microDAQ-Mk2
Pressure Scanner
Acquisition System
INSTALLATION
AND
OPERATING MANUAL
e-mail:- [email protected]
Visit the Chell website at:
http://www.chell.co.uk
900193-1.3
Please read this manual carefully before using the instrument.
Use of this equipment in a manner not specified in this
manual may impair the user’s protection.
Chell Document No. : 900193 Issue 1.3
ECO : 2230 Date: 9th May 2016
Chell’s policy of continuously updating and improving products means that this manual may contain
minor differences i
1 Description ................................................................................................................................. 3
1.1 General ....................................................................................................................................................................... 3
2
Specificatio
The microDAQ-Mk2 is a self contained acquisition system and combined pressure scanner that
acquires and transmits data to a host via Ethernet or a CAN bus. It is the newest version of the
popular Chell microDAQ acquisition system, now known as microDAQ-Mk1.
The microDAQ comes in several variants to accommodate both conventi
2.1 Power Supply:
Line voltage:
8-24 VDC
Absolute Max. Line voltage
25VDC
Consumption:
With a 64 channel scanner:
Max 4VA
2.2 CAN specifications:
CAN type
2.0B
CAN baudrate
Configurable (by internal switch) from 1M, 500K, 125K
and 100K.
Programmable variables:
Address 0x?nn
Most significant programmable device ID
Address 0xn?n
Next m
Installation and Interconnections
3.1 Connector – Mating connector: DEUTSCH ASDD606-09SN-HE
Pin Number
Designation
1
TX+
2
RX+
3
TRIGGER IN (TTL)
4
+8~25V SUPPLY
5
0V
6
CAN_H
7
CAN_L
8
TX-
9
RX-
Hot plugging the power to the microDAQ at the
connector can cause permanent damage to the unit.
Always switch the power at the power supply source.
3.3
Operation of the instrument
4.1 Connecting up the microDAQ.
The microDAQ has one connector and cable which supplies the unit with power and also provides
CAN and Ethernet comms. Ensure all the connections are made before powering up the
microDAQ. The microDAQ should not be hot plugged with the power connector. Doing so can
cause permanent damage to
4.4 Hardware Trigger
The microDAQ features a hardware trigger to enable the user to synchronise multiple microDAQ’s
and to calculate the timing of the measurements made. The hardware trigger takes the form of a
pulse train. Each time the microDAQ receives a positive edge, it will generate a set of
measurements for all the channels configured in the
5 microDAQ Configuration Webserver
5.1 Introduction.
The microDAQ web Configuration provides the means of setting up, calibrating and demonstrating
the microDAQ unit from a standard PC with an ethernet port and browser. It replaces the old
microDAQSetup program provided with the microDAQ-Mk1.
The software is divided by tabs into seven areas of functi