3, TUV OSP
70109107
O-ring, cover drum housing, D158-d3.5 PFA
52013654
Upgrade kit I/O-3
017803-0033
O-ring, calibration window, NBR
35 Electronic Module, A/O
70106018
O-ring, calibration window, CR
017800-0017
Output 4-20mA, I/O 5, TCB4
017803-0041
O-ring, calibration window, silicone
52013656
Upgrade kit I/O-5
017803-0036
O-ring, calibration window, FKM
36 Electronic Module, Ex [ia] HART Communication
56004518
O-ring, calibration window, PTFE
56004356
Communication module EXI HART, COM-4
017803-0030
O-ring, display cover, NBR
37 Electronic Module, I/O Spot Temperature Input
017803-0040
O-ring, terminal cover, NBR
56004442
Oper. I./C.O., spot temperature
71070894
PTFE packin+ring, calibr. window, Zone0
70106035
Upgrade kit oper. I./C.O., Ex d temp.
21 Cover
70106032
Upgrade kit oper. input/contact out
017800-0111
Cover, display module, Aluminium
56004441
Operation input/contact out, carrier
017800-0112
Cover, terminal box, Aluminum, NMS
56004491
Spot temp. with Chassis
017800-0113
Cover, drum housing, Alu, NMS531/NMS534
56004490
Spot temp. input T-AMP UP grade KIT
56004307
Cover, drum housing, SS, NMS535/536
70106027
Spot temp. Ex ia, carrier
56004308
Cover, drum housing, SS, NMS532
40 Fuse
56004295
Calibration window, Alu, NMS531/NMS534
70106027
Fuse 250VAC T2A50, NMS, 10 pcs.
56004296
Calibration window, SS, NMS532/NMS535
70106499
Fuse 20-62VDC T3A15, NMS, 10 pcs.
56004297
Calibration window, SS, NMS536
70106500
Fuse cover 10set, NMS
71070909
Calibration window, SS, Zone 0, NMS532/535/536
41 Connector, Cable
31 Electronic Module, Power Module
52013660
Wiring harness,Ter.-COM-1/RS485/BPM mod.
70103940
POW-6, HV, non Ex i
52013650
Wiring harness, Term.-HART passive mod.
70103942
POW-6, HV, Ex i
70106037
Wiring harness, Term.-HART active mod.
70103941
POW-6, LV, non Ex i
52013657
Wiring harness, Terminal-Mark Space mod.
70103943
POW-6, LV, Ex i
52013648
Wiring harness, Terminal-ENRAF BPM mod.
32 Electronic Module, Main CPU
52013653
Wiring harness, Terminal-I/O-3 module
70103937
TCB-6, CPU module, upgrade of TCB-4
52013652
Wiring harness I/O-3-TCB
33 Electronic Module, Remote Communication
52013655
Wiring harness, Terminal-I/O-5 module.
70103937
COM-1, Communication module, V1
70106036
Wiring harness, Terminal-COM-4 module
52013661
Upgrade kit V1
70106033
CPU, T-AMP Ex d module cable
71086763
Modbus module RS485, COM-5
70106034
Wiring harness,Terminal-T-AMP Ex d mod.
70106270
Upgrade kit Modbus RS485
70105996
Wiring harness,Terminal-T-AMP Ex ia mod.
56004484
COM-6 HART, passive remote commu. module
70105995
CPU, T-AMP Ex ia module cable
52013651
42 Terminal
56004485
56004508
��������������������������
56004509
�������������������������������
10 Diagnosis and Troubleshooting
PCB CPU Carrier set complete, spacer
56004257
5x measuring wire, 28m, d=0.15mm, 316L
017800-0021
Display module 2line, 3key optical
5x measuring wire, 16m, d=0.2mm, Alloy C
70103938
Display module 4line, 3key optical
56004914
Measuring wire, 0.2mm, AlloyC, 16m
56004256
5x measuring wire, 16m, d=0.4mm, PTFE
017800-0043
Detector unit NMS, TCC-3
56004912
Measuring wire, 0.4mm, PTFE coated, 16m
70103944
Detector unit NMS, TCC-3, V1
017800-0241
Wire ring, 316
017800-0242
Wire ring, PTFE coated
56004523
Wire drum + wire Mes., 28m
017800-0243
017800-0252Wire ring, AlloyC
56004524
�������������������������������
017800-0252
Displacer, 50mm, cylindrical, SUS316
56004520
Wire drum, wire Alloy, 16m
56004339
Displacer, 50mm, cylindrical, guide ring
71114404
Wire drum, 28m wire, CR
017800-0253
Displacer, 50mm, conical, PTFE solid
71114405
Wire drum, 28m wire, silicone
70105994
Displacer, 50mm, cylindrical, PTFE solid
71114406
������������������������������
017800-0271
Displacer, 50mm, cylindrical, AlloyC
71114407
������������������������������������
Displacer, 70mm, conical, SUS316, NMi
71114408
Wire drum, 16m AlloyC wire, CR
017800-0267
Displacer, 110mm, conical, SUS316, low
71114409
Wire drum, 16m AlloyC wire, silicone
70104684
Displacer, 110mm, conical, 316, high
71133077
Wire drum + 36m wire, SUS316L
71083820
Displacer, 50mm, cylindrical, 316, Zone0
71133079
Wire drum + 36m wire SUS316L, CR
71134069
Displacer, 30mm, cylindrical, 316, Zone0
71133078
Wire drum + 36m wire SUS316L, Silicone
71134070
Displacer, 40mm, cylindrical, 316, Zone0
71134099
Wire drum + 47m wire.
71134071
Displacer, ring, 50mm, cylind.,316, Zone0
71134100
Wire drum + 47m wire, CR.
71134067
Displacer, ring, 70mm, cylind.,316, Zone0
71134101
Wire drum + 47m wire, Silicone
Displacer, ring, 110mm, cylind.,316, Zone0
56004525
Wire drum bracket, Alu
Displacer, 50mm, cylind., AlloyC, Zone0
56004527
Wire drum bracket, SS
56004526
����������������������������������
71067825
Calibration weight 50gr, NMS5
017800-0221
Bearing, wire drum bracket, PTFE
71134095
Wire hook, SUS, L=100
Wire hook, SS400, L=100
71114403
Measuring wire 0.15mm 28m SUS316L, ring SUS316
44 Display Module
51 Detector, Servo Unit
52 Wire Drum
53 Measuring Wire, Displacer
56004255
017800-0263
71134068
71134072
75 Accessory
71134096
10 Diagnosis and Troubleshooting
The following procedure must be performed before returning NMS5 transmitter to Endress+Hauser e.g. for
repair or calibration.
•
Remove all residue. Pay special attention to the gasket grooves and crevices where fluid may be present.
This is especially important if the fluid is corrosive, poisonous, carcinogenic, radioactive, or otherwise
hazardous.
•
Always enclose a duly completed "Declaration of Hazardous Material and De-contamination" form (a copy
of the “Declaration of Hazardous Material and De-contamination” is included at the end of this operating
manual). Only then can Endress+Hauser transport, examine, and repair a returned device.
•
Enclose special handling instructions if necessary, for example a safety data sheet as per EN 91/155/EEC.
Additionally specify:
•
An exact description of the application
•
The chemical and physical characteristics of the instrument
•
A short description of the error that occurred (specify the error code where possible)
•
Operating time of the device
Caution!
•
Hazardous materials may be attached to damaged parts of NMS5 or its plastic material. Unless hazardous
materials are completely removed from NMS5, no repair request is accepted.
•
Incomplete cleaning of the instrument may result in waste disposal or cause harm to personnel (burns, etc.).
Any costs arising from this will be charged to the operator of the instrument.
10.8
Disposal
In case of disposal, separate the various components according to their materials.
10 Diagnosis and Troubleshooting
10.9 Gauge Software
10.10 Contact Addresses of Endress+Hauser
The addresses of Endress+Hauser are given on the back cover of this operating manual. If you have any
questions, do not hesitate to contact E+H representatives.
Software version / Date
Software changes
Documentation changes
V2.13, from 9.96
Original Release
BA001N/08/en/09.98
V2.20, from 10.97
T2:TCB-2, Modified V1: R&S command
BA001N/08/en/11.97
V4.06, from 04.98
HART master, T4:TCB-4
V4.06, from 04.98
HART master T&W, T4:TCB-4
V4.08X, from 08.98
upper stop speed reduction 3-step
V4.20, from 08.98
WM550, Commuwin II display, elem. Error fix
BA001N/08/en/12.99
V4.20, from 09.98
T&W
BA001N/08/en/12.99
V4.22, from 05.00
HART line selection add to static matrix: gas temp.
minus data remote communication level 6 digit data pro-
cessing memory clear volume calculation matrix deleted
V4.23B, from 09.00
standard release, alarm output modify, NMT level
selection
V4.24, from 04.01
standard release
BA001N/08/en/03.01
V4.24 OSP, from 06.01
Over Spill protection
V4.24T&W, from 07.01
T&W spec.
V4.24 T2, from 02.02
TCB-2 CPU, NMT, V1 level measurement
V4.25, from 09.02
Density profile function
BA001N/08/en/02.02
V4.27, from 09.04
Level hold matrix, Error display on Home screen only
BA001N/08/en/11.04
V4.27 T6/T&O, from11.05
W&M and Overspill prevention
BA001N/08/en/03.06
V4.27 A, from 05.07
standard release
V4.27 B, from 09.08
standard release
BA1001N/08/en/12.08
V4.27 C, from 04.09
Proactive Safety function
BA1001N/08/en/04.09
V4.27 E, from 03.10
DTM support
BA00401G/08/en/01.11
V4.27 F, from 05.11
SIL
BA00401G/08/en/02.11
V4.27 G, from 05.12
4-20mA long range support
BA00401G/08/en/03.13
10 Diagnosis and Troubleshooting
Displacer and Measuring Wire
11.1
Shape, Diameter, and Material
11.1.1
Displacer
There are several types of displacer available for NMS5.
•
The standard type has cylindrical shape and a diameter of 50 mm. Diameters from 30 to 50 mm are optional.
•
Cylindrical shape is used for viscous liquids. It is also effective if the stilling well is not smooth on its interior
surface.
Displacer weight and volume depend on the application. Thin displacers are suited for level measurement,
thicker ones for bottom level, interface level, and density measurement.
A counterweight is optional for heavy turbulence (inquire).
Displacers of three different materials are provided.
•
The standard material is stainless steel SUS316.
•
Alloy C and PFA are optional for corrosive liquids.
•
Solid PFA, however, is not applied for flammable liquids.
The following size of displacer will be supplied, when you order custody transfer approval.
Note!
NMi (Netherlands).................70mm
PTB (Germany).......................110mm
11.1.2
Measuring Wire
•
The standard material of the measuring wire is stainless steel SUS316L.
•
Alloy C and PFA coated stainless steel SUS316 are for corrosive liquids.
The following specification of measuring will be supplied when ordering the custody transfer approval.
Note!
•
Select SUS316L (0.15mm) for NMi and PTB specifications.
•
Refer to the separate "Techenical Information TI00452G" for displacer selection information and Appendix
"14.3 Displacer" in this chapter.
11 Displacer and Measuring Wire
Alarm History Display
Item
Procedure
Remarks
Matrix Group: DEVICE DATA
Figure. 88:
Device Data
Touch Control
Figure. 89:
LCD (Display)
1. Bring up Static Matrix
GVH=034 *ALARM
CONTACT."
Static Matrix
Figure. 90:
Static Matrix
2. Previous static matrix
records are sequentially
displayed in Static
Matrix screen
GVH=035, starting with
the latest record. Up to
100 alarm records can be
saved. If the number of
records exceeds 100, it is
sequentially
overwritten starting with
the oldest record. The
display includes the year,
month, day, hour, min-
ute, instrument
temperature, and an
error sequential
number, in this order.
For example, 97
3192238 2402 means
that an error occurred at
22:38 on March 19,
1997, when the
instrument temperature
was 24 oC and this is the
second error since
NMS5 was installed.
MORE
FUNCTION
ALARM CONTACT
0 3 4
MORE
FUNCTION
(Alarm Message)
0 3 5
11 Displacer and Measuring Wire
List of Alarm Messages
11.4
Hall Sensors
After exchanging CPU module, detector unit, wire drum, measuring wire (all or part), it is necessary to adjust
the hall sensors and make new weight table calibration.
1.
Move the displacer to the side of calibration window or the maintenance chamber window (top of the
tank) or put the displacer on a work bench so that the displacer can be touched.
Figure. 91:
Hall Sensor Adjustment 1
2.
Confirm that NMS5 is level and no vibration or other disturbance occurred.
• Access Code 530
• Set Operation = STOP
• Set GVH030 = Adj. Sensor
3.
Set A Span and Record Sensor Count (Wa2) at GVH74.
Figure. 92:
Hall Sensor 2
MESSAGE
Cause of Alarm
UPPER LIMIT LEVEL
The level has risen above the set alarm operation value.
LOWER LIMIT LEVEL
The level has fallen below the set alarm operation value.
UPPER LIMIT TEMP.
The temperature has risen above the set alarm operation value.
LOWER LIMIT TEMP.
The temperature has fallen below the set alarm operation value.
530
ACCESS CODE
STOP
OPERATION
ADJ. SENSOR
MATRIX OF
29XX 25000
ADJ. A SPAN
index
count
sensor
count
(Ex. Wa2= 25000)
29XX 15000
ADJ. A SPAN
(Ex. Wa1= 15000)
11 Displacer and Measuring Wire
Do not let wire fall off the wire drum.
4.
Lift the displacer until Sensor Count is stable and record the value of Sensor Count (Wa1).
• |Wa2 - Wa1 | = 10000 (+/-100)?
5.
If Yes, go to step 7 "GVH740 A Zero".
6.
If No, return the displacer to the original position at GVH741
• If |Wa2 - Wa1 |<10,000, increase Index Count.
• If |Wa2 - Wa1 |>10,000, decrease Index Count.
• Repeat until |Wa2 - Wa1 | = 10000 (+/-100).
7.
Adjust A Zero Sensor Count = 21000 (+/-100) at GVH740.
Figure. 93:
Hall Sensor Adjustment 3
8.
Record the value of Sensor Counter (Wa2)
9.
Lift the displacer until Sensor Count is stable and record the value of Sensor Count (Wa2).
• |Wa1 - Wa2 | = 10000 (+/-100)?
10.
If yes, go to step 12. "GVH743 B Span".
11.
If no, repeat Steps 3and 7.
12.
Set B Span at GVH743 and repeat steps above.
Figure. 94: Hall Sensor Adjustment 4
13.
Adjust B Zero Sensor Count = 11000 (+/- 100) at GVH742.
Figure. 95:
Hall Sensor Adjustment 5
This completes the hall sensor adjustment.
29XX 21000
ADJ. A ZERO
(Ex Wa2=2100)
(Ex Wa1=1100)
29XX 11000
ADJ. A ZERO
31XX 11000
ADJ. B SPAN
31XX 21000
ADJ. B SPAN
(Ex Wa1=1100)
31XX 21000
ADJ. B SPAN
11 Displacer and Measuring Wire