THORLABS OTKB Manual
Text of THORLABS OTKB Manual
Optical Trap Kit
19590-D04-V15; Apr 12, 2012
Page 2
www.thorlabs.com
Table of Contents
1.
Safety ............................................................................................................................................3
2.
Introduction ..................................................................................................................................4
3.
Setup Description and Segments .......................................
Optical Trap Kit
19590-D04-V15; Apr 12, 2012
Page 3
www.thorlabs.com
Part 1. Safety
All statements regarding safety of operation and technical data in this instruction manual will only apply
when the unit is operated correctly.
WARNING
This unit must not be operated in explosive environments
WARNING
Make sure the fuse setting at the back of the laser diode driver and temperature controller
are appropriate for your region. Failure to do this will result in
Optical Trap Kit
19590-D04-V15; Apr 12, 2012
Page 4
www.thorlabs.com
Part 2.
Introduction
In 2007, three researchers in the Department of Biological Engineering at MIT – Steve Wasserman,
David Appleyard, and Matthew Lang – built an optical trapping setup for use in teaching labs. Their
results were published in the American Journal of Physics [S. Wasserman, D. Appleyard, and M. Lang,
Optical Trapping for Undergraduates, Am. J. Phys. 75, (2007)]. Based
Optical Trap Kit
19590-D04-V15; Apr 12, 2012
Page 5
www.thorlabs.com
Figure 5.1: The Collimator Segment
3.2
The Beam Expander Segment:
This includes the beam expansion achromatic doublets and the cube assembly (with the dichroic) for
reflecting the trap laser and transmitting the visible spectrum for camera viewing. It also includes the
imaging lens for the camera. The Collimator segment, the Camera segment and the Vertical segment all
connect to this at i
Optical Trap Kit
19590-D04-V15; Apr 12, 2012
Page 6
www.thorlabs.com
3.3
The Vertical Segment
This segment holds the objective, condenser and light source as could be seen in Figure 5.3. To mount
this segment, first remove the top part of whole assembly by unlocking the two screws that connect this
to the vertical post. Next, release the screw attaching the lower C1500 to the vertical post and adjust its
height until the four ER1.5 rods can slide into the
Optical Trap Kit
19590-D04-V15; Apr 12, 2012
Page 7
www.thorlabs.com
3.4
The Stage Assembly
The stage assembly parts ship in their separate boxes and will have to be assembled as described in
Section 6.8.
3.5
The Camera Segment
This segment mainly has the CCD camera for live imaging. The imaging lens is already connected to the
camera port on the cube, C6W. To complete this segment, just slide the four ER1 rods into the four cage
rod holes on the C6W, such
Optical Trap Kit
19590-D04-V15; Apr 12, 2012
Page 8
www.thorlabs.com
Part 4.
Optical Trapping Kit OTKB(OTKB/M) Components
This section provides some technical details on the individual components used to build this system.
Like many optical trapping systems, this one is based on an inverted
microscope design. The structure of the inverted light microscope is
constructed using Thorlabs‟ 60mm cage system, which is supported by a
damped 1.5″ post (Item