AFL VS300 Series Quick Reference Manual
Text of the manual
VS300 Series Quick Reference Guide
The VS 300 is a complete video microscope system in a handheld and directly
viewed package. The VS 300 view safe microscope removes concerns for
eye safety while inspecting optical fiber connectors. The design of this video
microscope eliminates the optical path to the eye by utilizing a miniature
camera and a state-of-the-art microdisplay that achieves unparalleled clarity
and resolution.
Magnification of the VS 300 inspection scope is equivalent to 400X (23-
degree field of view comparable to an 8” monitor viewed at a distance of 20”)
and provides 600X when images are displayed on a 12” monitor connected to
the VS 300. The VS 300 video microscope uses thread-on adapter caps and
has an energy saving automatic shutoff.
The VS 300 is supplied with a 2.5mm Universal adapter cap and neck strap.
For greatest inspection flexibility additional adapter cap styles are available
from Noyes. For a list of available adapter caps contact Noyes, or visit our web
at www.AFLtele.com/go/Noyes.
Optical ports and connector end faces must be kept free from dirt or other
contaminates to ensure accurate measurements and operation. For cleaning
connector end faces on test jumpers and in fiber frames or adapters, use
optical quality cleaning fluid such as AFL FCC2 connector cleaning fluid and
AFL CCT molded cleaning tips. For cleaning the lenses and adapter caps, use
lint-free optical cleaning wipes such as AFL FiberWipes and optical quality
cleaning fluid such as AFL FCC2 connector cleaning fluid (or IPA - reagent
grade isopropyl alcohol 99% or better).
Visit our web at www.AFLtele.com/go/Clean for more information.
www.AFLtele.com /1.800.321.5298 /1.603.528.7780
VS 300 Features
#
Feature
Description
1
Power button
Pressing this button provides operation for
approximately 100 seconds. To conserve battery
life, an automatic shutoff will occur in 100 seconds.
2
Low battery indicator
When this indicator is on, two AA batteries require
replacement.
3
Eyepiece with diopter adjust Used to adjust the display focus to the user’s eye.
4
AC adapter port
This is an interface for the AC power adapter, which
provides continuous operation.
5
NTSC video out port
This port provides connection to an external video
monitor or to a PC.
L
O
C
K
U
N
L
O
C
K
4
1
2
3
5
Eyepiece lens
Objective lens
VS 300 Features
L
O
C
K
U
N
L
O
C
K
Battery compartment door
Retaining screw
Battery compartment -
for 2 AA batteries
Tripod mount
#
Feature
Description
6
Focus knob
Used to focus fiber-end images on the camera.
7
Fiber optic inspection port
(adapter cap mount)
This port (adapter cap mount) accepts Noyes
thread-on adapter caps.
8
Adapter cap
VS 300 inspection port must be equipped with an
adapter cap. Caps for different connector styles are
available from Noyes.
9
Locking ring
Used for image centering or ferrule panning.
8
7
9
6
Prior to inspecting fiber optic connectors, you need to perform a diopter
adjustment to provide proper viewing of the display for your vision. This is
accomplished by observing the display with no fiber in the inspection port.
Note: This adjustment needs to be performed once unless your vision
changes over time.
Turn on the VS 300 inspection scope.
1.
Slowly turn the eyepiece clockwise/counterclockwise until edges of the
2.
display becomes clear, sharp lines and pixels become visible.
This adjustment provides optimum focus of the display for your vision.
Inspecting Connectors
The VS 300 offers 3 methods of inspecting fiber optic connectors:
Viewing fiber-end images on the VS 300 display by looking directly into the
•
eyepiece. This method provides the 400X magnification (23-degree field
of view comparable to an 8” monitor viewed at a distance of 20”).
Viewing fiber-end images on the external video monitor connected to the
•
VS 300. This method provides the 600X magnification on a 12” monitor
and requires a video cable and external video monitor.
Transferring fiber-end images to a PC for viewing, saving, and printing.
•
This method provides the 600X magnification on a 12” monitor and
requires a video cable and computer that is equipped with an external or
internal NTSC video digitizer.
To Inspect Connectors:
Select an adapter cap that matches the connector to be inspected.
1.
Mount the selected adapter cap on the VS 300 inspection port.
2.
Insert the cable to be inspected in the adapter cap.
3.
Turn on the VS 300 either by pressing the power button or by connecting
4.
an AC adapter.
Slowly adjust the focus knob to focus the fiber-end image on the camera.
5.
L
O
C
K
U
N
L
O
C
K
1
5
3
2
Adapter cap
Focus knob
Adapter cap mount
Connector
Centering Images/Panning Ferrule
If the fiber-end image is not centered in the display, the adapter cap and
adapter cap mount may be moved by hand to correct the problem.
Loosen the locking ring.
1.
Apply left/right or front/back pressure on the adapter cap and the fiber-
2.
end image will move accordingly on the display (motion will be reversed
when viewed on an external monitor).
When centering a fiber-end image or viewing ferrule is complete, tighten
3.
the locking ring.
Note: The VS 300 is a complex imaging system. Occasionally, an image
may not appear after the instrument is powered up. To correct this problem,
remove and reinstall the batteries.
L
O
C
K
U
N
L
O
C
K
Locking ring
Adapter cap
Focus knob