WIKA 432.55 Operating Instructions Manual

Text of WIKA 432.55 Operating Instructions Manual

Manual de instrucciones
Flush diaphragm pressure gauge, model 432.55 Pressure gauge for sanitary applications, model 432.55 Druckmessgerät für die sterile Verfahrenstechnik, Typ 432.55 Manomètre pour applications sanitaires, type 432.55 Manómetro para procesos estériles, modelo 432.55
14055446.01 12/2012 GB/D/F/E
WIKA operating instructions pressure gauge, model 432.55 © 2012 WIKA Alexander Wiegand SE & Co. KG All rights reserved. / Alle Rechte vorbehalten. WIKA® is a registered trademark in various countries. WIKA® ist eine geschützte Marke in verschiedenen Ländern. Prior to starting any work, read the operating instructi
WIKA operating instructions pressure gauge, model 432.55
3 14055446.01 12/2012 GB/D/F/E 1. General information 4 2. Safety 6 3. Specifications 7 4. Design and function 8 5. Transport, packaging and storage 9 6. Commissioning, operation 9 7. Maintenance and cleaning 11 8. Faults 11 9. Dismounting and disposal 12 Appendix: Certificate 3-A Certificate of com
WIKA operating instructions pressure gauge, model 432.55
14055446.01 12/2012 GB/D/F/E 1. General information � The pressure gauge described in the operating instructions has been designed and manufactured using state-of-the-art technology. All components are subject to stringent quality and environmental criteria during production. Our management syst
Selection and installation recommendations for pressure
gauges EN 837-3 Diaphragm and capsule pressure gauges, dimensions, metrology, requirements and testing � Subject to technical modifications. � Further information: - Internet address: www.wika.de / www.wika.com - Relevant data sheet: PM 04.09 - Application consultant: T
WIKA operating instructions pressure gauge, model 432.55
14055446.01 12/2012 GB/D/F/E 2. Safety WARNING! Before installation, commissioning and operation, ensure that the appropriate pressure gauge has been selected in terms of measuring range, design and suitable wetted material (corrosion) for the specific measuring conditions. In order to guarantee t
WIKA operating instructions pressure gauge, model 432.55
7 14055446.01 12/2012 GB/D/F/E � The activities described in these operating instructions may only be carried out by skilled personnel who have the qualifications described below. Skilled personnel Skilled personnel are understood to be personnel who, based on their technical training, knowledge o
WIKA operating instructions pressure gauge, model 432.55
14055446.01 12/2012 GB/D/F/E Temperature effect When the temperature of the measuring system deviates from the reference temperature (+20 °C): max. ±0.8 %/10 K of full scale value Ingress protection IP 54 per EN 60529 / IEC 529 For further specifications see WIKA data sheet PM 04.09 and the orde
WIKA operating instructions pressure gauge, model 432.55
9 14055446.01 12/2012 GB/D/F/E 5. Transport, packaging ... / 6. Commissioning ... 5. Transport, packaging and storage 5.1 Transport Check the pressure gauge for any damage that may have been caused by transport. Obvious damage must be reported immediately. 5.2 Packaging Do not remove packaging until