TELEVAC MC300 Instruction Manual

Specifications

Brand: TELEVAC
Category: Controller
Model: MC300 (Instruction Manual), pr24TNruEtLdg
File Type: PDF
File Size: —
Pages: 24 pages

Download Document

Get a high-quality, printable PDF version of this manual for offline use.

Download PDF Manual (24 pages)

⚠️ Download Safety: This PDF is completely safe to download (last scan date: 09/04/2026).

Read Online

Read the TELEVAC MC300 Instruction Manual online using the built-in PDF viewer.

MC300 Instruction Manual: Key Pages of Document

The Fredericks Company

MC300 Vacuum Sensor Controller Instruction Manual The Fredericks Company - Televac 2400 Philmont Avenue Huntingdon Valley, PA 19006 web: www.frederickscompany.com email: [email protected] tel: +1 215 947 2500 fax: +1 215 947 7464 ...

Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3 1.1 Disclaimer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 1.2 Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 1.3 Operating Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 1.4 Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 1.5 Safety Information . . . . . . . . . . . . . . . . . . . . ...

LIST OF TABLES

LIST OF FIGURES 4.2.2 Sensitivity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23 4.2.3 Filament Select . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23 4.2.4 Filament Degas . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23 List of Tables 1 Operating Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 2 MC300 Power Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 3 2A Analog Output . . . . . . . . . . . . . . . . . . . . . . ...

All information in this manual is subject to change without notice. The Fredericks Company assumes

no responsibility for inaccuracies in product specifications or any liability arising from product use. Please contact Televac at [email protected] or call 215-947-2500 with comments or questions. 1.2 Description The MC300 is a rack-mountable vacuum controller, which operates up to three vacuum sensors, including cold cathode, mini Bayard-Alpert, thermocouple convection, and thermocouple sensors. The capability of controlling a wide range of sensors allows the MC300 to provide a maximum vacuum measurement rang ...

Figure 1: MC300 Dimensional Drawing

1.5 Safety Information START BY READING THESE IMPORTANT SAFETY INSTRUCTIONS AND NOTES In these instructions the word “product” refers to the MC300 and all of its approved parts and accessories. NOTE: These instructions do not and cannot provide for every contingency that may arise in connection with the installation, operation, or maintenance of this product. Should you require further assistance, please contact Televac at the email address found in the footer of this manual. This product has been designed and tested to offer reasonably safe service provided in it is i ...

ˆ RETAIN instructions – Retain the safety and operating instructions for future reference.

ˆ HEED warnings – Adhere to all warnings on the product and in the operating instructions. ˆ FOLLOW instructions – Follow all operating and maintenance instructions. ˆ ACCESSORIES – Do not use accessories not recommended in this manual as they may require a technician to restore the product to its normal operation. Do not substitute parts or modify instrument. Because of the danger onj introducing additional hazards, do not install substitute parts or perform any unauthorized modifications to the product. Return the product to Televac for ser ...

Quick Start Guide

2 SETUP 2.1.5 Check Display The unit will first display the Software version number and instrument type in a four-digit code. The first two digits refer to the software version and the second two digits refer to the Instrument Sensor type summarized in Table 9. 2.1.6 Gas Correction The MC300 sensors are calibrated under dry nitrogen gas. Each MC300 is shipped in the default nitrogen configuration. The MC300 includes built in correction factors for the 2A Thermocouple Sensor and 4A Thermocouple Convection Sensors for use in argon gas. To toggle between the default nitrogen setting and correction factors for argon, complete the f ...

Move the lock switch into the upper position to lock out all changes from the front panel, including

set point assignments and values as well as the Mini BA sensitivity adjustment. 2.1.10 Power down MC300 and connect peripherals Once you’ve confirmed that the MC300 is functional, power it down. Connect all sensors using sensor cables, and connect any analog output, or set point cables as necessary. Be sure to follow all safety precautions associated with these items and your system. 2.1.11 Restart MC300 Restart your MC300. It should be ready for use. Contact the Fredericks Company with any comments or quest ...

Figure 2: Relay Wiring Diagram in the OFF State

2.4.1 Assigning Set Points and Values The instrument has four relays that are assignable to any of the vacuum measurement channels. The default setting is thermocouple (or convection) Channel 1 is Channel 1 (CH 1), thermocouple (or convection) Channel 2 is Channel 2 (CH 2) and the ion sensor is Channel 3 (CH 3), if applicable. Set point SP 1 is relay 1, set point SP 2 is relay 2, and etc. The default assignments for each set point to each channel are SP 1 to Channel 1 and SP 2 to Channel 2 for all versions. For configurations with an ion sensor, SP 3 and SP 4 will be assigned to Ch ...

Each sensor on the MC300 ships with two different analog output formats for ease of use by the

customer. Consult the appropriate section below to learn more about each sensor output. The MC300 will return to its normal operating mode after the fourth relay has been set up or 60 seconds have elapsed since the last button press. Changes are stored after the complete set up of an individual relay channel. These set up parameters are stored in non-volatile memory and are not lost during power outages. 2.5.1 2A Thermocouple Sensor The 2A Thermocouple sensor has two analog output formats. The Traditional Output o ...

Table 3: 2A Analog Output

Pressure (Torr) Traditional Output (V) Linear Output (V) -0.005 10.59 -0.05 -0.001 10.16 -0.01 0 10 0 0.001 9.84 0.01 0.005 9.41 0.05 0.01 8.92 0.1 0.02 8.03 0.2 0.03 7.23 0.3 0.04 6.63 0.4 0.05 6.13 0.5 0.06 5.69 0.6 0.07 5.31 0.7 0.08 4.95 0.8 0.09 4.63 0.9 0.1 4.37 1 0.15 3.37 1.5 0.2 2.89 2 0.3 2.23 3 0.4 1.88 4 0.5 1.67 5 0.6 1.58 6 0.7 1.5 7 0.8 1.42 8 0.9 1.34 9 1 1.26 10 2 1.02 10.2 5 0.92 10.2 10 0.9 10.2 20 0.88 10.2 CABLE 0 10.2 email: [email protected] tel: +1 215 947 2500 web: frederickscompany.com mc300_im rev H 10 of 23 ...

4A Thermocouple Convection Sensor

The 4A Thermocouple Convection sensor has two standard analog output formats and possesses a third optional analog output format. The Traditional Output option mimics the output of the raw signal from the 4A sensor. The voltage can be converted to a pressure by using a linear interpolation table. To receive an MC300 with a 4A sensor analog output set to Traditional, the user must specify this change during ordering. The Linear Output option provides a linear signal from .001 Torr to 1 Torr. The voltage can be converted to a pressure via an equation. The Linear Output is included in stan ...

Table 4: 4A Analog Output

Pressure (Torr) Traditional Output (V) Linear by Decade Output (V) Linear Output (V) 0 0 4.0 0 0.001 0.013 4.1 0.01 0.005 0.065 4.5 0.05 0.01 0.124 5.1 0.1 0.02 0.238 5.2 0.2 0.03 0.344 5.3 0.3 0.04 0.45 5.4 0.4 0.05 0.546 5.5 0.5 0.06 0.638 5.6 0.6 0.07 0.73 5.7 0.7 0.08 0.816 5.8 0.8 0.09 0.898 5.9 0.9 0.1 0.98 6.1 1 0.15 1.35 6.15 1.5 0.2 1.67 6.2 2 0.3 2.285 6.3 3 0.4 2.63 6.4 4 0.5 2.992 6.5 5 0.8 3.824 6.8 8 1 4.226 7.1 10 2 5.44 7.2 13.5 3 6.08 7.3 13.5 4 6.44 7.4 13.5 5 6.772 7.5 13.5 10 7.412 8.1 13.5 20 7.782 8.2 13.5 30 7.91 8.3 13.5 40 7.936 8.4 13.5 50 7.976 8.5 13.5 60 8.006 8.6 13.5 80 ...

7B Penning Magnetron Cold Cathode

The 7B Penning Magnetron Cold Cathode sensor has two analog output settings. The Linear Output option provides a linear signal from 1 × 10−7 Torr to 1 × 10−3 Torr. The voltage can be converted to a pressure via an equation. Pressre (Torr) = V (Vots) × 10−4 (4) The Linear by Decade Output option provides a signal that is linear within each decade (power of ten) and uses the first digit to convey the decade of pressure. The signal can be converted to a pressure by using a function to parse the voltage measurement. For V = 2 × (NT.DDD) Pressre (Torr) = 0.DDD × 10(� ...

Table 5: 7B Analog Output

Pressure (Torr) Linear by Decade Output (V) Linear Output (V) 1.10E-03 10.11 10.11 1.00E-03 10 10 9.00E-04 9.8 9 8.00E-04 9.6 8 7.00E-04 9.4 7 6.00E-04 9.2 6 5.00E-04 9 5 4.00E-04 8.8 4 3.00E-04 8.6 3 2.00E-04 8.4 2 1.00E-04 8.2 1 9.00E-05 7.8 0.9 8.00E-05 7.6 0.8 7.00E-05 7.4 0.7 6.00E-05 7.2 0.6 5.00E-05 7 0.5 4.00E-05 6.8 0.4 3.00E-05 6.6 0.3 2.00E-05 6.4 0.2 1.00E-05 6.2 0.1 9.00E-06 5.8 0.09 8.00E-06 5.6 0.08 7.00E-06 5.4 0.07 6.00E-06 5.2 0.06 5.00E-06 5 0.05 4.00E-06 4.8 0.04 3.00E-06 4.6 0.03 2.00E-06 4.4 0.02 1.00E-06 4.2 0.01 9.00E-07 3.8 0 8.00E-07 3.6 0 7.00E-07 3.4 0 6.00E-07 3.2 0 5.00E ...

7E/7F/7FC Double Inverted Magnetron Cold Cathode

The 7E/7F/7FC Double Inverted Magnetron cold cathode sensor has two analog output settings. The Logarithmic Output option provides a logarithmic signal from 1 × 10−11 Torr to 1 × 10−2 Torr. The voltage can be converted to a pressure via an equation. Pressre (Torr) = 10V−11 (6) The Linear by Decade Output option provides a signal that is linear within each decade (power of ten) and uses the first digit to convey the decade of pressure. The signal can be converted to a pressure by using a function to parse the voltage measurement. For V = NT.DDD Pressre (T ...

Table 6: 7E/7F/7FC Analog Output

Pressure (Torr) Logarithmic Output (V) Linear by Decade Output (V) 1.00E-11 0 0.1 2.51E-11 0.4 0.251 6.31E-11 0.8 0.631 1.00E-10 1.0 1.1 2.51E-10 1.4 1.251 6.31E-10 1.8 1.631 1.00E-09 2.0 2.1 2.51E-09 2.4 2.251 3.98E-09 2.6 2.398 6.31E-09 2.8 2.631 1.00E-08 3.0 3.1 2.51E-08 3.4 3.251 3.98E-08 3.6 3.398 6.31E-08 3.8 3.631 1.00E-07 4.0 4.1 2.51E-07 4.4 4.251 3.98E-07 4.6 4.398 6.31E-07 4.8 4.631 1.00E-06 5.0 5.1 2.51E-06 5.4 5.251 3.98E-06 5.6 5.398 6.31E-06 5.8 5.631 1.00E-05 6.0 6.1 2.51E-05 6.4 6.251 3.98E-05 6.6 6.398 6.31E-05 6.8 6.631 1.00E-04 7.0 7.1 2.51E-04 7.4 7.251 3.98E-04 7.6 7.398 ...

Mini Bayard-Alpert Sensor

The Mini Bayard-Alpert sensor has two analog output settings. The Logarithmic Output option provides a logarithmic signal from 1 × 10−10 Torr to 1 × 10−2 Torr. The voltage can be converted to a pressure via an equation. Pressre (Torr) = 10V−11 (8) The second analog output format is the Millivolt Logarithmic Output. The voltage can be converted to a pressure via an equation. Pressre (Torr) = 10mV−11 (9) Figure 7: Mini Bayard-Alpert Analog Output email: [email protected] tel: +1 215 947 2500 web: frederickscompany.com mc300_im rev H 17 of 23 ...

Table 7: Mini Bayard-Alpart Analog Output

Pressure (Torr) Logarithmic Output (V) “CABLE” 0 1.00E-10 0 2.00E-10 0.3 1.00E-09 1 2.00E-09 1.3 5.00E-09 1.7 1.00E-08 2 2.00E-08 1.3 5.00E-08 1.7 1.00E-07 3 2.00E-07 3.3 5.00E-07 3.7 1.00E-06 4 2.00E-06 4.3 5.00E-06 4.7 1.00E-05 5 2.00E-05 5.3 5.00E-05 5.7 1.00E-04 6 2.00E-04 6.3 5.00E-04 6.7 1.00E-03 7 2.00E-03 7.3 5.00E-03 7.7 1.00E-02 8 1.40E-02 8.15 "OVER CURRENT" 8.15 email: [email protected] tel: +1 215 947 2500 web: frederickscompany.com mc300_im rev H 18 of 23 ...

The MC300 is factory calibrated to ensure that each sensor meets its specified accuracy. As such,

calibration of new units is unnecessary. However, with extended use, sensors characteristics may change and the user may wish to calibrate sensors to return the unit to maximum accuracy. Users may also wish to calibrate replacement sensors. Please see the below sections for more information on calibrating each specific sensor. The correct calibration of each sensor includes pumping down to one order of magnitude of pressure below the desired calibration point, using a controlled, dry nitrogen leak to increase the p ...

Controller PDF Instructions

Zamp Solar Pulse Tech CINDER 40 PWM Operation & user’s manual
Zamp Solar Pulse Tech CINDER 40 PWM
29.03.2026 27 pages

CINDER™40 PWM Charge Controller CINDER 40 PWM Part # SCC1005 Input: DC12V or DC24V solar panel (Max 50Voc) Output: DC 12V 40A DC 24V 40A User Manual

LOVATO ELECTRIC RGK600 Installation manual
LOVATO ELECTRIC RGK600
18.05.2026 12 pages

Doc: I379GBD02_14.doc 10/02/2014 p. 1 / 12 RGK600 - RGK601 RGK600 - RGK601 RGK600SA - RGK601SA RGK600SA - RGK601SA Aggregatesteuerungen Generating set controller INSTALLATIONSHANDBUCH INSTALLATION MANUAL ● Vor jedem Eingriff am Instrument die Spannungszufu

RDS Sands Delta 3400 Calibration
RDS Sands Delta 3400
12.04.2026 30 pages

1 Sands Delta 3400 Sprayer Controller Calibration RDS Part No.: Document Issue: Software Issue: S/DC/500-10-607 1.00 : 8/9/08 PS403-000 rev.09

Grundfos IO 113 Installation and operating instructions manual
Grundfos IO 113
05.04.2026 36 pages

IO 113 Installation and operating instructions GRUNDFOS INSTRUCTIONS Other languages http://net.grundfos.com/qr/i/98097396