SCI 260 Series Quick Start Manual
Text of SCI 260 Series Quick Start Manual
GROUND BOND TESTERS
QUICK START GUIDE
Models: 264, 266
SAFETY CHECKLIST
Survey the test station. Make sure it is safe & orderly.
Always keep unqualified/unauthorized personnel away from the test area.
Familiarize yourself with safety protocols in the event of a problem.
Exercise caution and never touch products or connections during a test.
Train operators. Connect the return lead first and never touch clips directly.
You should always know when
Connect the female end of the standard
NEMA style line power cord into the
input power receptacle on the rear panel
of the tester. Plug the male end of the
cord into a grounded power source.
Connect the Interlock Disable Key into
the Signal Input connector on the rear
panel of the tester.
This is required to run a test.
Turn the POWER switch ON.
Upon start up an initial screen will
appear briefly. After two seconds the
Home screen will appear as shown
bel
If you wish to have multiple test setups, you will need to begin by choosing a Memory Location
(M1 - M5) for each test.
3
hipot.com
SETTING TEST MEMORIES
EDIT TEST PARAMETERS
START
PROGRAM A MEMORY
Turn the yellow rotary knob to
scroll to M1 to edit Memory 1.
(M2 = Memory 2, etc.)
Home screen.
When M1 is selected. Press the
rotary knob to recall Memory 1.
EDIT TEST PARAMETERS
When the parameter is selected,
Press the rotary knob to edit the
parameter. The
Hook the black return test lead (P/N 99-10724-01) to the black RETURN terminal on the
front panel of the tester. You may need to unscrew the terminal slightly to make room for
the hook.
Clip the other end of the black return test lead to the chassis of the DUT.
Hook one end of the red high current test lead (P/N 99-10725-01) to the red CURRENT
terminal on the front panel of the tester.
Clip the other end of the red high current test lead to the ground pin
Connect the Interlock Disable Key (P/N 99-10040-01)
to the Signal Input connector on the rear panel of
the tester. If you’re not utilizing a DUT enclosure (P/N
99-10715-01 or P/N 99-10599-01) or other safety
device, the Interlock Disable Key is required in
order to run a test.
With the tester set to the desired parameter and
your DUT correctly connected to the tester, you are
now ready to start testing.
Push the green TEST button on the front
SAFETY MADE SIMPLE®
DO NOT set up electrical safety testing stations and ESD (electrostatic discharge) stations in the same area. ESD protocols are designed to protect a
component or device from static discharge (not the operator from high voltage hazards). DO NOT use anti-static robes, benches or floor mats during
electrical safety testing. All of these items are used to intentionally ground the tester operator which can cause injury or death to a high