iOptron iEQ30 Pro Instruction For Upgrading

Text of the manual

Internal Use Only

Instruction for Upgrading an iEQ30 Mount to an iEQ30Pro Mount January 2017 This instruction only provides a technical reference. It is at your own risk to perform the upgrading yourself. iOptron assumes no liability on any equipment damage or personal injury. The iEQ30 Pro upgrade kit includes following parts: 

Internal Use Only

3. Unscrew these two screws to remove the worm assembly supporting block from the. Unscrew two aluminum standoffs. 4. Clean and re-grease the ring gear. 5. Grease the R.A. worm of a new iEQ30 Pro RA motor assembly. 6. Place the iEQ30 Pro R.A. module onto the mount. Insert two new gear positioning screws that come with the package and tighte

Internal Use Only

Lock the R.A. axis and gently pull the belt to turn the worm in both directions. You should be able to pull rotate the R.A. gear smoothly and evenly. If you feel it is difficult to pull the belt in one direction, the worm/gear meshing is not evenly. Slightly release four gear positioning screws and adjust the worm assembly. 9. Gently push d

Internal Use Only

16. Install the main board and secure it with a small screw the small screw. Connect the wire between main board and R.A. board. 17. Install the cover for R.A. module and main board by screwing the four screws. 18. Replace the DEC module and cover. 19. Connect R.A. and DEC units, plug in HC and power supply. 20. Upgrade hand controller firm