MSA altair 5 Instruction Manual
Read Online
Read the MSA altair 5 Instruction Manual online using the built-in PDF viewer.
Text of the manual
ALTAIR® 5 Multigas Detector and
ALTAIR® 5 IR Multi Gas Detector
Instruction Manual
Détecteur multigaz ALTAIR® 5 et
Détecteur multigaz ALTAIR® 5 IR
Manuel d'utilisation
Detector Multigases ALTAIR® 5 y
Detector Multigases ALTAIR® 5 IR
Manual de funcionamiento
In North America, to contact your nearest stocking location, dial toll-free 1-800-MSA-2222
To contact MSA International, dial 1-412-967-3354 or 1-800-MSA-7777.
© MINE SAFETY APPLIANCES COMPANY 2009 - All Rights Reserved
This manual is available on the internet at www.msanet.com
En Amérique du Nord, pour contacter notre dépôt le plus proche, composez le numéro gratuit 1-800-
MSA-2222. Pour contacter MSA International, composez le 1-412-967-3354 ou le 1-800-MSA-7777.
© MINE SAFETY APPLIANCES COMPANY 2009 - Tous droits réservés
Ce manuel est disponible sur l’Internet au www.msanet.com
Para comunicarse con su localización de aprovisionamiento más cercana en América del Norte,
llame sin tarifa al 1-800-MSA-2222. Para comunicarse con MSA International, marque el 1-412-967-
3354 ó 1-800-MSA-7777.
© MINE SAFETY APPLIANCES COMPANY 2009 - Se reservan todos los derechos
Este manual puede obtenerse en la Internet en el sitio: www.msanet.com
Manufactured by/Fabriqué par/Fabricado por
MSA NORTH AMERICA
P.O. Box 427, Pittsburgh, Pennsylvania 15230
(L) Rev 2
10094176
Multigas Detector and
ALTAIR® 5 IR
Multi Gas Detector
Instruction Manual
In North America, to contact your nearest stocking location, dial toll-free 1-800-MSA-2222
To contact MSA International, dial 1-412-967-3354 or 1-800-MSA-7777.
© MINE SAFETY APPLIANCES COMPANY 2009 - All Rights Reserved
This manual is available on the internet at www.msanet.com
Manufactured by
MSA NORTH AMERICA
P.O. Box 427, Pittsburgh, Pennsylvania 15230
(L) Rev 2
10094176
THIS MANUAL MUST BE CAREFULLY READ BY ALL INDIVIDUALS WHO HAVE OR WILL
HAVE THE RESPONSIBILITY FOR USING OR SERVICING THE PRODUCT. Like any piece
of complex equipment, this instrument will perform as designed only if it is used and serv-
iced in accordance with the manufacturer’s instructions. OTHERWISE, IT COULD FAIL TO
PERFORM AS DESIGNED AND PERSONS WHO RELY ON THIS PRODUCT FOR THEIR
SAFETY COULD SUSTAIN SEVERE PERSONAL INJURY OR DEATH.
The warranties made by Mine Safety Appliances Company with respect to the product are
voided if the product is not used and serviced in accordance with the instructions in this
manual. Please protect yourself and others by following them. We encourage our cus-
tomers to write or call regarding this equipment prior to use or for any additional infor-
mation relative to use or repairs.
"! WARNING
Table of Contents
1. Instrument Safety . . . . . . . . . . . . . . . . . . . .1-1
1.1. Correct Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-1
1.2. Safety and Precautionary Measures . . . . . . . . . . .1-2
1.3. Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-5
1.3.1 Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-5
1.3.2 Exclusive Remedy . . . . . . . . . . . . . . . . . . . . . . .1-6
1.3.3 Exclusion of Consequential Damage . . . . . . . .1-6
2. Description . . . . . . . . . . . . . . . . . . . . . . . . .2-1
2.1. Instrument Overview. . . . . . . . . . . . . . . . . . . . . . . . .2-1
Figure 2-1. Instrument View . . . . . . . . . . . . . . . . . . .2-1
2.2 User Interfaces . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2
2.2.1 Button Definitions . . . . . . . . . . . . . . . . . . . . . . . .2-2
2.2.2 LED Definitions . . . . . . . . . . . . . . . . . . . . . . . . . .2-3
2.2.3 Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-3
2.3.3.1 Vibrating Alarm . . . . . . . . . . . . . . . . . . . . . . . .2-3
2.2.3.2 Horn . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-3
2.3.3.3 InstantAlert™ Alarm . . . . . . . . . . . . . . . . . . . . .2-3
2.3.3.4 Motion Alert Alarm . . . . . . . . . . . . . . . . . . . . . .2-3
2.2.3.5 Stealth Mode . . . . . . . . . . . . . . . . . . . . . . . . . .2-4
2.2.4 Backlight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-4
2.2.5 Confirmation Beep . . . . . . . . . . . . . . . . . . . . . . .2-4
2.3. On-Screen Indicators . . . . . . . . . . . . . . . . . . . . . . .2-5
2.3.1 Monochrome Display . . . . . . . . . . . . . . . . . . . . .2-5
Figure 2-2. Monochrome Display . . . . . . . . . . . . . . .2-5
2.3.2 Color Display . . . . . . . . . . . . . . . . . . . . . . . . . . .2-6
2.3.2.1 Wireless-Equipped Units . . . . . . . . . . . . . . . . .2-6
2.3.3 Battery Life Indicator . . . . . . . . . . . . . . . . . . . . . .2-6
Figure 2-3. Color Display . . . . . . . . . . . . . . . . . . . . .2-6
Figure 2-4. Battery Warning . . . . . . . . . . . . . . . . . . .2-7
2.3.4 Battery Charging . . . . . . . . . . . . . . . . . . . . . . . . .2-8
Figure 2-5. Battery Shut Down . . . . . . . . . . . . . . . .2-8
2.4. Viewing Additional Pages . . . . . . . . . . . . . . . . . . . .2-9
2.4.1 Peak Readings (PEAK page) . . . . . . . . . . . . . .2-10
2.4.2 Minimum Readings (MIN page) . . . . . . . . . . . .2-10
i
2.4.3 Short Term Exposure Limits (STEL page) . . . .2-11
2.4.4 Time Weighted Average (TWA Page) . . . . . . . .2-12
2.4.5 Date Display . . . . . . . . . . . . . . . . . . . . . . . . . . .2-13
2.4.6 Last Cal Page . . . . . . . . . . . . . . . . . . . . . . . . . .2-13
2.4.7 Cal Due Page . . . . . . . . . . . . . . . . . . . . . . . . . .2-13
2.4.8 Motion Alert Activation Page . . . . . . . . . . . . . . .2-13
2.5 Sensor Missing Alarm . . . . . . . . . . . . . . . . . . . . . .2-13
2.6 Monitoring Toxic Gases . . . . . . . . . . . . . . . . . . . . .2-14
2.7 Monitoring Oxygen Concentration . . . . . . . . . . . . .2-15
Figure 2-6. Alarm Conditions (High Alarm shown) 2-15
2.8. Monitoring Combustible Gases . . . . . . . . . . . . . . .2-16
3. Operation . . . . . . . . . . . . . . . . . . . . . . . . . . .3-1
3.1. Environmental Factors . . . . . . . . . . . . . . . . . . . . . .3-1
3.2. Turning ON and Fresh Air Setup . . . . . . . . . . . . . . .3-1
3.2.1 Sampling Safety Test . . . . . . . . . . . . . . . . . . . . .3-3
3.2.2 Fresh Air Setup (FAS) . . . . . . . . . . . . . . . . . . . . .3-3
Figure 3-1. Fresh Air Setup . . . . . . . . . . . . . . . . . . .3-4
3.3. Measurement Mode (Normal Operation) . . . . . . . .3-5
3.4 Instrument Setup . . . . . . . . . . . . . . . . . . . . . . . . . . .3-5
3.4.1 Calibration Options . . . . . . . . . . . . . . . . . . . . . . .3-6
Figure 3-2. Password Screens . . . . . . . . . . . . . . . .3-6
Figure 3-3. Calibration Options . . . . . . . . . . . . . . . .3-6
3.4.2 Alarm Options . . . . . . . . . . . . . . . . . . . . . . . . . . .3-8
Figure 3-4. Alarm Options . . . . . . . . . . . . . . . . . . . .3-8
Figure 3-5. Sensor Alarm Setup . . . . . . . . . . . . . .3-10
Figure 3-6. Sensor Alarm Setup . . . . . . . . . . . . . .3-10
3.4.3 Instrument Options . . . . . . . . . . . . . . . . . . . . . .3-11
Figure 3-7. Instrument Options . . . . . . . . . . . . . . .3-11
Figure 3-8. Sensor Options . . . . . . . . . . . . . . . . . .3-12
3.5. MSA Link Operation . . . . . . . . . . . . . . . . . . . . . . .3-14
3.6. Function Tests on the Instrument . . . . . . . . . . . . .3-14
3.7. Calibration Check . . . . . . . . . . . . . . . . . . . . . . . . .3-14
3.8. Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-15
3.8.1 Zero Calibration . . . . . . . . . . . . . . . . . . . . . . . .3-16
3.8.2 Span Calibration . . . . . . . . . . . . . . . . . . . . . . . .3-17
3.8.3 Finishing Calibration . . . . . . . . . . . . . . . . . . . . .3-19
3.8.4 Calibration Failure . . . . . . . . . . . . . . . . . . . . . . .3-19
ii
3.8.5 Calibration with GALAXY Test System . . . . . . .3-19
3.9 Instrument Shutdown . . . . . . . . . . . . . . . . . . . . . . .3-20
Figure 3-9. Instrument Shutdown . . . . . . . . . . . . .3-20
4. Maintenance . . . . . . . . . . . . . . . . . . . . . . . .4-1
4.1 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-2
4.2 Verifying Pump Operation . . . . . . . . . . . . . . . . . . . .4-2
4.3 Replacing the Battery . . . . . . . . . . . . . . . . . . . . . . . .4-4
Figure 4-1. Battery Replacement . . . . . . . . . . . . . . .4-4
Figure 4-2. Battery Replacement . . . . . . . . . . . . . . .4-4
4.4 Live Maintenance Procedure-
Replacing or Adding a Sensor . . . . . . . . . . . . . . . . . . . .4-5
Figure 4-3. Sensor Replacement . . . . . . . . . . . . . . .4-6
4.5. Replacing the Filter
(for instruments with integral pump) . . . . . . . . . . . . . . . .4-7
4.6. Cleaning the Instrument . . . . . . . . . . . . . . . . . . . . .4-7
4.7 Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-8
4.8. Shipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-8
5. Technical Specifications/Certifications . .5-1
5.1 Technical Specifications . . . . . . . . . . . . . . . . . . . . .5-1
5.2 Factory-set Alarm Thresholds . . . . . . . . . . . . . . . . .5-2
5.3 Sensor Performance Specifications . . . . . . . . . . . . .5-3
5.4 Calibration Specification . . . . . . . . . . . . . . . . . . . . . .5-5
5.5 Certifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-6
5.6 Combustible Gas Cross Reference Factors . . . . . .5-7
6. Order Information . . . . . . . . . . . . . . . . . . .6-1
6.1 Gas Cylinder Parts List . . . . . . . . . . . . . . . . . . . . . .6-1
6.2 Accessories Parts List . . . . . . . . . . . . . . . . . . . . . . .6-2
6.3 Instrument Replacement Parts . . . . . . . . . . . . . . . . .6-3
7. Appendix – Flow Charts . . . . . . . . . . . . . . .7-1
7.1. Start Up Sequence (Power ON) . . . . . . . . . . . . . . .7-1
7.2. Normal Operation . . . . . . . . . . . . . . . . . . . . . . . . . .7-2
7.3. Options from Main Page . . . . . . . . . . . . . . . . . . . . .7-3
iii
7.4. Password Protected Options . . . . . . . . . . . . . . . . . .7-4
7.5. Menu Sequence from Main page . . . . . . . . . . . . . .7-5
7.6. Calibration Sequence . . . . . . . . . . . . . . . . . . . . . . .7-7
7.7. Calibration Options . . . . . . . . . . . . . . . . . . . . . . . . .7-8
7.8. Alarm Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-9
7.8.1 Sensor Alarm Options . . . . . . . . . . . . . . . . . . . .7-10
7.9. Instrument Options . . . . . . . . . . . . . . . . . . . . . . . .7-11
iv
1. Instrument Safety
1.1. Correct Use
The ALTAIR 5 and ALTAIR 5 IR Multigas Detectors are for use by
trained and qualified personnel. They are designed to be used when
performing a hazard assessment to:
•
Assess potential worker exposure to combustible and toxic gases
and vapors as well as low level of oxygen.
•
Determine the appropriate gas and vapor monitoring needed for a
workplace.
The ALTAIR 5 and ALTAIR 5 IR Multigas Detectors can be equipped
to detect up to five gases in the ALTAIR 5 and up to six gases in the
ALTAIR 5 IR:
•
Combustible gases and certain combustible vapors
•
Oxygen-deficient or oxygen-rich atmospheres
•
Specific toxic gases for which a sensor is installed.
While the instrument can detect up to 25 % oxygen in ambient air, it is
approved for use only up to 21 % oxygen.
The ALTAIR 5 IR Gas Detector contains one Infrared sensor to detect
either CO2 or specific combustible gases up to 100% volume, in
addition to up to five other gases.
Read and follow all instructions carefully.
•
Perform a blocked flow test before each day's use if
equipped with an integral pump.
•
Check calibration before each day's use and adjust if
necessary.
•
Check calibration more frequently if exposed to silicone,
silicates, lead-containing compounds, hydrogen sulfide, or
high contaminant levels.
•
Recheck calibration if unit is subjected to physical shock.
•
Use only to detect gases/vapors for which a sensor is
installed.
•
Do not use to detect combustible dusts or mists.
" WARNING
Make sure adequate oxygen is present.
•
Never block sensors.
•
Never block pump inlet, except to perform sampling
safety test.
•
Have a trained and qualified person interpret instrument
readings.
•
Do not remove battery pack from instrument while in a
hazardous location.
•
Do not recharge Li Ion battery in a hazardous location.
Risk of explosion may occur.
•
Do not replace alkaline batteries in a hazardous location.
•
Do not alter or modify instrument.
•
Use only MSA-approved sampling lines.
•
Do not use silicone tubing or sampling lines.
•
Wait sufficient time for the reading; response times vary,
based on gas and length of sampling line.
•
Do not use the instrument with integral pump for
prolonged periods in an atmosphere containing a
concentration of fuel or solvent vapors that may be
greater than 10%LEL.
INCORRECT USE CAN CAUSE SERIOUS PERSONAL INJURY OR DEATH.
This Class A digital apparatus complies with Canadian ICES-003.
For users in Australia/ New Zealand: The charge cradle is a Class A
product. In a domestic environment, this product may cause radio
interference, in which case, the user may be required to take adequate
measures.
1.2. Safety and Precautionary Measures
Carefully review the following safety limitations and
precautions before placing this instrument in service.
Incorrect use can cause serious personal injury or death.
The Altair 5 Multigas Detector is designed to:
•
Detect gases and vapors in air only
" WARNING
1-2
Detect only specified toxic gases for which a sensor is installed.
(for available sensors,see Section 5.3)
•
Perform a blocked flow test (see section 4.2) before each day's
use if equipped with an integral pump.
•
MSA recommends to check the instrument's functionality
(see section 3.6) and carry out a routine inspection each day
before use.
•
Perform calibration check (see section 3.7) before each day's use
to verify proper instrument operation. Adjust calibration if the
readings are not within the specified limits.
•
Check calibration more frequently if unit is subjected to physical
shock or high levels of contaminants. Also, check calibration more
frequently if the tested atmosphere contains the following
materials, which may desensitize the combustible gas sensor and
reduce its readings:
•
Organic silicones
•
Silicates
•
Lead-containing compounds
•
Hydrogen sulfide exposures over 200 ppm or exposures over
50 ppm for one minute.
The minimum concentration of a combustible gas in air that can ignite is
defined as the Lower Explosive Limit (LEL). A combustible gas reading
of "100” or "5.00” indicates the atmosphere is above 100 % LEL or 5.00
%vol CH4, and an explosion hazard exists. Move away from hazardous
area immediately.
Do not use the ALTAIR 5 or ALTAIR 5 IR Multigas Detectors to test for
combustible or toxic gases in the following atmospheres as this may
result in erroneous readings:
•
Oxygen-deficient or oxygen-rich atmospheres
•
Reducing atmospheres
•
Furnace stacks
•
Inert environments (only IR sensors acceptable for use)
•
Atmospheres containing combustible airborne mists/dusts.
Do not use the ALTAIR 5 or ALTAIR 5 IR Multigas Detectors to test for
combustible gases in atmospheres containing vapors from liquids with a
high flash point (above 38 °C, 100°F) as this may result in erroneously
low readings.
1-3
Do not block sensor openings as this may cause inaccurate readings.
Do not press on the face of the sensors, as this may damage them and
cause erroneous readings. Do not use compressed air to clean the
sensor holes, as the pressure may damage the sensors.
Allow sufficient time for unit to display accurate reading. Response
times vary based on the type of sensor being utilized (see Section 5.3,
"Sensor Performance Specifications"). For units with a sample pump,
allow a minimum of 1 second per foot (3 seconds per meter) of sample
line to allow the sample to be drawn through the sensors.
Sampling lines made from 0.062 inch (1.57 mm) inner diameter tubing
provide fast transport times to the instrument; however, they must be
limited to 50 feet (15 m) in length.
Sampling of reactive toxic gases (Cl2, ClO2, NH3) must only be done
with the reactive gas sample line and probe kits listed in the
Accessories table in Section 6.2.
All instrument readings and information must be interpreted by
someone trained and qualified in interpreting instrument readings in
relation to the specific environment, industrial practice and exposure
limitations.
Observe proper battery maintenance
Use only battery chargers made available by MSA for use with this
instrument; other chargers may damage the battery pack and the
instrument. Dispose of in accordance with local health and safety
regulations.
Be aware of environmental conditions
A number of environmental factors may affect the oxygen and toxic
sensor readings, including changes in pressure, humidity and
temperature. Pressure and humidity changes affect the amount of
oxygen actually present in the atmosphere.
After exposure to negative (°C) temperatures, units with HCN (hydrogen
cyanide) sensors should be bump checked and calibrated, if necessary,
after returning to room temperature.
Be aware of the procedures for handling electrostatically sensitive
electronics
The instrument contains electrostatically sensitive components. Do not
open or repair the unit without using appropriate electrostatic discharge
(ESD) protection. The warranty does not cover damage caused by
electrostatic discharges.
1-4
Be aware of the product regulations
Follow all relevant national regulations applicable in the country of use.
Be aware of the warranty regulations
The warranties made by Mine Safety Appliances Company with respect
to the product are voided if the product is not used and maintained in
accordance with the instructions in this manual. Please protect yourself
and others by following them. We encourage our customers to write or
call regarding this equipment prior to use or for any additional
information relative to use or service.
1.3. Warranty
ITEM
WARRANTY PERIOD
Chassis and electronics
Two years*
COMB, O2, H2S, CO sensors
Two years*
Cl2, ClO2, NH3, HCN, NO2, PH3, SO2 sensors
One year*
Infrared Sensor
Two years*
*For extended warranty offerings please contact MSA
1.3.1 Warranty
This warranty does not cover filters, fuses, etc. Certain other
accessories not specifically listed here may have different warranty
periods. This warranty is valid only if the product is maintained and
used in accordance with Seller's instructions and/or recommendations.
The Seller shall be released from all obligations under this warranty in
the event repairs or modifications are made by persons other than its
own or authorized service personnel or if the warranty claim results from
physical abuse or misuse of the product. No agent, employee or
representative of the Seller has any authority to bind the Seller to any
affirmation, representation or warranty concerning this product. Seller
makes no warranty concerning components or accessories not
manufactured by the Seller, but will pass on to the Purchaser all
warranties of manufacturers of such components.
THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES,
EXPRESSED, IMPLIED OR STATUTORY, AND IS STRICTLY LIMITED
TO THE TERMS HEREOF. SELLER SPECIFICALLY DISCLAIMS ANY
WARRANTY OF MERCHANTABILITY OR OF FITNESS FOR A
PARTICULAR PURPOSE.
1-5
1.3.2 Exclusive Remedy
It is expressly agreed that Purchaser's sole and exclusive remedy for
breach of the above warranty, for any tortious conduct of Seller, or for
any other cause of action, shall be the replacement at Seller's option, of
any equipment or parts thereof, which after examination by Seller is
proven to be defective.
Replacement equipment and/or parts will be provided at no cost to
Purchaser, F.O.B. Seller's Plant. Failure of Seller to successfully replace
any nonconforming equipment or parts shall not cause the remedy
established hereby to fail of its essential purpose.
1.3.3 Exclusion of Consequential Damage
Purchaser specifically understands and agrees that under no
circumstances will seller be liable to purchaser for economic, special,
incidental or consequential damages or losses of any kind whatsoever,
including but not limited to, loss of anticipated profits and any other loss
caused by reason of nonoperation of the goods. This exclusion is
applicable to claims for breach of warranty, tortious conduct or any other
cause of action against seller.
1-6
2.1. Instrument Overview.
1
LEDs (2 red "Alarm", 1 green "Safe")
9
▼ Button
2
MSA LINK Communication
10
Belt Clip (Altair 5 only)
3
Sensor Inlets (Diffusion version)
11
Charging Port
4
Pump Cap (Pump version)
12
Screws
5
Horn
13
Pump inlet
6
Display
14
Filter
7
▲ Button
15
Charge Status LED
8
Button
The instrument monitors gases in ambient air and in the workplace.
The ALTAIR 5 is available with a maximum of four sensors, which can
display readings for five separate gases (one Dual Toxic Sensor
provides both CO and H2S sensing capabilities in a single package).
The detector is available in a diffusion or pumped configuration. When
using the pumped instrument, do not remove the pump cap
(item 4 on FIGURE 2-1).
Figure 2-1. Instrument View
2-1
The ALTAIR 5 IR is available with a maximum of five sensors, which
can display readings for six separate gases (one Dual Toxic Sensor
provides both CO and H2S sensing capabilities in a single package).
The ALTAIR 5 and ALTAIR 5 IR Detectors are available with either a
monochrome or color display.
While the instrument can detect up to 25 % oxygen in ambient air, it is
approved for use only up to 21 % oxygen.
The alarm levels for the individual gases are factory-set but can be
changed through the instrument Setup Menu. These changes can also
be made through MSA LinkTM Software. Ensure that the latest version of
the MSA Link software has been downloaded from MSA’s website
www.msanet.com. It is recommended that after making changes using
MSA Link, the unit should be turned OFF and ON.
2.2 User Interfaces
Instrument operation is dialog driven from the display with the aid of the
three function buttons (see FIGURE 2-1).
2.2.1 Button Definitions
The ALTAIR 5 and ALTAIR 5 IR units use three buttons for user
operation. Each button can function as a "soft key", as defined
immediately above the button.
BUTTON
DESCRIPTION
The
Button is used to turn instrument ON or OFF and to confirm
user action selections.
▼
The ▼ Button is used to page down through data screens or to
decrease the values in Setup mode.
▲
The ▲ button is used to reset peak, STEL TWA and alarms
(where possible) or perform calibration in measuring mode. It is also used
as page up or to increase the values in Setup mode.
When the ▲ button and the ▼ button are pressed simultaneously while
in normal measure mode, the Setup Mode can be entered after the
Password is confirmed.
2-2
2.2.2 LED Definitions
LED
DESCRIPTION
RED
The Alarm LEDs are visual indications of an alarm condition or any type of
(ALARM)
error in the instrument
GREEN
The Safe LED flashes once every 15 seconds to notify the user that the
(SAFE)
instrument is ON and operating under the conditions defined below:
• The green SAFE LED is enabled
• Combustible reading is 0% LEL or 0% VOL
• Oxygen (O2) reading is 20.8%
• All other sensor readings are 0 ppm
• No gas alarms are present (low or high)
• Instrument is not in Low Battery warning or alarm
• STEL and TWA readings are 0 ppm.
This option can be turned OFF through the MSA Link software
2.2.3 Alarms
The instrument is equipped with multiple alarms for increased user
safety:
2.3.3.1 Vibrating Alarm
The instrument vibrates when any alarm condition is active. This can be
turned OFF through the SETUP- ALARM OPTIONS menu
(Section 3.4.2).
2.2.3.2 Horn
This unit is equipped with an audible alarm. The horn can be turned
OFF through the SETUP- ALARM OPTIONS menu (Section 3.4.2).
2.3.3.3 InstantAlert™ Alarm
The InstantAlert exclusive feature allows the user to manually activate
an audible alarm to alert those nearby to potentially dangerous
situations. Holding the ▼ button for approximately 5 seconds while in
Normal Measure Mode activates the InstantAlert alarm.
2.3.3.4 Motion Alert Alarm
If Motion Alert is turned ON (+ = ON) and activated (see Section 3.4.2),
the instrument activates a "Man Down" alarm if motion is not detected
within 30 seconds. The Alarm LEDs flash, and the horn activates with
an increasing audible frequency.
2.2.3.5 Stealth Mode
Stealth Mode disables the visual, audible and vibrating alarms. MSA
recommends that this feature be left in its default "OFF" state. This can
be turned ON through the SETUP - INSTRUMENT OPTIONS menu
(Section 3.4.3).
The message "Alarms OFF" flashes on the monochrome display when
Stealth mode is ON. On the color display, all three alarm icons are
shown as OFF (
)
2.2.4 Backlight
The backlight automatically activates when any front panel button is
pressed and remains ON for the duration of user-selected timeout. This
duration can be changed using the Setup- Instrument Mode (Section
3.4.3) or through MSA Link software.
2.2.5 Confirmation Beep
This confirmation beep activates every 30 seconds by momentarily
beeping the horn and flashing the alarm LEDs under the following
conditions:
•
Confirmation beep is enabled
•
Instrument is on normal Measure Gases page
•
Instrument is not in battery warning
•
Instrument is not in gas alarm.
2-4
2.3. On-Screen Indicators
2.3.1 Monochrome Display
1
Gas Type
5
Battery Condition
2
Current Time
6
Motion Alert ON or Successful Cal Indicator
3
"Soft Key" ▼ Indicator
7
"Soft Key" ▲ Indicator
4
Gas Concentration
Combustible sensor reading
Oxygen sensor reading
CO sensor reading
H2S sensor reading
Toxic (fifth) sensor reading if installed
Indicates the name of next page
Initiates Calibration
IR Sensor - 6th sensor if installed
On a monochrome display, a message appears every 30 seconds
if the Vibration, Horn or LED alarms are turned OFF.
Figure 2-2. Monochrome Display
CAL
METH
2.3.2 Color Display
1
"Soft Key" ▼ Indicator
8
Vibration Alarm OFF
2
Gas Numeric Value
9
Horn OFF
3
Gas Concentration Units
10
LED OFF
4
Current Time
11
Battery Condition
5
Motion Alert (+ = ON)
12
Gas Type
6
Wireless ON
13
"Soft Key" ▲ Indicator
7
Combustible Gas type calibration
2.3.2.1 Wireless-Equipped Units
Color display units can be ordered with a wireless capability that allows
communication with the alphaScout Personal Network. This Network
capability is almost exclusively used in International markets. Contact
MSA for additional information.
2.3.3 Battery Life Indicator
The battery condition icon continuously displays in the upper right-hand
corner of the color display and in the lower right-hand corner of the
monochrome display. A bar represents the charging level of the battery.
The nominal run-time of the instrument (COMB, O2, H2S, CO, with
pump and monochrome display) at room temperature is 14 hours.
Actual run-time varies, depending on ambient temperature and alarm
conditions.
Figure 2-3. Color Display
2-6
Low Battery Warning
If battery warning alarm activates while using the
instrument, leave the area immediately as the end of
battery life is approaching. Failure to follow this
warning can result in serious personal injury or death.
The battery icon blinks when battery life is nearly expired. The Alarm
LEDs flash once per minute under this condition.
The duration of remaining instrument operation during the low battery
warning depends on ambient temperatures, battery condition and
alarm status.
When the instrument goes into battery warning the:
•
battery life indicator continuously blinks
•
alarm sounds and alarm LEDs flash once per minute
•
Safe LEDs no longer flash
•
instrument continues to operate until it is turned OFF or battery
shutdown occurs
Figure 2-4. Battery Warning
" WARNING
2-7
If Battery Alarm displays, stop using the instrument
as it no longer has enough power to indicate
potential hazards, and persons relying on this
product for their safety could sustain serious
personal injury or death.
The instrument goes into battery shutdown mode 60 seconds before
final shutdown (when the batteries can no longer operate the
instrument):
•
"BATTERY ALARM” and battery life indicator flash on the display
•
Alarm sounds
•
Alarm LEDs flash
•
No other pages can be viewed; after approximately one minute, the
instrument automatically turns OFF.
When battery shutdown condition occurs:
1. Leave the area immediately.
2. Recharge or replace the battery pack.
2.3.4 Battery Charging
Risk of explosion: Do not recharge instrument in hazardous
area.
Use of any charger, other than the Charger supplied with the
instrument, may damage or improperly charge the batteries.
" WARNING
Figure 2-5. Battery Shut Down
" WARNING
2-8
The charger is capable of charging a completely depleted pack in
less than six hours in normal, room-temperature environments.
NOTE: Allow very hot or cold instruments to stabilize for one hour
at room temperature before attempting to charge.
•
Minimum and maximum ambient temperature to charge the
instrument is 10°C (50°F) and 35°C (95°F).
•
For best results, charge the instrument at room temperature.
To Charge the Instrument
•
Firmly insert the charger connector into the charge port on the
back of the instrument.
•
An LED in the battery pack is used to indicate the charge status.
•
Red = charging
•
Green = charged
•
Amber = fault
•
If a problem is detected during charging (LED turns amber),
disconnect the charger momentarily to reset the charge cycle.
•
The charger MUST BE DISCONNECTED for the unit to operate.
2.4. Viewing Additional Pages
The Main Screen appears at instrument turn-ON.
Additional displays can be viewed by pressing the ▼ button to move to
the screen as indicated by the "soft key".
(For the monochrome display, the name of the page is displayed; for the
color display, it is represented by an icon.)
The sequence of pages are as follows and are described below:
2-9
2.4.1 Peak Readings (PEAK page)
MONOCHROME DISPLAY
COLOR DISPLAY
PEAK
This page shows the highest levels of gas recorded by the instrument
since turn-ON or since peak readings were reset.
To reset the peak readings:
1. Access the PEAK page.
2. Press the ▲ button.
2.4.2 Minimum Readings (MIN page)
MONOCHROME DISPLAY
COLOR DISPLAY
MIN
This page shows the lowest level of oxygen recorded by the instrument
since turn-ON or since the MIN reading was reset.
To reset the MIN reading:
1. Access the MIN page.
2. Press the ▲ button.
This page can be de-activated through MSA Link.
2-10
2.4.3 Short Term Exposure Limits (STEL page)
If the STEL alarm activates, leave the contaminated area
immediately; the ambient gas concentration has reached the
preset STEL alarm level. Failure to follow this warning will
cause over-exposure to toxic gases and persons relying on
this product for their safety could sustain serious personal
injury or death.
MONOCHROME DISPLAY
COLOR DISPLAY
STEL
This page shows the average exposure over a running 15-minute
period.
When the amount of gas detected by instrument is greater than the
STEL limit:
•
Alarm sounds
•
Alarm LEDs flash
•
“STEL ALARM” message flashes.
To reset the STEL:
1. Access the STEL page.
2. Press the ▲ button.
The STEL alarm is calculated over a 15-minute exposure.
STEL calculation examples:
Assume the instrument has been running for at least 15 minutes:
•
15-minute exposure of 35 ppm:
(15 minutes x 35 PPM)
15 minutes
= 35 PPM
•
10-minute exposure of 35 ppm and
5 minute-exposure of 15 ppm:
(10 minutes x 35 PPM) + (5 minutes x 15 PPM)
15 minutes
= 25 PPM
This page can be de-activated through MSA Link.
" WARNING
2-11
2.4.4 Time Weighted Average (TWA Page)
If the TWA alarm activates, leave the contaminated area
immediately; the ambient gas concentration has reached the
preset TWA alarm level. Failure to follow this warning will
cause over-exposure to toxic gases and persons relying on
this product for their safety could sustain serious personal
injury or death.
This page shows the average exposure over 8 hours since the
instrument was turned ON or the TWA reading was reset. When the
amount of gas detected by the instrument is greater than the TWA limit:
MONOCHROME DISPLAY
COLOR DISPLAY
TWA
•
Alarm sounds
•
Alarm lights flash
•
"TWA ALARM" message flashes.
To Reset the TWA Readings
1. Access the TWA Page.
2. Press the ▲button.
The TWA alarm is calculated over an eight-hour exposure.
TWA Calculation Examples:
•
1-hour exposure of 50 ppm:
(1 hour x 50 PPM) + (7 hours x 0 PPM)
8 hours
= 6.25 PPM
•
4-hour exposure of 50 ppm and
4-hour exposure of 100 ppm:
(4 hours x 50 PPM) + (4 hours x 100 PPM)
8 hours
= 75 PPM
•
12-hour exposure of 100 ppm:
(12 hours x 100 PPM)
8 hours
= 150 PPM
This page can be de-activated through MSA Link.
" WARNING
2-12
2.4.5 Date Display
Current date appears on the display in the format: MM-DD-YY.
2.4.6 Last Cal Page
Displays the instrument's last successful calibration date in the format:
MM-DD-YY. This page can be de-activated through MSA Link or the
Setup- Cal Options page.
2.4.7 Cal Due Page
Displays the days until the instrument's next calibration is due
(user selectable). This page can be de-activated through MSA Link or
the Setup- Cal Options page.
2.4.8 Motion Alert Activation Page
When the Motion Alert feature is active, the + symbol appears. The
instrument enters pre-alarm when no motion is detected for 20 seconds.
This condition can be cleared by moving the instrument. Motion Alert is
turned OFF each time the unit is powered OFF.
After 30 seconds of no motion, the full Motion Alert alarm is triggered.
This alarm can only be cleared by pressing the ▲ button. This page
displays if it was selected in Setup Mode.
To activate or deactivate the Motion Alert feature, press the ▲ button
while the Motion Alert Activation page is displayed.
2.5 Sensor Missing Alarm
The instrument enters the Sensor Missing alarm if the instrument
detects that an enabled sensor is not properly installed.
For the toxic sensors, the Sensor Missing feature is checked at
instrument turn-ON. The Sensor Missing function is not available for
HCN, Chlorine and Chlorine Dioxide sensors. The same functionality is
achieved by performing the recommended daily Calibration Check and
verifying that gas is detected.
The combustible and IR sensors are continually monitored.
2-13
If a sensor is detected as missing, the following occurs:
•
"SENSOR MISSING" flashes on the display.
•
The sensor detected as missing flashes on the display.
•
Alarm sounds and Alarm LEDs flash.
•
The alarm can be silenced by pressing the ▲ button; no other
pages can be viewed.
The instrument should be powered down and the sensor re-seated.
2.6 Monitoring Toxic Gases
The instrument can monitor the concentration of toxic gases in ambient
air. Which toxic gases are monitored depends on the installed sensors.
The instrument displays the gas concentration in parts per million (PPM)
or mg/m3 on the Measuring page.
If an alarm activates while using the instrument, leave the
area immediately.
Remaining in the area under such circumstances can cause
serious personal injury or death.
The instrument has four gas alarms:
•
HIGH Alarm
•
LOW Alarm
•
STEL Alarm
•
TWA Alarm
" WARNING
2-14
If the gas concentration reaches or exceeds the alarm setpoint or the
STEL or TWA limits, the:
•
alarm message displays and flashes in combination with the
corresponding gas concentration:
•
backlight turns ON
•
alarm sounds (if active)
•
alarm LEDs flash (if active)
•
vibrating alarm triggers (if active).
2.7 Monitoring Oxygen Concentration
The instrument monitors oxygen concentration in ambient air. The alarm
setpoints can be set to activate on two different conditions:
•
Enriched - oxygen concentration > 20.8 % or
•
Deficient - oxygen concentration < 19.5 %.
While the instrument can detect up to 25% oxygen in the ambient air, it
is approved for use only up to 21% oxygen-content.
If an alarm activates while using the instrument, leave the
area immediately.
Remaining in the area under such circumstances can cause
serious personal injury or death.
" WARNING
Figure 2-6. Alarm Conditions (High Alarm shown)
2-15
When the alarm setpoint is reached for either of the above conditions:
•
the alarm message displays and flashes in combination with the
oxygen gas concentration
•
backlight turns ON
•
alarm sounds (if active)
•
alarm LEDs flash (if active)
•
vibrating alarm triggers (if active).
The LOW alarm (oxygen deficient) is latching and will not reset even
when the O2 concentration rises above the LOW setpoint. To reset the
alarm, press the ▲ button. If the alarm is latching, the ▲ button
silences the alarm for five seconds. Alarms can be made latching or
unlatching via MSA Link
False oxygen alarms can occur due to changes in barometric pressure
(altitude), humidity, or extreme changes in ambient temperature.
It is recommended that an oxygen calibration be performed at the
temperature and pressure of use. Be sure that the instrument is in
known fresh air before performing a calibration.
2.8. Monitoring Combustible Gases
The instrument can be equipped with a catalytic combustible sensor
that detects a variety of combustible gases up to 100% LEL and
displays the reading as either %LEL or %VOL. The Altair 5IR can
contain both a catalytic combustible sensor and an IR combustible
sensor, if equipped. The IR sensor can display the reading in either
%LEL or %VOL, depending on the sensor chosen.
If an alarm activates while using the instrument, leave the
area immediately.
Remaining in the area under such circumstances can cause
serious personal injury or death.
The instrument has two alarm setpoints:
•
HIGH Alarm
•
LOW Alarm
" WARNING
2-16
If the gas concentration reaches or exceeds the alarm setpoint,
the instrument:
•
alarm message displays and flashes in combination with the
corresponding gas concentration:
•
backlight turns ON
•
alarm sounds (if active)
•
alarm LEDs flash (if active)
•
vibrating alarm triggers (if active).
When gas reading exceeds 100% of the Lower Explosive Limit (LEL),
the instrument enters a LockalarmTM state and displays “XXX” in place of
the actual reading.
A combustible gas reading of "XXX” indicates the
atmosphere may be above 100 % COMB or 5.00 %vol CH4
and an explosion hazard exists. Leave the contaminated
area immediately.
Certain IR sensors use a calibration gas >100% LEL. When the
calibration sequence shown in section 3.8.2 is followed, the combustible
LEL sensor will display the "XXX" reading. This state can only be reset
by turning the instrument OFF and then ON again.
" WARNING
2-17
Instrument operation is dialog driven from the display with the aid of the
three Function buttons (see Section 2.2.1).
3.1. Environmental Factors
A number of environmental factors may affect the gas sensor readings,
including changes in pressure, humidity and temperature. Pressure
and humidity changes affect the amount of oxygen actually present in
the atmosphere.
Pressure Changes
If pressure changes rapidly (e.g., stepping through airlock) the oxygen
sensor reading may temporarily shift, and possibly cause the detector to
go into alarm. While the percentage of oxygen may remain at or near
20.8 vol%, the total amount of oxygen present in the atmosphere
available for respiration may become a hazard if the overall pressure is
reduced by a significant degree.
Humidity Changes
If humidity changes by any significant degree (e.g., going from a dry, air
conditioned environment to outdoor, moisture laden air), oxygen
readings can be reduced by up to 0.5 %, due to water vapor in the air
displacing oxygen.
The oxygen sensor has a special filter to reduce the effects of humidity
changes on oxygen readings. This effect will not be noticed
immediately, but slowly impacts oxygen readings over several hours.
Temperature Changes
The sensors are temperature compensated. However, if temperature
shifts dramatically, the sensor's reading may shift temporarily.
3-1
3.2. Turning ON and Fresh Air Setup
Instrument operation is dialog driven from the display with the aid of the
three Function buttons (see Section 2.2.1). For more information, see
the flow charts in Section 7.
Turn the instrument ON with the
button.
•
The instrument performs a self test:
During the self test, the instrument checks alarm LEDs, audible
alarm, vibrating alarm, and for missing sensors.
In case of a missing sensor, the instrument displays the Sensor
Missing screen (see section 2.5) and alarms until it is turned OFF.
Otherwise, the turn-ON sequence continues.
•
The instrument displays:
•
Startup logo
•
Software version, instrument name and serial number,
company name, department, and user names
•
Sampling safety test (if equipped with an integral pump)
•
Combustible gas type and sensor units
(monochrome display only)
•
Low Alarm setpoints
•
High Alarm setpoints
•
STEL Alarm setpoints (if enabled)
•
TWA Alarm setpoints (if enabled)
•
Settings for calibration cylinder
•
Current date
•
Last calibration date (if enabled)
•
CAL due date. If the calibration due date is enabled,
the message "CAL DUE; X DAYS" appears on the
instrument display.
•
X = the number of days until a calibration is due, user
selectable for 1 to 180 days.
If the number of days until calibration is due reaches 0, an
alert occurs and "CAL DUE NOW" displays.
•
Press the ▲ button to clear the alert.
3-2
Sensor warm-up period
•
Fresh Air Setup option (if enabled).
Refer to flowchart in Appendix, Section 7.1
3.2.1 Sampling Safety Test
This test is only on units with software version 1.3 or higher for
instruments that are equipped with an integral pump.
At startup, an alarm (visual, audible, and vibrating) is triggered and the
customer is prompted to block the pumps/sampling system of the
instrument within 30 seconds.
When the instrument detects a pump flow block, it displays a PASS
message. The startup sequence resumes.
If the instrument does not detect a pump flow block, it displays an
error message.The instrument shuts OFF after the customer
acknowledges this message by pressing the ▲ button. If this occurs
check your sampling system and contact MSA as needed.
Users can check the operation of the sampling system anytime during
operation by blocking the sampling system to generate a pump alarm.
Do not use the pump, sample line, or probe unless the pump
alarm activates when the flow is blocked. Lack of an alarm is
an indication that a sample may not be drawn to the sensors,
which could cause inaccurate readings.
Failure to follow the above can result in serious personal
injury or death.
Never let the end of the sampling line touch or go under any
liquid surface. If liquid is drawn into the instrument,
readings will be inaccurate and instrument could be
damaged. We recommend the use of an MSA sample probe
containing a special membrane filter, permeable to gas but
impermeable to water, to prevent such an occurrence.
" WARNING
3-3
3.2.2 Fresh Air Setup (FAS)
The Fresh Air Setup (FAS) is for ZERO calibration of the instrument.
The FAS has limits. If a hazardous level of gas is present,
the instrument ignores the FAS command and the instrument
alarm activates.
NOTE: The Fresh Air Setup does not apply to the CO2 sensor.
Do not activate the Fresh Air Setup unless you are certain
you are in fresh, uncontaminated air; otherwise, inaccurate
readings can occur, which can falsely indicate that a
hazardous atmosphere is safe. If you have any doubts as to
the quality of the surrounding air, do not use the Fresh Air
Setup feature. Do not use the Fresh Air Setup as a substitute
for daily calibration checks. The calibration check is
required to verify span accuracy. Failure to follow this
warning can result in serious personal injury or death.
The instrument displays a blinking "FRESH AIR SETUP?", prompting
the user to perform a Fresh Air Setup:
1. Press the ▲ button to bypass the Fresh Air Setup.
•
The Fresh Air Setup is skipped
•
The instrument goes to the Measuring page (Main page).
2. Press the ▼ button to perform the Fresh Air Setup.
•
The instrument starts the FAS sequence and displays the
FAS screen
•
A progress bar shows the user how much of the FAS
is completed
Figure 3-1. Fresh Air Setup
" WARNING
3-4
At the end of the FAS, the instrument displays
"FRESH AIR SETUP PASS" or
"FRESH AIR SETUP FAIL".
If the FAS fails, perform a zero calibration per Section 3.8.1.
3.3. Measurement Mode (Normal Operation)
The following Options pages can be executed from the Normal
Operation screen (for further information see Section 2.4 and 7.5):
Peak Page*
shows the peak readings for all sensors
Min Page
shows the minimum reading for the oxygen sensor
STEL Page*
shows the calculated STEL readings of the instrument
TWA Page*
shows the calculated TWA readings of the instrument
Date Page
shows the date
Last Cal
shows the date of the last calibration. If the instrument does not
Date
have a valid calibration, it will display "LAST CAL INVALID"
Cal Due*
shows the set date for the next calibration
Motion AlertTM
+
if the Motion Alert Feature is activated
* The display of these pages can be de-activated through MSA Link (per Section 2.4)
3.4 Instrument Setup
The instrument allows the user to access and modify the following
parameters through direct button interface:
•
Calibration Options
•
Alarm Options
•
Instrument Options.
These menus can be accessed only from the Measure page by
pressing and holding the ▲ and ▼ buttons simultaneously until
prompted for a password.
The operation is as follows:
1. Turn the instrument ON and wait until the Measure page appears.
2. Press the ▲ and ▼ buttons simultaneously.
•
The default password is "672".
3-5
3. Enter the first digit by pressing the ▲ or ▼ button, and confirm
with the
button.
•
The cursor jumps to the second digit.
4. Enter the second and third digits.
•
Incorrect password:
instrument returns to the Main page.
•
Correct password:
user can set the instrument options.
The password can be changed with a PC through the
MSA Link software.
The following Options are available by pressing the ▲ and ▼ buttons:
•
Calibration Options - see Section 3.4.1
•
Alarm Options - see Section 3.4.2
•
Instrument Options - see Section 3.4.3.
3.4.1 Calibration Options
Figure 3-3. Calibration Options
Figure 3-2. Password Screens
3-6
The Calibration Options menu allows the user to:
•
modify the calibration cylinder settings
(CYLINDER SETUP)
•
enable/disable calibration due date and set the number of days
(CAL DUE OPTIONS)
•
enable/disable the option to show the last cal date at turn-ON
(LAST CAL DATE)
•
enable/disable the option for password-protected calibration
(CAL PASSWORD).
1. Press the:
•
▼button go to next page
•
▲button to go previous page
•
button to enter setup.
Setting Calibration Cylinder
This option has a dialog similar to the span calibration dialog.
The display shows all active sensors.
1. Press the
button to enter setup.
•
The screen for the first calibration cylinder displays.
2. Press the:
•
▼ or ▲ button to change the value.
•
button to confirm the setup.
With this confirmation, the instrument automatically moves to the
next cylinder setting.
3. Repeat the sequence for changing the required settings for all
necessary gas values.
•
After the last setting is performed, the instrument returns to the
Calibration Options menu.
3-7
Setting Cal Due Options
1. Press the
button to enter setup.
2. Press the ▼ or ▲ button to enable/disable this option.
3. Press the
button to confirm.
4. After confirmation, the instrument prompts the user to enter the
number of days for the reminder.
5. Change number of days by pressing the ▼ or ▲button.
6. Press the
button to go to the next menu.
Setting Last Cal Date
1. Press the
button to enable/disable this option.
2. Press the ▼button to go to the next page.
3. Press the ▲ button to go to the previous page.
Setting Calibration Password
1. Press the
button to enable/disable this option.
2. Press the to go to the next page.
3. Press the ▲button to go to the previous page.
Back To Main Menu
1. Press the
button to go to Instrument Setup Menu
•
The Cal Options screen displays
2. Press the ▼ button to go to the next (Alarm options) or the
▲ button to exit the Setup menu.
3.4.2 Alarm Options
Figure 3-4. Alarm Options
3-8
The Alarm Options Menu allows user to:
•
enable/disable the vibrating alarm
•
enable/disable the audible alarm (Horn)
•
enable/disable the Alarm LEDs
•
enable/disable the Motion Alert Selection page
•
set Sensor Alarms.
Press the:
•
▼ button go to next page
•
▲ button to go previous page
•
button to enter setup.
Setting Vibrating Alarm
1. Press the
button to enable/disable this option.
Setting Horn
1. Press
button to enable/disable this option.
Setting LEDs
1. Press the
button to enable/disable this option.
Setting Motion Alert
1. Press the
button to display the Motion Alert selection page.
2. To activate Motion Alert, from the Main Menu, press the ▼button
until the Motion Alert selection page is shown.
•
ALARM OFF displays.
3. Select ▲button to toggle to ON; then, press
button.
•
Unit returns to Main menu and Motion Alert is active (+ = ON).
Setting Sensor Alarms
This page allows user to modify the preset alarm values for:
•
LOW Alarm
•
HIGH Alarm
•
STEL Alarm
•
TWA Alarm.
The factory-set alarm values are shown in Section 5.2.
1. Press the
button to enter Sensor Alarm setup.
3-9
LOW Alarm Setup screen displays.
2. Press the:
•
▼ button to abort the operation or
•
▲ button to go to next alarm setup or
•
button to change the alarm setpoints.
• Alarm Value for the first Sensor displays.
3. Set values for Sensor Alarm by pressing the▼ or ▲ button.
4. Press the
button to confirm set value.
5. Repeat setting for all other sensors.
6. Press the ▲ button to return to the Alarm Options menu.
7. Repeat setting for all other alarm types.
60% LEL or 3.0% volume of methane is the maximum High Alarm
setpoint that can be programmed by the user for the combustible
LEL sensor.
The combustible alarm can be turned OFF by the user in the
instrument setup.
•
When the combustible alarm is turned OFF, the only indicator to
Figure 3-6. Sensor Alarm Setup
Figure 3-5. Sensor Alarm Setup
3-10
the user that the combustible alarm is turned OFF occurs during
instrument power-up, when a startup screen indicates that the
combustible alarm is turned OFF.
•
When turned ON, the combustible high alarm is latching. The
combustible alarm can be silenced momentarily by pressing the
button. However, if the gas concentration causing the alarm is still
present, the unit goes back into alarm.
3.4.3 Instrument Options
The Instrument Options Menu allows the user to modify different
instrument options:
•
Sensor Setup (enable/disable the channel)
•
Language
•
Time and Date Setup
•
Datalog Intervals
•
Stealth Mode
•
Confirmation Beep
•
Display Contrast (monochrome only)
•
Backlight Options.
1. Press the:
•
▼ button to go to next page
•
▲ button to go previous page
•
button to enter setup.
Figure 3-7. Instrument Options
3-11
Setting Sensor Options
1. Press the
button to enter setup.
•
The following screen displays:
2. Press the ▼ button to select sensor and the
button to
make changes.
•
The sensor information is displayed, and the sensor can be
enabled or disabled.
NOTE: Other operations such as changing the gas type (Methane,
Butane, Propane, etc. for the combustible sensor) and
units (ppm to mg/m3) are only possible using the
MSA Link software.
3. Change status by pressing the ▼ or ▲ button.
4. Press the
button to confirm and advance to the next screen
(next sensor).
5. Perform the sequence for all other sensors
•
After setting up the last sensor, the instrument
goes to the next Setup Page.
Language Setup
This option is for setting the language of the instrument.
1. Press
button to enter Setup.
2. Change language by pressing the ▼ or ▲ button.
3. Confirm by pressing the
button.
•
The instrument goes to the next Setup Page.
Time and Date Setup
The instrument first prompts the user to set the time;
then, it prompts for the date.
Figure 3-8. Sensor Options
3-12
NOTE: The time can be set for either regular AM/PM or military time
through MSA Link software. AM/PM time is the default setting.
1. Press the
button to enter setup.
2. Change hours by pressing the ▼ or ▲ button.
Increment through noon for PM hours.
3. Confirm by pressing the
button.
4. Change minutes by pressing the ▼ or ▲ button.
5. Confirm by pressing the
button.
•
The instrument goes to the Set Date Page.
6. Change month, date, and hours by pressing the ▼ or ▲ button:
7. confirming by pressing the
button.
•
The instrument goes to the next Setup Page.
Setting Datalog Intervals
This option is for setting the intervals at which all the readings
are logged.
1. Press the
button to enter Setup.
2. Change interval by pressing the ▼ or ▲ button.
3. Confirm by pressing the
button.
•
The instrument goes to the next Setup Page.
Setting Stealth Mode
Stealth mode disables the visual, audible and vibrating alarms.
1. Press the
button to change mode (ON/OFF).
2. Press the ▼ button to go to the ▲ button to return to the
previous page.
Setting Confirmation Beep
1. Press the
button to change mode (ON/OFF).
2. Press the ▼ button to go to the ▲ button to return to the
previous page.
Setting Contrast (monochrome display)
1. Press the ▼ or ▲ button to adjust the contrast levels.
2. Press the
button to confirm the contrast level.
3-13
1. Press the
button to enter Setup.
2. Change option by pressing the ▼ or ▲ button.
3. Press the
button to enter.
4. Change timeout by pressing the ▼ or ▲ button.
5. Press the
button to confirm timeout.
Back To Main Menu
There are three options at this point:
•
the ▼ button – to go to the Sensor Options menu
•
the ▲ button – to go to the Previous Setup page in the
Instrument Options menu
•
the
button – to go to the Instrument Options menu.
3.5. MSA Link Operation
Connecting Instrument to PC
1. Switch ON the ALTAIR 5 instrument and align the Datalink
Communication port on the ALTAIR 5 to the IR interface of the PC.
2. Start the MSA Link software on the PC and start the connection by
clicking the connect icon.
See MSA Link documentation for detailed instructions.
3.6. Function Tests on the Instrument
Alarm Test
1. Turn ON the instrument. The user should verify that:
•
alarm LEDs flash
•
horn sounds briefly
•
vibrating alarm triggers briefly.
3.7. Calibration Check
The calibration check is simple and should only take about one minute.
Perform this calibration check before each day’s use for each
installed sensor.
3-14
For Altair 5IR units with %volume IR sensors, the following gas levels
should not be exceeded when used for the daily Calibration Check:
•
IR Butane 25% vol
- 2% vol Butane Cal Check gas
•
IR Propane 25% vol
- 8% vol Propane Cal Check gas
•
IR Methane 25% vol
- 20% vol Methane Cal Check gas
•
IR Butane 100% vol
- 20% vol Butane Cal Check gas
•
IR Propane 100% vol
- 8% Propane Cal Check gas
•
IR Methane 100% vol
- 20% Methane Cal Check gas.
1. Turn ON the ALTAIR 5 or ALTAIR 5 IR Multigas Detector
in clean, fresh air.
2. Verify that readings indicate no gas is present.
3. Attach regulator (supplied with calibration kit) to the cylinder.
4. Connect tubing (supplied with calibration kit) to the regulator.
5. Attach other end of tubing to the instrument.
6. Open the valve on the regulator, if so supplied.
•
The reading of the ALTAIR 5 or ALTAIR 5 IR Multigas Detector
display should be within the limits stated on the calibration
cylinder or limits determined by your company.
•
If necessary, change cylinder to introduce other calibration.
•
If readings are not within these limits, the ALTAIR 5 Multigas
Detector requires recalibration (see Section 3.8).
If using the MSA GALAXY Test Stand for the Calibration (bump) check,
a check-mark appears on the ALTAIR 5 display for 24 hours following a
successful bump of all operational sensors.
3.8. Calibration
Span Calibration gases and Span Calibration values are shown in
Section 5.4, along with the appropriate MSA calibration
gas cylinder.
CSA requires (per 22.2 NO. 152) that the instrument's sensitivity be
tested before each day's use on a known concentration of methane
equivalent to 25 to 50% of full scale concentration. Accuracy must be
within 0 to +20% of actual. Correct accuracy by performing the
calibration procedure given in this manual.
3-15
The ALTAIR 5 instrument can be calibrated either manually by using this
procedure or automatically by using the GALAXYTM test stand. Refer to
Section 3.8.5 for additional GALAXY information.
NOTE: The GALAXY is not a CSA certified method of calibration.
Calibration must be performed by using a flow regulator with a flow rate
set to 0.25 liters per minute for a diffusion instrument. For instruments
with an integral pump, use of the demand regulators listed in Section
6.2 is recommended.
Attention! Special conditions with toxic gases!
Reactive toxic gases (chlorine, ammonia, chlorine
dioxide) have the property of diffusing into the rubber and
plastic tubes so that the volume of test gas available in the
instrument would no longer be sufficient to correctly
perform instrument calibration.
For this reason, when calibrating the instrument with
reactive toxic gases, certain prerequisites are required;
otherwise, incorrect calibration could result:
•
A special pressure regulator
•
The shortest possible connection tubes between the
pressure regulator and the instrument.
•
Connection tubes made from a material that does not absorb
the test gases (e.g., PTFE).
NOTE:
If using normal tubes and pressure regulators, expose them
to the required test gas for an extended time period. Keep
these materials dedicated for use with that test gas only;
do not use them for other gases.
For example: for chlorine, allow the entire contents of a test
gas cylinder to flow through the pressure regulator and
tubes before using to calibrate the instrument. Mark these
materials for use with chlorine only.
The phosphine (PH3) sensor used in the instrument is highly
cross-sensitive to hydrogen sulfide (H2S) gas. Do not use
calibration gas that contains H2S on an instrument that is
configured with a PH3 sensor.
" WARNING
3-16
3.8.1 Zero Calibration
1. Press the ▲ button for three seconds in the Normal Measurement
page.
•
ZERO screen displays
To skip the ZERO procedure and move directly to the calibration
span procedure, push the ▲ button. If no button is pushed for 30
seconds, the instrument prompts the user to perform a SPAN
calibration before the instrument returns to the Normal
Measurement page.
2. Press the ▲ button to confirm the ZERO screen, i.e., to execute
zero calibration.
•
LEDs flash
•
ZERO CALIBRATION" blinks
•
ZERO calibration starts
•
A progress bar shows the user how much of the calibration has
already been completed.
•
After the ZERO calibration is completed, the instrument
displays:
• "ZERO CALIBRATION PASS"
or
• "ZERO CALIBRATION FAIL".
3-17
• If the instrument passes the zero calibration, the SPAN
screen displays.
3.8.2 Span Calibration
To skip the Span calibration procedure, push the ▲ button.
If no button is pushed for 30 seconds, the instrument returns to the
Measuring mode without calibrating the instrument.
Because of the different combinations of gases, it is possible to perform
single-channel calibration for the additional toxic sensors, if installed.
1. The screen displays.
2. When calibrating with gases > 100% LEL, select the "Yes" option
to prompt "Span Calibration?" BEFORE applying gas to the
instrument. Otherwise, the LEL sensor reading could be shifted.
Following a successful calibration of >100% LEL IR sensors, the
unit must be turned OFF, then ON to clear the over-range alarm
"XXX" on the combustible LEL sensor.
3. Attach the calibration cap to the instrument (diffusion units only).
4. Connect one end of tubing to the cylinder regulator (supplied in the
calibration kit).
5. Noting the special exception in Step 2, connect the other end of the
tubing to the calibration cap or pump inlet.
6. Open the valve regulator (Fixed Flow regulators only).
3-18
7. Press the ▲ button to calibrate (span) the instrument.
•
LEDs flash
•
"SPAN CALIBRATION" blinks
•
SPAN calibration starts
•
A progress bar shows the user how much of the calibration has
already been completed
•
After the SPAN calibration is completed the instrument displays:
• "SPAN CALIBRATION PASS"
or
• "SPAN CALIBRATION FAIL"
•
The instrument returns to Measuring mode.
3.8.3 Finishing Calibration
1. Close the valve on the regulator.
2. Remove the calibration cap or tube from pump inlet.
The calibration procedure adjusts the span value for any sensor that
passes the calibration test. A check mark () displays for 24 hours after
a successful calibration. If the Horn Alarm is turned OFF, the calibration
does not appear on the color display.
Sensors that fail calibration are left unchanged. Since residual gas may
be present, the instrument may briefly go into an exposure alarm after
the calibration sequence is completed.
3-19
Following a successful calibration of >100% LEL IR sensors, the
unit must be turned OFF, then ON to clear the over-range alarm
"XXX" on the combustible LEL sensor.
3.8.4 Calibration Failure
If the instrument cannot calibrate one or more sensor(s), it goes into the
Calibration Failure page and remains in alarm for 10 seconds.
Sensors that could not be calibrated are indicated by dashed lines on
the concentration display.
If a sensor repeatedly fails calibration after the full calibration procedure
in this manual has been performed, replace the failed sensor.
3.8.5 Calibration with GALAXY Test System
The instrument can also be calibrated using the GALAXY Automated
Test System - contact MSA for a list of compatible gases and
concentrations. The instrument will return a on the display when all
sensors are successfully calibrated.
3.9 Instrument Shutdown
For instrument shutdown press and hold the
button.
•
The instrument:
•
displays a blinking "HOLD BUTTON FOR SHUTDOWN"
•
a progress bar showing the user how much longer to hold the
button to complete the shutdown.
Figure 3-9. Instrument Shutdown
3-20
If irregularities occur during operation, use the displayed error codes to
determine appropriate next steps.
Repair or alteration of the ALTAIR 5 or ALTAIR 5 IR Multigas
Detector, beyond the procedures described in this manual or
by anyone other than a person authorized by MSA, could
cause the instrument to fail to perform properly. Use only
genuine MSA replacement parts when performing any
maintenance procedures described in this manual.
Substitution or incorrect installation of components can
seriously impair instrument performance, alter intrinsic
safety characteristics or void agency approvals.
FAILURE TO FOLLOW THIS WARNING CAN RESULT IN
SERIOUS PERSONAL INJURY OR DEATH.
For all other errors the error code is shown, accompanied with the error
icon for the color display.
" WARNING
4-1
Temperature is below
Return to normal temperature range
-40 °F (-40 °C) or
and recalibrate.Contact MSA
above 167 °F (75 °C).
ERROR ADC
Sensor measurement error
Contact MSA
ERROR COMB
Combustible sensor
Contact MSA
power supply error
ERROR MEM
External Memory error
Contact MSA
ERROR PROG
Program error
Contact MSA
ERROR RAM
RAM error
Contact MSA
LOW
Battery warning
BATTERY
BATTERY
Battery is completely discharged.
ALARM
Shutdown sequence will begin
Instrument
Battery fully discharged
Remove from service as soon as
does not
possible and recharge or replace
turn ON
battery pack
MISSING
Sensor damaged or missing
Replace sensor
SENSOR
4.2 Verifying Pump Operation
Users can check operation of the sampling system any time during
operation by blocking the sampling system to generate a pump alarm.
Once gas readings display, plug the free end of the sampling line
or probe.
•
The pump motor shuts down and an alarm sounds.
•
PUMP ERROR flashes on the display.
4-2
When the pump inlet, sample line or probe is blocked, the pump alarm
must activate.
If the alarm does not activate:
1. Check the sample line and probe for leaks.
2. Once leak is fixed, recheck pump alarm by blocking the flow.
3. Press the ▲ button to reset the alarm and restart the pump.
Do not use the pump, sample line, or probe unless the pump
alarm activates when the flow is blocked. Lack of an alarm is
an indication that a sample may not be drawn to the sensors,
which could cause inaccurate readings.
Failure to follow the above can result in serious personal
injury or death.
Never let the end of the sampling line touch or go under any
liquid surface. If liquid is drawn into the instrument,
readings will be inaccurate and instrument could be
damaged. We recommend the use of an MSA sample probe
containing a special membrane filter, permeable to gas but
impermeable to water, to prevent such an occurrence.
During operation, a pump alarm may occur when the:
•
Flow system is blocked
•
Pump is inoperative
•
Sample lines are attached or removed.
To Clear Pump Alarm
1. Correct any flow blockage.
2. Press the ▲ button.
•
The Pump will now restart.
" WARNING
4-3
4.3 Replacing the Battery
Never replace the battery in a hazardous area.
1
Captive Screw
2
Battery Pack
1. Unscrew the two captive screws on the rear of the instrument.
2. Pull the battery pack out of the instrument by gripping the sides
and lifting it up and away from the instrument.
1
Battery Holder
Figure 4-2. Battery Replacement
Figure 4-1. Battery Replacement
" WARNING
4-4
3. For alkaline battery packs (ALTAIR 5 only):
a. Remove the battery holder circuit board from the instrument
b. Replace the three cells, using only those listed on the label.
• Be sure to observe proper polarity on the cells.
c. Place the battery holder circuit board back in the instrument
and reinstall the door.
d. Tighten the two screws.
4.4 Live Maintenance Procedure
- Replacing or Adding a Sensor
To add an additional toxic sensor to a unit, return it to an authorized
MSA Service Center for the necessary upgrades. Replacement of an
IR sensor should only be done at an authorized MSA Service Center.
Before handling the PC board, the user must be properly
grounded; otherwise, static charges could damage the
electronics. Such damage is not covered by the warranty.
Grounding straps and kits are available from electronics
suppliers.
Remove and reinstall sensors carefully, ensuring that the
components are not damaged; otherwise instrument
intrinsic safety may be adversely affected, wrong readings
could occur, and persons relying on this product for their
safety could sustain serious personal injury or death.
To add an LEL, O2, CO or H2S sensor to an instrument that is not already
equipped with a full array of sensors, remove the sensor plug from in
front of the formerly unused sensor housing.
While instrument case is open, do not touch any internal
components with metallic/conductive objects or tools.
Damage to the instrument can occur.
" CAUTION
" WARNING
" CAUTION
4-5
1. Verify that the instrument is turned OFF.
2. Remove the battery pack.
3. Remove the two remaining case screws, and remove the
case front.
4. Gently remove the sensor to be replaced.
•
Using fingers only, gently remove the tox, combustible or
oxygen sensor.
5. Carefully align the new sensor contact pins with the sockets on the
printed circuit board.
6. Press the new sensor into place.
•
Ensure groove in combustible sensor is aligned with notch in
sensor slot.
•
If a sensor is not to be installed, ensure that a sensor plug is
installed properly in its place.
7. If replacing sensor filter at this time:
•
Carefully peel off old filter, taking care not to damage the inside
of the case.
•
On new filter, peel off backing exposing adhesive. Note proper
orientation of filter and apply to inside of front case, adhesive
side against case.
•
Press filter into place, taking care not to damage filter surface.
8. Re-install the sensor gasket in the case front.
•
Ensure sensor gasket is properly installed.
9. Re-install the screws.
10. Go into the sensor setup and turn ON the sensor.
Figure 4-3. Sensor Replacement
4-6
11. After installation of new sensors, allow them to stabilize at
least 30 minutes before calibration. Calibrate instrument
before use.
Calibration is required after a sensor is installed; otherwise,
the instrument will not perform as expected and persons
relying on this product for their safety could sustain serious
personal injury or death.
4.5. Replacing the Filter
(for instruments with integral pump)
1. Turn OFF the instrument.
2. Unscrew the two captive screws from the clear filter cover on the
back of the instrument to access the filter.
3. Carefully lift out the O-ring and the filter disk.
4. Place the new filter disk in the recess.
5. Replace the O-ring in the recess.
6. If the instrument is used in dusty environments, the fibrous dust
filters supplied in Maintenance Kit P/N 10094829 should be used in
addition to the filter disk. Place the fibrous dust filter in the clear
filter cover.
If the instrument is configured for a reactive toxic gas (Cl2,
ClO2, NH3), use only the filters supplied in the Reactive Gas
Maintenance Kit P/N 10099533.
7. Re-install the clear filter cover on the back of the instrument.
4.6. Cleaning the Instrument
Clean the exterior of the instrument regularly using only a damp cloth.
Do not use cleaning agents.
" CAUTION
" WARNING
4-7
When not in use, store the instrument in a safe, dry place between
65°F and 86°F (18 °C and 30 °C). After storage, always recheck
instrument calibration before use. If not to be used in 30 days remove
battery pack.
4.8. Shipment
Pack the instrument in its original shipping container with suitable
padding. If the original container is unavailable, an equivalent container
may be substituted.
4-8
5. Technical Specifications/Certifications
5.1 Technical Specifications
WEIGHT
1 lb. (0.45 kg) for instrument with battery and clip (ALTAIR 5 unit)
WEIGHT (WITH
1.15 lb. (0.52 kg)
IR SENSOR)
DIMENSIONS
6.69" H x 3.49" W x 1.79" D -pumped, without belt clip (ALTAIR 5)
6.69" H x 3.49" W x 1.58" D diffusion, without belt clip (ALTAIR 5)
DIMENSIONS
6.68" H x 3.52" W x 1.92" D
WITH IR SENSOR
ALARMS
LEDs, audible alarm, vibrating alarm
ACOUSTIC
95 dB typical
ALARM
VOLUME
DISPLAYS
Monochrome or color
BATTERY TYPES Rechargeable Li ION battery (cannot be charged in Ex area)
Replaceable AA alkaline (ALTAIR 5 only)
CHARGING TIME ≤ 6 hours. Maximum safe area charging voltage; Um = 6.7 Volts DC
NORMAL
32 to 104°F (0 to 40°C)
TEMP RANGE
EXTENDED
-4 to +122°F (-20 to +50°C) MONOCHROME DISPLAY
TEMP RANGE
14°F – 122°F (-10° C to 50 °C) COLOR DISPLAY and HCN
- 4°F – 104°F (-20°C to 40°C) for instruments with
NH3 or ClO2 sensors
HUMIDITY
15 - 90% relative humidity, non-condensing;
RANGE
5 - 95% RH intermittent
ATMOSPHERIC
11.6 to 17.4 PSIA (80 to 120 kPA)
PRESSURE
RANGE
DUST & SPRAY
IP 65
PROTECTION
MEASURING
Combustible gas:
Catalytic or Infrared sensor
METHODS
O2 and Toxic gas:
Electrochemical or Infrared sensor
WARRANTY
See Section 1.3. Extended options available
MEASURING
H2S
CO
O2
Combustible
RANGE
0-200 ppm
0-1000 ppm
0-25 % Vol.
0-100% LEL
0-5.00% CH4
SO2
NO2
NH3
PH3
0-20.0 ppm
0-20 ppm
0-100 ppm
0-5.00 ppm
Cl2
ClO2
HCN
0-20.0 ppm
0-1.00 ppm
0-100 ppm
5-1
5.2 Factory-set Alarm Thresholds
NOTE: Units for Australia have slightly different alarm limits for COMB
and CO than shown below.
Sensor
LOW
HIGH
Setpoint
STEL
TWA
Default
alarm
alarm
Min
Max
Cal
Values
COMB
10% LEL
20% LEL
5% LEL
60% LEL
--1
--1
see
Section
5.6
CO
25 ppm
100 ppm
20 ppm
950 ppm
100 ppm
25 ppm
60 ppm
H2S
10 ppm
15 ppm
5 ppm
175 ppm
15 ppm
10 ppm
20 ppm
O2
19.5%
23.0%
5.0%
24.5%
--1
--1
15.0%
SO2
2.0 ppm
5.0 ppm
2.0 ppm
17.5 ppm
5 ppm
2.0 ppm
10 ppm
NO2
2.0 ppm
5.0 ppm
1.0 ppm
17.5 ppm
5.0 ppm
2.0 ppm
10 ppm
NH3
25 ppm
50 ppm
10 ppm
75 ppm
35 ppm
25 ppm
25 ppm
PH3
0.3 ppm
1.0 ppm
0.3 ppm
3.75 ppm
1.0 ppm
0.3 ppm
0.5 ppm
Cl2
0.5 ppm
1.0 ppm
0.3 ppm
17.5 ppm
1.0 ppm
0.5 ppm
10 ppm
ClO2
0.1 ppm
0.3 ppm
0.1 ppm
0.9 ppm
0.3 ppm
0.1 ppm
0.8 ppm
HCN
4.5 ppm
10.0 ppm
2.0 ppm
20.0 ppm
10 ppm
4.5 ppm
10 ppm
IR CO2
0.5% Vol
1.5%Vol
0.2%Vol
5%Vol
0.5% Vol
1.5%Vol
2.50%
(5%)
IRCO2
0.5%Vol
1.5%Vol
0.2%Vol
8%Vol
0.5%Vol
1.5%Vol
2.50%
(10 Vol%)
IR CO2
0.5%Vol
1.5%Vol
0.5%Vol
50%Vol
0.5%Vol
1.5%Vol
15%
(50 Vol%)
IR
15%LEL
29%LEL
9%LEL
100%LEL
--1
--1
35% LEL
Propane
(LEL)
IR
8% Vol
15% Vol
5%Vol
25%Vol
--1
--1
8%
Pro-pane
(25% Vol)
IR
21% LEL
50% LEL
11% LEL
100% LEL
--1
--1
36% LEL
Butane
(LEL )
IR
8% Vol
15% Vol
5% Vol
25% Vol
--1
--1
8%
Butane
(25% Vol)
IR
8% Vol
15% Vol
2.5% Vol
25% Vol
--1
--1
20%
Methane
(25% Vol)
1STEL and TWA not applicable for Combustible and oxygen gases.
5-2
NOTE: No alarm thresholds are possible for the 0-100% Volume
Methane, Propane and Butane IR sensors. In environments with
>100% LEL combustible gas present, units with a combustible
LEL sensor will be in a latching over-range alarm, and the
100% volume IR sensors will display the % volume gas reading.
5.3 Sensor Performance Specifications
PRIMARY SENSORS
Sensor
Display Range
Resolution
Reproducibility
Response Time
Com-
0-100% LEL
1% LEL or
Normal temp. range:
t(90) ≤30 sec
bustible
or 0-5% CH4
0.05% vol
<50% LEL:
3% LEL
(normal temp.)
(LEL)
CH4
50-100% LEL:
5% LEL
<2.5% CH4:
0.15% CH4
2.5-5.00% CH4: 0.25% CH4
Extended temp. range:
<50% LEL:
5% LEL
50-100% LEL:
8% LEL
<2.5% CH4:
0.25% CH4
2.5-5.00% CH4: 0.40% CH4
Oxygen
0-25% O2
0.1% O2
0.7% O2 for 2 - 25% O2
t(90) ≤ 30 sec
(normal temp.)
t(90) ≤ 3 min
(extended temp.)
Carbon
0-1000
1 ppm CO,
Normal temp. range:
t(90) ≤60 sec
Mon-
ppm CO
0-300 ppm CO:
(normal temp.)
oxide
±5 ppm or 10% of reading,
whichever is greater;
>300 ppm CO: ±15ppm
Extended temp. range:
±10 ppm or 20% of reading,
whichever is greater
Hydro-
0-200
1 ppm H2S,
Normal temp. range:
t(90) ≤60 sec
gen
ppm H2S
3-200
0-100 ppm H2S
(normal temp.)
Sulfide
ppm H2S
±2 ppm H2S
or 10 % of reading,
whichever is greater;
>100 ppm H2S:
±15 ppm H2S
Extended temp. range:
±10 ppm or 20% of reading,
whichever is greater
5-3
of the zero
of the
.
20°C
msd. value
t50
t90
CO2
0 - 5% Vol
0.01% Vol
≤ ± 0.05% Vol
≤ ±5%
≤12 s
≤36 s
CO2
0 - 10% Vol
0.01% Vol
≤ ± 0.1 % Vol
≤ ±4%
≤12 s
≤35 s
CO2
0 - 50% Vol
0.1% Vol
≤ ± 1.0 % Vol
≤ ±4%
≤14 s
≤36 s
CH4
0 - 25% Vol
0.2% Vol
≤ ± 0.5 % Vol
≤ ±5%
≤12 s
≤35 s
CH4
0 - 100% Vol
1% Vol
≤ ± 5 % Vol
≤ ±10%
≤11 s
≤34 s
C3H8
0 - 100% LEL
1% LEL
≤ ± 3 % LEL
≤ ±8%
≤11 s
≤32 s
C3H8
0 - 25% Vol
0.1% Vol
≤ ± 0.5 % Vol
≤ ±4%
≤12 s
≤35 s
C3H8
0 - 100% Vol
1% Vol
≤ ± 3 % Vol
≤ ±8%
≤12 s
≤36 s
C4H10
0 - 100% LEL
1% LEL
≤ ± 3 % LEL
≤ ±8%
≤11 s
≤32 s
C4H10
0 - 25 % Vol
0.1% Vol
≤ ± 0.5 % Vol
≤ ±4%
≤12 s
≤35 s
C4H10
0 - 100 % Vol
1% Vol
≤ ± 3 % Vol
≤ ±8%
≤12 s
≤36 s
ADDITIONAL TOXIC SENSORS
Sensor
Range
Resolution
SO2
0-20 ppm SO2
0.1 ppm SO2
NO2
0-20 ppm NO2
0.1 ppm NO2
NH3
0-100 ppm NH3
1 ppm NH3
PH3
0 - 5.0 ppm PH3
0.05 ppm PH3
HCN
0 - 30.0 ppm HCN
0.5 ppm HCN
Cl2
0-20.0 ppm Cl2
0.1 ppm Cl2
ClO2
0-1.00 ppm ClO2
0.02 ppm ClO2
5-4
5.4 Calibration Specification
See Section 6.1 for MSA gas cylinder part numbers.
SENSOR
ZERO
ZERO
SPAN
SPAN
CAL
GAS
CAL
CAL
CAL
DURA-
VALUE
GAS
VALUE
TION
(min-
utes)
COMB PENTANE
ROOM AIR
0
1.45% METHANE 58 LEL
1.5
COMB METHANE (0-5%v) ROOM AIR
0
2.5% METHANE
2.5%
1.5
COMB METHANE (4.4%v) ROOM AIR
0
1.45% METHANE 33 LEL
1.5
*COMB PROPANE (1.7%v) ROOM AIR
0
1.45% METHANE 58 LEL
1.5
*COMB PROPANE (2.1%v) ROOM AIR
0
1.45% METHANE 46 LEL
1.5
*COMB BUTANE
(1.4%v) ROOM AIR
0
1.45% METHANE 58 LEL
1.5
*COMB METHANE (5.0%v) ROOM AIR
0
1.45% METHANE 29 LEL
1.5
COMB HYDROGEN
ROOM AIR
0
1.45% METHANE 29 LEL
1.5
O2 Sensor
ROOM AIR
20.8% 15% O2
15.0%
1.5
CO Sensor
ROOM AIR
0
60 PPM CO
60 PPM
1.5
H2S Sensor
ROOM AIR
0
20 PPM H2S
20 PPM
1.5
SO2 Sensor
ROOM AIR
0
10 PPM SO2
10 PPM
3
Cl2 Sensor
ROOM AIR
0
10 PPM Cl2
10 PPM
4
NO2 Sensor
ROOM AIR
0
10 PPM NO2
10 PPM
4
NH3 Sensor
ROOM AIR
0
25 PPM NH3
25 PPM
4
PH3 Sensor
ROOM AIR
0
0.5 PPM PH3
0.5 PPM
4
HCN Sensor
ROOM AIR
0
10 PPM HCN
10 PPM
4
**ClO2 Sensor
ROOM AIR
0
0.8 PPM Cl2
0.8 PPM
4
IR CO2
300 PPM CO2
0.03% 2.5% CO2
2.50%
2
IR CO2
300 PPM CO2
0.03% 2.5% CO2
2.50%
2
IR CO2
300 PPM CO2
0%
15% CO2
15%
2
IR BUTANE (25%v)
ROOM AIR
0
8% BUTANE
8%
2
IR PROPANE (25%v)
ROOM AIR
0
8% PROPANE
8%
2
IR PROPANE (100%v)
ROOM AIR
0
50% PROPANE
50%
2
IR BUTANE (100%v)
ROOM AIR
0
20% BUTANE
20%
2
IR PROPANE (LEL)
ROOM AIR
0
0.6% PROPANE
29 LEL
2
IR BUTANE (LEL)
ROOM AIR
0
0.5% BUTANE
36 LEL
2
IR METHANE (25%v)
ROOM AIR
0
20% METHANE
20%
2
IR METHANE (100%v)
ROOM AIR
0
50% METHANE
50%
2
Span values can be changed if using different gas cylinders than those listed.
Changes can be made using MSA Link software.
*See Section 5.6.
**For most accurate results, calibration with ClO2 is recommended.
5-5
See instrument label for the approvals that apply to your specific unit.
USA and Canada
USA
Exia
Class I, Div. 1, Groups A, B, C, D, Class II, F, G
Ambient temperature: -4°F to +122°F; T4
Canada
EExia
Class I, Div. 1, Groups A, B, C, D
Ambient temperature: -20°C to +50°C; T4
ONLY THE COMBUSTIBLE DETECTION PORTION OF THIS
INSTRUMENT HAS BEEN ASSESSED FOR PERFORMANCE
BY THE CANADIAN STANDARDS ASSOCIATION (C.S.A.).
Other Countries
Australia
ANZEx: Ex ia s Zone 0 I/ IIC T4
IECEx: Ex ia d I/ IIC T4
Ambient Temperature: -20°C to +50°C
European
Community
Altair 5: II 2 G Ex ia d IIC T4
Altair 5 iR: II 2 G Ex ia d e IIC T4
-20°C ≤ Ta ≤+50°C
0080
5-6
5.6 Combustible Gas Cross Reference Factors
General purpose calibration using Calibration Cylinder P/N 10045035
set to 58% LEL Pentane Simulant.
Gas
Multiply LEL readings by:
Butane
1.0
Hydrogen
0.5
Methane (ATEX, 4.4% VOL)
0.6
Methane (UL, CSA 5.0% VOL)
0.5
Propane (ATEX, 1.7% VOL)
1.0
Propane (UL, CSA 2.1% VOL)
0.8
Pentane
1.0
Notes:
1. For an instrument calibrated on Pentane, multiply the displayed %LEL value by the
conversion factor above to get the true %LEL.
2. These conversion factors should be used only if the combustible gas is known.
3. These conversion factors are typical. Individual units may vary by +25% from these
values.
4. The results are intended for guidance only. For the most accurate measurements,
an instrument should be calibrated using the gas under investigation.
5-7
6. Order Information
6.1 Gas Cylinder Parts List
#
GAS MIX
MSA P/N
RECOMMENDED
GASES
ECONOCAL RP
CAL GAS FOR:
(34 L)
(58 L)
1
10% CO2 in N2
10081603
1
8% Butane in N2 (6L)
10075802
25% vol Butane IR
1
8% Propane in N2 (27L)
10075803
25% vol Propane IR
1
50% vol Methane in N2
10075804
100% vol Methane IR
1
100% vol Methane
711014
1
0.6% vol Propane (100 L)
493579
LEL Propane IR sensor
1
10 ppm NO2 in Air
711068
808977
NO2 sensor
1
10 ppm SO2 in Air
711070
808978
SO2 sensor
1
25 ppm NH3 in N2
711078
814866
NH3 sensor
1
10 ppm Cl2 in N2
711066
806740
Cl2 sensor
1
2 ppm Cl2 in N2
711082
10028080
ClO2 sensor
1
10 ppm HCN in N2
711072
809351
HCN sensor
1
0.5 ppm PH3 in N2
711088
710533
PH3 sensor
1
15% CO2 in N2
807387
50% CO2 IR
3
1.45% CH4, 15.0% O2,
10048790
10048788
20 ppm H2S
3
2.50% CH4, 15.0% O2,
10048888
10048889
20 ppm H2S
3
1.45% CH4, 15.0% O2,
10048789
478191
60 ppm CO
3
2.50% CH4, 15.0% O2,
10049056
813718
60 ppm CO
4
1.45% CH4, 15.0% O2,
10058034
60 ppm CO, 10 ppm NO2
4
1.45% CH4, 15.0% O2,
10048280
10045035
Big 4
60 ppm CO, 20 ppm H2S
4
2.50% CH4, 15.0% O2,
10048981
10048890
60 ppm CO, 20 ppm H2S
4
1.45% CH4, 15.0% O2,
10058036
10058171
60 ppm CO, 10 ppm NO2
4
2.50% CH4, 15.0% O2,
10058172
60 ppm CO, 10 ppm NO2
5
1.45% CH4, 15.0% O2,
10103262
Big 4+
60 ppm CO, 20 ppm H2S,
5% or 10% CO2 IR
2.5% CO2
6-1
6.2 Accessories Parts List
Calibration Accessories
DESCRIPTION
PART NO.
Calibration Assembly (Cap, Tube)
10093885
Regulator, Combination, 0.25 LPM, Model RP
467895
Fixed Flow Regulator, 0.25 LPM, Model RP
711175
Demand Regulator Kit
710288
Cl2/NH3/ClO2 Demand Regulator Kit
10034391
Interface & Carrying Accessories
MSA Link USB dongle
10082834
MSA Link Datalogging Software
10088099
Shoulder Strap
474555
Retractable Line with Belt Clip
10050976
Holster, leather
10099648
Charging Accessories
Charger only (North America)
10087913
Charger only (Global version)
10092936
Charging Cradle - Intrinsically safe (North America)
10093055
Charging Cradle - (North America)
10093054
Cradle Components Only - (North America)
10093053
Charging Cradle (EU)
10093057
Charging Cradle (Australia)
10093056
Vehicle Charger Cradle
10099397
Sample Lines & Probes
Probe, 1 ft. straight PEEK
10042621
Probe, 3 ft. straight PEEK
10042622
Polyurethane Sample Line, 10 ft.
10040665
Polyurethane Sample Line, 25 ft.
10040664
Polyurethane Sample Line, 3 ft. Coiled
10040667
(Cl2, ClO2, NH3) 5 ft.PU Coiled Sample line & probe, kit
10105210
(Cl2, ClO2, NH3) 5 ft. PU Sample line & probe, kit
10105251
(Cl2, ClO2, NH3) 10 ft. Teflon Sample line & probe, kit
10105839
Gortex replacement filters for probe, 10 pack
801582
6-2
Case assembly, upper, with label
10095081
Case assembly, upper, reactive gas (Cl2, ClO2, NH3) with label
10095082
2
Case, lower, pumped, ALTAIR 5
10083583
Case, lower, diffusion, ALTAIR 5
10083582
Case, lower, ALTAIR 5 IR
10088611
3
Battery pack, rechargeable, North America, ALTAIR 5
10090521
Battery pack, rechargeable, EU/Aus, ALTAIR 5
10083508
Battery pack, alkaline, North America, ALTAIR 5
10083507
Battery pack, alkaline, EU/Aus, ALTAIR 5
10093415
Battery pack, rechargeable, North America, ALTAIR 5 IR
10088522
Battery pack, rechargeable, EU/Aus, ALTAIR 5 IR
10093416
4
Kit, belt clip replacement, (ALTAIR 5 rechargeable)
10094830
5
Filter cover assembly
10083591
6
Display assembly, monochrome
10083589
Display assembly, color
10099650
7
Sensor bracket assembly with pump (includes vibrator motor)
10088609
Sensor bracket assembly - diffusion unit (includes vibrator motor)
10088523
8
Kit, pump cap replacement
10095051
9
Sensor, HCN
10106375
Sensor, Cl2
10080221
Sensor, ClO2
10080222
Sensor, SO2
10080223
Sensor, NO2
10080224
Sensor, NH3
10080225
Sensor, PH3
10080226
10 Sensor, COMB
10089116
11 Sensor, O2
10089163
12 Sensor, CO-H2S
10089117
Sensors, CO-H2S, O2, COMB
10095052
6-4
Kit, maintenance (includes filters, o-ring, screws)
10094829
Kit, maintenance, reactive gas (Cl2, ClO2, NH3)
10099533
(includes filters, o-ring, screws)
DESCRIPTION
PART NO.
IR sensors – must be repaired or replaced
at an MSA Authorized Service Center
CO2
0 - 5% Vol
10070758
CO2
0 - 10% Vol
10062209
CO2
0 - 50% Vol
10070757
CH4
0 - 25% Vol
10070759
CH4
0 - 100% Vol
10062205
C3H8
0 - 100% LEL
10062208
C3H8
0 - 25% Vol
10062202
C3H8
0 - 100% Vol
10062207
C4H10
0 - 100% LEL
10071381
C4H10
0 - 25 % Vol
10062201
C4H10
0 - 100 % Vol
10070756
6-5
ALTAIR® 5 et
Détecteur multigaz
ALTAIR® 5 IR
Manuel d'utilisation
En Amérique du Nord, pour contacter notre dépôt le plus proche, composez le numéro
gratuit 1-800-MSA-2222. Pour contacter MSA International, composez le 1-412-967-3354 ou
le 1-800-MSA-7777.
© MINE SAFETY APPLIANCES COMPANY 2009 - Tous droits réservés
Ce manuel est disponible sur l’Internet au www.msanet.com
Fabriqué par
MSA NORTH AMERICA
P.O. Box 427, Pittsburgh, Pennsylvania 15230
(L) Rev 2
10094176
CE MANUEL DOIT ÊTRE LU ATTENTIVEMENT PAR TOUTES LES PERSONNES RESPON-
SABLES ACTUELLEMENT OU DANS LE FUTUR DE L’UTILISATION ET DE L’ENTRETIEN
DE CE PRODUIT. Comme tout appareil complexe, ce détecteur ne fonctionne conformé-
ment à ses spécifications que s’il est utilisé et entretenu conformément aux instructions
de son fabricant. DANS LE CAS CONTRAIRE, IL RISQUE DE MAL FONCTIONNER ET LES
PERSONNES SE REPOSANT SUR CET APPAREIL POUR LEUR SÉCURITÉ RISQUENT UN
ACCIDENT GRAVE OU MORTEL.
Les garanties accordées par Mine Safety Appliances Company sur ce produit sont
annulées en cas d’utilisation et d’entretien non conforme aux instructions de ce mode
d’emploi. Protégez-vous, ainsi qu’autrui, en les respectant. Nous encourageons nos
clients à nous écrire ou à nous téléphoner avant d’utiliser cet appareil ou pour des ren-
seignements supplémentaires sur son utilisation ou sa réparation.
AVERTISSEMENT
!
1. Sécurité de l'appareil . . . . . . . . . . . . . . . . .1-1
1.1. Utilisation conforme . . . . . . . . . . . . . . . . . . . . . . . .1-1
1.2. Mesures de sécurité et de précaution à adopter . .1-3
1.3. Garantie . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-5
1.3.1 Garantie . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-6
1.3.2 Voie de recours exclusive . . . . . . . . . . . . . . . . . .1-6
1.3.3 Exclusion des dommages consécutifs . . . . . . .1-6
2. Description . . . . . . . . . . . . . . . . . . . . . . . . .2-1
2.1. Aperçu de l'appareil. . . . . . . . . . . . . . . . . . . . . . . . .2-1
Figure 2-1. Aperçu de l'appareil . . . . . . . . . . . . . . . .2-1
2.2. Interfaces matérielles . . . . . . . . . . . . . . . . . . . . . . .2-2
2.2.1 Description des touches . . . . . . . . . . . . . . . . . . .2-2
2.2.2 Description des DEL . . . . . . . . . . . . . . . . . . . . . .2-3
2.2.3 Alarmes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-3
2.3.3.1 Vibreur . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-3
2.2.3.2 Avertisseur . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-3
2.3.3.3 Alarme InstantAlert™ . . . . . . . . . . . . . . . . . . .2-3
2.3.3.4 Alerte déplacement . . . . . . . . . . . . . . . . . . . . .2-3
2.2.3.5 Mode furtif . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-4
2.2.4 Rétroéclairage . . . . . . . . . . . . . . . . . . . . . . . . . .2-4
2.2.5 Bip de confirmation . . . . . . . . . . . . . . . . . . . . . . .2-4
2.3. Indications sur l'écran . . . . . . . . . . . . . . . . . . . . . . .2-5
2.3.1 Écran monochrome . . . . . . . . . . . . . . . . . . . . . .2-5
Figure 2-2. Écran monochrome . . . . . . . . . . . . . . . .2-5
2.3.2 Écran couleur . . . . . . . . . . . . . . . . . . . . . . . . . . .2-6
2.3.2.1 Unités sans fil . . . . . . . . . . . . . . . . . . . . . . . . .2-6
2.3.3 Indicateur de charge de la pile . . . . . . . . . . . . . .2-6
Figure 2-3. Écran couleur . . . . . . . . . . . . . . . . . . . .2-6
Figure 2-4. Avertissement de la pile . . . . . . . . . . . .2-7
Figure 2-5. Arrêt de la pile . . . . . . . . . . . . . . . . . . . .2-8
2.3.4 Chargement de la pile . . . . . . . . . . . . . . . . . . . .2-9
2.4. Affichage des écrans des options . . . . . . . . . . . . . .2-9
2.4.1 Valeurs maximales affichées (page PEAK) . . .2-10
2.4.2 Valeurs minimales affichées (page MIN) . . . . .2-10
i
2.4.3 Limites d'exposition à court terme (page LECT) . .2-11
2.4.4 Moyenne pondérée dans le temps (page MPT) . .2-12
2.4.5 Affichage de la date . . . . . . . . . . . . . . . . . . . . .2-13
2.4.6 Page dernier cal. . . . . . . . . . . . . . . . . . . . . . . .2-13
2.4.7 Page échéance cal. . . . . . . . . . . . . . . . . . . . . .2-13
2.4.8 Page d'activation Alerte déplacement . . . . . . .2-13
2.5. Alarme de capteur manquant . . . . . . . . . . . . . . . .2-14
2.6. Surveillance des gaz toxiques . . . . . . . . . . . . . . . .2-14
2.7. Surveillance de la concentration d'oxygène . . . . .2-15
Figure 2-6. Conditions des alarmes
(Alarme Haute) . . . . . . . . . . . . . . . . . . . . . .2-15
2.8. Surveillance des gaz inflammables . . . . . . . . . . . .2-16
3. Fonctionnement . . . . . . . . . . . . . . . . . . . . .3-1
3.1. Facteurs environnementaux . . . . . . . . . . . . . . . . . .3-1
3.2. Mise en marche et réglage à l'air frais . . . . . . . . . .3-2
3.2.1 Test de sécurité d'échantillonnage . . . . . . . . . . .3-3
3.2.2 Réglage à l'air frais (FAS) . . . . . . . . . . . . . . . . . .3-4
Figure 3-1. Réglage à l'air frais . . . . . . . . . . . . . . . .3-4
3.3. Mode Mesure (fonctionnement normal) . . . . . . . . .3-5
3.4. Réglage de l'appareil . . . . . . . . . . . . . . . . . . . . . . . .3-5
3.4.1 Options de calibrage . . . . . . . . . . . . . . . . . . . . . .3-6
Figure 3-2. Réglage du mot de passe de calibrage 3-6
Figure 3-3. Options de calibrage . . . . . . . . . . . . . . .3-6
3.4.2 Options d'alarme . . . . . . . . . . . . . . . . . . . . . . . . .3-8
Figure 3-4. Options d'alarme . . . . . . . . . . . . . . . . . .3-8
Figure 3-5. Réglage des alarmes capteur . . . . . . .3-10
Figure 3-6. Réglage des alarmes capteur . . . . . .3-10
3.4.3 Options d'appareil . . . . . . . . . . . . . . . . . . . . . . .3-11
Figure 3-7. Options d'appareil . . . . . . . . . . . . . . . .3-11
Figure 3-8. Options de capteur . . . . . . . . . . . . . . .3-12
3.5. Fonctionnement de MSA Link . . . . . . . . . . . . . . . .3-14
3.6. Tests de fonctionnement sur l'appareil . . . . . . . . .3-14
3.7. Contrôle du calibrage . . . . . . . . . . . . . . . . . . . . . .3-15
3.8. Calibrage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-16
3.8.1 Calibrage zéro . . . . . . . . . . . . . . . . . . . . . . . . .3-17
3.8.2 Calibrage à gaz . . . . . . . . . . . . . . . . . . . . . . . .3-18
3.8.3 Fin du calibrage . . . . . . . . . . . . . . . . . . . . . . . .3-20
ii
3.8.4 Échec de l'autocalibrage . . . . . . . . . . . . . . . . . .3-20
3.8.5 Calibrage avec le banc d'essai GALAXY . . . . .3-20
3.9. Arrêt de l'appareil . . . . . . . . . . . . . . . . . . . . . . . . .3-21
Figure 3-9. Arrêt de l'appareil . . . . . . . . . . . . . . . .3-21
4. Maintenance . . . . . . . . . . . . . . . . . . . . . . . .4-1
4.1. Dépannage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-2
4.2. Vérification du fonctionnement de la pompe . . . . . .4-2
4.3. Remplacement des piles . . . . . . . . . . . . . . . . . . . . .4-4
Figure 4-1. Remplacement des piles . . . . . . . . . . . .4-4
Figure 4-2. Remplacement des piles . . . . . . . . . . . .4-4
4.4 Procédure de maintenance sous tension -
Remplacement ou ajout d'un capteur . . . . . . . . . . . . . .4-5
Figure 4-3. Remplacement de capteurs . . . . . . . . .4-6
4.5. Remplacement du filtre (pour les appareils équipés
d'une pompe intégrale) . . . . . . . . . . . . . . . . . . . . . . . . . .4-7
4.6. Nettoyage de l'appareil . . . . . . . . . . . . . . . . . . . . . .4-7
4.7. Entreposage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-8
4.8. Expédition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-8
5. Caractéristiques techniques/Certificats . .5-1
5.1. Caractéristiques techniques . . . . . . . . . . . . . . . . .5-1
5.2. Seuils d'alarme réglés à l'usine . . . . . . . . . . . . . . . .5-2
5.3. Spécifications de performance des capteurs . . . . .5-3
5.4. Spécification de calibrage . . . . . . . . . . . . . . . . . . . .5-5
5.5. Certificats . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-6
5.6. Facteurs de référence croisée des gaz
inflammables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-7
6. Informations pour les commandes . . . . .6-1
6.1. Liste des pièces de bouteille de gaz . . . . . . . . . . . .6-1
6.2. Liste des pièces accessoires . . . . . . . . . . . . . . . . .6-2
6.3. Pièces de rechange . . . . . . . . . . . . . . . . . . . . . . . .6-3
7. Annexe – Organigrammes . . . . . . . . . . . . .7-1
7.1. Séquence de démarrage (mise en marche) . . . . . .7-1
iii
7.2. Mode normal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-2
7.3. Options de la page principale . . . . . . . . . . . . . . . . .7-3
7.4. Options protégées par mot de passe . . . . . . . . . . .7-4
7.5. Séquence de menus à partir de la page principale 7-5
7.6. Séquence de calibrage . . . . . . . . . . . . . . . . . . . . . .7-7
7.7. Options de calibrage . . . . . . . . . . . . . . . . . . . . . . . .7-8
7.8. Options d'alarme . . . . . . . . . . . . . . . . . . . . . . . . . . .7-9
7.8.1 Options d'alarmes capteur . . . . . . . . . . . . . . . .7-10
7.9. Options d'appareil . . . . . . . . . . . . . . . . . . . . . . . . .7-11
iv
1. Sécurité de l'appareil
1.1. Utilisation conforme
Les détecteurs multigaz ALTAIR 5 et ALTAIR 5 IR doivent être utilisés
par du personnel formé et qualifié. Il sont conçus pour être utilisés dans
le cadre d'une évaluation de risques pour :
•
Évaluer l'exposition potentielle des ouvriers aux gaz et vapeurs
inflammables et toxiques ainsi qu'à un faible niveau d'oxygène.
•
Déterminer la surveillance appropriée des gaz et vapeurs
nécessaire sur un lieu de travail.
Les détecteurs multigaz ALTAIR 5 et ALTAIR 5 IR peuvent être
configurés pour détecter un maximum de cinq gaz pour l'ALTAIR 5 et un
maximum de six gaz pour l'ALTAIR 5 IR :
•
Les gaz inflammables et certaines vapeurs inflammables
•
Atmosphères pauvres en oxygène ou riches en oxygène
•
Les gaz toxiques spécifiques pour lesquels un capteur est installé.
Bien que l'appareil puisse détecter jusqu'à 25 % d'oxygène dans l'air
ambiant, il n’est homologué pour être utilisé que dans des atmosphères
ne contenant pas plus de 21 % d'oxygène.
Le détecteur de gaz ALTAIR 5 IR contient un capteur infrarouge pour
détecter le CO2 ou des gaz combustibles spécifiques jusqu'à une
concentration de 100 % en volume, en plus de jusqu'à cinq autres gaz.
Lire et observer attentivement toutes les instructions.
•
Effectuer un test d'écoulement bloqué avant chaque
utilisation quotidienne si l'appareil est équipé d'une
pompe intégrale.
•
Contrôler le calibrage avant chaque utilisation quotidienne
et l'ajuster si nécessaire.
•
Contrôler le calibrage plus fréquemment en cas
d'exposition au silicone, aux silicates, aux composés à
base de plomb, à l'hydrogène sulfuré ou à des niveaux
élevés de contaminants.
•
Contrôler de nouveau le calibrage si l'unité est soumise à
des chocs physiques.
AVERTISSEMENT
!
Utiliser uniquement pour détecter les gaz/vapeurs pour
lesquels un capteur est installé.
•
Ne pas utiliser pour détecter des poussières ou brumes
inflammables.
•
S'assurer qu'une quantité suffisante d'oxygène est
présente.
•
Ne pas obturer des capteurs.
•
Ne jamais obturer l'entrée de la pompe, sauf pour réaliser
un test de sécurité d'échantillonage.
•
Confier l'interprétation des valeurs affichées sur l'appareil
à une personne formée et qualifiée.
•
Ne pas retirer le bloc de piles de l'appareil dans une zone
dangereuse.
•
Ne pas recharger la pile lithium-ion dans une zone
dangereuse. Risque d'explosion.
•
Ne pas remplacer les piles alcalines dans une zone
dangereuse.
•
Ne pas modifier l'appareil.
•
Utiliser uniquement des lignes d'échantillonnage
homologuées par MSA.
•
Ne pas utiliser des tuyaux ou des lignes d'échantillonnage
en silicone.
•
Attendre suffisamment longtemps avant de lire les
valeurs ; les temps de réponse varient selon le gaz et la
longueur de la ligne d'échantillonnage.
•
Ne pas utiliser l'appareil avec la pompe intégrale pendant
une durée prolongée dans une atmosphère contenant une
concentration de vapeurs de carburant ou de solvant
pouvant dépasser 10 % de la LIE (limite inférieure
d'explosivité.
TOUTE UTILISATION NON CONFORME PEUT OCCASIONNER DES BLESSURES
GRAVES VOIRE MORTELLES.
Cet appareil numérique de classe A satisfait à la norme canadienne
ICES-003.
Pour les utilisateurs d'Australie/Nouvelle-Zélande : le berceau de
charge est un produit de classe A. Dans un environnement résidentiel,
ce produit peut causer des interférences radio, et dans ce cas
l'utilisateur peut devoir prendre des mesures adéquates.
1-2
1.2. Mesures de sécurité et de précaution à
adopter
Observer attentivement les limitations et précautions de
sécurité suivantes avant de mettre cet appareil en service.
Toute utilisation non conforme peut occasionner des
blessures graves voire mortelles.
Les détecteurs multigaz ALTAIR 5 et ALTAIR 5 IR sont conçus pour :
•
Détecter les gaz et vapeurs présents dans l'air uniquement.
•
Détecter seulement les gaz toxiques spécifiques pour lesquels un
capteur est installé (capteurs disponibles voir le Chapitre 6).
•
Effectuer un test d'écoulement bloqué (voir chapitre 4.2) avant
chaque utilisation quotidienne si l'appareil est équipé d'une pompe
intégrale.
•
MSA recommande de contrôler le fonctionnement (voir chapitre
3.6) et d'effectuer une révision de routine chaque jour avant
l'utilisation.
•
Réaliser le contrôle de calibrage (voir chap. 3.7) avant chaque
utilisation quotidienne afin de vérifier le fonctionnement correct de
l'appareil. Ajuster le calibrage si les valeurs affichées ne sont pas
comprises dans les limites spécifiées.
•
Contrôler le calibrage plus fréquemment si l'unité est soumise à
des chocs physiques ou à des niveaux élevés de contaminants.
Contrôler également le calibrage plus fréquemment si l'atmosphère
testée contient les matières suivantes, qui peuvent désensibiliser le
capteur de gaz inflammables et réduire les valeurs qu'il détecte :
•
Silicones organiques
•
Silicates
•
Composés à base de plomb
•
Expositions à l'hydrogène sulfuré de plus de 200 ppm (partie
par million) ou expositions de plus de 50 ppm pour une minute.
La concentration minimale d'un gaz inflammable dans l'air qui peut
s'enflammer est définie comme la Limite inférieure d'explosivité (LIE).
Une valeur de gaz inflammable affichée de « 100 » ou « 5,00 » indique
que l'atmosphère est au-dessus 100 % de la LIE ou 5,00 % vol. CH4, et
qu'un risque d'explosion existe. S'éloigner immédiatement de la zone
dangereuse.
Ne pas utiliser le détecteur multigaz ALTAIR 5 ou ALTAIR 5 IR pour
1-3
AVERTISSEMENT
!
tester les gaz inflammables ou toxiques dans les atmosphères
suivantes car cela peut entraîner des affichages erronés :
•
Atmosphères pauvres en oxygène ou riches en oxygène
•
Atmosphères réductrices
•
Cheminées de fourneaux
•
Environnements inertes (seuls les capteurs IR peuvent être utilisés)
•
Atmosphères contenant des brouillards/poussières aérogènes
inflammables.
Ne pas utiliser les détecteurs multigaz ALTAIR 5 et ALTAIR 5 IR pour tester
la présence de gaz inflammables dans des atmosphères contenant des
vapeurs issues de liquides avec une température d'éclair élevée (au-dessus
de 38 °C, 100 °F), car cela peut entraîner des détections faussement
basses.
Ne pas obstruer les ouvertures des capteurs, car cela peut occasionner des
détections imprécises. Ne pas appuyer sur la face des capteurs, car cela
peut les endommager et occasionner des détections erronées. Ne pas
utiliser de l'air comprimé pour nettoyer les trous des capteurs, car la pression
risque d'endommager les capteurs.
Laisser suffisamment de temps à l'unité pour afficher des valeurs précises.
Les temps de réaction varient selon le type de capteur utilisé (voir chapitre
5.3, « Spécifications de performance des capteurs »). Pour les unités avec
une pompe d'échantillonnage, attendre au moins 3 secondes/mètre de ligne
d'échantillonage (1 seconde/pied) pour permettre à l'échantillon d'être
amené sur les capteurs.
Les lignes d'échantillonnage composées de tuyaux à partir de 0,062 pouce
(1,57 mm) de diamètre intérieur permettent des temps de transport rapides
vers l'appareil ; toutefois, elles doivent être limitées à 50 pieds (15 m) en
longueur.
L'échantillonnage des gaz toxiques réactifs (Cl2, ClO2, NH3) doit être
effectué seulement avec les kits de ligne d'échantillonnage de gaz réactif et
de sonde listés dans le tableau Accessoires de la section 6.2.
Toutes les valeurs et informations affichées par l'appareil doivent être
interprétées par une personne formée et qualifiée pour l'interprétation des
valeurs affichées par rapport à l'environnement spécifique, à la pratique
industrielle et aux limitations d'expositions.
Observer la procédure de maintenance appropriée de la pile
Utiliser exclusivement des chargeurs de piles mis à disposition par la société
MSA pour être utilisés avec cet appareil ; l'utilisation d'autres chargeurs
risque d'endommager le bloc de piles et l'unité. Éliminer les piles
conformément aux réglementations locales en matière de santé et de
1-4
Prendre connaissance des conditions ambiantes
Un certain nombre de facteurs environnementaux peuvent affecter les
lectures des capteurs d'oxygène et de gaz toxiques, y compris les
changements de pression, d'humidité et de température. Les changements
de pression et d'humidité affectent la quantitéd'oxygène réellement présente
dans l'atmosphère.
Après l'exposition aux températures (°C) négatives, le fonctionnement des
unités équipées de capteurs HCN (acide cyanhydrique) doit être testé et, si
nécessaire, les unités doivent être calibrées après avoir été ramenées à la
température ambiante.
Prendre connaissance des procédures de manipulation des
composants électroniques sensibles aux décharges électrostatiques
L'appareil contient des composants sensibles aux décharges
électrostatiques. Ne pas ouvrir ni réparer l'unité sans utiliser une protection
contre les décharges électrostatiques adaptée. La garantie ne couvre pas
les dommages occasionnés par les décharges électrostatiques.
Prendre connaissance des réglementations relatives au produit
Observer toutes les réglementations nationales importantes applicables
dans le pays d'utilisation.
Prendre connaissance des règlements relatifs à la garantie
Les garanties offertes par la société Mine Safety Appliances Company par
rapport au produit sont nulles et non avenues si le produit n'est pas utilisé ni
entretenu conformément aux instructions contenues dans le présent manuel.
Veuillez vous protéger ainsi que les autres personnes en respectant ces
instructions. Nous encourageons nos clients à nous écrire ou nous appeler
concernant cet équipement avant de l'utiliser ou pour toutes informations
supplémentaires relatives à son utilisation ou sa réparation.
1.3. Garantie
ARTICLE
DUREE DE LA GARANTIE
Châssis et électronique
Deux ans*
Capteurs COMB, O2, H2S, CO
Deux ans*
Capteurs Cl2, ClO2, NH3, HCN, NO2, PH3, SO2
Un an*
Capteur infrarouge
Deux ans*
*Pour des extensions de garantie, veuillez contacter MSA
1-5
Cette garantie ne couvre pas les filtres, les fusibles, etc. D'autres
accessoires non listés spécifiquement peuvent avoir des périodes de
garanties différentes. Cette garantie est valide uniquement si le produit
est entretenu et utilisé conformément aux instructions et/ou
recommandations du vendeur.
Le vendeur sera dégagé de toutes les obligations liées à cette garantie
dans le cas où des réparations ou modifications sont réalisées par des
personnes autres que le personnel de maintenance propre et autorisé
ou bien si la demande en garantie résulte d'un abus physique ou d'une
utilisation incorrecte du produit. Aucun agent, employé ou représentant
du vendeur ne possède un quelconque pouvoir d'obliger le vendeur à
une quelconque déclaration, représentation ou garantie concernant ce
produit. Le vendeur ne garantit en aucune manière les composants ou
accessoires non fabriqués par le vendeur, mais transmettra à l'acheteur
toutes les garanties des fabricants de tels composants.
CETTE GARANTIE REMPLACE TOUTES LES AUTRES GARANTIES,
EXPRESSES, IMPLICITES OU LÉGALES, ET EST STRICTEMENT
LIMITÉE AUX TERMES DE CELLE-CI. LE VENDEUR RENONCE À
TOUTE GARANTIE DE QUALITE MARCHANDE OU D'APTITUDE À
UN OBJECTIF PARTICULIER.
1.3.2 Voie de recours exclusive
Il est expressément convenu que la voie de recours unique et exclusive
de l'acheteur pour violation de la garantie ci-dessus, pour toute conduite
délictuelle du vendeur, ou pour toute autre cause d'action, sera le
remplacement au gré du vendeur, de tout appareil ou pièces de celui-ci,
reconnu comme défectueux après examen par le vendeur.
L'appareil et/ou les pièces de remplacement seront fournis sans frais à
l'acheteur, franco à bord à l'usine du vendeur. Le manquement du
vendeur à remplacer avec succès tout appareil ou pièce non conforme
n'entraînera pas, pour la voie de recours établie par la présente, un
défaut de réalisation de son objectif initial.
1.3.3 Exclusion des dommages consécutifs
L'acheteur comprend et reconnaît spécifiquement que dans aucune
circonstance le vendeur ne sera responsable envers l'acheteur des
dommages consécutifs économiques, spéciaux, fortuits ou consécutifs
ou des pertes de quelque nature que ce soit, incluant mais non limitées
à la perte de profits anticipés et de toute autre perte causée par non-
1-6
des produits. Cette exclusion est applicable aux réclamations au titre de
la violation de garantie, de la conduite délictuelle ou de toute autre
cause d'action contre le vendeur.
1-7
2.1. Aperçu de l'appareil.
1
DEL (2 rouges « Alarme », 1 verte « Sécurité ») 9
Touche ▼
2
Communication MSA LINK
10
Clip ceinture (ALTAIR 5 seulement)
3
Entrées capteur (version Diffusion)
11
Port de chargement
4
Bouchon pompe (version avec pompe)
12
Vis
5
Avertisseur
13
Entrée pompe
6
Écran
14
Filtre
7
Touche ▲
15
Charge Status LED
8
Touche
L'appareil surveille les gaz présents dans l'air ambiant et sur le lieu de
travail.
Le détecteur ALTAIR 5 est disponible avec un maximum de quatre
capteurs qui peuvent afficher les valeurs détectées pour cinq gaz
séparés (un capteur de gaz toxiques double est capable de détecter à
la fois les concentrations de CO et de H2S en un seul capteur).
Le détecteur est disponible en version diffusion ou avec pompe. En cas
d'utilisation d'un appareil avec pompe, ne pas retirer le bouchon de la
pompe (pos. 4 sur la figure 2-1).
Figure 2-1. Aperçu de l'appareil
2-1
Le détecteur ALTAIR 5 IR est disponible avec un maximum de cinq
capteurs qui peuvent afficher les valeurs détectées pour six gaz
séparés (un capteur de gaz toxiques double est capable de détecter à
la fois les concentrations de CO et de H2S en un seul capteur).
Les détecteurs ALTAIR 5 et ALTAIR 5 IR sont disponibles avec un écran
monochrome ou couleur.
Bien que l'appareil puisse détecter jusqu'à 25 % d'oxygène dans l'air
ambiant, il n'est homologué pour être utilisé que dans des atmosphères
ne contenant pas plus de 21 % d'oxygène.
Les niveaux d'alarme pour les gaz individuels sont réglés en usine et
peuvent être modifiés dans le menu de réglage de l'appareil. Ces
modifications peuvent aussi être effectuées via le logiciel MSA LinkTM.
S'assurer d'avoir téléchargé la dernière version du logiciel MSA LinkTM à
partir du site web de www.msanet.com. Après avoir apporté des
changements à l'aide de MSA Link, il est recommandé d'arrêter (OFF)
et de remettre en marche (ON) l'unité.
2.2 Interfaces d'utilisateur
Le fonctionnement de l'appareil est géré par dialogue à partir de l'écran
à l'aide de trois touches de fonction (voir figure 2-1).
2.2.1 Description des touches
Les détecteurs ALTAIR 5 et ALTAIR 5 IR disposent de trois touches
destinées à l'utilisateur. Chaque touche peut fonctionner en tant que
« touche de fonction », comme défini juste au-dessus de la touche en
question.
TOUCHE
DESCRIPTION
La touche
permet de mettre en marche ou d'arrêter l'appareil et de
valider les sélections d'actions effectuées par l'utilisateur.
▼
La touche ▼ permet d'avancer dans les écrans de données ou de dimin-
uer les valeurs en mode Réglage.
▲
La touche ▲ permet de réinitialiser les valeurs maximales, LECT, MPT et
les alarmes (quand cela est possible) ou de réaliser le calibrage en mode
mesure. Elle sert également à revenir à la page précédante ou à aug-
menter les valeurs en mode réglage.
Lorsque le bouton ▲ et le bouton ▼ sont enfoncés simultanément dans
le mode Mesure normale, on peut entrer dans le mode Réglage après
avoir confirmé le mot de passe.
2-2
2.2.2 Description des DEL
DEL
DESCRIPTION
ROUGE
Les voyants DEL rouges sont les indications visuelles d'un état
(ALARME)
d'alarme ou d'un type quelconque d'erreur sur l'appareil
VERTE
La DEL de sécurité clignote une fois toutes les 15 secondes pour notifier
(DE SÉCURITÉ) l'utilisateur que l'instrument est en marche et opérationnel dans les condi-
tions indiquées ci-dessous :
• La DEL DE SECURITE verte est allumée
• La lecture de gaz inflammable est 0 % de la LIE ou 0 % VOL
• La lecture de l'oxygène (O2) est 20,8 %
• Toutes les autres lectures de capteur sont 0
• Aucune alarme de gaz n'est déclenchée (basse ou haute)
• L'instrument n'est pas dans un état d'avertissement ou d'alarme de
pile déchargée
• Les lectures de LECT et de MPT sont 0 ppm.
Cette option peut être arrêtée via le logiciel MSA Link
2.2.3 Alarmes
L'instrument est équipé de plusieurs alarmes pour une meilleure
sécurité de l'utilisateur :
2.3.3.Vibreur
L'instrument vibre lorsque toute condition d'alarme est active. Ceci peut
être arrêté dans le menu REGLAGE - OPTIONS D'ALARME (SETUP-
ALARM OPTIONS) (Section 3.4.2).
2.2.3.2 Avertisseur sonore
Cette unité est équipée d'une alarme sonore. L'avertisseur peut être
arrêté dans le menu REGLAGE - OPTIONS D'ALARME (SETUP-
ALARM OPTIONS) (Section 3.4.2).
2.3.3.3 Alarme InstantAlert™
La fonction InstantAlert exclusive permet à l'utilisateur d'activer
manuellement une alarme sonore pour avertir les personnes à proximité
des situations potentiellement dangereuses. On active l'alarme
InstantAlert en maintenant enfoncé le bouton ▼ pendant environ 5
secondes dans le mode Mesure normale.
2.3.3.4 Alerte déplacement
Si l'Alerte déplacement est mise en marche (+ = ON) ) et activée (voir
la section 3.4.2), l'instrument actionne une alarme de « détresse » si les
mouvements ne sont pas détectés dans un délai de 30 secondes. Les
DEL d'alarme clignotent et l'avertisseur sonore est actionné avec une
fréquence sonore progressive.
2.2.3.5 Mode furtif
Le Mode furtif désactive les alarmes visuelles, sonores et vibratoires.
MSA recommande de laisser cette fonction sur Arrêt (OFF), sa valeur
par défaut. Elle peut être mise en marche (ON) dans le menu
REGLAGE - OPTIONS D'INSTRUMENT (SETUP-INSTRUMENT
OPTIONS) (section 3.4.3).
Le message « Alarmes arrêtées » (Alarms OFF) clignote sur l'écran
monochrome lorsque le Mode furtif est en marche. Sur l'écran couleur,
toutes les trois icônes d'alarme sont montrées à l'arrêt (
).
2.2.4 Rétroéclairage
Le rétroéclairage est automatiquement activé lorsqu'un bouton
quelconque du panneau avant est enfoncé et reste allumé pendant la
durée de la temporisation sélectionnée par l'utilisateur. Cette durée peut
être changée en utilisant le Mode Réglage - Instrument (section 3.4.3)
ou à l'aide du logiciel MSA Link.
2.2.5 Bip de confirmation
Ce bip de confirmation est activé toutes les 30 secondes par l'émission
d'un bip temporaire par l'avertisseur sonore et le clignotement des DEL
d'alarme dans les conditions suivantes :
•
Le bip de confirmation est en service
•
L'instrument est sur la page de mesure normale des gaz
•
L'instrument n'est pas dans l'état d'avertissement de pile faible
•
L'instrument n'est pas dans l'état d'alarme de gaz.
2-4
2.3. Indications sur l'écran
2.3.1 Écran monochrome
1
Type de gaz
5 État de la pile
2
Heure actuelle
6 Indique Alerte de déplacement activé ou Cal réussie
3
« Touche de fonction » ▼
7 « Touche de fonction » ▲
4
Concentration de gaz
Lecture de capteur de gaz inflammable
Lecture de capteur d'oxygène
Lecture de capteur de CO
Lecture de capteur de H2S
Lecture de capteur toxique (cinquième capteur) s'il est installé
Indique le nom de la page suivante
Lance le calibrage
Capteur IR - sixième capteur, s'il est installé
Sur un écran monochrome, un message s'affiche toutes les 30
secondes si les alarmes de vibreur, d'avertisseur sonore ou à DEL sont
arrêtées.
Figure 2-2. Écran monochrome
CAL
METH
2.3.2 Écran couleur
1
Touche de fonction ▼
8
Vibreur désactivé
2
Indicateur numérique
9
Avertisseur désactivé
3
Unités de concentration en gaz
10
DEL éteinte
4
Heure actuelle
11
État de la pile
5
Symbole d'Alerte déplacement allumé 12
Type de gaz
(+ = ON)
13
Touche de fonction ▲
6
Force du signal sans fil
7
Type de calibration de gaz inflammable
2.3.2.1 Unités sans fil
Les unités à écran couleur peuvent être commandées avec une
fonctionnalité sans fil qui permet les communications avec le réseau
personnel alphaScout. Cette fonctionnalité de réseau est utilisée
presque exclusivement dans les marchés internationaux. Contacter
MSA pour des informations supplémentaires.
2.3.3 Indicateur de charge de la pile
L'icône d'état de la pile est affichée en continu dans l'angle supérieur
droit de l'écran couleur et dans l'angle inférieur droit de l'écran
monochrome. Une barre représente le niveau de charge de la pile.
La durée de fonctionnement nominale de l'instrument (COMB, O2, H2S,
CO, avec pompe et écran monochrome) à la température ambiante est
14 heures. La durée de fonctionnement réelle varie en fonction de la
Figure 2-3. Ecran couleur
2-6