Chell microDAQ-Mk2 Installation And Operating Manual

Text of the manual

Chell Instruments Ltd

Folgate House Folgate Road North Walsham Norfolk NR28 0AJ ENGLAND Tel: 01692 500555 Fax: 01692 500088 microDAQ-Mk2 Pressure Scanner Acquisition System INSTALLATION AND OPERATING MANUAL e-mail:- [email protected] Visit the Chell website at: http://www.chell.co.uk 900193-1.3

Please read this manual carefully before using the instrument.

Use of this equipment in a manner not specified in this manual may impair the user’s protection. Chell Document No. : 900193 Issue 1.3 ECO : 2230 Date: 9th May 2016 Chell’s policy of continuously updating and improving products means that this manual may contain minor differences i

1 Description ................................................................................................................................. 3

1.1 General ....................................................................................................................................................................... 3 2 Specificatio

The microDAQ-Mk2 is a self contained acquisition system and combined pressure scanner that

acquires and transmits data to a host via Ethernet or a CAN bus. It is the newest version of the popular Chell microDAQ acquisition system, now known as microDAQ-Mk1. The microDAQ comes in several variants to accommodate both conventi

2.1 Power Supply:

Line voltage: 8-24 VDC Absolute Max. Line voltage 25VDC Consumption: With a 64 channel scanner: Max 4VA 2.2 CAN specifications: CAN type 2.0B CAN baudrate Configurable (by internal switch) from 1M, 500K, 125K and 100K. Programmable variables: Address 0x?nn Most significant programmable device ID Address 0xn?n Next m

Installation and Interconnections

3.1 Connector – Mating connector: DEUTSCH ASDD606-09SN-HE Pin Number Designation 1 TX+ 2 RX+ 3 TRIGGER IN (TTL) 4 +8~25V SUPPLY 5 0V 6 CAN_H 7 CAN_L 8 TX- 9 RX- Hot plugging the power to the microDAQ at the connector can cause permanent damage to the unit. Always switch the power at the power supply source. 3.3

Operation of the instrument

4.1 Connecting up the microDAQ. The microDAQ has one connector and cable which supplies the unit with power and also provides CAN and Ethernet comms. Ensure all the connections are made before powering up the microDAQ. The microDAQ should not be hot plugged with the power connector. Doing so can cause permanent damage to

4.4 Hardware Trigger

The microDAQ features a hardware trigger to enable the user to synchronise multiple microDAQ’s and to calculate the timing of the measurements made. The hardware trigger takes the form of a pulse train. Each time the microDAQ receives a positive edge, it will generate a set of measurements for all the channels configured in the

5 microDAQ Configuration Webserver

5.1 Introduction. The microDAQ web Configuration provides the means of setting up, calibrating and demonstrating the microDAQ unit from a standard PC with an ethernet port and browser. It replaces the old microDAQSetup program provided with the microDAQ-Mk1. The software is divided by tabs into seven areas of functi