LabJack U12 Operation & User’s Manual
Text of the manual
Published on LabJack (http://labjack.com)
U12 User's Guide
The complete manual for the LabJack U12. To make a PDF of the whole manual, click "Export All" towards the upper-right of this
page. If you are looking at a PDF or hardcopy, realize that the original is an online document at
http://labjack.com/support/u12/users-guide.
User's Guide
1 - Installation
Th
Start => Programs => LabJack => LJtest.
Make sure “Test Fixture Installed” and “Continuous” are not selected, and press the “Run” button. LJtest will step through 8
separate tests and all should pass.
2 - Hardware Description
The external features of the LabJack U12 are:
USB connector,
DB25 digital I/O connector,
Status LED,
30 screw terminals.
T
G=2 ±10 volts
G=4 ±5 volts
G=5 ±4 volts
G=8 ±2.5 volts
G=10 ±2 volts
G=16 ±1.25 volts
G=20 ±1 volt
The reason the range is ±20 volts at G=1 is that, for example, AI0 could be +10 volts and AI1 could be -10 volts giving a difference
of +20 volts, or AI0 could be -10 volts and AI1 could be +10 volts giving a difference of -20 volts.
T
Figure 2-4 shows a single-ended connection used to measure the output voltage of a typical voltage-divider circuit. The voltage
divider circuit is a simple way to convert a varying resistance (thermistor, photoresistor, potentiometer, etc.) to a varying voltage.
With nothing connected to Va, the value of the unknown resistance, R2, can be calculated as:
R2 =
Hz per channel.
AIBurst acquires multiple samples of 1-4 channels at a hardware-timed sample rate of 400-8192 Hz. The acquisition can be
triggered based on a change of state on IO0 or IO1. This function also returns the states of the IO pins (which are read every 4
samples).
Internally, the actual number of samples collected and transferred by the LabJack du
Figure 2-5. IO used to detect the state of a switch.
While providing overvoltage/short-circuit protection, the 1.5 kΩ series resistor on each IO pin also limits the output current
capability. For instance, with an output current of 1 mA, the series resistor will drop 1.5 volts, resulting in an output voltage of about
3.5 volts.
Software
The easy functions E
Shorting a high output to ground (or any potential other than +5V).
Shorting a low output to a nonzero voltage (such as +5V).
Exceeding the voltage limits specified in Appendix A.
Software
The easy functions EDigitalIn or EDigitalOut are used to read or set the state of one digital line, and both take up to 20 ms to
execute.
The functions AOUpdate and Digita
The LabJack U12 is also available in 2 OEM (original equipment manufacturer) versions:
LJU12-PH: This is a populated LabJack U12 PCB with pin-headers installed (on the component side of the PCB) instead
of screw-terminals. Also, the LED is installed on the component side of the PCB, so nothing is installed on the solder side.
LJU12-NTH: This is a populated L
window shown in Figure 3-2 will appear.
Figure 3-2. LJconfig Change Local ID
Enter a new local ID between 0 and 255 and push the “Change” button. The new local ID will be written and the LabJack will be
forced to re-enumerate.
3.2 - LJcounter
Reads the LabJack counter and provides the current frequency or count.
Figure 3-3. LJcounter
Figure 3-3 shows the
Figure 3-4. LJfg
3.4 - LJlogger
LJlogger sends and receives data in command/response mode by making 2 DLL calls to “AISample” and 1 DLL call to
“AOUpdate”. It is capable of saving data to disk (2 Hz max) and performing various actions on trigger events.
Figure 3-5. LJlogger
The main window for LJlogger is shown in Figure 3-5. The white colored items