USER AND MAINTENANCE GUIDE FOR Samsoft Mini PATIENT LIFT

2 Samsoft Mini patient lift CONTENTS I - WARNINGS...........................................................................................2 II - SYMBOLS AND EXPLANATION......................................................2 III - TECHNICAL SPEFIFICATIONS.....................................................3 3.1 Samsoft Mini........................................................................
USER AND MAINTENANCE GUIDE FOR Samsoft Mini PATIENT LIFT 3 � Check the red emergency lowering ring is in the up position.  Check the elastic ring is correctly engaged in the groove of the axle. � Check the spring pin is correctly oriented on either side of the lug and correctly closed. � Check the beta pin in the lower part of the mast is correctly positioned.  Check the wing knob is correctly screwed in. . LABEL OF IDENTIFICATION The label of
USER AND MAINTENANCE GUIDE FOR Samsoft Mini PATIENT LIFT 4 Important: the patient lift should be used with slings (safety members) provided by Drive DeVilbiss Healthcare France exclusively Driven operating force: - with one fi nger: < 5N - with one hand: < 105 N - with one foot: < 300 N 2. Lifting assembly Autonomous box : (control box + battery pack) Dimensions L=272 l= 120 h=90 mm Weight 4,8 Kg Protection Current limiter Output voltage 24 V continuou
USER AND MAINTENANCE GUIDE FOR Samsoft Mini PATIENT LIFT 5  If the red indicator is lit up (under the small key), the device requires special maintenance, please contact Drive DeVilbiss Healthcare France customer service. CAUTION: before using the device, make sure all the shafts, screws, and pins handled during the installation are perfectly engaged in their respective supports, and they are correctly locked. Samsoft Mini is opr
USER AND MAINTENANCE GUIDE FOR Samsoft Mini PATIENT LIFT 6 Photo #°4 - Safety pin Photo #°5 - Mast safety shaft Photo #°6 - Position of the safety shaft when the mast is upright Photo #°7 - Unfolding the boom Photo #°8 - Pin of the jack 5.4) Folding The device should be folded to be transported. (Photo #°1). The folding allows to store vertically the device and to transport him in complete safety. To fold up the device, make the operations described
USER AND MAINTENANCE GUIDE FOR Samsoft Mini PATIENT LIFT 7 6.1) Commissioning When all the connections are made:  Use the patient lift with the remote control.  Check the battery charge (see colour code in Chapt. III. 3 - B-5).  The patient lift is ready to be used. 6.2) Charging THE POWER UNIT SHOULD BE CHARGED EVERY NIGHT. IT SHOULD BE CHARGED IN A WELL-VENTILATED ROOM. CHECK THE POWER SUPPLY CORD CONDITION BEFORE PLUGGING IT.  To charge the
USER AND MAINTENANCE GUIDE FOR Samsoft Mini PATIENT LIFT 8 Particular points:  When positioning the strap, check there is no fold at the leg level in order not to hurt the patient.  Do not engage the device brakes when lifting so that the device can correctly position with regard to the patient. CAUTION! WHEN LIFTING OR LOWERING A PATIENT, NE- VER ENGAGE THE WHEEL BRAKES SO THAT THE DEVICE CAN BE CORRECTLY POSITIONED, IN LINE WITH THE PATIENT
USER AND MAINTENANCE GUIDE FOR Samsoft Mini PATIENT LIFT 9 undone, the upholstery and sling weft is not visible, and there is no torn part. Never use a defective sling. The worn or malfunctioning parts should be replaced with original parts manufactured by Drive DeVilbiss Healthcare France. Have some tries with a load before commissioning the patient lift again. 9.5) Preventive maintenance Every year or when necessary, and to be performed by a comp
USER AND MAINTENANCE GUIDE FOR Samsoft Mini PATIENT LIFT 10 XI - WARRANTY CONDITIONS AND RESPONSIBILITIES Samsoft Mini s warranted for 5 years as far as the structure, the cylinder and the control box are concerned, and in case of normal use according to the instructions given in this guide. Negligence, handling errors and accidents are excluded from the warranty. The remote control is guaranteed for 3 months, and the battery pack is guaranteed for 6 month
USER AND MAINTENANCE GUIDE FOR Samsoft Mini PATIENT LIFT 11 XII - BEFORE CONTACTING DRIVE DEVILBISS HEALTHCARE FRANCE CUSTOMER SERVICE Calling the Customer Service can often be avoided! If the patient lift does not perfectly works, please perform the following checks fi rst: PROBLEM POSSIBLE CAUSE REMEDY Nothing happens when one of the remote control keys is pressed. ▷ The remote control cord is not correctly plugged in its slot. ▷ The remote control