Series Analog Output Series User Manual NI 6711/6713/DAQCard-6715, NI 6722/6723, and NI 6731/6733 Devices Analog Output Series User Manual January 2017 370735F-01 all-guides com ...
ni.com Worldwide Offices Visit ni.com/niglobal to access the branch office websites, which provide up-to-date contact information, support phone numbers, email addresses, and current events. National Instruments Corporate Headquarters 11500 North Mopac Expressway Austin, Texas 78759-3504 USA Tel: 512 683 0100 For further support information, refer to the NI Services appendix. To comment on NI docu ...
refer to ni.com/manuals. NI reviews this document carefully for technical accuracy; however, NI MAKES NO EXPRESS OR IMPLIED WARRANTIES AS TO THE ACCURACY OF THE INFORMATION CONTAINED HEREIN AND SHALL NOT BE LIABLE FOR ANY ERRORS. NI warrants that its hardware products will be free of defects i ...
DeviceNet™ and EtherNet/IP™ are trademarks of ODVA. Go!, SensorDAQ, and Vernier are registered trademarks of Vernier Software & Technology. Vernier Software & Technology and vernier.com are trademarks or trade dress. Xilinx is the registered trademark of Xilinx, Inc. Taptite and Trilobular are registered trademarks of Research Engineering & Manufacturing Inc. FireWire® is the regi ...
This hardware has been tested and found to comply with the applicable regulatory requirements and limits for electromagnetic compatibility (EMC) as indicated in the hardware’s Declaration of Conformity (DoC)1. These requirements and limits are designed to provide reasonable protection against harmful interference when the hardware is operated in the intended electromagnetic environment. In special case ...
| vii Contents Chapter 1 DAQ System Overview Safety Guidelines.............................................................................................................. 1-2 Electromagnetic Compatibility Guidelines ...................................................................... 1-2 DAQ Hardware................................................................................................................. 1-3 DAQ-STC...... ...
Waveform Generation Timing Summary .................................................................3-5 AO Start Trigger Signal............................................................................................3-5 Using a Digital Source............................................................................. ...
© National Instruments | ix Counter 0 Internal Output Signal.............................................................................. 5-4 CTR 0 OUT Pin................................................................................................ 5-5 Counter 0 Up/Down Signal ...................................................................................... 5-5 Counter 1 Source Signal.............................. ...
Contents x | ni.com Appendix A Device-Specific Information Appendix B Troubleshooting Appendix C NI Services Glossary Index ...
| 1-1 1 DAQ System Overview Figure 1-1 shows a typical DAQ system setup, which includes transducers, signal conditioning, cables that connect the various devices to the accessories, the analog output device, and the programming software. Refer to the Using Accessories with Devices section for a list of devices and their compatible accessories. Figure 1-1. DAQ System Setup 1 Sensors and Transducers 2 Terminal Block Accessory 3 Cable ...
Safety Guidelines Operate the device only as described in this document. Caution This icon denotes a caution, which advises you of precautions to take to avoid injury, data loss, or a system crash. Caution The protection provided by the device can be impaired if it is used in a manner not described in this document. Misuse of the device can result in a hazard. You can compromise the safety protection built into th ...
| 1-3 Analog Output Series User Manual Caution To ensure the specified EMC performance of the DAQCard-6715, the length of the I/O cable must be no longer than 3 m (10 ft). For all other products, the length of the I/O cable must be no longer than 30 m (100 ft). DAQ Hardware DAQ hardware digitizes signals, performs D/A conversions to generate analog output signals, and measures and controls digital I/O signals. Figure 1-2 shows the c ...
Calibration Circuitry Calibration is the process of making adjustments to a measurement device to reduce errors associated with measurements. Without calibration, the measurement results of your device will drift over time and temperature. Calibration adjusts for these changes to improve measurement accuracy and ensure that your product meets its required specifications. DAQ devices have high precision analog circ ...
| 1-5 Analog Output Series User Manual • RTSI bus cables • Low channel-count digital signal conditioning modules, devices, and accessories For more specific information about these products, refer to ni.com. The following sections contain information on how to select accessories for your AO Series device. Using Accessories with Devices Go to ni.com/info and enter the Info Code AOcables for the most current list of supported cabl ...
NI 6723 (AO 0–7 & DIGITAL connector) SH68-C68-S (Recommended, Shielded) RC68-68 (Low Cost) BNC-2110 CA1000 CB-68LP CB-68LPR SCB-68A SBC-68 TBX-68 NI 6723 (AO 8–31 connector) SH68-C68-S (Recommended, Shielded) RC68-68 (Low Cost) BNC-2115 CA-1000 CB-68LP CB-68LPR SCB-68A SCB-68 TBX-68 NI 6731/6733 SH68-68-EPM (Recommended) R6868 (Low Cost) BNC-2110 CA1000 CB-68LP CB-68LPR SCB-68 SCB-68A TBX-68 TB-2705 (PXI only) ...
| 1-7 Analog Output Series User Manual Custom Cabling Follow these guidelines if you want to develop your own cable. • Route the analog lines separately from the digital lines. • When using a cable shield, use separate shields for the analog and digital halves of the cable. Failure to do so results in noise coupling into the analog signals from transient digital signals. Table 1-3 shows the recommended connectors to use with the ...
Programming Devices in Software National Instruments measurement devices are packaged with NI-DAQmx driver software, an extensive library of functions and VIs you can call from your application software, such as LabVIEW or LabWindows/CVI, to program all the features of your NI measurement devices. Driver software has an application programming interface (API), which is a library of VIs, functions, classes, attribu ...
| 2-1 2 I/O Connector This chapter contains information about the AO Series I/O connectors. Note Some hardware accessories may not yet reflect the NI-DAQmx terminal names. If you are using an AO Series device in Traditional NI-DAQ (Legacy), refer to Table 2-1 for the Traditional NI-DAQ (Legacy) signal names. 68-Pin AO I/O Connector Pinouts Figure 2-1, Figure 2-2, and Figure 2-3 show the pinouts of 68-pin AO Series devices. ...
FREQ OUT CTR 0 OUT PFI 9/CTR 0 GATE D GND PFI 6/AO START TRIG PFI 5/AO SAMP CLK D GND +5 V D GND PFI 1 PFI 0 D GND D GND +5 V D GND P0.6 P0.1 D GND P0.4 AO EXT REF AO 1 AO 0 AO GND AO GND AO 3 AO GND AO GND NC AO GND NC AO GND AO GND NC AO GND D GND PFI 8/CTR 0 SOURCE PFI 7 CTR 1 OUT PFI 4/CTR 1 GATE PFI 3/CTR 1 SOURCE PFI 2 D GND D GND D GND EXT STROBE NC P0.3 P0 ...
| 2-3 TitleShort-Hidden (cross reference text) Figure 2-2. NI 6713/DAQCard-6715/NI 6733 68-Pin AO I/O Connector Pinout FREQ OUT CTR 0 OUT PFI 9/CTR 0 GATE D GND PFI 6/AO START TRIG PFI 5/AO SAMP CLK D GND +5 V D GND PFI 1 PFI 0 D GND D GND +5 V D GND P0.6 P0.1 D GND P0.4 AO EXT REF AO 1 AO 0 AO GND AO GND AO 3 AO GND AO GND AO 5 AO GND AO 6 AO GND AO GND NC AO GND D GND PFI 8/CTR 0 SOURCE PFI 7 CTR 1 OUT PFI 4/CTR 1 GATE PFI 3/CTR ...