Olympus ESG-400 Maintenance Manual

File Specifications

Brand: Olympus
Category: Inverter
Model: ESG-400 (Maintenance Manual), pr6OeTGbrgVE6
File Type: PDF
File Size: —
Pages: 110 pages

Download Document

Get a high-quality, printable PDF version of this manual for offline use.

Download PDF Manual (110 pages)

⚠️ Download Safety: This PDF is completely safe to download (last scan date: 20/06/2026).

Read Online

Read the Olympus ESG-400 Maintenance Manual online using the built-in PDF viewer.

Key Pages of Document

The intended use depends on the approval of the country. Refer to the instructions for use of the electrosurgical

unit. Maintenance instructions This maintenance manual contains essential information on using and maintaining this electrosurgical generator safely and effectively. Instructions for the operation of this electrosurgical generator and related danger, warnings and cautions concerning electrosurgery are beyond th ...

High frequency leakage current or spark discharge may cause user burns.

Follow the dangers, warnings and cautions given below when handling and servicing this electrosurgical unit. This information is to be supplemented by the dangers, warnings and cautions given in each chapter. User-related error prevention WARNING Improper use The safety and effectiveness of electrosurgical interventions depend not only on the design of the equipment ...

Do not apply excessive bending, straining, or squeezing force to any cords. It may cause

malfunction. CAUTION Non-compatible accessories and accessory damage The electrosurgical generator shall only be used with compatible accessories. When connecting accessories (cords, electrodes, HF instruments) avoid output settings where the maximum output voltage of the electrosurgical generator may exceed the rated acces ...

Fire / Explosion

DANGER Ignitable anaesthetics / fire supporting gases The risk of flammable gases or other materials being ignited exists with any contact of electrical energy. Precautionary measures must be taken to keep flammable materials and substances away from an active electrosurgical unit (do not use flammable anesthetics, nitrous oxide or oxygen). Otherwise, explosion or fire may result and cause serious injuries. This electrosurgical uni ...

©2011 Olympus Winter & Ibe GmbH. All rights reserved.

Unauthorized reproduction or distribution in part or in whole is prohibited. Trademarks OLYMPUS is a registered trademark of the Olympus Corporation. The company names, product names, and proprietary technical terms in this document are the trademarks or registered trademarks of their respective owners. 7.022.211 / ISSUE 5 6 / 110 Introduction ...

INTRODUCTION ................................................................................................................ 2

CONTENT .......................................................................................................................... 7 CHAPTER 1: PRODUCT SPECIFICATIONS .................................................................... 9 1 OUTLINE .................................................................... ...

1 ESG-400 MAIN UNIT .................................................................................................... 44

2 BOARD COMPATIBILITY ............................................................................................. 44 3 OPTIONAL ACCESSORIES ......................................................................................... 44 3-1 WB50402W (Footswitch with two pedals) ................................................... ...

CHAPTER 1: PRODUCT SPECIFICATIONS

1 OUTLINE ...................................................................................................................... 10 1-1 Intended Use .................................................................................................................................... 10 1-2 Compatibility ................................................................................................................... ...

The intended use depends on the approval of the country. Refer to the instructions for use of the electrosurgical

generator. 1-2 Compatibility This product can be used in combination with the products listed in compatibility table. 1-3 Expected service life The expected service life is 10 years. 7.022.211 / ISSUE 5 10 / 110 Chapter 1: PRODUCT SPECIFICATIONs ...

The ESG-400 is a reusable, non-sterile electrosurgical generator with different mono- and bipolar cutting and

coagulation modes. The maximum output power is 320 W. On the front side it features a touch screen display that displays the connection status of accessories and peripherals connected to the electrosurgical generator. It is also used to show and modify the output settings (e.g. mode, output power, effect) as well as to control o ...

BipolarCut (All bipolar cutting procedures of tissue structures; 3 Effects)

• SalineCut (Cutting in conductive fluid; 3 Effects; only available via UNIVERSAL socket) • PK PureCut (Cutting of varying tissue structures; 3 Effects; only available via UNIVERSAL socket) • PK SoftCut (Cutting of varying tissue structures; 3 Effects; only available via UNIVERSAL socket) • PK LoopCut (Cutting of varying tissue structures, especiall ...

(1) Use this product under the supervision of a doctor at a medical facility.

(2) Do not use this product in combination with the products other than those designated by Olympus. (3) This product shoul be used, transported or stored in the following environment. Operation environment Temperature + 10…+ 40°C Relative humidity 30…85%, non-condensing Atmospheric pressure 70…106 kPa Transportation and storage environment Temperatu ...

ELECTROSURGICAL GENERATOR ESG-400 (REF: WB91051W)

Power supply Voltage range 100…120 V~ / 220…240 V~ Frequency 50 / 60 Hz Maximum input power 1500 VA Power fuse 10 A (only FST-series from Schurter) Power connection line IEC 60320-1 / C13 Maximum length: 4.5 m Terminal for potential equalization Yes Size, weight and packaging Width x Depth x Height 370 × 465 × 156 mm Volume 25752 cm³ Weight of generator 12.5 kg Weight of ...

3-pin (∅ 4 mm),

Valleylab standard; coaxial (∅ 8 mm), Bovie standard MONOPOLAR 2 3-pin (∅ 4 mm), Valleylab standard; coaxial (∅ 5 / 9 mm), Erbe standard BIPOLAR 2-pin (∅ 4 mm, pin spacing 28.8 mm), Valleylab standard; coaxial (∅inner 8 mm, ∅outer 4 mm), Erbe standard UNIVERSAL 7-pin, Olympus standard Neutral electrode Single or split, 10 mm plug Contact quality monitor (CQM) Allowable resistance range for split typ ...

Footswitch (REF: WB50402W, double pedal)

Classification Protection class according to IEC 60529 IPX8 (except the plug section) Size, weight and packaging Width x Depth x Height 350 × 185 × 65 mm Weight of footswitch 1.9 kg Length of cord 4 m Weight of packaging 0.5 kg Type of packaging Cardboard material 4-4 Footswitch (REF: WB50403W, single pedal, optional) Classification Protection class according to IEC 60529 IPX8 (except the plug section) ...

Communication cable 0.25 m (REF: MAJ-1871, optional)

Size Weight 0.05 kg Length of cord 0.25 m 4-7 Communication cable 10 m (REF: MAJ-1872, optional) Size Weight 0.5 kg Length of cord 10 m 4-8 Adapter for UHI-2/3 (REF: MAJ-1873, optional) Size Width x Depth x Height 100 × 77 × 42 mm Weight 0.35 kg Compatible cables MAJ-1871, MAJ-1872 7.022.211 / ISSUE 5 17 / 110 Chapter 1: PRODUCT SPECIFICATIONs ...

5 Name and Function of each part

5-1 Symbols and descriptions 5-1-1 Safety related symbols Date of manufacture CE marking Non-ionizing electromagnetic radiation Output insulated from earth (connection for neutral electrode) Caution, read instructions Refer to instructions Defibrillation proof type CF applied part (cardiac application) Waste electrical and electronic equipment cTUVus marking Manufacturer Potential equalization terminal Fuse rating ...

Power on / off

Neutral electrode – non-split type Neutral electrode – split type Select procedure Footswitch Menu BIPOLAR socket MONOPOLAR 1 socket UNIVERSAL socket MONOPOLAR 2 socket 5-1-3 Touch screen Double footswitch Single footswitch Autostart Plus Minus Return OK Cancel 7.022.211 / ISSUE 5 19 / 110 Chapter 1: PRODUCT SPECIFICATIONs ...

Touch tone on

Touch tone off Software version Safety test Service Volume Brightness Select procedure (in title line) Menu (in title line) Toggle Previous Next Numeric Alphabetic Uppercase / lowercase Backspace Caution Communication indicator 7.022.211 / ISSUE 5 20 / 110 Chapter 1: PRODUCT SPECIFICATIONs ...