1
Introduction
The NGC2D process controller is a high-accuracy Bayard-Alpert (BA) ion gauge controller that offers integrated
pressure measurement and process control. The controller operates from electrical supply voltages in the
range 100 to 240 volts and frequencies of 48 or 65 Hertz.
The contr
© Arun Microelectronics Ltd 2015. All rights reserved
Page 3
2.4 Electrometer
Logarithmic conformance
2.5 Ion Gauge
Suitable type
UHV Bayard-Alpert gauges
Range
1 x 10 -3 to below 3 x 10 -11 mbar. Dependant on gauge-head
Grid
+200 V d.c. (+500 V at 60mA in Degas)
Filament
Tungsten: ≤12 V at ≤4.2 A
Coated iridium: ≤12 V at ≤2.6 A
Filament
© Arun Microelectronics Ltd 2015. All rights reserved
Page 4
3 Warnings and safety
WARNING – Safe operation of ion gauges requires grounding of all exposed
conductors of the vacuum system, gauges and controller. LETHAL VOLTAGES
may be established under some operating conditions unless correct grounding
is provided. All exposed and isolated metal parts
© Arun Microelectronics Ltd 2015. All rights reserved
Page 5
Ventilation
The instrument is forced-air ventilated through grilles on the sides and a vent in the rear panel. Mount it in a
location where there is an adequate supply of air as close as possible to cool room-ambient temperature.
The instrument is tolerant of, and is compensated for, operation at
© Arun Microelectronics Ltd 2015. All rights reserved
Page 6
Relays
Four single pole, change-over relays are provided for process control and can be independently assigned to
any gauge. Contacts are rated at 5A, 240V maximum. The contacts of the relays are shown diagrammatically
on the rear panel legend, in the de-energised condition, adjacent to the termin