Ins. No. B-149-4N-E

Meter Sizes 50, 52, 53, 55, 56, and 57 High Temp./Low Temp. Service and Jacketed SMART ULTRA OVAL FLOWMETERS Every SMART ULTRA OVAL is fabricated and shipped from our factory under stringent quality control. In order to maintain its design performance throughout the life of the meter, this manual offers the operator the necessary installation, operation and maintenance information. Be well familiar with these instructions before you place the meter in service and keep this manual at the field location for ready reference. The standard, high temperature-service, low temperature-service, and jacketed type meters all have in common the same specifications described in this instruction manual except for some differences in the jacketed type on its piping procedure and operating precautions. Be careful to observe the instructions given with a reminder "See also page 15 for the jacketed type". ◆ About Meter Size Designation ◆ The size of OVAL positive-displacement flowmeters is basically identified by a two-digit code. For complete details, see Section 17 PRODUCT CODE EXPLANATION. Meter size 50 Cooling tube provided Meter size 57 Cooling tube provided Meter sizes 52, 53, 55, 56 Heating tube provided

1. BEFORE YOU BEGIN ........................................................................................................................4

1.1 Confirming the Tag .........................................................................................................................4 1.2 Transportation Considerations .......................................................................................................4 1.3 Storage Considerations .................................................................................................................4 1.4 Structural Considerations ..............................................................................................................5 2. OPERATING CONDITIONS ...............................................................................................................5 3. GENERAL DESCRIPTION .................................................................................................................6 4. PART NAMES .....................................................................................................................................6 5. LCD COUNTER DISPLAY ..................................................................................................................7 5.1 About "MODE" Switch ...................................................................................................................7 5.2 Display Functions ...........................................................................................................................7 5.3 Display Mode Selection .................................................................................................................7 5.4 Total Counter Reset ........................................................................................................................8 5.5 Considerations with Pulse Output Type .........................................................................................8 6. INSTALLATION ...................................................................................................................................9 6.1 Considerations at Installation .........................................................................................................9 6.2 Standard Installation, Horizontal Line ............................................................................................9 6.3 Standard Installation, Vertical Line ........................................................................................................................................... 9 6.4 Example of Faulty Piping ...............................................................................................................9 7. HOW TO CHANGE FLOW DIRECTIONS ........................................................................................10 8. WIRING INSTRUCTIONS ................................................................................................................11 8.1 Wiring Guidelines ........................................................................................................................11 8.2 Terminal Connections ...................................................................................................................11 8.3 Preamplifier-to-Receiving Instrument Hookup .............................................................................12 9. OPERATING INSTRUCTIONS .........................................................................................................13 9.1 Operation .....................................................................................................................................13 9.2 Operating Precautions .................................................................................................................13 9.3 Precautions at Operation Shutdown ............................................................................................14 9.4 About Register Life ......................................................................................................................14 9.5 Piping Instructions and Operating Precautions of Jacketed Meters ...........................................15 10. DISASSEMBLY AND INSPECTION ...............................................................................................16 10.1 Size 50 Meter Body Disassembly and Inspection ......................................................................16 10.2 Size 50 Meter Body Assembly Procedure .................................................................................17 10.3 Sizes 52, 53, 55, 56, 57 Meter Body Disassembly and Inspection ............................................18 10.4 Sizes 52, 53, 55, 56, 57 Meter Body Assembly Procedure ........................................................19 11. REGISTER SWITCH FUNCTIONS AND PARAMETER SETTING ................................................20 11.1 Switch Names and Functions ....................................................................................................20 11.2 Individual Test Pin Functions ......................................................................................................21 11.3 About Meter Factor ....................................................................................................................21

12. SENSOR REPLACEMENT PROCEDURE .....................................................................................22

13. TROUBLESHOOTING ....................................................................................................................23 14. EXPLODED VIEWS AND PARTS LIST ..........................................................................................24 14.1 Meter Size 50 Exploded View and Parts List .............................................................................24 14.2 Meter Sizes 52 and 53 Exploded View and Parts List ...............................................................25 14.3 Meter Sizes 55 and 56 Exploded View and Parts List ...............................................................26 14.4 Meter Size 57 Exploded View and Parts List .............................................................................27 15. GENERAL SPECIFICATIONS ........................................................................................................28 16. OUTLINE DIMENSIONS ................................................................................................................29 17. PRODUCT CODE EXPLANATION .................................................................................................30 CONVENTIONS Shown in this manual are the signal words NOTE, CAUTION and WARNING, as described in the examples below: NOTE: Notes are separated from the general text to bring the user's attention to important information. CAUTION: Caution statements signal the user about hazards or unsafe practices which could result in minor personal injury or product or property damage. WARNING: Warning statements signal the user about hazards or unsafe practices which could result in severe personal injury or death.

1.2 Transportation Considerations

(1) To safeguard against damage during transit, transport your Smart Ultra OVAL to the installation site in the same container used for transportation from the factory if circumstances permit. (2) Smart Ultra OVAL is adjusted and inspected as an assembly consisting of the meter body, sensor and register. It should therefore be handled as an integral assembly. (3) The register is accurately configured and adjusted. Do not attempt to remove the front cover to gain access to its internal assembly. 1.3 Storage Considerations If your Smart Ultra OVAL is stored for long periods of time upon receipt before installation, unexpected faulty conditions could result. If a long-term storage is anticipated, take the following precautions: (1) Keep your Smart Ultra OVAL in store in the same shipping container used for transportation from OVAL if possible. (2) Place of storage should conform to the following requirements: - Free from rain and water. - Free from vibration and impact shocks. - Temperature and relative humidity in the storage place are at or near room temperature and humidity (around 25°C and 65%). (3) Purge the Smart Ultra OVAL that has once been placed in service with clean air, N2 gas, etc. to prevent the metered fluid from adhering to the meter connections, piping inner walls, housing, etc. before storage. (Wash clean with suitable detergent if necessary.) (4) In case of storage for extended periods of time, good practice is to keep in store in the same containers used for shipment from the factory. 1.1 Confirming the Tag Every Smart Ultra OVAL is assembled and adjusted according to individual specifications. Product code number and ratings are stated on the register tag. Make sure that, by referring to the general specifications on pages 28 and the product code explanation on page 29, the ratings shown conform to your particular specification. 1. BEFORE YOU BEGIN Every Smart Ultra OVAL is thoroughly tested before it leaves the factory. When received, it should be thoroughly inspected for indication of rough handling during transit. Necessary handling precautions are described in this section; read the instructions carefully. As for other information, find the respective sections from "CONTENTS" on pages 2. For any inquiries, contact your nearest OVAL designated sales office. CAUTION: When you make inquiries, include the product name, model number, product number, ratings and other pertinent information. NAMEPLATE (RATINGS SHOWN)

1.4 Structural Considerations

(1) The register is of waterproof construction for outdoor service. (2) The register is type approved. Do not attempt to replace component parts with others or make any circuit modification. (3) The pressuretight packing furnished constitutes part of the flameproof construction. Do not attempt to use any pressuretight packing other than those furnished. The pressuretight packing union should be tightened securely upon completion of connections. (4) Of the four different pressuretight packings (φ9, φ10, φ11 and φ12) furnished as standard accessories, a φ11 packing is tentatively installed in place. Select from these and install the one that best fits the finished outside diameter of your particular cable. Table 1.1 Applicable Cable Outside Dia. Unit in mm Packing Code Packing I.D. Cable O.D. 9 9.0 8.5 to 9.0 10 10.0 9.1 to 10.0 11 11.0 10.1 to 11.0 12 12.0 11.1 to 12.0 CAUTION Moisture allowed into the register wil cause blurred display glass and functional trouble. Precautions on cable lead-in fitting (pressuretight packing), both explosionproof and non-explosionproof models: - Use a pressuretight packing that fits the finished cable O.D. - Tighten up properly to allow no clearance between the pressuretight packing and the cable. - If a pressuretight packing is not used, apply some waterproof treatment to preclude any chance of moisture getting into the equipment. 2. OPERATING CONDITIONS To maintain the stated high accuracy and long service life of SMART ULTRA OVAL flowmeter, make sure that flowrate, pressure, temperature and viscosity are within the ratings as stamped on the meter register nameplate. Do not fail to confirm these ratings before placing it in service. CAUTION 1. Although allowable ambient temperature is up to +60 °C for explosionproof units, it is desirable that the meter be used at room temperature and humidity. 2. In cases where the register is exposed to elevated temperatures due to exposure to direct rays of the sun or to radiant heat, ensure, by providing a sunshade or similar protection, that the meter is used within the operating temperature range. NOTE: Avoid such installation locations where 1. Not easy to access for servicing. 2. Large temperature variation and vibration. 3. Potential risk of submersion in water. 4. Corrosive gas atmosphere. 5. Not compatible with ex. construction. 6. Near sources of strong magnetic noises. Register Cable Lead-in (Pressure-resistant packing) NOTE: Pressuretight packings are not supplied with nonexplosionproof models. ULTRA OVAL MADE IN JAPAN MNPJ-237 OVAL Corporation MODEL TAG. No MAX PRESS. FLOW RANGE INT. CONT. ~ ~ FULL SCALE TEMP. PULSE UNIT SERIAL No. FLUID DATE SIZE METER FACTOR NOTES 1. When measuring other liquids, consult us. 2. Place the meter body (outer case) in a horizontal position. For details, see instruction manual.

3. GENERAL DESCRIPTION

Smart Ultra OVAL has been developed to meet the needs of precise flowrate measurement. The local direct- reading total counter is an all-electronic register built around a single-chip CPU. With the latest electronic technologies used throughout, this versatile register displays accumulated total flow, instantaneous flowrate (digital readout) and provides, by option, a pulse and analog output proportional to the rate of flow. In this meter, fluid flow is detected by sensing with an amorphous sensor the magnetic fields of permanent magnets embedded in the oval rotors. As a result, high reliability is achieved. Features (1) Absence of any mechanically sliding components except for oval rotors contributes to long service life. (2) Small wetted parts count and pocketless design makes this meter ideally suited for flow measurement of chemical liquids. (3) You can monitor accumulated total flow and instantaneous flowrate locally on the digital display. (4) When coupled with a remotely located receiving instrument, output signals can readily and simply be used for applications including control, adjustment and recording. (5) IEC explosionproof construction offers increased safety and the design is compact. (6) A nonvolatile memory, both in the local totalizer type and externally powered type meters, retains accumulated total in the event of power failure or when de-energized. NOTE: With the use over an extended period of time, meter error will deviate more or less from the initial point. Upon request, we may conduct an instrumental error testing once again and establish a "new meter factor" in the scaler when your Smart Ultra OVAL is returned to the factory for periodic inspection or for other reasons. If you have any request, contact the supplier you purchased the product or our sales office. 4. PART NAMES MODE SWITCH LCD COUNTER FRONT COVER REGISTER FLOW DIRECTION REAR COVER SELECTOR MAGNET CONNECTING FLANGE METER BODY COOLING TUBE (OR HEATING TUBE) PRESSURETIGHT PACKING POWER SIGNAL Fig. 4.1

5. LCD COUNTER DISPLAY

5.1 About "MODE" Switch Removing the selector magnet inserted at the bottom of the register, apply it to the label "MODE" on the LCD counter face and the display will scroll forward through the available readings as shown in Fig. 5.1. CAUTION Do not fail to return the selector magnet to its holder after use lest you will not lose it. It uses an intensive magnet; never hold it close to floppy disks or other magnetic storage media. 5.2 Display Functions The display can show four different kinds of flow information - total flow, instantaneous actual flowrate, percent instantaneous flowrate, and 8-division percent bar graph. It also shows the following error messages: Full scale exceeded : ErrorFS Upper-limit flowrate exceeded : ErrorOF NOTE: Multiple errors will be indicated in priority order below: ErrorOF > ErrorFS 5.3 Display Mode Selection Two ways are available to select displays - with a display select switch inside the register, or through communications with the Smart Communication Unit Model EL2310 (or HHC: Model EK10). If your option is through communications, follow the instructions outlined in the Smart Communication Unit EL2310 (or HHC: EK10) instruction manual. NOTE: Show "Measure" window at "View" menu on the PC screen. Selection with display select switch requires access to the display select switch SW1 by opening the register cover facing its internal assembly and pressing this switch (see page 20). The display scrolls forward through available information each time you press this switch as shown in Fig. 5.1. SELECTOR MAGNET SELECTOR MAGNET "MODE" EXAMPLE OF READINGS MODE SYMBOL TOTALIZED FLOW INSTANTANEOUS ACTUAL FLOWRATE % INSTANTANEOUS FLOWRATE 8-DIVISION BAR GRAPH Fig. 5.1

5.4 Total Counter Reset

Displayed total flow can be reset either by the display select switch SW1 (see Fig. 11.1 on page 20) or through communications with the Smart Communication Unit. It your option is through communications, follow the instructions outlined in the Smart Communication Unit EL2310 (or HHC: EK10) instruction manual. NOTE: Show "Measure" window at "View" menu on the PC screen. With the display select switch, then, you can reset the totalizer by holding the display select switch SW1 depressed for more than 3 seconds while the total flow is shown. 5.5 Considerations with Pulse Output Type (1) If your model is of pulse output type, the pulse output and total counter remain inoperative for 15 seconds approx. after power on and while communications with the Smart Communication Unit continue. For 15 seconds approx. after termination of communications, the pulse output and total counter also remain inoperative. (2) Requirements for validating communications Communications are made valid only when the following requirements are met: - Flowrate at zero (There is no pulse output.) - Within 15 seconds after power turn-on NOTE: The period of 15 seconds immediately after power on is called "Communication standby mode." (The built-in indicator, if so equipped, will display as shown in ① in the figure below.) If communications are started during this time period, a switchover to "Communication mode" takes place, permitting you to communicate until power is turned off the next time. (The built-in indicator will display as shown in ② in the figure below.) To start flow measurement routine, turn power on again. (After power cycling, the pulse output and total counter will also remain inoperative for 15 seconds.) (3) While communications continue, the receiving instrument (total counter, etc.) may overcount under certain circumstances. To eliminate the possibility of such erratic extra counting, precautions should be taken by either disconnecting the receiving instrument before starting communications, or recording the most recent total reading and other important data on paper. (4) Except for the purpose of communications, do not attempt to connect the probe of Smart Communication Unit with the signal lines. If the probe remains connected, the influence of capacitive impedance the interface has could go to the point of producing distorted signals in waveform and, as a result, the receiving instrument would fail to receive pulse signals accurately. CAUTION: PRECAUTIONS with ANALOG OUTPUT TYPE The analog type permits communications with the Smart Communication Unit at any time. However, if, in an attempt to alter current parameters, the meter is configured by mistake such that the new parameters are incompatible with the specification, resultant sharp changes in output may disturb the behavior of the receiving instrument. It is good practice, therefore, to make parameter changes while the fluid flow is at zero. ② ① 15 seconds after power on. (Com Standby Mode) Communication is started within 15 seconds after power on. (Com Mode) Fig. 5.2 Display in Response to Power ON, Pulse Output Type

6.1 Considerations at Installation

(1) Flush the piping assembly. Flushing must be performed before meter installation. Install a short pipe in place of the meter at this time (see diagram below). BYPASS LINE STRAINER VALVE SHORT PIPE FLOW DIRECTION (1) In case flow direction is from R to L, switch the places of meter and strainer. (2) Arrange piping so as to facilitate drainage. (3) S t r a i n e r s h o u l d b e inspected on a regular basis. 6.3 Standard Installation, Vertical Line (1) If the flow direction is from top to bottom, switch the places of meter and strainer. (2) Installing the strainer at "A" will make net reinstallation difficult at cleaning; we recommend to install the strainer at "B". 6.4 Example of Faulty Piping NOTE: For outline dimensions and pipe connection dimensions, see approval drawing. OUT IN A B Marked * is the space required for disassembly and inspection. BYPASS STRAINER ULTRA OVAL OUT IN *(1100mm) WRONG! OUT IN (2) Install the meter free from pipe strains. (3) The meter must be installed on the discharge side of the pump. (4) If the meter is to be used under tank head, give a head pressure greater than the losses with the piping system, strainer, meter, etc. NOTE: Pressure losses of the Smart Ultra OVAL and strainers are stated in the GENERAL SPECIFICATIONS. (5) The meter installation is correct if the flow direction conforms to the arrow mark on the meter body. (6) The strainer should be located upstream of, and as close to the meter as possible. (7) Since the sensor in the Smart Ultra OVAL operates on the principle of sensing changes in magnetic flux density, it must be isolated from the influence of any external magnetic flux. In order to minimize the influence of external magnetic flux, select an installation location at least 5 meters away from existing power equipment and conductors - potential sources of creating strong magnetic and electric fields, such as motors and generators. (8) In case electric heating is desired, consult the factory. (9) In the heat retention work, aviod covering up the fins and heating tube with lagging material. - Typical Smart Ultra OVAL Installations 6.2 Standard Installation, Horizontal Line Do not install the meter in a position like this. "See also page 15 for the jacketed type".

7. HOW TO CHANGE FLOW DIRECTIONS

CAUTION: Do not fail to turn power OFF if the meter is of remote output type. To reverse the flow direction, change the orientation of both the register and flowmeter body to the new flow direction. (1) Take off four hex socket head screws (M6) with hex wrench. CAUTION 1. Rotate the register assembly without separating it from the skirt. 2. Do not attempt to rotate the register assembly more than one complete revolution. (2) Holding the register assembly in both hands, carefully rotate it and then secure it with hex socket head screws in line with the new direction of flow. Size 50 Sizes 52, 53, 55, 56, 57 REGETER ASSEMBLY CAREFULLY ROTATE. SKIRT Don’t separate! Do not attempt to change flow directions by separating the register assembly like this. NOTE: If you have separated the register assembly, first return it to the state before changing the flow direction, then follow the correct procedure in step (2) (without separating the register assembly) again to change flow directions.

8. WIRING INSTRUCTIONS

8.1 Wiring Guidelines (1) Cables for field wiring The following cables should be used unless otherwise specified: Use cables of conductor area 1.25mm2 and finished O.D. 12mm, selecting from 2 to 4pcs. according to your application. It is recommended that their shield be grounded at the receiving instrument. CAUTION: In case of TIIS explosionproof type used under the ambient temperature of 50°C or higher, use a cable resistant to the temperature of 70°C or higher. (2) Transmission length The maximum transmission length is typically one kilometer. NOTE: If it exceeds one kilometer, consult us. (3) Inductive interference prevention To minimize the possibility of stray current pickup, the field wiring should be routed sufficiently away from existing power cables or power circuits. (4) Considerations on connections 1) M3.5 terminal posts are used on the terminal block. Use crimp-style terminals that fit the conductors at the cable end. 2) Be sure to earth ground the preamplifier's ground terminal. 3) Pitch down the cable from the cable entry so that rainwater will not have a chance to enter the equipment through the cable. 4) In areas where lightning strokes are expected, provide a lightning arrestor for protection. CAUTION: Make sure of the validity of meter (register) and receiving instrument combination by referring to their model numbers, serial numbers, etc. before you make electrical connections. 8.2 Terminal Connections (1) Take off four hex socket head bolts (402) on the front of register assembly and remove the cover (401). (2) Remove the cover to gain access to a 4-post terminal block. Terminal identification is found on the back of cover. (3) Using crimped-lug terminals, ensure good electrical connections. HEX WRENCH COVER HEX SOCKET HEAD BOLTS O-RING REMOTE OUTPUT TERMINALS EXTERNAL GROUND TERMINAL INTERNAL GROUND TERMINAL NOTE: 1. Pressuretight packing is not furnished with non- explosionproof models and non-output models (see page 5). 2. Connect the external earth ground terminal to instrumentation earth ground before use. (Refer also to the wiring instructions of respective receiving instrument instruction manual.)

8.3 Preamplifier-to-Receiving Instrument Hookup

The 2-wire signal transmission system used in this flowmeter furnishes DC power to the meter. It serves as the power line and signal line as well with pulse or analog current output. An OVAL receiving instrument can be coupled directly, but instruments in general which are designed to accept a voltage signal input require a load resistor RL connected in series for voltage conversion. Since the voltage signal level varies with the load resistance value, determine the load resistance value by referring to the receiving instrument specifications and the acceptable load resistance range shown below. Communications with a PC (Smart Communication Unit) requires a 250 to 1000Ω load. ● In case a voltage input is fed to the receiving instrument Output Voltage Signal (V) Pulse Output ON/OFF = (20mA × RL) / (4mA × RL) Analog Output (4mA × RL) − (20mA × RL) at 0 ro FS NOTE: With the relationship with supply voltage E, select the load resistance value R L such that the current flowing into the transistor is held below 50mA. Fig. 8.1 Supply Voltage (VDC) Acceptable Load Resistance Range. RL(Load Resist.) 0 12V 17.25V 45V 1000Ω 1571Ω 250Ω 33V Area: Operating range Area: Acceptable Communication range R = L E−12V 21mA EXAMPLE: If supply voltage is 24V DC with load resistance 250Ω, then a 1 to 5V DC output is obtained. ( ( Fig. 8.2 L PC POWER R + − VOLT. SIGNAL TO RECEIVING INSTRUMENT + − OPEN COLLECTOR OUTPUT 30V, 50mA max. FLOWMETER REGISTER GND + − 2 1 4 3 + − (SMART COMMUNICATION UNIT) (OR HHC:EK10) PC (SMART COMMUNICATION UNIT) (OR HHC:EK10) PC RL

9. OPERATING INSTRUCTIONS

9.1 Operation At first-time operation, carefully operate in the following sequence, allowing the flow within the flow range specified. (Refer to the piping diagram below.) (1) Shut off the valve (A) on the inlet side and the valve (B) on the outlet side and then open the bypass line valve (C) to allow the fluid in the bypass line, thereby removing weld chips, scales and other foreign matter left in the piping assembly. CAUTION: This is particularly important on new installations. (2) Carefully and slightly at first, open the valve (A) upstream of the meter progressively and then, slightly at first, open the valve (B) downstream progressively. (3) Slowly close the bypass line valve (C) and make sure that the total counter in the register advances in response. Maintain a flowrate 10 to 20% of the maximum flowrate (confirm it in the instantaneous flowrate display mode) at this time, allow the flow for more than 15 minutes and make sure that air in the piping assembly has totally escaped. In applications where temperature exceeds 60°C, run the system at least for 30 minutes in this state to ensure uniform heat distribution in the measuring chamber. (4) Following the break-in operation (preheating), shut off the bypass line valve (C) completely and open the upstream valve (A) progressively until fully open and slowly open the downstream valve (B) until the rated flow is obtained. (5) Flowrate should be regulated with the valve (B) downstream of the meter and should be held within the rating. (6) The strainer net should be inspected for condition and cleaned on a regular basis. On a new installation, in particular, inspect daily first and, according to the clogged condition of the net being observed, inspection intervals may be reduced progressively to, say, once a week. A B C BYPASS LINE METER STRAINER VALVE FLOW DIRECTION 9.2 Operating Precautions (1) When changing flowrates In applications where the flowrate varies or where shutoff valve opening and closure takes place in batch operation, avoid rapid changes in flowrate across the meter. Operating the meter at flowrates in excess of the maximum allowable flowrate will nullify the guaranteed accuracy, reduce the meter life and may result in faulty conditions, such as the seizure of bearings or the rotor-to-measuring chamber contact. (2) Where the temperature of metered fluid changes Avoid rapid temperature changes in the meter. Temperature changes of the fluid in the meter should be held within 3°C per minute. Extra care should be used particularly when making a flow measurement in batch operation without the provision of heat tracing of the piping where the fluid temperature differs from atmospheric temperature. If rapid temperature changes are anticipated, heat trace the piping assembly as well as the meter. EXCESSIVE FLOW RAPID FLOW VARIATION RAPID TEMPERATURE CHANGES "See also page 15 for the jacketed type".

(3) Liquids of low steam pressure

Temperature and pressure of LPG, polyvinyl chloride monomers or anything with low viscosity and low steam pressure that are too ready to vaporize should strictly be controlled. During operation, the temperature of bearings in the meter is usually higher than that of the metered fluid. Vapors around the bearings can be causes of faulty conditions, including generation of unusual noise and bearing seizure. (4) Corrosive liquids When you make a measurement of highly corrosive liquids, such as nitric acid and sulfuric acid, appropriate materials should be used for tanks and piping assembly. Heterogeneous materials originally contained in the metered fluid or corrosive substances liquated out from tanks and pipes of inappropriate materials may lead to costly downtime, as a result of locked rotors, for example, when they are allowed into the measuring chamber. CAUTION: If it is a jacketed model, see page 15. 9.3 Precautions at Operation Shutdown (1) Valves should be closed progressively. Rapid valve closure could, under certain piping conditions, cause a sharp pressure rise by water hammer, or hydraulic shock, resulting in damage to the meter. (2) Precautions against pressure buildup on closure Complete closure of valves upstream and downstream of the meter makes the affected section a totally enclosed chamber and could cause a rise of internal pressure due to a rise of atmospheric temperature, resulting in an unexpected damage to the meter. (3) Liquids ready to adhere or gel at zero flow velocity Liquids that tend to adhere and solidify or gel at flow velocities around zero must thoroughly be washed away from the meter interior by running cleaning fluid before shutdown. Negligence of this instruction may leave the meter as an immovable unit when the operator attempts to resume meter operation the next time. CAUTION 9.4 About Register Life Due to the life expectancy of components incorporated, such as the LCD and nonvolatile memory, the internal electronics are designed to be replaced in about 10 years. The electronics life depends on the environment in which they are used; if any of the following operating conditions applies, we recommend you to replace them an assembly in five to six years: - High temperature environment. - High temperature process fluid measurement. - Field installation with widely varying temperatures. - LCD counter display is directly exposed to the rays of the sun (ultraviolet rays). - Frequent process fluid flow interruptions. TAKE PRECAUTIONS AGAINST BEARING SEIZURE!

9.5 Piping Instructions and Operating Precautions of Jacketed Meters

Hot water or steam jacketed meters require piping work in the following manner: (1) Piping to the meter body remains the same as that of standard meters (meters not jacketed). (2) Adhere to the following instructions for the piping to the jacket: Steam Jacketed Basically, the piping should be so arranged that steam flows in from the top and out from the bottom. Provide steam traps at outlet ports (see figure below). ⇩ Hot-water Jacketed Basically, the piping should be so arranged that hot water flows in from the top and out from the top (see figure below).⇩ (3) Although the measuring chamber is heat treated, due consideration should be taken not to give piping elongation or contraction by sharp temperature changes or pipe strains. Progressive heating is suggested at least four hours before commencing service operation. (4) As for outline dimensions and tubing connection dimensions, see approval drawings. (5) If heat tracing is desired, considerations should be taken to facilitate maintenance and servicing. Make sure, if it is the case, that there is no liquid leaks when related components are covered with lagging material. (6) Lagging should be so provided that the top cover of strainer is easily accessible. Also it is necessary that the strainer net be cleaned on a regular basis. (7) Never attempt to lag radiator fins. Lagging them will cause excessive heat buildup in the register and pulse generator, possibly leading to costly downtime. NOTE: As for outline dimensions and tube connection dimensions, see approval drawing. STEAM IN OUT FLOW DIRECTION Steam Jacketed Type, Horizontal Run IN OUT FLOW DIRECTION HOT WATER Hot Water Jacketed Type, Horizontal Run

10. DISASSEMBLY AND INSPECTION

Although it depends on individual operating conditions, periodic disassembly and inspection should be performed at least once a year. CAUTION: Because Smart Ultra OVAL is a precision industrial instrument, disassembly and inspection should be performed indoors as a rule. If it is desired to disassemble and inspect it as installed in the field, an important precaution to remember is to reduce the internal pressure of the piping assembly to a safe level, shut off valves upstream and downstream of the meter fully, drain the piping assembly and then place a suitable fluid receptacle directly below the Smart Ultra OVAL. Exercise care to keep individual components disassembled free from grit and dust. CAUTION: Be sure to deenergize before you disassemble. ◆ Meter Body Inspection ◆ Complaint and possible cause: Process fluid will not run. Jammed with scales, the oval rotors fail to rotate, resulting in an interrupted fluid flow. 10.1 Size 50 Meter Body Disassembly and Inspection METER BODY HEX SOCKET HEAD BOLTS FRONT COVER O-RING HEX SOCKET HEAD BOLTS (2) With the front cover (register) removed, the measuring chamber with oval rotors and other components are now accessible. (1) Take off all six front cover fitting bolts (104) (M6), using wrench. With the bolts removed, the register integral with the front cover is ready to be separated from the meter body. Be careful not to drop the register front cover assembly. ROTOR SHAFTS ROTORS MEASURING CHAMBER (3) Remove oval rotors (201) from the measuring chamber and inspect them for condition. (a) Inspect the oval rotors to see if they are jammed with foreign solids. (b) Check the rotors, shafts and other members for wear. (c) Inspect the measuring chamber and the inner wall of front cover for damage or any other sign of unusual condition. Upon completion of these inspections, clean the oval rotors, rotor shafts, measuring chamber and front cover with clean water or suitable solvent thoroughly. CAUTION 1. Score marks, scratches, high spots due to impressions, or other flaws should be reconditioned flat with oil stone or other tool. 2. If the areas which have been in contact with rear cover jack bolts are distorted outwardly, recondition it flat with oil stone.

10.2 Size 50 Meter Body Assembly Procedure

CAUTION: PRECAUTIONS BEFORE ASSEMBLY Oval rotors, inner walls of the rotor shafts, inner wall of the measuring chamber, inlet and outlet ports should be thoroughly washed clean, completely removing dust, grime and other foreign matter. MATCH MARKS 1ST ROTOR SIG. MAGNETS IN 1 2 OUT 2ND ROTOR FLOW DIR. ARROW AND 2ND ROTOR NO. FLOW DIRECTION FLOW DIR. ARROW AND 1ST ROTOR NO. (1) Rotor Installation Rotor installation is correct if the side where signal generating magnets are embedded faces the register (top in the measuring chamber) and the side with match marks ( • ) ( • • ) faces the front cover. Carefully install the rotors with the 1st rotor (match mark " • ") on the shaft with "1" stamped on the outside of measuring chamber and the 2nd rotor (match mark " • • ") on the shaft with "2" stamped. Ensure that the match marks are as shown in the inset. 回転子の回転方向 IN OUT Hot Water Jacketed Type, Horizontal Run (2) Confirming the Rotor Gear Engagement Hand rotate the rotors to make sure of correct gear mask. FRONT COVER O-RING HEX BOLTS (REGISTER SIDE) (3) Front Cover Installation Firstly install O-ring (108) on the front cover. If the O-ring is damaged or swollen with metered liquid, it will not fit in the groove. If such is the case, replace with a new one. Align the locating pin of front cover with its pin slot in the meter body and, fitting the rotor shafts into rotor shaft sockets in the front cover, force the meter body until the locating pin fits firmly in the pin slot in the meter body without slanting the front cover. Install six front cover fitting bolts (M6) and tighten them evenly until the front cover is closely in contact with the meter body. (4) Confirming Freely Rotation With air or water, make sure to see that the rotors move freely and that the register counts in response. CAUTION Rotation check should be performed at low rotor r.p.m. Violent rotor spinning may cause a damage to the components, such as bearing seizure. CAUTION: Since the registers are not compatible each other, do not replace one register with that of other meter.

10.3 Sizes 52, 53, 55, 56, 57 Meter Body Disassembly and Inspection

CAUTION: Be sure to deenergize before you disassemble. REAR COVER (1) Take off all rear cover fitting bolts, using wrench. REAR COVER HEX BOLTS TOP COVER (2) Remove the rear cover to gain access to the top cover. (3) Using cross slot screwdriver, remove all four bolts securing the top cover in place. Then, use two of them to jack up the top cover, eventually to remove it. CAUTION When jacking the top cover off, turn the jacking screws evenly and alternately so that the top cover not slant. REAR COVER GASKET END PLATE (4) With the top cover removed, the measuring chamber with oval rotors and other components are now accessible. ROTOR SHAFT REAR COVER TOP COVER GASKET ROTORS (5) Remove oval rotors and rotor shafts, in this order, from the measuring chamber and inspect them for condition. (The rotor shafts should be drawn out straight.) (a) Make sure to see if the oval rotors are jammed with foreign matter. (b) Check to see if the rotors, shafts and other components are worn. (c) Inspect if the measuring chamber and inner wall of the front cover have any sign of scratches or other damage. Upon completion of these inspections, clean the oval rotors, rotor shafts, measuring chamber and front cover with cleans water or suitable solvent thoroughly. CAUTION 1. Score marks, scratches, high spots due to impressions, or other flaws should be reconditioned flat with oil stone or other tool. 2. If the areas which have been in contact with rear cover forcing bolts are distorted outwardly, recondition it flat with oil stone.

10.4 Sizes 52, 53, 55, 56, 57 Meter Body Assembly Procedure

CAUTION: PRECAUTIONS BEFORE ASSEMBLY Oval rotors, inner walls of the rotor shafts, inner wall of the measuring chamber, inlet and outlet ports should be thoroughly washed clean, completely removing dust, grime and other foreign matter. NON TURN PIN ROTOR SHAFTS SHAFT SOCKET ROTORS 2ND ROTOR "2" MARKS 1ST ROTOR "1" MARKS FLOW IN OUT MATCH MARKS (2) Rotor Installation Rotor installation is correct if the side where signal generating magnets are embedded faces the register (bottom of the measuring chamber) and the sides with match marks ( • ) ( • • ) faces the rear cover. Carefully install the rotors with the 1st rotor (match mark " • ") on the shaft with "1" stamped on the outside of measuring chamber and the 2nd rotor (match mark " • •" ) on the shaft with "2" stamped. Ensure that the match marks are in alignment as shown in the inset. ROTORS TOP COVER (4) Installing top cover and confirming smooth rotor rotation With the locating pin of top cover in line with its pin slot in the meter body and fitting the rotor shafts into their sockets in the top cover, force them against the meter body until the locating pin fits in its slot in the meter body without slanting the top cover. Install top cover fitting bolts and tighten each of them evenly until the top cover is firmly in contact with the meter body. At this point, allowing air or water, make sure to see that the rotors move freely and that the register counts in response. CAUTION Rotation check should be performed at low rotor r.p.m. Violent rotor spinning may cause a damage to meter components, such as bearing seizure. REAR COVER GASKET (5) Rear Cover Installation Firstly install the gasket on the rear cover. If the gasket is found damaged, replace it with a new one. Install rear cover fitting bolts and tighten them evenly. NOTE: We suggest to replace the gasket with a new one at each disassembly. CAUTION: Since the registers are not compatible each other, do not replace one register with that of other meter. (1) Rotor Shaft Installation The non-turn pin end of each rotor shaft faces the register (bottom of the measuring chamber). Install the shafts, pin end first, into their sockets on the bottom of measuring chamber (bottom of meter body), forcing downward while turning each of them with hands until they are firmly installed in position. (Air in the sockets may resist to be compressed by the shafts. So good practice is to hold the shafts pressed down by hand for some time.) (3) Confirming the Rotor Gear Engagement Hand rotate the rotors to make sure of correct gear mesh. At this point, they may be shaky more or less as the shafts not fixed firmly. Exercise care not to damage the rotors and inner wall of the measuring chamber by forcing them to rotate.

11. REGISTER SWITCH FUNCTIONS AND PARAMETER SETTING

11.1 Switch Names and Functions (1) Using a hex key, take off four hex socket head bolts securing the front cover. O-RING FRONT COVER HEX SOCKET HEAD BOLTS (2) Removing the front cover provides access to the electronics unit. TP-PIN OUT 2 PIN OUT 1 PIN REV PIN FWD PIN MODE SELECT SWITCH SW1 0V PICKUP POINT EXTERNAL GND TERMINAL EXT. OUTPUT TERMINAL SELECTOR MAGNET HOLDER INTERNAL GND TERMINAL PARAMETER REWRITE PROTECT SWITCH (“ON” TO INHIBIT REWRITE) IN PIN PLS PIN Fig. 11.1 Part Names of Electronics Unit

11.2 Individual Test Pin Functions

NOTE: Connect the 0V end to the L.H. side fixed screw of the electronics unit. PIN NAME FUNCTION WAVEFORM FWD An output of No. 1 amorphous sensor waveform appears. 10μs 200mV 200mV REV An output of No. 2 amorphous sensor waveform appears. While the flowmeter is making forward revolutions, pulses are produced slightly behind the FWD pulses. Same as above. PLS A re c t a n g u l a r w a v e f o r m a f t e r F W D p u l s e waveshaping appears. Timing remains the same as that of FWD and its waveform is one before unfactored output amplification. 1 to 1.5ms 3V IN Accepts a square-wave pulse train from the pulse checker (OVAL Model PC2201, for example). Used for analog full scale adjustment, loop check, or other servicing. Input mode is Model PC2201's PG30 mode. Also accepts pulses with levels "0": 1V max. and "1": 7 to 12V min., or open collector. OUT1 Provides a waveform corresponding to the power signal which appears across remote output signal terminals 1 and 2. (1) Unfactored pulse 2ms (2) Factored pulse Depends on "Pulse widt"setting. (3) Analog output 20ms T OUT2 Produces a waveform corresponding to the open collector output which appears across remote output terminals 3 and 4. 11.3 About Meter Factor If it is desired to change meter factors in an instrumental error testing, for example, you may establish a new meter factor by the following procedure. Instrumental error testing must be conducted with proper facilities and procedures specified in the Weights and Measures Law, Japanese Industrial Instruments Federation, JIS standards, or other established standards. - How to Determine a New Meter Factor New Meter Factor = (Current meter factor) × (1− E 100 ) (mL/P) where current meter factor : Stated in the test report or on the tag of the product. E : Instrumental error determined by the test (% ) CAUTION: The new meter factor should be put on paper for later reference. T ≒ 2 to 10ms at 0 to FS

12. SENSOR REPLACEMENT PROCEDURE

ELECTRONICS UNIT ELECTRONICS UNIT REGISTER ASSEMBLY SKIRT NOTE: Size 55 meter body is shown here. The same procedure applies to other size models. HEX SOCKET HEAD BOLTS REGISTER ASSEMBLY SENSOR RETAINER FITTING SCREW (1) Take off four hex socket head screws on the register. (2) Carefully draw the register assembly out. Exercising care not to bump the sensor against adjacent components, draw it out in the horizontal direction. (3) Take off four hex socket head bolts and separate the cover. (4) With screwdriver, take off two fitting screws holding the electronics unit (5) Holding the electronics unit in both hands, carefully draw it out. CAUTION: Use extra care not to damage the leads by forcibly pulling the sensor fitting disc. (6) Uncouple the connector from the sensor unit at back of the electronics unit. (7) Using C-shaped stop ring pliers, remove the C-shaped stop ring for the shaft. The sensor unit is now separable from the register housing. (8) Loosen the sensor retainer fitting screws (M4) with screwdriver, remove the sensor retainer and draw out the sensor unit. (9) Install a new sensor unit through the opening through which the old sensor unit was removed and assemble in the reverse order of disassembly. (10) At register installation in the meter body, match the flow direction of the meter body with the arrow on the sensor fitting disc. ELECTRONICS UNIT CONNECTOR CONNECTOR C-SHAPED STOP RING SENSOR FITTING DISC SENSOR UNIT 153 152 158 FLOW DIRECTION ARROW LOCATION OF SENSOR INSTALLATION OPENING SENSOR FITTING DISC SENSOR OPENING BLIND PLUG (153)

1. Totalizer remains

inoperative. 1. Flowrate is low. 1. Open valves progressively. 2. Insufficient pump pressure or head pressure. 2. T aking pressure loss of entire piping assembly into consideration, correct pump pressure or head pressure. 3. Power line voltage is out of specification or current carrying capacity of power source is inadequate. 3. Provide a 14 to 45VDC power to the register. (12 to 24VDC for open collector or voltage pulse output.) Current carrying capacity 30mA min. is required for power. (With analog output, 24VDC, 60mA min. is required.) 4. Oval rotors jammed with foreign matter; rotors locked; metered liquid fails to run. 4. Referring to "Disassembly and Inspection" (Sec. 10), disassemble the meter body and clean the rotors thoroughly. 5. Oval rotors installed the wrong way. 5. Refer to 10.2 Meter Assembly Procedure in the "Disassembly and Inspection", and reassemble correctly. 6. Sensor installed out of position. 6. Refer to "How to Change Flow Directions" (Sec. 7) and reinstall the sensor. 2. Unusual noise. 1. Air is entrapped. 1. Reduce flowrate and eliminate air completely in the piping assembly. 2. Vaporized metered liquid in the piping assembly. 2. Reduce flowrate and control metered fluid temperature and pressure to prevent vaporization. 3. Oval rotors revolving in contact with measuring chamber. 3. Refer to "Disassembly and Inspection" (Sec. 10), and disassemble and inspect for condition. 3. Liquid leaks. 1. Incomplete seal of the pipeline. 1. Retighten bolts at pipeline connections or replace gaskets with new ones. 2. Incomplete seal on front/rear cover of meter body 2. Inspect front/rear cover fitting bolts for tightness and replace O-rings with new ones. 4. Counts while valves remain closed. 1. Valve and pipeline leaks. 1. Inspect valves and pipeline. 2. Air pockets between valve and Ultra OVAL; rotors in rocking motion in response to pump's pulsating pressure. 2. Vent air. Provide a check valve and accumulator. 3. Supply power voltage fluctuates. 3. Eliminate voltage fluctuation. 5. Analog output unusual. 1. Load resistance too great. 1. Referring to "Load Resistance Range" on page 12, check load resistance to power supply voltage relationship and keep them within the specified range. 6. Accumulated total too high. 1. Rotors in rocking motion in response to a pulsating flow. 1. Add a check valve and accumulator. 2. Influenced by external magnetic fields (Meter sensor picks up external magnetic fields created by a motor, generator, etc.) 2. Keep out external magnetic fields. 3. Air entrapped. 3. Provide an air vent. 7. Accumulated 1. Influenced by external magnetic fields. 1. Keep out external magnetic fields.

14. EXPLODED VIEWS AND PARTS LIST

14.1 Meter Size 50 Exploded View and Parts List - When you order replacement parts, specify the stock No., flowmeter model, instruction manual No., symbol No., part name and the quantity desired. EXPLODED VIEW 404 500 158 160 153 152 154 156 155 151 204 201 202 204 203 101 104 109 103 106 400 300 113 111 112 114 115 120 107 121 121 123A 123B 110 108 FLOW PARTS LIST, Meter Size 50 Symbol No. Part Name Q'ty Remarks 101 Meter Body 1 103 Rear Cover 1 104 Front Cover Fitting Bolt 6 M8×35 106 Rear Cover Fitting Bolt 4 M6×15 ▲ 107 O-Ring 1 G60 ▲ 108 O-Ring 1 G65 ▲ 109 O-Ring 1 G40 110 Front Cover Locating Pin 2 111 Skirt 1 112 Skirt Locating Pin 2 113 Skirt Fitting Bolt 4 M8×20 114 Register Fitting Bolt 4 M6×15 115 Gasket 1 120 Sealing Flange 1 121 Iron Slug 1 151 Sensor Fitting Disc 1 152 Sensor Retainer 1 153 Blind Plug 1 Symbol No. Part Name Q'ty Remarks 154 O-Ring B 2 S10 155 O-Ring C 1 G65 156 C-shaped Stop Ring 1 158 Cross Recess Pan Head Screw 2 M4 X 8 (w/washer) 160 Sensor 1 ※ 201 1st Rotor 1 w/Signal magnet ※ 202 2nd Rotor 1 w/Signal magnet 203 Rotor Shaft 2 ※ 204 Rotor Bearing 4 300 Sensor Assembly 1 set (151 to 158, 160) 400 Register Assembly 1 set 401 Register Cover 1 402 Cover Fitting Bolt 4 M6 (w/washer) 404 Selector Magnet 1 405 Internal Assembly 1 set 500 Pressuretight Packing 1 set 123 A Heating Tube 1 set (w/screen, fr. cover) 123 B Cooling Tube 1 set (w/screen, fr. cover) ▲: Recommended spare parts. ※: Rotors 201, 202 and rotor bearings 204 are available as an assembly.

14.2 Meter Sizes 52 and 53 Exploded View and Parts List

- When you order replacement parts, specify the stock No., flowmeter model, instruction manual No., symbol No., part name and the quantity desired. EXPLODED VIEW 4 PARTS LIST, Meter Sizes 52 and 53 Sym. No. Part Name Size 52 Size 53 Q'ty Remarks Q'ty Remarks 101 Meter Body 1 1 102 Rear Cover 1 1 103 Skirt 1 1 ▲104 Gasket 1 1 107 Rear Cover Fitting Bolt 4 M12×30 4 M12×30 108 Skirt Fitting Bolt 8 M10×20 8 M10×20 109 Gasket 1 t=1.5 1 t=1.5 111 Meter Body Skirt 1 1 112 Fitting Bolt 4 M8×12 4 M8×12 113 Gasket 2 t=1.5 2 t=1.5 114 Register Fitting Hex Socket Head Screw 4 M6×15 4 M6×15 151 Sensor Fitting Disc 1 1 152 Sensor Retainer 1 1 153 Blind Plug 1 1 154 O-Ring B 2 S10 2 S10 155 O-Ring C 1 G65 1 G65 156 C-shaped Stop Ring 1 1 158 Cross Recess Pan Head Screw 2 M4×8 (w/washer) 2 M4×8 (w/washer) 160 Sensor 1 1 ※201 First and Second Rotor 1 each w/signal magnet 1 each w/signal magnet 203 Rotor Shaft 2 2 ※204 Rotor Bearing 4 4 211 Top Cover 1 1 213 Top Cover Fitting Bolt 4 M6×12 4 M6×12 300 Sensor Assembly 1 set (151 to 158, 160) 1 set (151 to 158, 160) 400 Register Assembly 1 set 1 set 401 Register Cover 1 1 402 Cover Fitting Bolt 4 M6 (w/washer) 4 M6 (w/washer) 404 Selector Magnet 1 1 405 Internal Assembly 1 set 1 set 500 Pressuretight Packing 1 set Oprion 1 set Oprion 700A Cooling Tube 1 set 1 set 700B Heating Tube 1 set 1 set ▲: Recommended spare parts. ※: Rotors 201 and rotor bearings 204 are available as an assembly.

14.3 Meter Sizes 55 and 56 Exploded View and Parts List

- When you order replacement parts, specify the stock No., flowmeter model, instruction manual No., symbol No., part name and the quantity desired. EXPLODED VIEW PARTS LIST, Meter Sizes 55 and 56 Sym. No. Part Name Size 55 Size 56 ▲: Recommended spare parts. ※: Rotors 201 and rotor bearings 204 are available as an assembly. Q'ty Remarks Q'ty Remarks 101 Meter Body 1 1 102 Rear Cover 1 1 103 Skirt 1 1 ▲104 Gasket 1 1 107 Rear Cover Fitting Bolt 6 M12×40 8 M12×45 108 Skirt Fitting Bolt 8 M12×20 8 M12×20 113 Gasket 2 t=1.5 2 t=1.5 114 Register Fitting Hex Socket Head Screw 4 M6×15 4 M6×15 151 Sensor Fitting Disc 1 1 152 Sensor Retainer 1 1 153 Blind Plug 1 1 154 O-Ring B 2 S10 2 S10 155 O-Ring C 1 G65 1 G65 156 C-shaped Stop Ring 1 1 158 Cross Recess Pan Head Screw 2 M4×8 (w/washer) 2 M4×8 (w/washer) 160 Sensor 1 1 ※201 First and Second Rotor 1 each w/signal magnet 1 each w/signal magnet 203 Rotor Shaft 2 2 ※204 Rotor Bearing 4 4 211 Top Cover 1 1 213 Top Cover Fitting Bolt 4 M6×12 4 M6×12 300 Sensor Assembly 1 set (151 to 158, 160) 1 set (151 to 158, 160) 400 Register Assembly 1 set 1 set 401 Register Cover 1 1 402 Cover Fitting Bolt 4 M6 (w/washer) 4 M6 (w/washer) 404 Selector Magnet 1 1 405 Internal Assembly 1 set 1 set 500 Pressuretight Packing 1 set Option 1 set Option 700A Cooling Tube 1 set 1 set 700B Heating Tube 1 set 1 set

14.4 Meter Size 57 Exploded View and Parts List

- When you order replacement parts, specify the stock No., flowmeter model, instruction manual No., symbol No., part name and the quantity desired. EXPLODED VIEW PARTS LIST, Meter Sizes 57 Symbol No. Part Name Q'ty Remarks 101 Meter Body 1 102 Rear Cover 1 103 Skirt 1 ▲104 Gasket 1 107 Rear Cover Fitting Bolt 16 M12×45 108 Skirt Fitting Bolt 8 M12 ×20 113 Gasket 2 t=1.5 114 Register Fitting Hex Soc. Hd. Scr. 4 M6×15 151 Sensor Fitting Disc 1 152 Sensor Retainer 1 153 Blind Plug 1 154 O-Ring B 2 S10 155 O-Ring C 1 G65 156 C-shaped Stop Ring 1 158 Cross Recess Pan Head Screw 2 M4×8 (w/washer) 160 Sensor 1 ※201 First and Second Rotor 1 each w/signal magnet 203 Rotor Shaft 2 ※204 Rotor Bearing 4 211 Top Cover 1 213 Top Cover Fitting Bolt 4 M8×12 300 Sensor Assembly 1 set (151 to 158, 160) 400 Register Assembly 1 set 401 Register Cover 1 402 Cover Fitting Bolt 4 M6 (w/washer) 404 Selector Magnet 1 405 Internal Assembly 1 set 500 Pressuretight Packing 1 set Option 700A Cooling Tube 1 set 700B Heating Tube 1 set ▲: Recommended spare parts. ※: Rotors 201 and rotor bearings 204 are available as an assembly.

15. GENERAL SPECIFICATIONS

● Basic Meter Meter Size 50 52 53 55 56, 57 Nominal Diameter 20mm 25mm 40mm 50mm Flange Ratings 1 JIS 10K RF, ASME 150 RF, JPI 150 RF 3 JIS 16, 20, 30K RF, ASME 300 RF, JPI 300 RF, DIN 10, 16, 20, 25 Operating Temperature Range* - 60 to +60° C (low temp. service), +120°C to 350°C (high temp. service) Accuracy ± 0.5% and ± 0.2% Materials Meter Body SCS14 or SCS16 Rear Cover SUS316 or SUS316L Rotors SUS316 or SUS316L Bearings Special carbon or ceramics Shafts SUS316 or SUS316L Flow Direction Right → Left (standard), Left → Right, Top → Bottom, Bottom → Top External Finish Meter body and rear cover: Not painted; Skirt: Painted *: Size 50 operates in a range -30 to + 60° C (low temp. service) or +120 to + 200°C (high temp. service). ● Max. Operating Pressure per Flange Rarting Unit: MPa Flange Code JIS 10K RF ASME 150 RF JIS 20K RF JIS 30K RF ASME 300 RF 1 1.18 at 120°C 1.50 at 120°C …… …… …… 3 …… …… 1.96 at 120°C 2.94 at 120°C 2.94 at 120°C ● Register General Specifications Meter Size 50* 52 53* 55 56 57 Local Indicator (LCD) Mode Select Accumulated total, 8-digit reading units ×10mL (standard) ×0.1L (standard) ×1 L (standard) Instantaneous flowrate (b mode) ×1 L/h (standard) ×10L/h (standard) % Instantaneous flowrate (P mode) 0 to 100% Bargraph mode 8-segment bargraph Output Analog 0 to FS 4 to 20mA NOTE: One of them is chosen. Current pulse (factored) 0/1 4/20mA Current pulse (unfactored) 0/1 4/20mA Open collector factored pulse Open collector Max. allowable voltage: 30V DC Allowable current: 50mA NOTE: One of them is chosen. Open collector unfactored pulse Open collector alarm output Open collector FWD/REV flow discrimination Output Pulse Width (ON duration) 1 ms (standard) 1 ms to 1000 ms (resolution 1ms) (Notes 1 and 2) Power Supply 12 to 45VDC (Acceptable load resistance range exists. See page 12.) Signal Transmission Cable Shielded cable 1.25 mm2 Transmission length one kilometer (cable D.D. 8.5 to 12 mm) Operating Temperature Range -10 to + 50°C (Register itself -10 to + 60°C) Construction Explosionproof TIIS: Exd Ⅱ BT4 Explosionproof KCs: Exd Ⅱ BT4 Non-explosionproof construction Protection rating of Enclosure IP66 (dusttight/watertight) IEC/EN60529, JIS C0920 Housing Material Aluminum die casting Finish Munsell 2.5PB 5/8, baked NOTE: 1. If pulse width in excess of 1ms is desired for minimum factored pulse unit, the max. flowrate can possibly be restricted; consult the factory. 2. With meter size marked*, if standard factored pulse unit is selected, pulse width other than 1 ms is unacceptable. ● Nominal Meter Factors Size P/r Nominal Meter Factor 50 4 9.936 mL/P 52 4 9.664 mL/P 53 4 17.513 mL/P Size P/r Nominal Meter Factor 55 6 34.60 mL/P 56 8 74.66 mL/P 57 8 196.08 mL/P

b1: ×1 1/h

b2: ×10 s1/sis MIT ×10 ml MODE RESET MODE MODE RESET φ130 φ130 L 137 8 B D F B D E F E CABLE ENTRY G1/2 (Output signal type) Heating tube Cooling tube (1) 10K TYPE DIMENSIONS Size Flange Rating L (face-to-face) B (overall) D E F Approx. Weight 50 JIS 10K RF 200 296.8 43 φ 96 211.8 w/cool tube 14kg w/heat tube 13kg ASME, JPI, 150 RF 198 Size Flange Rating L (face-to-face) B (overall) D E F Approx. Weight 52 JIS 10K RF 200 353.3 47 □106 264.3 w/cool tube 17kg w/heat tube 16kg ASME 150 RF 200 53 JIS 10K RF 200 376.8 59.5 □106 275.3 w/cool tube 18kg w/heat tube 17kg ASME 150 RF 200 Size Flange Rating L (face-to-face) B (overall) D E F Approx. Weight 55 JIS 10K RF 230 371.3 71 φ 163 258.3 w/cool tube 23kg w/heat tube 22kg ASME 150 RF 233 56 JIS 10K RF 250 391.3 76 φ 193 273.3 w/cool tube 27kg w/heat tube 26kg ASME 150 RF 258 Unit: mm b1: ×1 1/h b2: ×10 s1/sis MIT ×10 ml MODE RESET MODE MODE RESET B D φ130 L 8 137 B D E φ130 G CABLE ENTRY G1/2 Heating tube Cooling tube E F G F (Output signal type) (1) 10K TYPE DIMENSIONS Size Flange Rating L (face-to-face) B (overall) D E F G Approx. Weight 57 JIS 10K RF 350 424.3 87 φ 260 295.3 171.5 w/cool tube 43kg w/heat tube 42kg ASME 150 RF 357 NOTE: For outline dimensions of jacketed meters and pipeline connections dimentions, see the approval drawing. 16. OUTLINE DIMENSIONS

17. PRODUCT CODE EXPLANATION

●Main code (Meter model: 50, 52, 53, 55, 56, 57) ① ② ③ Model L U S ULTRA OVAL (integral type) Single case all stainless L U J ULTRA OVAL (single standard jacket) Element: stainless steel ④ ⑤ Capacity (Nominal Diameter) 5 0 20mm (3/4") ND (Big) 5 2 25mm (1") ND (Big) 5 3 25mm (1") ND (Big) 5 5 40mm (1·1/2") ND (Big) 5 6 50mm (2") ND (Big) 5 7 50mm (2") ND (Big) ⑥ Nominal diameter 4 Nominal diameter (Big) integral type ⑦ — ⑧ Fluid category L Liquid ⑨ Temp. category 0 60℃ and lower 1 Over 60℃ up to 120℃ 2 Over 120℃ up to 200℃ w/Cooling fin 3 Over 200℃ and up to 260℃ w/Cooling fin 5 Sudden temp. change (120℃ and lower) 6 Low temp. (−30℃ and over and below −10℃) Max. temp. 60℃ w/Heating tube ⑩ Major material C SCS14 (SUS316) S SCS16 (SUS316L) Z Special ⑪ ⑫ Process connection J 1 JIS10K RF J B JIS16K RF J 2 JIS20K RF J 3 JIS30K RF A 1 ASME150 RF A 3 ASME300 RF P 1 JPI150 RF P 3 JPI300 RF D 1 DIN10 RF D B DIN16 RF D 3 DIN25 RF Z 9 Special ⑬ — ⑭ Explosion-proof 0 Non-explosionproof 2 TIIS 8 KCs ⑮ Ex-proof temp. class 0 Non-explosionproof 4 T4 ①②③④⑤⑥-⑧ ⑨⑩⑪⑫-⑭⑮⑯ - ⑳㉑- ㉓㉔ ㉕㉖ ⑲ ⑱ Model Fluid category Capacity (Nominal Diameter) Nominal diameter Register Regulations Power supply Bearing Special Seal material Viscosity category Version Output Explosion-proof Ex-proof temp. class Temp. category Process connection Major material Main code /// Additional code ⑯ Regulations 0 Standard G High Pressure Gas Safety Act (Approved product) w/Material test certificate H High Pressure Gas Safety Act (Individual test) w/Material test certificate (Designed on PO issued) J High Pressure Gas Safety Act (Completion inspection) w/Material test certificate L Gas Business Act (Approved product) w/Material test certificate (Designed on PO issued) M Gas Business Act w/Material test certificate Q Electricity Business Act (Certificate required) w/Material test certificate (Designed on PO issued) R Electricity Business Act w/Material test certificate T Fire Service Act w/Material test certificate A Nuclear power w/Material test certificate (Designed on PO issued) F w/Material test certificate Z Special ⑰ — ⑱ Register H Smart Type ULTRA register ⑲ Power supply 0 External power supply (standard) ⑳ Output A Analog D Current pulse M Current pulse + open collector pulse W Open collector pulse + analog Z Special � Version A Version code: A � — � Bearing 0 Standard (carbon bearing) 2 Standard (strong alkali carbon) 8 Polymerizing liquid (carbon bearing) 5 Highly corrosive liquid (ceramic bearing) 6 Hydrogen peroxide solution (ceramic bearing) 7 Highly corrosive polymerizing liquid (ceramic bearing) � Viscosity category 2 Below 1000mPa·s R3 sintered rotor 3 1000mPa·s and over R3 special cut sintered rotor 5 Below 200mPa·s Cut rotor (sintered rotors inapplicable) 6 200mPa·s and over Special cut cut rotor (sintered rotors inapplicable) 7 Below 1000mPa·s R3 cut rotor (sintered rotors inapplicable) 8 1000mPa·s and over R3 special cut cut rotor (sintered rotors inapplicable) � Seal material F O-ring (FPM), gasket (T#1120) C O-ring (IIR), gasket (T#1120) T O-ring (PTFE), gasket (V7020) P O-ring (PTFE), gasket (V7035) Z Special � Special 0 Standard Z Special

Category of High Pressure Gas

H P 0 Other than High Pressure Gas H P 1 Toxic gas and flammable gas H P 2 Toxic gas H P 3 Flammable gas H P 4 Other than toxic or flammable gas Accuracy R 0 5 ±0.50% ACCURACY L 0 1 ±0.15% LINEARITY ※Only for export L 0 3 ±0.35% LINEARITY ※Only for export R 0 2 ±0.20% ACCURACY R 9 9 Special Operating condition F C 0 Continuous F M 0 Intermittent Special test (instrumental error) A 1 0 Taxed custody transfer A 2 0 By certified measurer A 6 0 Standard oil meter According to JMIF standard (Bore size 80mm and over) A 7 0 Std. fuel oil meter, std. water meter A 8 0 Std. fuel oil meter, std. water meter A 9 9 Designation of instrumental error test method Addition of one (1) test point, etc. Flow direction F R 0 R→L F L 0 L→R F U 0 T→B Electric conduit at the bottom F D 0 B→T Electric conduit at the bottom Designated special paint on body B C 0 Corrosion proof B A 0 Salinity and acid tolerance 120℃ and lower B X 0 Customer designation Designated special paint on transmitter S F 0 Corrosion proof Special treatment S D 0 Salinity tolerance S E 0 Acid tolerance Special treatment S X 0 Customer designated paint Special treatment Cleansing T W 0 Non-oil and non-water treatment T W 1 Non-oil and non-water treatment equivalent T F 0 Food cleansing T A 0 Nuclear cleansing T C 0 CIP Choose countermeasure to sudden temp. change Label N P J Label (Japanese) N P E Label (English) Document D S J DWG and specifications for approval (Japanese) D S E DWG and specifications for approval (English) D R 0 Re-submission of DWG with specifications D C J Final DWG (Japanese) D C E Final DWG (English) D P J Calculation sheet (Japanese) D P E Calculation sheet (English) S E J Instrumental error test report (Japanese) S E E Instrumental error test report (English) S T J Pressure test report (Japanese) S T E Pressure test report (English) S A J Airtight test report (Japanese) S A E Airtight test report (English) D D J Dimensional check record (Japanese) D D E Dimensional check record (English) S P J Penetrant test report (Japanese) Welded part of pressure resistant vessel S P E Penetrant test report (English) Welded part of pressure resistant vessel S M J Magnetic particle inspection (Japanese) Welded part of pressure resistant vessel S M E Magnetic particle inspection (English) Welded part of pressure resistant vessel S R J Radiographic inspection (Japanese) Welded part of pressure resistant vessel S R E Radiographic inspection (English) Welded part of pressure resistant vessel S U J Ultrasonic inspection (Japanese) Welded part of pressure resistant vessel S U E Ultrasonic inspection (English) Welded part of pressure resistant vessel S X J PMI test report (Japanese) S X E PMI test report (English) S S J Impact test report (Japanese) S S E Impact test report (English) D Y J WPS/PQR (Japanese) D Y E WPS/PQR (English) D 9 J Photo (Japanese) D 9 E Photo (English) D T J Inspection procedure (Japanese) D T E Inspection procedure (English) C A J Inspection certificate: A set Only Japanese C B J Inspection certificate: B set Only Japanese C C J Inspection certificate: C set Only Japanese C D J Inspection certificate: D set Only Japanese Witnessed by customer V 1 0 Required ①②③④⑤⑥-⑧ ⑨⑩⑪⑫-⑭⑮⑯ - ⑳㉑- ㉓㉔ ㉕㉖ ⑲ ⑱ Model Fluid category Capacity (Nominal Diameter) Nominal diameter Register Regulations Power supply Bearing Special Seal material Viscosity category Version Output Explosion-proof Ex-proof temp. class Temp. category Process connection Major material Main code /// Additional code

All specifications are subject to change without notice for improvement.

2018.04 Revised△ 1998.12 Released B-149-4N-E (1)