Vaisala MI Instruction Manual
Read Online
Read the Vaisala MI Instruction Manual online using the built-in PDF viewer.
Text of the manual
Document/Revision No. IM-EN-MI/2.00
Effective: September 2, 2019
PUBLISHED BY
Vaisala Oyj
Vanha Nurmijärventie 21, FI-01670 Vantaa, Finland
P.O. Box 26, FI-00421 Helsinki, Finland
+358 9 8949 1
Visit our Internet pages at www.vaisala.com.
General safety considerations
Always power off the Multichannel user interface MI before accessing the modules
behind the display. Vaisala recommends the installation of an external power switch.
Warning! If this device is used against manufacturer’s instructions, user safety may be
compromised.
Warranty
For standard warranty terms and conditions, see www.vaisala.com/warranty.
Please observe that any such warranty may not be valid in case of damage due to nor-
mal wear and tear, exceptional operating conditions, negligent handling or installation,
or unauthorized modifications. Please see the applicable supply contract or Conditions
of Sale for details of the warranty for each product.
Technical support
Contact Vaisala technical support at [email protected]. Provide at least the fol-
lowing supporting information:
•
Product name, model, and serial number
•
Name and location of the installation site
•
Name and contact information of a technical person who can provide further infor-
mation of the problem
For more information, see www.vaisala.com/support.
© Copyright Vaisala 2019. All rights reserved.
Disposal
When wishing to dispose of an obsolete instrument or any parts of an in-
strument, please observe local and national regulations and requirements
for the disposal of electrical and electronic equipment.
Symbols and terms used in this manual:
This indicates a warning. It provides safety precaution information needed to
avoid injury while operating the refractometer system.
This indicates that something is important for the operation of the refractome-
ter system.
Note. Notes contain additional information and hints.
This product manual is delivered to the end user with a Vaisala K-PATENTS® product.
Information in this manual is subject to change without notice. When the manual is
changed, a revised copy is published at http://www.kpatents.com/
Table of contents
1
Description
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
2
Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
3
Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
3.1
Connecting the power cable
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
3.1.1
Power terminals for AC power
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
3.1.2
Power terminals for DC power
. . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
3.1.3
External mains switch
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
3.2
Refractometer cable connections
. . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3.3
Ethernet cable connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3.4
Module cards
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3.5
Connections for external control of Recipe app
. . . . . . . . . . . . . . . . 13
4
Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
4.1
Main display
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
4.2
Activating and removing apps
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
4.3
Managing users and passwords . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
4.4
System log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
4.5
Ethernet connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
4.5.1
MI ethernet connection
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
4.5.2
Refractometer connection
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
5
Apps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
5.1
Shared functionality . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
5.1.1
Application trend
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
5.1.2
Application log
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
5.2
Refractometer app
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
5.2.1
Diagnostics display
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
5.2.2
Viewing and adjusting parameters
. . . . . . . . . . . . . . . . . . . . . . . . . 34
5.2.3
Refractometer verification
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
5.2.4
Collecting field sample points . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
5.3
Recipe app
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
5.3.1
Requirements
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
5.3.2
Activating Recipe app
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Creating a new recipe
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
41
5.3.4
Modifying a recipe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
41
5.3.5
Selecting recipe switching mode . . . . . . . . . . . . . . . . . . . . . . . . . .
42
5.3.5.1
Switching recipe manually . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
42
5.3.5.2
Recipe identification numbers for remote control . . . . . . . . . . .
43
5.3.5.3
Switching recipe via General I/O module inputs
. . . . . . . . . . . .
44
5.3.5.4
Switching recipe via Modbus/TCP
. . . . . . . . . . . . . . . . . . . . . . .
45
5.3.5.5
Switching recipe via Ethernet/IP
. . . . . . . . . . . . . . . . . . . . . . . .
45
5.3.6
Using Recipe app to make unit conversions
. . . . . . . . . . . . . . . . .
46
5.3.6.1
Creating a recipe for unit conversion . . . . . . . . . . . . . . . . . . . . .
46
5.3.6.2
Displaying several unit conversions simultaneously
. . . . . . . . .
46
5.3.7
Configuring mA output for Recipe app
. . . . . . . . . . . . . . . . . . . . .
47
6
Module cards
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
49
6.1
Installation
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
49
6.2
Configuration
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
51
6.3
Relay module card PR-50023
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
53
6.3.1
Configuring relays
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
54
6.4
mA output module PR-50021 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
56
6.4.1
Configuring mA outputs
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
57
6.5
General I/O module PR-50024
. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
58
6.5.1
Connecting inputs/outputs
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
61
6.5.2
Configuring the general I/O module card
. . . . . . . . . . . . . . . . . . .
63
6.5.2.1
Using the general I/O module to trigger prism wash
. . . . . . . .
64
7
Prism wash . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
66
7.1
Configuring prism wash
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
66
7.1.1
Wash cycle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
66
7.1.2
Setting prism wash parameters . . . . . . . . . . . . . . . . . . . . . . . . . . .
70
8
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
72
8.1
Model code
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
74
8.2
Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
75
A
EU declaration of conformity
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
77
B
Software licence . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
79
© Copyright Vaisala 2019. All rights reserved.
Command selection tree for Refractometer application . . . . . . . . . . . . . . . . . . . . 81
Command selection tree for settings
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
© Copyright Vaisala 2019. All rights reserved.
1 Description
The Multichannel user interface MI is an industrial computer designed to add function-
ality to the PR-43 refractometers. The MI sends and receives data and it can also be
used to control the process.
The Multichannel user interface MI connects with up to 4 Vaisala K-PATENTS® PR-43
refractometers. With module cards, it can also be used to add external inputs and
outputs (e.g. relays and mA outputs) to your refractometer system.
The MI is operated by the touch-screen display. The stainless steel enclosure gives a
good environmental protection and allows also outdoor installations, especially with
the included rain cover. The outer door is easy to open by one hand and the handle
can be modified for padlock to prevent unauthorized access. The module cards can be
reached by opening the inner door. The MI has IP66 rating when the outer door is open
if the inner door (display panel) is fixed. With outer door closed the rating is IP67.
Figure 1.1
Multichannel user interface MI
MI instruction manual
Document/Revision No. IM-EN-MI/2.00
Effective: September 2, 2019
2 Mounting
The Multichannel user interface MI can be mounted on a wall with wall mounting or
on a table with a table stand. The mounting location should be easily accessible. If
mounted outdoors, the MI has to be protected from rain and direct sunlight.
Display on eye-level
External
mains switch
100-240VAC
4-20 mA
Ethernet
1-4 refractometers
Figure 2.1
Mounting guide
MI instruction manual
Document/Revision No. IM-EN-MI/2.00
Effective: September 2, 2019
for refractometers
Figure 2.3
Enclosure dimensions
© Copyright Vaisala 2019. All rights reserved.
1
1. Bracket screwed to wall by 4 screws.
Screw hole dimensions 325x254mm
325 mm (12.8 inches)
254mm
(10 inches)
1
1
1
Figure 2.4
Mounting the wall bracket
1. Four mounting screws
loosely attached
2. Top screws inserted into top bracket verticalmounting slots
3. Lower screws slide into bottom bracket horizontal mounting slots
4. Tighten mounting screws x4
width a 19 mm ring spanner
1
2
3
4
Figure 2.5
Mounting the MI onto the wall bracket
Note: Be careful when mounting, the MI weighs 13 kilos (28.5 lbs).
MI instruction manual
Document/Revision No. IM-EN-MI/2.00
Effective: September 2, 2019
Important: Do not drill mounting holes in the enclosure as that will affect the protec-
tion class of the enclosure and damage the electronics.
The MI is shipped with a rain cover. Vaisala recommends that it is installed even indoors
to protect the top of the door from dust and dirt.
1. Screw out top attaching screws to allow
clearance for rain cover to slot in place
1
2. Retighten screws
2
Figure 2.6
Mounting the rain cover
The table stand is very simple to mount.
Figure 2.7
Table stand
© Copyright Vaisala 2019. All rights reserved.
The MI door handle can be modified to make space for a padlock to prevent unautho-
rized access.
Remove handle center cover
and remove segment indicated
by break-line on the back of cover
Replace center cover once
indicated segment has been
removed, making sure the opening
aligns with the channel for the padlock shackle
Figure 2.8
Locking the MI
MI instruction manual
Document/Revision No. IM-EN-MI/2.00
Effective: September 2, 2019
3 Connections
Refractometer + MI
max. 200m (656ft)
PR-8441
Ethernet cable for interfaces
M12-4 pin, D-code, M+RS45
Ethernet
PR-8440
Ethernet cable for interfaces
M12-4 pin, D-code, M+M
Ethernet
mA cable
M
M
M
RS45
M
F
PR-8430
Platform 4 cable
M12-8 pin, A-code, F+M
Ethernet, mA, 24V
PR-43 REFRACTOMETER
M-12-8 pin, D code, M
non-Ex
Multi Channel User Interface, MI
1 x M12-4 pin, D-code, F (External Ethernet)
4 x M12-8 pin, A-code, F (System)
max. 90m (295ft)
max. 90m (295ft)
max. 90m (295ft)
100-240VAC or 24VDC
Figure 3.1
Multichannel user interface connections
© Copyright Vaisala 2019. All rights reserved.
3.1 Connecting the power cable
Warning!
Make sure there is no power in your cable. Then open the enclosure and
the inner door. You need a TX25 screwdriver to open the inner door retaining screws.
1
2
3
1. Push door gently
2. Lift handle 45 deg up
3. Open the door
Figure 3.2
Opening the enclosure
Remove inner door retaining screws
and open inner door
Figure 3.3
Opening the inner door
Unscrew one of the cable glands and push the power cable through. Connect the power
cable to the power terminals at lower right of the motherboard (see below). Screw on
the cable gland. Close the inner door and turn power on.
3.1.1 Power terminals for AC power
The primary AC power is connected to a separate terminal strip 1/2/3 in the lower right-
hand corner of the motherboard. The three terminals are marked 1.L, 2.N and 3
(protective earth). The power terminal 3/
is directly connected to the exposed metal
parts of the MI. Wiring to the terminals shall be 1.5mm2 minimum. The protective fuse
in the building system shall comply with the local requirements.
MI instruction manual
Document/Revision No. IM-EN-MI/2.00
Effective: September 2, 2019
1.L 2.N 3
Brown
Blue
Yellow-green
Figure 3.4
AC power terminals on the motherboard
3.1.2 Power terminals for DC power
The DC power is connected to a terminal stripe in the lower right-hand corner of the
motherboard. The three terminals are marked 1+, 2- and 3
(protective earth). The
power terminal
is directly connected to the exposed metal parts of the MI. The 24V
DC power to this terminal stripe shall be supplied from a secondary circuit which is
double or reinforced insulated from the mains supply within the limits for a limited-
energy circuit (maximum 200 VA/U) according to the IEC 61010-1.
3.1.3 External mains switch
External
mains switch
Figure 3.5
Installing external main switch
© Copyright Vaisala 2019. All rights reserved.
Important: The MI does not have a built-in power switch. The system is always pow-
ered when connected to a power source. Vaisala recommends mounting an external
mains switch to control the MI’s power supply (Figure 3.5). The external power switch
should be easy to access. It is also recommended to use a 10A/230V circuit breaker for
power supply.
MULTI CHANNEL USER INTERFACE MI
INSTRUMENTS
M12 CONNECTORS
L
N
MAIN POWER
SWITCH PR-10900
L
N
+
-
mA Output 1
4-20mA
POWER SUPPLY
100-240VAC/50-60Hz
4158
-
+
-
+
OPTIONAL 24V POWER
SUPPLY FOR MI
24VDC
ETHERNET
M12-4pin, D-code, M
ETHERNET
M12-4pin, D-code, M
PR-8440 Ethernet cable for interfaces
MI has 8 module slots. Any of module cards
can be installed into any free slot. There could
be multiple module cards in slots depending on
external I/O needs.
PR-50021 mA output module
PR-50023 Relay output module
PR-50021 mA output module
PR-50023 Relay output module
PR-8430
Platform
4 cable
SLOT 2
SLOT 4
SLOT 6
SLOT 8
SLOT 1
SLOT 3
SLOT 5
SLOT 7
1...4 x
L N
SOLENOID VALVE FOR
PNEUMATICALLY OPERATED
SHUT-OFF VALVE ASSEMBLY:
PR-3340-24/110/240
1...4 x
9
N
L
SAFETY SWITCH
PR-7060
8
7
6
5
4
3
2
1
N L
L
L
POWER SUPPLY
100-240VAC/50-60Hz
or 24VDC
L
N 1...4 x
1...4 x
Grounding screw
PR-43-...-UN/AX/IA refractometer
(For Intrinsic refractometer see further wiring drawing 4118)
1...4 x
Figure 3.6
Multichannel user interface wiring
MI instruction manual
Document/Revision No. IM-EN-MI/2.00
Effective: September 2, 2019
3.2 Refractometer cable connections
The refractometers are connected to the MI with a M12 male-female 8 pins A-coded
cable PR-8430, maximum length 90 m. The refractometer connectors are located un-
derneath the MI enclosure as a group of four. Refractometer connections can be made
while the MI is powered. A refractometer can be connected to any of the four connec-
tors.
3.3 Ethernet cable connection
The leftmost connector underneath the MI is for ethernet output. It is for a M12 male
4-pins D-coded cable. The cable can be connected while the MI is powered.
If you wish to connect the MI with e.g. a laptop, you can purchase Interconnecting
cable PR-8330 with a M12-4 pin connector in one end and a RJ-45 connector in the
other end.
3.4 Module cards
For the installation of the module cards, see Chapter 6 “Module cards”.
© Copyright Vaisala 2019. All rights reserved.
3.5 Connections for external control of Recipe app
1
GPIO
mA-output
6
2
3
4
24VDC or
240VAC
5
Item
Component
Part number Supplied by
QTY
1
PR-43 Refractometer
Vaisala
1-4
2
Platform 4 cable
PR-8430
Vaisala
1-4
3
Multichannel User Interface MI
Vaisala
1
4
Main power switch
PR-10900
Vaisala
1
5
mA-output cable
Customer
1
6
GPIO selection cable
Customer
1
Figure 3.7
Recipe app controlled via GPIO (SYS-605)
MI instruction manual
Document/Revision No. IM-EN-MI/2.00
Effective: September 2, 2019
1
24VDC or 240VAC
mA-output
5
2
3
6
4
Ethernet
Item
Component
Part number
Supplied by
QTY
1
PR-43 Refractometer
Vaisala
1-4
2
Platform 4 cable
PR-8430
Vaisala
1-4
3
Multichannel User Interface MI
Vaisala
1
4
Main power switch
PR-10900
Vaisala
1
5
mA-output cable
Customer
1
6
Ethernet cable
PR-8330
Vaisala
1
Figure 3.8
Recipe app controlled via Modbus (SYS-606)
© Copyright Vaisala 2019. All rights reserved.
4 Use
Power up the Multichannel user interface MI. A startup display with a progress bar is
shown for approximately one minute, then the Main display appears.
The MI is operated by tapping the touch screen. To get access to the MI touch screen,
open the enclosure door.
Note:
If you find that the touch screen is difficult to use with fingers, try tapping it
with e.g. the blunt end of a pen or use a fingernail instead of the whole fingertip. The
screen works even when you are wearing gloves.
4.1 Main display
The main view can show up to four application windows at the same time. You can
access measurement data and system settings via the Main display.
Tap anywhere
in the application
window to access
measurement
Tap for second
main view screen
Tap to access
system settings
Figure 4.1
Accessing measurement and system settings via Main display
The Main display setup is controlled via Settings > General. Display mode and display
language are chosen in drop-down menus, tap the little triangle to see the whole menu.
If you only have one application (usually a single refractometer), One maximized will
make the most of the display. Dual vertical allows you to see two application, Quadrant
and One on top shows up to of four applications at the same time. Note: If you have
three applications and choose One maximized, you’ll only see one application, but the
other two will continue to run in the background.
MI instruction manual
Document/Revision No. IM-EN-MI/2.00
Effective: September 2, 2019
Display
mode
Display
language
Figure 4.2
Selecting display mode and display language
Figure 4.3
Setting "One maximized"
© Copyright Vaisala 2019. All rights reserved.
Figure 4.4
Setting "Dual horizontal"
Figure 4.5
Setting "Quadrant"
MI instruction manual
Document/Revision No. IM-EN-MI/2.00
Effective: September 2, 2019
Figure 4.6
Setting "One on top"
Note: If the MI is left with a different display, e.g. Calibration, it will automatically go
back to the Main display after a timeout.
Note: You can adjust Refractometer app to show concentration only, temperature only
or both concentration and temperature. To change this setting, tap on the application
window (see Figure 4.1) and choose Calibration (You may need to enter your password
to get access to Calibration display). Then select the preferable display mode.
© Copyright Vaisala 2019. All rights reserved.
Figure 4.7
Calibration display
4.2 Activating and removing apps
The Main display may be empty even when one or several refractometers are con-
nected to the MI. This means that one or several applications must be activated. Go to
Settings > Applications, tap on the empty application slot and select the app type and
source for that slot. After choosing settings for the app, tap Done to exit the setup.
MI instruction manual
Document/Revision No. IM-EN-MI/2.00
Effective: September 2, 2019
Figure 4.8
Empty main display in quadrant mode
Figure 4.9
Selecting an empty application slot to activate the app
© Copyright Vaisala 2019. All rights reserved.
Figure 4.10
Selecting a source to activate the
app, here application type Ref is already selected
The empty application slot may contain an earlier app configuration. In that case a dia-
log will open during app activation and the earlier app configuration can be re-enabled.
Figure 4.11
Selecting an empty application slot
that contains old Refractometer app configuration
MI instruction manual
Document/Revision No. IM-EN-MI/2.00
Effective: September 2, 2019
Figure 4.12
The earlier app configuration can be re-enabled
If necessary the app can also be removed. Go to Settings > Applications and tap Remove
on the application slot. Accept the removal by tapping Ok. App’s configuration will be
saved until another kind of application is added to the same slot.
Figure 4.13
If necessary, an app can also be removed
© Copyright Vaisala 2019. All rights reserved.
4.3 Managing users and passwords
Some functions and settings, e.g. calibration, require a user to log in. Most functions
in the settings, including user management, are only available to the master user. The
master user password (pin code) is printed on the delivery data sheet shipped with
the MI.
Note: Controlling and switching recipes in Recipe app (see section 5.3) is not protected
by user passwords.
Go to Settings > Users to manage users. Vaisala recommends that a user name and
a password is set separately for each user who needs to have frequent access to the
restricted functions.
Figure 4.14
The Users display
Tap + Add User to add users. Tap Set name to get access to the keyboard, then type the
username and tap OK.
MI instruction manual
Document/Revision No. IM-EN-MI/2.00
Effective: September 2, 2019
Figure 4.15
Writing a username
Note: If you have entered a username, you must enter a password, too, you cannot
leave the password field empty.
Figure 4.16
New user added to the list
© Copyright Vaisala 2019. All rights reserved.
To remove a user, go to Users and tap the name of the user you want to remove, then
tap Remove user.
Note: The MI can store up to 38 users.
4.4 System log
To access a system log, go to Settings > Log. The system log lists status changes of the
system like booting up, adding an app and changing module card settings. The system
log is also accessed via http connection. Use the IP address found in the MI Settings >
Network menu, under Ethernet section. On the MI web page select files > system > log.
Figure 4.17
System log
4.5 Ethernet connections
4.5.1 MI ethernet connection
The MI has a built-in web server with a home page. The MI home page allows you to
browse files or upload an update or a license file. To access the MI web page use IP
address found under the title Ethernet in the Settings > Network menu.
MI instruction manual
Document/Revision No. IM-EN-MI/2.00
Effective: September 2, 2019
The default IP address mode setting for the MI is Auto. In this mode the MI gets the
IP address from an external DHCP server. If that fails the IP address will fall back to a
static IP setting. Alternatively IP address mode: Static configures the MI to use a static
IP address only. After choosing settings for the network, tap Apply to save all settings.
After factory reset the default IP address of the MI is 192.168.1.11.
Figure 4.18
To access the MI web page use IP
address found in the Settings > Network menu
© Copyright Vaisala 2019. All rights reserved.
Figure 4.19
Steps to configure a static IP address for an MI
4.5.2 Refractometer connection
Refractometer’s connection to the MI is configured under the title Instrument in the
Settings > Network menu. By default the refractometer’s IP address is configured via
BOOTP server running on the MI. To reconnect a refractometer you can just replug its
cable to the MI – the IP address is configured automatically. Refractometer connections
can be made while the MI is powered.
MI instruction manual
Document/Revision No. IM-EN-MI/2.00
Effective: September 2, 2019
5 Apps
Apps are small programs that give you different types of measurement data and func-
tionality. Vaisala is constantly developing new apps for the measurement, but some
are installed by default.
5.1 Shared functionality
While different applications work differently, there are some functions that are similar
to all applications.
5.1.1 Application trend
An MI app can display graphically one or two of its variables over selectable time period.
To access an app trend, tap application window on the main view and tap Trend. Save
the trend by tapping Save. Saved trend can be accessed via http connection. Use the
IP address found in the MI Settings > Network menu, under Ethernet section. On the
MI web page select files > trend > yearMonthDate-time_appType_slotNumber.csv.
Note: Only apps running on the first four app slots can display the graphical trend. MI
supports four apps on one main view screen and there are two main view screens that
can be switched, see figure 4.1.
© Copyright Vaisala 2019. All rights reserved.
Figure 5.1
MI app trend display
Figure 5.2
Saving app trend
MI instruction manual
Document/Revision No. IM-EN-MI/2.00
Effective: September 2, 2019
Figure 5.3
To access saved app trends, select files > trend on the MI web page
Figure 5.4
Select the trend to be viewed/downloaded
5.1.2 Application log
The app log lists status changes and possible error messages of apps in an app slot. To
access an app log, tap application window on the main view and tap Log. The app log is
addressed by the app slot number and lists status changes and possible error messages
of apps in that slot. App logs are also accessed via http connection. Use the IP address
found in the MI Settings > Network menu, under Ethernet section. On the MI web page
select files > app > slotNumber > log.
© Copyright Vaisala 2019. All rights reserved.
Figure 5.5
App log
Figure 5.6
On the MI web page select files > app
MI instruction manual
Document/Revision No. IM-EN-MI/2.00
Effective: September 2, 2019
Figure 5.7
On the MI web page select files > app > slot number
Figure 5.8
On the MI web page select files > app > slot number > log
5.2 Refractometer app
Refractometer app is included in your delivery. It gives you basic information about the
process, e.g. concentration and temperature.
5.2.1 Diagnostics display
Tap the application window to access the Diagnostics display. In the top part of the
window you see the optical and slope curves. By default both optical image and slope
are shown, but you can tap on the texts OPTICAL and SLOPE to turn these on and off.
© Copyright Vaisala 2019. All rights reserved.
Figure 5.9
Diagnostics display
The tabs in the lower part of the display contain detailed information about the mea-
surement. Tap the name of the tab to see the contents. Input tab shows the values
as sent by the refractometer. Output tab shows the results of the calculations by MI.
Identification tab contains various identification information and e.g. software version
information. Support tab is a summary of the information in the other tabs and other
information needed by technical support for troubleshooting.
Figure 5.10
Input tab
Figure 5.11
Output tab
MI instruction manual
Document/Revision No. IM-EN-MI/2.00
Effective: September 2, 2019
Figure 5.12
Identification tab
Figure 5.13
Support tab
5.2.2 Viewing and adjusting parameters
To view the parameters, go to the Calibration display and tap on Summary. Note that
the summary extends to a second page, a small arrow on the top right will take you to
the second (and third) page.
To adjust the parameters, go to the Calibration display, then choose appropriate para-
meter set, eg. Field. In the list, tap the parameter you want to change, type the new
numbers with the number pad and approve the change by tapping DONE. Note: The
parameter lists for Chemical and Field do not fit on the screen. Swipe down or use the
scroll bar on the right to view the bottommost parameters.
© Copyright Vaisala 2019. All rights reserved.
Figure 5.14
Calibration display
The settings in the upper part of the Calibration display are different types of lists that
sets parameters for showing the measurement data, e.g. measurement unit. In the
lower part of the window, the words are links to the given settings, e.g. by tapping
on Field you get access to the field calibration parameters. If you want to see all the
parameters, tap Summary.
MI instruction manual
Document/Revision No. IM-EN-MI/2.00
Effective: September 2, 2019
Figure 5.15
First page of calibration summary
(tap the little arrow in top right to see page 2)
Figure 5.16
Second page of calibration summary
(tap the little arrow in top left to go back to page 1)
© Copyright Vaisala 2019. All rights reserved.
5.2.3 Refractometer verification
An offline refractometer can be verified with standard refractive index liquids, available
from Vaisala. Depending on the refractometer model, a sample holder may be needed.
Go to the Diagnostics display (see Figure 5.9) and choose Verification. If you are not fa-
miliar with the verification procedure, tap the question mark on the Verification display
for additional instructions.
Figure 5.17
Verification in process
MI instruction manual
Document/Revision No. IM-EN-MI/2.00
Effective: September 2, 2019
Figure 5.18
Finished verification
Once you have finished the verification process, you can save the verification results
as pdf. The pdf is stored in the MI memory and it can be downloaded over the eth-
ernet connection. Each pdf report gets a unique name and thus you can have several
verification pdfs in the MI at the same time.
Figure 5.19
Pdf saved
© Copyright Vaisala 2019. All rights reserved.
5.2.4 Collecting field sample points
You can use the field sample functionality to acquire data points for Vaisala K-PATENTS®
field calibration service. Go to the Diagnostics display (see Figure 5.9) and tap Field
sample. Tap Collect field point to start collecting field samples and then tap Save file to
save the collected field points. These can be accessed via http connection. Use the IP
address found in the MI Settings > Network menu, under Ethernet section. On the MI
web page select files > fieldsamples > FieldSamples_SlotNumber.
Figure 5.20
Start collecting field samples by tapping Collect field point
MI instruction manual
Document/Revision No. IM-EN-MI/2.00
Effective: September 2, 2019
Figure 5.21
Field sampling in process
Figure 5.22
Save collected field points by tapping Save file
© Copyright Vaisala 2019. All rights reserved.
5.3 Recipe app
Recipe app provides a convenient way to control and switch different chemical curves.
Each Recipe app can provide up to 32 different recipes. Switching a recipe in Recipe app
means switching the chemical curve that is used in calculations. Each process medium
has its specific chemical curve and by switching it the same processing line can be used
to manufacture a variety of products.
5.3.1 Requirements
Using Recipe app requires a Refractometer app, which provides refractive index nD
and temperature as a source data to the connected Recipe app. A licence is required
to enable Recipe app. New Multichannel user interface MI:s are shipped with Recipe
app licence installed. Old MI:s require software update to version 1.07 or newer and
licence installation to enable Recipe app. Recipe app licences are granted free of charge
to older MI:s.
5.3.2 Activating Recipe app
To activate Recipe app go to Settings > Applications, tap on a free application slot and
select Recipe. Select a correct source Refractometer app with Select sources. The avail-
able Refractometer apps are listed based on their location in MI application slots. If
no Refractometer app is listed in Select sources, then activate a Refractometer app Ref
first.
5.3.3 Creating a new recipe
MI is provided with a Brix chemical curve as an example recipe for Recipe apps. A
new Recipe app has the example recipe activated by default. A new recipe is instantly
created when the example recipe is renamed. To rename a recipe go to Recipe app >
Calibration > Chemical and select Name. The new recipe is a copy of the old recipe with
the new name.
5.3.4 Modifying a recipe
A recipe can be modified e.g. by adjusting a field calibration parameter to offset the
measurement result according to recipes requirements. To adjust the field parameters
MI instruction manual
Document/Revision No. IM-EN-MI/2.00
Effective: September 2, 2019
of an active recipe go to Recipe app > Calibration > Field. In the list, tap the parameter
you want to change, type the new numbers with the number pad and approve the
change by tapping DONE. Note: The parameter lists for Chemical and Field do not fully
fit on the screen. Swipe down or use the scroll bar on the right to view the bottommost
parameters.
All recipes that are present in MI are visible and usable to all Recipe apps. If a specific
recipe is modified in some Recipe app, the modification propagates also to the other
Recipe app using the same recipe.
5.3.5 Selecting recipe switching mode
Recipes in the Recipe app can be switched manually, via General I/O module inputs
(Switch input) or via Modbus/TCP. To select the switching mode, tap the application
window to access Diagnostics display. Then tap Remote control and then select the
recipe switching mode.
Figure 5.23
Selecting how to switch recipes in the Recipe app
5.3.5.1 Switching recipe manually
To switch recipe manually, tap the application window to access Diagnostics display.
Then tap Change recipe, select the recipe by tapping its name and accept by tapping
OK.
© Copyright Vaisala 2019. All rights reserved.
Figure 5.24
Switching recipe manually
5.3.5.2 Recipe identification numbers for remote control
A recipe is activated remotely, when a remote signal corresponds to a recipe identifica-
tion number. To set the identification number go to Recipe App > Calibration > Remote
control, tap a recipe name and select the identification number for that recipe. The
identification number can have a value between 0 - 31. Different Recipe apps have
independent values for their recipe identification numbers and they need to be set in
each Recipe app that uses remote control functionality.
MI instruction manual
Document/Revision No. IM-EN-MI/2.00
Effective: September 2, 2019
Figure 5.25
Recipe identification numbers for remote control
Figure 5.26
Setting an identification number for a recipe
5.3.5.3 Switching recipe via General I/O module inputs
A recipe is activated via General I/O module inputs, when the inputs form a binary
number that corresponds to a recipe identification number. The recipe switching mode
Switch input must be selected, see figure 5.23. To configure General I/O module, go
© Copyright Vaisala 2019. All rights reserved.
to Settings > Modules, tap the appropriate General I/O module to select it and then
View module to access module configuration. Set the wanted inputs to control the
Recipe app, which has a number according to the appslot it is running on. Set the
purpose of inputs to implement a binary number with different multipliers for different
bits. Four digital inputs associated to a single Recipe app enable 24 different binary
number combinations that can address 16 different recipes. After choosing the settings
for inputs, tap Done to save changes and exit the setup.
5.3.5.4 Switching recipe via Modbus/TCP
In Modbus mode the recipe is switched by writing the recipe identification number to
a Modbus register. Each appslot has a specific Modbus register to switch recipe for
Recipe app running on that appslot. The register address is of the form 1n200, where
n is the appslot number. The register data type is a four-byte signed integer. The recipe
switching mode Modbus must be selected, see figure 5.23.
5.3.5.5 Switching recipe via Ethernet/IP
In Ethernet/IP mode the recipe is switched by writing the recipe identification number
to an Ethernet/IP object InputRecipeID, which is a four-byte integer. Ethernet/IP con-
nection parameters depend on what apps are running on MI. When Refractometer app
is on the first appslot, Recipe app is on the second appslot and all other appslots are
empty, the connection parameters are:
•
Port: 0xAF12 (44818)
•
O->T:
–
instance number: 119
–
data size: 4
–
real time format: modeless
–
packet rate: 1000 ms
•
T->O:
–
instance number: 109
–
data size: 120
–
real time format: modeless
–
packet rate: 1000 ms
–
connection type: point to point
MI instruction manual
Document/Revision No. IM-EN-MI/2.00
Effective: September 2, 2019
Also, an Electronic Data Sheet (EDS) file is available for easy setup of the Ethernet/IP
connection. The recipe switching mode Ethernet/IP must be selected, see figure 5.23.
5.3.6 Using Recipe app to make unit conversions
Recipe apps can perform unit conversions so that measurement results can be dis-
played in two or more units simultaneously. Unit conversions enable familiar readings
like numeric values and engineering units, and comparing the measurement values to
laboratory results gets easier.
5.3.6.1 Creating a recipe for unit conversion
A properly configured recipe is needed for each unit conversion. Create a new recipe
for each new unit to be displayed. For clarity it is recommended to name the new
recipe according to the unit. To implement the unit conversion, adjust a field calibration
parameter F 10 by setting its value to (M – 1) where M is the conversion multiplier. See
section 5.3.3 “Creating a new recipe” and section 5.3.4 “Modifying a recipe”.
5.3.6.2 Displaying several unit conversions simultaneously
To display several unit conversions simultaneously, run each unit conversion recipe in
its own Recipe app. If necessary, activate new Recipe apps, see section 5.3.2 “Activating
Recipe app”. Connect the new Recipe apps to the same source Refractometer app to
perform unit conversions of a single measurement. To change the primary unit marking
on the Recipe app display, go to Recipe app > Calibration and tap the little triangle of
drop-down menu Primary unit.
© Copyright Vaisala 2019. All rights reserved.
Figure 5.27
Changing the primary unit marking on the Recipe app display
5.3.7 Configuring mA output for Recipe app
If no mA output is configured for Recipe app, app status in Recipe app main view will
blink between current status and message ”mA output not configured”. This will pre-
vent false or omitted configuration of mA output. To configure mA output module, go
to Settings > Modules, tap the appropriate mA output module to select it and then View
module to access module configuration. Set the wanted output(s) for Recipe app, which
has a number according to the appslot it is running on. Set the purpose of output(s),
see section 6.4.1. After choosing the settings for outputs, tap Done to save changes
and exit the setup.
MI instruction manual
Document/Revision No. IM-EN-MI/2.00
Effective: September 2, 2019
Figure 5.28
Blinking app status message when no mA output is configured
Figure 5.29
Configuring mA output for Recipe app
6 Module cards
49
© Copyright Vaisala 2019. All rights reserved.
6 Module cards
A module card adds functions to the MI, e.g. a relay module card contains two relays
that can be used as wash relays or alarm relays. There are eight module card slots inside
an MI. Each Multichannel User Interface MI is shipped with one relay module card with
2 independent relays and one mA output module card with two independent 4-20 mA
outputs as a standard. These are already installed in the MI. In addition to the relay
module card and the mA output module card, there is a general input/output module
card with 8 digital inputs/outputs. These module cards have the capability to be used
in any combination as well as added and removed as needed.
6.1 Installation
To install a module card, turn off MI power first. Then open the enclosure, remove the
display screws and fold the display to the side (see Figure 6.1). Unscrew and remove
the module card cover to get access to the card slots. Insert the module card in an avail-
able card slot. To make adding cards easier, start from the left. Vaisala recommends for
safety reasons that relay modules are mounted in slots 2, 4, 6 and 8 (connector down-
wards). This way the relay power wiring is as short as possible. Other module cards
may be mounted in any slot.
Important: A relay module card may have an external power source. Make sure that
also this external power source connected to a relay module is OFF before installing or
removing module cards.
MI instruction manual
Document/Revision No. IM-EN-MI/2.00
Effective: September 2, 2019
Remove inner door screws
(TX25 screwdriver needed)
Open the inner door
and move it to the side
Open the four screws holding
the module card holder
and remove the holder
Insert the module card(s) and
replace the module card holder
Figure 6.1
Installing module cards
To make the connections, first remove the removable connector from the module card.
Bring the module connection cable to the MI and put it through the cable gland/conduit
hub (1). Then connect wires to the screw terminals on the removable connector (2), see
the module-specific sections for further information. Finally reconnect the removable
connector to the module card (3).
6 Module cards
51
© Copyright Vaisala 2019. All rights reserved.
1
2
3
4
5
6
7
8
1
2
3
4
5
6
7
8
1
2
3
4
5
6
7
8
1
2
3
4
5
6
7
8
1
2
3
3
MI-010
Figure 6.2
Connecting module connection cable to module card
6.2 Configuration
The MI automatically finds the installed module card(s). To configure a module card,
go to Settings > Modules, tap the appropriate module card to select it and then VIEW
MODULE to access module configuration. The selected module card is marked with a
border.
Note: There are two types of EMPTY slots: grey slots do not have module card, striped
slots are turned off (see Figures 6.3 and 6.4 on next page) and there may or may not be
a module card in that slot. If there is a module card, it will be automatically recognized
when the module slot is turned on by selecting the module and then tapping the TURN
ON command.
MI instruction manual
Document/Revision No. IM-EN-MI/2.00
Effective: September 2, 2019
Figure 6.3
Selecting a module card and viewing module card information
Figure 6.4
Selecting a turned-off module
card slot (slot 1 in this case) and turning it on
6 Module cards
53
© Copyright Vaisala 2019. All rights reserved.
6.3 Relay module card PR-50023
The PR-50023 relay module card adds two relays to the Multichannel user interface
MI. The relays can be configured independently to any connected refractometer. The
relays can be used e.g. for controlling prism wash.
Specifications
Relays per module
2
Voltage and current rating
Max. 250V/3A
Connection
Screw connection with tension sleeve
Connector
Phoenix 1900824 – MC 1,5/ 7-ST1-5,08
Contact type
Dry contact
Normal relay state
NO
Conductor cross section
0.14-1.5 mm2 (28-16 AWG)
Stripping length
7 mm (¼")
Tightening torque
0.25 Nm (0.18 lbf.ft)
7 Relay 1+
6 Relay 1-
5 Relay 2+
4 Relay 2-
3 no connection
1 no connection
2 no connection
LED
RLY1
LED
RLY2
Figure 6.5
Relay connections
MI instruction manual
Document/Revision No. IM-EN-MI/2.00
Effective: September 2, 2019
6.3.1 Configuring relays
On the relay display, first select the application the relay is used for, then select the
function. After the function is chosen, you get access to those relay parameters that
apply to this function.
Figure 6.6
Configuring a relay
Note:
Even though the relays can be configured independently, it is recommended
that if you e.g. set precondition and wash parameters for a certain refractometer, you
set those in the same relay module card.
Important: When you have chosen the settings for the relays, tap DONE to save changes
and exit the relay setup.
Parameters LIMIT and HYSTERESIS are used to switch the relay based on concentration
or temperature measurement values by the functions CONC - HIGH LIMIT, CONC - LOW
LIMIT, T - HIGH LIMIT and T - LOW LIMIT. The relay closes on high limit when the mea-
sured value exceeds HIGH LIMIT + HYSTERESIS. The relay opens on high limit when the
measured value goes below HIGH LIMIT - HYSTERESIS. Alternatively, the relay closes
on low limit when the measured value goes below LOW LIMIT - HYSTERESIS. The relay
opens on low limit when the measured value exceeds LOW LIMIT + HYSTERESIS.
6 Module cards
55
© Copyright Vaisala 2019. All rights reserved.
Relay test page
Relay parameters
Clear all settings
Finish and
save settings
Figure 6.7
Accessing relay functions
Figure 6.8
Relay module parameters
MI instruction manual
Document/Revision No. IM-EN-MI/2.00
Effective: September 2, 2019
6.4 mA output module PR-50021
The mA output module card adds mA outputs to Multichannel user interface MI. On
the module card there are two independent 4-20 mA current sources. Both current
outputs are configurable independently to indicate for example process concentration
or temperature or other diagnostic data of any connected refractometer. A mA output
has traceable calibration, accuracy is ±0.02 mA.
Maximum load (compliance) is 1000 Ohm. There is a place inside the Multichannel
user interface MI where the cable shield can be connected for grounding.
Cable specifications
Conductor cross section
0.14-1.5 mm2 (28-16 AWG)
Tightening torque
0.25 Nm (0.18 lbf.ft)
10 mA out 1+
9 mA out 1-
8 no connection
7 no connection
6 mA out 2+
5 mA out 2-
4 no connection
3 no connection
2 Frame ground
1 Frame ground
Figure 6.9
mA output pinout
The mA outputs are compliant to Namur NE43. NAMUR NE 43 uses the 3.8 to 20.5 mA
signal range for measurement information, with values ≥21 mA or ≤3.6 mA to indicate
diagnostic failures (see Figure 6.10). With that information, it is easier to detect a failure
condition on a refractometer, for example, it clearly tells you if you have an empty pipe.
6 Module cards
57
© Copyright Vaisala 2019. All rights reserved.
0
3.6
3.8
4
20
20.5
21
mA
failure
OK
Measurement Data
failure
Figure 6.10
Default mA output values
Certain malfunctions cause the mA measurements to return to mA default output
value, see Table 6.1 below. Factory setting for default output value is 3.4 mA.
Returns to default mA
Message
Conc
Temp
Short-circuit
X
X
No optical image
X
Temp measurement fault
X
X
No sample
X
Prism coated
X
No connection
X
X
Table 6.1
Messages causing mA output value to return to default output
6.4.1 Configuring mA outputs
First select the application for an output, then set function. There are five possible
functions, CALC, CONC, QF, TEMP and ND. After you’ve chosen a function, you can set the
parameters for that function.
•
CALC is the calculated concentration value without field calibration adjustment
•
CONC is the final concentration value including field calibration adjustment
•
QF is the quality factor measuring the image sharpness, available for Refractometer
app
•
TEMP is the process temperature
•
ND is the refractive index value nD from the sensor
MI instruction manual
Document/Revision No. IM-EN-MI/2.00
Effective: September 2, 2019
Finish by tapping here
to save settings
Tap for output
test page
Output parameters
Clears all settings
for this output
Figure 6.11
mA output settings
Important:
When you have chosen the settings for the outputs, tap DONE to save
changes and exit.
6.5 General I/O module PR-50024
The General I/O module card adds digital inputs and outputs to Multichannel user in-
terface MI. The module card provides 8 input/output channels.
The input and output terminals are galvanically isolated according to
•
UL 2500 V rms (UL1577)
•
CSA 800 V rms (1131V peak) maximum working voltage - basic insulation
•
CSA 400 V rms (566V peak) maximum working voltage - reinforced insulation
•
VDE Reinforced insulation 560V peak
•
TüV Reinforced insulation 400V rms maximum working voltage
Cable specifications
Conductor cross section
0.14-1.5 mm2 (28-16 AWG)
Tightening torque, max.
0.25 Nm (0.18 lbf.ft)
6 Module cards
59
© Copyright Vaisala 2019. All rights reserved.
Voltage reference:
IO voltage can be selected with a jumper in the module. IO voltage options are internal
5V, internal 24V and external voltage.
When external voltage is selected, a voltage of 3-30 VDC must be provided between
V_IO and GND pin.
When internal voltage is selected, voltage reference for switch input can be taken from
V_IO and GND pins.
MI instruction manual
Document/Revision No. IM-EN-MI/2.00
Effective: September 2, 2019
Electrical characteristics:
EXTERNAL IO VOLTAGE SUPPLY
External IO voltage input range (VIO)
3-30 VDC
Absolute minimum external IO voltage input
0 VDC
Absolute maximum external IO voltage input
36 VDC
OUTPUTS
Maximum combined* supply current when using internal
24V
20 mA
Maximum combined* supply current when using internal
5V
100 mA
Maximum individual channel supply current when using
external IO voltage
1 A
Maximum combined* supply current when using external
IO voltage
2 A
IO output voltage (voltage drop) when using external IO
voltage
VIN - 0.15 VDC @ 5 VDC
VIN - 0.20 VDC @ 24 VDC
INPUTS
Input type
Sinking
Logical 0: voltage range
0.0-0.7 VDC
Logical 1: voltage range
1.0-30 VDC
Input impedance
10 kohm
Absolute maximum input voltage
36 VDC
* current limit shared with all 8 channels
6 Module cards
61
© Copyright Vaisala 2019. All rights reserved.
6.5.1 Connecting inputs/outputs
Terminal no. Description
Function
1
Digital input/output
In/Out 1
2
Digital input/output
In/Out 2
3
Digital input/output
In/Out 3
4
Digital input/output
In/Out 4
5
Digital input/output
In/Out 5
6
Digital input/output
In/Out 6
7
Digital input/output
In/Out 7
8
Digital input/output
In/Out 8
9
Ground for outputs
GND
10
IO voltage input/output V_IO
1 2
3 4 5 6 7 8 9 10
Figure 6.12
General I/O module pinout
MI instruction manual
Document/Revision No. IM-EN-MI/2.00
Effective: September 2, 2019
Digital outputs:
Digital output voltages are selected with jumper. GND (pin 9) is used for ground. See
Figure 6.13 for connections.
Digital inputs:
Digital inputs can be used as switch inputs or controlled by external signal. Switch input
configuration is shown in Figure 6.14. When an external signal is used, an external
power supply is connected between input pin and GND (pin 9) as seen in figure 6.14.
1
2
3
4
5
6
7
8
9 10
GND
V_IO
PR-50024
Figure 6.13
Channels 1 and 2 configured as output
1
2
3
4
5
6
7
8
9 10
GND
V_IO
PR-50024
Figure 6.14
Channels 3 and 4 configured as input and connected as switch input
6 Module cards
63
© Copyright Vaisala 2019. All rights reserved.
1
2
3
4
5
6
7
8
9 10
GND
V_IO
PR-50024
+
Figure 6.15
Channel 5 configured as input from external power supply
6.5.2 Configuring the general I/O module card
To configure the inputs and outputs, select the module card in the module view. Select
the application to which the input/output is connected and then choose the desired
function from a list.
The digital outputs can be configured for
•
Concentration - high limit
•
Concentration - low limit
•
Error
•
Full wash (precondition, wash, recovery)
•
Normal
•
Precondition
•
Temperature - high limit
•
Temperature - low limit
•
Warning
•
Wash
The digital inputs can be configured for
•
External hold
•
External wash
•
External wash stop
MI instruction manual
Document/Revision No. IM-EN-MI/2.00
Effective: September 2, 2019
Tap to save all settings
Tap for test page
Change I/O 1
parameters
Clear I/O 1
settings
Figure 6.16
General I/O module card configuration
Parameters LIMIT and HYSTERESIS are used to switch the digital output based on con-
centration or temperature measurement values by the functions CONC - HIGH LIMIT,
CONC - LOW LIMIT, T - HIGH LIMIT and T - LOW LIMIT. Switching on high limit activates
when the measured value exceeds HIGH LIMIT + HYSTERESIS. The output switches off
on high limit when the measured value goes below HIGH LIMIT - HYSTERESIS. Respec-
tively, switching on low limit activates when the measured value goes below LOW LIMIT
- HYSTERESIS. The output switches off on low limit when the measured value exceeds
LOW LIMIT + HYSTERESIS.
Important: When you have chosen the settings for the inputs or outputs, tap DONE to
save changes and exit the setup.
6.5.2.1 Using the general I/O module to trigger prism wash
Prism wash can be started by configuring the general input for external wash request.
To activate the wash system a relay must be configured to be a wash relay. Select the
wash function on the relay display, see figure 6.6. To configure the general input, select
the function "IN: EXT WASH" on the I/O module display.
6 Module cards
65
© Copyright Vaisala 2019. All rights reserved.
Figure 6.17
Selecting the function "IN: EXT
WASH" to configure external wash request
MI instruction manual
Document/Revision No. IM-EN-MI/2.00
Effective: September 2, 2019
7 Prism wash
In some applications the process flow does not keep the prism clean because flow ve-
locity is not enough to keep the prism clean, or the process medium is sticky. In these
applications the prism can be automatically cleaned by installing a wash system (see
the refractometer manual).
7.1 Configuring prism wash
The prism wash settings for different refractometers are independent of each other.
The wash system is active if a relay has been configured to be a wash relay (see Sec-
tion 6.3.1) and the wash time is not zero. An automatic wash function can be configured
so that all refractometers connected to MI have different wash parameters. The wash
parameters can be configured at Refractometer app -> Configuration -> Wash.
7.1.1 Wash cycle
The wash logic is shown in Figure 7.2 as a flow diagram. The automatic prism wash cycle
(Figure 7.1) consists of three phases: precondition, wash and recovery. The optional
preconditioning function is used to e.g. blow out the condensate from the washing
line before washing. After the preconditioning there is a one-second pause to avoid
having both precondition and wash relays active at the same time.
The wash cycle is initiated when the wash interval has elapsed. The wash can also
be started by configuring a general input for external wash (Section 6.5.2) or manually
from the user interface at Refractometer app -> Calibration menu. The order of priority
for these wash triggers is
1. manual wash
2. external wash request
3. wash interval timer
The wash relay is closed for the wash time specified in the wash settings.
After the wash phase is completed, a recovery time is spent. During the wash cycle (pre-
condition, wash, recovery) the measurement result is in hold unless otherwise speci-
fied.
7 Prism wash
67
© Copyright Vaisala 2019. All rights reserved.
WASH
RECOVERY
INTERVAL
PRECOND.
Figure 7.1
The automatic prism wash cycle
Preventing automatic wash:
The precondition and wash relays are never activated by the automatic wash control:
–
If diagnostic message is NO SAMPLE as this indicates a clean prism in an empty process
line. The diagnostic message is WASH STOP/NO SAMPLE.
–
If a wash stop is present, indicating e.g. that there is no process flow. The diagnostic
message is EXTERNAL WASH STOP.
–
If the process temperature limit is activated and the temperature falls below the
limit, indicating that the process is not running. The diagnostic message is LOW TEMP
WASH STOP.
MI instruction manual
Document/Revision No. IM-EN-MI/2.00
Effective: September 2, 2019
MANUAL
WASH been
pressed?
NO
YES
External
wash
1
requested?
NO
YES
Wash
interval
reached?
YES
NO
Temp
limit
active?
YES
NO
Temp
above
limit?
YES
NO
Message
2
NO SAMPLE?
NO
YES
Wash cycle
Restart wash
interval timer
NOTES
1. External wash is triggered at
the general input. If the
input is held logical high,
only one wash cycle is
carried out.
2. The wash is inhibited if
there is no sample, no
refractometer or the
refractometer cannot
measure correctly.
Figure 7.2
Wash logic
7 Prism wash
69
© Copyright Vaisala 2019. All rights reserved.
Precondition
Start washing
1 s wait
Precondition
enabled?
YES
NO
Wash time
reached?
YES
NO
Stop washing
Recovery
PRECONDITION
RECOVERY
WASH
Figure 7.3
Wash cycle
MI instruction manual
Document/Revision No. IM-EN-MI/2.00
Effective: September 2, 2019
7.1.2 Setting prism wash parameters
To set prism wash parameters for a given refractometer, first select the Refractometer
app, then select Calibration and then Prism wash. This menu contains the alternatives
(factory settings are given in parentheses):
1. Precondition secs 0-60 s (0)
2. Recovery secs 0-120 s (20)
3. Wash secs 0-30 s (3)
4. Empty pipe check (enabled)
5. Wash check mode (disabled)
6. Wash verify delta nD limit 0.0-0.5 (0.0)
7. Wash inhibit temp enable (disabled)
8. Interval mins 0-1440 min (20)
9. Wash inhibit temp value 0-200 C (0)
10. Wash verify nD limit 1.15-1.80 (1.35)
11. Wash failure tolerance mins 0-1440 (0)
12. Measurement hold during wash (enabled)
13. Temperature hold during wash (disabled)
The prism wash cycle: The timing of the wash cycle is controlled by the WASH INTERVAL,
PRECONDITION TIME, WASH TIME and RECOVERY TIME settings. If the WASH INTERVAL is set to zero,
the wash can only be initiated by using the manual wash (PRISM WASH on Diagnostics
menu) or general input module configured for external wash.
If the PRECONDITION TIME is zero (or there is no relay configured for preconditioning), the
preconditioning phase is skipped. If the WASH TIME is zero (or there is no relay for wash),
the wash functionality is completely disabled.
The recommended wash times and wash medium pressures are given in the refrac-
tometer manual.
Wash check mode, standard: The prism wash check monitors automatically that the
prism wash really has an effect on the prism. In the WASH CHECK mode, prism wash is
accepted if the refractive index nD either falls below wash verify nD limit (default 1.35)
at NORMAL OPERATION or NO SAMPLE occurs. This is the indication of a successful wash
with water or steam.If the wash is not accepted, the diagnostic message PRISM WASH
WARNING will appear. If no wash is accepted during wash tolerance time, the message
7 Prism wash
71
© Copyright Vaisala 2019. All rights reserved.
becomes PRISM WASH FAILURE. Both messages and the wash tolerance counter are re-
set by a successful wash.The Wash check mode, automatic wash cut differs from the
standard mode by stopping the wash 2 seconds after the nD falls below the limit.
To stop the measurement for the duration of the prism wash, choose MEASUREMENT
HOLD DURING WASH and in that menu activate the hold function. The CONC reading and
mA output are held in the value they had immediately before starting the wash cycle.
You can also stop temperature measurement for the duration of the wash by setting
TEMPERATURE HOLD DURING WASH.
To set a low temperature limit, choose WASH INHIBIT TEMPERATURE ENABLE, enable it (fac-
tory setting is disabled) and then set the temperature with WASH INHIBIT TEMP VALUE. The
temperature is in the units that is chosen from the Calibration menu as the secondary
unit.
The empty pipe check prevents washing if message is NO SAMPLE, i.e. there’s no
process liquid in the pipe. To deactivate (or activate) the empty pipe check, choose
EMPTY PIPE CHECK and then the appropriate menu command.
To change the wash nD limit, select WASH VERIFY ND LIMIT to set the nD to be used by the
wash check functionality.
If you have set a wash verify delta nD limit, a temporary wash verify limit is calculated
in the beginning of wash: nD limit = nD current – ΔnD. This temporary limit is compared to
actual wash verify nD limit parameter, and the higher of these values takes precedence.
To set wash tolerance time, select WASH FAILURE TOLERANCE to set the time during which a
wash has to be accepted. If no wash is accepted during wash tolerance time, the mes-
sage becomes PRISM WASH FAILURE. The wash tolerance counter is reset by a successful
wash.
MI instruction manual
Document/Revision No. IM-EN-MI/2.00
Effective: September 2, 2019
8 Specifications
Model: Multichannel User Interface
Product code: MI-M-UN-AC
S/N: MI00103
Tag:
100-240V AC, 50/60 Hz, 60 VA
Made by VAISALA Oyj, Vantaa, Finland
www.vaisala.com
IP67
Type 4x
®
C
US
Model: Multichannel User Interface
Product code: MI-M-UN-DC
S/N: MI00103
Tag:
24V DC, 60 W
Made by VAISALA Oyj, Vantaa, Finland
www.vaisala.com
IP67
Type 4x
®
C
US
100-240 V AC
24 V DC
Figure 8.1
MI nameplates
© Copyright Vaisala 2019. All rights reserved.
MI specifications
Display:
xga 1024 x 768 pixels, touch screen
Power:
AC input 100-240 VAC, ±10%, 50/60 Hz, 60 VA, optional
24 VDC
Overvoltage category:
OVC II
Refractometer connectivity:
1-4 refractometers can be connected to the MI.
Refractometers independent of each other: own
parameter sets and usable in different applications.
Ethernet connection:
10/100 Mbit/s, data acquisition over UDP/IP Protocol
Modules:
1-8 modules for added connectivity
Apps:
1-4 apps
Protection class:
Enclosure IP66/67. Type 4X.
Installation:
Indoor/outdoor, max. 50°C (122°F), min. 0°C (32°F),
must be protected from rain and direct sunlight.During
installation highest relative humidity of ambient air 80%.
Highest altitude 2000 m above sea level.
Weight:
12.9 kg (28.5 lbs)
Regulatory compliance:
Safety: IEC 61010-1:2010/EN 61010-1:2010, CSA
certified to CSA-C.22.2 No. 61010.1-12, UL 61010-1:2012
Electromagnetic compatibility: IEC 61326-1:2012/EN
61326-1:2013.
Optional modules
Relay Module:
2 signal relays 150V/3A
mA Output Module:
2 x 4-10mA, max. load 1000 Ohm, galvanic isolation
1000VDC
Regulatory compliance
Safety: IEC 61010-1:2010/EN 61010-1:2010, CSA
certified to CSA-C.22.2 No. 61010.1-12, UL 61010-1:2012
Electromagnetic compatibility: IEC 61326-1:2012/EN
61326-1:2013
MI instruction manual
Document/Revision No. IM-EN-MI/2.00
Effective: September 2, 2019
8.1 Model code
MULTICHANNEL USER INTERFACE
Model and description
Model
MI = Multichannel user interface, stainless steel enclosure with wall mount.
Connectivity for up to 4 refractometers with Ethernet output, included
1 x mA output module card (2 x 4–20 mA), 1 x Relay module card (2 relays,
250 V/3A)
MI
Cable connection for power
-U = ½inch NPT type conduit hubs
-U
-M = M20x1.5 metric cable glands
-M
Electrical classification
-UN = Unclassified area, General Purpose, Ordinary location
-UN
Power supply
-AC = Power supply 100–240 VAC 50/60 Hz
-AC
-DC = Power supply 24 V DC
-DC
Module cards for MI (0-8 modules selectable in total, one or several)
-MO1 = 1 x mA output module card, for 2 x mA output (included in standard
delivery)
-MO1
-MO2 = 2 x mA output module card, for 4 x mA output
-MO2
-MO3 = 3 x mA output module card, for 6 x mA output
-MO3
-MO4 = 4 x mA output module card, for 8 x mA output
-MO4
-RM1 = 1 x Relay module card, for 2 x relay output (included in standard
delivery)
-RM1
-RM2 = 2 x Relay module card, for 4 x relay output
-RM2
-RM3 = 3 x Relay module card, for 6 x relay output
-RM3
-GO1 = 1 x General I/O module card, 4 x digital input, 4 x digital output
-GO1
-GO2 = 2 x General I/O module card, 8 x digital input, 8 x digital output
-GO2
A EU declaration of conformity
77
© Copyright Vaisala 2019. All rights reserved.
A EU declaration of conformity
2019-09-01B/JAMO
1 (1)
Vaisala Oyj | PO Box 26, FI-00421 Helsinki, Finland
Phone +358 9 894 91 | Fax +358 9 8949 2227
Email [email protected] | www.vaisala.com
Domicile Vantaa, Finland | VAT FI01244162 | Business ID 0124416-2
EU DECLARATION OF CONFORMITY
Manufacturer:
Vaisala Oyj
Mail address:
P.O. Box 26, FI-00421 Helsinki, Finland
Street Address:
Vanha Nurmijärventie 21, Vantaa, Finland
This declaration of conformity is issued under the sole responsibility of the manufacturer.
Object of the declaration:
K-Patents Multichannel User Interface MI
The object of the declaration described above is in conformity with Directives:
RoHS Directive (2011/65/EU)
EMC Directive (2014/30/EU)
Low Voltage Directive (2014/35/EU)
The conformity is declared using the following standards:
EN 50581:2012 Technical documentation for the assessment of electrical and
electronic products with respect to the restriction of hazardous substances
EN 61000-3-2:2006 + A1:2009 + A2:2009 Electromagnetic compatibility (EMC)
- Part 3-2: Limits - Limits for harmonic current emissions
EN 61000-3-3:2013 Electromagnetic compatibility (EMC) - Part 3-3: Limits -
Limitation of voltage changes, voltage fluctuations and flicker in public low-voltage
supply systems
EN 61010-1:2010 Safety requirements for electrical equipment for measurement,
control and laboratory use – Part 1: General requirements
EN 61326-1:2013 Electrical equipment for measurement, control and laboratory
use – EMC requirements – intended for use in industrial locations
Signed for and on behalf of Vaisala Oyj, in Vantaa, on 1st September 2019
______________________________________
Jukka Lyömiö
Standards and Approvals Manager
B Software licence
79
© Copyright Vaisala 2019. All rights reserved.
B Software licence
This software includes parts of software libraries included under the following licence
notices.
Copyright ©2012, Aaron Gifford All rights reserved.
Redistribution and use in source and binary forms, with or without modification, are
permitted provided that the following conditions are met:
•
Redistributions of source code must retain the above copyright notice, this list of
conditions and the following disclaimer.
•
Redistributions in binary form must reproduce the above copyright notice, this list
of conditions and the following disclaimer in the documentation and/or other ma-
terials provided with the distribution.
•
Neither the name of the nor the names of its contributors may be
used to endorse or promote products derived from this software without specific
prior written permission.
THIS SOFTWARE IS PROVIDED BY THE COPYRIGHT HOLDERS AND CONTRIBUTORS "AS
IS" AND ANY EXPRESS OR IMPLIED WARRANTIES, INCLUDING, BUT NOT LIMITED TO,
THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR
PURPOSE ARE DISCLAIMED. IN NO EVENT SHALL AARON GIFFORD BE LIABLE FOR ANY
DIRECT, INDIRECT, INCIDENTAL, SPECIAL, EXEMPLARY, OR CONSEQUENTIAL DAMAGES
(INCLUDING, BUT NOT LIMITED TO, PROCUREMENT OF SUBSTITUTE GOODS OR SER-
VICES; LOSS OF USE, DATA, OR PROFITS; OR BUSINESS INTERRUPTION) HOWEVER
CAUSED AND ON ANY THEORY OF LIABILITY, WHETHER IN CONTRACT, STRICT LIABILITY,
OR TORT (INCLUDING NEGLIGENCE OR OTHERWISE) ARISING IN ANY WAY OUT OF THE
USE OF THIS SOFTWARE, EVEN IF ADVISED OF THE POSSIBILITY OF SUCH DAMAGE.
Command selection tree for Refractometer application
81
© Copyright Vaisala 2019. All rights reserved.
Command selection tree for Refractometer appli-
cation
Main screen
REFR. APP
DIAGNOSTICS
CALIBRATION
TREND
LOG
PRISM WASH
VERIFICATION
FIELD SAMPLE
INPUT
OUTPUT
IDENTIFICATION
SUPPORT
SELECT DISPLAY
MODES
PRIMARY UNIT
PRIMARY DECIMALS
SECONDARY UNIT
CONCENTRATION
TEMPERATURE
BOTH
CELSIUS
FAHRENHEIT
CHEMICAL
FIELD
TAG
SENSOR
SOURCES
PRISM WASH
SUMMARY
CLEAR
Command selection tree for settings
83
© Copyright Vaisala 2019. All rights reserved.
Command selection tree for settings
Main screen
SETTINGS
GENERAL
DISPLAY SETUP
QUADRANT
ONE ON TOP
DUAL VERTICAL
ONE MAXIMIZED
SCREEN BRIGHTNESS
LANGUAGE
HIGH CONTRAST
MODE
DATE & TIME
PROTECT WITH
PASSWORD
BACKUP/RESTORE
FACTORY RESET
UPDATE
SYSTEM TAG
NETWORK
CURRENT IP ADDRESS
IP ADDRESS MODE
SET STATIC IP ADDRESS
SET STATIC IP ADDRESS
APPLICATIONS
APP SLOT X (1-8)
MODULES
REMOVE
ADD
MODULE SLOT X (1-8)
USERS
ADD USER
SET NAME
SET PASSWORD
LOG
CLEAR
START BOOTP SERVER
SET FIRST SENSOR
IP ADDRESS
CURRENT IP ADDRESS
IP ADDRESS MODE
Refractometer
Ethernet
Master user login required
(IP address can be viewed without login)
SET SUBNET MASK
SET SUBNET MASK
SET DEFAULT GATEWAY
STATIC
DYNAMIC
AUTO
YES
NO
AUTO
STATIC
DYNAMIC
AUTO