USER AND MAINTENANCE GUIDE FOR Samsoft Mini PATIENT LIFT
2
Samsoft Mini patient lift
CONTENTS
I - WARNINGS...........................................................................................2
II - SYMBOLS AND EXPLANATION......................................................2
III - TECHNICAL SPEFIFICATIONS.....................................................3
3.1 Samsoft Mini........................................................................
USER AND MAINTENANCE GUIDE FOR Samsoft Mini PATIENT LIFT 3
� Check the red emergency lowering ring is
in the up position.
Check the elastic ring is correctly
engaged in the groove of the axle.
� Check the spring pin is correctly oriented on
either side of the lug and correctly closed.
� Check the beta pin in the lower part of the mast
is correctly positioned.
Check the wing knob is correctly screwed in.
.
LABEL OF IDENTIFICATION
The label of
USER AND MAINTENANCE GUIDE FOR Samsoft Mini PATIENT LIFT 4
Important: the patient lift should be used with slings
(safety members) provided by Drive DeVilbiss Healthcare
France exclusively
Driven operating force:
- with one fi nger: < 5N
- with one hand: < 105 N
- with one foot: < 300 N
2. Lifting assembly
Autonomous box : (control box + battery pack)
Dimensions
L=272 l= 120 h=90 mm
Weight
4,8 Kg
Protection
Current limiter
Output voltage
24 V continuou
USER AND MAINTENANCE GUIDE FOR Samsoft Mini PATIENT LIFT 5
If the red indicator is lit up (under the small key), the device
requires special maintenance, please contact Drive DeVilbiss
Healthcare France customer service.
CAUTION: before using the device, make sure all the
shafts, screws, and pins handled during the installation
are perfectly engaged in their respective supports, and they
are correctly locked.
Samsoft Mini is opr
USER AND MAINTENANCE GUIDE FOR Samsoft Mini PATIENT LIFT 6
Photo #°4 - Safety pin
Photo #°5 - Mast safety shaft
Photo #°6 - Position of the safety shaft when the mast
is upright
Photo #°7 - Unfolding the boom
Photo #°8 - Pin of the jack
5.4) Folding
The
device
should
be
folded
to
be
transported.
(Photo #°1).
The folding allows to store vertically the device and to transport
him in complete safety. To fold up the device, make the operations
described
USER AND MAINTENANCE GUIDE FOR Samsoft Mini PATIENT LIFT 7
6.1) Commissioning
When all the connections are made:
Use the patient lift with the remote control.
Check the battery charge (see colour code in Chapt. III. 3 - B-5).
The patient lift is ready to be used.
6.2) Charging
THE POWER UNIT SHOULD BE CHARGED EVERY NIGHT.
IT SHOULD BE CHARGED IN A WELL-VENTILATED
ROOM. CHECK THE POWER SUPPLY CORD CONDITION
BEFORE PLUGGING IT.
To charge the
USER AND MAINTENANCE GUIDE FOR Samsoft Mini PATIENT LIFT 8
Particular points:
When positioning the strap, check there is no fold at the leg
level in order not to hurt the patient.
Do not engage the device brakes when lifting so that the
device can correctly position with regard to the patient.
CAUTION! WHEN LIFTING OR LOWERING A PATIENT, NE-
VER ENGAGE THE WHEEL BRAKES SO THAT THE DEVICE
CAN BE CORRECTLY POSITIONED, IN LINE WITH THE PATIENT
USER AND MAINTENANCE GUIDE FOR Samsoft Mini PATIENT LIFT 9
undone, the upholstery and sling weft is not visible, and there
is no torn part.
Never use a defective sling.
The worn or malfunctioning parts should be replaced
with original parts manufactured by Drive DeVilbiss
Healthcare France.
Have some tries with a load before commissioning the patient
lift again.
9.5) Preventive maintenance
Every year or when necessary, and to be performed by a
comp
USER AND MAINTENANCE GUIDE FOR Samsoft Mini PATIENT LIFT 10
XI - WARRANTY CONDITIONS AND
RESPONSIBILITIES
Samsoft Mini s warranted for 5 years as far as the structure,
the cylinder and the control box are concerned, and in case of
normal use according to the instructions given in this guide.
Negligence, handling errors and accidents are excluded from the
warranty. The remote control is guaranteed for 3 months, and the
battery pack is guaranteed for 6 month
USER AND MAINTENANCE GUIDE FOR Samsoft Mini PATIENT LIFT 11
XII - BEFORE CONTACTING DRIVE DEVILBISS HEALTHCARE FRANCE CUSTOMER
SERVICE
Calling the Customer Service can often be avoided!
If the patient lift does not perfectly works, please perform the following checks fi rst:
PROBLEM
POSSIBLE CAUSE
REMEDY
Nothing happens when one of the
remote control keys is pressed.
▷ The remote control cord is not
correctly plugged in its slot.
▷ The remote control
Drive DeVilbiss Lifting Systems Owners Instructions: