WIKA 432.55 Operating Instructions Manual
Text of WIKA 432.55 Operating Instructions Manual
Manual de instrucciones
Flush diaphragm pressure gauge, model 432.55
Pressure gauge for sanitary applications,
model 432.55
Druckmessgerät für die sterile Verfahrenstechnik,
Typ 432.55
Manomètre pour applications sanitaires,
type 432.55
Manómetro para procesos estériles,
modelo 432.55
14055446.01 12/2012 GB/D/F/E
WIKA operating instructions pressure gauge, model 432.55
© 2012 WIKA Alexander Wiegand SE & Co. KG
All rights reserved. / Alle Rechte vorbehalten.
WIKA® is a registered trademark in various countries.
WIKA® ist eine geschützte Marke in verschiedenen Ländern.
Prior to starting any work, read the operating instructi
WIKA operating instructions pressure gauge, model 432.55
3
14055446.01 12/2012 GB/D/F/E
1.
General information
4
2.
Safety
6
3.
Specifications
7
4.
Design and function
8
5.
Transport, packaging and storage
9
6.
Commissioning, operation
9
7.
Maintenance and cleaning
11
8.
Faults
11
9.
Dismounting and disposal
12
Appendix:
Certificate 3-A
Certificate of com
WIKA operating instructions pressure gauge, model 432.55
14055446.01 12/2012 GB/D/F/E
1. General information
� The pressure gauge described in the operating instructions has
been designed and manufactured using state-of-the-art technology.
All components are subject to stringent quality and environmental
criteria during production. Our management syst
Selection and installation recommendations for pressure
gauges
EN 837-3
Diaphragm and capsule pressure gauges, dimensions,
metrology, requirements and testing
� Subject to technical modifications.
� Further information:
- Internet address:
www.wika.de / www.wika.com
- Relevant data sheet:
PM 04.09
- Application consultant:
T
WIKA operating instructions pressure gauge, model 432.55
14055446.01 12/2012 GB/D/F/E
2. Safety
WARNING!
Before installation, commissioning and operation, ensure
that the appropriate pressure gauge has been selected
in terms of measuring range, design and suitable wetted
material (corrosion) for the specific measuring conditions.
In order to guarantee t
WIKA operating instructions pressure gauge, model 432.55
7
14055446.01 12/2012 GB/D/F/E
� The activities described in these operating
instructions may only be carried out by skilled
personnel who have the qualifications described
below.
Skilled personnel
Skilled personnel are understood to be personnel who, based on their
technical training, knowledge o
WIKA operating instructions pressure gauge, model 432.55
14055446.01 12/2012 GB/D/F/E
Temperature effect
When the temperature of the measuring system deviates from the
reference temperature (+20 °C): max. ±0.8 %/10 K of full scale value
Ingress protection
IP 54 per EN 60529 / IEC 529
For further specifications see WIKA data sheet PM 04.09 and the
orde
WIKA operating instructions pressure gauge, model 432.55
9
14055446.01 12/2012 GB/D/F/E
5. Transport, packaging ... / 6. Commissioning ...
5. Transport, packaging and storage
5.1 Transport
Check the pressure gauge for any damage that may have been
caused by transport. Obvious damage must be reported immediately.
5.2 Packaging
Do not remove packaging until