KAL Page 2 KAL 07/04 1. Contents 1. Contents........................................................................................................2 2. Note ..............................................................................................................3 3. Instrument Inspection.................................................................. ...

Please read and take note of these operating instructions before unpacking and

setting the unit for operation, and follow the instructions precisely as described herein. The devices are only to be used, maintained and serviced by persons familiar with these operating instructions and with the prevailing regulation applying to pro ...

4. Regulated Use

The KAL is to be installed only in the specified applications. Every usage which exceeds the specifications is considered to be non-specified. Any damages resulting therefrom are not the responsibility of the manufacturer. The user assumes all risk for such usage. The application specifications include the installation, ...

Attention! For the measuring ranges gives the flow speed was

converted to the nominal tube diameter. It should be noted in line with this that in the tube towards the wall section the flow speed approaches 0. Depending on nominal tube diameter, screw depth of the probe and flow profile, occasional deviations, some of them considerable, m ...

7. Electrical Connection

Warning! Ensure that the voltage values of your installation correspond to the voltage values given on the device´s specification plate. • Make sure that the electrical power supply is off. • Connect the sensor cable to terminals 6 & 8 on the setting system. • Connect the power supply (voltage supply) to t ...

9. Alignment / Balancing

Initial operation: 9.1. Fit sensor in pipe (ensure correct fitting location, see point 6 Mechanical Connection). 9.2. Connection sensor cables to electronic circuitry (see point 7 Electrical Connection). 9.3. Fill pipe with flow medium. Caution! There must be no air bubbles on the sensor. Flow speed must be at 0. ...

It is possible at any time to reset the electronic circuitry to the full range

• Flow potentiometer (1) at right-hand stop • (Only on KAL E3; Temperature potentiometer (2) to left hand- stop) • Press the SET key (5) and hold in position until the FLOW LED stops The device has now been reset into basic functioning mode. The device i ...

with more than 10 m cable between sensor and electronic circuitry, as well as

increased accuracy when measuring temperature. Temperature alignment • Caution! Disconnect from power supply. • Connect full length extention cable (>10 m) (terminals 6&8) • Connect 100 Ohm ±1% resistor (part of delivery package) in plac ...