Keysight Technologies 34134A Operation & User’s Manual
Text of the manual
Keysight 34134A User’s Guide
Notices
Copyright Notice
© Keysight Technologies 1999-2020
No part of this manual may be repro-
duced in any form or by any means
(including electronic storage and
retrieval or translation into a foreign
language) without prior agreement and
written consent from Keysight Technol-
ogies as governed by United States and
international copyright laws.
Manual Part Number
34134-90001
Edition
Edition 3, May 2020
Printed in:
Printed in Malaysia
Published by:
Keysight Technologies
Bayan Lepas Free Industrial Zone,
11900 Penang, Malaysia
U.S. Government Rights
The Software is “commercial computer
software,” as defined by Federal Acqui-
sition Regulation (“FAR”) 2.101. Pursu-
ant to FAR 12.212 and 27.405-3 and
Department of Defense FAR Supple-
ment (“DFARS”) 227.7202, the U.S.
government acquires commercial com-
puter software under the same terms
by which the software is customarily
provided to the public. Accordingly,
Keysight provides the Software to U.S.
government customers under its stan-
dard commercial license, which is
embodied in its End User License
Agreement (EULA), a copy of which can
be found at http://www.keysight.com/
find/sweula. The license set forth in the
EULA represents the exclusive authority
by which the U.S. government may use,
modify, distribute, or disclose the Soft-
ware. The EULA and the license set
forth therein, does not require or per-
mit, among other things, that Keysight:
(1) Furnish technical information
related to commercial computer soft-
ware or commercial computer software
documentation that is not customarily
provided to the public; or (2) Relinquish
to, or otherwise provide, the govern-
ment rights in excess of these rights
customarily provided to the public to
use, modify, reproduce, release, per-
form, display, or disclose commercial
computer software or commercial com-
puter software documentation. No
additional government requirements
beyond those set forth in the EULA
shall apply, except to the extent that
those terms, rights, or licenses are
explicitly required from all providers of
commercial computer software pursu-
ant to the FAR and the DFARS and are
set forth specifically in writing else-
where in the EULA. Keysight shall be
under no obligation to update, revise or
otherwise modify the Software. With
respect to any technical data as defined
by FAR 2.101, pursuant to FAR 12.211
and 27.404.2 and DFARS 227.7102, the
U.S. government acquires no greater
than Limited Rights as defined in FAR
27.401 or DFAR 227.7103-5 (c), as
applicable in any technical data.
Warranty
THE MATERIAL CONTAINED IN THIS
DOCUMENT IS PROVIDED “AS IS,”
AND IS SUBJECT TO BEING
CHANGED, WITHOUT NOTICE, IN
FUTURE EDITIONS. FURTHER, TO THE
MAXIMUM EXTENT PERMITTED BY
APPLICABLE LAW, KEYSIGHT DIS-
CLAIMS ALL WARRANTIES, EITHER
EXPRESS OR IMPLIED, WITH REGARD
TO THIS MANUAL AND ANY INFORMA-
TION CONTAINED HEREIN, INCLUD-
ING BUT NOT LIMITED TO THE
IMPLIED WARRANTIES OF MER-
CHANTABILITY AND FITNESS FOR A
PARTICULAR PURPOSE. KEYSIGHT
SHALL NOT BE LIABLE FOR ERRORS
OR FOR INCIDENTAL OR CONSE-
QUENTIAL DAMAGES IN CONNECTION
WITH THE FURNISHING, USE, OR
PERFORMANCE OF THIS DOCUMENT
OR OF ANY INFORMATION CON-
TAINED HEREIN. SHOULD KEYSIGHT
AND THE USER HAVE A SEPARATE
WRITTEN AGREEMENT WITH WAR-
RANTY TERMS COVERING THE MATE-
RIAL IN THIS DOCUMENT THAT
CONFLICT WITH THESE TERMS, THE
WARRANTY TERMS IN THE SEPARATE
AGREEMENT SHALL CONTROL.
Safety Information
CAUTION
A CAUTION notice denotes a hazard. It
calls attention to an operating proce-
dure, practice, or the like that, if not
correctly performed or adhered to,
could result in damage to the product
or loss of important data. Do not pro-
ceed beyond a CAUTION notice until
the indicated conditions are fully
understood and met.
WARNING
A WARNING notice denotes a hazard. It
calls attention to an operating proce-
dure, practice, or the like that, if not
correctly performed or adhered to,
could result in personal injury or death.
Do not proceed beyond a WARNING
notice until the indicated conditions are
fully understood and met.
Keysight 34134A User’s Guide
3
Safety Symbols
The following symbols on the instrument and in the documentation indicate
precautions which must be taken to maintain safe operation of the instrument.
Caution, risk of electric shock
Equipment protected throughout by
double insulation or reinforced
insulation
Protective earth (ground) terminal
Keysight 34134A User’s Guide
Safety Information
Read the information below before using this instrument.
The following general safety precautions must be observed during all phases of
operation, service, and repair of this instrument. Failure to comply with these
precautions or with specific warnings elsewhere in this manual violates safety
standards for design, manufacture, and intended use of the instrument. Keysight
Technologies assumes no liability for the customer’s failure to comply with these
requirements.
This apparatus has been designed and tested in accordance with IEC Publication
1010, Safety Requirements for Measuring Apparatus, and has been supplied in a
safe condition.
To ensure safe operation and to keep the product safe, the information, cautions,
and warnings in this operating manual must be heeded. In addition, note the
external markings on the instrument that are described under Safety Symbols.
WARNING
SHOCK HAZARD
– Connect the probe to the DMM or voltage measuring instrument before
clamping the probe around a conductor.
– Never use the probe on circuits rated higher than 660 V.
– Never leave the probe clamped around a conductor while not connected
to a voltage measuring instrument or DMM.
– Carefully center the conductor inside the probe jaws and make sure that
the probe is perpendicular to the conductor before closing the jaws.
– Avoid, if possible, the proximity of other conductors which may create
noise.
– Check the magnetic mating surfaces of the probe jaws; these should be
free of dirt, rust, or other foreign matter.
– Do not use a probe which is cracked, damaged or has defective leads.
Keysight 34134A User’s Guide
5
Table of Contents
Safety Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
1
Getting Started
34134A AC/DC DMM Current Probe . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
Compatibility . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
Inspect the Probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
Specifications and Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10
Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14
Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14
To check/replace the battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14
Cleaning the probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14
Keysight 34134A AC/DC DMM Current Probe
User’s Guide
7
1
Getting Started
34134A AC/DC DMM Current Probe 8
Inspect the Probe 9
Specifications and Characteristics 10
Operation 12
Repair 14
Keysight 34134A User’s Guide
34134A AC/DC DMM Current Probe
The 34134A Current Probe measures low DC or AC from 10 mA to 50 A DC, 40 A
AC. This battery-powered clamp-on probe may be used with digital multimeters,
voltmeters, or other voltage-measuring instruments. The probe jaws clamp round
the conductor under test, allowing current measurements without breaking the
circuit. Hall sensor technology senses the magnetic field produced by the current
measured, and generates a millivolt DC or AC output signal. The narrow jaw is
designed for optimum use in crowded wiring in industrial and automotive
environments.
Compatibility
The 34134A Current Probe is compatible with any DMM, or other voltage
measuring instrument which has the following features:
–
Range and resolution capable of displaying 1 mV of input.
–
Voltmeter accuracy (uncertainty) of 0.75% or better to take full advantage of
the accuracy of the probe.
–
Minimum input impedance of 100 kΩ (1 V/A range) or 10 kΩ (10 mV/A range).
–
Input jacks that accept 4mm shrouded banana plugs.
Figure 1-1
Control and connector identification
Keysight 34134A User’s Guide
9
Inspect the Probe
–
Inspect the shipping container for damage.
–
Keep a damaged shipping container or cushioning material until the
contents of the shipment have been checked for completeness and the
probe has been checked mechanically and electrically.
–
Inspect the instrument.
–
If there is mechanical damage or defect, or if the probe does not operate
properly or pass performance tests, notify your Keysight Sales Office.
–
If the shipping container is damaged, or the cushioning materials show
signs of stress, notify the carrier as well as your Keysight Sales Office. Keep
the shipping materials for the carrier’s inspection. The Keysight office will
arrange for repair or replacement at Keysight’s option without waiting for
claim settlement.
Keysight 34134A User’s Guide
Specifications and Characteristics
Operating and environmental characteristics of the 34134A Current Probes are
shown below.
Specifications
1 V/A Range (1 mV/mA)
10 mV/A Range
Current Range
DC: 10 mA to 2 A
AC: 10 mA to 1.5 A
DC: 100 mA to 80 A
AC: 100 mA to 60 A
Output Signal (AC/DC)
1 mV/mA (2 V @ 2 A)
10 mV/A
Accuracy[a]
[a]
Accuracies are given for an ambient temperature of 23 °C ±3 °C, relative humidity of 20 to 75%, conductor
centered in jaw window, probe zeroed, no hysteresis in DC, DC to 40 to 100Hz sine wave (1mV/mA range) or DC
and 40 to 1 kHz sine wave (10mV/A range), 1 minute warm-up, battery at 9V, appropriate load impedance,
magnetic field <40A/m and no common mode.
±2% reading 5 mA
500 mA to 50 A DC/40 A
AC: ±4% reading ±20 mA
50 to 80 A DC: ±12% reading
0 to 60 A AC: ±12% reading
Frequency Range
DC to 8 kHz @ -3 dB
DC to 8 kHz
Phase Shift
DC to 65 Hz: 1 °
DC to 65 Hz: 1 °
Minimum Input
Resistance
of Instrument
100 kΩ
100 kΩ
Noise
DC to 1 Hz: 1.5 mV
1 Hz to 10 kHz: 14 mV
0 kHz to 100 kHz: 18 mV
DC to 1 Hz: 15 μV
1 Hz to 10 kHz: 140 μV
0 kHz to 100 kHz: 180 μV
Slew Rate @ 5 A
120 mV/μs
5.5 mV/μs
Keysight 34134A User’s Guide
11
Typical Frequency Response Curve
Figure 1-2
Voltage vs frequency response curve
Environmental Characteristics
Operating temperature
0 to 50 °C
Storage temperature
-30 to 80 °C
Operating relative humidity
0 to 50 °C, 0 to 85% relative humidity
Temperature coefficient
800 ppm per °C +10 mA/°C typical from -10 to 50 °C
Altitude
Operating: 0 to 2000 m
Non-operating: 0 to 12,000 m
Mechanical Characteristics
Zero adjustment
20 turn potentiometer
Maximum cable diameter
11.8 mm
Dimensions
67 x 231 x 36 mm
Weight
330g with battery
Color
Grey
all-guides.co
m
Keysight 34134A User’s Guide
Operation
1
Connect the probe to the multimeter or other instrument.
Material
Fiberglass charged polycarbonate
Output cable
1.5m double insulated lead with shielded banana
plugs
Mechanical Characteristics
Working voltage
600 V CAT III
Case protection
IP20 per IEC 529
Drop test
1m according to IEC 68-2-32
Mechanical shock
100 G: test per IEC 68-2-27
Vibration
Test per IEC 68-2-6 frequency range 10 Hz to 55 Hz,
amplitude: 0.15mm
Environmental Characteristics
This symbol signifies that the 34134A AC/DC DMM Current Probe is protected by double or reinforced
insulation. Only use specified replacement parts when servicing the instrument.
This symbol signifies CAUTION! and requests that the user refer to the user manual before using the
instrument.
CAUTION
Be sure to observe the polarities: red = positive terminal, black = common
terminal.
Keysight 34134A User’s Guide
13
2
Select the multimeter range which best corresponds to the measured current.
For better reading stability, you may use the DMM 2 Volt range and let the
probe “warm up” for one minute before zeroing.
The Current Probe has a dual output, DC V output in DC and AC V output in
AC.
–
The “1 V/A” range has an output signal of 1 mV/mA AC/DC with an output
of 2 volts representing 2 amps present in the conductor being measured.
–
The “10 mV/A” range has an output signal of 10 mV/A AC/DC with an
output of 500 millivolts representing 50 amps present in the conductor
being measured.
3
Zero the probe in DC and in DC coupled AC measurements.
When the 34134A Current Probe has been used for a current measurement
and then removed from the conductor, a small amount of residual magnetism
often remains in the core. This residual magnetism will cause the voltmeter to
show a small DC reading even though there is no current passing through the
jaws of the probe. The residual magnetism should not cause a problem for AC
current measurement because the AC voltage function in most multimeters is
AC coupled. DC offset caused by the residual magnetism will contribute to
reading errors, but can be minimized by using the zero adjust knob to show a
reading of 0 a few counts on the multimeter (probe not clamped on a
conductor).
There will always be some instability and noise generated by the Hall sensor,
the earth magnetic field and environmental noise. This is particularly
noticeable on the most sensitive range, 1 V/A (1 mV/mA), where you may have
up to 5 mV of uncertainty, which cannot be “zeroed out.” “Zero” the probe
while it is connected to the DMM and on the range to be used. Let the probe
warm-up for one minute before zeroing. Turn the zero adjustment knob until
the probe is zeroed.
4
Observe the output polarities for DC measurement.
5
Clamp the probe on the conductor to be measured and read the current
flowing directly on your meter.
The output of the probe is 10 mV/A AC/DC or 1 mV/mA AC/DC. If your meter
indicates a negative reading during DC measurements, this simply means that
the current flow is in the opposite direction of the arrows marked “I” on the
probe or that the probe connections are reversed (polarity).
6
After measurement, turn the probe OFF.
Keysight 34134A User’s Guide
Calibration
To guarantee that your instrument complies with the factory specifications, we
recommend that the 34134A Current Probe be submitted to your local Keysight
Service Center at one year intervals for recalibration, or as required by other
standards.
Repair
This probe contains no user serviceable parts. Please return to your local Keysight
Service Center for repair.
To check/replace the battery
When the probe is turned on, the green LED should light up. If not, replace the 9 V
battery.
1
Disconnect the probe from the circuit and the DMM.
2
Turn the probe “OFF.”
3
Unscrew the battery compartment screw and pull out the battery
compartment cover.
4
Replace the battery and put the cover back up.
Cleaning the probe
Be sure that the mating surfaces of the jaw are free of dirt or foreign matter. If they
are rusted, gently clean with a soft, lightly oiled cloth.
CAUTION
Do not replace the battery while the probe is in use.
This information is subject to change
without notice. Always refer to the
English version at the Keysight
website for the latest revision.
© Keysight Technologies 1999-2020
Edition 3, May 2020
Printed in Malaysia
k
34134-90001
www.keysight.com