i1000SR System Quick Troubleshooting Guide
Error Code 3700 Unable to Process Test, (x)
Wash Aspiration Error for Probe(s) (y)
PN 205854-101
©2009, Abbott Laboratories, Abbott Park, IL, 60064
Refer to the Operations Manual for further instructions
Customer Support: 1-877- 4ABBOTT (1-877- 422-2688)
Instrument SN: ______________
•
Remove fibrin clots from
Customer Support: 1-877- 4ABBOTT (1-877- 422-2688)
i1000SR System Quick Troubleshooting Guide
Error Code 5209 No Reaction Vessel (RV)
Detected in Process path after RV Load
Attempt
•
If RV loader chute is not in position to load RVs,
reseat the RV loader chute into the wash zone
manifold.
Chute is in the UP position (incorrect).
Chute is in the DOWN pos