MTP MTP-3094 Instruction Manual
Text of the manual
Table of Contents
Introduction
Page 1
International Safety Symbols
Page 1
Safety Notes
Page 1
Warnings
Page 1
Caution
Page 2
Meter Description
Page 3
Meter’s Specifications
Page 4
Operation
Page 6
AC/DC Current Measurement
Page 6
AC/DC Voltage Measurement
Page 6
Resistance Measurement
Page 7
Diode and Continuity Test
Page 7
Capacitance Measurement
Page 8
Congratulations on your purchase of our AC/DC Clamp Meter, Model MTP-3094. Careful use
of this meter will provide years of reliable service.
International Safety Symbols
This symbol, adjacent to another symbol or terminal, indicates that the user must refer to
the manual for further information.
This symbol, adjacent to a terminal, indicates
Improper use of this meter can cause damage, shock, injury or death. Read and
understand this user’s manual before operating the meter.
Always remove the test leads before replacing the battery.
Inspect the condition of the test leads and the meter itself for any damage before operating
the meter. Repair or replace any damage before u
1. Current clamp
2. Non-contact AC voltage indicator light
3. Clamp trigger
4. Data Hold button
5. Rotary function switch
6. MODE select button
7. HZ% Hold button
8. Relative button
9. LCD display
10. COM input jack
11. V, Ω, CAP, Temp., Hz jack
12. Battery cover
Symbolic Display
1. AC DC
Alterning current (AC) and direct current (DC)
2.
Range & Resolution
Accuracy (% of reading)
AC Current
(50/60Hz)
40.00 A AC
(2.5% of reading + 8 digits)
400.0A AC
(2.8% of reading + 8 digits)
DC Current
40.00A DC
(2.5% of reading + 5 digits)
400.0A DC
(2.8% of reading + 5 digits)
DC Voltage
400.0 mV DC
(0.8% of reading + 2 digits)
4.000 V DC
(1.5%
Clamp size :
opening 1.2" (30mm) approx.
Diode test :
test current of 0.3mA typical ; open circuit voltage
1.5V DC typical
Continuity check:
treshold <150Ω ; test current <0.5mA
Low battery indication:
+ is displayed
Over range indication:
"OL" is displayed
Measurement rate:
2 readings per second, nominal
Input impedance :
10MΩ
Note: Read and understand all warning and precaution statements listed in the safety section of
this instruction manual prior to using the meter. Set the function select switch to the OFF
position when the meter is not in use.
AC/DC Current Measurement
WARNING : Ensure that the test leads are disconnected from the meter before making current
clamp
1. Insert the black test lead into the negative (COM) terminal and the red test lead into the
positive (V/Ω/Temp) terminal.
2. Set the function switch to the "Ω
CAP" position.
3. Touch the test probe tips across the circuit or component under test. It is best to
disconnect one side of the device under test so the rest of the circu