SCI 260 Series Quick Start Manual

Text of SCI 260 Series Quick Start Manual

GROUND BOND TESTERS

QUICK START GUIDE Models: 264, 266 SAFETY CHECKLIST Survey the test station. Make sure it is safe & orderly. Always keep unqualified/unauthorized personnel away from the test area. Familiarize yourself with safety protocols in the event of a problem. Exercise caution and never touch products or connections during a test. Train operators. Connect the return lead first and never touch clips directly. You should always know when

Connect the female end of the standard

NEMA style line power cord into the input power receptacle on the rear panel of the tester. Plug the male end of the cord into a grounded power source. Connect the Interlock Disable Key into the Signal Input connector on the rear panel of the tester. This is required to run a test. Turn the POWER switch ON. Upon start up an initial screen will appear briefly. After two seconds the Home screen will appear as shown bel

If you wish to have multiple test setups, you will need to begin by choosing a Memory Location

(M1 - M5) for each test. 3 hipot.com SETTING TEST MEMORIES EDIT TEST PARAMETERS START PROGRAM A MEMORY Turn the yellow rotary knob to scroll to M1 to edit Memory 1. (M2 = Memory 2, etc.) Home screen. When M1 is selected. Press the rotary knob to recall Memory 1. EDIT TEST PARAMETERS When the parameter is selected, Press the rotary knob to edit the parameter. The

Hook the black return test lead (P/N 99-10724-01) to the black RETURN terminal on the

front panel of the tester. You may need to unscrew the terminal slightly to make room for the hook. Clip the other end of the black return test lead to the chassis of the DUT. Hook one end of the red high current test lead (P/N 99-10725-01) to the red CURRENT terminal on the front panel of the tester. Clip the other end of the red high current test lead to the ground pin

Connect the Interlock Disable Key (P/N 99-10040-01)

to the Signal Input connector on the rear panel of the tester. If you’re not utilizing a DUT enclosure (P/N 99-10715-01 or P/N 99-10599-01) or other safety device, the Interlock Disable Key is required in order to run a test. With the tester set to the desired parameter and your DUT correctly connected to the tester, you are now ready to start testing. Push the green TEST button on the front

SAFETY MADE SIMPLE®

DO NOT set up electrical safety testing stations and ESD (electrostatic discharge) stations in the same area. ESD protocols are designed to protect a component or device from static discharge (not the operator from high voltage hazards). DO NOT use anti-static robes, benches or floor mats during electrical safety testing. All of these items are used to intentionally ground the tester operator which can cause injury or death to a high

SCI Test Equipment Owners Instructions: