Document/Revision No. IM-EN-MI/2.00

Effective: September 2, 2019 PUBLISHED BY Vaisala Oyj Vanha Nurmijärventie 21, FI-01670 Vantaa, Finland P.O. Box 26, FI-00421 Helsinki, Finland +358 9 8949 1 Visit our Internet pages at www.vaisala.com. General safety considerations Always power off the Multichannel user interface MI before accessing the modules behind the display. Vaisala recommends the installation of an external power switch. Warning! If this device is used against manufacturer’s instructions, user safety may be compromised. Warranty For standard warranty terms and conditions, see www.vaisala.com/warranty. Please observe that any such warranty may not be valid in case of damage due to nor- mal wear and tear, exceptional operating conditions, negligent handling or installation, or unauthorized modifications. Please see the applicable supply contract or Conditions of Sale for details of the warranty for each product. Technical support Contact Vaisala technical support at [email protected]. Provide at least the fol- lowing supporting information: • Product name, model, and serial number • Name and location of the installation site • Name and contact information of a technical person who can provide further infor- mation of the problem For more information, see www.vaisala.com/support.

© Copyright Vaisala 2019. All rights reserved.

Disposal When wishing to dispose of an obsolete instrument or any parts of an in- strument, please observe local and national regulations and requirements for the disposal of electrical and electronic equipment. Symbols and terms used in this manual: This indicates a warning. It provides safety precaution information needed to avoid injury while operating the refractometer system. This indicates that something is important for the operation of the refractome- ter system. Note. Notes contain additional information and hints. This product manual is delivered to the end user with a Vaisala K-PATENTS® product. Information in this manual is subject to change without notice. When the manual is changed, a revised copy is published at http://www.kpatents.com/

Table of contents

1 Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 2 Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 3 Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 3.1 Connecting the power cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 3.1.1 Power terminals for AC power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 3.1.2 Power terminals for DC power . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 3.1.3 External mains switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 3.2 Refractometer cable connections . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 3.3 Ethernet cable connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 3.4 Module cards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 3.5 Connections for external control of Recipe app . . . . . . . . . . . . . . . . 13 4 Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 4.1 Main display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 4.2 Activating and removing apps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19 4.3 Managing users and passwords . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23 4.4 System log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 4.5 Ethernet connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 4.5.1 MI ethernet connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 4.5.2 Refractometer connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27 5 Apps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28 5.1 Shared functionality . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28 5.1.1 Application trend . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28 5.1.2 Application log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30 5.2 Refractometer app . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32 5.2.1 Diagnostics display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32 5.2.2 Viewing and adjusting parameters . . . . . . . . . . . . . . . . . . . . . . . . . 34 5.2.3 Refractometer verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37 5.2.4 Collecting field sample points . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39 5.3 Recipe app . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41 5.3.1 Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41 5.3.2 Activating Recipe app . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41

Creating a new recipe

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41 5.3.4 Modifying a recipe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41 5.3.5 Selecting recipe switching mode . . . . . . . . . . . . . . . . . . . . . . . . . . 42 5.3.5.1 Switching recipe manually . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42 5.3.5.2 Recipe identification numbers for remote control . . . . . . . . . . . 43 5.3.5.3 Switching recipe via General I/O module inputs . . . . . . . . . . . . 44 5.3.5.4 Switching recipe via Modbus/TCP . . . . . . . . . . . . . . . . . . . . . . . 45 5.3.5.5 Switching recipe via Ethernet/IP . . . . . . . . . . . . . . . . . . . . . . . . 45 5.3.6 Using Recipe app to make unit conversions . . . . . . . . . . . . . . . . . 46 5.3.6.1 Creating a recipe for unit conversion . . . . . . . . . . . . . . . . . . . . . 46 5.3.6.2 Displaying several unit conversions simultaneously . . . . . . . . . 46 5.3.7 Configuring mA output for Recipe app . . . . . . . . . . . . . . . . . . . . . 47 6 Module cards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49 6.1 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49 6.2 Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51 6.3 Relay module card PR-50023 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53 6.3.1 Configuring relays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54 6.4 mA output module PR-50021 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56 6.4.1 Configuring mA outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57 6.5 General I/O module PR-50024 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58 6.5.1 Connecting inputs/outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61 6.5.2 Configuring the general I/O module card . . . . . . . . . . . . . . . . . . . 63 6.5.2.1 Using the general I/O module to trigger prism wash . . . . . . . . 64 7 Prism wash . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66 7.1 Configuring prism wash . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66 7.1.1 Wash cycle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66 7.1.2 Setting prism wash parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . 70 8 Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72 8.1 Model code . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74 8.2 Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75 A EU declaration of conformity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77 B Software licence . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79

© Copyright Vaisala 2019. All rights reserved.

Command selection tree for Refractometer application . . . . . . . . . . . . . . . . . . . . 81 Command selection tree for settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83

© Copyright Vaisala 2019. All rights reserved.

1 Description The Multichannel user interface MI is an industrial computer designed to add function- ality to the PR-43 refractometers. The MI sends and receives data and it can also be used to control the process. The Multichannel user interface MI connects with up to 4 Vaisala K-PATENTS® PR-43 refractometers. With module cards, it can also be used to add external inputs and outputs (e.g. relays and mA outputs) to your refractometer system. The MI is operated by the touch-screen display. The stainless steel enclosure gives a good environmental protection and allows also outdoor installations, especially with the included rain cover. The outer door is easy to open by one hand and the handle can be modified for padlock to prevent unauthorized access. The module cards can be reached by opening the inner door. The MI has IP66 rating when the outer door is open if the inner door (display panel) is fixed. With outer door closed the rating is IP67. Figure 1.1 Multichannel user interface MI

MI instruction manual

Document/Revision No. IM-EN-MI/2.00 Effective: September 2, 2019 2 Mounting The Multichannel user interface MI can be mounted on a wall with wall mounting or on a table with a table stand. The mounting location should be easily accessible. If mounted outdoors, the MI has to be protected from rain and direct sunlight. Display on eye-level External mains switch 100-240VAC 4-20 mA Ethernet 1-4 refractometers Figure 2.1 Mounting guide

MI instruction manual

Document/Revision No. IM-EN-MI/2.00 Effective: September 2, 2019 for refractometers Figure 2.3 Enclosure dimensions

© Copyright Vaisala 2019. All rights reserved.

1 1. Bracket screwed to wall by 4 screws. Screw hole dimensions 325x254mm 325 mm (12.8 inches) 254mm (10 inches) 1 1 1 Figure 2.4 Mounting the wall bracket 1. Four mounting screws loosely attached 2. Top screws inserted into top bracket verticalmounting slots 3. Lower screws slide into bottom bracket horizontal mounting slots 4. Tighten mounting screws x4 width a 19 mm ring spanner 1 2 3 4 Figure 2.5 Mounting the MI onto the wall bracket Note: Be careful when mounting, the MI weighs 13 kilos (28.5 lbs).

MI instruction manual

Document/Revision No. IM-EN-MI/2.00 Effective: September 2, 2019 Important: Do not drill mounting holes in the enclosure as that will affect the protec- tion class of the enclosure and damage the electronics. The MI is shipped with a rain cover. Vaisala recommends that it is installed even indoors to protect the top of the door from dust and dirt. 1. Screw out top attaching screws to allow clearance for rain cover to slot in place 1 2. Retighten screws 2 Figure 2.6 Mounting the rain cover The table stand is very simple to mount. Figure 2.7 Table stand

© Copyright Vaisala 2019. All rights reserved.

The MI door handle can be modified to make space for a padlock to prevent unautho- rized access. Remove handle center cover and remove segment indicated by break-line on the back of cover Replace center cover once indicated segment has been removed, making sure the opening aligns with the channel for the padlock shackle Figure 2.8 Locking the MI

MI instruction manual

Document/Revision No. IM-EN-MI/2.00 Effective: September 2, 2019 3 Connections Refractometer + MI max. 200m (656ft) PR-8441 Ethernet cable for interfaces M12-4 pin, D-code, M+RS45 Ethernet PR-8440 Ethernet cable for interfaces M12-4 pin, D-code, M+M Ethernet mA cable M M M RS45 M F PR-8430 Platform 4 cable M12-8 pin, A-code, F+M Ethernet, mA, 24V PR-43 REFRACTOMETER M-12-8 pin, D code, M non-Ex Multi Channel User Interface, MI 1 x M12-4 pin, D-code, F (External Ethernet) 4 x M12-8 pin, A-code, F (System) max. 90m (295ft) max. 90m (295ft) max. 90m (295ft) 100-240VAC or 24VDC Figure 3.1 Multichannel user interface connections

© Copyright Vaisala 2019. All rights reserved.

3.1 Connecting the power cable Warning! Make sure there is no power in your cable. Then open the enclosure and the inner door. You need a TX25 screwdriver to open the inner door retaining screws. 1 2 3 1.   Push door gently 2.   Lift handle 45 deg up 3.   Open the door Figure 3.2 Opening the enclosure Remove inner door retaining screws and open inner door Figure 3.3 Opening the inner door Unscrew one of the cable glands and push the power cable through. Connect the power cable to the power terminals at lower right of the motherboard (see below). Screw on the cable gland. Close the inner door and turn power on. 3.1.1 Power terminals for AC power The primary AC power is connected to a separate terminal strip 1/2/3 in the lower right- hand corner of the motherboard. The three terminals are marked 1.L, 2.N and 3 (protective earth). The power terminal 3/ is directly connected to the exposed metal parts of the MI. Wiring to the terminals shall be 1.5mm2 minimum. The protective fuse in the building system shall comply with the local requirements.

MI instruction manual

Document/Revision No. IM-EN-MI/2.00 Effective: September 2, 2019 1.L 2.N 3 Brown Blue Yellow-green Figure 3.4 AC power terminals on the motherboard 3.1.2 Power terminals for DC power The DC power is connected to a terminal stripe in the lower right-hand corner of the motherboard. The three terminals are marked 1+, 2- and 3 (protective earth). The power terminal is directly connected to the exposed metal parts of the MI. The 24V DC power to this terminal stripe shall be supplied from a secondary circuit which is double or reinforced insulated from the mains supply within the limits for a limited- energy circuit (maximum 200 VA/U) according to the IEC 61010-1. 3.1.3 External mains switch External mains switch Figure 3.5 Installing external main switch

© Copyright Vaisala 2019. All rights reserved.

Important: The MI does not have a built-in power switch. The system is always pow- ered when connected to a power source. Vaisala recommends mounting an external mains switch to control the MI’s power supply (Figure 3.5). The external power switch should be easy to access. It is also recommended to use a 10A/230V circuit breaker for power supply. MULTI CHANNEL USER INTERFACE MI INSTRUMENTS M12 CONNECTORS L N MAIN POWER SWITCH PR-10900 L N + - mA Output 1 4-20mA POWER SUPPLY 100-240VAC/50-60Hz 4158 - + - + OPTIONAL 24V POWER SUPPLY FOR MI 24VDC ETHERNET M12-4pin, D-code, M ETHERNET M12-4pin, D-code, M PR-8440 Ethernet cable for interfaces MI has 8 module slots. Any of module cards can be installed into any free slot. There could be multiple module cards in slots depending on external I/O needs. PR-50021 mA output module PR-50023 Relay output module PR-50021 mA output module PR-50023 Relay output module PR-8430 Platform 4 cable SLOT 2 SLOT 4 SLOT 6 SLOT 8 SLOT 1 SLOT 3 SLOT 5 SLOT 7 1...4 x L N SOLENOID VALVE FOR PNEUMATICALLY OPERATED SHUT-OFF VALVE ASSEMBLY: PR-3340-24/110/240 1...4 x 9 N L SAFETY SWITCH PR-7060 8 7 6 5 4 3 2 1 N L L L POWER SUPPLY 100-240VAC/50-60Hz or 24VDC L N 1...4 x 1...4 x Grounding screw PR-43-...-UN/AX/IA refractometer (For Intrinsic refractometer see further wiring drawing 4118) 1...4 x Figure 3.6 Multichannel user interface wiring

MI instruction manual

Document/Revision No. IM-EN-MI/2.00 Effective: September 2, 2019 3.2 Refractometer cable connections The refractometers are connected to the MI with a M12 male-female 8 pins A-coded cable PR-8430, maximum length 90 m. The refractometer connectors are located un- derneath the MI enclosure as a group of four. Refractometer connections can be made while the MI is powered. A refractometer can be connected to any of the four connec- tors. 3.3 Ethernet cable connection The leftmost connector underneath the MI is for ethernet output. It is for a M12 male 4-pins D-coded cable. The cable can be connected while the MI is powered. If you wish to connect the MI with e.g. a laptop, you can purchase Interconnecting cable PR-8330 with a M12-4 pin connector in one end and a RJ-45 connector in the other end. 3.4 Module cards For the installation of the module cards, see Chapter 6 “Module cards”.

© Copyright Vaisala 2019. All rights reserved.

3.5 Connections for external control of Recipe app 1 GPIO mA-output 6 2 3 4 24VDC or 240VAC 5 Item Component Part number Supplied by QTY 1 PR-43 Refractometer Vaisala 1-4 2 Platform 4 cable PR-8430 Vaisala 1-4 3 Multichannel User Interface MI Vaisala 1 4 Main power switch PR-10900 Vaisala 1 5 mA-output cable Customer 1 6 GPIO selection cable Customer 1 Figure 3.7 Recipe app controlled via GPIO (SYS-605)

MI instruction manual

Document/Revision No. IM-EN-MI/2.00 Effective: September 2, 2019 1 24VDC or 240VAC mA-output 5 2 3 6 4 Ethernet Item Component Part number Supplied by QTY 1 PR-43 Refractometer Vaisala 1-4 2 Platform 4 cable PR-8430 Vaisala 1-4 3 Multichannel User Interface MI Vaisala 1 4 Main power switch PR-10900 Vaisala 1 5 mA-output cable Customer 1 6 Ethernet cable PR-8330 Vaisala 1 Figure 3.8 Recipe app controlled via Modbus (SYS-606)

© Copyright Vaisala 2019. All rights reserved.

4 Use Power up the Multichannel user interface MI. A startup display with a progress bar is shown for approximately one minute, then the Main display appears. The MI is operated by tapping the touch screen. To get access to the MI touch screen, open the enclosure door. Note: If you find that the touch screen is difficult to use with fingers, try tapping it with e.g. the blunt end of a pen or use a fingernail instead of the whole fingertip. The screen works even when you are wearing gloves. 4.1 Main display The main view can show up to four application windows at the same time. You can access measurement data and system settings via the Main display. Tap anywhere in the application window to access measurement Tap for second main view screen Tap to access system settings Figure 4.1 Accessing measurement and system settings via Main display The Main display setup is controlled via Settings > General. Display mode and display language are chosen in drop-down menus, tap the little triangle to see the whole menu. If you only have one application (usually a single refractometer), One maximized will make the most of the display. Dual vertical allows you to see two application, Quadrant and One on top shows up to of four applications at the same time. Note: If you have three applications and choose One maximized, you’ll only see one application, but the other two will continue to run in the background.

MI instruction manual

Document/Revision No. IM-EN-MI/2.00 Effective: September 2, 2019 Display mode Display language Figure 4.2 Selecting display mode and display language Figure 4.3 Setting "One maximized"

© Copyright Vaisala 2019. All rights reserved.

Figure 4.4 Setting "Dual horizontal" Figure 4.5 Setting "Quadrant"

MI instruction manual

Document/Revision No. IM-EN-MI/2.00 Effective: September 2, 2019 Figure 4.6 Setting "One on top" Note: If the MI is left with a different display, e.g. Calibration, it will automatically go back to the Main display after a timeout. Note: You can adjust Refractometer app to show concentration only, temperature only or both concentration and temperature. To change this setting, tap on the application window (see Figure 4.1) and choose Calibration (You may need to enter your password to get access to Calibration display). Then select the preferable display mode.

© Copyright Vaisala 2019. All rights reserved.

Figure 4.7 Calibration display 4.2 Activating and removing apps The Main display may be empty even when one or several refractometers are con- nected to the MI. This means that one or several applications must be activated. Go to Settings > Applications, tap on the empty application slot and select the app type and source for that slot. After choosing settings for the app, tap Done to exit the setup.

MI instruction manual

Document/Revision No. IM-EN-MI/2.00 Effective: September 2, 2019 Figure 4.8 Empty main display in quadrant mode Figure 4.9 Selecting an empty application slot to activate the app

© Copyright Vaisala 2019. All rights reserved.

Figure 4.10 Selecting a source to activate the app, here application type Ref is already selected The empty application slot may contain an earlier app configuration. In that case a dia- log will open during app activation and the earlier app configuration can be re-enabled. Figure 4.11 Selecting an empty application slot that contains old Refractometer app configuration

MI instruction manual

Document/Revision No. IM-EN-MI/2.00 Effective: September 2, 2019 Figure 4.12 The earlier app configuration can be re-enabled If necessary the app can also be removed. Go to Settings > Applications and tap Remove on the application slot. Accept the removal by tapping Ok. App’s configuration will be saved until another kind of application is added to the same slot. Figure 4.13 If necessary, an app can also be removed

© Copyright Vaisala 2019. All rights reserved.

4.3 Managing users and passwords Some functions and settings, e.g. calibration, require a user to log in. Most functions in the settings, including user management, are only available to the master user. The master user password (pin code) is printed on the delivery data sheet shipped with the MI. Note: Controlling and switching recipes in Recipe app (see section 5.3) is not protected by user passwords. Go to Settings > Users to manage users. Vaisala recommends that a user name and a password is set separately for each user who needs to have frequent access to the restricted functions. Figure 4.14 The Users display Tap + Add User to add users. Tap Set name to get access to the keyboard, then type the username and tap OK.

MI instruction manual

Document/Revision No. IM-EN-MI/2.00 Effective: September 2, 2019 Figure 4.15 Writing a username Note: If you have entered a username, you must enter a password, too, you cannot leave the password field empty. Figure 4.16 New user added to the list

© Copyright Vaisala 2019. All rights reserved.

To remove a user, go to Users and tap the name of the user you want to remove, then tap Remove user. Note: The MI can store up to 38 users. 4.4 System log To access a system log, go to Settings > Log. The system log lists status changes of the system like booting up, adding an app and changing module card settings. The system log is also accessed via http connection. Use the IP address found in the MI Settings > Network menu, under Ethernet section. On the MI web page select files > system > log. Figure 4.17 System log 4.5 Ethernet connections 4.5.1 MI ethernet connection The MI has a built-in web server with a home page. The MI home page allows you to browse files or upload an update or a license file. To access the MI web page use IP address found under the title Ethernet in the Settings > Network menu.

MI instruction manual

Document/Revision No. IM-EN-MI/2.00 Effective: September 2, 2019 The default IP address mode setting for the MI is Auto. In this mode the MI gets the IP address from an external DHCP server. If that fails the IP address will fall back to a static IP setting. Alternatively IP address mode: Static configures the MI to use a static IP address only. After choosing settings for the network, tap Apply to save all settings. After factory reset the default IP address of the MI is 192.168.1.11. Figure 4.18 To access the MI web page use IP address found in the Settings > Network menu

© Copyright Vaisala 2019. All rights reserved.

Figure 4.19 Steps to configure a static IP address for an MI 4.5.2 Refractometer connection Refractometer’s connection to the MI is configured under the title Instrument in the Settings > Network menu. By default the refractometer’s IP address is configured via BOOTP server running on the MI. To reconnect a refractometer you can just replug its cable to the MI – the IP address is configured automatically. Refractometer connections can be made while the MI is powered.

MI instruction manual

Document/Revision No. IM-EN-MI/2.00 Effective: September 2, 2019 5 Apps Apps are small programs that give you different types of measurement data and func- tionality. Vaisala is constantly developing new apps for the measurement, but some are installed by default. 5.1 Shared functionality While different applications work differently, there are some functions that are similar to all applications. 5.1.1 Application trend An MI app can display graphically one or two of its variables over selectable time period. To access an app trend, tap application window on the main view and tap Trend. Save the trend by tapping Save. Saved trend can be accessed via http connection. Use the IP address found in the MI Settings > Network menu, under Ethernet section. On the MI web page select files > trend > yearMonthDate-time_appType_slotNumber.csv. Note: Only apps running on the first four app slots can display the graphical trend. MI supports four apps on one main view screen and there are two main view screens that can be switched, see figure 4.1.

© Copyright Vaisala 2019. All rights reserved.

Figure 5.1 MI app trend display Figure 5.2 Saving app trend

MI instruction manual

Document/Revision No. IM-EN-MI/2.00 Effective: September 2, 2019 Figure 5.3 To access saved app trends, select files > trend on the MI web page Figure 5.4 Select the trend to be viewed/downloaded 5.1.2 Application log The app log lists status changes and possible error messages of apps in an app slot. To access an app log, tap application window on the main view and tap Log. The app log is addressed by the app slot number and lists status changes and possible error messages of apps in that slot. App logs are also accessed via http connection. Use the IP address found in the MI Settings > Network menu, under Ethernet section. On the MI web page select files > app > slotNumber > log.

© Copyright Vaisala 2019. All rights reserved.

Figure 5.5 App log Figure 5.6 On the MI web page select files > app

MI instruction manual

Document/Revision No. IM-EN-MI/2.00 Effective: September 2, 2019 Figure 5.7 On the MI web page select files > app > slot number Figure 5.8 On the MI web page select files > app > slot number > log 5.2 Refractometer app Refractometer app is included in your delivery. It gives you basic information about the process, e.g. concentration and temperature. 5.2.1 Diagnostics display Tap the application window to access the Diagnostics display. In the top part of the window you see the optical and slope curves. By default both optical image and slope are shown, but you can tap on the texts OPTICAL and SLOPE to turn these on and off.

© Copyright Vaisala 2019. All rights reserved.

Figure 5.9 Diagnostics display The tabs in the lower part of the display contain detailed information about the mea- surement. Tap the name of the tab to see the contents. Input tab shows the values as sent by the refractometer. Output tab shows the results of the calculations by MI. Identification tab contains various identification information and e.g. software version information. Support tab is a summary of the information in the other tabs and other information needed by technical support for troubleshooting. Figure 5.10 Input tab Figure 5.11 Output tab

MI instruction manual

Document/Revision No. IM-EN-MI/2.00 Effective: September 2, 2019 Figure 5.12 Identification tab Figure 5.13 Support tab 5.2.2 Viewing and adjusting parameters To view the parameters, go to the Calibration display and tap on Summary. Note that the summary extends to a second page, a small arrow on the top right will take you to the second (and third) page. To adjust the parameters, go to the Calibration display, then choose appropriate para- meter set, eg. Field. In the list, tap the parameter you want to change, type the new numbers with the number pad and approve the change by tapping DONE. Note: The parameter lists for Chemical and Field do not fit on the screen. Swipe down or use the scroll bar on the right to view the bottommost parameters.

© Copyright Vaisala 2019. All rights reserved.

Figure 5.14 Calibration display The settings in the upper part of the Calibration display are different types of lists that sets parameters for showing the measurement data, e.g. measurement unit. In the lower part of the window, the words are links to the given settings, e.g. by tapping on Field you get access to the field calibration parameters. If you want to see all the parameters, tap Summary.

MI instruction manual

Document/Revision No. IM-EN-MI/2.00 Effective: September 2, 2019 Figure 5.15 First page of calibration summary (tap the little arrow in top right to see page 2) Figure 5.16 Second page of calibration summary (tap the little arrow in top left to go back to page 1)

© Copyright Vaisala 2019. All rights reserved.

5.2.3 Refractometer verification An offline refractometer can be verified with standard refractive index liquids, available from Vaisala. Depending on the refractometer model, a sample holder may be needed. Go to the Diagnostics display (see Figure 5.9) and choose Verification. If you are not fa- miliar with the verification procedure, tap the question mark on the Verification display for additional instructions. Figure 5.17 Verification in process

MI instruction manual

Document/Revision No. IM-EN-MI/2.00 Effective: September 2, 2019 Figure 5.18 Finished verification Once you have finished the verification process, you can save the verification results as pdf. The pdf is stored in the MI memory and it can be downloaded over the eth- ernet connection. Each pdf report gets a unique name and thus you can have several verification pdfs in the MI at the same time. Figure 5.19 Pdf saved

© Copyright Vaisala 2019. All rights reserved.

5.2.4 Collecting field sample points You can use the field sample functionality to acquire data points for Vaisala K-PATENTS® field calibration service. Go to the Diagnostics display (see Figure 5.9) and tap Field sample. Tap Collect field point to start collecting field samples and then tap Save file to save the collected field points. These can be accessed via http connection. Use the IP address found in the MI Settings > Network menu, under Ethernet section. On the MI web page select files > fieldsamples > FieldSamples_SlotNumber. Figure 5.20 Start collecting field samples by tapping Collect field point

MI instruction manual

Document/Revision No. IM-EN-MI/2.00 Effective: September 2, 2019 Figure 5.21 Field sampling in process Figure 5.22 Save collected field points by tapping Save file

© Copyright Vaisala 2019. All rights reserved.

5.3 Recipe app Recipe app provides a convenient way to control and switch different chemical curves. Each Recipe app can provide up to 32 different recipes. Switching a recipe in Recipe app means switching the chemical curve that is used in calculations. Each process medium has its specific chemical curve and by switching it the same processing line can be used to manufacture a variety of products. 5.3.1 Requirements Using Recipe app requires a Refractometer app, which provides refractive index nD and temperature as a source data to the connected Recipe app. A licence is required to enable Recipe app. New Multichannel user interface MI:s are shipped with Recipe app licence installed. Old MI:s require software update to version 1.07 or newer and licence installation to enable Recipe app. Recipe app licences are granted free of charge to older MI:s. 5.3.2 Activating Recipe app To activate Recipe app go to Settings > Applications, tap on a free application slot and select Recipe. Select a correct source Refractometer app with Select sources. The avail- able Refractometer apps are listed based on their location in MI application slots. If no Refractometer app is listed in Select sources, then activate a Refractometer app Ref first. 5.3.3 Creating a new recipe MI is provided with a Brix chemical curve as an example recipe for Recipe apps. A new Recipe app has the example recipe activated by default. A new recipe is instantly created when the example recipe is renamed. To rename a recipe go to Recipe app > Calibration > Chemical and select Name. The new recipe is a copy of the old recipe with the new name. 5.3.4 Modifying a recipe A recipe can be modified e.g. by adjusting a field calibration parameter to offset the measurement result according to recipes requirements. To adjust the field parameters

MI instruction manual

Document/Revision No. IM-EN-MI/2.00 Effective: September 2, 2019 of an active recipe go to Recipe app > Calibration > Field. In the list, tap the parameter you want to change, type the new numbers with the number pad and approve the change by tapping DONE. Note: The parameter lists for Chemical and Field do not fully fit on the screen. Swipe down or use the scroll bar on the right to view the bottommost parameters. All recipes that are present in MI are visible and usable to all Recipe apps. If a specific recipe is modified in some Recipe app, the modification propagates also to the other Recipe app using the same recipe. 5.3.5 Selecting recipe switching mode Recipes in the Recipe app can be switched manually, via General I/O module inputs (Switch input) or via Modbus/TCP. To select the switching mode, tap the application window to access Diagnostics display. Then tap Remote control and then select the recipe switching mode. Figure 5.23 Selecting how to switch recipes in the Recipe app 5.3.5.1 Switching recipe manually To switch recipe manually, tap the application window to access Diagnostics display. Then tap Change recipe, select the recipe by tapping its name and accept by tapping OK.

© Copyright Vaisala 2019. All rights reserved.

Figure 5.24 Switching recipe manually 5.3.5.2 Recipe identification numbers for remote control A recipe is activated remotely, when a remote signal corresponds to a recipe identifica- tion number. To set the identification number go to Recipe App > Calibration > Remote control, tap a recipe name and select the identification number for that recipe. The identification number can have a value between 0 - 31. Different Recipe apps have independent values for their recipe identification numbers and they need to be set in each Recipe app that uses remote control functionality.

MI instruction manual

Document/Revision No. IM-EN-MI/2.00 Effective: September 2, 2019 Figure 5.25 Recipe identification numbers for remote control Figure 5.26 Setting an identification number for a recipe 5.3.5.3 Switching recipe via General I/O module inputs A recipe is activated via General I/O module inputs, when the inputs form a binary number that corresponds to a recipe identification number. The recipe switching mode Switch input must be selected, see figure 5.23. To configure General I/O module, go

© Copyright Vaisala 2019. All rights reserved.

to Settings > Modules, tap the appropriate General I/O module to select it and then View module to access module configuration. Set the wanted inputs to control the Recipe app, which has a number according to the appslot it is running on. Set the purpose of inputs to implement a binary number with different multipliers for different bits. Four digital inputs associated to a single Recipe app enable 24 different binary number combinations that can address 16 different recipes. After choosing the settings for inputs, tap Done to save changes and exit the setup. 5.3.5.4 Switching recipe via Modbus/TCP In Modbus mode the recipe is switched by writing the recipe identification number to a Modbus register. Each appslot has a specific Modbus register to switch recipe for Recipe app running on that appslot. The register address is of the form 1n200, where n is the appslot number. The register data type is a four-byte signed integer. The recipe switching mode Modbus must be selected, see figure 5.23. 5.3.5.5 Switching recipe via Ethernet/IP In Ethernet/IP mode the recipe is switched by writing the recipe identification number to an Ethernet/IP object InputRecipeID, which is a four-byte integer. Ethernet/IP con- nection parameters depend on what apps are running on MI. When Refractometer app is on the first appslot, Recipe app is on the second appslot and all other appslots are empty, the connection parameters are: • Port: 0xAF12 (44818) • O->T: – instance number: 119 – data size: 4 – real time format: modeless – packet rate: 1000 ms • T->O: – instance number: 109 – data size: 120 – real time format: modeless – packet rate: 1000 ms – connection type: point to point

MI instruction manual

Document/Revision No. IM-EN-MI/2.00 Effective: September 2, 2019 Also, an Electronic Data Sheet (EDS) file is available for easy setup of the Ethernet/IP connection. The recipe switching mode Ethernet/IP must be selected, see figure 5.23. 5.3.6 Using Recipe app to make unit conversions Recipe apps can perform unit conversions so that measurement results can be dis- played in two or more units simultaneously. Unit conversions enable familiar readings like numeric values and engineering units, and comparing the measurement values to laboratory results gets easier. 5.3.6.1 Creating a recipe for unit conversion A properly configured recipe is needed for each unit conversion. Create a new recipe for each new unit to be displayed. For clarity it is recommended to name the new recipe according to the unit. To implement the unit conversion, adjust a field calibration parameter F 10 by setting its value to (M – 1) where M is the conversion multiplier. See section 5.3.3 “Creating a new recipe” and section 5.3.4 “Modifying a recipe”. 5.3.6.2 Displaying several unit conversions simultaneously To display several unit conversions simultaneously, run each unit conversion recipe in its own Recipe app. If necessary, activate new Recipe apps, see section 5.3.2 “Activating Recipe app”. Connect the new Recipe apps to the same source Refractometer app to perform unit conversions of a single measurement. To change the primary unit marking on the Recipe app display, go to Recipe app > Calibration and tap the little triangle of drop-down menu Primary unit.

© Copyright Vaisala 2019. All rights reserved.

Figure 5.27 Changing the primary unit marking on the Recipe app display 5.3.7 Configuring mA output for Recipe app If no mA output is configured for Recipe app, app status in Recipe app main view will blink between current status and message ”mA output not configured”. This will pre- vent false or omitted configuration of mA output. To configure mA output module, go to Settings > Modules, tap the appropriate mA output module to select it and then View module to access module configuration. Set the wanted output(s) for Recipe app, which has a number according to the appslot it is running on. Set the purpose of output(s), see section 6.4.1. After choosing the settings for outputs, tap Done to save changes and exit the setup.

MI instruction manual

Document/Revision No. IM-EN-MI/2.00 Effective: September 2, 2019 Figure 5.28 Blinking app status message when no mA output is configured Figure 5.29 Configuring mA output for Recipe app

6 Module cards

49 © Copyright Vaisala 2019. All rights reserved. 6 Module cards A module card adds functions to the MI, e.g. a relay module card contains two relays that can be used as wash relays or alarm relays. There are eight module card slots inside an MI. Each Multichannel User Interface MI is shipped with one relay module card with 2 independent relays and one mA output module card with two independent 4-20 mA outputs as a standard. These are already installed in the MI. In addition to the relay module card and the mA output module card, there is a general input/output module card with 8 digital inputs/outputs. These module cards have the capability to be used in any combination as well as added and removed as needed. 6.1 Installation To install a module card, turn off MI power first. Then open the enclosure, remove the display screws and fold the display to the side (see Figure 6.1). Unscrew and remove the module card cover to get access to the card slots. Insert the module card in an avail- able card slot. To make adding cards easier, start from the left. Vaisala recommends for safety reasons that relay modules are mounted in slots 2, 4, 6 and 8 (connector down- wards). This way the relay power wiring is as short as possible. Other module cards may be mounted in any slot. Important: A relay module card may have an external power source. Make sure that also this external power source connected to a relay module is OFF before installing or removing module cards.

MI instruction manual

Document/Revision No. IM-EN-MI/2.00 Effective: September 2, 2019 Remove inner door screws (TX25 screwdriver needed) Open the inner door and move it to the side Open the four screws holding the module card holder and remove the holder Insert the module card(s) and replace the module card holder Figure 6.1 Installing module cards To make the connections, first remove the removable connector from the module card. Bring the module connection cable to the MI and put it through the cable gland/conduit hub (1). Then connect wires to the screw terminals on the removable connector (2), see the module-specific sections for further information. Finally reconnect the removable connector to the module card (3).

6 Module cards

51 © Copyright Vaisala 2019. All rights reserved. 1 2 3 4 5 6 7 8 1 2 3 4 5 6 7 8 1 2 3 4 5 6 7 8 1 2 3 4 5 6 7 8 1 2 3 3 MI-010 Figure 6.2 Connecting module connection cable to module card 6.2 Configuration The MI automatically finds the installed module card(s). To configure a module card, go to Settings > Modules, tap the appropriate module card to select it and then VIEW MODULE to access module configuration. The selected module card is marked with a border. Note: There are two types of EMPTY slots: grey slots do not have module card, striped slots are turned off (see Figures 6.3 and 6.4 on next page) and there may or may not be a module card in that slot. If there is a module card, it will be automatically recognized when the module slot is turned on by selecting the module and then tapping the TURN ON command.

MI instruction manual

Document/Revision No. IM-EN-MI/2.00 Effective: September 2, 2019 Figure 6.3 Selecting a module card and viewing module card information Figure 6.4 Selecting a turned-off module card slot (slot 1 in this case) and turning it on

6 Module cards

53 © Copyright Vaisala 2019. All rights reserved. 6.3 Relay module card PR-50023 The PR-50023 relay module card adds two relays to the Multichannel user interface MI. The relays can be configured independently to any connected refractometer. The relays can be used e.g. for controlling prism wash. Specifications Relays per module 2 Voltage and current rating Max. 250V/3A Connection Screw connection with tension sleeve Connector Phoenix 1900824 – MC 1,5/ 7-ST1-5,08 Contact type Dry contact Normal relay state NO Conductor cross section 0.14-1.5 mm2 (28-16 AWG) Stripping length 7 mm (¼") Tightening torque 0.25 Nm (0.18 lbf.ft) 7 Relay 1+ 6 Relay 1- 5 Relay 2+ 4 Relay 2- 3 no connection 1 no connection 2 no connection LED RLY1 LED RLY2 Figure 6.5 Relay connections

MI instruction manual

Document/Revision No. IM-EN-MI/2.00 Effective: September 2, 2019 6.3.1 Configuring relays On the relay display, first select the application the relay is used for, then select the function. After the function is chosen, you get access to those relay parameters that apply to this function. Figure 6.6 Configuring a relay Note: Even though the relays can be configured independently, it is recommended that if you e.g. set precondition and wash parameters for a certain refractometer, you set those in the same relay module card. Important: When you have chosen the settings for the relays, tap DONE to save changes and exit the relay setup. Parameters LIMIT and HYSTERESIS are used to switch the relay based on concentration or temperature measurement values by the functions CONC - HIGH LIMIT, CONC - LOW LIMIT, T - HIGH LIMIT and T - LOW LIMIT. The relay closes on high limit when the mea- sured value exceeds HIGH LIMIT + HYSTERESIS. The relay opens on high limit when the measured value goes below HIGH LIMIT - HYSTERESIS. Alternatively, the relay closes on low limit when the measured value goes below LOW LIMIT - HYSTERESIS. The relay opens on low limit when the measured value exceeds LOW LIMIT + HYSTERESIS.

6 Module cards

55 © Copyright Vaisala 2019. All rights reserved. Relay test page Relay parameters Clear all settings Finish and save settings Figure 6.7 Accessing relay functions Figure 6.8 Relay module parameters

MI instruction manual

Document/Revision No. IM-EN-MI/2.00 Effective: September 2, 2019 6.4 mA output module PR-50021 The mA output module card adds mA outputs to Multichannel user interface MI. On the module card there are two independent 4-20 mA current sources. Both current outputs are configurable independently to indicate for example process concentration or temperature or other diagnostic data of any connected refractometer. A mA output has traceable calibration, accuracy is ±0.02 mA. Maximum load (compliance) is 1000 Ohm. There is a place inside the Multichannel user interface MI where the cable shield can be connected for grounding. Cable specifications Conductor cross section 0.14-1.5 mm2 (28-16 AWG) Tightening torque 0.25 Nm (0.18 lbf.ft) 10 mA out 1+ 9 mA out 1- 8 no connection 7 no connection 6 mA out 2+ 5 mA out 2- 4 no connection 3 no connection 2 Frame ground 1 Frame ground Figure 6.9 mA output pinout The mA outputs are compliant to Namur NE43. NAMUR NE 43 uses the 3.8 to 20.5 mA signal range for measurement information, with values ≥21 mA or ≤3.6 mA to indicate diagnostic failures (see Figure 6.10). With that information, it is easier to detect a failure condition on a refractometer, for example, it clearly tells you if you have an empty pipe.

6 Module cards

57 © Copyright Vaisala 2019. All rights reserved. 0 3.6 3.8 4 20 20.5 21 mA failure OK Measurement Data failure Figure 6.10 Default mA output values Certain malfunctions cause the mA measurements to return to mA default output value, see Table 6.1 below. Factory setting for default output value is 3.4 mA. Returns to default mA Message Conc Temp Short-circuit X X No optical image X Temp measurement fault X X No sample X Prism coated X No connection X X Table 6.1 Messages causing mA output value to return to default output 6.4.1 Configuring mA outputs First select the application for an output, then set function. There are five possible functions, CALC, CONC, QF, TEMP and ND. After you’ve chosen a function, you can set the parameters for that function. • CALC is the calculated concentration value without field calibration adjustment • CONC is the final concentration value including field calibration adjustment • QF is the quality factor measuring the image sharpness, available for Refractometer app • TEMP is the process temperature • ND is the refractive index value nD from the sensor

MI instruction manual

Document/Revision No. IM-EN-MI/2.00 Effective: September 2, 2019 Finish by tapping here to save settings Tap for output test page Output parameters Clears all settings for this output Figure 6.11 mA output settings Important: When you have chosen the settings for the outputs, tap DONE to save changes and exit. 6.5 General I/O module PR-50024 The General I/O module card adds digital inputs and outputs to Multichannel user in- terface MI. The module card provides 8 input/output channels. The input and output terminals are galvanically isolated according to • UL 2500 V rms (UL1577) • CSA 800 V rms (1131V peak) maximum working voltage - basic insulation • CSA 400 V rms (566V peak) maximum working voltage - reinforced insulation • VDE Reinforced insulation 560V peak • TüV Reinforced insulation 400V rms maximum working voltage Cable specifications Conductor cross section 0.14-1.5 mm2 (28-16 AWG) Tightening torque, max. 0.25 Nm (0.18 lbf.ft)

6 Module cards

59 © Copyright Vaisala 2019. All rights reserved. Voltage reference: IO voltage can be selected with a jumper in the module. IO voltage options are internal 5V, internal 24V and external voltage. When external voltage is selected, a voltage of 3-30 VDC must be provided between V_IO and GND pin. When internal voltage is selected, voltage reference for switch input can be taken from V_IO and GND pins.

MI instruction manual

Document/Revision No. IM-EN-MI/2.00 Effective: September 2, 2019 Electrical characteristics: EXTERNAL IO VOLTAGE SUPPLY External IO voltage input range (VIO) 3-30 VDC Absolute minimum external IO voltage input 0 VDC Absolute maximum external IO voltage input 36 VDC OUTPUTS Maximum combined* supply current when using internal 24V 20 mA Maximum combined* supply current when using internal 5V 100 mA Maximum individual channel supply current when using external IO voltage 1 A Maximum combined* supply current when using external IO voltage 2 A IO output voltage (voltage drop) when using external IO voltage VIN - 0.15 VDC @ 5 VDC VIN - 0.20 VDC @ 24 VDC INPUTS Input type Sinking Logical 0: voltage range 0.0-0.7 VDC Logical 1: voltage range 1.0-30 VDC Input impedance 10 kohm Absolute maximum input voltage 36 VDC * current limit shared with all 8 channels

6 Module cards

61 © Copyright Vaisala 2019. All rights reserved. 6.5.1 Connecting inputs/outputs Terminal no. Description Function 1 Digital input/output In/Out 1 2 Digital input/output In/Out 2 3 Digital input/output In/Out 3 4 Digital input/output In/Out 4 5 Digital input/output In/Out 5 6 Digital input/output In/Out 6 7 Digital input/output In/Out 7 8 Digital input/output In/Out 8 9 Ground for outputs GND 10 IO voltage input/output V_IO 1 2 3 4 5 6 7 8 9 10 Figure 6.12 General I/O module pinout

MI instruction manual

Document/Revision No. IM-EN-MI/2.00 Effective: September 2, 2019 Digital outputs: Digital output voltages are selected with jumper. GND (pin 9) is used for ground. See Figure 6.13 for connections. Digital inputs: Digital inputs can be used as switch inputs or controlled by external signal. Switch input configuration is shown in Figure 6.14. When an external signal is used, an external power supply is connected between input pin and GND (pin 9) as seen in figure 6.14. 1 2 3 4 5 6 7 8 9 10 GND V_IO PR-50024 Figure 6.13 Channels 1 and 2 configured as output 1 2 3 4 5 6 7 8 9 10 GND V_IO PR-50024 Figure 6.14 Channels 3 and 4 configured as input and connected as switch input

6 Module cards

63 © Copyright Vaisala 2019. All rights reserved. 1 2 3 4 5 6 7 8 9 10 GND V_IO PR-50024 + Figure 6.15 Channel 5 configured as input from external power supply 6.5.2 Configuring the general I/O module card To configure the inputs and outputs, select the module card in the module view. Select the application to which the input/output is connected and then choose the desired function from a list. The digital outputs can be configured for • Concentration - high limit • Concentration - low limit • Error • Full wash (precondition, wash, recovery) • Normal • Precondition • Temperature - high limit • Temperature - low limit • Warning • Wash The digital inputs can be configured for • External hold • External wash • External wash stop

MI instruction manual

Document/Revision No. IM-EN-MI/2.00 Effective: September 2, 2019 Tap to save all settings Tap for test page Change I/O 1 parameters Clear I/O 1 settings Figure 6.16 General I/O module card configuration Parameters LIMIT and HYSTERESIS are used to switch the digital output based on con- centration or temperature measurement values by the functions CONC - HIGH LIMIT, CONC - LOW LIMIT, T - HIGH LIMIT and T - LOW LIMIT. Switching on high limit activates when the measured value exceeds HIGH LIMIT + HYSTERESIS. The output switches off on high limit when the measured value goes below HIGH LIMIT - HYSTERESIS. Respec- tively, switching on low limit activates when the measured value goes below LOW LIMIT - HYSTERESIS. The output switches off on low limit when the measured value exceeds LOW LIMIT + HYSTERESIS. Important: When you have chosen the settings for the inputs or outputs, tap DONE to save changes and exit the setup. 6.5.2.1 Using the general I/O module to trigger prism wash Prism wash can be started by configuring the general input for external wash request. To activate the wash system a relay must be configured to be a wash relay. Select the wash function on the relay display, see figure 6.6. To configure the general input, select the function "IN: EXT WASH" on the I/O module display.

6 Module cards

65 © Copyright Vaisala 2019. All rights reserved. Figure 6.17 Selecting the function "IN: EXT WASH" to configure external wash request

MI instruction manual

Document/Revision No. IM-EN-MI/2.00 Effective: September 2, 2019 7 Prism wash In some applications the process flow does not keep the prism clean because flow ve- locity is not enough to keep the prism clean, or the process medium is sticky. In these applications the prism can be automatically cleaned by installing a wash system (see the refractometer manual). 7.1 Configuring prism wash The prism wash settings for different refractometers are independent of each other. The wash system is active if a relay has been configured to be a wash relay (see Sec- tion 6.3.1) and the wash time is not zero. An automatic wash function can be configured so that all refractometers connected to MI have different wash parameters. The wash parameters can be configured at Refractometer app -> Configuration -> Wash. 7.1.1 Wash cycle The wash logic is shown in Figure 7.2 as a flow diagram. The automatic prism wash cycle (Figure 7.1) consists of three phases: precondition, wash and recovery. The optional preconditioning function is used to e.g. blow out the condensate from the washing line before washing. After the preconditioning there is a one-second pause to avoid having both precondition and wash relays active at the same time. The wash cycle is initiated when the wash interval has elapsed. The wash can also be started by configuring a general input for external wash (Section 6.5.2) or manually from the user interface at Refractometer app -> Calibration menu. The order of priority for these wash triggers is 1. manual wash 2. external wash request 3. wash interval timer The wash relay is closed for the wash time specified in the wash settings. After the wash phase is completed, a recovery time is spent. During the wash cycle (pre- condition, wash, recovery) the measurement result is in hold unless otherwise speci- fied.

7 Prism wash

67 © Copyright Vaisala 2019. All rights reserved. WASH RECOVERY INTERVAL PRECOND. Figure 7.1 The automatic prism wash cycle Preventing automatic wash: The precondition and wash relays are never activated by the automatic wash control: – If diagnostic message is NO SAMPLE as this indicates a clean prism in an empty process line. The diagnostic message is WASH STOP/NO SAMPLE. – If a wash stop is present, indicating e.g. that there is no process flow. The diagnostic message is EXTERNAL WASH STOP. – If the process temperature limit is activated and the temperature falls below the limit, indicating that the process is not running. The diagnostic message is LOW TEMP WASH STOP.

MI instruction manual

Document/Revision No. IM-EN-MI/2.00 Effective: September 2, 2019 MANUAL WASH been pressed? NO YES External wash 1 requested? NO YES Wash interval reached? YES NO Temp limit active? YES NO Temp above limit? YES NO Message 2 NO SAMPLE? NO YES Wash cycle Restart wash interval timer NOTES 1. External wash is triggered at the general input. If the input is held logical high, only one wash cycle is carried out. 2. The wash is inhibited if there is no sample, no refractometer or the refractometer cannot measure correctly. Figure 7.2 Wash logic

7 Prism wash

69 © Copyright Vaisala 2019. All rights reserved. Precondition Start washing 1 s wait Precondition enabled? YES NO Wash time reached? YES NO Stop washing Recovery PRECONDITION RECOVERY WASH Figure 7.3 Wash cycle

MI instruction manual

Document/Revision No. IM-EN-MI/2.00 Effective: September 2, 2019 7.1.2 Setting prism wash parameters To set prism wash parameters for a given refractometer, first select the Refractometer app, then select Calibration and then Prism wash. This menu contains the alternatives (factory settings are given in parentheses): 1. Precondition secs 0-60 s (0) 2. Recovery secs 0-120 s (20) 3. Wash secs 0-30 s (3) 4. Empty pipe check (enabled) 5. Wash check mode (disabled) 6. Wash verify delta nD limit 0.0-0.5 (0.0) 7. Wash inhibit temp enable (disabled) 8. Interval mins 0-1440 min (20) 9. Wash inhibit temp value 0-200 C (0) 10. Wash verify nD limit 1.15-1.80 (1.35) 11. Wash failure tolerance mins 0-1440 (0) 12. Measurement hold during wash (enabled) 13. Temperature hold during wash (disabled) The prism wash cycle: The timing of the wash cycle is controlled by the WASH INTERVAL, PRECONDITION TIME, WASH TIME and RECOVERY TIME settings. If the WASH INTERVAL is set to zero, the wash can only be initiated by using the manual wash (PRISM WASH on Diagnostics menu) or general input module configured for external wash. If the PRECONDITION TIME is zero (or there is no relay configured for preconditioning), the preconditioning phase is skipped. If the WASH TIME is zero (or there is no relay for wash), the wash functionality is completely disabled. The recommended wash times and wash medium pressures are given in the refrac- tometer manual. Wash check mode, standard: The prism wash check monitors automatically that the prism wash really has an effect on the prism. In the WASH CHECK mode, prism wash is accepted if the refractive index nD either falls below wash verify nD limit (default 1.35) at NORMAL OPERATION or NO SAMPLE occurs. This is the indication of a successful wash with water or steam.If the wash is not accepted, the diagnostic message PRISM WASH WARNING will appear. If no wash is accepted during wash tolerance time, the message

7 Prism wash

71 © Copyright Vaisala 2019. All rights reserved. becomes PRISM WASH FAILURE. Both messages and the wash tolerance counter are re- set by a successful wash.The Wash check mode, automatic wash cut differs from the standard mode by stopping the wash 2 seconds after the nD falls below the limit. To stop the measurement for the duration of the prism wash, choose MEASUREMENT HOLD DURING WASH and in that menu activate the hold function. The CONC reading and mA output are held in the value they had immediately before starting the wash cycle. You can also stop temperature measurement for the duration of the wash by setting TEMPERATURE HOLD DURING WASH. To set a low temperature limit, choose WASH INHIBIT TEMPERATURE ENABLE, enable it (fac- tory setting is disabled) and then set the temperature with WASH INHIBIT TEMP VALUE. The temperature is in the units that is chosen from the Calibration menu as the secondary unit. The empty pipe check prevents washing if message is NO SAMPLE, i.e. there’s no process liquid in the pipe. To deactivate (or activate) the empty pipe check, choose EMPTY PIPE CHECK and then the appropriate menu command. To change the wash nD limit, select WASH VERIFY ND LIMIT to set the nD to be used by the wash check functionality. If you have set a wash verify delta nD limit, a temporary wash verify limit is calculated in the beginning of wash: nD limit = nD current – ΔnD. This temporary limit is compared to actual wash verify nD limit parameter, and the higher of these values takes precedence. To set wash tolerance time, select WASH FAILURE TOLERANCE to set the time during which a wash has to be accepted. If no wash is accepted during wash tolerance time, the mes- sage becomes PRISM WASH FAILURE. The wash tolerance counter is reset by a successful wash.

MI instruction manual

Document/Revision No. IM-EN-MI/2.00 Effective: September 2, 2019 8 Specifications Model: Multichannel User Interface Product code: MI-M-UN-AC S/N: MI00103 Tag: 100-240V AC, 50/60 Hz, 60 VA Made by VAISALA Oyj, Vantaa, Finland www.vaisala.com IP67 Type 4x ® C US Model: Multichannel User Interface Product code: MI-M-UN-DC S/N: MI00103 Tag: 24V DC, 60 W Made by VAISALA Oyj, Vantaa, Finland www.vaisala.com IP67 Type 4x ® C US 100-240 V AC 24 V DC Figure 8.1 MI nameplates

© Copyright Vaisala 2019. All rights reserved.

MI specifications Display: xga 1024 x 768 pixels, touch screen Power: AC input 100-240 VAC, ±10%, 50/60 Hz, 60 VA, optional 24 VDC Overvoltage category: OVC II Refractometer connectivity: 1-4 refractometers can be connected to the MI. Refractometers independent of each other: own parameter sets and usable in different applications. Ethernet connection: 10/100 Mbit/s, data acquisition over UDP/IP Protocol Modules: 1-8 modules for added connectivity Apps: 1-4 apps Protection class: Enclosure IP66/67. Type 4X. Installation: Indoor/outdoor, max. 50°C (122°F), min. 0°C (32°F), must be protected from rain and direct sunlight.During installation highest relative humidity of ambient air 80%. Highest altitude 2000 m above sea level. Weight: 12.9 kg (28.5 lbs) Regulatory compliance: Safety: IEC 61010-1:2010/EN 61010-1:2010, CSA certified to CSA-C.22.2 No. 61010.1-12, UL 61010-1:2012 Electromagnetic compatibility: IEC 61326-1:2012/EN 61326-1:2013. Optional modules Relay Module: 2 signal relays 150V/3A mA Output Module: 2 x 4-10mA, max. load 1000 Ohm, galvanic isolation 1000VDC Regulatory compliance Safety: IEC 61010-1:2010/EN 61010-1:2010, CSA certified to CSA-C.22.2 No. 61010.1-12, UL 61010-1:2012 Electromagnetic compatibility: IEC 61326-1:2012/EN 61326-1:2013

MI instruction manual

Document/Revision No. IM-EN-MI/2.00 Effective: September 2, 2019 8.1 Model code MULTICHANNEL USER INTERFACE Model and description Model MI = Multichannel user interface, stainless steel enclosure with wall mount. Connectivity for up to 4 refractometers with Ethernet output, included 1 x mA output module card (2 x 4–20 mA), 1 x Relay module card (2 relays, 250 V/3A) MI Cable connection for power -U = ½inch NPT type conduit hubs -U -M = M20x1.5 metric cable glands -M Electrical classification -UN = Unclassified area, General Purpose, Ordinary location -UN Power supply -AC = Power supply 100–240 VAC 50/60 Hz -AC -DC = Power supply 24 V DC -DC Module cards for MI (0-8 modules selectable in total, one or several) -MO1 = 1 x mA output module card, for 2 x mA output (included in standard delivery) -MO1 -MO2 = 2 x mA output module card, for 4 x mA output -MO2 -MO3 = 3 x mA output module card, for 6 x mA output -MO3 -MO4 = 4 x mA output module card, for 8 x mA output -MO4 -RM1 = 1 x Relay module card, for 2 x relay output (included in standard delivery) -RM1 -RM2 = 2 x Relay module card, for 4 x relay output -RM2 -RM3 = 3 x Relay module card, for 6 x relay output -RM3 -GO1 = 1 x General I/O module card, 4 x digital input, 4 x digital output -GO1 -GO2 = 2 x General I/O module card, 8 x digital input, 8 x digital output -GO2

A EU declaration of conformity

77 © Copyright Vaisala 2019. All rights reserved. A EU declaration of conformity 2019-09-01B/JAMO 1 (1) Vaisala Oyj | PO Box 26, FI-00421 Helsinki, Finland Phone +358 9 894 91 | Fax +358 9 8949 2227 Email [email protected] | www.vaisala.com Domicile Vantaa, Finland | VAT FI01244162 | Business ID 0124416-2 EU DECLARATION OF CONFORMITY Manufacturer: Vaisala Oyj Mail address: P.O. Box 26, FI-00421 Helsinki, Finland Street Address: Vanha Nurmijärventie 21, Vantaa, Finland This declaration of conformity is issued under the sole responsibility of the manufacturer. Object of the declaration: K-Patents Multichannel User Interface MI The object of the declaration described above is in conformity with Directives: RoHS Directive (2011/65/EU) EMC Directive (2014/30/EU) Low Voltage Directive (2014/35/EU) The conformity is declared using the following standards: EN 50581:2012 Technical documentation for the assessment of electrical and electronic products with respect to the restriction of hazardous substances EN 61000-3-2:2006 + A1:2009 + A2:2009 Electromagnetic compatibility (EMC) - Part 3-2: Limits - Limits for harmonic current emissions EN 61000-3-3:2013 Electromagnetic compatibility (EMC) - Part 3-3: Limits - Limitation of voltage changes, voltage fluctuations and flicker in public low-voltage supply systems EN 61010-1:2010 Safety requirements for electrical equipment for measurement, control and laboratory use – Part 1: General requirements EN 61326-1:2013 Electrical equipment for measurement, control and laboratory use – EMC requirements – intended for use in industrial locations Signed for and on behalf of Vaisala Oyj, in Vantaa, on 1st September 2019 ______________________________________ Jukka Lyömiö Standards and Approvals Manager

B Software licence

79 © Copyright Vaisala 2019. All rights reserved. B Software licence This software includes parts of software libraries included under the following licence notices. Copyright ©2012, Aaron Gifford All rights reserved. Redistribution and use in source and binary forms, with or without modification, are permitted provided that the following conditions are met: • Redistributions of source code must retain the above copyright notice, this list of conditions and the following disclaimer. • Redistributions in binary form must reproduce the above copyright notice, this list of conditions and the following disclaimer in the documentation and/or other ma- terials provided with the distribution. • Neither the name of the nor the names of its contributors may be used to endorse or promote products derived from this software without specific prior written permission. THIS SOFTWARE IS PROVIDED BY THE COPYRIGHT HOLDERS AND CONTRIBUTORS "AS IS" AND ANY EXPRESS OR IMPLIED WARRANTIES, INCLUDING, BUT NOT LIMITED TO, THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE DISCLAIMED. IN NO EVENT SHALL AARON GIFFORD BE LIABLE FOR ANY DIRECT, INDIRECT, INCIDENTAL, SPECIAL, EXEMPLARY, OR CONSEQUENTIAL DAMAGES (INCLUDING, BUT NOT LIMITED TO, PROCUREMENT OF SUBSTITUTE GOODS OR SER- VICES; LOSS OF USE, DATA, OR PROFITS; OR BUSINESS INTERRUPTION) HOWEVER CAUSED AND ON ANY THEORY OF LIABILITY, WHETHER IN CONTRACT, STRICT LIABILITY, OR TORT (INCLUDING NEGLIGENCE OR OTHERWISE) ARISING IN ANY WAY OUT OF THE USE OF THIS SOFTWARE, EVEN IF ADVISED OF THE POSSIBILITY OF SUCH DAMAGE.

Command selection tree for Refractometer application

81 © Copyright Vaisala 2019. All rights reserved. Command selection tree for Refractometer appli- cation Main screen REFR. APP DIAGNOSTICS CALIBRATION TREND LOG PRISM WASH VERIFICATION FIELD SAMPLE INPUT OUTPUT IDENTIFICATION SUPPORT SELECT DISPLAY MODES PRIMARY UNIT PRIMARY DECIMALS SECONDARY UNIT CONCENTRATION TEMPERATURE BOTH CELSIUS FAHRENHEIT CHEMICAL FIELD TAG SENSOR SOURCES PRISM WASH SUMMARY CLEAR

Command selection tree for settings

83 © Copyright Vaisala 2019. All rights reserved. Command selection tree for settings Main screen SETTINGS GENERAL DISPLAY SETUP QUADRANT ONE ON TOP DUAL VERTICAL ONE MAXIMIZED SCREEN BRIGHTNESS LANGUAGE HIGH CONTRAST MODE DATE & TIME PROTECT WITH PASSWORD BACKUP/RESTORE FACTORY RESET UPDATE SYSTEM TAG NETWORK CURRENT IP ADDRESS IP ADDRESS MODE SET STATIC IP ADDRESS SET STATIC IP ADDRESS APPLICATIONS APP SLOT X (1-8) MODULES REMOVE ADD MODULE SLOT X (1-8) USERS ADD USER SET NAME SET PASSWORD LOG CLEAR START BOOTP SERVER SET FIRST SENSOR IP ADDRESS CURRENT IP ADDRESS IP ADDRESS MODE Refractometer Ethernet Master user login required (IP address can be viewed without login) SET SUBNET MASK SET SUBNET MASK SET DEFAULT GATEWAY STATIC DYNAMIC AUTO YES NO AUTO STATIC DYNAMIC AUTO