AmScope 490 Series Operator's Manual
Read Online
Read the AmScope 490 Series Operator's Manual online using the built-in PDF viewer.
Text of the manual
(Please Read This Manual Before Using the
Microscope)
490 Series
Compound Microscope
490 Series Microscope
AmScope Proprietary
Page 2 of 64
Copyright © 2015 United Scope Inc. DBA AmScope. All rights reserved.
Copyright Notice
This document, which is in draft form and subject to change without notice, is an
unpublished work containing confidential, trade secret information that is proprietary to
AmScope, and is protected by laws pertaining to such materials. This document, the
information in this document and all rights thereto are the sole and exclusive property of
AmScope and are not to be copied, used or disclosed to anyone, in whole or in part,
without the express written permission of AmScope.
AmScope assumes no liability for any damages incurred, directly or indirectly, from any
errors, omissions or discrepancies between the use of this microscope and the information
contained in this document.
AmScope
Purchasing: 949-333-0001 or 1-888-950-2888 FREE (toll free)
Customer Service: 949-333-0002
Technical Support: 949-333-0004
Shipping: 949-333-0005
Accounting: 949-333-0008
Fax: 949-333-0007
490 Series Microscope
AmScope Proprietary
Page 3 of 64
Contents
BEFORE USE ..................................................................................................................................................................... 4
Introduction ................................................................................................................................................................. 4
Precautions ................................................................................................................................................................. 4
1. MICROSCOPE PARTS ..................................................................................................................................................... 5
2. OPERATION ................................................................................................................................................................... 6
2.1 Unpacking ............................................................................................................................................................. 6
2.2 Assembly ............................................................................................................................................................... 7
2.3 Adjusting the View .............................................................................................................................................. 10
2.4 Specimen Set Up .................................................................................................................................................. 11
2.5 Focusing .............................................................................................................................................................. 12
2.6 Using the Trinocular Port ................................................................................................................................... 13
2.7 Attaching a Camera ............................................................................................................................................ 13
2.8 Setting the Stage’s Stop-Limit ............................................................................................................................. 14
2.9 Adjusting Focus Tension ..................................................................................................................................... 14
2.10 Setting the Condenser Lens Adjustment Knob................................................................................................... 15
2.11 Adjusting the Iris ............................................................................................................................................... 15
2.12 Changing a Filter .............................................................................................................................................. 15
2.13 Changing the Halogen Light Bulbs ................................................................................................................... 16
2.14 Changing the Fuse ............................................................................................................................................ 16
2.15 Using Oil Immersion ......................................................................................................................................... 17
2.16 Using Darkfield Condenser Lenses (Optional Accessories) ............................................................................. 19
2.17 Using a Dry Darkfield Condenser Lens (Optional Accessory) ......................................................................... 20
2.18 Using Oil Darkfield Condenser Lens (Optional Accessory) ............................................................................. 21
2.19 Using a Phase Contrast Kit (Optional Accessory) ............................................................................................ 24
2.20 Using a Simple Phase Contrast Kit (Optional Accessory) ................................................................................ 28
2.21 Using a Turret Phase Contrast Kit (Optional Accessory) ................................................................................. 33
2.22 Microscope Maintenance .................................................................................................................................. 38
2.23 Darkfield Condenser Kit & Phase Contrast Kit Maintenance .......................................................................... 39
3. GENERAL SPECIFICATIONS .......................................................................................................................................... 40
3.1 General Model Comparison ................................................................................................................................ 40
3.2 Objectives ............................................................................................................................................................ 41
3.3 Eyepieces ............................................................................................................................................................. 41
3.4 Darkfield Condensers.......................................................................................................................................... 41
3.5 Phase Contrast Kits............................................................................................................................................. 42
4. PARAMETERS .............................................................................................................................................................. 43
4.1 Electrical System ................................................................................................................................................. 43
4.2 Microscope Parameters ...................................................................................................................................... 43
5. RECOMMENDED ACCESSORIES ................................................................................................................................... 44
5.1 Eyepieces ............................................................................................................................................................. 44
5.2 Darkfield Condensers.......................................................................................................................................... 44
5.3 Objective Lenses ................................................................................................................................................. 45
5.4 Phase Contrast Kits............................................................................................................................................. 45
5.5 Cameras and Accessories ................................................................................................................................... 46
5.6 Halogen Light Bulbs and Fuses .......................................................................................................................... 47
5.7 Cases and Bags ................................................................................................................................................... 47
5.8 Stage Warmer ...................................................................................................................................................... 47
5.9 Slides ................................................................................................................................................................... 48
5.10 Cleaners ............................................................................................................................................................ 48
5.11 Books ................................................................................................................................................................. 48
6. TROUBLESHOOTING .................................................................................................................................................... 49
6.1 Optical Issues ...................................................................................................................................................... 49
6.2 Mechanical Issues ............................................................................................................................................... 50
6.3 Electrical Issues .................................................................................................................................................. 51
7. GENERAL MICROSCOPY GUIDE ................................................................................................................................... 52
490 Series Microscope
AmScope Proprietary
Page 4 of 64
Before Use
Introduction
Congratulations on the purchase of your new AmScope Microscope! This manual is
designed for 490 series microscopes. Though the microscope’s body and
magnification levels may vary depending on your series model number, the 490
series functions remain the same.
Please take a few minutes to familiarize yourself with the features and functions of
your new microscope. If you want more information on microscope, parts, and
accessories, please visit our website at:
www.AmScope.com
We recommend that you study this manual thoroughly before operating the
microscope and that you keep it on hand for future reference. If you have additional
questions or need assistance, please send us an email at:
[email protected]
Please include the microscope model number in your email so that we can identify
your model and provide immediate help.
Precautions
1. As the microscope is a precision instrument, always handle it with care, avoiding
impact or abrupt movement during transportation. Do not shake the package.
2. Do not place the microscope in direct sunlight or in high heat. Keep it indoors in a
dry and clean place with temperatures between 32°-100° F (0°-40°C), maximum
relative humidity: 85%
3. Avoid touching the lenses on the objective and the eyepiece so that oil and dirt
from your fingerprints do not obstruct your view.
4. Before turning the power on, make sure that the power supply voltage is
consistent with the voltage of your microscope.
490 Series Microscope
AmScope Proprietary
Page 6 of 64
2. Operation
2.1 Unpacking
2.1.1 Unpack the Box
1. Very carefully slide the Styrofoam container out of the cardboard carton.
2. Lay the Styrofoam container on its side. Make sure the side labeled up is up.
3. Remove the tape.
4. Carefully open the Styrofoam container, avoid dropping and damaging the
optical items.
2.1.2 Verify Packing List
Check the packing list to ensure that you’re received all items:
� One Microscope Body
� One Binocular (B490) or One Trinocular (T490) Rotating Gemel Head
� One Adjustable Trinocular Photo Port (for T490 only)
� Four Objectives (4x, 10x, 40x, 100x)
� Pair of 10X Widefield Eyepieces
� Pair of 16X Widefield Eyepieces (for -A models only)
� Pair of 20X Widefield Eyepieces (for -B models only)
� Pair of 25X Widefield Eyepieces (for -C models only)
� Three Color Filters (Blue, Green, & Yellow), (for non-LED models only)
� One Bottle of Immersion Oil
� One 6V/20W Spare Halogen Bulb (for 20W models only)
� One 6V/30W Spare Halogen Bulb (for 30W models only)
� One Spare Fuse
� One Dust Cover
Note: LED models do not have a spare bulb, as the LED bulbs do not need
replacing.
490 Series Microscope
AmScope Proprietary
Page 7 of 64
2.2 Assembly
2.2.1 Carefully Remove Microscope from Box
Remove the microscope body from the box and
remove the plastic protective covering.
The body of the microscope is composed of the base,
the stage, the arm, and the nosepiece.
2.2.2 Loosen the Head Lock Screw
Loosen the metal knob located directly above the
nosepiece (the head lock screw) completely.
2.2.3 Install the Microscope Head
Carefully take the microscope head out of the plastic bag,
making sure not to touch glass.
Place the bottom side of the microscope head (flat,
circular side down) into the circular opening on top of the
arm (just above the nosepiece and lock screw).
Tighten the head lock screw to secure the head in
place.
490 Series Microscope
AmScope Proprietary
Page 8 of 64
2.2.4 Install the Trinocular Port and C-Mount (For Trinocular Microscopes
Only)
Unscrew and remove the photo port cap from the top side of the
microscope head.
Screw in the Trinocular port
If your camera has a c-mount, insert the c-mount adapter onto
the Trinocular port. If your camera does not have a c-mount, you
can skip this step.
2.2.5 Insert Eyepieces
Remove the eyetube caps.
Place the desired eyepieces into the empty eye/ocular tubes.
Be sure to avoid touching the lenses.
490 Series Microscope
AmScope Proprietary
Page 9 of 64
2.2.6 Insert Objective Lenses
Unpack the objective lenses. Screw the
objectives into the microscope nosepiece.
Start with the lowest magnification, and in order
of magnification, insert each lens, without
touching the glass.
2.2.7 Plug it in, turn it on, and adjust the dimmer if more light is needed.
Plug in the microscope and turn it on. If no light
emerges from the light source, adjust the dimmer
knob on the side of the base.
490 Series Microscope
AmScope Proprietary
Page 10 of 64
2.3 Adjusting the View
2.3.1 Choose forward or reverse viewing
The microscope head rotates to allow for forward or
reverse viewing.
There is no difference as far as the optics; it depends
on what’s comfortable. In forward position you are
looking over the stage, in reverse position you are
looking at the stage.
Once you’ve chosen your preferred microscope head
position, lock the head-lock screw.
2.3.2 Focus the distance between your eyes.
(The interpupillary distance, the distance between your eye pupils.)
With both eyes open, look into the eyepieces. Adjust the interpupillary distance
by holding the eyetubes and rotating the tubes either towards or away from each
other until only one circle of light is seen by both eyes.
490 Series Microscope
AmScope Proprietary
Page 11 of 64
2.4 Specimen Set Up
2.4.1 Prepare the Specimen
Place the specimen to be studied on a glass slide
(or use a prepared slide, sold separately).
2.4.2 Secure the Slide
Place the slide on the stage, holding it snugly in
place with the metal slide holders (clips) of the
mechanical stage.
2.4.3 Center the Specimen
Using the mechanical stage controls, center the
slide over the stage opening, lining it up with the
light and the objective lens.
The top ring moves the stage forward and
backward. The bottom ring moves the stage from
left to right
2.4.4 Adjust the dimmer if more light is needed.
To adjust the illumination, slowly turn the dimmer on
the right side of the base until the desired intensity
of light is achieved.
490 Series Microscope
AmScope Proprietary
Page 12 of 64
2.5 Focusing
2.5.1 Choose an Objective Lens
Turn the nosepiece to choose an objective. It is easiest to use
the lowest magnification first (4X objective) to locate and focus
on the specimen. As you move up in magnification you will
need to refocus the image a little each time.
2.5.2 Parfocal
Under the diopters on the eyepieces, there are numbered
rings. Set these rings to 0 and the image should remain in
focus for the entire magnification range. This will enable you to
stay more in focus when you add a camera or switch objective
lenses.
2.5.3 Focus one eye at time
1. Look with one eye through the right eyetube. Or,
if you have two diopters, select either eyepiece.
2. Close your other eye.
3. Focus the image using the coarse focusing knob.
Adjust the height of the stage until the sample
comes into clear focus.
Note: You can loosen the limit-stop knob (located on the inside of the coarse
focusing knob on the left side of the microscope) in order to give yourself the
full range of motion.
2.5.4 Focus your other eye
1. Close the first eye. Open the other one.
2. Look with the second eye through the
eyepiece.
3. Turn the diopter until the image is clear through
the side as well. Do not turn the focusing knob
when you are adjusting the diopter or you will
have difficulty focusing properly.
2.5.5 Re-center Specimen and Re-focus
When you change magnification levels, you may need to re-center the specimen.
When you change magnification, you will need to re-focus.
490 Series Microscope
AmScope Proprietary
Page 13 of 64
2.6 Using the Trinocular Port
The AmScope T490 models are uniquely designed so
that you can view the image through the eyepieces and
the Trinocular port simultaneously, as well as fine tune
the focus of the camera with a C-mount focus adjustment.
You do not need an adapter to attach your AmScope
camera to the Trinocular port; however you may need one
if you have a non-AmScope camera. Our photo port is a
23mm size.
2.7 Attaching a Camera
Most AmScope cameras come with a reduction lens.
Reduction lenses align the magnification of the
microscope’s image with the capacity of the digital
camera’s sensor.
The specimen’s image seen through the magnified
objective lens is very large. The reduction lens renders
the image in a size compatible with the smaller size of
the digital camera’s sensor.
1. If your camera has a 23mm mounting size, you do not
need the C-mount adapter. If installed, remove the C-
mount from the top of the Trinocular port by loosening
the screw.
2. Place the camera (mounted on a reduction lens)
directly into the Trinocular port. It should slide in
without issue.
3. Focus through the Trinocular port, by turning the
middle portion of the tube.
4. Open up your AmScope Digital Camera Solution
software on your computer and let it connect to your
microscope.
If your camera needs a C-mount adapter, attach the adapter to your camera and screw the
adapter into the Trinocular port. Plug one end of the cord directly into the camera and the
USB end into your computer.
490 Series Microscope
AmScope Proprietary
Page 14 of 64
2.8 Setting the Stage’s Stop-Limit
1. Unlock the stop-limit on the stage.
2. Adjust the stage to the desired maximum
height.
3. Lock the stop-limit. This allows you to limit
the movement of the stage from the bottom of
the range up to the point it is set at.
4. To reset it, unlock the stage and reset the
stage to the new height.
If no limit is desired, simply unlock the stop-limit.
2.9 Adjusting Focus Tension
1. To adjust the tension of the focusing knobs,
first locate the black ridged tension knob on
the inside of the coarse focusing knob.
2. To decrease tension, rotate the adjustment
forward, towards the stage
(counterclockwise). To increase, rotate away
from the stage (clockwise).
Note: If your stage is slipping down after setting the focus, you need to increase the
tension.
490 Series Microscope
AmScope Proprietary
Page 15 of 64
2.10 Setting the Condenser Lens Adjustment Knob
The condenser adjustment knob raises and lowers the
condenser’s distance to the base lens, varying light
delivery.
2.11 Adjusting the Iris
By changing the aperture (hole size) of the iris/diaphragm
of the condenser, you can adjust the background
brightness. Adjust the aperture of the iris diaphragm using
the iris adjustment slider located directly under the stage.
2.12 Changing a Filter
1. At the bottom of the condenser lens, there is a small
ring that slides out. Swivel the ring as far out as
possible.
2. Insert the preferred colored glass filter.
3. Slide the ring back into place.
Note: The filter holder is placed in from the factory in a manner in which it swings out
and hits the arm of the microscope (backwards). If this happens, simply grip
the condenser assembly and rotate it. It may take a small amount of force to
rotate it, but after doing so, you will be able to swing the filter holder out
towards the front of the unit for easier operation.
490 Series Microscope
AmScope Proprietary
Page 16 of 64
2.13 Changing the Halogen Light Bulbs
1. Turn off and unplug the microscope.
2. Wait until the bulb cools down.
3. Do not touch the bulb directly, the oil from
your skin will damage the bulb.
Use a cloth or paper to handle.
(Microscope Base)
1. Be sure to remove the eyepieces from the unit
before turning the microscope over to prevent
them from falling and breaking.
2. Carefully turn the microscope on its side,
exposing the base.
3. Loosen the door screw on the bottom of the
microscope and pull the door open.
4. Remove the old bulb and replace with a new
one.
5. Close door and tighten the screws of the
door.
2.14 Changing the Fuse
(Microscope Base)
1. Turn off and unplug the microscope.
2. Be sure to remove the eyepieces from the unit
before turning the microscope over to prevent
them from falling and breaking.
3. Carefully turn the microscope on its side,
exposing the base.
4. Using a small screwdriver, unscrew the cover and
replace the fuse.
5. Screw the fuse cover back in.
490 Series Microscope
AmScope Proprietary
Page 17 of 64
2.15 Using Oil Immersion
When using the 100X objective, a drop of immersion oil should be
placed between the cover slip and the objective to minimize distortion
caused by air.
1. Place the slide on the stage and center the
specimen.
2. Close the condenser’s aperture iris and ensure that the
specimen focused and centered as possible. Increase
light intensity once the specimen is clearly illuminated.
3. Move the objectives out of the way to enable access
to the slide. Place a drop of immersion oil directly
onto the slide’s cover slip, not directly on the
specimen; a cover slip MUST be used.
4. Release the stage stop limit.
5. Lower your oil immersion lens (100X objective) into
position so that the front lens is immersed in the oil. If
your lens is not immersed, use the coarse focusing
knobs to adjust the height of the stage to the proper
levels. The oil lens should be immersed in oil, but not
touching the slide.
490 Series Microscope
AmScope Proprietary
Page 18 of 64
6. Adjust the focusing knob, watch for the field of view to
come into focus. This is best seen when the iris of the
condenser is smaller than the field of the objective. Use
the fine focusing knob to make adjustments.
7. Re-engage the stage stop limit.
8. When the specimen is in focus, adjust the iris to open
just larger than the condenser’s diameter.
9. After using the oil immersion lens, you will need to clean the lens and
specimen before the oil dries.
Note: Please be careful when using the stage stop limit. The stage limit
stop is designed to prevent impact between objective and slide, so when it
is off you can damage the microscope or the slide.
490 Series Microscope
AmScope Proprietary
Page 19 of 64
2.16 Using Darkfield Condenser Lenses (Optional Accessories)
Darkfield Illumination is a lighting method for increasing contrast. A darkfield
condenser lens produces bright images against a dark background for low contrast
specimens.
The dry darkfield condenser is for low power magnifications,
such as with the 4X, 10X or 40X objective lenses, with
numeric apertures of .65 or less.
The oil darkfield condenser is for high power magnifications,
the 40X and 100X objective lenses, with numeric apertures
up to 1.25 or less.
2.16.1 Installing a Darkfield Condenser Lens
1. Use the condenser lens adjustment knob to lower the condenser so you have
easy access to the condenser knob.
2. Loosen the condenser lock screw and replace the
condenser lens.
3. Install the darkfield condenser lens and tighten the
condenser lock thumb screw on the condenser holder.
When choosing a darkfield condenser, be sure the
condenser’s numerical aperture is larger than the
numerical aperture of the objective lens.
If the objective lens aperture is larger than the aperture of
the darkfield condenser, direct light will be transmitted and
a darkfield will not be achieved.
490 Series Microscope
AmScope Proprietary
Page 20 of 64
2.17 Using a Dry Darkfield Condenser Lens (Optional Accessory)
2.17.1 Using the Dry Darkfield Condenser
1. The first step is centering the condenser. Move the 10X objective lens into the
light path.
2. Remove an eyepiece and look through
the empty eye tube, to see the
condenser’s image directly.
3. Use the condenser lens adjustment
knob to adjust the height of the
condenser until the opaque disc is
slightly smaller than the entire image
circle, leaving a ring of light.
4. Turn the condenser’s centering screws
to move the dark spot to the center.
5. Replace the eyepieces.
6. Raise the condenser to the top of its
range, just short of the slide on the
stage. The is the optimal position for
darkfield viewing, but can be adjusted
later, if necessary.
7. Place a slide on the stage and focus. Adjust
the height of the stage up or down slightly
for optimal viewing, but the condenser lens
should be relatively close to the slide.
8. If you are using a variable numerical
aperture objective lens, adjust the ring
at the bottom to fine tune the darkfield.
490 Series Microscope
AmScope Proprietary
Page 21 of 64
2.18 Using Oil Darkfield Condenser Lens (Optional Accessory)
The condenser works with 100X oil objective and low power objectives (4X, 10X, etc) as
well. Immersion oil is needed on top of the condenser lens for darkfield observation.
Immersion oil is needed between the 100X oil objective and slide.
2.18.1 Using the Oil Darkfield Condenser
1. The first step is centering the condenser. Move the 40X objective lens into the
light path.
2. Remove an eyepiece and look through
the empty eye tube, to see the
condenser’s image directly.
3. Use the condenser lens adjustment
knob to adjust the height of the
condenser until the opaque disc is
slightly smaller than the entire image
circle, leaving a ring of light.
Note: The dark spot has a fuzzy edge and will become larger when the condenser
gets closers to the objective. The distance between the top of the condenser
and the objective is about 8mm, when the dark spot shows in the field of view.
4. Turn the condenser’s centering screws to move
the dark spot to the center.
5. Replace the eyepieces.
6. Raise the condenser to the top of its range, just
short of the slide on the stage. The is the optimal
position for darkfield viewing, but can be adjusted
later, if necessary.
7. Adjust the dimmer so the lamp is at maximum
brightness.
8. Place a slide on the stage and focus. Adjust the height of the stage up or
down slightly for optimal viewing, but the condenser lens should be relatively
close to the slide.
490 Series Microscope
AmScope Proprietary
Page 22 of 64
2.18.2 Using Oil Immersion
1. Before you install the oil darkfield condenser lens,
apply a drop of immersion oil on the top lens of
condenser. If you’ve already installed it, remove
the condenser to apply the drop of oil to the lens,
and re-install it. You might need to re-center the
condenser.
2. Place the slide on the stage.
3. Raise the condenser and let the oil drop contact
the underside of the slide. If air bubbles exist in
the oil, clean the oil from the condenser lens and
bottom of slide and repeat the procedures.
4. Apply a drop of immersion oil to the slide’s
coverslip.
5. Carefully raise the stage until the oil makes
contact with the objective lens.
6. Adjust the iris ring to get proper brightness and
contrast of view field.
7. If you are using a variable numerical aperture
objective lens, adjust the ring at the bottom to
fine tune the darkfield.
490 Series Microscope
AmScope Proprietary
Page 23 of 64
2.18.3 Cleaning the Oil Darkfield Condenser
1. When using oil immersion, once you have finished using the oil darkfield lens,
you will need to clean the lens and specimen of the oil before it dries.
2. You can wipe the lens with a soft nonabrasive cloth to remove the oil. If you
want to use a cleaner, you can use cigarette lighter fluid to safely remove the
oil, as it will break it down. Do not use alcohol based solvents, as they
dissolve the cement used to in assembly of the lens (thus breaking your
objective).
3. The specimen itself can be discarded, or if desired, you could clean it up and
store it for later use. The specimen may also be left with the oil on it, as you
will still need to apply more for use later, however it may dry up and cloud
your specimen. Cleaning is recommended.
490 Series Microscope
AmScope Proprietary
Page 24 of 64
2.19 Using a Phase Contrast Kit (Optional Accessory)
Brightfield
Phase Contrast
Some specimens may offer little in the way of color-variance or opacity and can
appear essentially invisible when using basic, transmitted light. Phase-contrast is a
technique used in microscopy to enhance the contrast of an image, when observing
a specimen which offers little contrast in normal bright-field conditions.
Light waves have several
measurable components,
the most-common of which
are length, frequency and
amplitude. While length and
frequency translate into the
color of light we see,
amplitude translates into
brightness.
One component that we do
not perceive is phase, which
can be described as “how
much of the wave’s cycle
has elapsed at a given
time.”
If you imagine light as a
small dot (a photon), moving
along the cyclical arcs of a
light wave, the phase is
current position of the
photon, relative to the
wave’s entire cycle.
490 Series Microscope
AmScope Proprietary
Page 25 of 64
As waves of light move
through space, if their cycles
move in a perfect parallel
and are at the same point in
their cycles, they are in-
phase.
If some of the light waves hit
an obstacle and the phases
of the waves are altered, so
they are no longer in parallel
and at the same point in
their cycles, they are out-of-
phase.
When a light wave passes
through any medium, the
medium can affect the light
wave’s length, amplitude,
frequency and phase. There
are shifts in phase, due to
changes in light’s speed.
As the speed of light changes when hitting or missing obstacles, like a specimen, the
rays are bent, or refracted. The amount of refraction when a light wave hits an obstacle
is measured as its refractive index. It’s this refractive index which creates the variations
in phase as light passes through a specimen.
Even though a specimen may appear transparent, as long as its refractive index is
different from the baseline index of air, and the specimen has thickness, then there will
be changes in phase to any light which passes through the specimen.
490 Series Microscope
AmScope Proprietary
Page 26 of 64
A phase contrast system incorporates two
components to segregate light, so out-of-
phase waves can be realigned. The two
components are the annulus in the
condenser and the phase ring in the objective
lens.
The condenser’s annulus separates light into
surround light and diffracted light. Light
passing directly through the annulus creates
a hollow cone called surround light. This light
passes directly through the optical system
and is intercepted by the phase ring in the
objective lens.
Light diffracted by the edges of the annulus
passes more centrally through the specimen
in the objective lens.
As the light collectively passes through the objective lens, it is further segregated by the
phase ring. The phase ring shifts the phase of the surround light either positively or
negatively so the waves are in-phase.
Since human sight is incapable of perceiving light phase, the aim of phase contrast is to
translate the disparities in phase into perceivable variations in brightness.
There are several techniques
available for increasing the contrast
between the specimen and the
background, each tailored for a
particular purpose. Phase-contrast
does so by manipulating the phase
of light, and is useful for observing
the inner mechanics of cellular
structures, especially live cells,
without the need for staining.
490 Series Microscope
AmScope Proprietary
Page 27 of 64
AmScope’s PCS Simple Phase Contrast Kit
includes three phase-contrast objective lenses,
three phase contrast condenser lenses, and a
centering telescope.
AmScope’s PCT Turret Phase Contrast Kit
includes four phase-contrast objective lenses, a
condenser turret, and a centering telescope. The
turret has built-in annuli, designed to work with
each of the lenses, as well as a brightfield iris.
490 Series Microscope
AmScope Proprietary
Page 28 of 64
2.20 Using a Simple Phase Contrast Kit (Optional Accessory)
2.20.1 Unpacking the Simple Phase Contrast Kit
1. Take the aluminum case containing your kit out of the cardboard carton.
Remove the tape and open the container carefully so as to avoid dropping
and damaging the optical items. You may need to use the key taped onto the
box. Check carefully to ensure that all parts and accessories are intact.
Note: You must match finite tube-length kits to finite tube-length microscopes
and infinity-corrected units to infinity-corrected microscopes for optimal
results.
2. Check the packing list to ensure that you’re received all the items. All units
should come with:
� 10X, 40X, & 100X Phase Contrast Condenser Lenses
� 10X, 40X, & 100X Phase Contrast Objective Lenses
� Condenser Assembly
� Centering Telescope Lens
� Green filter
490 Series Microscope
AmScope Proprietary
Page 29 of 64
2.20.2 Focus the Specimen
1. Once you have installed the phase contrast kit, you will need to calibrate and
center the light perfectly for optimal results. You will initially calibrate with the
Abbe brightfield condenser.
2. Place a slide on the microscope stage.
Make sure the specimen is covered
with a cover slip (a good example for a
phase contrast specimen is human
saliva). Clean any excess material
from the slide.
3. Make sure your Abbe brightfield condenser is installed.
4. Focus the specimen using the 10X
objective. Adjust the iris ring on the
condenser so the light field is just
larger than the field of view in the
microscope. The condenser itself
should be moved to the normal height,
which is close to the bottom of the
stage of the microscope (the top of the
range of the condenser).
2.20.3 Installing a Simple Phase Contrast Condenser
1. Loosen the thumbscrew and remove the
existing Abbe brightfield condenser from
your microscope.
2. Replace the brightfield objective on nosepiece with the 10X phase contrast
objective.
3. Screw the 10X condenser ring plate onto
the condenser assembly. Install the ring
plate and assembly on the microscope’s
condenser ring.
490 Series Microscope
AmScope Proprietary
Page 30 of 64
4. Change the objective lens to the 10X
phase contrast lens. Objective lenses
designed for phase contrast will be
marked “Ph”, this denotes the objective
lens has a built-in phase ring.
Note: there are 3 phase contrast objectives: 10X, 40X and 100X, and
there are 3 condenser ring plates: 10X, 40X and 100X. The
corresponding objective and ring plate must work together. For
example, the 10X phase contrast objective must work with the 10X
condenser ring plate.
2.20.4 Calibrating the Phase Contrast Objective Lens and Condenser Lens
1. Remove one eyepiece from the
microscope eyetube and insert the
centering telescope.
2. Turn the microscope’s light on and look
through the telescope lens.
3. While looking through the eyepiece,
extend the eyepiece by rotating the top
counter clockwise until an image of the
objective’s phase ring and the
condenser’s annulus are clearly in
focus. Since they are both part of the
same conjugate image plane, they can
be simultaneously in focus.
490 Series Microscope
AmScope Proprietary
Page 31 of 64
4. If the bright ring is still obscure, raise or
lower the condenser by adjusting the
condenser focusing knob (or the
microscope focusing knob if necessary)
until the bright ring is in focus and the
dark ring is visible.
5. If the two ring images are not
positioned so one is inside the other,
adjust the centering screws on the
condenser ring plate until the two rings
center in parallel.
The annulus’ bright ring of
light should overlap the
dark grey ring, which is the
objective’s phase ring.
6. Place the green filter in the filter holder
or directly onto the center of the base
lens of your microscope (where the
light emerges from). Remove the
green filter if doing darkfield
microscopy.
7. Remove the centering telescope and replace the eyepiece.
Note: Once aligned, repeat the process for each magnification.
490 Series Microscope
AmScope Proprietary
Page 32 of 64
2.20.5 Perform the Phase Contrast Observation
After you center the condenser, you can perform the phase contrast observation
the same way as a normal brightfield microscope.
Note: When changing to another phase contrast objective and corresponding
condenser ring plate, the focusing and centering of bright ring and dark
ring should be repeated following the procedures above.
Tips:
1. Make the illumination as bright as possible.
2. The thinner the specimen, the better the image.
490 Series Microscope
AmScope Proprietary
Page 33 of 64
2.21 Using a Turret Phase Contrast Kit (Optional Accessory)
2.19.1 Unpacking the Turret Phase Contrast Kit
1. Take the aluminum case containing your kit out of the cardboard carton.
Remove the tape and open the container carefully so as to avoid dropping
and damaging the optical items. You may need to use the key taped onto the
box. Check carefully to ensure that all parts and accessories are intact.
Note: You must match finite tube-length kits to finite tube-length microscopes and
infinity-corrected units to infinity-corrected microscopes for optimal results.
2. Check the packing list to ensure that you’re received all the items. All units
should come with:
� Turret with 10X, 20X, 40X &100X Phase Contrast Condensers
� 10X, 20X, 40X & 100X Phase Contrast Objectives
� Centering Telescope Lens
� Green filter
� Small flathead screwdriver
490 Series Microscope
AmScope Proprietary
Page 34 of 64
2.21.2 Installing the Turret Phase Contrast Kit
1. Loosen the lock screw and remove the Abbe
brightfield condenser from the holder.
2. Insert the phase contrast condenser turret
into the condenser holder and tighten the lock
screw with the screw driver. Due to the
weight of the turret, it’s best to use the screw
driver to securely fasten the turret to the
condenser holder.
2.21.3 Focusing the Turret Phase Contrast Kit
1. Rotate the microscope’s nosepiece to the 10X
phase contrast objective lens. Objective
lenses designed for phase contrast will be
marked “Ph”, this denotes the objective lens
has a built-in phase ring.
Note: There are 4 phase contrast objectives:
10X, 20X, 40X and 100X, and there are 5
settings on the condenser’s turret ring: 10, 20,
40, 100 and B. The numbers on the turret ring
must always correspond with objective’s
numbers.
For example, when using a 10X phase contrast
objective the condenser’s turret ring setting
must be 10. The B setting will make the phase
contrast condenser work as a conventional
brightfield condenser.
490 Series Microscope
AmScope Proprietary
Page 35 of 64
2. Place a slide on the microscope stage.
Make sure the specimen is covered with a
cover slip (a good example for a phase
contrast specimen is human saliva). Clean
any excess material from the slide.
3. Turn the phase contrast condenser’s turret to the B setting. Focus the
specimen. Before aligning the phase contrast lens and condenser, you will
need to bring the specimen into focus with the brightfield setting.
4. Once the specimen is in focus with the brightfield setting, rotate the
condenser’s turret ring so it is set to 10.
2.21.4 Calibrating the Turret Phase Contrast Objectives and Condenser Lenses
The turret has a built-in annulus for each
objective magnification. Once the turret
and objective lens are set with matching
numbers, they will need to be aligned.
You will need to do this again each time
you change the objective lens.
Each annulus in the turret is easy to
reposition by using a finger.
1. Remove one eyepiece from the
microscope eyetube and insert the
centering telescope.
2. Turn the microscope’s light on and
look through the telescope lens.
490 Series Microscope
AmScope Proprietary
Page 36 of 64
3. While looking through the telescope
lens, extend the eyepiece by rotating the
top counter clockwise.
4. Keep rotating the telescope lens until
an image of the objective’s phase ring
and the condenser’s annulus are clearly
in focus.
5. By slowly moving the annulus on the
underside of the turret, now in-line with
the condenser’s lens, the annulus’ bright
ring of light should overlap the dark grey
ring, which is the objective’s phase ring
6. If the bright ring is still obscure, raise or
lower the condenser by adjusting the
condenser focusing knob (or the
microscope focusing knob if necessary)
until the bright ring is in focus and the
dark ring is visible.
7. Place the green filter in the filter holder
or directly onto the center of the base
lens of your microscope (where the
light emerges from). Remove the
green filter if doing darkfield
microscopy.
8. Remove the centering telescope and replace the eyepiece.
Note: Once aligned, repeat the process for each magnification.
490 Series Microscope
AmScope Proprietary
Page 37 of 64
2.21.5 Perform the Phase Contrast Observation
After you center the ring plate, you can perform the phase contrast observation
the same way as a normal brightfield microscope.
Note: When changing to another phase contrast objective and corresponding
condenser ring plate, the focusing and centering of bright ring and dark
ring should be repeated following the procedures above.
Tips:
1. Make the illumination as bright as possible.
2. The thinner the specimen, the better the image.
490 Series Microscope
AmScope Proprietary
Page 38 of 64
2.22 Microscope Maintenance
2.22.1 Handle with Care
As the microscope and all accessories are precision instruments, always handle
them with care, avoiding impact or abrupt movement during transportation. Do not
shake the package.
2.22.2 No Sun/Heat
Do not place the microscope and accessories in direct sunlight or in
high heat.
Keep the microscope and all accessories indoors in a dry and clean
place with temperatures between 32°-100° F (0°-40°C), maximum
relative humidity: 85%.
2.22.3 No Dust
Always keep the microscope covered by the dust cover when not in
use. The glass will attract dust, obscuring your view. Make sure to
keep it in a dry and clean place in order to prevent rust.
2.22.4 Keep Glass Clean
All glass surfaces must always be kept clean.
Oil and dirt from your fingerprints will obstruct your view.
Fine dust on the optical surface should be blown off using a hand
blower or gently wiped off with a soft lens tissue.
2.22.5 Use an optical lens cleaner.
Carefully wipe off oil or fingerprints on lens surfaces using a tissue
moistened with a small amount of 3:7 alcohol to ether mixture.
2.22.6 Do not use lens cleaner on the non-glass parts of the microscope.
Do not use the optical lens cleaning solution to wipe the surfaces of the
other components of the microscope.
All other parts, especially those made of plastic, should be cleaned with
a mild detergent.
2.22.7 Do not touch the glass.
Avoid touching the lenses on the objective, the eyepiece and all the
accessories with glass lenses.
490 Series Microscope
AmScope Proprietary
Page 39 of 64
2.22.8 Do not take apart the microscope.
Do not assemble or disassemble the microscope’s electrical components
yourself.
.
2.23 Darkfield Condenser Kit & Phase Contrast Kit Maintenance
2.23.1 Secure Setting
Ensure that the microscope is located on a smooth, level and firm
surface.
2.23.2 Do Not Take Apart
Do not attempt to disassemble any components.
2.23.3 Keep Kits Clean
Keep the kits clean; remove dirt and debris regularly.
Accumulated dirt on outer surfaces should be cleaned with a
damp cloth. More persistent dirt should be removed using a mild
soap solution. Do not use organic solvents for cleansing.
2.23.4 Keep Lens Clean
The outer surface of the optics should be inspected and cleaned periodically using
an air stream from an air bulb. If dirt remains on the optical surface, use a soft cloth
or cotton swab dampened with a lens cleaning solution (available at the camera
stores).
All optical lenses should be swabbed using a circular motion. A small amount of
absorbent cotton wound on the end of a tapered stick makes a useful tool for
cleaning recessed optical surfaces.
Avoid using an excessive amount of solvents as this may cause problems with
optical coatings or cemented optics or the flowing solvent may pick up grease
making cleaning more difficult.
2.23.5 Cool and Dry Storage
Store the instrument in a cool, dry environment. Put the condenser back in the
storage box when not in use.
490 Series Microscope
AmScope Proprietary
Page 40 of 64
3. General Specifications
3.1 General Model Comparison
PARTS
SPECIFICATIONS
B490
T490
B490-
30W
B490-
LED
T490-
30W
T490-
LED
WF Eyepiece
WF10X/18mm
●
●
●
●
●
●
WF10X/18mm w/ Pointer
WF10X/18mm w/ Reticle
WF16X/13mm
● (if -A)
● (if -A)
● (if -A)
● (if -A)
● (if -A)
● (if -A)
WF20X/10mm
● (if -B)
● (if -B)
● (if -B)
● (if -B)
● (if -B)
● (if -B)
WF25X/9mm
● (if -C)
● (if -C)
● (if -C)
● (if -C)
● (if -C)
● (if -C)
DIN Achromatic Objectives
4X/0.10
●
●
●
●
●
●
10X/0.25
●
●
●
●
●
●
40X (spring)/0.65
●
●
●
●
●
●
100X (spring, oil)/1.25
●
●
●
●
●
●
30 Degree Viewing Head
Binocular Sliding
360 Degree Swiveling
●
●
●
Trinocular Sliding
360 Degree Swiveling
●
●
●
Trinocular Port
C-Mount Adjustable Photo port
●
●
●
23mm Adjustable Photo Port
●
●
●
Simultaneous Viewing
●
●
●
Focusing Nosepiece
Coaxial Focusing System
●
●
●
●
●
●
Quadruple Nosepiece
●
●
●
●
●
●
Stage
130mm x 140mm
Mechanical Stage
●
●
●
●
●
●
Movement Range
50mm x 76mm
●
●
●
●
●
●
Condenser
Abbe, NA= 1.25
●
●
●
●
●
●
Diaphragm
Iris Diaphragm
●
●
●
●
●
●
Illumination
Halogen Light w/ Dimmer
●
●
●
●
LED Light w/ Dimmer
●
●
Lamp
6V/20W
●
●
6V/30W
●
●
LED
●
●
Filter
Blue/Yellow/Green
●
●
●
●
490 Series Microscope
AmScope Proprietary
Page 41 of 64
3.2 Objectives
Type
Model
Number
Magnification
Numerical
Aperture
(N.A.)
Rating
Medium
Magnification
Marks
(Color Ring)
DIN Achromatic
Objective with
Associated N.A.
Ratings
(195mm)
A4X
4X
0.10 Air
Red
A10X
10X
0.25 Air
Yellow
A40X
40X
0.65 Air
Light Blue
A60X
60X
0.85 Air
Dark Blue
A100X
100X
1.25 Cedar Oil White
Plan Objective
(195mm)
PA4X
Plan 4X
0.10 Air
Red
PA10X Plan 10X
0.25 Air
Yellow
PA40X Plan 40X
0.65 Air
Light Blue
PA100X Plan 100X
1.25 Cedar Oil White
3.3 Eyepieces
Eyepieces
Magnification
5X
10X
15X
16X
20X
25X
Field of View
Φ18
Φ18
Φ13
Φ15
Φ11
Φ9
3.4 Darkfield Condensers
Specifications
Dry
Darkfield
Condenser
Oil
Darkfield
Condenser
Model
DK-DRY100
DK-OIL100
Numerical Aperture
(N.A.) Rating
0.7 – 0.9
1.36 – 1.25
Mounting Size (Diameter)
37mm
37mm
490 Series Microscope
AmScope Proprietary
Page 42 of 64
3.5 Phase Contrast Kits
Specifications
Simple Phase Contrast
Kit
Turret Phase Contrast
Kit
Model
PCS
PCT
Phase Contrast Objective
Achromatic 10X/0.25, 160/0.17
with built-in phase plate
Plan phase contrast 10X/0.25,
160/0.17
Plan phase contrast 20X/0.40,
160/0.17
Achromatic 40X/0.65, 160/0.17
with built-in phase plate, spring
Plan phase contrast 40X/0.65,
160/0.17
Achromatic 100X/1.25, 160/0.17
with built-in phase plate, spring, oil
Plan phase contrast 100X/1.25,
160/0.17, Oil
Numerical Aperture (N.A.)
Rating
1.25
1.25
Turret Ring Plates
1 For 10X phase contrast objective
10 for 10X phase contrast objective
20 for 20X phase contrast objective
1 For 40X phase contrast objective
40 for 40X phase contrast objective
1 For 100X phase contrast
objective
100 for 100X phase contrast
objective
Centering Telescope
Focusing adjustable
Focusing adjustable
490 Series Microscope
AmScope Proprietary
Page 43 of 64
4. Parameters
4.1 Electrical System
There are two options for electrical systems for this series of microscope. The light
source is dependent on which model you have, but can either be a 6V/20W halogen, a
6V/30W halogen, or an LED system.
1. 100V-120V power supply: 100V-120V ±10%, 50/60Hz.
2. 220V-240V power supply: 220V-240V ±10%, 50/60Hz.
All units come standard as 110V units unless an upgrade to a 220V system is
requested. Upgrade fee is dependent on which unit is purchased.
4.2 Microscope Parameters
Magnification
-A Model: 40x-1600x
-B Model: 40x-2000x
-C Model: 40x-2500x
Field of View
Φ0.8mm~Φ4.5mm
Mechanical Tube Length
160mm
Object to Primary Image Distance
195mm
Fine Focusing Sensitivity
0.002mm
490 Series Microscope
AmScope Proprietary
Page 44 of 64
5. Recommended Accessories
(Purchase separately. Please visit www.amscope.com and search with SKU #, for more
information.)
5.1 Eyepieces
5X Eyepieces
20X Eyepieces
25X Eyepieces
SKU: EP5X23
SKU: EP20X23
SKU: EP25X23
10X w/ Pointer
10X w/ Reticle
SKU: EP10X23P
SKU: EP10X23R
5.2 Darkfield Condensers
Dry Darkfield Condenser
Oil Darkfield Condenser
SKU: DK-DRY100
SKU: DK-OIL100
490 Series Microscope
AmScope Proprietary
Page 45 of 64
5.3 Objective Lenses
Achromatic
2X
5X
20X
60X
SKU: A2X
SKU: A5X
SKU: A20X
SKU: A60X
Plan Achromatic
Plan 4X
Plan 10X
Plan 40X
Plan 100X
SKU: PA4X
SKU: PA10X
SKU: PA40X
SKU: PA100X
5.4 Phase Contrast Kits
Simple
Turret
SKU: PCS
SKU: PCT
490 Series Microscope
AmScope Proprietary
Page 46 of 64
5.5 Cameras and Accessories
Cameras
To capture images, video, or view live display on a computer (PC/Mac OS X). Real-
Time Live Video Microscope Digital Camera, Reduction Lens, High Sensitivity
Sensor and Deluxe Calibration Kit.
350K
Pixel
1.3 Mega
Pixel
3 Mega
Pixel
5 Mega
Pixel
8 Mega
Pixel
9 Mega
Pixel
10 Mega
Pixel
SKU:
SKU:
SKU:
SKU:
SKU:
SKU:
SKU:
MU035
MU130
MU300
MU500
MU800
MU900
MU1000
Calibration Micrometer
Calibrate the camera software for on-screen measurements.
SKU: MR400
Charge-Coupled Device Television
CCD Digital
(VGA, Trinocular Only)
CCD TV/Video
(Trinocular Only)
1920x1080 Full HD HDMI
(Trinocular Only)
SKU: CCD-MT
SKU: CCD-NP
SKU: HD1080
To view live display on a
computer monitor (VGA).
To view live display on a
television (RCA).
To view live display on a
computer monitor (VGA).
490 Series Microscope
AmScope Proprietary
Page 47 of 64
5.6 Halogen Light Bulbs and Fuses
6V 20W
Halogen Bulb
6V 30W
Halogen Bulb
Fuses For
Microscopes
SKU: BH-
6V20W
SKU: BH-
6V30W
SKU: FS-M
T1AL 220V
5.7 Cases and Bags
Aluminum Case
Hard Wood Case
Vinyl Bag
SKU: AC-B400
SKU: WC-B400
SKU: CB-B490
5.8 Stage Warmer
Stage Warmer
TCS-100
Microscope Temperature
Control Stage Slide
Warmer
490 Series Microscope
AmScope Proprietary
Page 48 of 64
5.9 Slides
Prepared
25 Slides
50 Slides
100 Slides
200 Slides
SKU: PS25W
SKU: PS50
SKU: PS100A
SKU: PS200
Blank
50 Blank Slides
100 Blank Slides
144 Blank Slides
300 Blank Slides
SKU: BS-50P-C
SKU: BS-100P-
100S-22
SKU: BS-144P-
200S-22
SKU: BS-300P-
300S
5.10 Cleaners
Sparkle Microscope
3 in 1 Professional
3 in 1 Cleaning Kit
Optical Lens Cleaner
Cleaning Kit for
for Microscopes,
SKU: CLS
Microscopes, Cameras
Cameras and
and Laptops
LCD Screens
SKU: CK-I
SKU: CK-II
5.11 Books
Complete Book of the
Microscope
The Ultimate Guide to
Your Microscope
The World of the
Microscope
(What to look for)
Ages 8-Adult
(Science and
Experiments)
Ages 8-Adult
Ages 8-12
SKU: BK-CM
SKU: BK-UG
SKU: BK-WM
490 Series Microscope
AmScope Proprietary
Page 49 of 64
6. Troubleshooting
6.1 Optical Issues
SYMPTOM
SYMPTOM
CAUSE
REMEDY
OPTICAL ISSUES
One side of the field of view is
darker
The nosepiece is misaligned.
Turn the nosepiece until it clicks into
place.
Stains, dust, or dirt has
accumulated on the objectives
or eyepieces.
Clean all lenses with lens cleaner or a lint
free non-abrasive cloth.
Obstructions are observed in
the field of view
Stains, dust, or dirt has
accumulated on the specimen.
Clean the slide or use a new specimen if
sample is destroyed.
Stains, dust, or dirt has
accumulated on the objective,
eyepieces, or Barlow lens.
Clean the lens.
Unclear Image
There is no cover slip on the
slide.
Add a cover slip. The objectives are
designed for use with a 0.17mm cover
slip, so it is a requirement to use one for
proper images.
The cover slip is not standard
sized.
Replace the cover slip with the
appropriate 0.17mm thickness slip.
The immersion oil has
accumulated on the dry
objective.
Thoroughly clean the objective lens with
lens cleaner or a lint free non-abrasive
cloth
No immersion oil is used with
the 100X objective.
Use immersion oil for better clarity and
resolution.
Used wrong oil.
Use standard cedar wood oil.
The aperture is not open to an
appropriate diameter.
Adjust the aperture to have the light just
larger than
the size of the condenser.
Stains, dust, or dirt has
accumulated on the inlet of the
head.
Clean the lens with lens cleaner or a non-
abrasive lint free cloth, as well as spray
with compressed air.
The condenser is not in the
right position.
Adjust condenser height to the top of the
travel range, and then adjust down to
focus image.
Initial focus was at too high a
magnification.
Start focusing with the lowest objective
and then switch to the higher ones.
The slide is upside down
The slide thickness at higher
magnifications will impact clarity if the
slide is upside down.
The color of the image is not
accurate
The brightness adjustment
knob is not in the right position.
Adjust the brightness knob to a higher or
lower setting for color
clarity.
No filter is used or filter is in
use.
Remove color filter if natural light is
desired, or insert desired
filter.
490 Series Microscope
AmScope Proprietary
Page 50 of 64
SYMPTOM
CAUSE
REMEDY
OPTICAL ISSUES
One side of the field of
view is dark or the image
moves while focusing
The specimen slide is not fixed.
Secure the slide to the stage with
clips.
The nosepiece is not in the right
position.
Turn the nosepiece until it clicks into
place.
The field of view is not
bright enough
The iris diaphragm is too small.
Adjust the iris diaphragm to allow the
light to be just larger than the
condenser.
The condenser is not in the right
position.
Adjust condenser height to the top of
the travel range, and then adjust
down to focus image.
Stains, dust or dirt has accumulated
on the condenser, objective,
eyepieces, or base lens.
Clean the lens with lens cleaner or a
non-abrasive lint free cloth.
6.2 Mechanical Issues
SYMPTOM
CAUSE
REMEDY
MECHANICAL ISSUES
Unable to move the
slide smoothly.
The slide is not secured
correctly.
Adjust the slide to use the stage clips and secure
the sample.
The mechanical stage is not
properly secured.
Tighten the mechanical stage screws to better
secure the stage.
The objective
touches the cover
slip.
The cover slip is not standard
sized.
Replace the cover slip with the appropriate 0.17mm
thickness slip.
The limit-stop is set too high
or not engaged.
Be careful to avoid contact between objective and
the slide when the limit stop is not engaged (unless
using the 100X objective with oil). To re-engage,
focus the sample, then lock the limit stop into place
to set maximum height at a safe but usable
distance.
Focus-knob does
not turn.
The tension knob is too tight.
Loosen it by adjusting the tension ring inside the
coarse focus knob counterclockwise (close to the
arm of the microscope on the left of the
microscope).
Stage declines by
itself.
The tension knob is too loose.
Tighten it by adjusting the tension ring inside the
coarse focus knob clockwise (close to the arm of
the microscope on the left of the microscope).
The coarse
focusing knob won’t
raise the stage.
Limit-stop is engaged.
Disengage the limit stop on the left side of the
microscope inside the coarse focusing knob.
The fine focusing
knob won’t raise the
stage.
Limit-stop is engaged.
Disengage the limit stop on the left side of the
microscope inside the coarse focusing knob.
490 Series Microscope
AmScope Proprietary
Page 51 of 64
6.3 Electrical Issues
SYMPTOM
CAUSE
REMEDY
ELECTRICAL ISSUES
The bulb/light source
flickers.
The bulb is close to
burning out.
Replace the bulb. This unit uses our BH-
6V20W with our 20w unit or our BH-
6V30W for the 30W unit.
The microscope does not
light up.
The microscope is
unplugged.
Insert the plug into the wall socket to
achieve electrical illumination.
The bulb is not inserted
correctly.
Check the bulb by unscrewing the base
(remove eyepieces first to prevent falling
out) door and ensuring that the bulb is
inserted.
The bulb burned out.
Replace the bulb. This unit uses our BH-
6V20W with our 20w unit or our BH-
6V30W for the 30W unit.
The fuse burned out.
Replace with fuse on the bottom of the
microscope.
The fuse burns out
frequently.
The voltage is too high.
Use the correct power supply (110v if
110v unit, 220v if 220v unit), or get a
voltage adapter to convert to the proper
electrical system.
The bulb burns out
frequently.
The voltage is too high.
Use the correct power supply (110v if
110v unit, 220v if 220v unit), or get a
voltage adapter to convert to the proper
electrical system.
Used wrong bulb.
Use the correct wattage bulb for the unit.
Using a higher wattage than it is rated for
can damage your unit (melt components
with additional heat), so please be sure to
use the correct one. Damage from
incorrect usage is not covered under
warranty.
490 Series Microscope
AmScope Proprietary
Page 52 of 64
7. General Microscopy Guide
Microscopes come in a wide variety of types with many different features. Each
AmScope model is designed for specific uses and specific users. This glossary will help
illustrate what the variations mean and why they are useful.
Compound and Stereo Microscopes
Using a compound or a stereo microscope depends on the specimen being studied.
Compound microscopes are best for smaller
transparent specimens, like slides and biological
subjects.
Compound microscopes show a two-dimensional image
of the specimen (usually reversed and upside-down)
The common magnification range of our compound
microscopes is between 40X-1000X, could be up to
2500X. You need a minimum of 400X to study cell
structure.
Stereo microscopes are best for larger specimens you
cannot see through.
Stereo microscopes show a 3D image. Three
dimensional imaging is perfect for performing
dissections, repairing circuit boards, studying fossils
and gems or examining any specimen where you want
to use your hands.
The magnification range of our stereo microscopes is
between 2X to 225X.
490 Series Microscope
AmScope Proprietary
Page 53 of 64
Base Lens
Base lenses direct the light source towards the specimen.
Binocular Head-Rotating
A rotating binocular head is designed for maximum
flexibility in viewing options.
Binocular microscope heads allow for the easy
adjustment of the interpupillary distance (the distance
between the eyes), by moving the eyepieces toward
or away from each other...
The rotating feature allows for either forward or
reverse view, depending on whether you’d rather look
at the specimen or over it.
C-Mount
C-Mounts are adapters with a standard thread for mounting a
lens to a camera. C-Mounts fit onto a Trinocular port.
Condenser Lens
A condenser lens
concentrates light on a
specimen and increases the
resolution. An adjustable iris
controls the diameter of the
beam of light entering the
lens system. Abbe
Condenser lenses are
specially designed to mount
under the stage.
490 Series Microscope
AmScope Proprietary
Page 54 of 64
Condenser Lens Adjustment Knob
The Condenser Lens Adjustment Knob changes the
distance between the light condenser and the base lens.
This allows you to control the concentration of the light
hitting your slide.
Darkfield Condenser Lens
A darkfield condenser lens produces bright images against a dark background for low
contrast biological specimens.
With brightfield lenses, a solid
beam of light illuminates the
objective lens. Darkfield condenser
lenses block the center of the
beam of light to produce a hollow
cone of light. Light does not enter
the objective lens directly. The
specimen scatters light, which then
enters the objective lens.
Brightfield
Darkfield
Dimmer
Dimmers control the amount of light escaping from
the base lens.
DIN/JIS
DIN is the Deutsches Institut für Normung, the German Institute for Standardization, an
international standards organization determining the "standard" for many types of
technology. JIS is the Japanese Industrial Standard, specifying the standards used for
industrial activities in Japan.
490 Series Microscope
AmScope Proprietary
Page 55 of 64
Diopters
Small rings on the eyetubes, used to focus ocular lenses
(eyepieces).
Eye Guards
Eye Guards fit over the eyepieces, they are for comfort and to
protect the glass in the eyepiece.
Eyepieces
Eyepieces are also called ocular lenses. Eyepieces come in
many magnifications and you replace them by swapping them
out of the eyetubes.
The eyepieces magnify the specimen further from the first
magnification through the objective lens. Each eyepiece is
marked with 2 numbers. The first number is the magnification and
the second is the field of view.
For example, an eyepiece marked WF10/20 has a magnification
of 10X and a field of view of 20mm.
Eyetubes
Eyetubes are also called ocular tubes. Eyetubes house the
eyepieces and camera adapters. You change eyepeices by
removing them from the eyetubes and replacing them with the new
ones.
490 Series Microscope
AmScope Proprietary
Page 56 of 64
Field of View and Working Distance
Field of view is how much of the specimen you can see
through the eyepiece. The linear field of view of the eyepiece
is divided by the magnification of the objective. The higher the
magnification, the smaller the field of view.
Working distance is the distance between the
bottom of objective lens and the stage. You
change the working distance when you use the
coarse focusing knob.
Together, working distance and field of view
determine how much of the specimen you see
and how closely.
Filters
Optical filters selectively transmit light of different wavelengths.
Blue filters compensate for low light levels in a brightfield application. (Like
dry layer blood analysis.) With low light, you see a yellowing effect. The blue
filter brings the light back to a more natural white light.
Green filters optimize the performance of achromatic and plan achromatic
objectives. When using phase contrast the best images are often produced
in green light.
Yellow filters compensate for blue images. The yellow filter brings the light
back to a more natural white light.
490 Series Microscope
AmScope Proprietary
Page 57 of 64
Focusing Knobs
Coarse Focus: This is the large knob on the side of the
microscope that moves the objective lens up and down.
It is used in conjunction with the fine focus. Do not use
this knob with the 100X objective lens.
Fine Focus: The outer knob, used to fine-tune the focus
of a specimen in conjunction with the coarse focus.
Head Lock Screw
The head lock screw secures the microscope head to
the microscope. Loosen the head to rotate between
forward and reverse mounting. Tighten the screw to
lock the microscope head position into place.
Head Orientation
Microscope head
orientation is for user
preference only; it
makes no difference in
imaging.
Forward
Reverse
490 Series Microscope
AmScope Proprietary
Page 58 of 64
Head Types
Monocular
Monocular microscope heads have one eyepiece.
Binocular
Binocular microscope heads have two eyepieces.
Trinocular
Trinocular microscope heads have two eyepieces and a photo port
for a camera. This allows you to look at a specimen with a camera
without removing an eyepiece.
Multi-head
Multi head microscopes are specially designed for three-person
observation. Ideal for teaching and/or training purposes. It has a
binocular head and two monocular heads.
Simul-Focal
Simul-Focal heads mitigate the amount of re-focusing needed when
you change your view from the eyepieces to the monitor.
Perfect for lectures, teaching demonstrations, clinical examinations
and laboratory applications.
490 Series Microscope
AmScope Proprietary
Page 59 of 64
Immersion Oil
A special oil used with the 100X objective to concentrate the light and increase
the resolution of the image. At high magnifications the field of view is so small
that very little light passes. The light refracts, scatters and produces a dark
image.
A drop of oil is placed on the slide’s cover slip and the objective is lowered until
it touches the oil. The oil will match the specimen’s light refracting index to the
lens, adding more light to the specimen, which improves resolution and clarity.
Interpupillary Distance
Interpupillary Distance is the distance between your eyes. AmScope uses 2 types of
systems to change interpupillary distance.
Siedentopf heads adjust the interpupillary distance by rotating the
eyetubes around a central axis. A Siedentopf head changes the
interpupillary distance without changing the tube distance.
Jentsch heads adjust the interpupillary distance by moving the
eyetubes closer together or farther apart in a linear fashion. The
tube length changes when changing interpupillary distance, so the
user compensates by adjusting the focus of the eyepieces to
correspond to the new interpupillary distance.
Iris or Diaphragm
The diaphragm or iris is located under the stage and is an
apparatus that can be adjusted to vary the intensity, and
size, of the cone of light that is projected through the slide.
The iris adjustment slider allows you to adjust the opening of
the iris.
Lenses
Microscopes have a two lens system, objective and ocular.
The Objective Lens is the lens closest to the specimen or object.
The Ocular Lens (Eyepiece) is the lens closest to the eye.
490 Series Microscope
AmScope Proprietary
Page 60 of 64
Light Bulbs-Caution
1. Turn off the microscope when not in use to preserve the life of the bulb.
2. Never touch a light bulb directly, especially when it’s turned on.
a. The light bulbs can get hot enough to burn skin. Wait until the bulb has cooled
completely before handling it.
b. The oil from your skin will create a hot spot on the bulb which will damage the
glass, dramatically shortening the life of the bulb.
c. When changing light bulbs cover skin with cloth or paper to handle the bulb.
This will protect both you and the bulb.
3. Hot bulbs will not damage the microscope. AmScope’s microscopes are designed
specifically to handle the heat output of the model’s bulb.
Lighting-Types
Tungsten
Tungsten light bulbs are the most economical, providing a reliable
source of light. They burn hotter than the other bulbs, but are the best
option for certain microscopes.
Fluorescent
Fluorescent bulbs burn cooler and brighter than tungsten bulbs. The
bulbs are more expensive than tungsten, but
last longer.
LED
LED light bulbs burn the coolest of all the bulbs and last the longest.
The bulbs are also the most expensive. Depending on use, LED
lights should last about 40,000 hours.
Halogen
Halogen light bulbs produce the most amount of light. The light is
very white and concentrated; the bulbs can get very hot and must be
cooled regularly. Depending on use, halogen light bulbs can last
anywhere from 1 week to 6 months.
Fiber Optic
Fiber optical lighting uses a system of flexible transparent fibers
made of plastic or glass. The light is transmitted between the two
ends of the fiber, allowing for greater illumination in confined spaces.
490 Series Microscope
AmScope Proprietary
Page 61 of 64
Limit Stop Knob
The limit stop is designed to prevent impact between objective
and slide. You may loosen the limit-stop knob in order to give
yourself the full range of motion for fine tuning the focus when
using the 40X and 100X objective lenses.
When the limit stop is off you will be able to damage the microscope or the glass
slide. For safety, engage the limit stop once you have it close or in contact with the
objective (if using oil contact is required).
Magnification
Microscopes have a two lens system. Total Magnification is the power of two lenses
multiplied together.
E.g.: (10X Eyepieces) x (4X Objective) = 40X Total Magnification
Mechanical Stage
A Mechanical Stage is a flat mechanism that sits on top of
the stage and allows the viewer to move a specimen small
distances - a task that is otherwise difficult at higher
magnifications. Most mechanical stages are equipped with
an X and Y axis so the viewer can see how far the slide has
moved.
Mechanical Stage Control
The mechanical stage control moves the stage,
mechanically moving the slide along an X and Y axis for
optimal positioning.
The top ring moves the stage forward and backward. The
bottom ring moves the stage from left to right.
Nosepiece
This circular structure is where the different objective lenses are
screwed in. To change the magnification power, rotate the turret.
490 Series Microscope
AmScope Proprietary
Page 62 of 64
Numerical Aperture (N.A.)
N.A. is a rating for the resolution of the objective lens. The N.A. rating ranks the
objective lens’ ability to capture light and show fine detail.
Lens with larger numerical apertures capture finer details than lenses with smaller
numerical apertures. Generally, lenses with larger numerical apertures create a brighter
image, but the depth of field will be shallower.
N.A. Rating
Objective
Lens
Average
Numerical
Aperture
4X
0.1
10X
0.25
40X
0.65
100X
1.25
Parcentered
When changing objectives, the image of the specimen stays centered. Most compound
microscopes are parcentered.
Parfocal
When changing objectives, the image of the specimen stays in focus without
needing to adjust the coarse focusing knob. Not all compound microscopes are
parfocal.
Phase Contrast
Many biological specimens are virtually transparent when viewed with a typical
compound microscope. Phase contrast microscopy is a method that collects light from
specific incidents (phases) of light only to allow better contrast in viewing transparent
specimens.
Phase Contrast improves visibility by manipulating direct and diffracted light to produce
greater contrast without losing resolution. The major benefit is the ability to study living cells
in their natural state instead of killing them with stain.
Brightfield
Phase Contrast
490 Series Microscope
AmScope Proprietary
Page 63 of 64
Phase Contrast Condenser Lenses
A specific condenser with a ring printed on it, used for phase contrast. The
magnification of the objective must match the magnification of the condenser.
Condensers are used to concentrate the light from a transmitted light on a sample for
illumination and viewing.
Phase Contrast Condenser Assembly
For phase contrast kits without a turret, the condenser assembly is the
mount for the condenser lenses.
Phase Contrast Objective
A specific objective with a ring printed on it that is used for phase
contrast.
The objective is the bottom lens that determines the resolution of
the image in the microscope, and is one of the two lenses that
compound to make total magnification.
Phase Contrast Turret
The phase contrast condensers are mounted on a rotating
turret much like the objectives on a microscope are. The
turret also has brightfield and darkfield condensers.
490 Series Microscope
AmScope Proprietary
Page 64 of 64
Telescope Lens-Centering
A lens that telescopes in one way or another, used in an eyepiece
slot to assist in centering the light in phase contrast microscopy.
Tension Knob
The large, innermost ring on the focusing knob. Adjusts the
tension of the focusing knobs.
Trinocular Port
A trinocular microscope has three viewing ports. There are
two eyepieces and a third port for photography and video.
The trinocular port is a 23mm adjustable photo port and
uses a C-Mount adapter.