Abbott i1000SR Quick Troubleshooting Manual

Text of Abbott i1000SR Quick Troubleshooting Manual

i1000SR System Quick Troubleshooting Guide

Error Code 3700 Unable to Process Test, (x) Wash Aspiration Error for Probe(s) (y) PN 205854-101 ©2009, Abbott Laboratories, Abbott Park, IL, 60064 Refer to the Operations Manual for further instructions Customer Support: 1-877- 4ABBOTT (1-877- 422-2688) Instrument SN: ______________ • Remove fibrin clots from

Customer Support: 1-877- 4ABBOTT (1-877- 422-2688)

i1000SR System Quick Troubleshooting Guide Error Code 5209 No Reaction Vessel (RV) Detected in Process path after RV Load Attempt • If RV loader chute is not in position to load RVs, reseat the RV loader chute into the wash zone manifold. Chute is in the UP position (incorrect). Chute is in the DOWN pos

Abbott Laboratory Equipment Owners Instructions: