OTT PLS Supplement To Operating Instructions
Text of the manual
Table of contents
1 Scope of supply
4
2 Order numbers
4
3 Introduction
5
4 Please note – important information
5
5 Separating/inserting plug-in connection
5
6 Information on repair
7
7 Technical Data
7
1 Scope of supply
– 1 pressure probe with ceramic, capacitive relative pressure measuring cell and
plug-in, shielded, pressure probe cable with pressure compensation capillary
and Kevlar core for length stabilization. Pre-fabricated cable end with trans-
port protection against moisture. Optionally with 4 … 20 mA or RS-485 inter-
face (SDI-12 protoco
The "OTT PLS plug-in type" is equipped with a plug-in pressure probe cable, in
contrast to the standard model. In the case of damage to the device, the pressure
probe can thus be replaced more easily and deinstalling the pressure probe cable
is not necessary. In the same way, during the initial installation, first the pressure
probe cable (wit
Pressure probe cable
O-ring
Pressure
probe
Connector
(connector
sleeve on
connector
body)
Protective cap
Coding lug
Protective cap
Caution:
There must not be
a gap here when
connected and
screwed together
(as far as possible)!
Caution:
Avoid the entry of
damp and contamination
when disconnected!
Put on protective caps!
Coding lug
How to join the plug-in connection:
If necessary: dry and clean the connector, pressure probe and protective cap*.
Take the protective cap off the connector and the pressure probe.
Check the O-ring on the pressure probe for damage and correct seating.
Replace O-ring if necessary* (spare part - temporarily cover the thread with
adhesive tape before repl