AMI Watchdog Manual
Text of AMI Watchdog Manual
Watchdog Oxygen Analyzer
5
Introduction
5
Features:
5
Oxygen sensor:
4
Sensor Warranty:
4
Instrument Warranty:
4
Installation and Operation
5
Receiving the analyzer
5
Installation.
5
-------------------------Points to consider first! ------------------------------
5
Location:
6
Safety Considerations:
6
Flashing Display:
6
AMI Analyzer Manual
2
Short-form Installation Procedure
7
Laptop set up procedure:
8
Expanded Installation Procedure
10
Analyzer Description
18
Basic description:
18
General Information
20
Basic knowledge about oxygen measurement
20
Sample tubing and components
21
Leaks
21
Exhaust
21
Calibrating the analyzer
21
Gas Connections:
23
Power connections:
24
Output connections:
24
Sample Handling:
25
Gas pressures:
26
Sensor Installation:
26
Hydrogen Sulfide:
AMI Analyzer Manual 3
Basics of trace oxygen troubleshooting
30
Specifications and Disclaimer
34
Specifications:
34
Disclaimer
35
AMI Analyzer Manual Preface 4
Preface
Thank you!
We would like to thank you for purchasing the most advanced trace oxygen analyzer available. We have
gone to great lengths to make this analyzer as simple as possible. It includes our patented cell block,
(patent numbers 5,728,289 and 6,675,629), and our patented sensor. With the optional liquid rejection
probe accessory and demister it provides a simple, low cost system for monitoring natural gas w
AMI Analyzer Manual Watchdog Oxygen Analyzer 5
Watchdog Oxygen Analyzer
Introduction
The Watchdog provides the essential elements of a complete oxygen analyzer, but omits features that are
duplicated by the flow computers with which it is intended to be used. If provides a complete sample
control system, a means of calibration and a user-configurable analog and digital output.
This manual is divided into two major sections: a quick reference section
AMI Analyzer Manual Watchdog Oxygen Analyzer 4
Oxygen sensor:
AMI manufactures its own electrochemical sensor. It measures the concentration of oxygen in a gas
stream, using an oxygen specific chemistry. It generates an output current in proportion to the amount of
oxygen present, and has zero output in the absence of oxygen, thus avoiding any requirement to zero the
analyzer. The cell is linear throughout its range. The span calibration may be p
AMI Analyzer Manual Installation and Operation 5
Installation and Operation
Receiving the analyzer
When you receive the instrument, check the package for evidence of damage and if any is found, contact
the shipper.
Do not install the sensor until the analyzer is completely installed, the gas lines are plumbed and the
electrical connections are all made; and sample or zero gas is ready to flow into it.
Installation.
This section contains important in
AMI Analyzer Manual Installation and Operation 6
Power supply – if you are going to run the unit off solar power, using a battery, you need to size both the
battery and the solar panel correctly. The analyzer draws about 30mA at 14.4 volts, but if you select the 4-
20mA output option this will go up to about 60mA.
Location:
The unit is designed to be mounted on a wall or on a 2” pipe in a general purpose, Class 1 division 2 Group
C,D area. It m