ABB 8037 Application Manual
Text of ABB 8037 Application Manual
in Water Treatment Processes
Application Guide
AG/AI–024_4
Cation Bed
Reduction of
positively
charged
dissolved
solids
Anion
Exchange
Bed
Reduction of
negatively
charged
dissolved
solids
Mixed Bed
Further reduction
of any remaining
concentrations of
both positively and
negatively
charged dissolved
solids to produce pure
water
Raw
Water
In
High Purity
Water Out
8037 Sodium Monitor
Typical Range: 0.1ppb to 10ppm
Normal sodium concentrati
Sodium Monitoring in Water Treatment Processes
AG/AI-024_4
2
Why use Sodium
Monitoring in Water
Treatment Processes?
The customer needs:
L
To detect sodium breakthrough and to ensure the process water is pure.
L
To ensure the plant operates at the maximum efficiency.
L
To allow the plant to be maintained to specified standards.
Why use a
Sodium Monitor?
L
Wide sample and ambient temperature tolerances as standard minimise the need
for additional conditioni
Sodium Monitoring in Water Treatment Processes AG/AI-024_4
3
8241
Silica
Monitor
4600 Series
Conductivity
Transmitters
Portable Rack
(can also be fixed)
8037
Sodium
Monitor
Sodium monitor in typical rack-mounted arrangement
Standard solution
bottle, mounted on
shelf level with the
top of the liquid handling
(sensor) unit
8037 Liquid
handling
(sensor) unit
8037 Transmitter
unit
Drain
Model 8037 Sodium monitor – wall-mounted
Installation
L
ABB Sodium monit
Sodium Monitoring in Water Treatment Processes AG/AI-024_4
4
Process Description
Basic Measurement/ Analysis Theory
All on-line sodium monitors are based on the use of a sodium ion selective
electrode and a reference electrode. The measurement is very similar to that of pH,
with the electrode pair also responding to hydrogen ions and changes in sample
temperature.
The hydrogen ion interference is controlled by raising the pH of the sample
(reducing the hyd
Oldends Lane, Stonehouse
Gloucestershire
GL10 3TA
UK
Tel:
+44 (0)1453 826661
Fax: +44 (0)1453 827856
ABB Inc.
125 E. County Line Road
Warminster
PA 18974
USA
Tel:
+1 215 674 6000
Fax: +1 215 674 7183
ABB has Sales & Customer Support
expertise in over 100 countries worldwide
www.abb.com
The Company’s policy is one of continuous product
improvement and the right is reserved to modify the
information contained herein without
ABB Measuring Instruments Owners Instructions:
| Device Model | Type of Document | Pages | Size | Views | Downloads | Updated | For Device |
|---|---|---|---|---|---|---|---|
C13 |
ABB Measuring Instruments Operation & user’s manual (File: PDF, 12-06-2026) |
44 | 531 | 80 | 12-06-2026 | view all | |
ACF-NT |
Guide: ABB ACF-NT (1U4G8A, Upd.25-05-2026) |
28 | 817 | 131 | 25-05-2026 | view all | |
AC211 |
AC211 Installation manual - 1H6UB1 |
2 | 1430 | 244 | 25-08-2026 | view all | |
Endura AZ20 series |