Vernacare Vortex+ Technical Manual

File Specifications

Brand: Vernacare
Category: Medical Equipment
Model: Vortex+ (Technical Manual), prwnCDF7trFGE
File Type: PDF
File Size: —
Pages: 31 pages

Download PDF

Download Vernacare Vortex+ Manual (31 pages)

Read Online:

Read the Vernacare Vortex+ Technical Manual online using the built-in PDF viewer.

Key Pages of Document

Vernacare Vortex+ Technical Manual

Contents Declaration of Conformity Page 01 Warranty Page 02 Section 1 - Specifications Page 05 Section 2 - Installation & Commissioning Page 08 Section 3 - General Description Page 15 Section 4 - Outer Casing Page 17 Section 5 - Motor Assembly Page 18 Section 6 - Diaphragm Valve Assembly Page 19 Section 7- Lid Arm Assembly ...

Page | 3

Declaration of Conformity VERNACARE Declare that this declaration has been issued under the sole responsibility of the manufacturer. The object of the declaration is in conformity with the relevant Union harmonisation PRODUCT Vernacare Vortex / Vortex + Macerator DIRECTIVE EC Directive 2014/30/EU EC Directive 2006/42/EC EC Directive 2014/35/EU EC Di ...

Page | 4

Warranty A. Warranty terms 1. The warranty is for 12 months from the date of commissioning or 18 months from the date of delivery, whichever is the shortest period. 2. The warranty will cover parts and labour. 3. The warranty will not cover customer damage or abuse. 4. Warranty requests should be submitted in writing, quoting the machine number, dat ...

Page | 5

1 Specifications 1.1 Capacities and usage The “Vernacare Vortex +” macerator will dispose of a maximum of four “Vernacare” disposable products such as bedpan liners, bottles etc. together with their contents in a single cycle. Please note due to their physical size only two Vernacare wash bowls may be disposed of per cycle. Products should b ...

Page | 6

except in case of emergency, as this will impair the effective operation of the machine and may cause damage to the components. 1.4.1 Motor 1.1 KW; 3 Phase; 220-240v 50Hz/ 380-420v 50Hz; IP55; D90 frame size. Rating at peak load 4.57 amps (nominal). 1.4.2 Invertor AC invertor with either 110v or 230v supply voltage, 230v motor output and utilising c ...

Page | 7

150mm Clear each side 100mm Clear at rear Deodoriser fill cap NOTE: Water inlet & warning pipe connections can be made to either the left or right hand side of the machine. 2" Drain Connector ALL DIMENSIONS IN mm 2 Mounting Holes Ø9 Dimensions of the Vernacare Vortex + Water Inlet ½’’ BSP Male Warning Pipe Connection (drain) with lid open !IMP ...

Page | 8

2 Installation / Commissioning Installation Note: Installation of this equipment should only be carried out by a suitable qualified person. 2.1 Siting and fixing The “Vernacare Vortex +” disposal unit should be positioned adjacent to the following: a) A suitable power supply point.* b) An adequate water supply line. c) A suitable drainage connec ...

Page | 9

2.2 Plumbing 2.2.1 Water supply The machine should be connected to a suitable cold water supply as follows:- The connection to the machine water tank is made at the upper rear left or right hand side corner of the machine. It is a standard connection to a 1/2 inch BSP float-type cistern valve. This can be made in rigid or flexible piping, according ...

Page | 10

2.2.3 Overflow - Warning Pipe The Vortex Is supplied with a 32mm (1 ¼“) tank connector which should be installed using one of the predrilled holes in either the Left Hand side or Right Hand side of the water tank. It is Important that any discharge from the warning pipe is visible. A standard warning pipe installation kit is available. Contact V ...