Door Top Alarm

Manual Contents 1. Introduction 2. Components 3. Architrave detailing (single swing doors) 4. Premanufacturing Information - MUST RETURN TO HIPAC 5. Preparing the Door/Frame 6. Sign off Sheet - MUST RETURN TO HIPAC 7. Wiring Stages Connections 8. Testing/Checks 9. Changing of the Alarm delay time 10. Fail safe 11. Failsafe Alarm sounding 12. Operating Instructions 13. Trouble Shooting 14. Contact Details 15. HFM Electrical Transfer Hinge

Hipac Door Top Alarms are available in 44m and 54mm thicknesses and to a width to suit the

door (made to measure). The solution is available for Non Fire and FD30 fire door application. All components are quality checked, tested and signed off before dispatch. The importance of correct installation, handling with care and connecting in the correct sequence cannot be emphasised too strongly. 2. Components

Frame Architrave Detailing - Single Swing Doors

For reduced ligature when the Door Top Alarm is used on single swing doors (with SECURAHinge), please note the below architrave detailing / modification is required to allow for the Domed Cap on top of the hinge. If new door-set, it is recommended this is planned in at design stage. Domed Cap Modified Frame Architrave Detail Domed Cap New Frame Architrave Detail

Door Top Alarm

Ordering Form Project: Company: Checked By: Duress System: Date: Signed: Door Reference# Door Width (mm) Door Thickness (mm) Hinge Type Single Action Double Action Hinge Fitment Face Fixed Single Rebate Double Rebate Door Type Single Door Double Door Fire Rating Yes No Condition New Door Existing Door Door Handing (Refer Diagram) Figure 1 or 3 Figure 2 or 4 Please use this diagram to determine handing of doors Submit completed form to [email protected] 4. Pre Manufacturing Information (Multiple Doors)

Preparing the door

Hipac Healthcare Pty Ltd can take no responsibility for incorrect operation of Door Top Alarms that are not fitted as per directed in these instructions. If retrofitting, then removal of the door is recommended. Remove 12mm from the top of the door Bore down 69mm from above at 29mm from the hinged end. Meet this bore with another from the hinged side. This is a channel for routing the cables. Make channel down centerline on the top of the door: • 26x26mm (Not Fire Rated) • 28x28mm (FD30) Feed cable and place DTA. Screw in with fixings provided. Wire Door top alarm and test. Recess 3mm: • HINGED END: From bottom of channel 18mm and 25mm wide • FREE END: From bottom of channel 18mm and full width

Please note for the FD30 applications, the below door-set layout must be applied

8 Frame intumescent 1 Supplier: Reference: Description: Lorient Polyproducts Limited LP1504DS Type 617 2NO. PVC encased sodium silicate based intumescent strips incorporating elastometric fins, set 6 and 26 from the exposed edge of frame head. 15 x 4 12 Alarm bar intumescent Supplier: Reference: Description: Overall size (t): Intumescent Seals** Therm-A-Strip** A white phosphate-based intumescent compound with self- adhesive tape on one side lining the walls and the bottom of the alarm bar rebate in the head of the leaf. 1** 6. Door Top Alarm Wiring Test Sign Off Sheet Overleaf is the completion sign off sheet for each stage of the installation. A completed signed copy of the above must be returned to Hipac for their records. Failure to return will invalidate the warranty. NOTE - To be used in conjunction with the loaned Tested Box. NOTE 2 - Sheet(s) to be controlled by Project Manager.

Sales Order Number:

Room No. Door Top Alarm (Connect Alarm to the hinge) Electrically Modified Hinge (Connect the control cable to the hinge) DTA Wired into Electronics Box (to match the connections in the O&M manual) Mains power Staff Attack Interface Unit Magnet test on Door Name Date Name Date Name Date Name Date Name Date Name Date 7. Wiring Stages / Connections 1) Door Top Alarm 2) Electrically Modified Hinge 3) Cable fed into the ceiling 4) Electronics Box (by others - electrician or staff attack co.) 5) Mains Power (by others - electrician or staff attack co.) 6) Staff Attach Interface Unit (by others - staff attack co.) 7) Staff Attack mains computer / nurses station / fobs (by others - staff attack) NOTE - Each stage must be signed off on the ‘Sign Off Sheet’ before handing over a copy returned to Hipac. A battery powered tested unit is loaned to the Door Top Alarm installers so they can sign off stages 1-3 before other companies get involved.

1. Door Top Alarm

The 4 coloured wires from the DTA cable fed into the ceiling, should be connected into the 4 connector block on the left hand side of the electronics boxes as shown. It is extremely important the wires are connected in the correct junction box: Green LED = Green Wire Red LED = Red Wire Switch = Blue Wire Switch Common = Yellow Wire Wiring into the electronics box should be as follows: Green Red Blue Yellow Cables codes Green - Green LED Red - Red LED Blue - Trigger switch Yellow - Common Staff attack wiring Wire to normally closed (recommended) or normally open. Tamper loop Tamper loop Normally closed Alarm contacts (NC) PSU - (Black) PSU + (Red) Normally open Alarm contacts (NO) External sounder PSU- External sounder open drain 2. Power The power cable has 2 cables. The black cable is connected to the PSU (-) and the red wire is connected to the PSU (+). The PSU connectors are located on the right hand side of the box. The Power can either be from the Staff Attach System (if available)or mains power. If mains power, it MUST BE transformed down to 12V.DC.

Once power is connected, the red LED on the outside of the box should light up.

If the power is connected without the Door Top Alarm in place, the unit will cycle through the following sequence - Power LED (red) comes on and the Status LED will cycle green, amber, and red then extinguish. When the pre-set delay time has elapsed, the LED will turn red and the internal alarm will sound for approximately 5 seconds. After a total of 20 seconds, the unit will return to standby mode. 3. Staff attack The staff attack is connected into EITHER the Normally Closed OR Normally Open connectors on the right hand side of the box. Hipac recommend, where possible, the Normally Closed connections are used as this will allow a notification alarm to sound if there is a power failure or wire breakage. In the Normally Open connections re used, it may still be possible to obtain a non-audible notification if there is a failure e.g. messages appears on nurse’s station display unit. Please consult the Staff Attack supplier. NOTE - The staff attack should not be connected until the Door Top Alarm and Power stages have been signed off. (The alarm could go off as the Door Top Alarm has failsafe system built in). NOTE 2 - All the wiring should be checked and signed off at each stage on the supplied sign off sheet during installation by the various contractors. If any problems occur during the installation stages, Hipac will not attend site to carry out an investigation without proof that all the stages have been signed (or stages up to when the problem has occurred). If Hipac attend site and it is proved a stage was signed off in error, this will be chargeable at a day’s rate.

Testing / Checks

The Door Top Alarms should be tested on a regular basis. Frequency of the testing is as per your risk assessments policy. To carry out the tests, place the supplied magnet just under the LED on the Door Top Alarm. (The LED only lights up when a magnet is placed against it so it is not visible to patients). Press and release the top bar on the DTA a couple of times and the LED should be green. Press and hold the bar for 3 seconds and the LED should change to orange. When the orange light is continuously on for a set period of time (factory setting = 10 seconds), the LED will change to Red and the alarm will be registered. After approximately 20 seconds, the unit will reset itself and the red LED will disappear. For risk assessment checks you could go to the orange light state where the alarm does not go off and probably only allow 1 or 2 on a random basis to go to red stage. On the outside of the electronics box there is a matching tricolored LED labeled ‘Status’ which should also light up every time the top bar is pressed. This LED does not need a magnet to light up. 9. Changing of the Alarm delay time The factory setting is set at a 10 second delay (continuous pressure on the door top bar). Other standard options are 5, 20 or 30 seconds. Inside the electronics box, as shown on the below photo, is a red or a blue box with 2 white switches.

The Door Top Alarm includes a failsafe system, where the alarm will sound if there is a power

cut or a wire breakage. NOTE - this is only applicable if the staff attack system has been wired into the Normally Closed terminals in the Hipac electronics box. It is possible to get notification if the Normally Open terminals have to be used but this will not be an audible alarm and is dependent on your staff attack company being able to send a failure message back to say the main control panel at the nurses’ station. The hinge is supplied in 2 pieces, the main section of the hinge which takes most of the doors weight and the electrically modified section, at the top, which has the cables running through it. The reason for this is for ease of service, so if at any time there is a problem, the electrically modified piece can simply be taken out without having to take the door off. The wiring to the electronics box should be checked (and signed off at each stage on the supplied sign off sheet) with the tester box at each stage during installation by the contractor. 5 Seconds 10 Seconds 20 Seconds 30 Seconds

Failsafe Alarm sounding

If this situation occurs, the alarm will keep sounding until action is taken. To switch off, locate the bedroom referenced electronics box in the loft and disconnect the power. Either take out the fusing the closely located transformer or if the power is via the staff attack system, unscrew the power connection wires (shown previously) in the electronics box with a small Phillips screwdriver. If the alarm continues to go off this is outside of the Hipac solution and down to the programming of the staff attack system. Try disconnecting the staff attack wires (also as shown previously) and if this fails to stop the alarm, immediately contact the staff attack company to discuss and solve. Once under control, please contact Hipac’s operations department to notify the problem, explain what actions you have taken to silence and discuss / agree action plan to repair the problem. 12. Operating Instructions The Door Top Alarm will be passive the majority of the time, if there is nothing placed over the door then the alarm will not be in use. When pressure is applied to the top of the door for a preset period of time between 5 & 30 seconds this will trigger a relay inside the door top alarm. The relay inside will send a signal to our control unit situated out of sight. This will trigger the staff attack system which will inform staff of an incident in accordance with the staff attack system protocol. Once the ligature has been removed from the top of the door, the alarm will automatically be reset after 20 seconds and normal use will resume. Please Note: The Door Top Alarms should be tested on a regular basis. Frequency of the testing is as per your risk assessments policy. To carry out the tests, place the supplied magnet onto the edge of the LED on the Door Top Alarm. (The LED only lights up when a magnet is placed against it so not visible to patients.) Press and release the top once and the LED should shine green. Press and release the bar again whilst the orange light is still on, and the LED should change to Orange. Then hold the bar for a predetermined period (factory setting = 10 seconds), the LED will change to Red and the alarm will be registered. Note the alarm will sound if weight is applied to the sensor; edges 1.3kg centre 2.6kg dry towel OK but wet towel not - suggest a ‘click’ style towel rail/rails. Power, Connections and DTA working -

Green Light - Press the top bar once and move magnet around LED

Amber LED - indicates correct protocol for tamper setting Red LED - DTA Alarm signal sent to Staff Attack System After approximately 20 seconds, the unit will reset itself and the red LED will disappear. The DTA resets but the staff attack system still needs resetting as per Staff Attack Instructions

Trouble Shooting Guide

Please Note: This guide is intended to help you diagnose a fault on the door top alarm, it may not tell you how to repair it. In normal operation the Door top alarm should remain passive until the bar on the top is activated. To test the alarm is working properly, you will need a magnet Begin by closing the door and re-opening it. • Without pressing on th bar put a magnet to the LED on the leading edge of the door. The LED should remain clear and unilluminated. If at this point the LED is presenting green then the fitting of the door is wrong and the DTA is clashing with the frame causing it to activate. This can be fixed by wither channeling the DTA deeper into the door or by slightly dropping the hinge. In this case Hipac do not need to attend site. • If when the door is first opened, the magnet remains clear then quickly press and release the bar once (Keeping the magnet next to the LED). The LED should turn green. This signals that you have power to the DTA. If once pressed and released the LED doesn’t illuminate, then there is a power supply issue somewhere on the circuit. The DTA control box should be located in the ceiling or a cupboard. On the front of this is a red power LED, this should be illuminated. It this is not lit then there is an issue with your power supply to thee box. In this case Hipac should not need to attend site. If the red LED is lit up but the LED on the DTA is not illuminating then there is an issue between the control box and the DTA. Please ensure that the cables at the back of the hinge and the DTA are both connected as they should be. If they are both connected correctly and the LED on the box is lit, the Hipac may need to attend site. If the LED does illuminate but the light is flashing or no a solid display in anyway, you may need to check your power supply is running off DC and not AC. Hipac do not need to attend site. • If the LED did illuminate Green, then whilst it is still displaying green the quickly press and release it once again. The LED should momentarily change from green to orange. Once the bar is released the LED will change back to green. All the time you are doing this, the magnet should be kept next to the LED. If the LED stays orange once you have released the bar then the micro switches may be over sensitive and will need adjusting. You can test this by rolling the bar to the side and seeing if

it sticks to one side, if it does Hipac will need to visit site.

If the bar is sticking but there is no side to side movement in the bar then you will need to check the channeling the door to make sure it’s not too tight on the alarm. Hipac should not need to visit site. • Once you have seen the LED change to orange, hold the bar down for 10 seconds and the LED will change to red and the alarm will sound. If the LED changes to red but the alarm doesn’t sound, you need to speak to your staff attack providers to make sure their system is wired correctly. Hipac do not need to attend site. If the Led doesn’t cycle through the colours of green, amber, red as it should then the system may be wired incorrectly. Please consult the Hipac O&M manual to make sure it’s wired as it should be. Please Note: Once the alarm has been activated it will take approximately 30 seconds to reset. If you keep the magnet to the LED, whilst it is still in alarm is will display red. If it has gone back to clear then it has reset. 14. Contact Information Hipac Healthcare Pty Ltd 36 Long Street Goulburn NSW 2580 Telephone 02 4823 0000 Visit our website at www.hipac.com.au

15. HFM Electrical Transfer Hinge

Fitting Instructions Fixings (2x) M3 x 6mm (AUS 12mm), Countersunk, Security Torx screws in A2 Stainless steel. (5x) No.8 x 1 1/2”, Countersunk, Security Torx wood screws in A2 Stainless steel. Domed Cap (Optional extra). Note: If the hinge is to be rebated, attach the EM hinge to the DTA using the M3 x 6mm screws. Trace the profile and rebate before assembly. Instructions 1. Fit the Door top alarm to the door. 2. Connect the DTA plug and the hinge plug on the side of the hinge which has been profiled with extra holes. Ensure the plug is connected in the correct orientation with the clip, passing over the latching arm. 3. Place the electrical transfer hinge against the door edge. 4. Screw the two M3 x 6mm fixings into the DTA first to position the hinge correctly. The DTA should follow the profile of the hinge. Ensure that the cable from the EM hinge isn’t trapped between the door and the hinge. The cabling should sit within the recess in the door.

We are an Australian healthcare company that guarantees to stand

behind all our medical products, both manufactured and imported, because we’re here for the life of our products and clients. PO Box 797, 36 Long St. Goulburn NSW 2580, Australia P. +61 2 4823 0000 E. [email protected] SYDNEY | MELBOURNE | BRISBANE | ADELAIDE | PERTH hipac.com.au ABN 27 600 353 688 5. Use the hinge fixing holes as guidance to pilot drill holes using a 3.5mm drill. 6. Screw the EM hinge section to the door. 7. Then fit the full length hinge section referring to the HFM/HAL fitting instruc- tions ensure the bottom of the EM hinge and the top of the full length hinge are butted together leaving no more than a 0.5mm gap. 8. Push the EM hinge section up to within 0.5mm of the top of the frame. 9. Mark then pilot the fixing holes for the screws for both hinge sections. 10. Fix the door to the frame with the main section of the hinge, leaving the EM section detached from the frame. 11. Connect the control cable from the frame to the other side of the hinge. 12. Open the hinge against the frame, ensuring the cable is not trapped then fix the EM hinge to the frame. 13. Once all hinge sections are fitted, fit the domed cap on top of the hinge knuckle, following domed cap fitting instructions. 14. Once installed, Hipac should be contacted to carry out the commissioning to validate the warranty.