Arun Microelectronics LTD NGC2D Operation & User’s Manual

Text of the manual

© Arun Microelectronics Ltd 2015. All rights reserved

Page 2 1 Introduction The NGC2D process controller is a high-accuracy Bayard-Alpert (BA) ion gauge controller that offers integrated pressure measurement and process control. The controller operates from electrical supply voltages in the range 100 to 240 volts and frequencies of 48 or 65 Hertz. The contr

© Arun Microelectronics Ltd 2015. All rights reserved

Page 3 2.4 Electrometer Logarithmic conformance 2.5 Ion Gauge Suitable type UHV Bayard-Alpert gauges Range 1 x 10 -3 to below 3 x 10 -11 mbar. Dependant on gauge-head Grid +200 V d.c. (+500 V at 60mA in Degas) Filament Tungsten: ≤12 V at ≤4.2 A Coated iridium: ≤12 V at ≤2.6 A Filament

© Arun Microelectronics Ltd 2015. All rights reserved

Page 4 3 Warnings and safety WARNING – Safe operation of ion gauges requires grounding of all exposed conductors of the vacuum system, gauges and controller. LETHAL VOLTAGES may be established under some operating conditions unless correct grounding is provided. All exposed and isolated metal parts

© Arun Microelectronics Ltd 2015. All rights reserved

Page 5 Ventilation The instrument is forced-air ventilated through grilles on the sides and a vent in the rear panel. Mount it in a location where there is an adequate supply of air as close as possible to cool room-ambient temperature. The instrument is tolerant of, and is compensated for, operation at

© Arun Microelectronics Ltd 2015. All rights reserved

Page 6 Relays Four single pole, change-over relays are provided for process control and can be independently assigned to any gauge. Contacts are rated at 5A, 240V maximum. The contacts of the relays are shown diagrammatically on the rear panel legend, in the de-energised condition, adjacent to the termin