1 ClassicSeries™ Compressor Installation Applies to Models: CL21- 115 Volts CL22, CL32, CL52 - 208- 230 Volts 10540700 Rev. O (11/11) Model CL32 Shown Equipment Alert Compressor system must be installed per local plumbing and electrical codes. Components Shown are Typical on all ClassicSeries™ Compressor Models AA18110 ...
Supply CL21 CL22 CL32 CL52 P21 P22 P32 P52 P72 Voltage 115 208-230 208-230 208-230 115 208-230 208-230 208-230 208-230 Single Phase Disconnect Switch Box - Fused - User Supplied - Box(es) must be located within 3 ft. of compressor Fresh Air Intake Line Type Line Sizing Compressor Connections Drain Type Delivery Line Main Tru ...
3 AA223800 Compressor CFM @ 80 psig Main Header Line Size 5 - 10 3/8" ODT 10 - 20 1/2" ODT 20 - 30 5/8" ODT 30 - 40 3/4" ODT AIR LINE SIZING CHART AA189200 Specification Sheet Compressor Site Layout Elevation View Sample Plumbing Layout Main Trunk Line Equipment Room Branch Line “Typical” 3/8” to 3/4” 12’ 12’ 6’ A ...
AA2369 Keep dry Maximum stacking height (palletted units) Fuse rating specification Caution hot surface Ordinary Equipment IPXO Intended Use To provide compressed air during general examinations and procedures conducted by qualified dental professionals. Electromagnetic Interference This Midmark Lubricated Compressor is d ...
5 AA181600 Before you Install... Step 1: Remove compressor and accessories from shipping skid. Step 4: Install feet. Oil Fill Feet Motor Mount Nuts Oil Site Gauge Exhaust Silencer Out to Drain Step 6: Attach exhaust silencer into bottom of solenoid. Step 3: Loosen all motor mount nuts three full turns. Equipment Alert Compressor s ...
AA181701 Installation To Operatory Step 10: Install elbow to valve. Connect hose assembly to valve. Connect office air supply line. Note: Hose assembly included in the installation kit. Top View Shown Plumbing Connections Model CL21 & CL22 Connect Hose and Fitting Supplied with Installation Kit. Connect Hose Supplied with Inst ...
7 AA181800 Low Voltage Control Wires Installation Electrical Connections *Refer to Specification Sheet for Electrical Ratings in this manual. Step 12: Connect conduit cable to user supplied electrical box. Green/Yellow =Ground (White-115V) (Red -230V) and Black Connect to Electrical Supply* Control Panel Reference Cross Wi ...
AA181900 Installation Check & Test Step 14: Turn power on. Note: Compressor should run quietly and storage tank will begin to pressurize. Step 13: Verify shut-off valve is closed. Step 15: Use soapy water to check for pressure leaks. Shut-off Valve in Closed Position Soapy Water Gauge Note: Compressor will shut-off auto ...
9 AA182000 Step 17: Allow office air line to pressurize for 20-30 seconds. Check storage pressure gauge reading. Step 18: Check gauge reading after 10 minutes. Note: If the storage pressure gauge registers a lower reading, an air leak exists. Locate the leak(s) and repair. Step 16: Open the shut-off valve slowly and allow ai ...
Specification Sheet www.Midmark.com • 1-800-Midmark • www.Documark.com ClassicSeries™ Compressors Model CL32 Shown AA181500 Classifications: Class 1, No Applied Part, Continuous Operation Model # of Users CFM @ 80 PSI Total HP Tank Capacity Voltage Total Amps Breaker Size (Amps) Sound Level (dbA) Dimensions H x W x D ...
⚠️ Download Safety
The Compressor Installation manual (11 pages) is completely safe to download (last scan date: 31/07/2026).
Related PDF Instructions for Midmark Classic Series:
-
Midmark 398 Medical Equipment Installation and operation manual, 8 pages
-