BROOKFIELD KU-2
Viscometer
Operating Instructions
Manual No. M04-242-E1213
Brookfield KU-2 Table of Contents
I. INTRODUCTION......................................................................................4
I.1 Components...........................................................................................................4
I.2.1 Air Purge........................................................................................................5
I.2.2 Optional Paste Spindle................................................................
Brookfield KU-2 Brookfield Engineering Labs., Inc.
Page 4
Manual No. M04-242-E1213
I. INTRODUCTION
The Brookfield KU-2 Viscometer measures fluid viscosity in Krebs units. A paddle type spindle
is driven at 200 rpm by a constant speed motor. The reaction torque of the spindle rotating at
200 rpm is converted to viscosity in Krebs Units. The digital display of the Viscometer shows
viscosity in Krebs Units (KU), the associated grams value (gm), and viscosity in centipois
Brookfield Engineering Labs., Inc. Page 5
Manual No. M04-242-E1213
I.2 Options
I.2.1
Air Purge
The air purge option, Part No. KU-2A (Figure 2),
allows the interior of the KU-2 to be pressurized with
air (or an inert gas) when the instrument is to be used
in a hazardous environment. A built-in pressure switch
turns the instrument off when the pressure drops below
2.0 PSI (falling). The switch comes on at 10.0 PSI
(rising).
Connect a regulated clean air
Brookfield Engineering Labs., Inc. Page 6
Manual No. M04-242-E1213
I.3 Specifications
Range:
40–141 KU
32–1,099 gm
27 - 5,274 cP*
Accuracy:
± 1% of full scale range
Repeatability:
± 0.5% of full scale range
Paddle speed:
200 rpm ± 0.1 rpm
Printer output:
25D-Type Centronics Parallel Printer Output
Net weight:
22 lb. (10 kg)
Gross weight:
25 lb. (11 kg)
Dimensions:
15 x 11 x 20
Operating Environment: 0°C (32°F) to 40°C (104°F)
20% - 80% R.H.: no
Brookfield Engineering Labs., Inc. Page 7
Manual No. M04-242-E1213
Parallel Printer Output Connection
Pin
Function
1
STB
2
D0
3
D1
4
D2
5
D3
6
D4
7
D5
8
D6
9
D7
25
GND
Figure 4
I.6 Safety Symbols and Precautions
Safety Symbols
The following explains safety symbols which may be found in this operating manual.
Indicates hazardous voltages may be present.
Refer to the manual for specific warning or caution information to avoid personal injury
or damage to th
Brookfield Engineering Labs., Inc. Page 8
Manual No. M04-242-E1213
I.8 Viscosity Units
The Krebs Unit is a special measure of viscosity that is not based on the Newtonian model of
flow. The ASTM test method D562 was originally developed around the special conditions of an
instrument that used gravity to drive a paddle spindle at 200 rpm. The weight required to achieve
200 rpm varied depending upon the viscosity of the fluid under test. The Krebs unit w
Brookfield Engineering Labs., Inc. Page 9
Manual No. M04-242-E1213
II. OPERATION
II.1 Set-Up
1) Be sure that the power switch is off. Attach the power cord to the appropriate power source.
2) Move the operating handle to the top (upper most) position.
3) Connect the printer (if used) to the parallel port. Be sure the printer power is off.
4) Loosen the thumb screw on the Viscometer shaft; insert the paddle spindle into the Viscometer
shaft, as far as i
Brookfield Engineering Labs., Inc. Page 10
Manual No. M04-242-E1213
CAUTION:
When using the 1/2 pint can, do not lower the spindle directly into the
container. The narrow diameter of the can requires the spindle to be
introduced at an angle. Tilt the 1/2 pint can while lowering the viscometer.
8) The spindle will begin to rotate once the handle is within 1/2 inch of the lowest position.
9) Wait 5 seconds for the display reading to stabilize. A display
Brookfield Engineering Labs., Inc. Page 11
Manual No. M04-242-E1213
Appendix A - KU-2 Calibration Information
The accuracy of the KU-2 is verified using viscosity standard fluids calibrated in Krebs Units, which
are available from Brookfield Engineering Laboratories or your local Brookfield dealer/distributor/
representative. Note: Calibration should not be verified using the centipoise scale. The centipoise
scale is for reference use only and is based
Brookfield Measuring Instruments Owners Instructions:
| Device Model | Type of Document | Pages | Size | Views | Downloads | Updated |
|---|---|---|---|---|---|---|
RST |
RST (Measuring Instruments ePDF User Guide, #9UQJYR) |
69 | 900 | 207 | 20-06-2026 | |
CAP 2000+L |
Manual: Brookfield CAP 2000+L (T17BZY, Upd.01-09-2026) |
36 | 1331 | 240 | 01-09-2026 | |
RVDV-II+Pro |
Measuring Instruments #34N82N |
88 | 81 | 19 | 30-06-2026 | |
LV |
LV Operating instructions manual - X3644B |
36 | 133 | 23 | 29-09-2026 | |
KU-2 |
Measuring Instruments #8YA97G |
18 | 738 | 163 | 13-05-2026 |