Getting Ready
1. Assemble all accessories
so that you have easy
access to them
6. Remove the lenses cover
2. Adjust the shoulder
straps to fit your body
by fastening the straps
3. Fasten the front pad
4. Lean the camera against
the front pad. Use the
grooves to adjust the
viewing angle of the
camera.
5.
U S E R M A N U A L D A Y C O R ® S U P E R B
Operating tips *
Status Line
This function can be turned on/off through the Status function-key. The status line,
positioned at the LCD bottom, shows the current status of the display mode, the focus
mode, the UV gain, if long-integration is on and the counting value. This data is
valuable when processing a
U S E R M A N U A L D A Y C O R ® S U P E R B
6
• Visible only mode
• UV only mode
• Combined mode
While in the UV only mode, the status line displays the UV gain value only .
The visible only mode is handy when there is a reflection on glass insulators that might
be mistaken for corona. The camera enters the visible only mode during visible
zoomi
U S E R M A N U A L D A Y C O R ® S U P E R B
7
LI – Long Integration
Long Integration supports inspection of very weak signals or inspection in areas with
abundant non-corona UV background signals. This function affects the read-out
frequency of frames. Below is an example taken from a distribution line in Costa Rica
(a 35 kV line) where a nearby weld
U S E R M A N U A L D A Y C O R ® S U P E R B
8
How to count
Enter the counting mode by pressing on the counting function-key. A small frame will
appear on the display. By pressing the up and down arrows the size of the frame will
change. Position the corona image in the center of the frame and wait until the count
value stabilizes with only small fluctu