Tenmars TRIAXIAL ELF Manual
Read Online
Read the Tenmars TRIAXIAL ELF Manual online using the built-in PDF viewer.
Text of the manual
Table of Contents
1.
Introduction ............................................... 1
2.
Application ................................................ 2
3.
Features.................................................... 3
4.
Identifying Parts ........................................ 4
5.
Measurement Procedures......................... 5
6.
Clock setup ............................................... 7
7.
Auto power off setup................................. 8
8.
Continuity data logging setup.................... 9
9.
Clear data logger memory ...................... 10
10.
Clear all data logger memory...................11
11.
Single data memory................................ 12
12.
Viewing logged reading control key ........ 12
13.
Specifications.......................................... 13
14.
Battery replacement................................ 14
15.
External DC Power ................................. 14
16.
Safety Precaution.................................... 15
17.
Software installation (192D).................... 15
18.
Note ........................................................ 17
Safety precautions should be taken against electric
appliances in places including medical institutions,
schools and residential districts, where people
usually stay for a long time, to prevent patients,
babies and senior citizens from exposure to high
electromagnetic waves.
An electromagnetic wave simply means the wave
motion of the electromagnetic field (EMF).
The change in electric fields produces magnetic
fields, and the change in magnetic fields can also
generate electric fields. The fluctuation of
correlation between each other is known as
“electromagnetic waves", which is a form of energy
similar to light and heat that can be transmitted
either by radiation in the air or by an electric
conductor.
It is suggested that rearrangements should be
made in homes and work locations where the
strongest electromagnetic fields are detected in
order to avoid prolonged exposure to excessive
electromagnetic fields.
z This meter is applied to measuring magnetic
fields of extremely low frequency (ELF) of 30 to
2000Hz.
z It is capable of measuring the magnetic field
radiation intensity that is produced from electric
transmission equipment, power lines, microwave
ovens, air conditioners, refrigerators, computer
monitors, video/audio devices and so forth.
z The magnetic field unit is Tesla (T), Gauss (G),
milli-Gauss (mG) or micro-Tesla (μT).
1 T=10,000 G
1 G=1,000 mG
1μT=10 mG
z Uses three internal orthogonal sensors to test a
wide range of ELF magnetic fields, independent
of measurement angle.
z The tester is designed to provide a quick, reliable
and easy way to measure magnetic field radiation
levels around power lines, home appliances and
industrial devices.
z The tester is a cost-effective hand-held
instrument that was designed and calibrated to
measure magnetic field radiation at different
bandwidths from 30Hz to 2000Hz.
z Display microTesla(uT) or milliGauss(mG) units.
z Data hold (HOLD), maximum Hold (MAX), and
minimum Hold(MIN) function.
z Auto range or manual range select mode.
z Datalogging capacity with Memory Size:500 data
sets (Model 192) or 9999 data sets(Model 192D).
z USB PC interface (192D).
z Time and calendar function.
z Low battery indication “
”.
z Over load display “OL”.
z Auto power OFF function .
4. Identifying Parts
1. LCD display.
2. Power key.
3. maximum Hold and
minimum Hold key.
4. Data Hold key.
5. Record data and
Setup function left
key.
6. Range and Setup
function up key.
7. Reading and Setup
function right key.
8. Setup function key.
9. Units and Setup
function down key.
10. Enter key.
11. Calibration point for
12. Triple axis.
13. External power DC
9V.
14. USB interface
(TM-192D).
15. Tripod mounting
screw.
16. Battery cover.
5. Measurement Procedures
z Press “
” key turn on the meter, press “
”
key again to turn OFF the meter.
z Press “
” key to select milli-gauss (mG) or
micro-tesla (µT) unit.
z Position the front of the meter to measure the
magnetic waves.
z Read the measured value. When manual range
mode is
selected, LCD will shown the
mark, The
display of “OL” on the highest position indicates
there is an overload on the reading. Press “
”
to select a higher range for measurement again.
z The default value display is triaxial total magnetic
field reading. Press “
” key to toggle between
independent three single-axis magnetic field
readings and triaxial total magnetic field reading.
z To lock and keep the reading displayed on the
LCD, press “
” or press “
” again to
unlock.
z To retain the maximum and minimum value,
press “
” key and the reading value displayed
on the LCD will keep
z updating to the maximum value. Press again to
select minimum value, press and hold down
“
” key 2 seconds to exit the maximum and
minimum mode.
z Due to the magnetic interference of the
environment field factors, this magnetic field
meter could display a reading value that is
greater than 0.5 mG prior to measuring. This is
not a malfunction of the tester.
z With the tester in hand, move slowly towards to
the object under measurement until it is
physically touched.
z Notice how the field intensity increases as you
move closer to the object.
z If the power of object was turned off during the
measurement, the reading of tester should return
to zero, unless an electromagnetic field from
other sources is detected.
6. Clock setup
z This meter clock uses 24 hour time.
z Press “
” key to enter clock setting mode
(1.SET).
z Press “
” or “
” key to select option to
adjust.
z Press “
” or “
” key to change the digit.
z Press “
” key to store the setup, exit the
mode.
7. Auto power off setup
z Press “
” key again to auto power off time
setting mode (2.SET).
z Press “
” or “
” key to change the auto
power off time.
z Press “
” key to store the setup, exit the
mode.
z The auto power off time default value is 5 minutes.
Range is 0 to 99 minutes. To cancel auto power
off, please set time to 0 minutes.
8. Continuity data logging setup
The data logging interval is the time between data
records. To begin data logging, set the logging
interval as per instructions below. To exit data
logging, set logging interval back to 0.
z Press “
” key again to enter continuous data
logging interval setting mode (3.SET).
z Press “
” or “
” key to select option to
adjust.
z Press “
” or “
” key to change the value.
z Press “
” key to store the setup, exit the
mode.
Interval range is 0 seconds (no data logging) to 99
hours, 99 minutes and 99 seconds.
Clear data logger memory (last data entry)
z Press “
” key again to clear data logger
memory for last record setting mode (4.CLA)
z Press “
” key to clear data logger memory for
last record and exit the mode.
9. Clear ALL data logger memory
z Press “
” key again to clear data logger
memory for all record setting mode. (5.CLA)
z Press “
” key to clear data logger memory for
all record.
z Press “
” key to store the setup, exit the
mode.
10. Single data memory record
z Press “
” key each time to store the display
reading and time stamp in memory.
11. Viewing logged reading control
key
z Press “
” key to view logged readings.
Press
or
key to scroll through the
readings, Press “
” key to exit this mode.
z Display : 4 digits Triple LCD display.
z Range : 20/200/2000 mG, 2/20/200µT.
z Resolution : 0.01/0.1/1 mG or 0.001/0.01/0.1 µT.
z Frequency response : 30Hz to 2000Hz.
z Sensor : Triple Axis (X, Y, Z).
z Accuracy : 20 mG ±(3.0%+30dgt) at 50Hz/60Hz.
or 3 µT ±(3.0%+30dgt) at 50Hz/60Hz.
z Over load : LCD display “OL”.
z Sample rate : 2.5 times per second.
z Battery : 9V NEDA 1604, IEC 6F22 or JIS 006P.
z Battery life : Approximate 100 hours.
z Operating temperature & humidity :
5℃to 40℃, below 80% RH.
z Storage temperature & humidity :
-10℃ to 60℃, below 70%.
z Weight : About 230g.
z Dimensions : 173(L) X80(W) X32(H) mm.
z Accessories : User’s manual, 9V battery,
Carrying case.(MINI USB 4P(MALE) to USB A
Type cable, CD disk 192D).
13. Battery replacement
z Turn off the instrument.
z Open the battery cover and remove the battery.
z Replace with four-9V NEDA 1604, IEC 6F22 or
JIS 006P size battery.
z Install the battery cover.
14. External DC Power
z External AC to DC adapter :
Voltage 9VDC (8~14VDCMax).
z Supply current:>300Madc.
z Socket:pin Positive, Ground Casing External.
z Diameter 6.3mm ; internal Diameter 2.0 mm.
WARNING
If the symbol “
” appears on the
LCD, please replace the battery
immediately